Specificatioas: Harm O N Ic Distortion Meter

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 14

specificatioas

FORTHE

Pioneer And World-Leaderin Eleclronic Kits

H A R MO N I C
-
ir'1," '" - ." )
DISTORTION METER
MODEL IM-58

O I S T O F T I O N
H A F M O N i O

All prices are subject to change without notice.


The Heath Companyreserves the right to discon- HEATH COMPANY
tinue instruments and to changespecijications at BENTON HARBOR,
any time. MICHIGAN

596 - 1003 PRINTEO IN tJ.S. A.


SPECIFICATIONS
Frequency. 20 cycles to 20,000 cycles in three ranges

Distortion. 0-1, -3, -10, -30, -10070full scale

Voltmeter. 0 - 1- 3 - 10 - 30 volts full scale

Input Resistance. 300 l(f,l

Minimum Input Voltage for Distortion


Measurement. 0.3 volts

ouq)ut voltage for Monitoring

Accuracy:
Voltmeter. +570of full scale
Distortion. *5% of fuU scale, +0.1%

?ube Complement. .. OA2, 6)f4, 5879, LZAT?, LZAr{I, I?BYI

Power Requirementg. . . 105-125 or 210-25Ovolts' 50/60 cycles, 30 watts

Dimensions. 13" wide x 8-l/2" high x 7" deep

Net Weight. 11 lbs

The Heath Companyreserves the right to discon-


tinue instruments and to changespecifications at
any time without incurring any obligation to
incorporate nev/ features in instruments pre-
viously sold.
}tEAl}I COMPANY ' i.i.,.
BENTOTI MICHIGAII'g}A
HARBOR.
PHON€St6-8S3-3961TWX6le0$'3097

April 17,1973

I M P O R T A N TN O T I C E
yq|r Hatmonic
your Ltlrnualbgton you start to sssomblo
Plo'le make the lollor/ing chang€sin
Di3tdtion lr€t€r.

- Lc-ftcolumn,undr '4Jvi]?sls'ine"'
ESg-l
Add: 346'2tt I Slesvin!'cloar

- Undor 'Terminal StriPs,"

D6lott: 431-11 I 5'lug$fminalsuiP

Change: 431-55 1 6-lugterminalstrip


To: 431'@ 2 6'lugterminalstrip

- Rightcolumn,undor"Miscsllaneous,"

Ad,d:421'17 1 Fuse3/4 A
Add: 390362 1 Fuselab€l

d|€ tirst lineto read:


- Rightcolumn'Change
BE-g
( ) h6talttwo l'lug, on€z'lugandono6'lu9' ' '

of the eighlhstspto read:


t'l€ first two sentenc€s
Pags15 - Rightcolumn'ChangE

( ) Form two 11" wires into a twistedpair' fush these


wire to
wiresthroughthe grommet'and conn€ctone
to lus 6 (S)on terminal
lug t (NS)and tre otrer wire
Fi6-a.
your
dr€rn o''$ trs correspondirua'€esin
Cut fi€ pordons from Pag€2 of tlis notice and tapo
Manusl.
or'er
nevvPictorisl2 {tor th€ fold'out from Page21)
Similarly,tapethe newPage7 and 16,andrhe t
Ce corresponOlng Pag€sin your Manual'
Thankyou,

HEAT}ICOMPANY

PageI of 5
003-02
rM.68/595-r
591.tr 9l
Q O
v€o

I
I
I
I

F,ffi
res+++++++
iltl ii*u'o"^*^*'* I if-;
imrtf,,ffis:ilii""^"".^^.' !,**" i
J'**'-spni'rcwrsnen
s,ti iFj 7**. i
ii
ii lIl\"-."\
lli
il'"gi+t**"li
l i iril-
l T f l * "I i,"-
ll
iI
.,ssrD----l
I tt tii i F;6a ;

ill-i---'-o--E- ii 5$::r;r:;:' i

r- ,^.
i / - \ # i r P l ii i lF".^.,,,-
c-r i ii l!
FoRroP,'ir-=---l ffi
Liir.111o_
iW- f glffiffiHi'f
'#'ffii\\
"ory119F"
ff"S;r^"J"I\\
i\\ I
ffim1ffi}"..,
t\Wtrtrlilli"'
- i i ilililL"'"
!i
d
.- o i I
i l
\\\Hl---- =il:: ii

