Omron RS6 Wrist Blood Pressure Monitor Instruction Manual (English)
Omron RS6 Wrist Blood Pressure Monitor Instruction Manual (English)
3. Press the START/STOP button to store the setting. After the Positioning symbol lights for more than 2 seconds,
Notes: the wrist cuff will start to inflate automatically.
• If the batteries have been removed, the date and time setting Alternating date/time display
will need to be reset. Cuff Wrapping Guide
• If the date and time are not set, “-:--” appears during or after
measurement.
12K2520
HEM-6221-E.book Page 1 Tuesday, November 6, 2012 2:06 PM HEM-6221-E(A)M05_120926.pdf
3. Press the START/STOP button to store the setting. After the Positioning symbol lights for more than 2 seconds,
Notes: the wrist cuff will start to inflate automatically.
• If the batteries have been removed, the date and time setting Alternating date/time display
will need to be reset. Cuff Wrapping Guide
• If the date and time are not set, “-:--” appears during or after
measurement.
12K2520
HEM-6221-E.book Page 1 Tuesday, November 6, 2012 2:06 PM HEM-6221-E(A)M05_120926.pdf
3. Press the START/STOP button to store the setting. After the Positioning symbol lights for more than 2 seconds,
Notes: the wrist cuff will start to inflate automatically.
• If the batteries have been removed, the date and time setting Alternating date/time display
will need to be reset. Cuff Wrapping Guide
• If the date and time are not set, “-:--” appears during or after
measurement.
12K2520
HEM-6221-E.book Page 1 Tuesday, November 6, 2012 2:06 PM HEM-6221-E(A)M05_120926.pdf
3. Press the START/STOP button to store the setting. After the Positioning symbol lights for more than 2 seconds,
Notes: the wrist cuff will start to inflate automatically.
• If the batteries have been removed, the date and time setting Alternating date/time display
will need to be reset. Cuff Wrapping Guide
• If the date and time are not set, “-:--” appears during or after
measurement.
12K2520
HEM-6221-E.book Page 2 Tuesday, November 6, 2012 2:06 PM
3.5 Using the Memory Function 4.2 Troubleshooting Do not store the unit in the following situations:
Important information regarding Electro Magnetic
The monitor automatically stores the result up to 90 sets.
Problem Cause Remedy • If the unit is wet. Compatibility (EMC)
It can also calculate an average reading based on the
measurements from the last three readings taken within 10 • Locations exposed to extreme temperatures, With the increased number of electronic devices such as PC’s and
minutes. If there are only two readings in memory for that Measure while in the humidity, direct sunlight, dust or corrosive mobile (cellular) telephones, medical devices in use may be
period, the average will be based on two readings. If there is The wrist cuff is not vapours. susceptible to electromagnetic interference from other devices.
correct posture.
one reading in memory for that period, the average will be at heart level. Electromagnetic interference may result in incorrect operation of
Refer to section 3.2. • Locations exposed to vibrations, shocks or
based on one reading. the medical device and create a potentially unsafe situation.
where it will be at an angle. Medical devices should also not interfere with other devices.
Notes:
• If the memory is full, the monitor will delete the oldest The cuff is not In order to regulate the requirements for EMC (Electro Magnetic
readings. wrapped snugly
Wrap the cuff correctly. Calibration and Service Compatibility) with the aim to prevent unsafe product situations, the
• When viewing the reading taken without setting the date and around the wrist.
Refer to section 3.1. • The accuracy of this blood pressure monitor has EN60601-1-2:2007 standard has been implemented. This standard
The reading is been carefully tested and is designed for a long defines the levels of immunity to electromagnetic interferences as
time, “-:--” is displayed instead of the date and time.
extremely low
service life. well as maximum levels of electromagnetic emissions for medical
(or high). devices.
To View the Average Value
The arms and Relax and try taking the • It is generally recommended to have the unit
shoulders are measurement again. inspected every two years to ensure correct This medical device manufactured by OMRON HEALTHCARE
tense. Refer to section 3.3. conforms to this EN60601-1-2:2007 standard for both immunity and
Press the button. functioning and accuracy. Please consult your
emissions.
authorised OMRON dealer or the OMRON Nevertheless, special precautions need to be observed:
Movement or Remain still and do not Customer Service at the address given on the • Do not use mobile (cellular) telephones and other devices, which
talking during talk during Measurement. packaging or attached literature. generate strong electrical or electromagnetic fields, near the
measurement. Refer to section 3.3. medical device. This may result in incorrect operation of the unit
and create a potentially unsafe situation. Recommendation is to
keep a minimum distance of 7 m. Verify correct operation of the
device in case the distance is shorter.
