HEM-6221-E.book Page 1 Tuesday, November 6, 2012 2:06 PM HEM-6221-E(A)M05_120926.

pdf

Cuff Wrapping Guide


1. Overview 3. Using the Unit The Cuff Wrapping Guide is a unique feature that
Main unit: 3.1 Applying the Wrist Cuff indicates if the cuff is not wrapped tightly enough
Do not apply over clothing. around the wrist. Even when the is displayed, a
A
1. Put your arm through the blood pressure reading will be taken.
D cuff loop. Note: This reading is NOT reliable due to the incorrect
Wrist Blood
wrapping of the cuff. Please wrap the cuff again, taking
Pressure Monitor Your palm should face upward. care to wrap it correctly and take the measurement
E again. When the is displayed, the cuff is correctly
Model RS6 wrapped tightly enough on the wrist and the reading is
Instruction Manual accurate and reliable.
IM-HEM-6221-E-01-08/2012 F
1883267-9A
B 2. Wrap the wrist cuff around 2. Undo the wrist cuff and remove the unit.
C
Introduction your wrist.
3. Press the START/STOP button to turn off the monitor.
Thank you for purchasing the OMRON RS6 Wrist Blood Pressure A. Wrist cuff D. Memory ( ) button 1 The monitor automatically stores the measurement in its memory.
Wrap the wrist cuff securely
Monitor. It will automatically turn off after two minutes.
B. Battery compartment E. Date/Time setting ( ) button around the wrist for taking
accurate measurements.
C. Display F. START/STOP button
The OMRON RS6 is a compact and easy to use blood pressure Important:
monitor, operating on the oscillometric principle. It measures your • Recent research suggests that the
blood pressure and pulse rate simply and quickly. For comfortable Display: following values can be used as a
controlled inflation without the need of pressure pre-setting or guide to high blood pressure for
P
re-inflation the device uses its advanced “IntelliSense” technology. Q measurements taken at home.
G R 2
Intended Use
This product is designed to measure the blood pressure and pulse H S
Systolic Blood Pressure Above 135 mmHg
rate of people within the range of the designated wrist cuff, following
I Diastolic Blood Pressure Above 85 mmHg
the instructions in this instruction manual.
It is mainly designed for general household use. Please read the This criteria is for home blood pressure measurement.
J T
Important Safety Information in this instruction manual before using Make sure that the wrist cuff
the unit. K does not cover the protruding • Your blood pressure monitor includes an
U part of the wrist bone (ulna) on irregular heartbeat feature. Irregular
L M N O the outside of the wrist. heartbeats can influence the results of the
Please read this instruction manual thoroughly before using the measurement. The irregular heartbeat
unit. G. Memory symbol N. Movement error symbol algorithm automatically determines if the
Please keep for future reference. H. Systolic blood pressure O. Deflation symbol measurement is usable or needs to be
For specific information about your own blood pressure, Notes:
I. Diastolic blood pressure P. Average value symbol repeated. If the measurement results are
CONSULT YOUR DOCTOR. • You can take a measurement on
J. Battery low symbol Q. Date/Time display affected by irregular heartbeats but the result
either your left or right wrist.
K. Heartbeat symbol R. Blood pressure level indicator • If the wrist cuff comes off, is valid, the result is shown together with the
assemble as shown in the figure. irregular heartbeat symbol ( ).
Important Safety Information (Flashes during measurement) S. Positioning indicator
If the irregular heartbeats cause the measurement to be
Consult your doctor prior to using in pregnancy or if diagnosed with L. Pulse display and Memory T. Positioning symbol
number invalid, no result is shown. If the irregular heartbeat symbol
arrhythmia or arteriosclerosis. U. Cuff wrapping guide
M. Irregular heartbeat symbol ( ) is shown after you have taken a measurement, repeat
Please read this section carefully before using the unit.
the measurement. If the irregular heartbeat symbol ( ) is
Warning: shown frequently, please make your doctor aware of it.
• Indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury. 2. Preparation 3.2 How to Sit Correctly • If you move during measurement, the
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results 2.1 Installing/Replacing the Batteries movement error symbol ( ) will appear on the
To take a measurement, you need display. Keep still and repeat the measurement.
and self-treatment are dangerous. 1. Remove the battery cover. to be relaxed and comfortably
• People with severe blood flow problems, or blood disorders, should Notes:
seated, under comfortable room
consult a doctor before using the unit, as cuff inflation can cause • Wait 2 - 3 minutes before taking another
internal bleeding.
temperature. No bathing, drinking
alcohol or caffeine, smoking, blood pressure measurement. Waiting
(Battery Usage) between readings allows the arteries to
exercising or eating 30 minutes
• If battery fluid should get in your eyes, immediately rinse with return to the condition prior to taking the
plenty of clean water. Consult a doctor immediately.
before taking a measurement.
blood pressure measurement.
Caution:
2. Insert two 1.5V “AAA” • The blood pressure can differ between the right arm and the
alkaline (LR03) batteries • Sit on a chair with your feet flat on the
left arm, and therefore also the measured blood pressure
• Indicates a potentially hazardous situation which, if not avoided, as indicated in the battery floor.
may result in minor or moderate injury to the user or patient or values can be different. OMRON recommends to always use
compartment and then • Sit upright with your back straight.
damage to the equipment or other property. the same arm for measurement. If the values between the
• The cuff should be at the same level
replace the battery cover. two arms differ substantially, please check with your
(General Usage) as your heart.
physician which arm to use for your measurement.
• Do not leave the unit unattended with infants or persons who Notes:
cannot express their consent. • Relax your wrist and hand. Do not bend your wrist back, clench
• If the battery low symbol ( ) appears on the display, turn off
• Do not use the unit for any purpose other than measuring blood
your fist, or bend your wrist forward. Warning:
the unit then replace both batteries at the same time.
Self-diagnosis of measured results and treatment are
pressure. • The measurement values continue to be stored in memory
dangerous. Please follow the instructions of your doctor.
• Do not use a mobile phone or other devices that emit even after the batteries are replaced.
electromagnetic fields, near the unit. This may result in incorrect
operation of the unit. Disposal of used batteries should be carried out in accordance 3.4 Setting the Positioning Sensor
with the national regulations for the disposal of batteries.
• Do not disassemble the unit or wrist cuff. 1. Press and hold the
• Do not operate the unit in a moving vehicle (car, airplane).
3.3 Taking a Reading button to reset the
(Battery Usage) 2.2 Setting the Date and Time Notes: Positioning sensor.
• If battery fluid should get on your skin or clothing, immediately rinse
with plenty of clean water.
1. Press the button. • To cancel a measurement, press the START/STOP button at The Positioning sensor
any time during measurement.
• Use only two “AAA” alkaline (LR03) batteries with this unit. Do not is set "on" as default.
use other types of batteries.
2. Set the monitor to the correct date and time • Remain still while taking a measurement.

• Do not insert the batteries with their polarities incorrectly aligned.


before taking a measurement for the first time.
1. Press the START/STOP button. 2. Press the START/STOP button to store the setting.
• Replace old batteries with new ones immediately. Replace both 1) Push the button to change. All the symbols appear on the display.
batteries at the same time. • Hold down to advance the digit While your position is improper, the Positioning symbol and the
• Remove the batteries if the unit will not be used for three months or rapidly. Positioning indicator flash, when you are in proper position the
more. Positioning symbol lights on the display.
• Do not use new and used batteries together. 2) Push the button to confirm.
The monitor is in The monitor is higher The monitor is lower
General Precautions • Next setting appears. proper position. than the heart. than the heart.
• Do not apply strong shocks and vibrations to or drop the unit.
• Do not take measurements after bathing, drinking alcohol or
caffeine, smoking, exercising or eating.
• Do not inflate the wrist cuff when it is not wrapped around your Year
wrist.
Month
• Read and follow the “Important information regarding Electro
Magnetic Compatibility (EMC)” in the Technical Data Section. Day
• Read and follow the “Correct Disposal of This Product” in the
Technical Data Section when disposing of the device and any used Hour
accessories or optional parts. Minute

3. Press the START/STOP button to store the setting. After the Positioning symbol lights for more than 2 seconds,
Notes: the wrist cuff will start to inflate automatically.
• If the batteries have been removed, the date and time setting Alternating date/time display
will need to be reset. Cuff Wrapping Guide
• If the date and time are not set, “-:--” appears during or after
measurement.

