6. Specifications
1. Recommendation for an efficient operation
1) Please try to keep in ‘Recharging’ after using as the below table.
Exposing Value
No Exposing Times
(Sec.)
1 0.06 Max. 24 times
2 0.10 Max. 16 times
3 0.16 Max. 10 times
4 0.19 Max. 8 times
5 0.24 Max. 6 times
6 0.29 Max. 5 times
7 0.50 Max. 2 times
8 1.00 Max. 1 times
9 1.50 Max. 1 times
2) Please try to keep on the provided ‘Cradle’ for ‘Re-charging’ when it be
'Non-usage’.
3) Under malfunction, please try to keep press ‘Power Button’ for 4(four) sec.,
and Power on again.
2) Precaution
- Please try to keep away the device from ‘Water and Moisture’. An
unexpected or unknown damage would be able to affect a device.
- Please do not use the current device at near flammable materials.
- Please do not remove any plastic cover without a supplier confirmation.
- Please do not modify and repair before a supplier confirmation.
Because, an unexpected issue would be able to affect a patient and
user.
- Please use a lead apron as a mandatory rule to protect a patient from
an unnecessary radiation’s circumstance.
- Under a patient especially a pregnant and child examination case,
please try to explain them before X-Ray exposing.
3. Components and Parts Name
1) Components
No Photo Item Name
1 Main Frame
Cradle for
3
charging
2) Parts Name
Scroll
Click Wheel
No Content Description
1 Click Adjust an exposing mode by one click
2 Scroll Adjust an exposing value by one scroll
2) Display Panel
No Display Description
1 Display icon for Charging and Waiting mode
2 Display icon under charging
3 Display icon during X-Ray exposing
4 Display icon for exposing times
5 Display icon for cooling time
6 Display icon to be selected ‘Adult and Child’
Display icon for exposing value(sec.), and
7
‘Charging Status’through a letter sign
4) Battery’s Status
No Display Description
1 Display icon under ‘Low Batter’
2 Display icon under ‘Charging’
3 Display icon under ‘Ready’
Blink Scroll
Click
- Bisecting radiography :
A technique to radiography while the examinee(Patient) holds ‘Intra-
Oral Sensor’ or ‘Film’ in place with his/her finger. The X-Ray source is
directed perpendicularly towards an imaginary line, which bisects the
angle between plane and the long axis of the teeth.
Exposure Switch
7) Standard ‘Exposing Value(Sec.)’ and ‘Exposing Times’ by mode
- APossible maximum examination per minute : Duty ratio 30:1
Exposing Value
No Exposing Times
(Sec.)
1 0.06 Max. 24 times
2 0.10 Max. 16 times
3 0.16 Max. 10 times
4 0.19 Max. 8 times
5 0.24 Max. 6 times
6 0.29 Max. 5 times
7 0.50 Max. 2 times
8 1.00 Max. 1 times
9 1.50 Max. 1 times
Maxilla: +45o
Incisor 70 2 0.06
Mandible: -25o
Maxilla: +45o
Canine 70 2 0.09
Mandible: -20o
Child
Molar/ Maxilla: +30o
70 2 0.12
Premolar Mandible: -5o
Maxilla: +45o
Canine 70 2 0.39
Mandible: -20o
Adult
Molar/ Maxilla: +30o
70 2 0.42
Premolar Mandible: -5o
Maxilla: +45o
Incisor 70 2 0.28
Mandible: -25o
Maxilla: +45o
Canine 70 2 0.31
Mandible: -20o
Child
Molar/ Maxilla: +30o
70 2 0.34
Premolar Mandible: -5o
Maxilla: +45o
Canine 70 2 1.00
Mandible: -20o
Adult
Molar/ Maxilla: +30o
70 2 1.30
Premolar Mandible: -5o
Maxilla: +45o
Incisor 70 2 0.56
Mandible: -25o
Maxilla: +45o
Canine 70 2 0.80
Mandible: -20o
Child
Molar/ Maxilla: +30o
70 2 1.10
Premolar Mandible: -5o
Blink
5 seconds
5. Storage and Cleaning
1) Storage and Using condition
- All provided item should be placed in a suitable place to avoid ‘High
humidity and direct sunlight and etc’.
- Please try to store all provided items separately under non-using
cases.
2) Cleaning
- When cleaning the product, turn off the product and disconnect the
charge cradle and charger from the power source.
- Wipe the surface of the product with a soft cloth dampened with non-
corrosive alcohol or disinfectant.
- Be careful that the solution used for cleaning does not enter the
inside of the product, the charger and the charging cradle.
- Don’t use spray type cleaner as it may cause fire and danger when
using spray typeCleaning solution or disinfectant.
6. Product Specifications
Items Description
Model NR-F100
Factory
248-20. Mojeon 1-gil, Mojeon-ri
Seobuk-gu, Cheonan-si,
Chungcheongnam-do
Republic of Korea
NANORAY
Patent Number 10-1684400
Title of the invention Portable CNT and Filament type X-Ray Device