Professional Documents
Culture Documents
Performance of A Portable Near Infrared Instrument For Brix Value Determination of Intact Mango Fruit
Performance of A Portable Near Infrared Instrument For Brix Value Determination of Intact Mango Fruit
The performance of a commercially-available, portable NIR instrument “FT20” for fruit quality evaluation was in-
vestigated using mango fruits. The calibration result for the FT20 was compared with the result for the Foss
NIRSystems 6500. Partial least squares (PLS) regression was used to make the calibration equations. It was found
that the calibration equation using spectral data obtained with the FT20 had similar accuracy to that using Foss
NIRSystems 6500 spectra. The SEPs of calibration equations were 0.40 °Brix for both instruments.
Keywords: near infrared (NIR) spectroscopy, non-destructive quality evaluation, portable NIR instrument, partial
least squares (PLS) regression, Brix value, mango
Spectral acquisition
NIR spectra of mango fruits were acquired with
the portable NIR instrument “Fruit Tester 20” 28 mm
(FANTEC, Kosai-city, Japan) and the research type
NIR instrument “Model 6500” (Foss NIRSystems,
Silver Spring, USA), hereafter referred to as the
“FT20” and “NIRS6500”, respectively. The spectral 9.3 mm
range of the FT20 was 600–1000 nm while that of 14 mm
25.4 mm
the NIRS6500 was 400–1100 nm. For comparison,
certain characteristics and measuring conditions of
the two instruments are given in Table 1 and Table 2, (a) (b)
respectively.
Figure 1. Sample presentation of mango fruit for FT20
Both the FT20 and the NIRS6500 utilised inter- instrument (a) and NIRS6500 instrument (b). The area
actance probes. The FT20 used a probe in which the filled with crossed lines is the illuminator, while the area
illuminator portion and detector portion were side filled with dots is the detector.
0.60 1.20
(a) Original spectra (a) Original spectra M
0.40 1.00 H
L
0.20 0.80
log (1/R)
log (1/R)
0.00 0.60
M
-0.20 H
0.40
L
-0.40 0.20
-0.60 0.00
750 800 850 900 950 1000 750 800 850 900 950 1000
0.15 0.15
(b) Second derivative spectra
0.10 (b) Second derivative spectra
0.10
d2log (1/R)
d2log (1/R)
0.05 0.05
0.00 0.00
-0.05 -0.05
-0.10 844 978 -0.10 840
962 962
-0.15 -0.15
750 800 850 900 950 1000 750 800 850 900 950 1000
Wavelength (nm) Wavelength (nm)
Figure 2. Original spectra (a) and second derivative spec- Figure 3. Original spectra (a) and second derivative spec-
tra (b) of typical mango fruits having low (L), medium (M) tra (b) of typical mango fruits having low (L), medium (M)
and high (H) Brix values measured with FT20 instrument. and high (H) Brix values measured with NIRS6500 instru-
ment.
Table 4. Calibration and validation results for Brix value using spectral data measured with FT20 in the wavelength region
from 900 to 1000 nm and from 850 to 1000 nm.
850 to 1000 nm, which was wider than the result 200
Regression coefficient (K )
(900–1000 nm) reported in our previous work.6 This 150 (a) 900 ~ 1000nm
is discussed below. 100
In order to explain the result from PLS regression, 984
50
the regression coefficient plots of the FT20 spectra 0
using the wavelength region from 900 to 1000 nm,
-50
and from 850 to 1000 nm are shown in Figure 4(a)
-100
and Figure 4(b), respectively. Figure 4(a) shows that 913 951
-150
the wavelength of 978 nm had an important role in 978
-200
calibration A in Table 4 as was the case for our previ-
ous result using the NIRS6500 with the same wave- 850 900 950 1000
length region (900–1000 nm).6 However, due to the 80
Regression coefficient (K )
second derivative treatment and the spectral limita- 60 (b) 850 ~ 1000nm
tion of 1000 nm for the FT20, there was no signal 40
984
from 978 to 1000 nm as shown in Figure 2(b). There- 20
0 878
fore, using spectral data from 978 to 984 nm as indi-
cated in Figure 4(a) was considered to cause errors -20
giving high SEP value. On the other hand, Figure -40 978
4(b), the regression coefficient plot for calibration -60 954
913
B, shows that the wavelength of 878 nm had an im- -80
portant role while the peak at 978 nm became 850 900 950 1000
smaller than that in calibration A. These plots indi- Wavelength (nm)
cated that using spectral data at 878 nm in place of
978 nm could compensate for the missing data at Figure 4. Regression coefficient plots for PLS calibration
of mango spectra measured with the FT20 over the wave-
978 nm and improve the accuracy of the FT20 cali- length region of 900–1000 nm (a) and 850–1000 nm (b).
bration equation.
The calibration and validation results for the
FT20 and the NIRS6500 are shown in Table 5. Scat-
ter plots between actual and predicted Brix values of
the validation sample set are also shown in Figure 5. to 1000 nm would have similar accuracy to the
The result showed that the calibration equation us- NIRS6500 equation. Both equations had SEPs of
ing FT20 spectra in the wavelength region from 850 0.40 °Brix. Using the 95% pair t-test, there were no
180 Performance of Portable NIR Instrument for Brix Determination of Intact Mango
Table 5. Calibration and validation results for Brix value using spectral data measured with FT20 and NIRS6500 instru-
ments.
Conclusion
The calibration equations developed using spec-
tra measured with the portable NIR instrument
“FT20” and the research NIR instrument
“NIRS6500” had the same accuracy in predicting
Brix value of intact mango. With its compact size,
fast scan, low cost and sufficient accuracy, it was
concluded that the portable NIR instrument “FT20”
was the promising instrument for non-destructive
quality evaluation of fruits such as mango, espe-
cially on trees.
Acknowledgement
This research was supported by the Royal Golden
Jubilee PhD Program under the Thailand Research
Fund (TRF).
References