Download as docx, pdf, or txt
Download as docx, pdf, or txt
You are on page 1of 3

Standard Calibration Procedure

Test Subject: Welding Machines


Procedure #:QSCP-92
Revision #: 1.0
Date: 03/01/2012
Approval: James Cagle

1.0 Purpose

1.1 To establish a standard procedure for calibration of Welding Machines

2.0 Requirements

2.1 The calibration laboratory shall be furnished with all items of equipment (including
reference materials) required for the correct performance of calibrations and tests. In
those cases where the laboratory needs to use equipment outside its permanent control, it
shall ensure that the relevant requirements of this section are met.

2.2 All measuring and testing equipment having an effect on the accuracy or validity of
calibrations or tests shall be calibrated and/or verified before being put into service. The
laboratory shall have an established program for the calibration and verification of its
measuring and test equipment. The program will ensure the recall or removal from
service of any standard or equipment that has exceeded its calibration interval or is
otherwise judged to be unreliable.

2.3 The laboratory shall use appropriate methods and procedures for all calibrations and tests
and related activities within its responsibility (including sampling, handling, transport and
storage, preparation of items, estimation of uncertainty of measurement and analysis of
calibration and/or test data). They shall be consistent with the accuracy required and with
any standard specifications relevant to the calibrations or tests concerned.

2.4 The laboratory shall have documented instructions on the use and operation of all
relevant equipment, on the handling and preparation of items, and for calibration and/or
testing, where the absence of such instructions could jeopardize the calibrations or tests.
All instructions, standards, manuals and reference data relevant to the work of the
laboratory shall be maintained up-to-date and be readily available to the staff.

3.0 Referenced Documents

3.1 ISO 17025

3.2 NIST Handbook 150

3.3 NIST Technical Note 1297

3.4 ANSI/NCSL Z540-1

3.5 MILLER- Welding Equipment Meter Calibration

4.0 Materials Required

4.1 LBP-350- Miller Load Bank

Page 1 of 3
QualiCal, LLC Quality Management System_Calibration Procedures
Standard Calibration Procedure
Test Subject: Welding Machines
Procedure #:QSCP-92
Revision #: 1.0
Date: 03/01/2012
Approval: James Cagle

4.2 Fluke 8808A 5.5 Digit MultiMeter

4.3 600A Shunt

4.4 Fluke Software

4.5 Leads

5.0 Tolerance / Measurement Range / Measurement Uncertainties

5.1 Tolerance will be + 10%.

5.2 Measurements will be taken at regular intervals Min to Max scale

5.3 If the tolerance is specified by the customer or if there is some functional characteristics
that require tolerances other than specified above, these considerations should govern the
acceptability of the results.

NOTE: where applicable, the methodology of the Guide to Expression of Uncertainty in


Measurement 1993 shall be used as the basis for expression of uncertainty of the
measurement. Where detailed procedures are not used to quantify and combine
uncertainties (i.e. use of test accuracy ratio concepts) the sources of uncertainty shall be
tabulated and demonstrated to be acceptable for the measurement undertaken.

7.0 Procedure

A. Voltmeter Certification
1. Voltage readings shall be taken at the output studs of unit under test. Voltage readings
shall be taken at open circuit voltage with the Fluke 8808A.
2. Verify that welder voltmeter reading is within + or - 10% of standard meter reading. (If
adjustment is required, see technical manual for each make and model.)
3. Take readings at 10 volt intervals from min to max open circuit voltage.
4. Ensure readings are recorded on Fluke View Forms Log.

B. Ammeter Certification
1. Put certified shunt in series with Miller Load Bank #043329 and the output of unit under
test.
2. Monitor output of shunt with Fluke 45.
3. Verify that welder ammeter reading is within + or - 10% of standard meter reading. (If
adjustment is required, see technical manual for each make and model.)
4. Take readings at 100 amp intervals from 100 amps to max amperage output of unit under
test.

Page 2 of 3
QualiCal, LLC Quality Management System_Calibration Procedures
Standard Calibration Procedure
Test Subject: Welding Machines
Procedure #:QSCP-92
Revision #: 1.0
Date: 03/01/2012
Approval: James Cagle

10.0 Records

10.1 The records for item requiring calibration shall include an individual record of calibration
or verification, or other means of control. This record shall provide a description or
identification of the item, calibration interval, date calibrated, identification of the
calibration source, calibration results (data and/or condition status) and calibration action
taken (adjusted, repaired, new value assigned, de-rated, etc.).

11.0 Calibration Status

11.1 Item shall be labeled to indicate calibration or verification status. The label shall identify
specific date calibrated (date, month, year, Julian date, or equivalent) and the specific
calibration due date or usage equivalent. Items not calibrated to their full capability or
which have other limitations of use, shall be labeled or otherwise identified as to the
limitations. When it is impractical to apply a label directly to an item, the label may be
affixed to the instrument container or some other suitable means may be used to reflect
calibration status. Tamper resistant seals are affixed to operator accessible controls or
adjustments, which if moved will invalidate the calibration. The quality system shall
provide instructions for disposition of equipment with broken tamper resistant seals.

12.0 Storage and Handling

12.1 Items shall be handled, stored, and transported in a manner which shall not adversely
affect the calibration or condition of the equipment.

Page 3 of 3
QualiCal, LLC Quality Management System_Calibration Procedures

You might also like