BS-120

Chemistry Analyzer
Technical Specifications

System Function: ISE Module (optional):


Automatic, discrete, random access, Measure parameter: K+, Na+, Cl−
STAT sample priority
Throughput: Up to 100 tests/hour (without ISE), up to Reaction System:
300 tests/hour with ISE (3 ions) Reaction rotor: Rotating tray, containing 40 cuvettes
Measuring principles: Cuvette: Optical length 5mm
Absorbance photometry, turbidimetry, Reaction volume: 180~500μl
Ion Selective Electrode technology Reaction temperature: 37±0.1˚C
Methodology: End-point, fixed-time, kinetic, optional Mixing system: Independent mixing bar
ISE
Single/dual reagent chemistries, Optical System:
monochromatic/bichromatic Light Source: Halogen-tungsten lamp
linear/non-linear multipoint calibration Wavelength: 340nm, 405nm, 450nm, 510nm,
Programming: Open system with user defined profiles 546nm, 578nm, 630nm, 670nm
and calculations Linear range: 0~3.5Abs

Reagent/Sample Handling: Control and Calibration:


Reagent/Sample tray: Calibration mode: Linear (one-point, two-point and multi-point),
Up to 33 positions for sample, Logit-Log 4P, Logit-Log 5P, Spline,
up to 35 positions for reagent; Exponential, Polynomial, Parabola
24 hour non-stop refrigerated Control rules: Westgard multi-rule, Cumulative
compartment (4~15˚C) sum check, Twin plot
Reagent volume:
R1: 180~450μl, step by 1μl Operation Unit:
R2: 30~250μl, step by 1μl Operation system: Windows® XP Professional/Home SP2, Windows® 7
Sample volume: 3~45μl, step by 0.5μl or above Windows® VISTA Home/Business
Reagent/Sample probe: Interface: RS-232
Liquid level detection, collision
protection and inventory checking Working Conditions:
Probe cleaning: Automatic washing both interior and exterior Power Supply: AC 200~240V, 50/60Hz, 800W or
Carry-over < 0.1% AC 100~130V, 50/60Hz, 800W
Automatic sample dilution: Temperature: 15~30˚C
Pre-dilution and post-dilution Humidity: 35~85%
dilution ratio up to 1: 150 Water consumption: 2.5L/hour
Dilution vessel: Disposable cuvette Dimension: Bench top: 690mm(W)x570 mm(D)x595 mm(H)
Weight: 75 Kg
External Bar Code Reader (optional):
Used for sample and reagent programming;
Applicable to various bar code systems
BS-120
including Codabar, ITF (Interleaved Two of
Five), Code128, Code39, UPC/EAN, Code93;
Chemistry Analyzer
capable to communicate with LIS in a
bi-directional mode

is a trademark of Shenzhen Mindray Bio-Medical Electronics Co., Ltd.


Mindray Building, Keji 12th Road South,
2013 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights reserved. Specifications subject to changes without prior notice.
High-tech Industrial Park, Nanshan, Shenzhen 518057, P.R. China
P/N: ENG-BS120-210285x6-20130303
Tel: +86 755 8188 8998 Fax: +86 755 26582680
E-mail: intl-market@mindray.com www.mindray.com
Mindray is listed on the NYSE under the symbol”MR”
BS-120 Dynamic and Real-time display of running status

Chemistry Analyzer Running status of reagent/sample tray and reaction tray

Real-time monitoring of reagent residual volume

Real-time diagnosis of system working status

Discrete, random access, fully automated

100 tests per hour, up to 300 tests per hour with ISE

Up to 33 onboard chemistries and 3 ions Original reaction data record


Real-time monitoring of reaction

Refrigerated reagent compartment Bichromatic testing to avoid interference

Simultaneously display primary and secondary wavelengths


Flexible configuration for sample/reagent positions Detailed profile of alert messages

Automatic probe cleaning, liquid level detection & collision protection

8 wavelengths: 340~670nm
Optimum calibration curve
Automatic dilution for abnormal sample
Linear curve types: One-point linear, Two-point linear and Multi-point linear

Nonlinear curve types: Logistic-Log 4P, Logistic-Log 5P, Exponential 5P,


External bar code reader (optional)
Polynomial 5P and Spline

Bi-directional LIS interface

Flexible sample/reagent tray High quality ISE module (optional)


