Professional Documents
Culture Documents
ILYTEAutospl Serv Manu R2
ILYTEAutospl Serv Manu R2
Na+/K+
Analyzer
Na+/K+/Cl-
Analyzer
Na+/K+/Li+
Analyzer
Autosampler
Service Manual
Service Manual Contents
Product Description
Troubleshooting
Repair
Repair Overview........................................................................................................................33
Tools ..........................................................................................................................................34
Rear Housing Removal ............................................................................................................35
Rear Housing Installation ..........................................................................................................37
Power Supply Removal ............................................................................................................39
Power Supply Installation..........................................................................................................41
Printer Removal ........................................................................................................................43
Printer Installation ....................................................................................................................47
CPU Board Removal ................................................................................................................50
CPU Board Installation..............................................................................................................54
Display Removal ......................................................................................................................57
Display Installation ..................................................................................................................59
Pump Motor Removal ..............................................................................................................62
Pump Motor Installation ............................................................................................................63
Probe Motor Assembly Removal ..............................................................................................64
Probe Motor Assembly Installation............................................................................................67
Guide Rod Removal..................................................................................................................73
Guide Rod Installation ..............................................................................................................75
Electrode Jack Removal ..........................................................................................................76
Electrode Jack Installation ........................................................................................................77
Autosampler
Specifications
Appendix
Display Messages......................................................................................................................111
QC TEST? Flowcharts ..............................................................................................................118
Software Flowcharts ................................................................................................................120
Schematics
CPU Board
CPU Board – Analyzer Series below 6100
Power Supply – Analyzer Series below 6100
Transformer Assembly
AutoSampler PCB
Product Description 1
Product Overview
The ILyte analyzer is a direct measurement ion-selective electrode (ISE) system which uses
Sodium (Na), Chloride (Cl), Potassium (K), and Lithium (Li) sensitive components to measure
electrolytes in either whole blood, serum, plasma or urine.
The analyzer software allows the operator to perform sample analysis, analyzer diagnostics and
maintenance by simply pressing the "YES" or "NO" buttons in response to prompts displayed on
the LCD screen. The ILyte analyzer performs the analysis both quickly and cost effectively.
Analysis takes only one minute and utilizes approximately one hundred microliters of whole
blood, serum or plasma; or four hundred microliters of diluted urine.
The analyzer can re-calibrate automatically every four hours and can perform emergency (STAT)
analysis on demand. Results are always displayed on the LCD screen and can also be docu-
mented on paper by an integrated printer.
The analyzer contains a patented reagent system including a modular reagent pack. This pack
provides a minimum of two months supply of solution and also serves as a waste receptacle for
solution and samples after analysis.
System Overview
The ILyte analyzer is engineered to make local service feasible. Modular construction is used
throughout the design. The basic operating system consists of a motor which drives a peristaltic
pump. The peristaltic pump pulls the desired fluid through the system, disposing waste into a
designated compartment inside the solutions pack. A second motor drives a lead screw which in
turn drives a mechanical link-arm assembly. This link-arm assembly moves the sample probe
vertically allowing the probe inside the solutions valve to aspirate the appropriate fluid for pump-
ing: sample; standards; or air. The sample detector in the fluid line positions the fluids.
Please note that mechanical and electrical assemblies can be replaced. Follow the instructions in
the TROUBLE SHOOTING SECTION to determine if assembly replacement is necessary. Refer
to the Replacement List in this section which lists replaceable assemblies, components, and
appropriate catalog numbers.
Table of Symbols
6100
ILYTE ANALYZER’S Information for ILyte analyzers with serial numbers
BELOW 6100 below 6100.
CPU Assembly contains the CPU board with digital components and analog sensor inputs
Firmware (EPROM) contains the software
Interface Ports for Autosampler and RS-232 for serial devices
Probe Arm holds and moves the probe to aspirate samples and calibrants
Guide Rod allows probe arm to move vertically
Printer Assembly
Front Housing
Display
Assembly
Electrode Jack(s)
Printer Cable
Sample
Detector Cable
Display Cable
P6
00XXX RY
Firmware
(EPROM)
Probe Position
Sensor
Pump Motor
Cable AUTOSAMPLER
Interface Ports
Sensor
Cable
Power Supply
Cable
P3 P1 P2 P4 P13
Probe Motor
Cable
Printer Cable
Voltage Selector
Switch
Entry Module
Fuse Bay
Front Housing
Probe
Position Scale
Probe
Motor
Assembly
Lead Screw
Probe
Motor
Mounting
Bracket
Front Housing
Display Cable
Pump Head
Assembly
Pump
Assembly
Printer
Cable
Pump Motor
This section describes the main assemblies and components of the ILyte front assembly, analyz-
er series below 6100. Refer to Figures P4, P5, P6 for the location of each component.
CPU Assembly contains the CPU board with digital components and analog sensor inputs
Firmware (EPROM) contains the software
Interface Ports for Autosampler and RS-232 for serial devices
Probe Arm holds and moves the probe to aspirate samples and calibrants
Guide Rod allows the probe arm to move vertically
Printer Assembly
Front Housing
Printer Cable
Sample
Detector Cable
Electrode Jack(s) Display Cable
P6 Display
Assembly
00XXX RY Firmware
(EPROM)
Probe Position
Sensor
Pump Motor
Cable AUTOSAMPLER
Interface Ports
Sensor
Cable
Power Supply
Cable
P3 P1 P2 P4 P13
Probe Motor
Cable
Printer Cable
Entry Module
Voltage Selector
Switch Fuse Bay
Front Housing
Printer
Assembly
Guide Rod
Probe
Position Scale
Probe
Motor
Assembly
Lead Screw
Probe
Motor
Mounting
Bracket
Voltage Selector
Switch
Front Housing
Pump Head
Assembly
Pump
Printer Assembly
Cable
Pump Motor
Sample Detector
Fill Plug
Potassium Electrode Internal Filling
(Red Dot) Solution
Sodium Electrode
(Blue Dot)
Chloride Electrode
(Green Dot) or
Lithium Electrode
(Yellow Dot)
Reference Electrode
Membrane Assembly
Rear Housing
Label with
Serial Number
Autosampler
Serial Port
Port
Voltage
Power Cord Selector
Entry Module Switch
6100 On ILyte Analyzers with serial numbers below 6100 the voltage selector is located inside
the front housing on the power supply.
Product Description 13
Replacement List
Consumable and assemblies items can be purchased directly from IL. When ordering these
items in the U.S. replace the prefix “05” with the prefix “ME“.
CONSUMABLE ITEMS
Description Catalog No.
Na Electrode 05002101
K Electrode 05002102
Cl Electrode 05002113
Li Electrode 05002106
Reference Electrode 05002103
Sample Detector 05002257
Sample Probe 05002107
Solutions Valve 05002108
Fill Plug 05000195
Membrane Assembly 05002258
Tubing Kit 05002104
ILyte Operator's Manual 05002277
ASSEMBLIES
Description Part No.
Electrode Housing Na/K/Cl 05000885
Electrode Housing Na/K/Li 05000985
Electrode Housing Na/K 05000177
Power Supply 05001676
with CPU 05001937-001 (Accepts 800ml Sol. Pack only)
with CPU 05001937-002 (Accepts 400ml and 800ml Sol. Pack)
Power Supply 05001065 6100
with CPU 05001063
Display, 1 Line 05000186
Display, 2 Line 05001901
Pump 05001067
Probe Motor 05001392
Probe Position Sensor 05001306
Motor Driver 05001148
(CPU, 05001937,
positions U35, U36)
(CPU, 05001063,
positions U22, U23)
Printer 05001296
Electrode Jack 05000173
ILyte Service Kit 05001950-001(Accepts 800ml Sol. Pack only)
Includes: 05001950-002 (Accepts 400ml and 800ml Sol. Pack)
Power Supply (05001676)
CPU (05001937)
Service Kit Installation Label
6100 Please note that ILyte Analyzers with serial Numbers below 6100 are assembled using Power
Supply, 05001065 and CPU, 05001063. These two parts cannot be mixed with
Power Supply, 05001676 and CPU, 05001937. To replace either Power Supply, 05001065
or CPU 05001063, use ILyte Service Kit 05001950.
14 Product Description
Troubleshooting 15
Troubleshooting Overview
This section describes troubleshooting the mechanical and electrical functions of the ILyte ana-
lyzer. This section is based on the principle of replacing suspected faulty assemblies with new
assemblies. IL recommends keeping an inventory of functioning test assemblies for troubleshoot-
ing the ILyte. Refer to the Troubleshooting Guide in this section to determine the cause and cor-
rective action that must be taken to properly service the ILyte Analyzer.
6100 Since ILyte Analyzers with serial numbers below 6100 are assembled using Power
Supply, part no. 05001065 and CPU, part no. 05001063 the troubleshooting instructions
differ in certain areas. Please refer to the serial number, located on the back panel of the
analyzer, to identify which instructions to follow.
For information on troubleshooting the ILyte Analyzer flow system and consumable items, consult
the ILyte Operator’s Manual. See the Replacement List in the PRODUCT DESCRIPTION
SECTION of this manual for a complete listing of all consumable items and catalog numbers.
For information on troubleshooting the ILyte Autosampler consult Troubleshooting in the
AUTOSAMPLER SECTION.
