Certificate of Quality

Pall Corporation certifies the specified product meets all manufacturing specifications and criteria for release.

Pall ® : MEMBRANE DISC FILTERS


Part Number: 66191
Membrane Type: GN-6 Metricel Physical Tests
® We verified the
(Hydrophilic mixed Cellulose Esters) Criteria Specification Results
Pore Size: 0.45 µm Depending on
Lot Number: T71454 P/N:
Expiration Date: September/2020 Thickness 114.3 - 190.5 µm Conform
(4.5 - 7.5 mils)
Product Description Diameter 47 mm Conform
Pall Laboratory microbiology hydrophilic membranes
Bubble Point Minimum of 1.8 Conform
are available as single packed membranes or as 200
bar
membranes for use with Sentino® Filter Dispenser.
(180 kPa, 1.8 psi)
Both configurations are optimal for microbiological
in water
testing as membrane filtration is the method of choice
for testing. The dotted gridlines provide guidance for Water Flow Rate 65 mL/min/cm² Conform
qualification of bacterial or fungal colonies without at 0.7 bar (70 kPa,
growth inhibition or enhancement. 65 psi)
Wet Time 15 seconds Conform
Materials of Construction
Biological Tests Growth Recovery
Filter Media: GN-6 Metricel (Hydrophilic mixed
The membranes demonstrate typical recovery of the
Cellulose Esters) target microorganism listed below as directed per ISO
7704 and Standard Methods for the Examiniation of
Gamma Irradiation Information Water and Wastewater.
All S-pack membranes and Sentino Dispenser Pack Microorganism Specification - Results
Refills are gamma irradiated with a maximum dosage of Acceptable
30 kGys. A Quarterly Dose Audit is conducted to growth
demonstrate continued effectiveness of the dose Bacterial 90% batch Complies
delivered as required in ANSI/AAMI/ISO 11137-2. Recovery average of E. coli
ATCC 11229 on
Quality Assurance Lot Release Criteria mFC and
This product is tested and monitored for compliance to mENDO
ISO 7704 and Standard Methods for the Examination of microbiological
Water and Wastewater with the following characteristics: media
In addition to the above tests, this product met manufacturing inspection standards and requirements for full
traceability by the lot number. This product was manufactured in a Pall facility under a Quality Management
System and Environmental Management program approved by an accredited registering body to ISO 9001 and ISO
14001, respectively. Consider only unopened, undamaged packages for use. Further information is available by
contacting Pall.
13/September/2017

James Merrill, Quality Manager, Pall Pensacola Date of Manufacture


Pall Pensacola, 8780 Ely Road, Pensacola, FL 32514
CoQ0047E rev 01 www.pall.com
Filtration. Separation. Solution.SM © Copyright 2017, Pall Corporation. Pall, , Sentino, and Metricel are
trademarks of Pall Corporation.