AX Systems

AX

Startup Instructions
Table for Ground Wire Measurement(s)

Alternative to the "Safety-technical © S iem ens AG 2003


Checks" Protocol (AX00.000.820.01....) The reproduction, transmission or
use of this document or its contents
is not permitted without express
The safety-technical check is mandatory in only a written authority. Offenders will be
liable for damages. All rights,
few countries! including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.

English
Print No.: AX00-000.815.02.01.02 Doc. Gen. Date: 11.03
Replaces: n.a.
2 Ground Wire Measurement Table 1

System- System System


Type:................................................... Part No.: ............................................ Serial No.:.......................................

Test Equipment Used: (ground wire test meter)

Manufacturer: ....................................... Model: ................................................... Serial No.: ...........................................

During startup, if a safety-technical check does not have to be performed, the individual
measurement results must be entered in this table.

Ground Wire Test: Table of Individually Measured Values


From the power distribu- Enter the max. measured value or check n.a.
tor/ generator bus rail or (for the tolerances, see the Startup Instructions.)
power connector to: n.a. means: component not configured.

No. 1 *1) No. 2 *1) No. 3 *1) No. 4 *1)

‡ ‡ ‡ ‡
Unit *2) ......... Ω ......... Ω ......... Ω ......... Ω
n.a. n.a. n.a. n.a.

‡ ‡ ‡
Collimator *2) ......... Ω ......... Ω ......... Ω
n.a. n.a. n.a.

‡ ‡ ‡
Tube unit housing ......... Ω ......... Ω ......... Ω
n.a. n.a. n.a.

Monitor(s) (1 to 4) in the ‡ ‡ ‡ ‡
......... Ω ......... Ω ......... Ω ......... Ω
examination room n.a. n.a. n.a. n.a.

Monitor(s) (5 to 8) in the ‡ ‡ ‡ ‡
......... Ω ......... Ω ......... Ω ......... Ω
examination room n.a. n.a. n.a. n.a.

Monitor(s) in the ‡ ‡ ‡ ‡
......... Ω ......... Ω ......... Ω ......... Ω
control room n.a. n.a. n.a. n.a.

‡
Image system ......... Ω
n.a.

‡ ‡
Generator ......... Ω ......... Ω
n.a. n.a.

Monitor trolley system, ‡ ‡


......... Ω ......... Ω
DCS System n.a. n.a.

‡
Other components ......... Ω .............................................................. Remark
n.a.

*1) Component No.: systems sometimes include more than one examination unit or e.g., more than
one monitor. Fill out a separate column for each examination unit or monitor.
*2) The resistance is measured to all metal parts of the components that can be touched,
Exception: accessory rail on the collimator and tabletop.

Tester: ....................................... Date: ...................................... Signature: ...............................

AX Systems AX00-000.815.02 Page 2 of 2 Siemens AG


Rev. 01 11.03 CS PS 21 Medical Solutions