16 Troubleshooting
Troubleshooting Guide
Refer to the chart below to help determine the problem and corrective action needed.
AC Power Verification
1. Check that the voltage selector switch is set at the proper voltage (110V or 220V). This volt-
age must match the power outlet voltage.
2. Unplug the power cord from the entry module and plug into the outlet.
3. Check the voltage at hot and neutral at the end of the power cord. Refer to Figure T1.
Neutral Hot
Earth Ground
Figure T1
For ILyte analyzers set to 110V, the voltage at hot should be 90V–125V.
For ILyte analyzers set to 220V, the voltage at hot should be 210V–240V.
If the voltage is incorrect, replace the power cord and return to normal operation. Otherwise,
proceed to STEP 4.
4. Using a small flat-tip screwdriver pry open the fuse bay. Refer to Figure T2 for the location of
fuse bay. Remove and inspect both fuses as follows:
For 110V ILyte analyzers, the fuses must be rated 0.8A, 250 Volts
For 220V ILyte analyzers, the fuses must be rated 0.4A, 250 Volts
P3 P1 P2 P4 P13
Entry Module
Fuse Bay
Figure T2
Visually inspect the wire conductor inside the fuses. If the wire is broken or missing,
replace the fuse.
18 Troubleshooting
5. Remove the rear housing. Refer to Rear Housing Removal in the REPAIR SECTION.
6. Check for short circuits between the test points listed below:
TP12 and TP2
TP12 and TP8
TP12 and TP6
TP12 and TP7
TP12 and TP11
If a short exists, replace the CPU board. Otherwise, return to normal operation.
7. Disconnect the Power Supply cable (6-pin connector) from the CPU board, see Figure T3.
Connect the ILyte entry module to a wall outlet with a known good power cord.
AUTOSAMPLER
Power Supply
Pin 6 –> 1 Cable
P4 P13
Figure T3
8. Connect the reference side of the voltmeter to Pin 1 of the power supply cable (6-pin
connector).
9. Check the AC voltages at Pins 2 and 5. see Figure T3 for the location of the pins. The voltage
for each should be 7.6 ± 2V.
10. Check the voltaage across Pins 3 and 4. The voltage should be 26± 2V.
11. Measure the resistance between Pin 6 and power supply chassis. The resistance should be
zero.
If any of the voltages are incorrect, replace the power supply module and return to normal
operation.
Troubleshooting 19
12. Check for short circuits between the test points below:
TP11 and TP5
6100 TP11 and TP8
TP9 and TP6
TP9 and TP7
If a short exists, replace the CPU Board and return to normal
13. Disconnect the power supply cable (13-pin connector) from the CPU board, see Figure T4.
Power Supply
Pin 13 –> 1 Cable
P4 P13
6100
Figure T4
14. Connect the reference side of the voltmeter to Pin 5 of the powr supply cable. Avoid
shorting the adjacent pins.
15. Check the DC voltages at the pin placements listed below. See Figure T4 for the pin
locations.
If any of the voltages are incorrect, replace the power supply and return to normal operation.
20 Troubleshooting
DC Power Verification
1. Remove the rear housing. See Rear Housing Removal in the REPAIR SECTION for details.
2. Connect the power cord first to an outlet, then to the entry module.
3. Connect the reference side of the voltage meter to TP12 on the CPU board.
6100 5. For ILyte serial numbers below 6100, use TP11 as the reference point. Verify the
voltage at each test point listed below.
6. If any of the voltage values in STEP 4 OR STEP 5 are not within specification, AND
the AC power verification is successful, replace the CPU board.
Troubleshooting 21
Display Verification
1. Disconnect the display cable from the CPU board. See Display Assembly Removal in the
REPAIR SECTION for details.
2. Connect a known good display assembly to the CPU board and connect the main power cord.
Verify that the test display is operative. Based on the test results, take the appropriate
corrective action.
1. Disconnect the power cord, wait 5 seconds, then reinstall the power. The probe arm should
move down and back up. If there are no error messages, resume normal operation. If an
“MTS” error occurs, proceed to STEP 2.
2. Remove the sample probe from the probe arm. For details refer to ILyte Operator’s Manual.
Remove and reinstall the power cord. The probe arm should move down and back up. If there
are no error messages examine and if necessary replace the sample probe, solutions valve
and probe wiper. If an “MTS” error occurs, proceed to STEP 3.
3. If the probe arm does not move at all perform the following operations.
A. Visually inspect that the actuator and retaining ring are correctly installed on the probe
motor lead screw. Refer to Figure R64 in the REPAIR SECTION. If the installation is
incorrect, refer to CPU Removal and Probe Motor/Lead Screw Installation in the
REPAIR SECTION for details.
B. Check that the probe motor wires are completely seated inside the 4-pin connector as
follows.
– Remove the probe motor connector from the bottom left of the CPU Board
(position P3).
– Compress the strain-relief cap against the probe motor connector to firmly seat
the wires.
– Reinstall the probe motor connector into the CPU Board.
C. Repeat STEP 1. If no errors occur, resume normal operation.
4. If the probe arm moves and stops anywhere except the top position, or if the probe arm
vibrates against the top of the ILyte housing, replace the probe position sensor. To replace the
sensor, first remove the CPU Board and the probe position scale. Refer to Probe Motor/Lead
Screw in the REPAIR SECTION for details.
5. If the Probe does not move at all, check for a short circuit on the CPU Board between TP12
and TP2. If there is a short circuit, replace the motor driver at position U36.
6100 For ILyte serial numbers below 6100, check for a short circuit on the CPU Board
between TP8 and TP11. If there is a short circuit, replace the motor driver at position
U23.
6. If no short circuit is found, replace the probe motor assembly. Refer to the Probe Motor
Assembly in the REPAIR SECTION for details.
Troubleshooting 23
1. Enter the QC TEST? function and proceed to the PUMP TEST? Refer to QC Tests in this
section for further instructions.
4. If the pump vibrates or does not run smoothly, ensure that the pump motor wires are properly
seated inside the 6 pin connector as detailed below.
A. Remove the pump motor connector from the bottom left of the CPU board
See Figure P1 in the PRODUCT DESCRIPTION SECTION.
B. Compress the strain-relief cap against the pump motor connector to firmly seat the wires.
5. Repeat the Pump Test. If the pump malfunctions, replace the pump motor with a test motor.
See Pump Motor in the REPAIR SECTION for futher instructions.
6. Repeat the Pump Test with the test motor. Based on the test results, take the appropriate cor-
rective action.
• Motor Operative: Replace the test motor with a new motor. See Pump Motor in the
REPAIR SECTION.
• Motor Malfunction: Check for a short on the CPU board between TP12 and TP2. If there is
a short circuit, replace the pump motor driver at U35.
6100 7. For ILytes with serial numbers below 6100, check for a short circuit between TP8 and
TP11.
• No Short Circuit: Replace the pump motor assembly. See Pump Motor in the
REPAIR SECTION for details.
24 Troubleshooting
Printer Verification
1. Press the printer paper advance button. If the printer head moves proceed to STEP 2. If the
printer head does not move, proceed to STEP 3.
2. Verify that the printer is on. Turn the printer on by pressing NO to the PRINTER OFF? ques-
tion in the USER OPTIONS section.
3. If the printer is still not working, remove the power cord from the analyzer and wait for ten sec-
onds, then install the power cord back into the entry module. Press the printer paper advance
button. The printer head should move across the paper and the paper should advance for-
ward.
4. If the printer is still not working, remove the printer cover and check for any paper obstruction
in the printer head area. Remove any obstructions if possible, without damaging the printer
assembly or paper tear bar. If an obstruction cannot be removed, replace the Printer
Assembly. Refer to Printer Removal and Installation in the REPAIR SECTION for details.
5. If there are no obstructions, but the printer does not work, perform the following steps.
B. Remove the printer cable at the bottom right of the CPU board.
C. Compress the strain-relief cap against the probe motor connector to firmly seat the printer
cable.
6. If the printer still does not print, connect a test printer assembly to the CPU Board. If the test
printer works, install a new printer assembly. If the test printer does not work, replace the CPU
board. Refer to CPU Installation in the REPAIR SECTION for details.
Troubleshooting 25
1. Remove the electrode housing from the analyzer. See the ILyte Operators Manual for further
details.
4. Proceed to the DIAGNOSTICS? menu and perform the SAMPLE DETECTOR TEST?
• With the sample detector installed, the value on the display screen should be less than 200.
• With the test in progress, remove the sample detector. The display value should rapidly
change to 255.
If the sample detector test meets these specifications, resume normal operation.
A value of 255 with the sample detector installed indicates a bad connection. Proceed to
STEP 5 to check the sample detector circuit.
5. Reinstall the sample detector. Look at the left side of the sample detector. A red light must
shine from the LED hole. If there is no red light, install a known good sample detector. If there
is still no red light, replace the Display Assembly. Refer to the REPAIR SECTION for informa-
tion.
6. If the TEST sample detector red light is on, check the four connecting pins on the ORIGINAL
sample detector. The pins must face straight out forming a rectangle. If the pins are bent
against the black plastic body, straighten the pins using a small flat head screwdriver.
1. Remove and inspect electrodes and electrode housing for liquid leakage and salt build-up. All
electrodes and the electrode housing surfaces MUST be completely dry. See ILyte Operator’s
Manual for further information on consumable items.
