Professional Documents
Culture Documents
13 Biomedical Instrumentation - Ultrasound
13 Biomedical Instrumentation - Ultrasound
13 Biomedical Instrumentation - Ultrasound
Biomedical Instrumentation
Medical applications of ultrasound
24 weeks
atrium
heart valves
ventricle
2D ultrasound images
US imaging of the heart in real time
US including Doppler measurement of blood flow
Doppler measurement of heart muscle movements
Ultrasound propagation
➢ Mechanical displacement of particles of the media may be
described by the equation
combining displacement s of a particle induced by ultrasound
propagating through the media at velocity c and at the
distance x from the position where US entered the media
and at the instantaneous time t from the moment US
entered the media, where ω is the US frequency
Matter Velocity c (m/s)
x
s A sin t Water 1480
c Air (normal pressure) 331
Plexiglass 2680
Aluminium 6400
Muscles 1585
Blood 1570
Bones of the skull 4080
Brain 1541
11
Ultrasound propagation
➢ The velocity v of the particle
ds
v A cos t
dt
➢ i.e. the displacement velocity of the particle
is lower that the velocity of US along the
directin x.
➢ The motion of the particles is longitudinal
(in nthe direction of US wave propagation).
➢ Transverse oscillations are considered
negligible.
12
Ultrasound propagation
13
Ultrasound propagation
➢ Local density x
0 m sin t
c
➢ Local pressure x
p p0 P sin t
c
➢ where ρo is mean
density, and po mean
pressure
➢ Δρm is the maximal
change in density, ΔP
maximal change in
pressure caused by US
14
US interaction with tissue
15
US interaction with tissue
➢ Scattering
16
US interaction with tissue
➢ Refraction and Reflection
17
Acoustic Impedance
Matter
➢ Acoustic impedance is the
Air (normal 0.0004
resistance to sound wave
conditions)
propogation. The denser the
Water 1.48
material, the larger is the
Aluminium 17.4
acoustic impedance
Steel 45.7
Muscles (middle 1.70
➢ Acoustic impedance of tissue type)
Blood 1.61
P Ns kg Bones of the 7.80
Za c m3 m 2 s skull
v
Brain 1.58
Fat 1.38
Kidney 1.62
Liver 1.65
Aorta 1.69
Lung 0.18
Reflection coeficient
➢ at the boundary of the media (internal organs), a part of the emitted
energy is transmitted to the second layer, and a part of the energy is
reflected (back to the US transducer)
➢ Reflection coeficient R is defined as:
2
I R Z1 Z 2
R
I1 Z1 Z 2
Ultrasound intensity I is defined as
ultrasound energy E that passes
through volume A in a time unit t
E W
I m3
At
19
Transmission coeficient
➢ Transmission coeficient T is defined as:
IT 4 Z1Z 2
T
I1 Z1 Z 2 2
where
T R 1
20
Example:
➢ Reflection coeficient at the boundary Matter
between muscle (mi) and fat (ma) tissue: Air (normal 0.0004
conditions)
2
2
IR Z mi Z ma 1, 7 1,38 Water 1.48
R 0, 01
I1 Z mi Z ma 1, 7 1,38 Aluminium 17.4
Steel 45.7
and between bone (ko) and soft tissue (mt): Muscles (middle 1.70
2 type)
I R Z ko Z mt 7,8 1, 7
2
R 0, 42 Blood 1.61
I1 Z ko Z mt 7,8 1, 7
Bones of the 7.80
skull
Brain 1.58
Fat 1.38
Kidney 1.62
Liver 1.65
Aorta 1.69
Lung 0.18
Attenuation in tissue
➢ US intensity by passing through the homogeneous substance (tissue) is
reduced by the exponential law :
I I 0e x
The reflected intensity of US equals
I2
R e d
I0
where a is the attenuation coeficient
which is different for different tissue
22
Attenuation in tissue
The transducer ha the function of transmitter
and receiver of US waves
US divergence
➢ The angle of divergence of the US wave equals
arcsin 0, 61 0, 61 za << D
D D
➢ Compromise in design: D /
25
US resolution
➢ Resolution depends on:
c
• Diameter of the US beam (d) d
f
• Wavelength of US()
m
➢ Npr. c 1500
s
f 2 MHz
d 0, 75 mm
➢ v is particle velocity,
➢ ρo average density,
➢ c US velocity
➢ The pressure change causes vibration of the US
sensor (having the capacity C) and induces charge
q
q u C
29
US transducers
➢ piezoelectric
(do 1GHz)
➢ Magnetostrictive (10-100kHz)
➢ Materials:
• Crystals
• Quartz
• Lithium niobate
• Amorphous materials:
• Microcrystal struktures
• Ceramics
30
Quartz crystal
31
US transducers - principle
q d x Fx u d x Fx
Ex g x
S 0 S
'
C 0 '
x 0 ' 1
d d Fx Ex Ex
u x Fx x 0 Em d x Em h
C 0 S
32
US transducers
➢ Equivalent circuit
➢ Quality factor
RP L
Q
L RS
➢ Resonant frequency
c
f0
2
33
US transducer
➢ Single sensor & US lens
34
A mode (obsolite)
➢ Propagation through the scull and brain, in oftalmology:
A A0 e
k1u p
A0 e k1k2t
AI r A0 I0 R konst.
