Download as pdf or txt
Download as pdf or txt
You are on page 1of 7

Research Article

Annals of Clinical Biochemistry


2016, Vol. 53(4) 452–458
! The Author(s) 2015
Reprints and permissions:
sagepub.co.uk/journalsPermissions.nav
DOI: 10.1177/0004563215602028
acb.sagepub.com

The influence of storage time and temperature on the


measurement of serum, plasma and urine osmolality

Karla Bezuidenhout1, Megan A Rensburg2, Careen L Hudson3, Younus Essack4 and


M Razeen Davids1

Abstract
Background: Many clinical laboratories require that specimens for serum and urine osmolality determination be
processed within 3 h of sampling or need to arrive at the laboratory on ice. This protocol is based on the World
Health Organization report on sample storage and stability, but the recommendation lacks good supporting data. We
studied the effect of storage temperature and time on osmolality measurements.
Methods: Blood and urine samples were obtained from 16 patients and 25 healthy volunteers. Baseline serum, plasma
and urine osmolality measurements were performed within 30 min. Measurements were then made at 3, 6, 12, 24 and
36 h on samples stored at 4–8 C and room temperature. We compared baseline values with subsequent measurements
and used difference plots to illustrate changes in osmolality.
Results: At 4–8 C, serum and plasma osmolality were stable for up to 36 h. At room temperature, serum and plasma
osmolality were very stable for up to 12 h. At 24 and 36 h, changes from baseline osmolality were statistically significant
and exceeded the total allowable error of 1.5% but not the reference change value of 4.1%. Urine osmolality was
extremely stable at room temperature with a mean change of less than 1 mosmol/kg at 36 h.
Conclusions: Serum and plasma samples can be stored at room temperature for up to 36 h before measuring osmo-
lality. Cooling samples to 4–8 C may be useful when delays in measurement beyond 12 h are anticipated. Urine osmolality
is extremely stable for up to 36 h at room temperature.

Keywords
Analytes, laboratory methods
Accepted: 31st July 2015

1
Introduction Division of Nephrology, Department of Medicine, Stellenbosch
University and Tygerberg Hospital, Cape Town, South Africa
2
Serum and urine osmolality are frequently measured in Division of Chemical Pathology, Stellenbosch University and National
clinical practice to assist with the diagnosis and manage- Health Laboratory Service, Cape Town, South Africa
3
National Health Laboratory Service, Greenpoint Complex, Cape Town,
ment of polyuria, hypernatraemia and hyponatraemia. It is
South Africa
also applied in the calculation of serum or urine osmolal 4
PathCare Laboratories, Cape Town, South Africa
gaps and can be used in the estimation of urinary ammo-
nium excretion.1 Specimens for serum and urine osmolality Corresponding author:
Razeen Davids, Division of Nephrology at the Faculty of Medicine and
determination are often required to be processed within 3 h Health Sciences of Stellenbosch University, Francie van Zijl Avenue,
of sampling or must be transported and delivered to the Tygerberg 7505, Cape Town, South Africa.
laboratory on ice. This protocol has been in place in many Email: mrd@sun.ac.za
Bezuidenhout et al. 453

