How To Calibrate The Accuracy For DI-2000
How To Calibrate The Accuracy For DI-2000
How To Calibrate The Accuracy For DI-2000
Please try to check out the accuracy error with analyzer or measuring cylinder according to the below.
First of all, you have to set "USER MODE" and please check out the accuracy of the pump 3 times.
If you can see "U001" on third display, press the button "MODE" during 3 seconds.
* How to enter the "USER MODE" of the DI-2000 *
If you can see "U001" on third display, press the button "MODE" during 3 seconds.
1. Please press button “Stop” and "Sel" at the same time during about 3 seconds.
2. If you can see "I V.St" on first display, press button “MODE” 2 times and then you can see "Set 3" on second display.
3. Please press button “BOULS” 1 time and check out the accuracy against below procedure.
* Procedure *
1. Please check out the measure flow rate(50ml/h or 100ml/h) during 1 hour by using analyzer or measuring cylinder.
2. Please repeat 3 times every one pump.
3. Please calculate the everage of accuracy error.
4. Please adjust the calibration value of USER MODE in the pump.
5. Please check out the measure flow rate(50ml/h or 100ml/h) during 1 hour by using the pump analyzer or measuring cylinder.
1. Please press the buton "Stop" and "Sel" at the same time during 3 seconds.
2. If you can see "U001" on third display, press the button "MODE" during 3 seconds.
3. Please press the button "BOLUS" after adjusting the calibration value.
(Example)
- Error rate = (Set the Flow rate – Measure Flow rate) / Set Flow rate * 100
Error rate: (Set Flow rate - Measure Flow rate) / Set Flow rate * 100
If you will get the result like above, you will be able to adjust the calibration value in the pump.
Error rate: (Set Flow rate - Measure Flow rate) / Set Flow rate * 100
Measure Flow
After Calibration Flow
NO rat/volume(ml/ Error rate
(CLAC)Value rate(ml/h)
h,ml)
1 -14.0 100 100 0.00%
Average 0.00%
And you should check out the accuracy of DI-2000 once again.