l
ie;t
L,J-{-!
.lle=rtr- H >
: (.)
6 A NP E
t1':
I : Y
tr
" o
c

!
o
o
,

,
z
6 N
, D
o

e
=
I
z
o

Pas3ofS
lM-58/59$1@3.0i1
591-1r9t
a

t> t!
I{ 11

-l

iE N
5

o
=
u
z
o
1 2 l ?A X 7

A L L S W I I C H E S S H O W N I N C O U N T E I C L O C K I Y I S EP O S I T I O N
FFACTIOI,'AL CAPACITI€S
OIHEF CAPACITISS AiE IN MMF. UNLESS I{OTEO
t(e looo
ilEC.= IOOO,OOO
w w = w t FE w o u N o

SCHEMATICOF THE
HEATHKITO
HARMONIC DISTORTIONMETER
M O D E TI M - 5 8
sEltstTtvtTY

567e r2^r7 llfj ,.""t t241,

I GIEEX

3 O OM F D
t50v

f---;;;ilffiil-t
l l
is%r
l)- I
iI
TESTS AND CALIBRATION
\-',
Set the voltmeter calibration control H (nearest NOTE: It may be necessary to turn the sensi-
the meter) and the coarse balance control G tivity switch to lower percentage values to show
(by the power switch) at approximately the mid- the minimum more clearly.
dle position of their rotation. Set the hum bal-
ancing control F fully counterclockwise. Set With the sensitivity switch at 10070,iurn the
the RANGE switch to SET LEVEL position. Set range switch to 200-2000. (1) Turn the tuing
the SENSITIyITY switch to 10070.Set the level conirol for a minimum meter indication. (2) Ad-
control fully counterclockwise.Connectthe line iust the trimmer condenser(throughthe hole in
cord to the proper voltage outlet. DO NOT PLUG ihe top shield) with an insulated screwdriver
INTO A DC OUTLET. Like all transformer (alignment tool) for a minimum indication. Re-
operateddevices, this instrument will not oper' peat these two steps until no further improve-
ate on DC and the transformer will be damaged ment is obtained.
if connectedto a DC source.
NOTE: It may be necessary to turn the sensi-
Turn the power switch on and see if ihe pilot tivity switch to lower percentage values to show
light and tube heaters light up, The OA2 should the minimum more clearly.
show a faint purple glow inside the tube elements
after about 30 seconds. Now double cheek the setting of the coarse bal-
ance control by returning to the 2000-20,000
range as before. Finally, recheck the setting
V O L T M E T E RC A L I B R A T I O N of t-he trimmer condenser by going back to the
200-2000range again.
Set the sensitivity switch to 30 volts. Connect
a piece of wire between +}Ie"hot" input binding If the frequencypoi.nteris more than 10degrees
post and touch it to one ot the pilot light termi- off, readjust it by using a frequency of known
\ nals. Trv both terminals and use the one that accuracy. Connect this signal to the input ter-
gives an indication on the meter. Set the sen- minals and obtain a null as before. Then loosen
sitivity switch to 10volts. Observethat the meter the setscrew of the insulatedcoupling(Figure 10,
r eadingincreases. Adjust the voltm eter calibra- Page 13) and adjust the knob pointer to indicate
tion control to show a 6.3 volt reading on the the correct frequency. Make sure the insulated
0-10 volt scale. NOTE: ff a standardmeter of coupling does not turn. Retightenthe setscrew
known accuracy is available, it may be used to before continuing.
provide a more exact calibration.
HUM BALANCEADJUSTMENT

