Lansion Biotechnology Co., Ltd.

Maintenance Manual
Dry Fluorescence Immunoassay Analyzer

Model: LS-1000

(Version 1.0)

1
Catalogue

Chapter One Inspection..........................................................................................4


1.1 Abstract...................................................................................................................4

1.2 Visual Inspection................................................................................................... 4

1.3 Power On Self Test............................................................................................... 4

Chapter Two Troubleshooting................................................................................6


2.1 Abstract...................................................................................................................6

2.2 Software Version Check........................................................................................ 6

2.3 Fault Code Alarm and Inspection........................................................................ 6

2.4 Troubleshooting..................................................................................................... 8

2.4.1 Switch fault................................................................................................... 8

2.4.2 Display screen fault.................................................................................... 8

2.4.3 Power supply fault....................................................................................... 8

2.4.4 Input/output fault........................................................................................ 9

2.4.5 Software upgrade fault............................................................................... 9

2.4.6 Scan fault 9

2.4.7 Step motor fault......................................................................................... 10

Chapter Three Repair and Disassembly............................................................... 11


3.1 Tool 11

3.2 Notes for disassembly and installation............................................................... 11

3.3 Disassembly and Circuit Port................................................................................11

3.3.1 Disassemble rear cover.............................................................................. 11

3.3.2 Disassemble screen and switch keys....................................................... 12

3.3.3 Disassemble mainbard, printer, outer scanner.......................................13

3.3.4 Disassemble power mainboard and battery........................................... 14

3.3.5 Disassemble transfer module..................................................................... 14

2
3.4 Disassembly Procedure.......................................................................................... 15

3.4.1 Install the screen and power key............................................................... 15

3.4.2 Install the drive assembly with the battery and reset keys................... 15

3.4.3 Assemble power board................................................................................. 16

3.4.4 Install mainboard, printer, outer scanner and cable............................... 17

3.4.5 Combine the front and rear covers to complete the installation........... 18

3.5 Motor Maintenance.................................................................................................. 19

3
Chapter One
Inspection

1.1 Abstract
To ensure the long-term stability of the analyzer, maintenance personnel need to check, test and
maintain it regularly.This chapter provides the basic test method for the analyzer. Please ask the
maintenance personnel to check and test the analyzer according to the actual situation.
The test and inspection methods provided in this chapter are mainly used to verify whether the
performance of the analyzer can meet the specifications requirements.In the test, if the test
result fails to meet the requirements, it indicates that some functional module of the analyzer is
broken and needs to be repaired or replaced immediately.If you have any other questions,
please contact our after-sales service department.

1.2 Visual inspection


Visual inspection mainly carries out the comprehensive appearance inspection on the analyzer. If
there is no obvious physical damage to the tester, visual inspection will pass.Specific inspection
contents are as follows:
(1) Whether there is any physical damage to the shell, display screen and buttons of the analyzer.
(2) Whether the external connection cable is worn or not, and whether the connector pin is loose
or distorted.
(3) Whether the external port of the analyzer is loose or the pin is distorted.

1.3 Power On Self Test


The power on test is used to determine whether the analyzer can start up normally. If the
analyzer can start up by following steps, the power on test will pass.
(1) Remove the battery adjustment plate and load the lithium battery into the analyzer; Connect
the battery cable inside the instrument.
(2) Press the power switch to turn on the analyzer.The key is always flickering, which will make a
4
brief motor sound and enter the startup interface.
(3) The startup interface disappears and enters the system's main interface, normal startup is
completed.
(4) Connect the analyzer to the power adapter and power on. At the top right corner of the
display, where the battery power is displayed, a lightning flag is displayed indicating that the
battery is being charged.

5
Chapter Two
Troubleshooting
2.1 Abstract
The chapter classifies the fault components and fault phenomena of the analyzer. When
troubleshooting, please refer to the corresponding fault table, and check, locate and
troubleshoot the fault in order.
The recommended solutions provided in this chapter can solve most of the equipment failures
you encounter, but do not include any possible problems. If any failure is not included in this
chapter, please contact our after-sales service department.

2.2 Software Version Check


Some troubleshooting may involve software version compatibility. At this time, you need to
know the analyzer configuration and software version information. For detailed information
about version compatibility, please contact our after-sales service department.Please follow the
following steps to check the software version: click in turn, [Menu]→ [System Settings] →
[Analyzer information] , the software version and hardware version will appear.