START INFLATE DEFLATE COMPLETE

12K2520
HEM-6221-E.book Page 1 Tuesday, November 6, 2012 2:06 PM HEM-6221-E(A)M05_120926.pdf

Cuff Wrapping Guide


1. Overview 3. Using the Unit The Cuff Wrapping Guide is a unique feature that
Main unit: 3.1 Applying the Wrist Cuff indicates if the cuff is not wrapped tightly enough
Do not apply over clothing. around the wrist. Even when the is displayed, a
A
1. Put your arm through the blood pressure reading will be taken.
D cuff loop. Note: This reading is NOT reliable due to the incorrect
Wrist Blood
wrapping of the cuff. Please wrap the cuff again, taking
Pressure Monitor Your palm should face upward. care to wrap it correctly and take the measurement
E again. When the is displayed, the cuff is correctly
Model RS6 wrapped tightly enough on the wrist and the reading is
Instruction Manual accurate and reliable.
IM-HEM-6221-E-01-08/2012 F
1883267-9A
B 2. Wrap the wrist cuff around 2. Undo the wrist cuff and remove the unit.
C
Introduction your wrist.
3. Press the START/STOP button to turn off the monitor.
Thank you for purchasing the OMRON RS6 Wrist Blood Pressure A. Wrist cuff D. Memory ( ) button 1 The monitor automatically stores the measurement in its memory.
Wrap the wrist cuff securely
Monitor. It will automatically turn off after two minutes.
B. Battery compartment E. Date/Time setting ( ) button around the wrist for taking
accurate measurements.
C. Display F. START/STOP button
The OMRON RS6 is a compact and easy to use blood pressure Important:
monitor, operating on the oscillometric principle. It measures your • Recent research suggests that the
blood pressure and pulse rate simply and quickly. For comfortable Display: following values can be used as a
controlled inflation without the need of pressure pre-setting or guide to high blood pressure for
P
re-inflation the device uses its advanced “IntelliSense” technology. Q measurements taken at home.
G R 2
Intended Use
This product is designed to measure the blood pressure and pulse H S
Systolic Blood Pressure Above 135 mmHg
rate of people within the range of the designated wrist cuff, following
I Diastolic Blood Pressure Above 85 mmHg
the instructions in this instruction manual.
It is mainly designed for general household use. Please read the This criteria is for home blood pressure measurement.
J T
Important Safety Information in this instruction manual before using Make sure that the wrist cuff
the unit. K does not cover the protruding • Your blood pressure monitor includes an
U part of the wrist bone (ulna) on irregular heartbeat feature. Irregular
L M N O the outside of the wrist. heartbeats can influence the results of the
Please read this instruction manual thoroughly before using the measurement. The irregular heartbeat
unit. G. Memory symbol N. Movement error symbol algorithm automatically determines if the
Please keep for future reference. H. Systolic blood pressure O. Deflation symbol measurement is usable or needs to be
For specific information about your own blood pressure, Notes:
I. Diastolic blood pressure P. Average value symbol repeated. If the measurement results are
CONSULT YOUR DOCTOR. • You can take a measurement on
J. Battery low symbol Q. Date/Time display affected by irregular heartbeats but the result
either your left or right wrist.
K. Heartbeat symbol R. Blood pressure level indicator • If the wrist cuff comes off, is valid, the result is shown together with the
assemble as shown in the figure. irregular heartbeat symbol ( ).
Important Safety Information (Flashes during measurement) S. Positioning indicator
If the irregular heartbeats cause the measurement to be
Consult your doctor prior to using in pregnancy or if diagnosed with L. Pulse display and Memory T. Positioning symbol
number invalid, no result is shown. If the irregular heartbeat symbol
arrhythmia or arteriosclerosis. U. Cuff wrapping guide
M. Irregular heartbeat symbol ( ) is shown after you have taken a measurement, repeat
Please read this section carefully before using the unit.
the measurement. If the irregular heartbeat symbol ( ) is
Warning: shown frequently, please make your doctor aware of it.
• Indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury. 2. Preparation 3.2 How to Sit Correctly • If you move during measurement, the
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results 2.1 Installing/Replacing the Batteries movement error symbol ( ) will appear on the
To take a measurement, you need display. Keep still and repeat the measurement.
and self-treatment are dangerous. 1. Remove the battery cover. to be relaxed and comfortably
• People with severe blood flow problems, or blood disorders, should Notes:
seated, under comfortable room
consult a doctor before using the unit, as cuff inflation can cause • Wait 2 - 3 minutes before taking another
internal bleeding.
temperature. No bathing, drinking
alcohol or caffeine, smoking, blood pressure measurement. Waiting
(Battery Usage) between readings allows the arteries to
exercising or eating 30 minutes
• If battery fluid should get in your eyes, immediately rinse with return to the condition prior to taking the
plenty of clean water. Consult a doctor immediately.
before taking a measurement.
blood pressure measurement.
Caution:
2. Insert two 1.5V “AAA” • The blood pressure can differ between the right arm and the
alkaline (LR03) batteries • Sit on a chair with your feet flat on the
left arm, and therefore also the measured blood pressure
• Indicates a potentially hazardous situation which, if not avoided, as indicated in the battery floor.
may result in minor or moderate injury to the user or patient or values can be different. OMRON recommends to always use
compartment and then • Sit upright with your back straight.
damage to the equipment or other property. the same arm for measurement. If the values between the
• The cuff should be at the same level
replace the battery cover. two arms differ substantially, please check with your
(General Usage) as your heart.
physician which arm to use for your measurement.
• Do not leave the unit unattended with infants or persons who Notes:
cannot express their consent. • Relax your wrist and hand. Do not bend your wrist back, clench
• If the battery low symbol ( ) appears on the display, turn off
• Do not use the unit for any purpose other than measuring blood
your fist, or bend your wrist forward. Warning:
the unit then replace both batteries at the same time.
Self-diagnosis of measured results and treatment are
pressure. • The measurement values continue to be stored in memory
dangerous. Please follow the instructions of your doctor.
• Do not use a mobile phone or other devices that emit even after the batteries are replaced.
electromagnetic fields, near the unit. This may result in incorrect
operation of the unit. Disposal of used batteries should be carried out in accordance 3.4 Setting the Positioning Sensor
with the national regulations for the disposal of batteries.
• Do not disassemble the unit or wrist cuff. 1. Press and hold the
• Do not operate the unit in a moving vehicle (car, airplane).
3.3 Taking a Reading button to reset the
(Battery Usage) 2.2 Setting the Date and Time Notes: Positioning sensor.
• If battery fluid should get on your skin or clothing, immediately rinse
with plenty of clean water.
1. Press the button. • To cancel a measurement, press the START/STOP button at The Positioning sensor
any time during measurement.
• Use only two “AAA” alkaline (LR03) batteries with this unit. Do not is set "on" as default.
use other types of batteries.
2. Set the monitor to the correct date and time • Remain still while taking a measurement.

• Do not insert the batteries with their polarities incorrectly aligned.


before taking a measurement for the first time.
1. Press the START/STOP button. 2. Press the START/STOP button to store the setting.
• Replace old batteries with new ones immediately. Replace both 1) Push the button to change. All the symbols appear on the display.
batteries at the same time. • Hold down to advance the digit While your position is improper, the Positioning symbol and the
• Remove the batteries if the unit will not be used for three months or rapidly. Positioning indicator flash, when you are in proper position the
more. Positioning symbol lights on the display.
• Do not use new and used batteries together. 2) Push the button to confirm.
The monitor is in The monitor is higher The monitor is lower
General Precautions • Next setting appears. proper position. than the heart. than the heart.
• Do not apply strong shocks and vibrations to or drop the unit.
• Do not take measurements after bathing, drinking alcohol or
caffeine, smoking, exercising or eating.
• Do not inflate the wrist cuff when it is not wrapped around your Year
wrist.
Month
• Read and follow the “Important information regarding Electro
Magnetic Compatibility (EMC)” in the Technical Data Section. Day
• Read and follow the “Correct Disposal of This Product” in the
Technical Data Section when disposing of the device and any used Hour
accessories or optional parts. Minute

3. Press the START/STOP button to store the setting. After the Positioning symbol lights for more than 2 seconds,
Notes: the wrist cuff will start to inflate automatically.
• If the batteries have been removed, the date and time setting Alternating date/time display
will need to be reset. Cuff Wrapping Guide
• If the date and time are not set, “-:--” appears during or after
measurement.