Note: If there are no measurements
results stored in the memory, the
5. Technical Data Further documentation in accordance with EN60601-1-2:2007 is
Wrist cuff pressure Air is leaking from Consult your OMRON
screen to the right is displayed. does not rise. the wrist cuff. retail outlet or distributor. Product Description Wrist Blood Pressure Monitor available at OMRON HEALTHCARE EUROPE at the address
mentioned in this instruction manual.
Model OMRON RS6 (HEM-6221-E) Documentation is also available at www.omron-healthcare.com.
Display LCD Digital Display
Apply the cuff correctly so
Wrist cuff deflates The wrist cuff is that it is firmly wrapped Measurement Method Oscillometric method Correct Disposal of This Product
too soon. loose. around the wrist. Refer to (Waste Electrical & Electronic Equipment)
Measurement Range Pressure: 0 to 299 mmHg
To View the Readings Stored in Memory section 3.1. Pulse: 40 to 180 beats/min. This marking shown on the product or its
literature, indicates that it should not be
1. Press the button, while the average value is Blood pressure readings Accuracy Pressure: ±3 mmHg
disposed of, with other household wastes at the
displayed. constantly vary with time Pulse: ±5% of display reading
end of its working life.
The blood pressure is different of day and how relaxed Inflation Automatic inflation by pump To prevent possible harm to the environment or human health from
The Memory number each time. The reading is you are. Take several uncontrolled waste disposal, please separate this product from
extremely low (or high). deep breaths and try to Deflation Automatic rapid deflation
appears for a second other types of wastes and recycle it responsibly to promote the
before the pulse rate is remain relaxed before Memory 90 Measurements sustainable reuse of material resources.
displayed. The newest taking a measurement.
Alternating Power Source 2 x 1.5V (LR03, AAA alkaline batteries) Household users should contact either the retailer where they
set is numbered “1”.
date/time Battery Life Approx. 300 measurements with new purchased this product, or their local government office, for details
The unit loses
display alkaline batteries at a room temperature of of where and how they can return this item for environmentally safe
power during Replace the batteries with
The batteries are new ones. 23°C recycling.
measurement.
exhausted. Refer to section 2.1. Business users should contact their supplier and check the terms
2. Press the button repeatedly to view the Applied Part = Type B and conditions of the purchase contract. This product should not be
readings stored in memory. mixed with other commercial wastes for disposal.
Nothing happens Protection Against Internally powered ME equipment
To Delete All the Values Stored in Memory when you press The batteries have Insert the batteries with Electric Shock Manufacturer OMRON HEALTHCARE Co., Ltd.
the buttons. been inserted the correct (+/-) polarity. 53, Kunotsubo, Terado-cho, Muko, Kyoto,
Operating temperature/ +10 to +40°C / Maximum: 30 to 85% RH 617-0002 JAPAN
When the memory symbol ( ) appears, first press incorrectly. Refer to section 2.1.
Humidity
the button. Then while holding it down, press the Storage temperature/ -20 to +60°C / Maximum: 10 to 95% RH / EU-representative OMRON HEALTHCARE EUROPE B.V.
START/STOP button simultaneously for about 2 - 3 • Press the START/STOP button and repeat Scorpius 33, 2132 LR Hoofddorp
Humidity/Air pressure 700 to 1060hPa
seconds. measurement. THE NETHERLANDS
• If the problem continues, try replacing the Console Weight Approximately 85 g without batteries www.omron-healthcare.com
Other problems.
batteries with new ones. Outer Dimensions Approximately 87 (w) mm x 64 (h) mm x Production Facility OMRON (DALIAN) CO., LTD.
If this still does not solve the problem, contact 14 (d) mm (without the wrist cuff) Dalian, CHINA
your OMRON retail outlet or distributor. Subsidiary OMRON HEALTHCARE UK LTD.
Measurable Approximately 13.5 to 21.5 cm Opal Drive, Fox Milne, Milton Keynes, MK15 0DG
circumference U.K.
First Second Cuff Material Nylon and polyester OMRON MEDIZINTECHNIK
Note: You cannot partially delete the stored readings.
4.3 Maintenance HANDELSGESELLSCHAFT mbH
To protect your unit from damage, please observe the Package Content Main unit, storage case, battery set, John-Deere-Str. 81a 68163 Mannheim, GERMANY
instruction manual, guarantee card, blood www.omron-medizintechnik.de
following:
pressure pass OMRON SANTÉ FRANCE SAS
14, rue de Lisbonne, 93561 Rosny-sous-Bois
4. Troubleshooting and Maintenance • Do not subject the main unit and the cuff to Cedex, FRANCE
Note: Subject to technical modification without prior notice.
