Reagent prime

Turn on the instrument.


◆Press the «Stop» button when the startup cycle starts, the error message «Command has been cancelled» is
displayed.
◆Press the «Valid» button.
◆Enter the Technician menu by pressing the «Tech» button.
◆Enter the password: 01hb05 using the virtual keyboard or the external keyboard, if connected.
◆Press the «Valid» button twice.
◆Enter: Maintenance/Service/Adjustment/Prod
Press the «Serv.» tab.
◆Press the «Service cycle» button.
◆Select «All reagents priming» then press the «Start cycle» button.

Startup
When all reagents are primed, press the «Status» button.
◆Press the «Startup» button and wait for the end of the cycle.