2.3 Fault Code Alarm and Inspection


Before troubleshooting, please note whether the analyzer displays error code alarm
information.If yes, please remove the technical alarm firstly, and then troubleshoot the following
to avoid unnecessary equipment disassembly.LS-1000 fault code and related explaination are
shown in the following table:

6
Fault
Code Definition Fault Type Recovery method
code

Unrecoverable
100 Initialization error Restart the analyzer
error

Unrecoverable
101 AD Initialization error Restart the analyzer
error

Scan time out (sample code Unrecoverable


102 Rescan the code
error) error

Inner scan time out (item Unrecoverable


103 Rescan the code
batch number error) error

Enter the item management


No parameter on item (not Unrecoverable
104 interface to add the
read item parameter ) error
corresponding item and batch
Read the wrong item
parameter or unscanned Unrecoverable Check the code with no issue and
105
(wrong or unscanned item error scan the code again
code )

Reagent not inserted into the Unrecoverable Insert the reagent to the
200
detection place error detection place and retest

Abnormal reagent (cannot Unrecoverable Confirm whether the sample is


201
detect C line) error added to the reagent and retest

Confirm whether the date setting


Unrecoverable of the analyzer is wrong, and
202 Reagent expire
error replace a batch of reagent that
are not expired after confirmation

The currently measured Unrecoverable


203 Take QC card to retest
reagent is not a QC reagent error

7
2.4 Troubleshooting
2.4.1 Switch fault

Phenomenon Possible cause Troubleshooting


Battery low Check that the battery is fully charged

Battery connection problem Check the battery to make sure the analyzer
is in good connection to the battery

Cannot switch Power protection Abnormal power supply, check whether the
on (no power power supply voltage meets the
adapter used) requirements
Power panel breakdown Replace power board

Mainboard breakdown Replace mainboard

Battery breakdown Remove the battery and connect the power


adapter to switch on

Power key breakdown Replace power key

2.4.2 Display screen fault

Phenomenon Possible cause Troubleshooting


Screen displays Connection cable breakdown Check whether the cable and the interface to
abnormally, the screen are damaged or not connected
others run properly
normally
Display screen breakdown Replace display screen

Mainboard breakdown Replace mainboard

2.4.3 Power supply fault

Phenomenon Possible cause Troubleshooting


Battery cannot Power board breakdown Replace power board
be fully charged
Power adapter breakdown Replace power adapter

Battery cannot Battery breakdown Replace battery


be charged
Power board breakdown Replace power board

8
2.4.4 Input/output fault

Phenomenon Possible cause Troubleshooting

Cannot perform Innter scanner fault Replace innter scanner


the test
Mainboard breakdown Replace mainboard

Printer cannot The printer's wire cannot Replace the printer wire or reconnect the wire
print connect to the mainboard
tightly
printing paper in reverse Change the direction of the printing paper

Mianboard breakdown Replace mainboard

2.4.5 Software upgrade fault

Phenomenon Possible cause Troubleshooting


Connection Fault USB port fault

Cannot software Wrong program upgrade Choose the right program upgrade package
upgrade package

2.4.6 Scan fault

Phenomenon Possible cause Troubleshooting


Outer scanner breakdown Replace outer scanner

Mainboard breakdown Replace mainboard


Cannot scan
Connection cable damage Replace connection cable

9
2.4.7 Step motor fault

Phenomenon Possible cause Troubleshooting

The step motor of the Step motor fault Replace step motor
drive component is
unable to move or
sound abnormal The stepper motor is not on the same Adjust the level of step motor
level as the card carrier and card carrier

Cannot insert reagent Step motor damage Replace step motor


card

The reagent card was taken by mistake Use reagent card with same
(the reagent card was pushed to the batch
internal scan code and was
automatically withdrawn)

10
Chapter Three
Repair and Disassembly
3.1 Tool
You may need the following tools during the disassembling and replacement of spare parts: cross
screwdriver, hexagon screwdriver, oblique nose pliers, tweezers, etc.

3.2 Notes for disassembly and installation


(1) Disconnect the power adapter and shut down before disassembling the tester.
(2) When reinstallation, please pay attention to the line sequence and insert the connecting line,
and place the connecting line well to avoid short circuit caused by the crushing of the connecting
line.
(3) When reinstallation, choose the screw suitable for the model.If the screw type is not correctly
selected and forced into, the equipment may be damaged;And during use after reinstallation,
screws or components may come off and cause unpredictable product damage.
(4) Please disassemble the equipment in a certain order. If the order is wrong and the equipment
is forcibly disassembled, it may cause irreversible damage to the equipment.
(5) Before disassembling the components, please make sure that all connection wires have been
removed. Please take care to avoid pulling the connection wire or damaging the connection head.
(6) Please sort the disassembled screws and other parts so that they can be taken away when
reinstalled and avoid falling, pollution or loss.