4.1 The Icons and Error Messages extreme temperatures, humidity, moisture or Made in China
direct sunlight.
Error Display Cause Remedy
• Do not disassemble the unit.
Remove the wrist cuff. Wait • Do not subject the unit to strong shocks or • This device fulfils the provisions of EC directive 93/42/EEC
2 - 3 minutes and then take vibrations (for example, dropping the unit on the (Medical Device Directive).
Irregular heartbeats another measurement.
are detected. Repeat the steps in section floor). • This blood pressure monitor is designed according to the European
3.3. If this error continues to • Do not use volatile liquids to clean the main unit. Standard EN1060, Non-invasive sphygmomanometers Part 1:
appear, contact your doctor. General Requirements and Part 3: Supplementary requirements
• Do not wash the cuff or immerse it in water. for electromechanical blood pressure measuring systems.
• Do not use petrol, thinners or similar solvents to • This OMRON product is produced under the strict quality system of
Movement during Carefully read and repeat the clean the cuff. OMRON HEALTHCARE Co. Ltd., Japan. The Core component for
measurement. steps in section 3.3.
• Do not carry out repairs of any kind yourself. If a OMRON blood pressure monitors, which is the Pressure Sensor, is
defect occurs, consult your OMRON retail outlet produced in Japan.
You should replace them with
The batteries are or distributor as mentioned on the packaging.
Blink low. new ones ahead of time.
Refer to section 2.1.
3.5 Using the Memory Function 4.2 Troubleshooting Do not store the unit in the following situations:
Important information regarding Electro Magnetic
The monitor automatically stores the result up to 90 sets.
Problem Cause Remedy • If the unit is wet. Compatibility (EMC)
It can also calculate an average reading based on the
measurements from the last three readings taken within 10 • Locations exposed to extreme temperatures, With the increased number of electronic devices such as PC’s and
minutes. If there are only two readings in memory for that Measure while in the humidity, direct sunlight, dust or corrosive mobile (cellular) telephones, medical devices in use may be
period, the average will be based on two readings. If there is The wrist cuff is not vapours. susceptible to electromagnetic interference from other devices.
correct posture.
one reading in memory for that period, the average will be at heart level. Electromagnetic interference may result in incorrect operation of
Refer to section 3.2. • Locations exposed to vibrations, shocks or
based on one reading. the medical device and create a potentially unsafe situation.
where it will be at an angle. Medical devices should also not interfere with other devices.
Notes:
• If the memory is full, the monitor will delete the oldest The cuff is not In order to regulate the requirements for EMC (Electro Magnetic
readings. wrapped snugly
Wrap the cuff correctly. Calibration and Service Compatibility) with the aim to prevent unsafe product situations, the
• When viewing the reading taken without setting the date and around the wrist.
Refer to section 3.1. • The accuracy of this blood pressure monitor has EN60601-1-2:2007 standard has been implemented. This standard
The reading is been carefully tested and is designed for a long defines the levels of immunity to electromagnetic interferences as
time, “-:--” is displayed instead of the date and time.
extremely low
service life. well as maximum levels of electromagnetic emissions for medical
(or high). devices.
To View the Average Value
The arms and Relax and try taking the • It is generally recommended to have the unit
shoulders are measurement again. inspected every two years to ensure correct This medical device manufactured by OMRON HEALTHCARE
tense. Refer to section 3.3. conforms to this EN60601-1-2:2007 standard for both immunity and
Press the button. functioning and accuracy. Please consult your
emissions.
authorised OMRON dealer or the OMRON Nevertheless, special precautions need to be observed:
Movement or Remain still and do not Customer Service at the address given on the • Do not use mobile (cellular) telephones and other devices, which
talking during talk during Measurement. packaging or attached literature. generate strong electrical or electromagnetic fields, near the
measurement. Refer to section 3.3. medical device. This may result in incorrect operation of the unit
and create a potentially unsafe situation. Recommendation is to
keep a minimum distance of 7 m. Verify correct operation of the
device in case the distance is shorter.
Note: If there are no measurements
results stored in the memory, the
5. Technical Data Further documentation in accordance with EN60601-1-2:2007 is
Wrist cuff pressure Air is leaking from Consult your OMRON
screen to the right is displayed. does not rise. the wrist cuff. retail outlet or distributor. Product Description Wrist Blood Pressure Monitor available at OMRON HEALTHCARE EUROPE at the address
mentioned in this instruction manual.