START INFLATE DEFLATE COMPLETE

12K2520
HEM-6221-E.book Page 1 Tuesday, November 6, 2012 2:06 PM HEM-6221-E(A)M05_120926.pdf

Cuff Wrapping Guide


1. Overview 3. Using the Unit The Cuff Wrapping Guide is a unique feature that
Main unit: 3.1 Applying the Wrist Cuff indicates if the cuff is not wrapped tightly enough
Do not apply over clothing. around the wrist. Even when the is displayed, a
A
1. Put your arm through the blood pressure reading will be taken.
D cuff loop. Note: This reading is NOT reliable due to the incorrect
Wrist Blood
wrapping of the cuff. Please wrap the cuff again, taking
Pressure Monitor Your palm should face upward. care to wrap it correctly and take the measurement
E again. When the is displayed, the cuff is correctly
Model RS6 wrapped tightly enough on the wrist and the reading is
Instruction Manual accurate and reliable.
IM-HEM-6221-E-01-08/2012 F
1883267-9A
B 2. Wrap the wrist cuff around 2. Undo the wrist cuff and remove the unit.
C
Introduction your wrist.
3. Press the START/STOP button to turn off the monitor.
Thank you for purchasing the OMRON RS6 Wrist Blood Pressure A. Wrist cuff D. Memory ( ) button 1 The monitor automatically stores the measurement in its memory.
Wrap the wrist cuff securely
Monitor. It will automatically turn off after two minutes.
B. Battery compartment E. Date/Time setting ( ) button around the wrist for taking
accurate measurements.
C. Display F. START/STOP button
The OMRON RS6 is a compact and easy to use blood pressure Important:
monitor, operating on the oscillometric principle. It measures your • Recent research suggests that the
blood pressure and pulse rate simply and quickly. For comfortable Display: following values can be used as a
controlled inflation without the need of pressure pre-setting or guide to high blood pressure for
P
re-inflation the device uses its advanced “IntelliSense” technology. Q measurements taken at home.
G R 2
Intended Use
This product is designed to measure the blood pressure and pulse H S
Systolic Blood Pressure Above 135 mmHg
rate of people within the range of the designated wrist cuff, following
I Diastolic Blood Pressure Above 85 mmHg
the instructions in this instruction manual.
It is mainly designed for general household use. Please read the This criteria is for home blood pressure measurement.
J T
Important Safety Information in this instruction manual before using Make sure that the wrist cuff
the unit. K does not cover the protruding • Your blood pressure monitor includes an
U part of the wrist bone (ulna) on irregular heartbeat feature. Irregular
L M N O the outside of the wrist. heartbeats can influence the results of the
Please read this instruction manual thoroughly before using the measurement. The irregular heartbeat
unit. G. Memory symbol N. Movement error symbol algorithm automatically determines if the
Please keep for future reference. H. Systolic blood pressure O. Deflation symbol measurement is usable or needs to be
For specific information about your own blood pressure, Notes:
I. Diastolic blood pressure P. Average value symbol repeated. If the measurement results are
CONSULT YOUR DOCTOR. • You can take a measurement on
J. Battery low symbol Q. Date/Time display affected by irregular heartbeats but the result
either your left or right wrist.
K. Heartbeat symbol R. Blood pressure level indicator • If the wrist cuff comes off, is valid, the result is shown together with the
assemble as shown in the figure. irregular heartbeat symbol ( ).
Important Safety Information (Flashes during measurement) S. Positioning indicator
If the irregular heartbeats cause the measurement to be
Consult your doctor prior to using in pregnancy or if diagnosed with L. Pulse display and Memory T. Positioning symbol
number invalid, no result is shown. If the irregular heartbeat symbol
arrhythmia or arteriosclerosis. U. Cuff wrapping guide
M. Irregular heartbeat symbol ( ) is shown after you have taken a measurement, repeat
Please read this section carefully before using the unit.
the measurement. If the irregular heartbeat symbol ( ) is
Warning: shown frequently, please make your doctor aware of it.
• Indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury. 2. Preparation 3.2 How to Sit Correctly • If you move during measurement, the
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results 2.1 Installing/Replacing the Batteries movement error symbol ( ) will appear on the
To take a measurement, you need display. Keep still and repeat the measurement.
and self-treatment are dangerous. 1. Remove the battery cover. to be relaxed and comfortably
• People with severe blood flow problems, or blood disorders, should Notes:
seated, under comfortable room
consult a doctor before using the unit, as cuff inflation can cause • Wait 2 - 3 minutes before taking another
internal bleeding.
temperature. No bathing, drinking
alcohol or caffeine, smoking, blood pressure measurement. Waiting
(Battery Usage) between readings allows the arteries to
exercising or eating 30 minutes
• If battery fluid should get in your eyes, immediately rinse with return to the condition prior to taking the
plenty of clean water. Consult a doctor immediately.
before taking a measurement.
blood pressure measurement.
Caution:
2. Insert two 1.5V “AAA” • The blood pressure can differ between the right arm and the
alkaline (LR03) batteries • Sit on a chair with your feet flat on the
left arm, and therefore also the measured blood pressure
• Indicates a potentially hazardous situation which, if not avoided, as indicated in the battery floor.
may result in minor or moderate injury to the user or patient or values can be different. OMRON recommends to always use
compartment and then • Sit upright with your back straight.
damage to the equipment or other property. the same arm for measurement. If the values between the
• The cuff should be at the same level
replace the battery cover. two arms differ substantially, please check with your
(General Usage) as your heart.
physician which arm to use for your measurement.
• Do not leave the unit unattended with infants or persons who Notes:
cannot express their consent. • Relax your wrist and hand. Do not bend your wrist back, clench
• If the battery low symbol ( ) appears on the display, turn off
• Do not use the unit for any purpose other than measuring blood
your fist, or bend your wrist forward. Warning:
the unit then replace both batteries at the same time.
Self-diagnosis of measured results and treatment are
pressure. • The measurement values continue to be stored in memory
dangerous. Please follow the instructions of your doctor.
• Do not use a mobile phone or other devices that emit even after the batteries are replaced.
electromagnetic fields, near the unit. This may result in incorrect
operation of the unit. Disposal of used batteries should be carried out in accordance 3.4 Setting the Positioning Sensor
with the national regulations for the disposal of batteries.
• Do not disassemble the unit or wrist cuff. 1. Press and hold the
• Do not operate the unit in a moving vehicle (car, airplane).
3.3 Taking a Reading button to reset the
(Battery Usage) 2.2 Setting the Date and Time Notes: Positioning sensor.
• If battery fluid should get on your skin or clothing, immediately rinse
with plenty of clean water.
1. Press the button. • To cancel a measurement, press the START/STOP button at The Positioning sensor
any time during measurement.
• Use only two “AAA” alkaline (LR03) batteries with this unit. Do not is set "on" as default.
use other types of batteries.
2. Set the monitor to the correct date and time • Remain still while taking a measurement.

• Do not insert the batteries with their polarities incorrectly aligned.


before taking a measurement for the first time.
1. Press the START/STOP button. 2. Press the START/STOP button to store the setting.
• Replace old batteries with new ones immediately. Replace both 1) Push the button to change. All the symbols appear on the display.
batteries at the same time. • Hold down to advance the digit While your position is improper, the Positioning symbol and the
• Remove the batteries if the unit will not be used for three months or rapidly. Positioning indicator flash, when you are in proper position the
more. Positioning symbol lights on the display.
• Do not use new and used batteries together. 2) Push the button to confirm.
The monitor is in The monitor is higher The monitor is lower
General Precautions • Next setting appears. proper position. than the heart. than the heart.
• Do not apply strong shocks and vibrations to or drop the unit.
• Do not take measurements after bathing, drinking alcohol or
caffeine, smoking, exercising or eating.
• Do not inflate the wrist cuff when it is not wrapped around your Year
wrist.
Month
• Read and follow the “Important information regarding Electro
Magnetic Compatibility (EMC)” in the Technical Data Section. Day
• Read and follow the “Correct Disposal of This Product” in the
Technical Data Section when disposing of the device and any used Hour
accessories or optional parts. Minute

3. Press the START/STOP button to store the setting. After the Positioning symbol lights for more than 2 seconds,
Notes: the wrist cuff will start to inflate automatically.
• If the batteries have been removed, the date and time setting Alternating date/time display
will need to be reset. Cuff Wrapping Guide
• If the date and time are not set, “-:--” appears during or after
measurement.

START INFLATE DEFLATE COMPLETE

12K2520
HEM-6221-E.book Page 1 Tuesday, November 6, 2012 2:06 PM HEM-6221-E(A)M05_120926.pdf