laboratories since 2009 and is based on the World Health Most laboratories consider plasma and serum osmo-
Organization (WHO) report on sample storage and stabil- lality to be identical.6,9 Serum is considered to be more
ity.2 Implementation of this protocol results in the rejection stable than plasma for the measurement of many ana-
of many samples with obvious cost implications and a lytes, when there is prolonged contact with cells. In
negative impact on effective clinical decision making. Boyanton and Blick’s stability study,4 the changes
According to the WHO report, serum/plasma osmo- seen over time in lactate, glucose, potassium, pH,
lality is stable for 3 h at 20 to 25 C, for 1 day at 4–8 C CO2 and HCO3 were more pronounced in plasma
and for three months at 20 C. It also states that urine than in serum. This suggests that clotted serum samples
osmolality is stable for 3 h at 20–25 C, for seven days at might be the preferred medium for measuring osmolal-
4–8 C and for more than three months at 20 C. The ity when separation is delayed.
recommendations are based on a study by Zhang et al.3 To the best of our knowledge, the only stability data
on the effects of prolonged serum-clot contact time on available for urine osmolality is a white paper from an
the determination of commonly measured analytes. osmometer manufacturing company, which reported
They found that potassium and glucose concentrations good stability over 24 h.10 The WHO report refers to
had clinically relevant changes at 3 h and that phosphate Cooper et al.,11 but this study has no data on the sta-
concentration was elevated by 6 h. Sodium, calcium, bility of urinary analytes or osmolality. It is generally
urea and chloride concentrations were stable over 24 h. advised that urine specimens be refrigerated or chem-
The WHO recommendations for plasma and serum ically treated to prevent bacterial growth if samples are
osmolality stability are therefore based on the instability to be stored before analysis. Bacteria may cause con-
of analytes that contribute to osmolality, rather than on version of glucose into acids and alcohols, and urea
stability studies which measured osmolality directly.3 into ammonia. They also utilize glucose and ketones,
Boyanton and Blick4 demonstrated a decrease in glu- causing a decrease in their concentration. For determin-
cose and HCO3 concentration and an increase in lac- ing urine osmolality, refrigeration, but not the addition
tate and CO2 concentrations when serum and plasma of preservatives, is advised.12
separation is delayed. These changes are attributed to In summary, guidelines for the handling of samples
on-going glycolysis by red blood cells in vitro. The for- for osmolality lack good supporting data. The stability
mation of lactate during glycolysis yields Hþ ions that of chemical analytes contributing to serum and plasma
are buffered by HCO3. HCO3 concentration conse- osmolality has been studied well, but only one study
quently decreases and CO2 concentration increases over directly investigated the effect of temperature and
time. Changes in the concentrations of HCO3 and delayed serum separation on the determination of
CO2 were shown to influence plasma osmolality by osmolality, and then only over a 5-h period. No studies
Boning and Maassen.5 It is therefore reasonable to have directly examined the stability of urine and plasma
expect osmolality to change over time when separation osmolality. In view of this knowledge gap, we investi-
from cellular components is delayed. gated the effect of delayed serum and plasma separ-
Osmolality stability studies where serum or plasma is ation, and cooling, on osmolality measurements. We
separated from cellular components almost immedi- also compared serum with plasma osmolality to deter-
ately after sampling and then stored have limited rele- mine the most stable medium. Last, we investigated the
vance for clinical practice where there are often delays effect of storage time and temperature on the stability
in getting samples to the laboratory and where plasma of urine osmolality.
or serum has not yet been separated.6,7 The direct effect
on osmolality of delaying serum separation was exam-
ined by Redetzki et al.8 who showed that the osmolality Materials and methods
of serum samples stored for 1–4 h at room temperature Assessment of the sample rejection rate based on
before separation was, on average, 2.8 mosmol/kg
current laboratory protocols
higher than that of baseline samples which were
cooled immediately and measured within 20 min of The sample rejection rate for serum and urine osmolal-
sampling. They demonstrated that this difference was ity was determined by a search of the Tygerberg
attributable to the formation of lactate. In another set Hospital NHLS database for the period 1 January to
of samples, they also demonstrated an average increase 31 December 2012.
in serum osmolality of 3.5 mosmol/kg over 5 h at room
temperature and 1.5 mosmol/kg at 0 C. The changes Participants and samples for osmolality
from baseline correlated very well with the changes in
measurements
lactate concentration. The effect of storage temperature
and delayed serum separation was, however, not stu- We obtained a total of 30 blood samples and 40 urine
died beyond 5 h. samples over a period of six weeks. These were
454 Annals of Clinical Biochemistry 53(4)