Model OMRON RS6 (HEM-6221-E) Documentation is also available at www.omron-healthcare.com.
Display LCD Digital Display
Apply the cuff correctly so
Wrist cuff deflates The wrist cuff is that it is firmly wrapped Measurement Method Oscillometric method Correct Disposal of This Product
too soon. loose. around the wrist. Refer to (Waste Electrical & Electronic Equipment)
Measurement Range Pressure: 0 to 299 mmHg
To View the Readings Stored in Memory section 3.1. Pulse: 40 to 180 beats/min. This marking shown on the product or its
literature, indicates that it should not be
1. Press the button, while the average value is Blood pressure readings Accuracy Pressure: ±3 mmHg
disposed of, with other household wastes at the
displayed. constantly vary with time Pulse: ±5% of display reading
end of its working life.
The blood pressure is different of day and how relaxed Inflation Automatic inflation by pump To prevent possible harm to the environment or human health from
The Memory number each time. The reading is you are. Take several uncontrolled waste disposal, please separate this product from
extremely low (or high). deep breaths and try to Deflation Automatic rapid deflation
appears for a second other types of wastes and recycle it responsibly to promote the
before the pulse rate is remain relaxed before Memory 90 Measurements sustainable reuse of material resources.
displayed. The newest taking a measurement.
Alternating Power Source 2 x 1.5V (LR03, AAA alkaline batteries) Household users should contact either the retailer where they
set is numbered “1”.
date/time Battery Life Approx. 300 measurements with new purchased this product, or their local government office, for details
The unit loses
display alkaline batteries at a room temperature of of where and how they can return this item for environmentally safe
power during Replace the batteries with
The batteries are new ones. 23°C recycling.
measurement.
exhausted. Refer to section 2.1. Business users should contact their supplier and check the terms
2. Press the button repeatedly to view the Applied Part = Type B and conditions of the purchase contract. This product should not be
readings stored in memory. mixed with other commercial wastes for disposal.
Nothing happens Protection Against Internally powered ME equipment
To Delete All the Values Stored in Memory when you press The batteries have Insert the batteries with Electric Shock Manufacturer OMRON HEALTHCARE Co., Ltd.
the buttons. been inserted the correct (+/-) polarity. 53, Kunotsubo, Terado-cho, Muko, Kyoto,
Operating temperature/ +10 to +40°C / Maximum: 30 to 85% RH 617-0002 JAPAN
When the memory symbol ( ) appears, first press incorrectly. Refer to section 2.1.
Humidity
the button. Then while holding it down, press the Storage temperature/ -20 to +60°C / Maximum: 10 to 95% RH / EU-representative OMRON HEALTHCARE EUROPE B.V.
START/STOP button simultaneously for about 2 - 3 • Press the START/STOP button and repeat Scorpius 33, 2132 LR Hoofddorp
Humidity/Air pressure 700 to 1060hPa
seconds. measurement. THE NETHERLANDS
• If the problem continues, try replacing the Console Weight Approximately 85 g without batteries www.omron-healthcare.com
Other problems.
batteries with new ones. Outer Dimensions Approximately 87 (w) mm x 64 (h) mm x Production Facility OMRON (DALIAN) CO., LTD.
If this still does not solve the problem, contact 14 (d) mm (without the wrist cuff) Dalian, CHINA
your OMRON retail outlet or distributor. Subsidiary OMRON HEALTHCARE UK LTD.
Measurable Approximately 13.5 to 21.5 cm Opal Drive, Fox Milne, Milton Keynes, MK15 0DG
circumference U.K.
First Second Cuff Material Nylon and polyester OMRON MEDIZINTECHNIK
Note: You cannot partially delete the stored readings.
4.3 Maintenance HANDELSGESELLSCHAFT mbH
To protect your unit from damage, please observe the Package Content Main unit, storage case, battery set, John-Deere-Str. 81a 68163 Mannheim, GERMANY
instruction manual, guarantee card, blood www.omron-medizintechnik.de
following:
pressure pass OMRON SANTÉ FRANCE SAS
14, rue de Lisbonne, 93561 Rosny-sous-Bois
4. Troubleshooting and Maintenance • Do not subject the main unit and the cuff to Cedex, FRANCE
Note: Subject to technical modification without prior notice.
4.1 The Icons and Error Messages extreme temperatures, humidity, moisture or Made in China
direct sunlight.
Error Display Cause Remedy
• Do not disassemble the unit.
Remove the wrist cuff. Wait • Do not subject the unit to strong shocks or • This device fulfils the provisions of EC directive 93/42/EEC
2 - 3 minutes and then take vibrations (for example, dropping the unit on the (Medical Device Directive).