Cuff Wrapping Guide


1. Overview 3. Using the Unit The Cuff Wrapping Guide is a unique feature that
Main unit: 3.1 Applying the Wrist Cuff indicates if the cuff is not wrapped tightly enough
Do not apply over clothing. around the wrist. Even when the is displayed, a
A
1. Put your arm through the blood pressure reading will be taken.
D cuff loop. Note: This reading is NOT reliable due to the incorrect
Wrist Blood
wrapping of the cuff. Please wrap the cuff again, taking
Pressure Monitor Your palm should face upward. care to wrap it correctly and take the measurement
E again. When the is displayed, the cuff is correctly
Model RS6 wrapped tightly enough on the wrist and the reading is
Instruction Manual accurate and reliable.
IM-HEM-6221-E-01-08/2012 F
1883267-9A
B 2. Wrap the wrist cuff around 2. Undo the wrist cuff and remove the unit.
C
Introduction your wrist.
3. Press the START/STOP button to turn off the monitor.
Thank you for purchasing the OMRON RS6 Wrist Blood Pressure A. Wrist cuff D. Memory ( ) button 1 The monitor automatically stores the measurement in its memory.
Wrap the wrist cuff securely
Monitor. It will automatically turn off after two minutes.
B. Battery compartment E. Date/Time setting ( ) button around the wrist for taking
accurate measurements.
C. Display F. START/STOP button
The OMRON RS6 is a compact and easy to use blood pressure Important:
monitor, operating on the oscillometric principle. It measures your • Recent research suggests that the
blood pressure and pulse rate simply and quickly. For comfortable Display: following values can be used as a
controlled inflation without the need of pressure pre-setting or guide to high blood pressure for
P
re-inflation the device uses its advanced “IntelliSense” technology. Q measurements taken at home.
G R 2
Intended Use
This product is designed to measure the blood pressure and pulse H S
Systolic Blood Pressure Above 135 mmHg
rate of people within the range of the designated wrist cuff, following
I Diastolic Blood Pressure Above 85 mmHg
the instructions in this instruction manual.
It is mainly designed for general household use. Please read the This criteria is for home blood pressure measurement.
J T
Important Safety Information in this instruction manual before using Make sure that the wrist cuff
the unit. K does not cover the protruding • Your blood pressure monitor includes an
U part of the wrist bone (ulna) on irregular heartbeat feature. Irregular
L M N O the outside of the wrist. heartbeats can influence the results of the
Please read this instruction manual thoroughly before using the measurement. The irregular heartbeat
unit. G. Memory symbol N. Movement error symbol algorithm automatically determines if the
Please keep for future reference. H. Systolic blood pressure O. Deflation symbol measurement is usable or needs to be
For specific information about your own blood pressure, Notes:
I. Diastolic blood pressure P. Average value symbol repeated. If the measurement results are
CONSULT YOUR DOCTOR. • You can take a measurement on
J. Battery low symbol Q. Date/Time display affected by irregular heartbeats but the result
either your left or right wrist.
K. Heartbeat symbol R. Blood pressure level indicator • If the wrist cuff comes off, is valid, the result is shown together with the
assemble as shown in the figure. irregular heartbeat symbol ( ).
Important Safety Information (Flashes during measurement) S. Positioning indicator
If the irregular heartbeats cause the measurement to be
Consult your doctor prior to using in pregnancy or if diagnosed with L. Pulse display and Memory T. Positioning symbol
number invalid, no result is shown. If the irregular heartbeat symbol
arrhythmia or arteriosclerosis. U. Cuff wrapping guide
M. Irregular heartbeat symbol ( ) is shown after you have taken a measurement, repeat
Please read this section carefully before using the unit.
the measurement. If the irregular heartbeat symbol ( ) is
Warning: shown frequently, please make your doctor aware of it.
• Indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury. 2. Preparation 3.2 How to Sit Correctly • If you move during measurement, the
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results 2.1 Installing/Replacing the Batteries movement error symbol ( ) will appear on the
To take a measurement, you need display. Keep still and repeat the measurement.
and self-treatment are dangerous. 1. Remove the battery cover. to be relaxed and comfortably
• People with severe blood flow problems, or blood disorders, should Notes:
seated, under comfortable room
consult a doctor before using the unit, as cuff inflation can cause • Wait 2 - 3 minutes before taking another
internal bleeding.
temperature. No bathing, drinking
alcohol or caffeine, smoking, blood pressure measurement. Waiting
(Battery Usage) between readings allows the arteries to
exercising or eating 30 minutes
• If battery fluid should get in your eyes, immediately rinse with return to the condition prior to taking the
plenty of clean water. Consult a doctor immediately.
before taking a measurement.
blood pressure measurement.
Caution:
2. Insert two 1.5V “AAA” • The blood pressure can differ between the right arm and the
alkaline (LR03) batteries • Sit on a chair with your feet flat on the
left arm, and therefore also the measured blood pressure
• Indicates a potentially hazardous situation which, if not avoided, as indicated in the battery floor.
may result in minor or moderate injury to the user or patient or values can be different. OMRON recommends to always use
compartment and then • Sit upright with your back straight.
damage to the equipment or other property. the same arm for measurement. If the values between the
• The cuff should be at the same level
replace the battery cover. two arms differ substantially, please check with your
(General Usage) as your heart.
physician which arm to use for your measurement.
• Do not leave the unit unattended with infants or persons who Notes:
cannot express their consent. • Relax your wrist and hand. Do not bend your wrist back, clench
• If the battery low symbol ( ) appears on the display, turn off
• Do not use the unit for any purpose other than measuring blood
your fist, or bend your wrist forward. Warning:
the unit then replace both batteries at the same time.
Self-diagnosis of measured results and treatment are
pressure. • The measurement values continue to be stored in memory
dangerous. Please follow the instructions of your doctor.
• Do not use a mobile phone or other devices that emit even after the batteries are replaced.
electromagnetic fields, near the unit. This may result in incorrect
operation of the unit. Disposal of used batteries should be carried out in accordance 3.4 Setting the Positioning Sensor
with the national regulations for the disposal of batteries.
• Do not disassemble the unit or wrist cuff. 1. Press and hold the
• Do not operate the unit in a moving vehicle (car, airplane).
3.3 Taking a Reading button to reset the
(Battery Usage) 2.2 Setting the Date and Time Notes: Positioning sensor.
• If battery fluid should get on your skin or clothing, immediately rinse
with plenty of clean water.
1. Press the button. • To cancel a measurement, press the START/STOP button at The Positioning sensor
any time during measurement.
• Use only two “AAA” alkaline (LR03) batteries with this unit. Do not is set "on" as default.
use other types of batteries.
2. Set the monitor to the correct date and time • Remain still while taking a measurement.

• Do not insert the batteries with their polarities incorrectly aligned.


before taking a measurement for the first time.
1. Press the START/STOP button. 2. Press the START/STOP button to store the setting.
• Replace old batteries with new ones immediately. Replace both 1) Push the button to change. All the symbols appear on the display.
batteries at the same time. • Hold down to advance the digit While your position is improper, the Positioning symbol and the
• Remove the batteries if the unit will not be used for three months or rapidly. Positioning indicator flash, when you are in proper position the
more. Positioning symbol lights on the display.
• Do not use new and used batteries together. 2) Push the button to confirm.
The monitor is in The monitor is higher The monitor is lower
General Precautions • Next setting appears. proper position. than the heart. than the heart.
• Do not apply strong shocks and vibrations to or drop the unit.
• Do not take measurements after bathing, drinking alcohol or
caffeine, smoking, exercising or eating.
• Do not inflate the wrist cuff when it is not wrapped around your Year
wrist.
Month
• Read and follow the “Important information regarding Electro
Magnetic Compatibility (EMC)” in the Technical Data Section. Day
• Read and follow the “Correct Disposal of This Product” in the
Technical Data Section when disposing of the device and any used Hour
accessories or optional parts. Minute

3. Press the START/STOP button to store the setting. After the Positioning symbol lights for more than 2 seconds,
Notes: the wrist cuff will start to inflate automatically.
• If the batteries have been removed, the date and time setting Alternating date/time display
will need to be reset. Cuff Wrapping Guide
• If the date and time are not set, “-:--” appears during or after
measurement.

