USERMANUAL

Ver.1.1.1_Rev.Oct.2017

TheRay (Model : NR-F100)


Contents

1. Recommendation for an efficient operation

2. Guidelines and Precautions

3. Components and Parts Name

4. Instruction for Use

5. Storage and Cleaning

6. Specifications
1. Recommendation for an efficient operation
1) Please try to keep in ‘Recharging’ after using as the below table.
Exposing Value
No Exposing Times
(Sec.)
1 0.06 Max. 24 times
2 0.10 Max. 16 times
3 0.16 Max. 10 times
4 0.19 Max. 8 times
5 0.24 Max. 6 times
6 0.29 Max. 5 times
7 0.50 Max. 2 times
8 1.00 Max. 1 times
9 1.50 Max. 1 times
2) Please try to keep on the provided ‘Cradle’ for ‘Re-charging’ when it be
'Non-usage’.
3) Under malfunction, please try to keep press ‘Power Button’ for 4(four) sec.,
and Power on again.

2. Guidelines and Precautions


1) For safety usage
- As a recommendation, the current device is being recommendedfor
recharging more than the previous exposing time.
- The current device is designed and manufactured for maximum usage
in safety.
Therefore, for an operation and maintenance cases, please try to keep
in our guidelines for safety usage.
- The current device should be operated by a qualified person.
- Please try to use the current device at near a fire extinguisher to avoid
any unexpected case under a local regulation.
- For more safety, please try to power off under cleaning.

2) Precaution
- Please try to keep away the device from ‘Water and Moisture’. An
unexpected or unknown damage would be able to affect a device.
- Please do not use the current device at near flammable materials.
- Please do not remove any plastic cover without a supplier confirmation.
- Please do not modify and repair before a supplier confirmation.
Because, an unexpected issue would be able to affect a patient and
user.
- Please use a lead apron as a mandatory rule to protect a patient from
an unnecessary radiation’s circumstance.
- Under a patient especially a pregnant and child examination case,
please try to explain them before X-Ray exposing.
3. Components and Parts Name
1) Components
No Photo Item Name

1 Main Frame

Adapter with 220


2 voltage
and Power Cable

Cradle for
3
charging

2) Parts Name

No Part Name Description


1 Display Panel Display LCD panel about a selected status
Change an exposing mode through ‘control
2 Control Wheel
wheel’, and provide ‘Power on/off function’
3 X-Ray Cone Radiation aperture
Exposure
4 A button for X-Ray exposing
Switch
5 Handle Frame Hand's grab point for X-Ray exposing
4. Introduction for use
1) Control wheel
(1) Power On/Off
No Content Description
1 Power On Press ‘Control Wheel’ for 1(one) second
2 Power Off Press ‘Control Wheel’ for 4(four) seconds
(2) Adjustment

Scroll
Click Wheel

No Content Description
1 Click Adjust an exposing mode by one click
2 Scroll Adjust an exposing value by one scroll
2) Display Panel

No Display Description
1 Display icon for Charging and Waiting mode
2 Display icon under charging
3 Display icon during X-Ray exposing
4 Display icon for exposing times
5 Display icon for cooling time
6 Display icon to be selected ‘Adult and Child’
Display icon for exposing value(sec.), and
7
‘Charging Status’through a letter sign

8 Display icon to select on each ‘Teeth Mode’


3) Battery’s Display
- Provide a visible battery’s status in ‘Super Capacitor’
The balance is 0 to 25%. The balance is 50 to 75%.
The balance is 25 to 50%. The balance is 75 to 100%.

4) Battery’s Status
No Display Description
1 Display icon under ‘Low Batter’
2 Display icon under ‘Charging’
3 Display icon under ‘Ready’

5) Selection Mode and Adjustment


No Contents Description
‘Bitewing–Mandible Moral–Maxillary Moral –
1 Sequence
Canine–Incisor’
Refer to the below photo to select a patient
2 Patient Type
type

Blink Scroll

Click

6) How to make an optimal ‘Exposure Angel’ to take an image


Target Angle Exposure Time
Tooth
Upper Lower Adult / Child
Incisor 45º -25º 0.13 / 0.1
Canine 45º -20º 0.2 / 0.16
Molar 30º -5º 0.24 – 0.26 / 0.19 – 0.22
Bitewing 5~8º 0.29 / 0.24
- Paralleling radiography :
A technique to radiography by maintaining the X-Ray collimator(CONE)
in parallel with the axis of a teeth using thesupport.

- Bisecting radiography :
A technique to radiography while the examinee(Patient) holds ‘Intra-
Oral Sensor’ or ‘Film’ in place with his/her finger. The X-Ray source is
directed perpendicularly towards an imaginary line, which bisects the
angle between plane and the long axis of the teeth.