Irregular heartbeats another measurement.
are detected. Repeat the steps in section floor). • This blood pressure monitor is designed according to the European
3.3. If this error continues to • Do not use volatile liquids to clean the main unit. Standard EN1060, Non-invasive sphygmomanometers Part 1:
appear, contact your doctor. General Requirements and Part 3: Supplementary requirements
• Do not wash the cuff or immerse it in water. for electromechanical blood pressure measuring systems.
• Do not use petrol, thinners or similar solvents to • This OMRON product is produced under the strict quality system of
Movement during Carefully read and repeat the clean the cuff. OMRON HEALTHCARE Co. Ltd., Japan. The Core component for
measurement. steps in section 3.3.
• Do not carry out repairs of any kind yourself. If a OMRON blood pressure monitors, which is the Pressure Sensor, is
defect occurs, consult your OMRON retail outlet produced in Japan.
You should replace them with
The batteries are or distributor as mentioned on the packaging.
Blink low. new ones ahead of time.
Refer to section 2.1.
3.5 Using the Memory Function 4.2 Troubleshooting Do not store the unit in the following situations:
Important information regarding Electro Magnetic
The monitor automatically stores the result up to 90 sets.
Problem Cause Remedy • If the unit is wet. Compatibility (EMC)
It can also calculate an average reading based on the
measurements from the last three readings taken within 10 • Locations exposed to extreme temperatures, With the increased number of electronic devices such as PC’s and
minutes. If there are only two readings in memory for that Measure while in the humidity, direct sunlight, dust or corrosive mobile (cellular) telephones, medical devices in use may be
period, the average will be based on two readings. If there is The wrist cuff is not vapours. susceptible to electromagnetic interference from other devices.
correct posture.
one reading in memory for that period, the average will be at heart level. Electromagnetic interference may result in incorrect operation of
Refer to section 3.2. • Locations exposed to vibrations, shocks or
based on one reading. the medical device and create a potentially unsafe situation.
where it will be at an angle. Medical devices should also not interfere with other devices.
Notes:
• If the memory is full, the monitor will delete the oldest The cuff is not In order to regulate the requirements for EMC (Electro Magnetic
readings. wrapped snugly
Wrap the cuff correctly. Calibration and Service Compatibility) with the aim to prevent unsafe product situations, the
• When viewing the reading taken without setting the date and around the wrist.
Refer to section 3.1. • The accuracy of this blood pressure monitor has EN60601-1-2:2007 standard has been implemented. This standard
The reading is been carefully tested and is designed for a long defines the levels of immunity to electromagnetic interferences as
time, “-:--” is displayed instead of the date and time.
extremely low
service life. well as maximum levels of electromagnetic emissions for medical
(or high). devices.
To View the Average Value
The arms and Relax and try taking the • It is generally recommended to have the unit
shoulders are measurement again. inspected every two years to ensure correct This medical device manufactured by OMRON HEALTHCARE
tense. Refer to section 3.3. conforms to this EN60601-1-2:2007 standard for both immunity and
Press the button. functioning and accuracy. Please consult your
emissions.
authorised OMRON dealer or the OMRON Nevertheless, special precautions need to be observed:
Movement or Remain still and do not Customer Service at the address given on the • Do not use mobile (cellular) telephones and other devices, which
talking during talk during Measurement. packaging or attached literature. generate strong electrical or electromagnetic fields, near the
measurement. Refer to section 3.3. medical device. This may result in incorrect operation of the unit
and create a potentially unsafe situation. Recommendation is to
keep a minimum distance of 7 m. Verify correct operation of the
device in case the distance is shorter.
Note: If there are no measurements
results stored in the memory, the
5. Technical Data Further documentation in accordance with EN60601-1-2:2007 is
Wrist cuff pressure Air is leaking from Consult your OMRON
screen to the right is displayed. does not rise. the wrist cuff. retail outlet or distributor. Product Description Wrist Blood Pressure Monitor available at OMRON HEALTHCARE EUROPE at the address
mentioned in this instruction manual.
Model OMRON RS6 (HEM-6221-E) Documentation is also available at www.omron-healthcare.com.
Display LCD Digital Display
Apply the cuff correctly so
Wrist cuff deflates The wrist cuff is that it is firmly wrapped Measurement Method Oscillometric method Correct Disposal of This Product
too soon. loose. around the wrist. Refer to (Waste Electrical & Electronic Equipment)
Measurement Range Pressure: 0 to 299 mmHg
To View the Readings Stored in Memory section 3.1. Pulse: 40 to 180 beats/min. This marking shown on the product or its
literature, indicates that it should not be
1. Press the button, while the average value is Blood pressure readings Accuracy Pressure: ±3 mmHg
disposed of, with other household wastes at the
displayed. constantly vary with time Pulse: ±5% of display reading
end of its working life.