START INFLATE DEFLATE COMPLETE

12K2520
HEM-6221-E.book Page 2 Tuesday, November 6, 2012 2:06 PM

3.5 Using the Memory Function 4.2 Troubleshooting Do not store the unit in the following situations:
Important information regarding Electro Magnetic
The monitor automatically stores the result up to 90 sets.
Problem Cause Remedy • If the unit is wet. Compatibility (EMC)
It can also calculate an average reading based on the
measurements from the last three readings taken within 10 • Locations exposed to extreme temperatures, With the increased number of electronic devices such as PC’s and
minutes. If there are only two readings in memory for that Measure while in the humidity, direct sunlight, dust or corrosive mobile (cellular) telephones, medical devices in use may be
period, the average will be based on two readings. If there is The wrist cuff is not vapours. susceptible to electromagnetic interference from other devices.
correct posture.
one reading in memory for that period, the average will be at heart level. Electromagnetic interference may result in incorrect operation of
Refer to section 3.2. • Locations exposed to vibrations, shocks or
based on one reading. the medical device and create a potentially unsafe situation.
where it will be at an angle. Medical devices should also not interfere with other devices.
Notes:
• If the memory is full, the monitor will delete the oldest The cuff is not In order to regulate the requirements for EMC (Electro Magnetic
readings. wrapped snugly
Wrap the cuff correctly. Calibration and Service Compatibility) with the aim to prevent unsafe product situations, the
• When viewing the reading taken without setting the date and around the wrist.
Refer to section 3.1. • The accuracy of this blood pressure monitor has EN60601-1-2:2007 standard has been implemented. This standard
The reading is been carefully tested and is designed for a long defines the levels of immunity to electromagnetic interferences as
time, “-:--” is displayed instead of the date and time.
extremely low
service life. well as maximum levels of electromagnetic emissions for medical
(or high). devices.
To View the Average Value
The arms and Relax and try taking the • It is generally recommended to have the unit
shoulders are measurement again. inspected every two years to ensure correct This medical device manufactured by OMRON HEALTHCARE
tense. Refer to section 3.3. conforms to this EN60601-1-2:2007 standard for both immunity and
Press the button. functioning and accuracy. Please consult your
emissions.
authorised OMRON dealer or the OMRON Nevertheless, special precautions need to be observed:
Movement or Remain still and do not Customer Service at the address given on the • Do not use mobile (cellular) telephones and other devices, which
talking during talk during Measurement. packaging or attached literature. generate strong electrical or electromagnetic fields, near the
measurement. Refer to section 3.3. medical device. This may result in incorrect operation of the unit
and create a potentially unsafe situation. Recommendation is to
keep a minimum distance of 7 m. Verify correct operation of the
device in case the distance is shorter.
Note: If there are no measurements
results stored in the memory, the
5. Technical Data Further documentation in accordance with EN60601-1-2:2007 is
Wrist cuff pressure Air is leaking from Consult your OMRON
screen to the right is displayed. does not rise. the wrist cuff. retail outlet or distributor. Product Description Wrist Blood Pressure Monitor available at OMRON HEALTHCARE EUROPE at the address
mentioned in this instruction manual.
Model OMRON RS6 (HEM-6221-E) Documentation is also available at www.omron-healthcare.com.
Display LCD Digital Display
Apply the cuff correctly so
Wrist cuff deflates The wrist cuff is that it is firmly wrapped Measurement Method Oscillometric method Correct Disposal of This Product
too soon. loose. around the wrist. Refer to (Waste Electrical & Electronic Equipment)
Measurement Range Pressure: 0 to 299 mmHg
To View the Readings Stored in Memory section 3.1. Pulse: 40 to 180 beats/min. This marking shown on the product or its
literature, indicates that it should not be
1. Press the button, while the average value is Blood pressure readings Accuracy Pressure: ±3 mmHg
disposed of, with other household wastes at the
displayed. constantly vary with time Pulse: ±5% of display reading
end of its working life.
The blood pressure is different of day and how relaxed Inflation Automatic inflation by pump To prevent possible harm to the environment or human health from
The Memory number each time. The reading is you are. Take several uncontrolled waste disposal, please separate this product from
extremely low (or high). deep breaths and try to Deflation Automatic rapid deflation
appears for a second other types of wastes and recycle it responsibly to promote the
before the pulse rate is remain relaxed before Memory 90 Measurements sustainable reuse of material resources.
displayed. The newest taking a measurement.
Alternating Power Source 2 x 1.5V (LR03, AAA alkaline batteries) Household users should contact either the retailer where they
set is numbered “1”.
date/time Battery Life Approx. 300 measurements with new purchased this product, or their local government office, for details
The unit loses
display alkaline batteries at a room temperature of of where and how they can return this item for environmentally safe
power during Replace the batteries with
The batteries are new ones. 23°C recycling.
measurement.
exhausted. Refer to section 2.1. Business users should contact their supplier and check the terms
2. Press the button repeatedly to view the Applied Part = Type B and conditions of the purchase contract. This product should not be
readings stored in memory. mixed with other commercial wastes for disposal.
Nothing happens Protection Against Internally powered ME equipment
To Delete All the Values Stored in Memory when you press The batteries have Insert the batteries with Electric Shock Manufacturer OMRON HEALTHCARE Co., Ltd.
the buttons. been inserted the correct (+/-) polarity. 53, Kunotsubo, Terado-cho, Muko, Kyoto,
Operating temperature/ +10 to +40°C / Maximum: 30 to 85% RH 617-0002 JAPAN
When the memory symbol ( ) appears, first press incorrectly. Refer to section 2.1.
Humidity
the button. Then while holding it down, press the Storage temperature/ -20 to +60°C / Maximum: 10 to 95% RH / EU-representative OMRON HEALTHCARE EUROPE B.V.
START/STOP button simultaneously for about 2 - 3 • Press the START/STOP button and repeat Scorpius 33, 2132 LR Hoofddorp
Humidity/Air pressure 700 to 1060hPa
seconds. measurement. THE NETHERLANDS
• If the problem continues, try replacing the Console Weight Approximately 85 g without batteries www.omron-healthcare.com
Other problems.
batteries with new ones. Outer Dimensions Approximately 87 (w) mm x 64 (h) mm x Production Facility OMRON (DALIAN) CO., LTD.
If this still does not solve the problem, contact 14 (d) mm (without the wrist cuff) Dalian, CHINA
your OMRON retail outlet or distributor. Subsidiary OMRON HEALTHCARE UK LTD.
Measurable Approximately 13.5 to 21.5 cm Opal Drive, Fox Milne, Milton Keynes, MK15 0DG
circumference U.K.
First Second Cuff Material Nylon and polyester OMRON MEDIZINTECHNIK
Note: You cannot partially delete the stored readings.
4.3 Maintenance HANDELSGESELLSCHAFT mbH
To protect your unit from damage, please observe the Package Content Main unit, storage case, battery set, John-Deere-Str. 81a 68163 Mannheim, GERMANY
instruction manual, guarantee card, blood www.omron-medizintechnik.de
following:
pressure pass OMRON SANTÉ FRANCE SAS
14, rue de Lisbonne, 93561 Rosny-sous-Bois
4. Troubleshooting and Maintenance • Do not subject the main unit and the cuff to Cedex, FRANCE
Note: Subject to technical modification without prior notice.
4.1 The Icons and Error Messages extreme temperatures, humidity, moisture or Made in China
direct sunlight.
Error Display Cause Remedy
• Do not disassemble the unit.
Remove the wrist cuff. Wait • Do not subject the unit to strong shocks or • This device fulfils the provisions of EC directive 93/42/EEC
2 - 3 minutes and then take vibrations (for example, dropping the unit on the (Medical Device Directive).
Irregular heartbeats another measurement.
are detected. Repeat the steps in section floor). • This blood pressure monitor is designed according to the European
3.3. If this error continues to • Do not use volatile liquids to clean the main unit. Standard EN1060, Non-invasive sphygmomanometers Part 1:
appear, contact your doctor. General Requirements and Part 3: Supplementary requirements
• Do not wash the cuff or immerse it in water. for electromechanical blood pressure measuring systems.
• Do not use petrol, thinners or similar solvents to • This OMRON product is produced under the strict quality system of
Movement during Carefully read and repeat the clean the cuff. OMRON HEALTHCARE Co. Ltd., Japan. The Core component for
measurement. steps in section 3.3.
• Do not carry out repairs of any kind yourself. If a OMRON blood pressure monitors, which is the Pressure Sensor, is
defect occurs, consult your OMRON retail outlet produced in Japan.
You should replace them with
The batteries are or distributor as mentioned on the packaging.
Blink low. new ones ahead of time.
Refer to section 2.1.

You should replace them with


The batteries are
Lit new ones at once.
exhausted.
Refer to section 2.1.

Wrist cuff not Apply the wrist cuff correctly.


applied correctly. Refer to section 3.1.

Repeat Measurement while


remaining still and refraining
Movement during
from talking during the
measurement.
measurement.
Refer to section 3.3.
• The unit should be cleaned with a soft, dry cloth.
Apply the wrist cuff correctly • Use a soft, moistened cloth and neutral soap to
Wrist cuff not and repeat measurement clean the cuff.
applied correctly, or while remaining still and
movement during refraining from taking during • Keep the unit in its storage case when not in
measurement. the measurement. use.
Refer to section 3.1 and 3.3. • Fold the cuff into the storage case.
Arm position Remain still until the
changed during measurement is complete.
measurement. Refer to section 3.3.

Contact your OMRON retail


Device error.
outlet or distributor.

Note: The irregular heartbeat symbol ( ) and the movement error


symbol ( ) may also be displayed with error messages.
HEM-6221-E.book Page 2 Tuesday, November 6, 2012 2:06 PM