2. Inspect the electrode pin receptacles on the ILyte analyzer front housing. Remove any liquid
or salt from these areas.
3. Enter * DIAGNOSTICS? * and answer YES to ELECTRODE VALUES? The display most
likely will show “249 249 249”.
The following steps require the use of the Mini-Test High Impedance Voltage Source,
part no. 05001374. If this tool is not available, the remaining tests cannot be
performed.
4. Connect the Voltage Source to the ILyte analyzer as follows. Refer to Figure T5.
A. Connect the three red wires from the Voltage Source to the K+, Na+, and CL-/Li+
electrode receptacles
B. Connect the black wire from the Voltage Source to the reference electrode receptacle.
C. Connect the remaining black wire into the back connector on the Cl-/Li+ red wire.
This step will short out all three electrode channels.
Figure T5
6100 For analyzers with serial numbers below 6100, the electrode specification is
-12 mV to +12 mV.
Troubleshooting 27
6. Disconnect the black wire from the back connector on the Cl-/Li+ red wire.
8. Set the Voltage Source to the group 1 settings listed in the table below. Three new individual
values will appear on the screen. Wait a few seconds for the values to stabilize then verify that
the values are within specification. Press YES to print the values.
9. Set the Voltage Source to the group 2 settings, listed in the table below, and repeat STEP 8.
10. Calculate the DIFFERENCE between the group 1 values and group 2 values. Verify that the
values are within specification.
11. If any of the voltage values are out of specification, examine the electrode jacks for corrosion
and verify the connection between the electrode jack and CPU board. Refer to
Electrode Jack Removal in the REPAIR SECTION. If the electrode jacks are not corroded
and are properly connected, replace the CPU board with a new CPU board. See CPU
Assembly in the REPAIR SECTION for details.
If the technician has any questions about the reliability of the mini-tester, install a new 9
volt battery and repeat the above procedure.
28 Troubleshooting
1. Connect the ILyte analyzer to a device with a known working serial port. Use a known good
ILyte serial cable, part no. 05001334. Make cross connections as required (e.g. 2 to 3;
4 to 5).
Pin assignments
Pin 1 Chassis, Not Connected
Pin 2 TxD
Pin 3 RxD
Pin 4 RTS
Pin 5 CTS
Pin 7 Signal Ground
Transmission Rate 1200 baud
RS-232 Protocols
3. When correctly linked the ILyte will transmit to the external device all of the data that is printed
by the ILyte printer, in the same format.
Troubleshooting 29
QC Tests
The QC TEST? functions are similar to the Service Diagnostics program used with other IL
systems.
1. To access the QC TEST? functions, answer the questions that appear on the ILyte display, as
shown in the chart below.
2. Two numbers appear on the display screen separated by an arrow. A cursor will be posi-
tioned underneath the first digit of the first number.
3. Change the first number to “246.0” as follows. Press NO to change the first digit. When the
digit is correct press YES to advance to the next digit.
5. Press YES when finished. QC TEST? will appear on the screen. If an erroneous value is
input the screen displays CORRECT?. Press NO to return to STEP 2 and correct the
number.
7. Consult the table below which lists all of the QC tests and their functions. To select a test
press YES. To bypass a test press NO.
Repair Overview
Always consult the TROUBLESHOOTING SECTION first to determine the problem before
attempting to repair the analyzer.
Always remove the solutions pack and the electrode housing from the analyzer before
beginning any repair work. Refer to the ILyte Operator’s Manual for detailed instructions.
Follow the assembly and removal instructions in the order outlined in this section unless other-
wise noted. Always unplug the power cord from the analyzer before beginning any repair work.
Note that only complete mechanical and electronic assemblies can be replaced. Refer to the
Replacement List in the PRODUCT DESCRIPTION SECTION which lists replaceable assem-
blies as well as consumable items. Consult the list to determine the appropriate replacement
item and catalog number.
If repair to any electrical assembly is necessary, consult the Schematics in the APPENDIX SEC-
TION for further detail. Service personnel are encouraged to contact IL directly with any prob-
lems or concerns relating to electrical assemblies.
34 Repair
Recommended Tools
The following is a list of the tools required to perform repairs on the ILyte analyzer. Please refer
to this list to determine the appropriate tool and size needed.
Optional Tools
The following is a list of tools recommended to help perform repairs on the ILyte analyzer. Please
refer to this list to determine the tool and part number when ordering.
1. Disconnect the power cord from the entry module, see Figure R1.
2. Detach the four screws from the rear housing using a #2 Phillips screw driver, as shown in
Figure R2.
Figure R1
Screws Screws
Figure R2
36 Repair
Figure R3
Repair 37
1. Align and attach the front and back housing. See Figure R4. Verify that all cable
connectors are attached properly. Refer to Figures P1 and P6 in the PRODUCT
DESCRIPTION SECTION for detail. Verify that all wires are contained within the
front housing before proceeding.
Figure R4
38 Repair
2. Attach the four screws to the rear housing using a #2 Phillips screw driver, as shown in
Figure R5.
3. Connect the power cord to the entry module, see Figure R6.
Screws Screws
Figure R5
Figure R6
Repair 39
1. Disconnect the power supply cable from the CPU board, see Figure R7. Refer to
Figure R8 for ILyte analyzers with serial numbers below 6100.
2. Detach the four screws from the bottom of the front housing, using a #2 Phillips screw driver.
See Figure R9.
Connector to
CPU Board
Power Supply
Cable
Figure R7
Connector to
CPU Board
Power Supply
Cable
6100
Figure R8
Screws
Figure R9
40 Repair
3. Lift and remove the power supply assembly from under the CPU board, as shown in
Figure R10. Refer to Figure R11 for ILyte analyzers with serial numbers below 6100.
Power Supply
Assembly
Figure R10
Power Supply
Assembly
6100
Figure R11
Repair 41
1. Slide the power supply assembly under the CPU board, as shown in Figure R12. Fit into the
alignment holes in the bottom of the front housing. Refer to Figure R13 for ILyte
analyzers with serial numbers below 6100.
2. Attach the four screws through the bottom of the front housing using a #2 Phillips screw
driver. See Figure R14.
Power Supply
Assembly
Figure R12
Power Supply
6100 Assembly
Figure R13
Screws
Figure R14
42 Repair
3. Connect the power supply cable to the CPU board, as shown in Figure R15. Refer to
Figure R16 for ILyte analyzers with serial numbers below 6100.
Connector to
CPU Board
Power Supply
Cable
Figure R15
Connector to
CPU Board
Power Supply
Cable
6100
Figure R16
Repair 43
Printer Removal
1. Remove the printer lid from the printer, see Figure R17.
2. Remove the paper roll from the printer bay, as shown in Figure R18.
Printer Lid
Printer
Figure R17
Paper Roll
Printer Bay
Figure R18
44 Repair
3. Disconnect the printer cable from the CPU board, see Figure R19. Refer to Figure R20 for
ILyte analyzers with serial numbers below 6100.
Connector to
CPU Board
Printer Cable
Figure R19
Connector to
CPU Board
Printer Cable
6100
Figure R20
Repair 45
4. Detach the printer cable from the retaining clamps located along the right side of the CPU
board, as shown in Figure R21.
5. Detach the two screws using a # 2 Phillips screw driver. See Figure R22.
Retaining
Printer Cable Clamps
Figure R21
Screws
Figure R22
46 Repair
7. Pull the printer cable up from behind the CPU board, while holding the printer body as
shown in Figure R24.
Printer Body
Figure R23
Printer Body
Printer Cable
CPU Board
Figure R24
Repair 47
Printer Installation
1. Slide the printer cable behind the CPU board, as shown in Figure R25.
3. Attach the two screws using a #2 Phillips screw driver. See Figure R27.
Printer Body
Printer Cable
CPU Board
Figure R25
Printer Body
Figure R26
Screws
Figure R27
48 Repair
4. Connect the printer cable to the CPU board, see Figure R28. Refer to Figure R29 for ILyte
analyzers with serial numbers below 6100.
5. Using scissors, cut the beginning of the paper roll into a “V” shape. See Figure R30. Insert
the paper into the feeder slot inside the printer bay, and press the paper feed button.
6. Place the printer roll into the printer bay, as shown in Figure R31.
Connector to
CPU Board
Printer Cable
Figure R28
Connector to
CPU Board
Printer Cable
6100
Figure R29
Figure R30
Paper Roll
Printer Bay
Figure 31
Repair 49
7. Place the printer lid onto the printer, see Figure R32.
8. Attach the printer cable to the retaining clamps located along the right side of the CPU
board, as shown in Figure R33.
Printer Lid
Printer
Figure R32
Printer Cable
Retaining Clamps
Figure R33
50 Repair
1. Disconnect the probe motor, sensor, pump motor, power supply and printer cables from the
CPU board. See Figure R34. Refer to Figure R35 for ILyte analyzers with serial numbers
below 6100.
2. Remove the display cable from the top of the CPU board as shown in Figure R36.
Power Supply
Cable
Figure R34
Figure R35
Display Cable
Connector to
CPU Board
CPU Board
Figure R36
Repair 51
3. For ILyte Analyzers with serial numbers below 6100, push the display cable tabs to the side
to unlock. Remove the cable from the connector. See Figure R37.
4. Disconnect the electrode jacks from behind the CPU board. See Figure R38 for the location
of the electrode jacks. To disconnect, gently pull electrode jack straight away from the con-
nector pin of the CPU Board.
