Professional Documents
Culture Documents
980504B Multi Variable FT
980504B Multi Variable FT
Contents
User Manual Reference Scope ................................................................................................................... 1
This technical bulletin Abstract ............................................................................................................... 1
complements the information
contained in the User Improving Accuracy and Performance............................................................. 2
Manual, applicable to all Multiple Sensors and Parameters ................................................................................. 2
revision .74+.
Scan Interval ................................................................................................................. 2
Time Lag ....................................................................................................................... 2
Multivariable Flow
Transmitters Transferring Flow Rate and Totalizer Data ...................................................... 3
These are a special type of Serial Data Communications ............................................................................. 4
smart digital instrumentation The Flowmeter Device as a Communication Slave ....................................................... 4
device that incorporates
multiple sensors. The sensors The Flowmeter Device as a Communication Pseudo Master ........................................ 4
are controlled by The Flowmeter Device as a Full Communication Master .............................................. 4
microprocessors. Point-to-Point Configurations......................................................................................... 5
Advantages and Disadvantages of a Serial Data Link................................................... 6
Direct Pulse Train ............................................................................................... 7
Figures
Figure. 1. Point-to-Point Wiring Configuration.................................................5
Scope
This technical bulletin applies to all firmware revisions versions .74+ of OMNI
6000/OMNI 3000 Flow Computers.
Abstract
The term ‘multivariable flow transmitter’ describes a class of smart digital
instrumentation devices. This class of device incorporates multiple sensors
controlled by either one or more microprocessors. Coriolis and ultrasonic liquid
and gas flowmeters are examples of current multivariable transmitter
technology. These devices use some form of serial data communication link to
transfer data to and from the OMNI flow computer, requiring an ‘SV’
multivariable communication combo module. In addition, the devices provide an
output pulse train which is proportional to the flow (either mass or volume).
52-0003-0004 Rev B 1
OMNI 6000 / OMNI 3000 Flow Computers Technical Bulletin
Scan Interval
All microprocessor controlled multivariable flowmeter devices operate on some
scan interval; i.e., input parameters are measured on a scan interval (fixed or
variable). The measured parameters are then input into a calculation sequence
which produces a resultant flow rate, (either mass / unit time or volume / unit
time).
Time Lag
Note that sensor measurements must be taken before a result can be
calculated. The calculated flow rate represents that which existed for the
previous scan interval; i.e., there is a time lag from when the measurement is
taken to when the flow rate is calculated. Fast scan intervals are typically used
to minimize any uncertainty that may be introduced by this time lag.
2 52-0003-0004 Rev B
TB-980504B MV Flow Transmitter Interfaces: Connectivity and Data Transfer Issues
52-0003-0004 Rev B
3
OMNI 6000 / OMNI 3000 Flow Computers Technical Bulletin
4 52-0003-0004 Rev B
TB-980504B MV Flow Transmitter Interfaces: Connectivity and Data Transfer Issues
Point-to-Point Configurations
Point-to-point configurations (Figure. 1), with both the flow computer and
transmitter terminated, are the only acceptable wiring configuration. Each
flowmeter transmitter is connected to an independent ‘SV’ port of the flow
computer. Because each ‘SV’ port is connected to only one flowmeter
transmitter, each transmitter can now use a default communication address of
‘1’, greatly simplifying flow transmitter replacement issues. Data transfers are
much faster than in a multi-drop mode and the likelihood of transmitter
interaction is greatly minimized.
Omni Flow
Computer
(Using
Independent
SV Ports)
Modbus IDs of Multivariable All Four Multivariable
Devices can be the same in this Devices must be
Point-to-Point Configuration Terminated
52-0003-0004 Rev B
5
OMNI 6000 / OMNI 3000 Flow Computers Technical Bulletin
6 52-0003-0004 Rev B
TB-980504B MV Flow Transmitter Interfaces: Connectivity and Data Transfer Issues
Disadvantages of totalizing the flow in the flow computer via direct pulse train
are:
• Two extra wires are needed to transmit the pulse signal.
• Need to configure (scale) the digital I/O point pulse train frequency.
52-0003-0004 Rev B
7