3.5 Using the Memory Function 4.2 Troubleshooting Do not store the unit in the following situations:
Important information regarding Electro Magnetic
The monitor automatically stores the result up to 90 sets.
Problem Cause Remedy • If the unit is wet. Compatibility (EMC)
It can also calculate an average reading based on the
measurements from the last three readings taken within 10 • Locations exposed to extreme temperatures, With the increased number of electronic devices such as PC’s and
minutes. If there are only two readings in memory for that Measure while in the humidity, direct sunlight, dust or corrosive mobile (cellular) telephones, medical devices in use may be
period, the average will be based on two readings. If there is The wrist cuff is not vapours. susceptible to electromagnetic interference from other devices.
correct posture.
one reading in memory for that period, the average will be at heart level. Electromagnetic interference may result in incorrect operation of
Refer to section 3.2. • Locations exposed to vibrations, shocks or
based on one reading. the medical device and create a potentially unsafe situation.
where it will be at an angle. Medical devices should also not interfere with other devices.
Notes:
• If the memory is full, the monitor will delete the oldest The cuff is not In order to regulate the requirements for EMC (Electro Magnetic
readings. wrapped snugly
Wrap the cuff correctly. Calibration and Service Compatibility) with the aim to prevent unsafe product situations, the
• When viewing the reading taken without setting the date and around the wrist.
Refer to section 3.1. • The accuracy of this blood pressure monitor has EN60601-1-2:2007 standard has been implemented. This standard
The reading is been carefully tested and is designed for a long defines the levels of immunity to electromagnetic interferences as
time, “-:--” is displayed instead of the date and time.
extremely low
service life. well as maximum levels of electromagnetic emissions for medical
(or high). devices.
To View the Average Value
The arms and Relax and try taking the • It is generally recommended to have the unit
shoulders are measurement again. inspected every two years to ensure correct This medical device manufactured by OMRON HEALTHCARE
tense. Refer to section 3.3. conforms to this EN60601-1-2:2007 standard for both immunity and
Press the button. functioning and accuracy. Please consult your
emissions.
authorised OMRON dealer or the OMRON Nevertheless, special precautions need to be observed:
Movement or Remain still and do not Customer Service at the address given on the • Do not use mobile (cellular) telephones and other devices, which
talking during talk during Measurement. packaging or attached literature. generate strong electrical or electromagnetic fields, near the
measurement. Refer to section 3.3. medical device. This may result in incorrect operation of the unit
and create a potentially unsafe situation. Recommendation is to
keep a minimum distance of 7 m. Verify correct operation of the
device in case the distance is shorter.
Note: If there are no measurements
results stored in the memory, the
5. Technical Data Further documentation in accordance with EN60601-1-2:2007 is
Wrist cuff pressure Air is leaking from Consult your OMRON
screen to the right is displayed. does not rise. the wrist cuff. retail outlet or distributor. Product Description Wrist Blood Pressure Monitor available at OMRON HEALTHCARE EUROPE at the address
mentioned in this instruction manual.
Model OMRON RS6 (HEM-6221-E) Documentation is also available at www.omron-healthcare.com.
Display LCD Digital Display
Apply the cuff correctly so
Wrist cuff deflates The wrist cuff is that it is firmly wrapped Measurement Method Oscillometric method Correct Disposal of This Product
too soon. loose. around the wrist. Refer to (Waste Electrical & Electronic Equipment)
Measurement Range Pressure: 0 to 299 mmHg
To View the Readings Stored in Memory section 3.1. Pulse: 40 to 180 beats/min. This marking shown on the product or its
literature, indicates that it should not be
1. Press the button, while the average value is Blood pressure readings Accuracy Pressure: ±3 mmHg
disposed of, with other household wastes at the
displayed. constantly vary with time Pulse: ±5% of display reading
end of its working life.
The blood pressure is different of day and how relaxed Inflation Automatic inflation by pump To prevent possible harm to the environment or human health from
The Memory number each time. The reading is you are. Take several uncontrolled waste disposal, please separate this product from
extremely low (or high). deep breaths and try to Deflation Automatic rapid deflation
appears for a second other types of wastes and recycle it responsibly to promote the
before the pulse rate is remain relaxed before Memory 90 Measurements sustainable reuse of material resources.
displayed. The newest taking a measurement.
Alternating Power Source 2 x 1.5V (LR03, AAA alkaline batteries) Household users should contact either the retailer where they
set is numbered “1”.
date/time Battery Life Approx. 300 measurements with new purchased this product, or their local government office, for details
The unit loses
display alkaline batteries at a room temperature of of where and how they can return this item for environmentally safe
power during Replace the batteries with
The batteries are new ones. 23°C recycling.
measurement.
exhausted. Refer to section 2.1. Business users should contact their supplier and check the terms
2. Press the button repeatedly to view the Applied Part = Type B and conditions of the purchase contract. This product should not be
readings stored in memory. mixed with other commercial wastes for disposal.
Nothing happens Protection Against Internally powered ME equipment
To Delete All the Values Stored in Memory when you press The batteries have Insert the batteries with Electric Shock Manufacturer OMRON HEALTHCARE Co., Ltd.
the buttons. been inserted the correct (+/-) polarity. 53, Kunotsubo, Terado-cho, Muko, Kyoto,
Operating temperature/ +10 to +40°C / Maximum: 30 to 85% RH 617-0002 JAPAN
When the memory symbol ( ) appears, first press incorrectly. Refer to section 2.1.
Humidity
the button. Then while holding it down, press the Storage temperature/ -20 to +60°C / Maximum: 10 to 95% RH / EU-representative OMRON HEALTHCARE EUROPE B.V.
START/STOP button simultaneously for about 2 - 3 • Press the START/STOP button and repeat Scorpius 33, 2132 LR Hoofddorp
Humidity/Air pressure 700 to 1060hPa
seconds. measurement. THE NETHERLANDS
• If the problem continues, try replacing the Console Weight Approximately 85 g without batteries www.omron-healthcare.com
Other problems.
batteries with new ones. Outer Dimensions Approximately 87 (w) mm x 64 (h) mm x Production Facility OMRON (DALIAN) CO., LTD.
If this still does not solve the problem, contact 14 (d) mm (without the wrist cuff) Dalian, CHINA
your OMRON retail outlet or distributor. Subsidiary OMRON HEALTHCARE UK LTD.
Measurable Approximately 13.5 to 21.5 cm Opal Drive, Fox Milne, Milton Keynes, MK15 0DG
circumference U.K.
First Second Cuff Material Nylon and polyester OMRON MEDIZINTECHNIK
Note: You cannot partially delete the stored readings.
4.3 Maintenance HANDELSGESELLSCHAFT mbH
To protect your unit from damage, please observe the Package Content Main unit, storage case, battery set, John-Deere-Str. 81a 68163 Mannheim, GERMANY
instruction manual, guarantee card, blood www.omron-medizintechnik.de
following:
pressure pass OMRON SANTÉ FRANCE SAS
14, rue de Lisbonne, 93561 Rosny-sous-Bois
4. Troubleshooting and Maintenance • Do not subject the main unit and the cuff to Cedex, FRANCE
Note: Subject to technical modification without prior notice.
4.1 The Icons and Error Messages extreme temperatures, humidity, moisture or Made in China
direct sunlight.
Error Display Cause Remedy
• Do not disassemble the unit.
Remove the wrist cuff. Wait • Do not subject the unit to strong shocks or • This device fulfils the provisions of EC directive 93/42/EEC
2 - 3 minutes and then take vibrations (for example, dropping the unit on the (Medical Device Directive).
Irregular heartbeats another measurement.
are detected. Repeat the steps in section floor). • This blood pressure monitor is designed according to the European
3.3. If this error continues to • Do not use volatile liquids to clean the main unit. Standard EN1060, Non-invasive sphygmomanometers Part 1:
appear, contact your doctor. General Requirements and Part 3: Supplementary requirements
• Do not wash the cuff or immerse it in water. for electromechanical blood pressure measuring systems.
• Do not use petrol, thinners or similar solvents to • This OMRON product is produced under the strict quality system of
Movement during Carefully read and repeat the clean the cuff. OMRON HEALTHCARE Co. Ltd., Japan. The Core component for
measurement. steps in section 3.3.
• Do not carry out repairs of any kind yourself. If a OMRON blood pressure monitors, which is the Pressure Sensor, is
defect occurs, consult your OMRON retail outlet produced in Japan.
You should replace them with
The batteries are or distributor as mentioned on the packaging.
Blink low. new ones ahead of time.
Refer to section 2.1.

You should replace them with


The batteries are
Lit new ones at once.
exhausted.
Refer to section 2.1.

Wrist cuff not Apply the wrist cuff correctly.


applied correctly. Refer to section 3.1.

Repeat Measurement while


remaining still and refraining
Movement during
from talking during the
measurement.
measurement.
Refer to section 3.3.
• The unit should be cleaned with a soft, dry cloth.
Apply the wrist cuff correctly • Use a soft, moistened cloth and neutral soap to
Wrist cuff not and repeat measurement clean the cuff.
applied correctly, or while remaining still and
movement during refraining from taking during • Keep the unit in its storage case when not in
measurement. the measurement. use.
Refer to section 3.1 and 3.3. • Fold the cuff into the storage case.
Arm position Remain still until the
changed during measurement is complete.
measurement. Refer to section 3.3.

Contact your OMRON retail


Device error.
outlet or distributor.