The blood pressure is different of day and how relaxed Inflation Automatic inflation by pump To prevent possible harm to the environment or human health from
The Memory number each time. The reading is you are. Take several uncontrolled waste disposal, please separate this product from
extremely low (or high). deep breaths and try to Deflation Automatic rapid deflation
appears for a second other types of wastes and recycle it responsibly to promote the
before the pulse rate is remain relaxed before Memory 90 Measurements sustainable reuse of material resources.
displayed. The newest taking a measurement.
Alternating Power Source 2 x 1.5V (LR03, AAA alkaline batteries) Household users should contact either the retailer where they
set is numbered “1”.
date/time Battery Life Approx. 300 measurements with new purchased this product, or their local government office, for details
The unit loses
display alkaline batteries at a room temperature of of where and how they can return this item for environmentally safe
power during Replace the batteries with
The batteries are new ones. 23°C recycling.
measurement.
exhausted. Refer to section 2.1. Business users should contact their supplier and check the terms
2. Press the button repeatedly to view the Applied Part = Type B and conditions of the purchase contract. This product should not be
readings stored in memory. mixed with other commercial wastes for disposal.
Nothing happens Protection Against Internally powered ME equipment
To Delete All the Values Stored in Memory when you press The batteries have Insert the batteries with Electric Shock Manufacturer OMRON HEALTHCARE Co., Ltd.
the buttons. been inserted the correct (+/-) polarity. 53, Kunotsubo, Terado-cho, Muko, Kyoto,
Operating temperature/ +10 to +40°C / Maximum: 30 to 85% RH 617-0002 JAPAN
When the memory symbol ( ) appears, first press incorrectly. Refer to section 2.1.
Humidity
the button. Then while holding it down, press the Storage temperature/ -20 to +60°C / Maximum: 10 to 95% RH / EU-representative OMRON HEALTHCARE EUROPE B.V.
START/STOP button simultaneously for about 2 - 3 • Press the START/STOP button and repeat Scorpius 33, 2132 LR Hoofddorp
Humidity/Air pressure 700 to 1060hPa
seconds. measurement. THE NETHERLANDS
• If the problem continues, try replacing the Console Weight Approximately 85 g without batteries www.omron-healthcare.com
Other problems.
batteries with new ones. Outer Dimensions Approximately 87 (w) mm x 64 (h) mm x Production Facility OMRON (DALIAN) CO., LTD.
If this still does not solve the problem, contact 14 (d) mm (without the wrist cuff) Dalian, CHINA
your OMRON retail outlet or distributor. Subsidiary OMRON HEALTHCARE UK LTD.
Measurable Approximately 13.5 to 21.5 cm Opal Drive, Fox Milne, Milton Keynes, MK15 0DG
circumference U.K.
First Second Cuff Material Nylon and polyester OMRON MEDIZINTECHNIK
Note: You cannot partially delete the stored readings.
4.3 Maintenance HANDELSGESELLSCHAFT mbH
To protect your unit from damage, please observe the Package Content Main unit, storage case, battery set, John-Deere-Str. 81a 68163 Mannheim, GERMANY
instruction manual, guarantee card, blood www.omron-medizintechnik.de
following:
pressure pass OMRON SANTÉ FRANCE SAS
14, rue de Lisbonne, 93561 Rosny-sous-Bois
4. Troubleshooting and Maintenance • Do not subject the main unit and the cuff to Cedex, FRANCE
Note: Subject to technical modification without prior notice.
4.1 The Icons and Error Messages extreme temperatures, humidity, moisture or Made in China
direct sunlight.
Error Display Cause Remedy
• Do not disassemble the unit.
Remove the wrist cuff. Wait • Do not subject the unit to strong shocks or • This device fulfils the provisions of EC directive 93/42/EEC
2 - 3 minutes and then take vibrations (for example, dropping the unit on the (Medical Device Directive).
Irregular heartbeats another measurement.
are detected. Repeat the steps in section floor). • This blood pressure monitor is designed according to the European
3.3. If this error continues to • Do not use volatile liquids to clean the main unit. Standard EN1060, Non-invasive sphygmomanometers Part 1:
appear, contact your doctor. General Requirements and Part 3: Supplementary requirements
• Do not wash the cuff or immerse it in water. for electromechanical blood pressure measuring systems.