The connector pins are extremely delicate. To avoid pin breakage when removing the
electrode jacks, gently pull the electrode jacks straight away from the pins. If a pin breaks
the entire CPU board may require replacement.
Display Cable
6100
Figure R37
Connector Pin
Electrode Jacks to CPU Board
CPU Board
Figure R38
52 Repair
5. Disengage the CPU board from the four plastic mounting posts, located at each corner of the
CPU board. To disengage, press the mounting post tab in and pull the CPU board over the
tab, as shown in Figure R39.
6. Disengage the center plastic mounting post from the CPU board, located above the serial
and autosampler ports.
7. For ILyte analyzers with serial numbers below 6100, detach the screw from the center of
the board using a # 2 Phillips screw driver. See Figure R40.
CPU
Board
Press in Tab
Pull Board
Plastic Over Tab
Mounting Post
Figure R39
Screw
6100
Figure R40
Repair 53
7. Slide the CPU board off the mounting posts. See Figure R41.
Mounting Mounting
Posts Posts
Figure R41
54 Repair
1. Align the CPU board with the plastic mounting posts. Snap the right side mounting posts of
the CPU board into place. See Figure R42.
Do not snap the left or center mounting posts into place at this time.
Mounting Posts
Figure R42
Repair 55
2. Connect the electrode jacks to the back of the CPU board. To connect, gently push the elec-
trode jack straight into the connector pin. See Figure R43 to determine the proper connector
pin position on the CPU board and the recommended installation order.
See Figure R44 for ILyte Analyzer with serial numbers below 6100. On some earlier ILyte
analyzers below 6100, the reference electrode jack is connected to R42 in place of P11.
Do not bend the connector pin when installing the electrode jack. If a pin breaks the
entire CPU board may require replacement.
3. Snap the left and center mounting posts of the CPU board into place. For ILyte analyzers
with serial numbers below 6100, attach the center screw to the CPU board using a
#2 Phillips screw driver. See Figure R45.
P10 (K+) Connect 2nd
P8
(Cl-/Li+) Connect 4th
P9 P10
P8
P11
AGND
(Reference)
Connect 1st
Figure R43
P10
P10
(Na+) Connect 3rd
P9 P9
P8
P8
Screw
6100
Figure R45
56 Repair
4. Attach the display cable to the CPU board. Push the display cable tabs to the front to secure
cable in place. See Figure R46.
5. Connect the probe motor, sensor, pump motor, power supply and printer cables to the CPU
board, see Figure R47. Refer to Figure R48 for ILyte analyzers with serial numbers below
6100.
Display Cable
Figure R46
Figure R47
Power Supply
6100
Cable
Figure R48
Repair 57
Display Removal
1. Detach the two short self tapping screws from the sample detector board using a #2 Phillips
screw driver. See Figure R49.
2. Move and hold the display cable to expose the three display assembly screws. Detach the
three short screws from the display assembly using a #2 Phillips screw driver. See
Figure R50.
Sample Detector
Board
Screws
Figure R49
Display Cable
Screws
Figure R50
58 Repair
3. Carefully remove the display assembly and the LCD screen, holding both items together.
See Figure R51.
Notice the orientation of the LCD screen which is extremely important when installing the
display assembly. The one-line display has soldered leads located on the top of the LCD
screen; the two-line display has soldered leads located on the bottom of the LCD screen.
Refer to Figure R52 in the Display Installation for further details.
4. Remove the display adapter from the front housing window. Refer to Figure R51 for the loca-
tion of the display adapter.
For two-line displays there is an orientation notch which is important when installing the
display adapter. The orientation notch is located on the top inside edge of the adapter.
Display
Adapter
Display
Assembly LCD Screen
Figure R51
Repair 59
Display Installation
1. Place the display adapter into the front housing window. The smooth edge should be facing
the front housing of the analyzer.
For two-line displays there is an additional orientation notch. To install correctly, arrange
the adapter so that the notch is located on the top inside edge of the adapter.
2. Place the foam pad on the display board above the YES/NO buttons.
3. Place the LCD screen on top of the foam pad. The orientation of the LCD screen is extreme-
ly important, see Figure R52.
The one-line display has soldered leads located on the top of the LCD screen, the two-
line display has soldered leads located on the bottom of the LCD screen.
Soldered Leads
LCD Screen
4. Position the front housing so that the front of the analyzer is facing down. Align the display
assembly and insert into place, as shown in Figure R53, using the YES/NO buttons as a
guide.
5. Hold the display in place and attach the three screws using a #2 Phillips screw driver. See
Figure R54.
Display
Assembly
Figure R53
Screws
Figure R54
Repair 61
6. Align the sample detector board so the sample detector cable is positioned on the right side.
Attach the two screws lightly, using a #2 Phillips screw driver, see Figure R55.
7. Install the sample detector in the front of the analyzer to verify the sample detector board
alignment. See Figure R56. When properly aligned fully tighten the sample detector screws.
Sample Screws
Detector
Board
Sample
Detector
Cable
Figure R55
Sample Detector
Figure R56
62 Repair
2. Using a wire cutter, cut the tie wrap that secures the pump motor cable wires, as shown in
Figure R57.
3. Detach the four screws from the front of the analyzer, using a #1 Phillips screw driver. See
Figure R58. Some earlier ILyte analyzers may use socket screws that require a 1/16 inch
Allen wrench for removal.
Pump Motor
Assembly
Tie Wrap
Cable
Figure R57
Screws
Screws
Figure R58
Pump Motor
Assembly
Figure R59
Repair 63
1. Install the pump motor assembly through the front housing, as shown in Figure R60. The
pump motor wires are positioned below the pump when properly aligned.
2. Attach the four screws into the front of the analyzer, using a #2 Phillips screw driver. See
Figure R61. Earlier ILyte analyzers may use socket screws that require a 1/16 inch Allen
wrench for installation.
3. Using a tie wrap cable, secure the loose wires under the pump motor to the lead screw/probe
motor wires and the sensor wires.
Pump Motor
Assembly
Pump Motor
Wires
Figure R60
Screws
Screws
Figure R61
64 Repair
1. Detach the screw at the top of the probe position scale, using a #2 Phillips screw driver, as
shown in Figure R62. Remove the washer. Some ILyte analyzers use a white spacer that
should be removed also.
Screw
Probe Position
Scale
Figure R62
Probe Position
Scale
Figure R63
Repair 65
3. Remove the C-shaped retaining ring from the actuator, using the retaining ring pliers. DO
NOT stretch out the retaining ring with the pliers. Disengage the actuator from the top of the
lead screw assembly. See Figure R64.
4. Cut the tie wrap using a wire cutter, see Figure R65. Remove the probe motor cable wires
from the retaining clamp. See Figure R66.
Actuator
Retaining Ring
Top of
Lead Screw
Figure R64
Figure R65
Probe Motor
Cable Wires
Retaining
Clamp
Figure R66
66 Repair
5. Detach the three screws from the probe motor bracket, using a #2 Phillips screw driver, as
shown in Figure R67. Remove the washers.
The other two screws are located on the bottom edge of the bracket.
Probe Motor
Bracket
Figure R67
Probe Motor/
Lead Screw
Assembly
Figure R68
Repair 67
1. Place the probe motor assembly into place. See Figure R74.
2. Attach the three washers and screws to the probe motor assembly bracket, using a #2
Phillips screw driver. See Figure R75.
The other two screws are located on the bottom edge of the bracket.
Probe Motor
Assembly
Figure R74
Screw
Probe Motor
Bracket
Figure R75
68 Repair
3. Insert the actuator through the top of the lead screw assembly as shown in Figure R76.
Install the retaining ring onto the actuator using the retaining ring pliers. DO NOT stretch out
the retaining ring with the pliers.
4. Place the probe motor cable wires into the retaining clamp. See Figure R77. Secure the
wires to the sensor cable wires with a tie wrap cable. The optimum location for the tie wrap
cable is slightly to the left of the retaining clamp.
5. Install the alignment valve tool, part number 05000104, in place of the solutions valve.
Actuator
Retaining Ring
Lead Screw
Assembly
Figure R76
Probe Motor
Retaining Cable
Clamp Wires
Tie Wrap
Cable
Location
Figure R77
Repair 69
6. Install the sample probe as follows. Refer to Figure R78 for details.
A. With the probe hole facing out, align the probe collar with the white line on the top of the
probe arm. When aligned, snap the probe into the probe arm.
B. Manually move the probe arm until the sample probe hole is aligned with the top line
marked on the alignment valve.
When the sample probe is aligned, DO NOT move the probe arm until the Probe Motor
installation is complete.
Probe Arm
A. Probe Collar
aligned with Slot
Sample
Probe
Alignment
Valve
Figure R78
70 Repair
7. Hold the sample probe against the alignment valve tool to keep in place. Install the probe
position scale through the top of the housing, see Figure R79. Verify the probe position
alignment slot is located on the bottom right side of the scale.
Probe Position
Scale
Probe Position
Alignment Slot
Bottom of
Scale
Figure R79
Repair 71
8. Place the scale into position. The scale should be seated inside the probe position sensor
and inside the mounting bracket slot. See Figure R80.
