Spesifikasi Teknis Hyperion X9

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 4

Spesifikasi Teknis Hyperion X9

1. ELECTRICAL FEATURES

Input voltage 115 – 240 VAC


Power supply frequency 50 / 60 Hz
Current absorbed in rest conditions 1A @ 115V; 0.5A @ 240V
Maximum current absorbed in operating conditions 14A @ 115V; 7A @ 240V
Column movement operating cycle 25 s ON, 400 s OFF
Maximum apparent line resistance 0.5 Ω @ 240V – 0.25 Ω @ 115V
Technical factors corresponding to the maximum 85 kV 10 mA
power absorption

2. RADIOLOGICAL FEATURES

Generator voltage 60 – 85 kV continuous mode (2D MODE)


90 kV pulsed mode (3D CB3D MODE)
(selectable in steps of 1 kV)
Anode current 1 – 10 mA continuous mode (2D MODE)
selectable in steps of 1 mA
3 – 10 mA pulsed mode (3D CB3D MODE)
with automatic selection
Maximum deviations from declared values: kV: < 10%
mA: < 10%
Linearity < 5 %
Coefficient of variation < 0.05
X-ray tube CEI OPX/105
Or
TOSHIBA D-054SB
Focal spot size 0.5 mm according to IEC 60336
Anode material Tungsten (W)
Anode inclination 5°
Anode thermal capacity 30 kJ (CEI)
35 kJ (Toshiba)
Maximum continuous heat dissipation 270 W

Inherent filtration 3.2 mm Al @ 85 kV (2D mode)


6.2 mm Al @ 85kV (CB3D mode)
Half Value Layer (HVL) 2D (PAN,CEPH): HVL=3.6mm @85kV
3D (CBCT mode): HVL=5.0mm @90kV (with
automatic added filter), with CB3D mode
Dispersed radiation < 0.88 mGy/h to 1m in any direction (1:20 duty cycle)
@ 90 kV
Exposure time (PAN/CEPH) 2.6 – 14 s, continuous radiation
Load factor of X-ray unit Max 20 s ON, 400 sec. OFF
Source distance - primary collimator 100 mm
Source - skin distance (PAN) > 150 mm
Source - skin distance (Ceph) > 1000 mm
Source - skin distance (CB3D) > 150 mm
Source - PAN sensor distance (SID) 55 cm
Source - CEPH sensor distance (SID) 151 cm
Output radiation intensity (Output dose) typical 0.30 mGy/s (+/- 30%) @ 70 kV, 10 mA, 1 m -
1.0 mGy/s (+/- 30%) @ 70kV 10 mA 55 cm
X-RAY TUBE HEATING AND COOLING CURVE (CEI OPX/105)

X-RAY TUBE HEATING AND COOLING CURVE (TOSHIBA D-054SB)

TUBE HOUSING HEATING AND COOLING CURVE


PAN beam opening

3. RADIOLOGICAL FEATURES IN CB3D MODE

Operating voltage 90kV with all projections


Generator voltage at maximum current 90kV - 10mA
Load factors for the maximum output power 90kV - 10mA
Maximum output rated power (value averaged with 810 VA (90kV 10mA variable duty cycle)
T>4s)
Exposure time (CB3D) 18 – 36 sec pulsed mode, with actual X-ray
emission time 3.6-9 sec
Operating cycle in pulsed mode for CB3D exposure 20% - 90%

4. PAN / CEPH SENSOR FEATURES

Detectors PAN CCD


CEPH CCD
Pixel size 48 m
Pixel binning mode Binning hardware 2X2: giving pixel size 96 m
Number of tiles 2 for PAN projections
3 for CEPH projections
Effective image reception area 146 x 6 mm for PAN detector
220 x 6 mm for CEPH detector
Number of active pixels per tile 1536 x 128 per tile (not binned)
Dead area at tile junction ≤ 192 m (corresponding to 2 rows in 2X2 binning)
Detector life cycles > 30'000 exposures depending on type of exam