35
Exponentila amplifier
➢ With diodes
➢ With MOSFET
36
B mode
Darkness
Brightness
Brightn ess
38
Medical US diagnostic device
➢ Block diagram
39
Medical US diagnostic device
➢ Block diagram
40
US imaging devices
➢ Bedside US device
41
US imaging devices
➢ Handheld or point-of-care
ultrasound imaging device
➢ Ultrasoundprobe compatible
with smart phone
42
US probes
Generation of the
ultrasound wave:
44
US probes
45
US probes
46
US probe with linear array
US lens
Side view
US lens
47
Electronic focusing
Transducers
Variable Variable Transducers
delay delay
Transversal shift
Focus
49
B mode images
➢ US image of abdomen
50
B mode images
Images
obtained with
curvilinear
probe and
with linear
array
transducer
51
M – mode images
➢ US of the heart
52
M – mode images
53
M – mode images
54
Color Doppler images
58
Color Doppler images
59
Conclusions
• Continuously improving image quality =>increased clinical use
• Relatively inexpensive compared to CT and MR – 25% of all
imaging exams in hospitals is US
• Many clinical applications
Soft tissues (fetal, liver, kidneys)
Dynamics of blood flow (heart, circulatory system)
60
US Lithotripsy
• electrohydraulic shock-wave lithotripsy (EHL) -
a shockwave is generated immediately
adjacent to a urolith within the urinary bladder
• extracorporeal shock-wave lithotripsy (ESWL) -
the shock-waves are generated outside the
body and directed toward the urolith
•
61
US Lithotripsy
Extracorporeal shock wave lithotripsy (ESWL)
A nonsurgical technique for treating stones in the bladder,
urethra, kidney, or gallbladder using high-energy US shock
waves generated extracorporealy.
The shock waves, which are generated by a lithotripter, are
focused by x-ray onto the kidney stone. Stones are broken into
"stone dust" or fragments that are small enough to pass in
urine.
Stones that are smaller than 2 cm in diameter are the best size
for SWL. The treatment might not be effective in very large
ones.
Introduced clinically in 1980s.
62
Lithotripter
The technique is based on the
principle that shock waves are not
destructive until they reach a
surface in which there is a change in
acoustical impedance, which is a
form of resistance to the passage of
sound waves. The impedance of
calculi is different from that of
water, bone, and soft tissue;
therefore, tissue through which the
wave travels as well as tissues
surrounding the stone are not
harmed.
Percutaneous ultrasonic lithotripsy (PUL) surgical removal of kidney stones via an incision and insertion of a
nephroscope into the portion of the kidney where the stone is lodged.
An attempt is first made to remove the calculus through the endoscope by basket or forceps.
If this is not successful, an ultrasonic lithotrite that sends out high-frequency sound waves is used to break
up the stone.
A continuous saline irrigation flushes out the particles, which are removed by suction.
65
Literatura:
➢ Šantić, A., “Biomedicinska elektronika”,
Školska knjiga, Zagreb, 1995
➢ Breyer, B., “Medicinski dijagnostički
ultrazvuk”, Školska knjiga, Zagreb, 1991
➢ Brown, BH., Smallwood, RH., et al.,
“Medical Physics and Biomedical
Engineering, IoP Press, Bristol, 1999
➢ Chaussy, C. et al, „Extracorporeal Shock
Wave Lithotripsy in a Nutshell”, Dornier
MedTech Europe GmbH, 2015
66