B A L A N C EA D J U S T M E N T Install the lnstrument in its case with two #6


sheet metal screws through the rear of the
cabinet. With no connections made to the input
Connect the output from an audio generator to terminal, set the range switch to 20-200, the
the input terminals of the'instrument. Set the tuning control to show about 60 on thefrequency
sensitivity switch to 10070.Set the range switch dial ind the sensitivity switch to 170.Turn the
to one of the SET LEVEL positions. Set the hum balancing control with a screwdriver to give
audio generator to 2000 cycles with at least a minimum indication on the meter.
0.5 volt output. Advancethe level control to give
a full scale meter reading.Turn the rangeswitch This minimum reading may be expectedto fall
to ihe 2000-20,000position. Set the balance around 1/10 of full scale and usually dependson
control at the middle position of its rota- the 58?9 tube heater to cathode leakage. Read-
tion. (1) Turn the tuning control for a minimum ings above 1/10 full scale may drop below that
meter indication. (2) Adjust the coarse balance value after a few hours of continuous operation.
control with a screwdriver for minimum indi-
\- cation. Repeat these two steps until no further This completes the test and calibration. The
improvement is obtained. instrument is ready for use.
P'euzo
@@l
IN CASE OF DIFFICULTY
ff any difftculty is eq)erlenced durlng test and meter, may reveal improper construction or
calibration, proceed as follows: faulty components.

1. Read the circuit description to understand 4, Frequently much time may be savedbyhav-
the principles of operation. ing a friend check the wiring with you. sur-
prisingly often a simple mistake becomes
2, Recheck the specific instructlons for the invisible to the constructor h$ is clearly
test and calibraiion procedure. evident to another, even unskilled, person.

3. Check the operating voltages against the NOTE: In an extreme case where youare unable
voltage chart. Any large discrepanciesmay to resolve a difficulty, refei to the Service and
give a clue to the source of tlte difficulty end Warranty sections of the "Kit Builders Guide",
furtier investigation in tlte particular part of and to the rrFactory Repah Service" information
the circuit, both visually and \ ith a volt- on Page 28 of this Manual.

TABLEOF TYPICALSOCKETVOLTAGES

TUBE Pin I Pin 2 Pin 3 Pin 4 Pin 5 Pin 6 Pin 7 Pin I Pin 9

58?9 0 0 1.7 H H 0 80 33 0

t2AX1 125 33 34 H 103 c.o L2.5 H

12BY? I 2 H H 98 98 o

12ATjI db -0.? 0 H 58 -0.8 0 H

oA2 145 0 145 0

6X4 220 Ac H H 220 AC 280

Heater voltage betweenpins H is 6-7 volts AC.


All DC voltages measured between pin and
chassis using a VTVM with 11 megohm input
resigtance.
P"g"zz
@EEE@r
C I R C U I TD E S C R I P T I O N
The Heathkit Harmonic Distortion Meter indi- wttHour rrroglcx ---
cates the remains of the signal under test, after /
the fundamental frequency ls eliminated, as a
percentage of the signal under test. The "indi-
cated remains" include all frequencies in the
audio range including hum, noise andharmonics
from the generator, from the equipmentunder
test and from the instrument itself.

