PROBLEM:
It is possible to use the Key Sequence "Ctrl+Alt+Del" with the Service Dongle to go to Maintenance
Mode. This is a wrong Process to go to maintenance mode and can cause system data corruption.
SOLUTION:
It is best to use the Service Dongle with the following process to go to the maintenance mode.
1.) In the imaging mode, to go to the Maintenance mode, insert the Service Dongle and then press
“Ctrl+Alt+R” keys
2.) Click on “Yes” as shown in the figure below.
Technical Support