provided by 16 patients and 25 healthy volunteers. The and 0.36% for urine. Each measurement was per-
patients were selected on the basis of having conditions formed in duplicate. When measurements differed by
such as hyperglycaemia, hyponatraemia and renal fail- more than 5 mosmol/kg, a third measurement was
ure, which are known to affect osmolality. To ensure a made. The average of the two closest values was
good spread of urine osmolality, some volunteers were taken to be the true value.
sampled after an overnight fast and others after taking
a water load.
Statistical analysis
Repeated-measures analysis of variance was used to
Specimen collection and handling compare baseline values with subsequent measurements
After obtaining written informed consent, 45–50 mL of at each time point, to compare serum and plasma, and
blood was drawn from a forearm vein. The sample was cooled and room temperature specimens. We used
immediately divided among 11 serum-separating tubes Bland–Altman difference plots to illustrate the devi-
and 11 lithium heparin tubes (BD VacutainerÕ , Becton, ation from baseline osmolality at different time
Dickinson South Africa, Gauteng, South Africa). points. The total allowable error for serum osmolality
Serum-separating tubes were each filled with 1 mL of of 1.5% as specified in the Westgard Biological
blood and heparin lithium tubes with 3 mL of blood. Variation Database, and the reference change value of
One serum and one plasma tube were centrifuged, and 4.1% was used to determine serum and plasma sample
osmolality was determined within 30 min to provide a stability at the various time points.13 The reference
baseline value for each medium. Five tubes from each change value (RCV) indicates whether a true clinical
medium were stored at 4–8 C and the remaining five change in the value has been observed. The RCV was
tubes at room temperature. Pairs of serum and plasma calculated using a bidirectional Z-score of 1.96, a
samples from the cooled and room temperature sets of within-subject biological variation of 1.3% and an ana-
samples were centrifuged and measured at 3, 6, 12, 24 lytical imprecision of 0.7%.14 The biological variation
and 36 h. in urine osmolality is large, hence the large total allow-
The stability of urine osmolality was assessed in a able error of 39.5% quoted in the Westgard Database13
similar manner. A fresh urine sample was divided and the RCV of 88%. This is too large to be clinically
amongst 11 non-sterile, plastic 5 mL tubes containing useful and, in the absence of other guidelines, we con-
no additives. A baseline sample was centrifuged and sidered a deviation from baseline values of >5% to be
measured within 30 min. The remaining samples were of clinical significance. A P-value of <0.05 was used to
split into a group of tubes stored at 4–8 C and a group indicate statistical significance.
stored at room temperature. Samples from each group
then had osmolality determinations at 3, 6, 12, 24 and
36 h.
Results
There were 483 requests for serum osmolality and 628
requests for urine osmolality in 2012 at Tygerberg
Osmometry
Hospital. One-hundred and nine (22%) of the serum
All osmolality measurements were determined by freez- samples and 132 (21%) of the urine samples were
ing-point depression. Samples from the 16 patients were rejected because they were not transported on ice and
analysed on the Advanced Micro-Osmometer 3320 had not reached the laboratory within 3 h of sampling.
(Advanced Instruments, Inc., Norwood, MA, USA). In the samples obtained from our participants, blood
The instrument has a within-day coefficient of variation osmolality measurements ranged from 252 to 311
(CV) of 0.9% for measurements below 400 mosmol/kg mosmol/kg (n ¼ 30). No haemolysed specimens were
and a CV of 0.5% for measurements between 400 and received. Urine osmolality ranged from 49 to 1157
2000 mosmol/kg. Baseline measurements falling outside mosmol/kg (n ¼ 40).
the 280–295 mosmol/kg reference intervals for serum/ Table 1 shows the changes in serum, plasma and
plasma osmolality were repeated as per the standard urine osmolality over time. The average room tempera-
operating procedure in our laboratory. Measurements ture during the period of data collection was 22 C.
at later time points were repeated if they differed by Figure 1 demonstrates the serum and plasma osmo-
more than 10 mosmol/kg from the baseline value to lality at 4–8 C and room temperature at each time
minimize random errors. point. In serum samples kept at room temperature,
A Micro-Osmometer 3300 (Advanced Instruments, there was no difference between measurements at base-
Incorporated Norwood, MA, USA) was used to meas- line and those made at 3, 6 or 12 h. The average change
ure the samples from the healthy volunteers. This from baseline was 0.4 mosmol/kg at 3 h (P ¼ 0.60),
instrument has a within-day CV of 0.87% for serum 1.0 mosmol/kg at 6 h (P ¼ 0.20) and 1.1 mosmol/kg
Bezuidenhout et al. 455