Generally the remains are predominantly the


harmonics (multiples of the fundamental test
frequency) produced by the equipmentunder test.
The hum, noise and distortion in the instrument
h + .
are generally less than 0.170 F igure 16
The circuit may be considered to have three
parts: (a) The fundamental-suppressioncircuit. appears) the responseis still considerably atten-
(b) The voltmeter circuit. (c) Thepower supply. uated. Through use of negative feedback around
the bridge, from the l/2 12AX1cathodefollower
to the input 1/2 l2AX1 cathode,the bridge re-
-SUPPRESSION
FUNDAMENTAL C IRCUIT sponse is altered to a sharp notchat the null fre-
quencywith nearly flat response abovetwice the
This section consists of a triode voltage ampli- null frequency and below half the null frequency.
f.ier (l/2 12AX?) driving a type of phasesplitter
using a triode connectedpentode(128Y?).The The range switch selects the resistancevalues
phase splitter feedsthe Wien Bridge null network in the Wien Brldge frequeneydetermlnlngarms.
to suppress the fundamentaltest frequency. the The two intermediateswitch positions openone
signal voltages that do not cancelin the null net- arm and short out the other arm. This permits
work are amplified by a pentodevoltage ampli- the 12BY? tube to function as a cathodefollower
fier (5879),which is directly coupledto atriode only and drive the 58?9 grid directly. Then this
(1/2 l2A]{7) connected as a cathode follower. section acts as a normal amplifier having the
same gain and frequency response but without
The Wien Bridge requires two signal voltagesof the Wien Bridge null in the responsecharacter-
opposite phase at a 2:1 voltage ratio to produce istic.
a null, The plate and cathode signal voltages of
the phase splitter are of opposite phase. The
voltages are approximately 2:1 because of the In operation, the entire input signal is fed
values of the plate load and cathodeload resis- through this section to the voltmeter circuit. The
tor s. operating level of this section is adjustedwith the
input level control to give a full scale reading on
The frequency determining elements of theWien the meter.
Bridge are the resistance-capacityseries arm
and the resistance-capacity parallel arm. The Then the Wien Bridge is sr ritchedinto the circuit
resistances are equal, as are the capacities. and adjusted for elimination of the fundamental
Minor variations within the tolerance of the pre- and the remaining voltage is read on the meter
cision resistors and variations dueto mistrack- as distortion in percentage of full scale.
ing of the two-gang tuning condenser are com-
pensatedby the balancecontrol, which varies the The physical construction of the Wien Bridge,
,..
I
I
2:1 voltage ratio a small amount. such as the "hot" tuning condenserframe andthe
; - l high impedance above chassis ground level (20
I The natural response of a Wien Bridge nullnet- megohms on the 20-200 cycle range), tend to
I
work is too broad for this instrument, for at twice make the instrument subject tohum pickup. This
the null frequency (wtrere the secondharmonic inherent senbitivity is substantially reduced by
l@EE@l n"e"zt
the feedback loop around the bridge. But during t}le voltage desired as well as the source. Full
operation at or near line frequency, the null counterclockwise rotation connects t}te volt-
network opens the feedbeck loop andt}le hum re- meter to the input binding posts for observation
duction is tltus ineffective. Therefore, shieldlng of the signal level. Clockwise rotation reduces
the tuning condenser, shielding the range s$ritch, the meter reading in steps of 10 db.
insulating tlle condenserehaft, balancedfllament
wiring and a wired groundcircuit are usedto re- In the fifth position, the voltmeter is connected
duce the residual hum level in tlte instrument. to t}le cathodefollower outputofthe Fundamental
Suppression circuit. Clockwise rotation in-
VOLTMETERCIRCUIT creases the meter reading in stepsofl0db. The
frequency response at the various positions of
This section consists of a two-stage arnplifler, the sensitivity switch are compensatedwith fixed
using a 12AT7twin triode, driving a crystal diode condensers. At 1ow sensitlvity position, such
haH-bridge. Good linearity and frequency re- compensationis not required.
sponse are obtained by negatlve feedbackfrom
the hau bridge to the input catlode. The callbra-
tion control permits adjustmentof t]tefeedbackto
give tlre desired sensitivity. The output voltage POWERSUPPLY
of this two-stage amplifier is also available at
the output terminals, so the wave shapesof tbe The power supply is transformer operated. It
voltages read on the meter maybe vi8uallymoni- provides tlle filament current for all the tube
tored with an oscilloscope. heaters and the pilot light, as weu as highly fil-
tered, voltage regulated plate supply voltagefor
The input to the voltmeter is determined by ttre the instrument. A 6X4rectifier andan OA2regu-
sensitivity switch which Eelects tbe tractlon of lator are used.

APPLICATIONS
This instrument is usedprimarlly ln contunction adeguatepicture to the design or test engineer,
wit}t audio frequency slne wave generators to who is coneerned with the capabilities under
determine tlle distortion characteristics of audlo less favorable but realistic conditions. Thus not
amplifiers and componentsuged in audloampli- only a single distortion figure but families of
fiers. distortion curves are desirablefor proper evalu-
ation of equipmentunder tegt.
The distortion of a particular device under test
generally varies with output level. Often other The versatility of ttte Heathkit Harmonic Distor-
factors, such as test frequency, gain control tion Meter makes it possible to obtain adequate
setting and loading of the output may seriously data for guch curveg conveniently and rapidly.
affect the distortion readings obtalned.
NOTE: Be sure tlat youonly connectto one chan-
While frequently a single distoruon figure is nel at a time when making measurements on
quoted for a specific test frequency, at a rated stereo amplifiers. Also, if there is a speaker
output level with a standarddummyload, operat- reversal switch, be sure it is in the "Normal"
ing at nominal line voltage, thls rarely gives an position.