3.3 Disassembly and Circuit Port


3.3.1 Disassemble rear cover
Remove the six screws at the red marked position as shown in figure 3.3.1 with a cross
screwdriver. After removing the screws, remove the flexible row of connecting screen and
mainboard. Remove the connecting line of switch machine key from the interface of mainboard
to separate the front and rear cover.

11
Figure 3.3.1

3.3.2 Disassemble screen and switch keys


Remove the screen by removing the four screws marked red with the cross screwdriver.Remove
the four yellow cross screws with the cross screwdriver to remove the switch keys.Figure 3.3.2
shows:

Figure 3.3.2

12
3.3.3 Disassemble mainbard, printer, outer scanner
Remove the flat line marked yellow to remove the outer scan code.Remove the green label
printer cable to remove the printer.Remove the connecting lines of each interface on the
mainboard successively, and then remove the four cross screws marked in red to remove the
mainboard, as shown in figure 3.3.3

Figure 3.3.3

(1)Separate the dismantled printer and the connecting cable as shown in figure 3.3.3.1
(2)Separate the removed outer scanner and the connected flat wire as shown in figure 3.3.3.2

Figure 3.3.3.1 Figure 3.3.3.2

13
3.3.4 Disassemble power mainboard and battery
Remove the yellow-marked power connection cable first, and then remove the battery cover of
the rear case, then remove the battery.Remove the supporting screws of the four red marked
mainboards, and then remove the network port board from the power supply board to remove
the power supply board.As shown in figure 3.3.4

Figure 3.3.4

3.3.5 Disassemble transfer module


Remove rear cover to remove the transfer module. As shown in figure 3.3.5

Step
Motor

Inner
Scanner

Inner
Scanner

Figure 3.3.5

14
3.4 Disassembly Procedure
3.4.1 Install the screen and power key
Place the display in the black area, note the direction of the screen, and secure it with four cross
screws.Put the switch into the red mark position, pay attention to the direction of the switch key,
and fix it with four cross screws.Connect one end of the flat wire to the display screen and one
end of the four-pin cable to the switch button.As shown in figure 3.4.1:

Figure 3.4.1

3.4.2 Install the drive assembly with the battery and reset keys
Put the battery into the rear cover and install the battery cover.Fix the reset key with four cross
screws and connect the connecting line of the two pins, and fix the step motor with six cross
screws at the red mark on the rear shell. As shown in figure 3.4.2:

15
Reset key
connection line

Power line

Step motor
connection line

Scanner
connection line

Figure 3.4.2

3.4.3 Assemble power board


Fix the power supply board to the rear cover through four support screws, as shown infigure
3.4.3

Figure 3.4.3

16
3.4.4 Install mainboard, printer, outer scanner and cable
Firstly, the mainboard is fixed on the support screw by four cross screws as shown in figure 3.4.4,
and then the printer and external scanner are installed.
(1) The printer is connected with the wire of five plugs and four pins. The printer is connected
with five pins and the mainboard is connected with four pins, as shown in the green mark.
(2) Connect one end of the flat line, and then install the outer scanner on the rear cover. Connect
the other end of the flat line to the mainboard at the yellow mark as shown in the figure. Refer
to the wiring diagram of the mainboard and the power supply board in 3.4. As shown in figure
3.4.4

17
Figure 3.4.4

3.4.5 Combine the front and rear covers to complete the installation

Connect the screen to the mainboard with soft platen firstly, and then connect the switch
machine key to the mainboard and merge the front and rear cover. Fix with six cross screws at
the marked red position as shown in figure 3.4.5 to complete the assembly.

18
F
Figure 3.4.5

3.5 Motor Maintenance


The motor is maintained for about six months. After the transmission component is removed,
the metal shaft and metal guide rail of the analyzer motor are coated with lubricant (vaseline).In
red mark locations as shown in figure 2.1 and 2.2, put on your glove to take appropriate lubricant
and evenly apply it on the metal shaft and metal guide rail.

Motor screw

Partial Enlarged
Drawing

Figure 3.5.1

19
Metal rail
Partial Enlarged
Drawing

Figure 3.5.2

20