Note: The irregular heartbeat symbol ( ) and the movement error


symbol ( ) may also be displayed with error messages.
HEM-6221-E.book Page 2 Tuesday, November 6, 2012 2:06 PM

3.5 Using the Memory Function 4.2 Troubleshooting Do not store the unit in the following situations:
Important information regarding Electro Magnetic
The monitor automatically stores the result up to 90 sets.
Problem Cause Remedy • If the unit is wet. Compatibility (EMC)
It can also calculate an average reading based on the
measurements from the last three readings taken within 10 • Locations exposed to extreme temperatures, With the increased number of electronic devices such as PC’s and
minutes. If there are only two readings in memory for that Measure while in the humidity, direct sunlight, dust or corrosive mobile (cellular) telephones, medical devices in use may be
period, the average will be based on two readings. If there is The wrist cuff is not vapours. susceptible to electromagnetic interference from other devices.
correct posture.
one reading in memory for that period, the average will be at heart level. Electromagnetic interference may result in incorrect operation of
Refer to section 3.2. • Locations exposed to vibrations, shocks or
based on one reading. the medical device and create a potentially unsafe situation.
where it will be at an angle. Medical devices should also not interfere with other devices.
Notes:
• If the memory is full, the monitor will delete the oldest The cuff is not In order to regulate the requirements for EMC (Electro Magnetic
readings. wrapped snugly
Wrap the cuff correctly. Calibration and Service Compatibility) with the aim to prevent unsafe product situations, the
• When viewing the reading taken without setting the date and around the wrist.
Refer to section 3.1. • The accuracy of this blood pressure monitor has EN60601-1-2:2007 standard has been implemented. This standard
The reading is been carefully tested and is designed for a long defines the levels of immunity to electromagnetic interferences as
time, “-:--” is displayed instead of the date and time.
extremely low
service life. well as maximum levels of electromagnetic emissions for medical
(or high). devices.
To View the Average Value
The arms and Relax and try taking the • It is generally recommended to have the unit
shoulders are measurement again. inspected every two years to ensure correct This medical device manufactured by OMRON HEALTHCARE
tense. Refer to section 3.3. conforms to this EN60601-1-2:2007 standard for both immunity and
Press the button. functioning and accuracy. Please consult your
emissions.
authorised OMRON dealer or the OMRON Nevertheless, special precautions need to be observed:
Movement or Remain still and do not Customer Service at the address given on the • Do not use mobile (cellular) telephones and other devices, which
talking during talk during Measurement. packaging or attached literature. generate strong electrical or electromagnetic fields, near the
measurement. Refer to section 3.3. medical device. This may result in incorrect operation of the unit
and create a potentially unsafe situation. Recommendation is to
keep a minimum distance of 7 m. Verify correct operation of the
device in case the distance is shorter.
Note: If there are no measurements
results stored in the memory, the
5. Technical Data Further documentation in accordance with EN60601-1-2:2007 is
Wrist cuff pressure Air is leaking from Consult your OMRON
screen to the right is displayed. does not rise. the wrist cuff. retail outlet or distributor. Product Description Wrist Blood Pressure Monitor available at OMRON HEALTHCARE EUROPE at the address
mentioned in this instruction manual.
Model OMRON RS6 (HEM-6221-E) Documentation is also available at www.omron-healthcare.com.
Display LCD Digital Display
Apply the cuff correctly so
Wrist cuff deflates The wrist cuff is that it is firmly wrapped Measurement Method Oscillometric method Correct Disposal of This Product
too soon. loose. around the wrist. Refer to (Waste Electrical & Electronic Equipment)
Measurement Range Pressure: 0 to 299 mmHg
To View the Readings Stored in Memory section 3.1. Pulse: 40 to 180 beats/min. This marking shown on the product or its
literature, indicates that it should not be
1. Press the button, while the average value is Blood pressure readings Accuracy Pressure: ±3 mmHg
disposed of, with other household wastes at the
displayed. constantly vary with time Pulse: ±5% of display reading
end of its working life.
The blood pressure is different of day and how relaxed Inflation Automatic inflation by pump To prevent possible harm to the environment or human health from
The Memory number each time. The reading is you are. Take several uncontrolled waste disposal, please separate this product from
extremely low (or high). deep breaths and try to Deflation Automatic rapid deflation
appears for a second other types of wastes and recycle it responsibly to promote the
before the pulse rate is remain relaxed before Memory 90 Measurements sustainable reuse of material resources.
displayed. The newest taking a measurement.
Alternating Power Source 2 x 1.5V (LR03, AAA alkaline batteries) Household users should contact either the retailer where they
set is numbered “1”.
date/time Battery Life Approx. 300 measurements with new purchased this product, or their local government office, for details
The unit loses
display alkaline batteries at a room temperature of of where and how they can return this item for environmentally safe
power during Replace the batteries with
The batteries are new ones. 23°C recycling.
measurement.
exhausted. Refer to section 2.1. Business users should contact their supplier and check the terms
2. Press the button repeatedly to view the Applied Part = Type B and conditions of the purchase contract. This product should not be
readings stored in memory. mixed with other commercial wastes for disposal.
Nothing happens Protection Against Internally powered ME equipment
To Delete All the Values Stored in Memory when you press The batteries have Insert the batteries with Electric Shock Manufacturer OMRON HEALTHCARE Co., Ltd.
the buttons. been inserted the correct (+/-) polarity. 53, Kunotsubo, Terado-cho, Muko, Kyoto,
Operating temperature/ +10 to +40°C / Maximum: 30 to 85% RH 617-0002 JAPAN
When the memory symbol ( ) appears, first press incorrectly. Refer to section 2.1.
Humidity
the button. Then while holding it down, press the Storage temperature/ -20 to +60°C / Maximum: 10 to 95% RH / EU-representative OMRON HEALTHCARE EUROPE B.V.
START/STOP button simultaneously for about 2 - 3 • Press the START/STOP button and repeat Scorpius 33, 2132 LR Hoofddorp
Humidity/Air pressure 700 to 1060hPa
seconds. measurement. THE NETHERLANDS
• If the problem continues, try replacing the Console Weight Approximately 85 g without batteries www.omron-healthcare.com
Other problems.
batteries with new ones. Outer Dimensions Approximately 87 (w) mm x 64 (h) mm x Production Facility OMRON (DALIAN) CO., LTD.
If this still does not solve the problem, contact 14 (d) mm (without the wrist cuff) Dalian, CHINA
your OMRON retail outlet or distributor. Subsidiary OMRON HEALTHCARE UK LTD.
Measurable Approximately 13.5 to 21.5 cm Opal Drive, Fox Milne, Milton Keynes, MK15 0DG
circumference U.K.
First Second Cuff Material Nylon and polyester OMRON MEDIZINTECHNIK
Note: You cannot partially delete the stored readings.
4.3 Maintenance HANDELSGESELLSCHAFT mbH
To protect your unit from damage, please observe the Package Content Main unit, storage case, battery set, John-Deere-Str. 81a 68163 Mannheim, GERMANY
instruction manual, guarantee card, blood www.omron-medizintechnik.de
following:
pressure pass OMRON SANTÉ FRANCE SAS
14, rue de Lisbonne, 93561 Rosny-sous-Bois
4. Troubleshooting and Maintenance • Do not subject the main unit and the cuff to Cedex, FRANCE
Note: Subject to technical modification without prior notice.
4.1 The Icons and Error Messages extreme temperatures, humidity, moisture or Made in China
direct sunlight.
Error Display Cause Remedy
• Do not disassemble the unit.
Remove the wrist cuff. Wait • Do not subject the unit to strong shocks or • This device fulfils the provisions of EC directive 93/42/EEC
2 - 3 minutes and then take vibrations (for example, dropping the unit on the (Medical Device Directive).
Irregular heartbeats another measurement.
are detected. Repeat the steps in section floor). • This blood pressure monitor is designed according to the European
3.3. If this error continues to • Do not use volatile liquids to clean the main unit. Standard EN1060, Non-invasive sphygmomanometers Part 1:
appear, contact your doctor. General Requirements and Part 3: Supplementary requirements
• Do not wash the cuff or immerse it in water. for electromechanical blood pressure measuring systems.
• Do not use petrol, thinners or similar solvents to • This OMRON product is produced under the strict quality system of
Movement during Carefully read and repeat the clean the cuff. OMRON HEALTHCARE Co. Ltd., Japan. The Core component for
measurement. steps in section 3.3.
• Do not carry out repairs of any kind yourself. If a OMRON blood pressure monitors, which is the Pressure Sensor, is
defect occurs, consult your OMRON retail outlet produced in Japan.
You should replace them with
The batteries are or distributor as mentioned on the packaging.
Blink low. new ones ahead of time.
Refer to section 2.1.

You should replace them with


The batteries are
Lit new ones at once.
exhausted.
Refer to section 2.1.

Wrist cuff not Apply the wrist cuff correctly.


applied correctly. Refer to section 3.1.

Repeat Measurement while


remaining still and refraining
Movement during
from talking during the
measurement.
measurement.
Refer to section 3.3.
• The unit should be cleaned with a soft, dry cloth.
Apply the wrist cuff correctly • Use a soft, moistened cloth and neutral soap to
Wrist cuff not and repeat measurement clean the cuff.
applied correctly, or while remaining still and
movement during refraining from taking during • Keep the unit in its storage case when not in
measurement. the measurement. use.
Refer to section 3.1 and 3.3. • Fold the cuff into the storage case.
Arm position Remain still until the
changed during measurement is complete.
measurement. Refer to section 3.3.

Contact your OMRON retail


Device error.
outlet or distributor.