Table 1. Changes in serum, plasma and urine osmolality in stored samples over time.

Serum Plasma Urine


Time from
sampling (h) 4–8 C Room temperature 4–8 C Room temperature 4–8 C Room temperature

3 0.2  3.3 0.4  2.9 2.0  4.5 1.1  4.9 1.6  4.0 0.1  6.1
6 0.5  3.0 1.0  4.4 1.8  4.3 0.1  4.7 1.0  4.6 1.7  4.4
12 0.7  3.1 1.1  2.5 2.1  4.3 0.7  4.2 0.3  4.5 0.1  5.1
24 0.2  3.7 3.5  3.6 a 0.6  5.7 3.5  5.7 a 1.5  3.7 0.7  4.3
36 0.9  4.4 6.4  4.2 a 0.9  10.1 3.3  3.3 a 1.2  3.2 0.5  5.2
Mean differences from baseline values  SD are shown in mosmol/kg.
a
Indicates statistically significant changes from baseline values.

Figure 1. Serum and plasma osmolality at each time point. Mean values and 95% confidence intervals are shown. In serum samples at
room temperature, there was an increase of 3.5 mosmol/kg (P < 0.005) at 24 h and 6.4 mosmol/kg (P < 0.005) by 36 h. In cooled
samples, there were no significant changes from baseline measurements for up to 36 h. In plasma samples at room temperature, the
average change from baseline at 24 h was 3.5 mosmol/kg (P ¼ 0.001), with no further increase at 36 h. In cooled samples, there was a
fall in plasma osmolality at 12 h of 2.1 mosmol/kg (P ¼ 0.04) but no significant changes at the other time points.

at 12 h (P ¼ 0.15). Statistically significant changes were average change from baseline in the cooled samples
present at 24 and 36 h. At 24 h, the average change was less than 1 mosmol/kg at all time points. For
from baseline was an increase of 3.5 mosmol/kg plasma, there was a small decease in osmolality from
(P < 0.05), and by 36 h, the mean increase was 6.4 baseline values of approximately 2 mosmol/kg at 3,
mosmol/kg (P < 0.05). 6 and 12 h, and then at 24 and 36 h small increases
In plasma samples kept at room temperature, no of less than 1 mosmol/kg from baseline values were
significant change from baseline measurements was recorded.
observed up to 12 h. Like serum, the average change Figures 2 and 3 illustrate the changes in serum osmo-
from baseline at 24 h was 3.5 mosmol/kg, but at 36 h, lality at 24 and 36 h for samples stored at room tem-
there was no further increase. perature. Figure 4 illustrates the changes in plasma
Serum and plasma samples stored at 4–8 C osmolality at 36 h at room temperature. Dashed lines
showed no significant change from baseline osmolal- indicate the limits of the total allowable error and
ity measurements (see Table 1). For serum, the dotted lines indicate the reference change value.
456 Annals of Clinical Biochemistry 53(4)

Figure 2. Serum osmolality of samples at room temperature: deviation from baseline at 24 h. Dashed lines indicate the limits of the
total allowable error and dotted lines indicate the reference change value.