- X-Ray should be released after pressing ‘Exposure Switch’

Exposure Switch
7) Standard ‘Exposing Value(Sec.)’ and ‘Exposing Times’ by mode
- APossible maximum examination per minute : Duty ratio 30:1

Exposing Value
No Exposing Times
(Sec.)
1 0.06 Max. 24 times
2 0.10 Max. 16 times
3 0.16 Max. 10 times
4 0.19 Max. 8 times
5 0.24 Max. 6 times
6 0.29 Max. 5 times
7 0.50 Max. 2 times
8 1.00 Max. 1 times
9 1.50 Max. 1 times

8) Standard Exposing Value(Sec.) by each equipment

Exposing Value(Sec.) under 200mm Cone


Mode and Equipment Min. Max.
DR 0.12 0.40
Adult CR 0.36 0.76
FL(Film) 0.76 1.16
DR 0.06 0.34
Child CR 0.28 0.60
FL(Film) 0.56 0.96

- The above table is a tested value internally as a reference to take an


optimal image. Therefore, it will be able to affect a different
circumstance by ‘Patient tooth anatomy’ and ‘Each Manufacturer’
Table of Exposure Time (DR)>
SSD: 200mm
Angle of
Patient Teeth
inclination kV mA Sec.
Maxilla:+45o
Incisor 70 2 0.12
Mandible: -25o
Maxilla: +45o
Canine 70 2 0.15
Mandible: -20o
Adult
Molar/ Maxilla: +30o
70 2 0.18
Premolar Mandible: -5o

Bitewing +5~8o 70 2 0.2

Maxilla: +45o
Incisor 70 2 0.06
Mandible: -25o
Maxilla: +45o
Canine 70 2 0.09
Mandible: -20o
Child
Molar/ Maxilla: +30o
70 2 0.12
Premolar Mandible: -5o

Bitewing +5~8o 70 2 0.14

<Table of Exposure Time (CR)>


SSD: 200mm
Angle of
Patient Teeth
inclination kV mA Sec.
Maxilla: +45o
Incisor 70 2 0.36
Mandible: -25o

Maxilla: +45o
Canine 70 2 0.39
Mandible: -20o
Adult
Molar/ Maxilla: +30o
70 2 0.42
Premolar Mandible: -5o

Bitewing +5~8o 70 2 0.45

Maxilla: +45o
Incisor 70 2 0.28
Mandible: -25o

Maxilla: +45o
Canine 70 2 0.31
Mandible: -20o
Child
Molar/ Maxilla: +30o
70 2 0.34
Premolar Mandible: -5o

Bitewing +5~8o 70 2 0.36


<Table of Exposure Time (Film)>
SSD: 200mm
Angle of
Patient Teeth
inclination kV mA Sec.
Maxilla: +45o
Incisor 70 2 0.76
Mandible: -25o

Maxilla: +45o
Canine 70 2 1.00
Mandible: -20o
Adult
Molar/ Maxilla: +30o
70 2 1.30
Premolar Mandible: -5o

Bitewing +5~8o 70 2 1.50

Maxilla: +45o
Incisor 70 2 0.56
Mandible: -25o

Maxilla: +45o
Canine 70 2 0.80
Mandible: -20o
Child
Molar/ Maxilla: +30o
70 2 1.10
Premolar Mandible: -5o

Bitewing +5~8o 70 2 1.30

9) How to adjust a factory’s setting value


- Automatic Setting :
An adjusted value will be saved automatically when it be power off, and
will be applied when it be power on as an initial value.

- Exposing value’s Initialization


Under ‘Patient type’s selection mode’, when ‘Exposing Switch’ is being
pressed for 5(five) second.It will be returned back to ‘Factory’s setting
value’.

Blink
5 seconds
5. Storage and Cleaning
1) Storage and Using condition
- All provided item should be placed in a suitable place to avoid ‘High
humidity and direct sunlight and etc’.
- Please try to store all provided items separately under non-using
cases.
2) Cleaning
- When cleaning the product, turn off the product and disconnect the
charge cradle and charger from the power source.
- Wipe the surface of the product with a soft cloth dampened with non-
corrosive alcohol or disinfectant.
- Be careful that the solution used for cleaning does not enter the
inside of the product, the charger and the charging cradle.
- Don’t use spray type cleaner as it may cause fire and danger when
using spray typeCleaning solution or disinfectant.

6. Product Specifications

Items Description

Model NR-F100

Total Filtration 1.5 ( 0.8/0.7mm Al) @ 75kV

Tube Voltage 70kV

Tube current 2mA

Focal Spot 0.4mm

Target material Tungsten

Target Solidity 12º

Main Frame : 1.6 kg/3.5 lbs


Weight
Cradle for Charging : 540g / 1.19 lbs
Nanoray Co., Ltd
Office
Medical Innovation Foundation
601-604, 88, Dongnae-ro, Dong-gu,
Daegu, Republic of Korea

Factory
248-20. Mojeon 1-gil, Mojeon-ri
Seobuk-gu, Cheonan-si,
Chungcheongnam-do
Republic of Korea

Tel +82 53 962 4900


Fax +82 53 962 4902
Web : www.nano-ray.com

NANORAY
Patent Number 10-1684400
Title of the invention Portable CNT and Filament type X-Ray Device