Laboratory testing of
ion exchange and
adsorbent resins
INTRODUCTION
Founded in 1981, Purolite is a leading manufacturer of ion exchange,
catalyst, adsorbent and specialty resins. With global headquarters X]
the United States, Purolite is the only company that focuses 100%^U
its resources on the development and production of resin technology.
This guide describes how to perform a lab-scale used, configure the column with a rubber stopper
process evaluation on the effectiveness of ion at each end, with a glass tube running through
exchange (IEX) and adsorbent resins. each, see Figure 1. A nylon cloth or screen should
be placed over the surface of each stopper to retain
Test results obtained and the depth of work the IEX or adsorbent resin in the bed. Then, fill the
performed may enable further operating condition column to a depth of 15 – 25 mm with small
optimization. While the test may not provide diameter glass beads (3 mm diameter) to help
sufficient information to design a full-scale plant, distribute the fluid and hold the resin above the
it should show if a process is viable and enable the bottom of the column, preventing it from blocking
design of a larger pilot plant, if necessary. the outlet.
Purolite is sensitive to the amount of time and If the column does not have a valve connection to
expense dedicated to developing new and novel control the flow rate at the exit end of the column,
products and processes. If assistance is needed for rubber tubing with a screw clamp assembly can be
resin configuration or system design optimization used as an on/off valve.
during the lab study, Purolite technical experts can
be utilized, and confidentiality agreements can Depending on the laboratory set-up, additional
be arranged. equipment may include a funnel, tubing, a pump,
nylon cloth or glass wool, and a graduated cylinder.
Equipment In sensitive applications, the equipment should be
sterilized and then fully rinsed with high-quality
To carry out initial trials, a 25 mm diameter glass demineralized water before loading the resin.
column with a sintered glass bottom is ideal.
However, a 25 mm diameter glass or plastic column
is suitable to create a basic test system. If this is
Table 1 – Suggested parameters for testing viability of ion exchange and adsorbent resin
Resin regeneration is normally carried out at Three consecutive cycles producing consistent
relatively low flow rates (1 – 6 BV/h) to achieve results should be obtained before changing any
maximum removal from the beads, and is followed operating conditions to optimize performance.
by a slow rinse at a similar flow rate to maximize
removal of the regenerant. It normally takes two or three cycles to obtain
reliable test information.
The final rinse is carried out using a higher flow rate
than the in-service flow rate.