Figure 3. Serum osmolality of samples at room temperature: deviation from baseline at 36 h. Dashed lines indicate the limits of the
total allowable error and dotted lines indicate the reference change value.

Urine osmolality was found to be extremely stable. deviate by more than 2% (or 10 mosmol/kg) from base-
Analysis of variance with repeated measures for the line values.
factor time indicated no significant differences between
baseline measurements and measurements made at 3, 6,
12, 24 and 36 h at 4–8 C or at room temperature
Discussion
(P ¼ 0.21). The average deviations from baseline osmo- The investigation of salt and water disorders and the
lality measurements did not exceed 2 mosmol/kg at any calculation of serum and urine osmolal gaps depend on
time point (see Table 1). Figure 5 illustrates percentage the accurate measurement of osmolality. Our study has
changes from baseline urine osmolality at 36 h. demonstrated good stability of osmolality in blood and
Absolute differences are shown in Figure S4 of the sup- urine samples stored at room temperature for up to
plementary materials. Most measurements do not 36 h. The wide range of blood and urine osmolality
Bezuidenhout et al. 457

Figure 4. Plasma osmolality of samples at room temperature: deviation from baseline at 36 h. Dashed lines indicate the limits of the
total allowable error and dotted lines indicate the reference change value.

Figure 5. Urine osmolality of samples at room temperature: percentage deviation from baseline at 36 h. Most measurements do not
deviate by more than 2% from baseline values.

included in the study makes the results applicable to a temperature. At 24 h, the average increase for both
variety of clinical settings, and it therefore appears that serum and plasma osmolality was 1.2%. At 36 h,
the rejection of so many samples based on the current there was no further change for plasma samples but
recommendations can be avoided. the osmolality of the serum samples increased to
There were no changes in the osmolality of serum 2.1% above baseline values, exceeding the total allow-
and plasma samples stored at 4–8 C for up to 36 h. In able error. Importantly, these changes did not exceed
the samples stored at room temperature, we found no the reference change value and were not regarded as
changes in the osmolality of samples stored for up to clinically significant.
12 h. At 24 and 36 h, there were statistically significant In our setting, most samples for osmolality reach the
increases in the osmolality of the samples kept at room laboratory within 12 h. Sending samples on ice or
458 Annals of Clinical Biochemistry 53(4)