9. Attach and slightly tighten the washer and screw at the top of the position scale. Do not fully
tighten the screw. Some ILyte analyzers use a white spacer that should be attached to the
back surface of the scale.
10. Align the probe position scale slot within the center of the probe position sensor. Gently
move the scale to the left and visually confirm the alignment. Refer to Figure R80 for cor-
rect alignment.
Confirm that the sample probe hole is aligned with the top line marked on the alignment
valve, on the front of the analyzer, before continuing.
Probe Position
Screw – Keep
loose until
scale in align-
ment
Probe Position
Scale CPU Board
Probe Position
Sensor
Mounting Bracket
Note:
Alignment of
Scale and
Sensor
Probe Scale
Alignment Slot
Figure R80
72 Repair
11. When alignment is complete, fully tighten the probe position screw into place using a #2
Phillips screw driver, as shown in Figure R81.
Screw
Figure R81
Repair 73
1. Remove the sample probe from the probe arm and the solutions valve. See Figure R69.
Remove the solutions valve.
2. Detach the screw from the bottom of the guide rod, using a #2 Phillips screw driver, as
shown in Figure R70. Remove the washer.
Probe Arm
Sample
Probe
Solutions
Valve
Figure R69
Guide Rod
Screw
Figure R70
74 Repair
3. Remove the guide rod and probe arm assembly, down and away from the housing.
See Figure R71.
Guide Rod
Probe Arm
Figure R71
Repair 75
1. Install the guide rod as shown in Figure R72. To install, first insert the guide rod into the
alignment hole located in the top wall of the front assembly.
2. Attach the washer and screw to the bottom of the guide rod, using a #2 Phillips screw
driver. See Figure R73.
Guide Rod
Alignment Hole
Probe Arm
Figure R72
Guide Rod
Screw
Figure R73
76 Repair
1. Insert the electrode jack wires into the 11/32 inch nut driver. See Figure R82.
2. Remove the hex nuts using the nut driver. See Figure R83.
3. Pull the electrode jacks from the front housing of the analyzer, as shown in Figure R84.
Electrode Jack
Wires
Nut Driver
Figure R82
Figure R83
Analyzer Front
Housing
Electrode Jack
Figure R84
Repair 77
1. Push the electrode jacks through the holes in the front of the analyzer, see Figure R85.
2. Insert the electrode jack wires into the 11/32 inch nut driver. See Figure R86.
3. Attach the hex nut using the nut driver, and tighten, as shown in Figure R87.
Analyzer Front
Housing
Electrode Jack
Holes
Electrode Jack
Figure R85
Electrode Jack
Wires
Nut Driver
Figure R86
Figure R87
Autosampler 79
Product Description
The ILyte autosampler when used with the ILyte analyzer provides automated batch sampling of
serum, plasma or diluted urine for electrolyte analysis. The autosampler can also perform emer-
gency (STAT) analysis of blood samples on demand.
The ILyte autosampler contains a 25 position sample tray which includes a STAT and two quality
control positions. The tray holds either 500 µl or 2 ml sample cups.
The ILyte analyzer software allows the operator to perform autosampler diagnostics by simply
pressing the “YES” or “NO” buttons in response to prompts displayed on the analyzer LCD
screen. Results are always displayed on the LCD screen and can also be documented on paper
by the ILyte analyzer printer.
80 Autosampler
The ILyte autosampler is made up of the following mechanical assemblies and associated com-
ponents. Refer to Figure A1 to determine the location of each assembly or component in relation
to the autosampler.
Sample Tray
IR Windows
Autosampler
Housing
IR Receiver IR Transmitter
Spindle Column
PC Board Assembly
Assembly
Drive Belt
Sensor
Motor Assembly
Assembly
Chassis
Assembly
Replacement List
The following assemblies and components can be purchased directly from IL.
Refer to the chart below to determine the problem and corrective action needed.
Alignment Verification
1. Remove and reinstall the power cord. The sample tray will rotate and then stop.
2. ALIGN PROBE/CUP? appears on the LCD screen. Press YES to verify the probe
alignment. The sample probe will descend and stop above the “Daily Cleaner” cup position.
3. ALIGN CUP!/CUP ALIGNED? appears on the LCD screen; do not press YES or NO at this
time.
4. Verify that the tip of the sample probe is centered over the “Daily Cleaner” cup position.
Based on the inspection results, take the appropriate corrective action.
• Probe Aligned: If sample probe is aligned in the proper position, press YES to the
ALIGN CUP!/CUP ALIGNED? display. ALIGNMENT OK? appears on
the LCD screen; press YES. CALIBRATE NOW? appears on the LCD
screen; press YES. Return to normal operation.
5. Grasp the autosampler and analyzer at the base and carefully slide forward until the
autosampler extends over the edge of the work surface.
6. Turn the knob underneath the autosampler base clockwise to loosen. DO NOT remove the
knob.
7. Locate the hub on top of the autosampler tray. Move the hub and the knob together, until
the sample probe aligns with the “Daily Cleaner” cup position.
8. When aligned, hold the hub in place and turn the knob counterclockwise to tighten.
9. Grasp the autosampler and analyzer at the base and carefully slide back to the original work
area.
10. Press YES to the ALIGN CUP!/CUP ALIGNED? display. ALIGNMENT OK? appears on the
LCD screen; press YES. CALIBRATE NOW? appears on the LCD screen; press YES.
84 Autosampler
1. Advance to the DAILY CLEANER? display; press YES. TRAY IN PLACE? appears on the
LCD screen; press YES.
2. The sample tray should rotate and stop with the probe at cup position “12”.
4. The sample tray should rotate and then stop with the probe at “Daily Cleaner” position.
• Correct Position: If the tray stops at the correct position, return to normal operation.
• Incorrect Position: If the tray stops at an incorrect position, perform the Carousel Test.
Autosampler 85
Carousel Test
1. Enter the QC TEST? menu and proceed to TEST CAROUSEL?. Refer to QC Test in this
section for further instructions.
3. CALIB CAROUSEL? appears on the LCD screen; press YES. The sample tray should spin
one revolution; then stop.
4. FIND HOME? appears on the LCD screen; press YES. The sample tray should spin and
stop with the probe at the “Daily Cleaner” position.
5. Based on the CALIB CAROUSEL? test results, take the appropriate corrective action.
1. Enter the QC TEST? menu and proceed to the TEST CAROUSEL? test. Refer to QC Test
in this section for further instructions.
2. At the TEST CAROUSEL? display, press YES. At the CALIB CAROUSEL? display, press
NO. At the FIND HOME? display, press YES. The tray should spin and stop with the sample
probe at the “Daily Cleaner” position.
4. The sample tray should rotate from the “Daily Cleaner” position to each cup position and
back for cup positions “NORMAL through 10”. Then the sample tray should reverse direction
and rotate from the “Daily Cleaner” position to each cup position and back for cup positions
“11 through STAT”.
5. The sample tray will perform four complete rotation cycles as described above.
6. The analyzer will print the test results during the fourth cycle, as illustrated in the
example below.
7. Verify that the number of steps between adjacent cup positions is 0096 +/-0006 for both
columns of test data.
8. Based on the tray cycle test results, take the appropriate corrective action.
• Error Message: If “MTC” or “CAROUSEL PROBLEM” appears on the LCD screen during
testing, perform Sensor Verification.
• In Specification: If tray cycle readings are in specification, exit the QC TEST? menu and
return to normal operation.
Autosampler 87
Sensor Verification
1. Remove the rear housing of the ILyte analyzer as detailed in Rear Housing Assembly
Removal in the REPAIR SECTION.
2. Connect a voltage meter to the ILyte analyzer. Attach one reference lead to TP11 (DGND)
and the other to U13 PIN #2 on the analyzer CPU board.
3. Manually turn the sample tray so the probe is at the “Daily Cleaner” position.
5. Manually turn the sample tray off the “Daily Cleaner” position.
7. Based on the voltage test results, take the appropriate corrective action.
• In Specification: If voltage readings are in specification, adjust the drive belt tension and
repeat Tray Cycle Verification. If “MTC” or “CAROUSEL PROBLEM”
errors continue, replace the spindle column assembly. Refer to Spindle
Column Removal and Installation in this section for details.
• Out of Specification: If voltage readings are out of specification, replace the sensor assem-
bly. Refer to Sensor Removal and Installation in this section for
details.
88 Autosampler
QC Tests
1. To access the QC TEST? functions, answer the questions that appear on the ILyte display,
as shown on the chart below.
2. Two numbers appear on the screen separated by an arrow. A cursor will be positioned
underneath the first digit of the first number.
3. Change the first number to “246.0” as follows. Press NO to change the first digit. When the
digit is correct press YES to advance to the next digit.
5. Press YES when finished. QC TEST? will appear on the screen. If a value other than “246”
is input the screen displays CORRECT?. Press NO to return to STEP 2 and correct any
error.
7. Consult the table below which lists all the QC tests and their functions. To select a test press
YES. To bypass a test press NO.
Repair Overview
Always consult Troubleshooting in this section first to determine the problem before attempting to
service the autosampler.
Please note that complete mechanical assemblies can be replaced. Refer to the Replacement
List in this section which lists replaceable assemblies and appropriate catalog numbers.
Recommended Tools
The following is a list of the tools required to perform repairs on the autosampler. Please refer to
this list to determine the appropriate tool and size needed.
