1.

Maintenance
1.1. Introduction
1.2. Daily customer maintenance
Only a Startup and a Shutdown are required at the beginning and the end of the day.
Those cycles are described in the User manual.
1.3. Weekly customer maintenance
An overall check for cleanliness of the system is recommended every week.
All traces of blood or reagent have to be wiped off as soon as possible using a piece of cloth and
distilled water.
1.4. Maintenance periodicity
2. Maintenance kit
Yearly maintenance kit (XEC061AS) contents:
Customer maintenance has to be carried out according to the recommended
frequency, after having attended a HORIBA Medical approved customer
training course. The system warranty may be affected if damage occurs after
a non trained technician intervenes or if replaced spare parts and
consumables do not come from a HORIBA Medical approved origin.
Never use solvent or abrasive cleaning material to clean the system.
Number of cycles Maintenance
< 6000/year 1/year
> 6000/year every 6 month
Part numbers Designation Qty
FAA017A O’ring, 30,8x3,6 NIT. 70SH (vacuum syringe O’ring) 1
FAA023A O’ring, 16x1,25 NIT. 70SH (not used on ABX Micros ES60) 1
FAA029A O’ring, 15x1,5 NIT. 70SH (Diluent piston O’ring) 1
FAA036A O’ring, 6x1,5 SILIC. 60SH (Lyse piston O’ring) 2
FAA046A O’ring, 2,75x1,6 VITON (Coaxial O’ring) 2
FAA049A O’ring, 2,2x1 SILIC. 60SH (not used on ABX Micros ES60) 2
FAA053A O’ring, 1,4x1,25 FLUOCARBO. 1
FAA054A O’ring, 0,74x1 FLUOCARBO 2
FAA055A O’ring, 1,07x1,27 FLUOCARBO (Sampling syringe O’ring) 2
GBC236A Diluent piston 1
GBG275A Flat seal, ep = 0.5 (Aperture seal) 4
GBC364A Insulator 1