• Do not use petrol, thinners or similar solvents to • This OMRON product is produced under the strict quality system of
Movement during Carefully read and repeat the clean the cuff. OMRON HEALTHCARE Co. Ltd., Japan. The Core component for
measurement. steps in section 3.3.
• Do not carry out repairs of any kind yourself. If a OMRON blood pressure monitors, which is the Pressure Sensor, is
defect occurs, consult your OMRON retail outlet produced in Japan.
You should replace them with
The batteries are or distributor as mentioned on the packaging.
Blink low. new ones ahead of time.
Refer to section 2.1.
3.5 Using the Memory Function 4.2 Troubleshooting Do not store the unit in the following situations:
Important information regarding Electro Magnetic
The monitor automatically stores the result up to 90 sets.
Problem Cause Remedy • If the unit is wet. Compatibility (EMC)
It can also calculate an average reading based on the
measurements from the last three readings taken within 10 • Locations exposed to extreme temperatures, With the increased number of electronic devices such as PC’s and
minutes. If there are only two readings in memory for that Measure while in the humidity, direct sunlight, dust or corrosive mobile (cellular) telephones, medical devices in use may be
period, the average will be based on two readings. If there is The wrist cuff is not vapours. susceptible to electromagnetic interference from other devices.
correct posture.
one reading in memory for that period, the average will be at heart level. Electromagnetic interference may result in incorrect operation of
Refer to section 3.2. • Locations exposed to vibrations, shocks or
based on one reading. the medical device and create a potentially unsafe situation.
where it will be at an angle. Medical devices should also not interfere with other devices.
Notes:
• If the memory is full, the monitor will delete the oldest The cuff is not In order to regulate the requirements for EMC (Electro Magnetic
readings. wrapped snugly
Wrap the cuff correctly. Calibration and Service Compatibility) with the aim to prevent unsafe product situations, the
• When viewing the reading taken without setting the date and around the wrist.
Refer to section 3.1. • The accuracy of this blood pressure monitor has EN60601-1-2:2007 standard has been implemented. This standard
The reading is been carefully tested and is designed for a long defines the levels of immunity to electromagnetic interferences as
time, “-:--” is displayed instead of the date and time.
extremely low
service life. well as maximum levels of electromagnetic emissions for medical
(or high). devices.
To View the Average Value
The arms and Relax and try taking the • It is generally recommended to have the unit
shoulders are measurement again. inspected every two years to ensure correct This medical device manufactured by OMRON HEALTHCARE
tense. Refer to section 3.3. conforms to this EN60601-1-2:2007 standard for both immunity and
Press the button. functioning and accuracy. Please consult your
emissions.
authorised OMRON dealer or the OMRON Nevertheless, special precautions need to be observed:
Movement or Remain still and do not Customer Service at the address given on the • Do not use mobile (cellular) telephones and other devices, which
talking during talk during Measurement. packaging or attached literature. generate strong electrical or electromagnetic fields, near the
measurement. Refer to section 3.3. medical device. This may result in incorrect operation of the unit
and create a potentially unsafe situation. Recommendation is to
keep a minimum distance of 7 m. Verify correct operation of the
device in case the distance is shorter.
Note: If there are no measurements
results stored in the memory, the
5. Technical Data Further documentation in accordance with EN60601-1-2:2007 is
Wrist cuff pressure Air is leaking from Consult your OMRON
screen to the right is displayed. does not rise. the wrist cuff. retail outlet or distributor. Product Description Wrist Blood Pressure Monitor available at OMRON HEALTHCARE EUROPE at the address
mentioned in this instruction manual.
Model OMRON RS6 (HEM-6221-E) Documentation is also available at www.omron-healthcare.com.
Display LCD Digital Display
Apply the cuff correctly so
Wrist cuff deflates The wrist cuff is that it is firmly wrapped Measurement Method Oscillometric method Correct Disposal of This Product
too soon. loose. around the wrist. Refer to (Waste Electrical & Electronic Equipment)
Measurement Range Pressure: 0 to 299 mmHg
To View the Readings Stored in Memory section 3.1. Pulse: 40 to 180 beats/min. This marking shown on the product or its
literature, indicates that it should not be
1. Press the button, while the average value is Blood pressure readings Accuracy Pressure: ±3 mmHg
disposed of, with other household wastes at the
displayed. constantly vary with time Pulse: ±5% of display reading
end of its working life.
The blood pressure is different of day and how relaxed Inflation Automatic inflation by pump To prevent possible harm to the environment or human health from
The Memory number each time. The reading is you are. Take several uncontrolled waste disposal, please separate this product from
extremely low (or high). deep breaths and try to Deflation Automatic rapid deflation
appears for a second other types of wastes and recycle it responsibly to promote the
before the pulse rate is remain relaxed before Memory 90 Measurements sustainable reuse of material resources.
displayed. The newest taking a measurement.