Housing Removal
1. Turn the autosampler upside down and gently place on a flat surface.
2. Remove the two screws from the cable retaining plate using a #2 Phillips screw driver, as
shown in Figure A2.
Screws
Cable Retaining
Plate
Figure A2
92 Autosampler
3. Remove the autosampler cable from the groove in the bottom of the housing.
4. Remove the three screws from the metal base using a #2 Phillips screw driver, as shown in
Figure A3.
6. Lift the housing, as shown in Figure A4, to remove from the chassis assembly.
Metal Base
Screws
Figure A3
Housing
Chasis Assembly
Figure A4
Autosampler 93
Housing Installation
1. Align and place the housing over the chassis assembly, as shown in Figure A5.
Be careful not to damage the IR receiver or transmitter on the PCB.
2. Turn the autosampler upside down and gently place on a flat surface.
3. Attach the three screws to the metal base using a #2 Phillips screw driver, see Figure A6.
Housing
Chasis
Assembly
Figure A5
Screws
Metal Base
Figure A6
94 Autosampler
4. Thread the autosampler cable into the groove in the bottom of the housing. See Figure A7.
5. Attach the two screws to the cable retaining plate using a #2 Phillips screw driver, as shown
in Figure A7.
Cable Retaining
Plate
Autosampler
Cable
Screws
Figure A7
Autosampler 95
1. Disconnect the sensor cable and the motor cables from the PC Board. See Figure A8.
2. Turn the adjustment knob counterclockwise to loosen. Remove the adjustment knob from
the bottom of the chassis assembly, as shown in Figure A9.
Some autosamplers use a spacing disk that should be removed and retained.
PC Board
Sensor Cable
Motor Cable
Figure A8
Adjustment
Knob Chassis
Assembly
Figure A9
96 Autosampler
Spindle Column
Assembly
Figure A10
Autosampler 97
1. Place the spindle column assembly down into place, as shown in Figure A11.
2. Attach the adjustment knob, see Figure A12, and turn clockwise to tighten.
Some autosamplers use a spacing disk that should be attached before the knob is
installed.
Spindle Column
Assembly
Figure A11
Adjustment Knob
Figure A12
98 Autosampler
3. Attach the sensor cable and the motor cable to positions J1 and J3 on the PC board, see
Figure A13.
Sensor Connector
(J3)
Sensor
Cable Motor Connector (J1)
Motor
Cable
Figure A13
Autosampler 99
Sensor Removal
1. Loosen the two slotted screws from the motor bracket, using a #2 Phillips screw driver, as
shown in Figure A14.
3. Remove the two slotted screws from the mounting block, using a #2 Phillips screw driver.
Screws
Figure A14
Sensor
and Mounting
Block
Motor
Bracket
Figure A15
100 Autosampler
Sensor Installation
2. Attach the two slotted screws to the mounting block, using a #2 Phillips screw driver.
3. Place the sensor and mounting block into the motor bracket. See Figure A16.
4. Attach the two slotted screws to the motor bracket, using a #2 Phillips screw driver, as shown
in Figure A17.
Sensor
and Mounting
Block
Motor
Bracket
Figure A16
Screws
Figure A17
Autosampler 101
Motor Removal
1. Loosen the socket screw from the brass damper using a 1/16 inch Allen wrench. See Figure
A18.
2. Remove the damper from the top of the motor shaft, as shown in Figure A19.
Damper
Socket Screw
Figure A18
Damper
Motor
Shaft
Figure A19
102 Autosampler
3. Remove the four slotted screws from the top of the motor using a 7/32” flat head screw driv-
er. See Figure A20.
4. Remove the motor from the motor shelf, See Figure A21.
Slotted
Screws
Motor
Figure A20
Figure A21
Autosampler 103
Motor Installation
2. Pull the belt around the gear of the motor, see Figure A22 for location of gear and belt.
3. Attach the four slotted screws to the top of the motor using a 7/32” flat head screw driver, as
shown in Figure A22.
4. Attach the damper to the top of the motor shaft, as shown in Figure A23.
Motor Gear
Slotted
Screws Belt
Motor
Figure A22
Damper
Motor
Shaft
Figure A23
104 Autosampler
5. Tighten the socket screw to the damper using a 1/16 inch Allen wrench. See Figure A24.
Damper
Socket Screw
Figure A24
Autosampler 105
PC Board Removal
1. Disconnect the power cable from the PC board. See Figure A25.
2. Remove the hex nut from the heat-sink plate using a 1/4 inch open end wrench.
See Figure A26.
Power Cable
Figure A25
Hex Nut
Heat Sink
Plate
Figure A26
106 Autosampler
3. Remove the four screws from the PC board using a #2 Phillips screw driver.
See Figure A27.
The lower left screw may be hidden by the EMI suppressor box. If so, peel back the tape
to reveal the screw.
4. Lift the PC Board, as shown in Figure A28, to remove from the base housing.
PC
Board
Screws
Screws
Figure A27
PC
Board
Base
Housing
Figure A28
Autosampler 107
PC Board Installation
1. Place the PC Board, as shown in Figure A29, onto the base housing.
2. Install the four screws to the CPU board using a #2 Phillips screw driver,
see Figure A30.
Figure A29
Screws
Screws
Figure A30
108 Autosampler
3. Install the hex nut to the heat-sink plate using a 1/4 inch open end wrench,
see Figure A31.
4. Connect the power cable to position J2 on the PC board, see Figure A32.
Hex Nut
Heat Sink
Plate
Figure A31
PC Board
Power
Cable Power
Connector
(J2)
Figure A32
Specifications 109
109
Analyzer Specifications
Sample Size: 100 µl Whole Blood, Serum, Plasma or 400 µL diluted (1:10) Urine
Resolution: Na+: 0.1 mmol/L K+: 0.01 mmol/L Li+: 0.01 mmol/L Cl-:0.1mmol/L
BLOOD, SERUM, PLASMA 80-200 mmol/L 1-10 mmol/L 0.2-2.0 mmol/L 80-200 mmol/L
Reproducibility (typical)
within run (n=20) C.V. <1% C.V. <2% SD≤.03 mmol/L C.V. <2%
between run (10 day) C.V. <2% C.V. <2.5% SD≤.05 mmol/L C.V. <2.5%
URINE
Reproducibility (typical)
within run (n=20) C.V. <5% C.V. <5% Not C.V. <5%
between run (10 day) C.V. <5% C.V. <5% Applicable C.V. <5%
Accuracy (typical)
Correlation coeff 1.00 1.00 Not 0.98
slope 1.05 1.00 Applicable 1.07
Power: 100/115~VAC, 50-60 Hz, 0.8A/250 V fuse or 220~VAC, 50-60 Hz 0.4A/250 V fuse
Refer to the label on the back of the analyzer for preset power and fuse
settings.
Autosampler Specifications
ABNORMAL CONTROL?
Allows operator to select analysis of abnormal control.
ABNORMAL STATS?
Allows operator to display or print statistics for stored abnormal control results.
AIR IN CLEANER
Cleaning solution not detected correctly.
AIR IN CONTROL
Control sample not detected correctly.
AIR IN SAMPLE
Sample not detected correctly.
AIR IN STD A
Standard A not detected correctly.
AIR IN STD B
Standard B not detected correctly.
AIR IN FLUSH
Flush not detected correctly.
ALL ABNORMAL?
Allows operator to delete all stored abnormal control results.
ALL NORMAL?
Allows operator to delete all stored normal control results.
ALL RESULTS?
Allows operator to display all stored results.
ANALYZE BLOOD?
Sample analysis mode for Blood, Serum or Plasma. The message will only appear
when the analyzer is calibrated.
*ANALYZE URINE?*
Urine analysis mode for Na/K/Cl results only.
ANALYZING...
Analyzer is measuring sample.
ASPIRATING...
Sample is being drawn into probe.
AUTOCAL?
Allows operator to activate and to set time for initial daily calibration.
BLOOD Cl LIMITS?
Allows operator to set normal Cl range.
112 Appendix
BLOOD K LIMITS?
Allows operator to set normal K range.
BLOOD Li LIMITS?
Allows operator to set normal Li range.
BLOOD NA LIMITS?
Allows operator to set normal Na range.
CAL VALUES?
Allows operator to display last calibration values.
CALIBRATE NOW?
Allows operator to calibrate analyzer.
CALIBRATING...
Analyzer is calibrating.
CALIBR’TING STD A
Electrodes being calibrated in relation to Standard A.
CALIBR’TNG STD B
Electrodes being calibrated in relation to Standard B.
CHANGE DAY/TIME?
Allows operator to change date and time.
CHANGE PACK!
Occurs at 99% pack usage. Pack must be replaced!
CLEANING...
Analyzer is performing daily cleaning.
Cl HIGH...
Chloride result obtained is higher than the preset normal range (flashing display).
Cl LOW...
Chloride result obtained is lower than the preset normal range (flashing display).
CONTINUE?
Analyzer requests confirmation to proceed.
C.V. VALUES?
Allows operator to display % C.V.’s of Quality Control results.
DAILY CLEANER?
Allows operator to perform daily cleaning.
DAILY SUMMARY?
Allows operator to print summary of daily results.
*DATE/RANGES?*
Allows operator to set date, time, and ranges.
DELETE CONTROLS?
Allows operator to delete stored control results from memory.
DELETE RESULTS?
Allows operator to delete patient sample results from memory.