Optional external reagent/sample bar code Measurements of K+, Na+, Cl

reader 6 months shelf life

Up to 33 positions for sample,

up to 35 positions for reagent

Up to 20/10 virtual sample/reagent trays can

be programmed

24 hour non-stop cooling with Peltier elements

Disposable reaction cuvettes High performance mixer design


Disposable cuvettes to avoid carry-over and Avoid cross contamination

to save testing costs Optimal homogenization in minimum time

Automatic cuvettes blank testing to assure Thoroughly mixes after dispending of sample

precise results or second reagent


Mindray solution for clinical chemistry
Chemistry Reagents
After more than 10 years of research and development on reagents, Mindray can now provide 48
Hepatic Lipids
parameters of dedicated reagents(more than 17 others are coming), covering hepatic, renal, cardiac, Alanine Aminotransferase (ALT) Total Cholesterol (TC)
Aspartate Aminotransferase (AST) Triglycerides (TG)
lipids, diabetes, pancreatitis, inorganic ions and immunalassays, etc.,together with original calibrators
Alkaline Phosphatase (ALP) HDL-Cholesterol (HDL-C)
with metrological traceability as well as controls for BS-120 chemistry analyzer. γ-GlutamylTransferase (γ-GT) LDL-Cholesterol (LDL-C)
Direct Bilirubin (D-Bil) DSA Method Apolipoprotein A1 (ApoA1)
Direct Bilirubin (D-Bil)VOX Method Apolipoprotein B (ApoB)
Total Bilirubin (T-Bil) DSA Method Lipoportein(a) [LP(a)]
Total Bilirubin (T-Bil)VOX Method
Total Protein (TP) Pancreatitis
Albumin (ALB) α-Amylase (α-AMY)
Auto Chemistry Total Bile Acids (TBA) Lipase (LIP)
Analyzer Prealbumin (PA)
Original Quality Adenosine deaminase (ADA) * Diabetes
Control α-L-fucosidase (AFU) * Glucose (Glu) GOD-POD Method
5'-nucleotidase (5'-NT) * Glucose (Glu) HK Meth
Fructosamine (FUN)
Mindray
solution for Renal
clinical chemistry Urea (UREA) Inorganic ions
Creatinine (CREA) Modified JafféMethod Calcium (Ca)
Creatinine (CREA)Sarcosine OxidaseMethod Magnesium (Mg)
Uric Acid (UA) Phosphate Inorganic (P)
Microalbumin*
β2-Microglobulin (β2-MG) * Rheumatism
Cystatin C (CysC) * High sensitivity C-reactive protein (hs-CRP) *
Rheumatoid Factor (RF)
Original System
Antibodies Against Streptolysin O (ASO)
Pack Reagent Original
Cardiac
Calibrators
Creatine Kinase (CK) Immune
with Traceability
Creatine Kinase-MB (CK-MB) Immunoglobulin A (IgA)
Lactate Dehydrogenase (LDH) Immunoglobulin G (IgG)
Original Calibrators with traceability :
α-Hydroxybutyrate Dehydrogenase(α-HBDH) Immunoglobulin M (IgM)
Myoglobin* Immunoglobulin E (IgE) *
Reference Method (Certified by ‘Joint Committee for Traceability in Laboratory Medicine’ (JCTLM))
Complement C3 (C3)
International Federation of Clinical Chemistry and Laboratory Medicine (IFCC)
Complement C4 (C4)
National Institute of Standards and Technology(NIST)
Ferrum C-Reactive Protein (CRP)
Centers for Disease Control and Prevention (CDC, USA) Iron (Fe)
American Association for Clinical Chemistry (AACC) Ferritin (FER) * Others
Transferrin (TRF) * Glucose-6-phosphate dehydrogenase (G6PD) *
Reference Material Total iron binding capacity / unsaturated iron D-dimer*
Institute for Reference Materials and Measurements (IRMM) standards Binding capacity (TIBC/UIBC) * Angiotensin converting enzyme (ACE) *
National Institute of Standards and Technology (NIST) standards Retinol binding protein (RBP) *
World Health Organization (WHO) standards D3-hydroxybutyric acid (D3-HB) *

Japan Committee for Clinical Laboratory (JCCLS) standards


* Coming soon