Alternating Power Source 2 x 1.5V (LR03, AAA alkaline batteries) Household users should contact either the retailer where they
set is numbered “1”.
date/time Battery Life Approx. 300 measurements with new purchased this product, or their local government office, for details
The unit loses
display alkaline batteries at a room temperature of of where and how they can return this item for environmentally safe
power during Replace the batteries with
The batteries are new ones. 23°C recycling.
measurement.
exhausted. Refer to section 2.1. Business users should contact their supplier and check the terms
2. Press the button repeatedly to view the Applied Part = Type B and conditions of the purchase contract. This product should not be
readings stored in memory. mixed with other commercial wastes for disposal.
Nothing happens Protection Against Internally powered ME equipment
To Delete All the Values Stored in Memory when you press The batteries have Insert the batteries with Electric Shock Manufacturer OMRON HEALTHCARE Co., Ltd.
the buttons. been inserted the correct (+/-) polarity. 53, Kunotsubo, Terado-cho, Muko, Kyoto,
Operating temperature/ +10 to +40°C / Maximum: 30 to 85% RH 617-0002 JAPAN
When the memory symbol ( ) appears, first press incorrectly. Refer to section 2.1.
Humidity
the button. Then while holding it down, press the Storage temperature/ -20 to +60°C / Maximum: 10 to 95% RH / EU-representative OMRON HEALTHCARE EUROPE B.V.
START/STOP button simultaneously for about 2 - 3 • Press the START/STOP button and repeat Scorpius 33, 2132 LR Hoofddorp
Humidity/Air pressure 700 to 1060hPa
seconds. measurement. THE NETHERLANDS
• If the problem continues, try replacing the Console Weight Approximately 85 g without batteries www.omron-healthcare.com
Other problems.
batteries with new ones. Outer Dimensions Approximately 87 (w) mm x 64 (h) mm x Production Facility OMRON (DALIAN) CO., LTD.
If this still does not solve the problem, contact 14 (d) mm (without the wrist cuff) Dalian, CHINA
your OMRON retail outlet or distributor. Subsidiary OMRON HEALTHCARE UK LTD.
Measurable Approximately 13.5 to 21.5 cm Opal Drive, Fox Milne, Milton Keynes, MK15 0DG
circumference U.K.
First Second Cuff Material Nylon and polyester OMRON MEDIZINTECHNIK
Note: You cannot partially delete the stored readings.
4.3 Maintenance HANDELSGESELLSCHAFT mbH
To protect your unit from damage, please observe the Package Content Main unit, storage case, battery set, John-Deere-Str. 81a 68163 Mannheim, GERMANY
instruction manual, guarantee card, blood www.omron-medizintechnik.de
following:
pressure pass OMRON SANTÉ FRANCE SAS
14, rue de Lisbonne, 93561 Rosny-sous-Bois
4. Troubleshooting and Maintenance • Do not subject the main unit and the cuff to Cedex, FRANCE
Note: Subject to technical modification without prior notice.
4.1 The Icons and Error Messages extreme temperatures, humidity, moisture or Made in China
direct sunlight.
Error Display Cause Remedy
• Do not disassemble the unit.
Remove the wrist cuff. Wait • Do not subject the unit to strong shocks or • This device fulfils the provisions of EC directive 93/42/EEC
2 - 3 minutes and then take vibrations (for example, dropping the unit on the (Medical Device Directive).
Irregular heartbeats another measurement.
are detected. Repeat the steps in section floor). • This blood pressure monitor is designed according to the European
3.3. If this error continues to • Do not use volatile liquids to clean the main unit. Standard EN1060, Non-invasive sphygmomanometers Part 1:
appear, contact your doctor. General Requirements and Part 3: Supplementary requirements
• Do not wash the cuff or immerse it in water. for electromechanical blood pressure measuring systems.
• Do not use petrol, thinners or similar solvents to • This OMRON product is produced under the strict quality system of
Movement during Carefully read and repeat the clean the cuff. OMRON HEALTHCARE Co. Ltd., Japan. The Core component for
measurement. steps in section 3.3.
• Do not carry out repairs of any kind yourself. If a OMRON blood pressure monitors, which is the Pressure Sensor, is
defect occurs, consult your OMRON retail outlet produced in Japan.
You should replace them with
The batteries are or distributor as mentioned on the packaging.
Blink low. new ones ahead of time.
Refer to section 2.1.