Professional Documents
Culture Documents
Check List Preventive Maintenance
Check List Preventive Maintenance
SAFETY CHECK
1 Cek Tahanan Ground **)
2 Cek Kebocoran Arus Ground Pada Kondisi Normal **)
MAINTENANCE
1 Periksa Tegangan Listrik
2 Periksa Kabel Kelistrikan
3. Periksa Dan Ukur Heather Pemanas
4 Periksa Fungsi Parameter Pada Autoclave
5 Periksa Tutup Autoclave
6 Periksa Pembuangan Air
7 Periksa Selang Pembuangan Air, Ganti Bila Perlu
8 Periksa Kebersihan Alat
9 Uji Kinerja Alat
No Measure
Measure step Norm unity 1 2
MONITOR
1 Mains Voltage 22010% Volts
2 Power -----* VA
3 Ground Leakage Current ≤ 500 µA
3.1 Single fault condition (Open feed line) ≤ 1000 µA
4 Insulated Resistance between line conductors to ground 2 M
5 Ground Resistance ≤ 0.2
OBSERVATIONS
1.
2.
RECOMMENDATION 1 2
The equipment has no problems. Result of safety test showed no deviation from the required
reference values
The equipment has minor problems that do restrict its use. However they should be corrected as the
preventive measure. Result of safety test showed no deviation from the required reference values
The equipment has major problems. For safety reasons, the equipment may not be used until the
problem has been resolved.