Note: The irregular heartbeat symbol ( ) and the movement error


symbol ( ) may also be displayed with error messages.
HEM-6221-E.book Page 2 Tuesday, November 6, 2012 2:06 PM

3.5 Using the Memory Function 4.2 Troubleshooting Do not store the unit in the following situations:
Important information regarding Electro Magnetic
The monitor automatically stores the result up to 90 sets.
Problem Cause Remedy • If the unit is wet. Compatibility (EMC)
It can also calculate an average reading based on the
measurements from the last three readings taken within 10 • Locations exposed to extreme temperatures, With the increased number of electronic devices such as PC’s and
minutes. If there are only two readings in memory for that Measure while in the humidity, direct sunlight, dust or corrosive mobile (cellular) telephones, medical devices in use may be
period, the average will be based on two readings. If there is The wrist cuff is not vapours. susceptible to electromagnetic interference from other devices.
correct posture.
one reading in memory for that period, the average will be at heart level. Electromagnetic interference may result in incorrect operation of
Refer to section 3.2. • Locations exposed to vibrations, shocks or
based on one reading. the medical device and create a potentially unsafe situation.
where it will be at an angle. Medical devices should also not interfere with other devices.
Notes:
• If the memory is full, the monitor will delete the oldest The cuff is not In order to regulate the requirements for EMC (Electro Magnetic
readings. wrapped snugly
Wrap the cuff correctly. Calibration and Service Compatibility) with the aim to prevent unsafe product situations, the
• When viewing the reading taken without setting the date and around the wrist.
Refer to section 3.1. • The accuracy of this blood pressure monitor has EN60601-1-2:2007 standard has been implemented. This standard
The reading is been carefully tested and is designed for a long defines the levels of immunity to electromagnetic interferences as
time, “-:--” is displayed instead of the date and time.
extremely low
service life. well as maximum levels of electromagnetic emissions for medical
(or high). devices.
To View the Average Value
The arms and Relax and try taking the • It is generally recommended to have the unit
shoulders are measurement again. inspected every two years to ensure correct This medical device manufactured by OMRON HEALTHCARE
tense. Refer to section 3.3. conforms to this EN60601-1-2:2007 standard for both immunity and
Press the button. functioning and accuracy. Please consult your
emissions.
authorised OMRON dealer or the OMRON Nevertheless, special precautions need to be observed:
Movement or Remain still and do not Customer Service at the address given on the • Do not use mobile (cellular) telephones and other devices, which
talking during talk during Measurement. packaging or attached literature. generate strong electrical or electromagnetic fields, near the
measurement. Refer to section 3.3. medical device. This may result in incorrect operation of the unit
and create a potentially unsafe situation. Recommendation is to
keep a minimum distance of 7 m. Verify correct operation of the
device in case the distance is shorter.
Note: If there are no measurements
results stored in the memory, the
5. Technical Data Further documentation in accordance with EN60601-1-2:2007 is
Wrist cuff pressure Air is leaking from Consult your OMRON
screen to the right is displayed. does not rise. the wrist cuff. retail outlet or distributor. Product Description Wrist Blood Pressure Monitor available at OMRON HEALTHCARE EUROPE at the address
mentioned in this instruction manual.
Model OMRON RS6 (HEM-6221-E) Documentation is also available at www.omron-healthcare.com.
Display LCD Digital Display
Apply the cuff correctly so
Wrist cuff deflates The wrist cuff is that it is firmly wrapped Measurement Method Oscillometric method Correct Disposal of This Product
too soon. loose. around the wrist. Refer to (Waste Electrical & Electronic Equipment)
Measurement Range Pressure: 0 to 299 mmHg
To View the Readings Stored in Memory section 3.1. Pulse: 40 to 180 beats/min. This marking shown on the product or its
literature, indicates that it should not be
1. Press the button, while the average value is Blood pressure readings Accuracy Pressure: ±3 mmHg
disposed of, with other household wastes at the
displayed. constantly vary with time Pulse: ±5% of display reading
end of its working life.
The blood pressure is different of day and how relaxed Inflation Automatic inflation by pump To prevent possible harm to the environment or human health from
The Memory number each time. The reading is you are. Take several uncontrolled waste disposal, please separate this product from
extremely low (or high). deep breaths and try to Deflation Automatic rapid deflation
appears for a second other types of wastes and recycle it responsibly to promote the
before the pulse rate is remain relaxed before Memory 90 Measurements sustainable reuse of material resources.
displayed. The newest taking a measurement.
Alternating Power Source 2 x 1.5V (LR03, AAA alkaline batteries) Household users should contact either the retailer where they
set is numbered “1”.
date/time Battery Life Approx. 300 measurements with new purchased this product, or their local government office, for details
The unit loses
display alkaline batteries at a room temperature of of where and how they can return this item for environmentally safe
power during Replace the batteries with
The batteries are new ones. 23°C recycling.
measurement.
exhausted. Refer to section 2.1. Business users should contact their supplier and check the terms
2. Press the button repeatedly to view the Applied Part = Type B and conditions of the purchase contract. This product should not be
readings stored in memory. mixed with other commercial wastes for disposal.
Nothing happens Protection Against Internally powered ME equipment
To Delete All the Values Stored in Memory when you press The batteries have Insert the batteries with Electric Shock Manufacturer OMRON HEALTHCARE Co., Ltd.
the buttons. been inserted the correct (+/-) polarity. 53, Kunotsubo, Terado-cho, Muko, Kyoto,
Operating temperature/ +10 to +40°C / Maximum: 30 to 85% RH 617-0002 JAPAN
When the memory symbol ( ) appears, first press incorrectly. Refer to section 2.1.
Humidity
the button. Then while holding it down, press the Storage temperature/ -20 to +60°C / Maximum: 10 to 95% RH / EU-representative OMRON HEALTHCARE EUROPE B.V.
START/STOP button simultaneously for about 2 - 3 • Press the START/STOP button and repeat Scorpius 33, 2132 LR Hoofddorp
Humidity/Air pressure 700 to 1060hPa
seconds. measurement. THE NETHERLANDS
• If the problem continues, try replacing the Console Weight Approximately 85 g without batteries www.omron-healthcare.com
Other problems.
batteries with new ones. Outer Dimensions Approximately 87 (w) mm x 64 (h) mm x Production Facility OMRON (DALIAN) CO., LTD.
If this still does not solve the problem, contact 14 (d) mm (without the wrist cuff) Dalian, CHINA
your OMRON retail outlet or distributor. Subsidiary OMRON HEALTHCARE UK LTD.
Measurable Approximately 13.5 to 21.5 cm Opal Drive, Fox Milne, Milton Keynes, MK15 0DG
circumference U.K.
First Second Cuff Material Nylon and polyester OMRON MEDIZINTECHNIK
Note: You cannot partially delete the stored readings.
4.3 Maintenance HANDELSGESELLSCHAFT mbH
To protect your unit from damage, please observe the Package Content Main unit, storage case, battery set, John-Deere-Str. 81a 68163 Mannheim, GERMANY
instruction manual, guarantee card, blood www.omron-medizintechnik.de
following:
pressure pass OMRON SANTÉ FRANCE SAS
14, rue de Lisbonne, 93561 Rosny-sous-Bois
4. Troubleshooting and Maintenance • Do not subject the main unit and the cuff to Cedex, FRANCE
Note: Subject to technical modification without prior notice.
4.1 The Icons and Error Messages extreme temperatures, humidity, moisture or Made in China
direct sunlight.
Error Display Cause Remedy
• Do not disassemble the unit.
Remove the wrist cuff. Wait • Do not subject the unit to strong shocks or • This device fulfils the provisions of EC directive 93/42/EEC
2 - 3 minutes and then take vibrations (for example, dropping the unit on the (Medical Device Directive).
Irregular heartbeats another measurement.
are detected. Repeat the steps in section floor). • This blood pressure monitor is designed according to the European
3.3. If this error continues to • Do not use volatile liquids to clean the main unit. Standard EN1060, Non-invasive sphygmomanometers Part 1:
appear, contact your doctor. General Requirements and Part 3: Supplementary requirements
• Do not wash the cuff or immerse it in water. for electromechanical blood pressure measuring systems.
• Do not use petrol, thinners or similar solvents to • This OMRON product is produced under the strict quality system of
Movement during Carefully read and repeat the clean the cuff. OMRON HEALTHCARE Co. Ltd., Japan. The Core component for
measurement. steps in section 3.3.
• Do not carry out repairs of any kind yourself. If a OMRON blood pressure monitors, which is the Pressure Sensor, is
defect occurs, consult your OMRON retail outlet produced in Japan.
You should replace them with
The batteries are or distributor as mentioned on the packaging.
Blink low. new ones ahead of time.
Refer to section 2.1.

You should replace them with


The batteries are
Lit new ones at once.
exhausted.
Refer to section 2.1.

Wrist cuff not Apply the wrist cuff correctly.


applied correctly. Refer to section 3.1.

Repeat Measurement while


remaining still and refraining
Movement during
from talking during the
measurement.
measurement.
Refer to section 3.3.
• The unit should be cleaned with a soft, dry cloth.
Apply the wrist cuff correctly • Use a soft, moistened cloth and neutral soap to
Wrist cuff not and repeat measurement clean the cuff.
applied correctly, or while remaining still and
movement during refraining from taking during • Keep the unit in its storage case when not in
measurement. the measurement. use.
Refer to section 3.1 and 3.3. • Fold the cuff into the storage case.
Arm position Remain still until the
changed during measurement is complete.
measurement. Refer to section 3.3.

Contact your OMRON retail


Device error.
outlet or distributor.

Note: The irregular heartbeat symbol ( ) and the movement error


symbol ( ) may also be displayed with error messages.