Appendix 113
*DIAGNOSTICS?*
Allows operator to check key analyzer functions.
DILUTED 1:10?
Analyzer requests confirmation that urine sample has been diluted.
DRIFT Cl
Chloride electrode is drifting outside of preset electrode operating range.
DRIFT K
Potassium electrode is drifting outside of preset electrode operating range.
DRIFT Li
Lithium electrode is drifting outside of preset electrode operating range.
DRIFT NA
Sodium electrode is drifting outside of preset electrode operating range.
DRIFT, RECAL..
Analyzer is automatically recalibrating because of electrode drift.
ELECTR’D VALUES?
Allows operator to display real time electrode millivolt readings.
ENTIRE SUMMARY?
Allows operator to display summary of all results in storage.
***ERROR***
Denotes an error has occurred. Press YES or NO to determine error.
FLOW TEST FAIL
Unsuccessful Fluid Flow Test.
FLUID FLOW OK
Successful Fluid Flow Test.
FLUID FLOW TEST?
Allows operator to run red Test Dye to check fluid flow.
FLUID PATH!
Analyzer is not detecting fluid and/or air segments when expected.
FLUSH?
Allows operator to perform a Flush cycle on demand.
FLUSHING...
Analyzer is flushing electrodes.
FLUSH, RECAL ...
If calibration fails, analyzer does a Flush and recalibration automatically.
I.D. #- - - -
Allows operator to assign an identification number to patient result.
K HIGH...
Potassium result obtained is higher than preset normal range (flashing display).
K LOW...
Potassium result obtained is lower than preset normal range (flashing display).
LAST CONTROL?
Allows operator to display last control result.
114 Appendix
LAST RESULT?
Allows operator to display last analysis result.
Li BELOW 0.2mM/L
Lithium sample concentration is below ILyte measurement range.
Li HIGH..
Lithium result obtained is higher than the preset normal range (flashing display).
Li LOW...
Lithium result obtained is lower than the preset normal range (flashing display).
MEAN VALUES?
Allows operator to display mean values of quality control results.
MTP
Mechanical problem with pump.
MTS
Mechanical problem with probe.
MUST USE CLEANR!
Cleaning Solution must be used before operating the analyzer.
MV RANGE Cl
Measured Chloride electrode voltage is outside operating limits.
MV RANGE K
Measured Potassium electrode voltage is outside operating limits.
MV RANGE Li
Measured Lithium electrode voltage is outside operating limits.
MV RANGE Na
Measured Sodium electrode voltage is outside operating limits.
NA HIGH...
Sodium result obtained is higher than the preset normal range (flashing display).
NA LOW...
Sodium result obtained is lower than the preset normal range (flashing display).
**NA/K*** or ***Na/K/Cl*** or ***Na/K/Li***
Appears when power is connected or restored after interruption.
Na/K/Cl or Na/K/Li RANGE?
Allows operator to set normal ranges.
NEW PACK INSTLD?
Asks operator to confirm that a new pack has just been installed.
NEXT SAMPLE?
Allows operator to proceed to next analysis.
NOISE
Erratic signal coming from electrodes.
NOISE, RECAL..
Analyzer automatically recalibrates after detecting noise.
NORMAL CONTROL?
Allows operator to select analysis of normal control.
Appendix 115
NORMAL STATS?
Allows operator to display or print statistics for stored normal control results.
NOT ENOUGH DATA!
In order to calculate statistics, the Analyzer requires 5 stored control results of both normal
and abnormal levels.
*OPER FUNCTS?*
Allows operator to select “On Demand” flush, purge or calibration functions.
*PACK USAGE?*
Allows operator to display % pack expended.
PATIENT ID#?
Allows operator to assign identification number to patient’s result.
% OF PACK USED
Percentage of solutions pack that has been expended.
PRINTER OFF?
Allows operator to turn printer function on or off.
PRINTING...
Printer is printing requested data.
PRINT RESULTS?
Allows operator to print all results stored in memory.
PRINT STAT’TICS?
Allows operator to print normal or abnormal control summaries.
PRINT SUMMARY?
Prints daily summary of results.
PRINT WKLY SUM?
Prints summary of last five days results.
PROBE IN BLOOD?
Requests operator to confirm that probe is in blood.
PROBE IN CLEANR?
Requests operator to confirm that probe is in daily cleaning solution.
PROBE IN CONTROL?
Requests operator to confirm that probe is in control sample.
PROBE IN DYE?
Requests operator to confirm that probe is in Test Dye solution.
PROBE IN URINE?
Requests operator to confirm that probe is in urine.
PROBE IN FLUSH?
Requests operator to confirm that probe is in Flush Solution.
PROBE WIPING?
Allows operator time to manually wipe the probe after sample is removed.
PUMP CAL...
Pump is being calibrated.
PUMPCAL FAILURE!
Failure of Pumpcal Test.
116 Appendix
PUMPCAL OK
Successful completion of Pumpcal Test.
PUMPCAL TEST?
Allows operator to test pump calibration.
PURGING STD A
Standard A fluid line is being purged.
PURGING STD B
Standard B fluid line is being purged.
PURGING FLUSH
Flush Solution fluid line is being purged.
*QUAL CONTROL?
Allows operator to choose quality control functions.
REMOVE CONTROL
Requests operator to remove control sample from probe.
REMOVE SAMPLE
Requests operator to remove sample from probe.
RESET TO 0%?
Sets pack usage percentage counter to 0%.
RUN CONTROL?
Quality control analysis mode.
SAMPLE CAL...
Analyzer is calibrating the Sample Detector.
SAMPLE DETECT OK
Successful completion of Sample Detector Test.
SAMPLE DETECTOR!
Sample Detector readings out of range.
SAMPLE DET. TEST?
Allows operator to test sample detector.
S.D. VALUES?
Allows operator to display Standard Deviation values of Quality control results.
*SECOND MENU?*
Allows operator access to secondary menu.
SEE MANUAL!
See Operator’s Manual for assistance or explanation.
*SEE RESULTS?*
Allows operator to display or print stored results.
SEE STATISTICS?
Allows operator to display Mean, S.D., C.V. values of Quality Control results.
SET CORRELATION?
Allows operator to select dilution equivalent correlations.
SOL’N PACK LOW!
Appears when solutions pack is 88% expended. Continue operating analyzer.
Appendix 117
SOL’N PURGE?
Allows operator to purge air from fluid lines.
***STANDBY***
Appears when Analyzer is in Standby mode.
STANDBY IN 1 MIN
Analyzer will go into standby in 1 minute (Flashing Display).
STANDBY MODE?
Allows operator to put analyzer in standby mode to conserve fluids.
STATISTICS?
Allows operator to display statistics for all control results.
STORE RESULT?
Allows operator to store results in memory.
SYRINGE SAMPLE?
Allows operator to adjust probe extension during analysis to accommodate a syringe.
TESTING...
Analyzer in process of completing a Fluid Flow Test.
TRY PURGE AGAIN?
Allows operator to perform additional purge after initial failure.
URINE Cl LIMITS?
Allows operator to set normal Cl range for urine.
URINE K LIMITS?
Allows operator to set normal K range for urine.
URINE NA LIMITS?
Allows operator to set normal Na range for urine.
USE EXT. FLUSH!
Flush solution in pack is empty. Operator should use external flush.
USER OPTIONS
Allows operator to access the following functions: Probe wiping, Patient I.D. #,
Syringe Sample, Set Correlation, Autosampler and Printer Off.
WEEKLY SUMMARY?
Allows operator to print last five days of stored results.
WIPE PROBE
Allows operator time to wipe off sample from exterior of probe.
400 mL PACK?
Allows operator to select for correct solutions pack size.
800 mL PACK?
Allows operator to select for correct solutions pack size.
118 QC TEST? Flowchart (Service Diagnostics)
DATE/RANGES?
BLOOD Na LIMITS?
246.0
Yes/No
QC TEST?
TURN ON MV’S?
CHECK SUM
CHECK CAL
CONSTS?
RELEASE DATE
SIPPER TEST?
CALIB CAROUSEL?
CYCLE TRAY?
TEST CAROUSEL?
ELECTRODES VALUES?
CALIBRATE NOW?
FLUID FLOW TEST?
PUMPCAL TEST?
TEST VALVE?
8 NORMAL AND
RUN CONTROLS? NORMAL CONTROL?
STATS
8 ABNORMAL AND
ABNORMAL CONTROL? STATS
TEST ALIGNMENT?
AUTOSAMPLER
5 NORMAL AND 5 ABNORMAL CONTROLS
AT CUP #1 TO #10 IN SAMPLE TRAY
BURN IN?
QC TEST?
QC TEST? Flowchart (Continued) 119
QC TEST?
Y TURNS ON
TURN ON MV’S?
PRINTED MV’S
PERFORMS NON-DESTRUCTIVE
Y RAM TEST AND CALCULATES
RAM/PROM TEST?
PROGRAM CHECKSUM
***NA/K/CL***
N
(PROGRAM DATE)
N Y MOVES PROBE
UP AND DOWN
PUMP TURNS SLOWLY FOR 10
REVOLUTIONS TO PERMIT
PUMP TEST? NOISE AND STEP ACCURACY
Y CHECK
N
ELECTRD VALS?
