Item Department/Section Equiptment Description Job Request NO

You might also like

Download as xlsx, pdf, or txt
Download as xlsx, pdf, or txt
You are on page 1of 12

JOB REQUEST DEPARTMENT/SECTION

ITEM EQUIPTMENT DESCRIPTION


NO.

1 19-0357 NICU Weighing Scale

2 19-0372 OR Cautery Machine

3 19-0338 ICU Complex ECG Machine

4
19-0359 ER Droplight
DATE/TIME DATE
BRAND/MODEL SERIAL COMPLAINT
ASSESSED: REQUESTED

Detecto/ Neonate N/A 5/20/2019 For Repair

Bovie/ AARON A2350 AA3116068 5/23/2019 Switch Intermittent

Carewell/ ECG-1103 123L18700773E13 5/14/2019 Failed to print ECG Leads

5/22/2019
N/A N/A Not Functional
ASSESSMENT PARTS NEEDED QTY. ASSESSED BY

Unit is out of order Defective


counter weight For Prieto/ Lloren
further evaluation and assessment and
possible repair/ Correction

Defective power switch for purchase of Prieto/ Lloren


defective parts

Under Warranty
For referral to supplier (centramed) For Prieto/ Lloren
further assessment and repair

Prieto/ Lloren
Defective receptacle for purchase order receptacle 1
ASSIGNED TO
JOB REQUEST DEPARTMENT/SECTION
ITEM EQUIPTMENT DESCRIPTION
NO.

Aircon Unit/ Package Type/Floor


1 1 19-0009 Laboratory/ BloodBank Mounted 3T

2 2 19-0370 MMS/ Supply Bio- Refrigerator

Aircon Unit/
3 3 Radiology Wall Mounted 2HP

4 4

19-0393 ENT-OPD Aircon Unit/ Wall mounted

5
19-0385 Emergency Room/ I.M Aircon Unit/ Floor mounted/ 5T
DATE/TIME DATE
BRAND/MODEL SERIAL ASSESSED: REQUESTED

42BFMFSUCAR03TEA 3/28/2019
Carrier/ ASBFM360BA 1/7/2019
A12PC0021545

FUJI Chiller 5/22/2019 5/22/2019

Media D2102002912157021 2/4/2019

6/26/2019

Carrier 596083 6/20/2019 6/20/2019


COMPLAINT ASSESSMENT PARTS NEEDED QTY.

Compressor Motor
For Check-up NH52NXBT
LRA. 91

Defective filter drier, advice toreplace filter Filter drier 1 pc.


not functioning properly drier and new recharge refrigerant #22 refrigerant #22 2kg

Defective compressor motor


Grounded/ shorted starting starting and
common winding. Recommended for Comp. Motor
For Check-up and repair replacement of compressor motor. Model: PH240X2C- 1 pc.
3FTUI LRA: 45

Defective Motor Terminal Winding, Motor Compressor 1 pc.


commom, starting, running, grounded Mode:
For Check-up No Cooling advice replacement new motor compressor 2P19S3R236A

Model: ZR54KS-TF5-
522 LRA. 1 pc.
non- functional 125/135
ASSESSED BY ASSIGNED TO

Baldon

Tran/ Baldon
JOB REQUEST DEPARTMENT/SECTION
ITEM EQUIPTMENT DESCRIPTION
NO.

1 19-0408 HIMS Filimh Room Stand Fan

2 19-0409 CSS Stand Fan


DATE/TIME DATE
BRAND/MODEL SERIAL COMPLAINT
ASSESSED: REQUESTED

Astron/ Jaguar 81-1635 OF4-16-267889 6/14/2019 not functioning

National/ F 401M 571404135 6/14/2019 6/11/2019 defective


ASSESSMENT PARTS NEEDED QTY. ASSESSED BY

Winding was burned and grounded, Unit


is beyoned repair and recommended for Tahud
CONDEMN.

Winding was burned and grounded, Unit


is beyoned repair and recommended for Tahud
CONDEMN.
ASSIGNED TO