February 2, 2004 SERVICE NOTES Page 1 of 2

FILE TITLE LOGIQTM 3 SN 73050


TAB 133

Maintenance Mode in LOGIQTM 3

EFFECTIVITY:ALL LOGIQTM 3 SYSTEMS

PROBLEM:
It is possible to use the Key Sequence "Ctrl+Alt+Del" with the Service Dongle to go to Maintenance
Mode. This is a wrong Process to go to maintenance mode and can cause system data corruption.

SOLUTION:
It is best to use the Service Dongle with the following process to go to the maintenance mode.

1.) In the imaging mode, to go to the Maintenance mode, insert the Service Dongle and then press
“Ctrl+Alt+R” keys
2.) Click on “Yes” as shown in the figure below.

Figure 1-1 Log Off Screen

Copyright© 2004 by General Electric Company


February 2, 2004 SERVICE NOTES Page 2 of 2

FILE TITLE LOGIQTM 3 SN 73050


3.) Now, Enter the service password and enter the maintenance mode.

Submitted by Ashwin S. Bangalore,


Service Engineering,
Wipro GE Medical Systems, India.