CALIBRATE NOW SEE PAGE 121
Y
N N
ILyte Na/K/Li
>88% Y >99% Y
PACK USED PACK USED CHANGE PACK
? ?
N N
Y N
JUST
CALIBRATED CALIBRATED SOL’N PACK LOW
? ? Y
N
SEE PAGE 121
Y N
N
N N
CALIBRATE NOW?
N
DAILY CLEANER? N
Y
CALIBRATING... TRY PURGE AGAIN?
SEE MANUAL!
Y
PURGE FLAG PURGING STD B
?
“ERROR”
N N
AIR IN STD B
N
OK? N N
Y
FIRST
PURGE ERROR
? Y
PURGING WASH
“ERROR”
N DAYS
FIRST CAL
? N
OK?
AIR IN WASH
Y Y
“ERROR”
N
N Y N
OK? OK?
AIR IS STD A
Y N
Y DAYS Y N
FIRST CAL OK? “ERROR” (SETS PURGE FLAG)
?
FLUSHING...
PROBE
AIR ININWASH
FLUSH?
N
Y Y
CALIBRATING STD A OK?
USE EXT FLUSH? CALIBRATE NOW?
N
N
CALIBRATING STD A
CALIBRATE NOW?
NOISE, RECAL...
NOISE Na OR K
Y
N N FIRST N
OK? NOISE? AIR ERROR FIRST NOISE? “ERROR”
?
Y
Y N AIR IN STD A
(SETS
OR PURGE
N FLAG)
“ERROR” AIR IN STD B
FLUSH, RECAL...
CALIBRATING STD B
N
OK?
N
OK?
Y
FLUSHING
CALIBRATING STD A
N Y
DRIFT RECAL...
Y
BAD FIRST BAD
SLOPES ? N “ERROR”
Y N
OK?
Y
N
Y N
DRIFT? FIRST DRIFT DRIFT Na OR K
MAIN MENU ?
LOOP
SEE PAGE 124
124 Software Flowchart
USER ENTRY
Y Y Y
ANALYZE BLOOD? ID SET? PATIENT ID? ID XXXX
N N Y
N
DAILY CLEANER? CONTINUE?
N Y
LOWEST SYRINGE SET? MAIN MENU
POSITION LOOP
M
A
AIR IN SAMPLE I
N
PROBE IN BLOOD?
N L
Y FLUID PATH O
AIR IN SAMPLE
? Y O
ASPIRATING... P
N
Y
Y 4TH
PROBE Y
WIPE SET? REMOVE SAMPLE AIR IN STD A THIS DAY
? ?
N N
Y
NOISE? NOISE Na OR K
N
OK? “ERROR”
C N
Y DRIFT Na OR K
L
E
A
N Na XXXX K XXX Na XXXX KXXX
E N OR
Y
R NEXT SAMPLE? Na/K HIGH/LOW
?
Software Flowchart 125
N
DAILY CLEANER? STANDBY MODE?
Y SEE BELOW
N
PROBE IN CL’NR? CLEANING...
ASPIRATING...
N OK?
Y Y
OK?
N “STANDBY“
SEE BELOW
“ERROR”
AIR IN CLEANER
N
STANDBY MODE?
Y
NO ACTION
STANDBY IN 1 MIN
N
N Y N
MAIN MENU SECOND MENU? MAIN MENU
“STANDBY”
LOOP LOOP
Y CLEARS CALIBRATION
ANALYZE URINE
N Y Y
ANALYZE URINE SEE RESULTS? PATIENT ID#?
ID SET?
Y
N
N
ID# XXXX
Y
N
DILUTED 1:10? CONTINUE?
Y N
Y
Y
N
PROBE IN URINE? ASPIRATING: WIPING SET REMOVE SAMPLE
Y ?
N
ANALYZING... WIPE PROBE
Y
C
L AIR IN SAMPLE AIR IN SAMPLE
E N ?
OK? N
A “ERROR”
N
E Y FLUID PATH
R Y
?
Y 4TH N
AIR AIR IN STD A
IN STD A THIS DAY
? ?
Y NOISE Na OR K
N Y
Na XXXX K XXX Na XXXX K XXX
OR
NEXT SAMPLE? Na/K HIGH/LOW NOISE? DRIFT Na OR K
N
N
SEE RESULTS?
Y
Y
Y
DAILY SUMMARY? PRINT SUMMARY? PRINTING...
N N
Y N
WEEKLY SUMMARY PRINT WKLY SUM? OK?
“ERROR”
N N
Y PRINTER ERROR
Y NEXT SAMPLE?
ALL RESULTS? PRINT RESULTS?
Y
N
N N
Y MAIN MENU
DELETE RESULTS? ALL RESULTS?
LOOP
N Y ALL RESULTS BUT LAST RESULT CLEARED.
N
SAMPLE COUNTER RESET TO ZERO.
N
PACK USAGE? QUAL CONTRL?
Y
RESET TO 0%? “STANDBY” MAIN MENU
N N LOOP
Y
WIPING
QUAL CONTROLS? OPER FUNCTS? SET?
N
Y
N
N N
Y
CALIBRATED ABNORM CONTRL? PROBE IN CONTRL? REMOVE SAMPLE
N ?
Y
Y
N
Y
RUN CONTROL? NORM CONTROL? ASPIRATING... WIPE PROBE
Y
N
ANALYZING... “ERROR”
Y N AIR N N
OK? NOISE?
ERROR
(5 MIN ?
TOGGLE)
MAIN MENU Y
Na XXXX K XXX LOOP
N Y AIR IN CONTROL
(RESULT IN MEM)
N O
NOISE Na/K
NEXT SAMPLE R
AIR IN STD A
Y N
LAST CONTROL? NORMAL CONTRLS ABNORM CONTROLS
N N Y
Y
STATISTICS?
Y N
N
SEE STATISTICS? NORMAL STATS? ABNORMAL STATS?
N
Y Y
N
>1
CONTROL QUAL CONTROLS
N >5 N
DELETE CONTRLS? PRNT STATISTICS? CONTROLS NOT ENOUGH DATA
?
Y
RETURN
Y
ALL NORMAL?
Y
N MEAN VALUES? Na XXXX KXXXX
Y
ALL NORM
N
ERASED
ALL ABNORMAL?
Y S.D. VALUES? Na XXXX KXXXX
N Y
N
ALL ABNRML
ERASED C.V. VALUES? Na XXXX KXXXX
Y
N
RET
FLUSHING... OK?
CLEANED? CALIBRATE NOW?
Y N
N
N N
Y FLUSH?
USE EXT. AIR IN FLUSH
Y
Y Y
FLUSH? PROBE IN FLUSH “ERROR”
N
N
USER OPTIONS?
N AIR IN STD B FIRST TRY
OK? ?
Y
N Y N
SEE MANUAL!
“ERROR” N
DRIFT Na/K
OK?
N
Y
PROBE WIPING?
N Y
WIPING
WIPING SET
CLEARED
PATIENT ID#?
N Y
PRINTER OFF PRINTER OFF
ID MODE CLEARED
ID MODE SET
CLEARED
N Y
N Y
SYR POS
CLEARED SYR POS SET
SET CORRELATION
N Y
CORRELATE
BLOOD?
N Y
CORRELATED
ISE UNITS SET
VALUES SET
(OPERATOR ENTERS
Na = ___+___ VALUES
K = ___+___ FOR CORRELATION)
CORRELATE URINE?
N Y
CORRELATED
ISE UNITS SET
VALUES SET
AUTOSAMPLER?
RESET
N Y Y
SAMPLER CONNECTED? Na/K...
132 Software Flowchart
DIAGNOSTICS? DATE/RANGES?
N
N
ELECTRD VAL’S? Na XXXXK XXXX (LIVE mM READINGS)
N CONTINUE?
Y
N
NEXT SAMPLE?
N Y
Y
CAL FREQUENCY? TIME SPAN X HRS AUTOCAL?
Y
N (USER ENTRY) (TURNS OFF (USER
AUTOCAL) ENTRY)
NEXT SAMPLE?
N Y
134 Software Flowchart
CPU Board – Analyzer Series under 6100
CPU Board
Power Supply – Analyzer Series under 6100
Transformer Assembly
Autosampler PCB
Worldwide Locations
Instrumentation Laboratory
Corporate Headquarters
Aragón 90-08015 Barcelona, Spain
P.O. Box 35027 (08080)
Telephone: 34-3-4010101
Fax: 34-3-4513745
001462-002 R2
Fax: (617) 861-1908
Italy Switzerland
Instrumentation Laboratory SpA Instrumentation Laboratory AG
Divisione Commerciale Italia Giessenstrasse 15 - Postfach - CH-8952 Schlieren (ZH)
Viale Monza 338-20128 Milan, Italy Telephone: 41-1-7423030 - Fax: 41-1-7423035
Telephone: 39-2-25221-Telex: 330112 ILSpA I
Fax: 39-2-2575250 United Kingdom
Instrumentation Laboratory (U.K.) Ltd.
The Netherlands Kelvin Close - Birchwood Science Park
Instrumentation Laboratory (Netherlands) BV Warrington - Cheshire WA3 7PB - England
P.O. Box 253 Ambachtsweg 3-4 Telephone: 44-925-810141 - Telex: 627713 ILUK G
3400 AG IJsselstein, The Netherlands Fax: 44-925-826708
Telephone: 31-3408-86824 - Fax: 31-3408-87973