Funding
storing them at 4–8 C before measurement is therefore
The osmolality measurements were provided by the National Health
not necessary. In rural clinical settings, the time from Laboratory Service and PathCare Laboratories of South Africa.
sampling to measurement may well exceed 24 or even
36 h. In such cases, it would be acceptable to do serum Ethical approval
osmolality measurements up to 36 h, but cooling serum Ethical approval was granted by the Health Research Ethics Committee of
specimens to 4–8 C or sending a heparinized specimen Stellenbosch University (reference number S13/03/052).
for measurement of plasma osmolality may be prefer-
able if delays beyond 12 h are anticipated. Guarantor
Our data revealed excellent stability of urine osmo- MRD.
lality over 36 h. There was no difference in osmolality
measurements between urine samples stored at 4–8 C Contributorship
and those kept at room temperature. The average devi- KB and MRD conceived the study, developed the initial protocol and gained
ation from baseline at 36 h was 0.46 mosmol/kg in sam- ethical approval. MAR, CLH and YE contributed to refining the protocol. KB,
MAR, CLH and YE were involved with data collection. KB, MRD and MAR
ples at room temperature. The total allowable error for were involved with data analysis. KB wrote the first draft of the manuscript. All
urine osmolality is 39.5% and the RCV is 88%.12 Most authors reviewed earlier drafts of the manuscript and approved the final version.
of our samples measured at 36 h were within 2%
(or within 10 mosmol/kg) of baseline osmolality References
values. Sending urine samples on ice or storing them 1. Halperin ML and Goldstein MB. Fluid, electrolyte, and acid-base physi-
ology: a problem-based approach, 3rd ed. Philadelphia, PA: W.B.
at 4–8 C before measurement is therefore not required. Saunders, 1999.
The effect on osmolality of storing samples at higher 2. World Health Organization. Use of anticoagulants in diagnostic laboratory
ambient temperatures was not evaluated in this study. investigations and stability of blood, plasma and serum samples. Geneva:
World Health Organization, 2002.
We therefore cannot comment on the stability of sam- 3. Zhang DJ, Elswick RK, Miller WG, et al. Effect of serum-clot contact time
ples when very warm transportation or storage condi- on clinical chemistry laboratory results. Clin Chem 1998; 44: 1325–1333.
tions exist. 4. Boyanton BL Jr and Blick KE. Stability studies of twenty-four analytes in
human plasma and serum. Clin Chem 2002; 48: 2242–2247.
In conclusion, we suggest that existing recommenda- 5. Boning D and Maassen N. Blood osmolality in vitro: dependance on pCO2,
tions for sample stability and storage for osmolality be lactic acid concentration, and O2 saturation. J Appl Physiol Respir Environ
amended. Serum and plasma samples can be stored for Exerc Physiol 1983; 54: 118–122.
6. Seifarth CC, Miertschischk J, Hahn EG, et al. Measurement of serum and
up to 36 h before separation and measurement of osmo- plasma osmolality in healthy young humans – influence of time and storage
lality. Cooling blood samples to 4–8 C is advised when conditions. Clin Chem Lab Med 2004; 42: 927–932.
delays in measurement beyond 12 h are anticipated. 7. Bohnen N, Terwel D, Markerink M, et al. Pitfalls in the measurement of
plasma osmolality pertinent to research in vasopressin and water metabol-
Samples for urine osmolality can be stored for up to ism. Clin Chem 1992; 38: 2278–2280.
36 h before measurement and cooling of these samples 8. Redetzki HM, Hughes JR and Redetzki JE. Differences between serum and
is not required. plasma osmolalities and their relationship to lactic acid values. Proc Soc
Exp Biol Med 1972; 139: 315–318.
9. Bunting PS, Reid SR and Ngan M. Comparison of serum and plasma
Note osmolalities in a surgical ICU. Crit Care Med 1986; 14: 650.
Since the acceptance of this article the authors have become aware of the letter 10. Curria A. Refrigerated and room temperature storage stability of urine osmo-
by Abbadi et al.15 reporting good stability of osmolality in heparinised whole lality measurements. Norwood, MA: Advanced Instruments, Inc., 2011.
blood samples kept at room temperature for up to 24 h. 11. Cooper GR, Sampson EJ and Smith SJ. Preanalytical, including biological,
variation in lipid and apolipoprotein measurements. Curr Opin Lipidol
1992; 3: 365–371.
Acknowledgements 12. Garcia LS. Clinical laboratory management. Washington, DC: ASM Press,
We thank Wessel Kleinhans for his help in determining the specimen rejection 2004.
rate as well as Tarryn Abrahams and the technicians at the Tygerberg Hospital 13. Ricos C, Alvarez V, Cava F, et al. Desirable specifications for total error,
NHLS and the PathCare laboratories for the osmolality measurements. We imprecision, and bias, derived from intra- and inter-individual biologic
also thank Justin Harvey of the Stellenbosch University Centre for Statistical variation. http://www.westgard.com/biodatabase1.htm (accessed 23 May
Consultation for assistance with data analysis. 2015).
14. Fraser CG. Biological variation: from principles to practice. Washington,
DC: AACC Press, 2001.
Declaration of conflicting interests 15. Abbadi A, El-Khoury JM and Wang S. Stability of serum and plasma
The author(s) declared no potential conflicts of interest with respect to the osmolality in common clinical laboratory storage conditions. Clin
research, authorship, and/or publication of this article. Biochem 2014; 47: 686–687.

You might also like