A U O I OC E N E R A T O F E O U I P M E N TU I I O E SI E S T osctLLoscoPE
H€AtriKtT tM-56

F igure l7
Page 24
@
|

EXAMPLEOF TESTPROCEDURE ground loop currents that will make your distor-
tion readings invalid. This applies particularly
Connect an audio generator with sine waveoutput to solid state amplifiers, and all amplifiers that
to the input of the equipmentunder test. Connect have current feedback through the speaker to
the Harmonic Distortion Meter across the output the "common" output terminal.
load of this equipment. Set the rarge switch to
SET LEVEL position. Set ttle sensitivity switch With the range switch in the SET LEVELposition,
to 100%. Adjust the level control for full scale the bridge amplifier may be usedas apreampli-
indication on the 0-10 scale. fier for the voltmeter. The level control maybe
used to select a usable sensitivity. A simple way
Reset the range switch to the range appropriate to setthe level controlfor a convenienthigh value
for the generator frequency. Adiust the tuning of sensitivity is as follows:
control for a minimum reading. Adjust the bal-
ance control for a minimum reading. Reset the Connect a generator to the input of the instrument.
sensitivity switch clockwise to provide a more Using the 1 volt scale, adjust the generator output
readable indication. Readjust tuning andbalance to 0.3 volts. Sv/itehto 10070 andadjust Ievel con-
controls for a minimum reading. Read the total trol for full scale. The instrument will nowread:
harmonic distortion on the meter andsensitivity
switch. 0.3 volts full scale on 100ft setting
0.1 volts full scale on 30fl6setting
0.03 volts full scale on 10%setting
Rotate tlle sensitivity switch counterclockwiseto 0.01 volts full scale on 370setting
a voltage position that gives a suitable meter 0.003 volts full scale on 17osetting
reading to note the input level to the distortion
meter. This voltage reading may be used to If the signal source has an impedance of 600
calculate the output power level of the equipment ohms, the 0 db marks correspondto:
under test or the reading on the db scale may be
used for comparative purposes. -10 dbm at 100%
-20 dbm at 30%
While tire majority of mdasurements will be -30 dbm at 1070
made on complete amplifiers, often effective use -40 dbm at 370
mav be made of this instrument to check por- -50 dbm at 170
tions of amplifiers during tlle development.The
effect of changed operating voltages or circuit The -10 db mark at the 170scalecorrespondsto
constants is readily observed and thus optimum -60 dbm.
operating conditions may be established. This
instrument lends itself also to other applications Noise levels are frequently expressedasanum-
such as measuring the distortion in receivers ber of db below a certain output level. Thus an
and transmission lines. with a high quality amplifier with a nominal outputof 5.4volts would
microphone and amplifier connected to the read -3 db on the red scale, using the 10 volt
instrument, the distortion introducedby a speak- position. Turning off the generator, the output
er may be determined. level would drop. Assuming the level control on
the instrum ent was adiusted as previously de-
CAUTION: The circuit ground and the case of scribed, turning sensitivity switch to 100lomakes
this instrument are both connectedto the power the range 30 db less.
line ground through the green wire of the line
cord. Always connect the ground test lead of this If a usable reading is found on the 3% scale of
device to the chassis, or ground, of the circuit +1 db for instance,thenoiselevel is then -3' +60
being tested or measured. -1 = 56 db. In this measurement, noise level is
purely a ratio of signal voltage to noise voltage
When making harmonic distortion measure- and is expressedas db,notdbm,anda statement
ments, do not use anyaccessory instruments that of reference level is generally desirable: noise
have an earth ground connection (3-wire power tevel 56 db below 3.6 watts (5.4 volts across 8
plug). Suchan earth groundconnectioncan cause ohms).
----_ Dodd ni

L[.=*ryl|l
ACCURACY DISTORTION: ?he accuracy of the distortlon
measurement depends primarily on tlle com-
pleteness of the null and the residual noise,
VOLTMETER: The full scale accuracy of tlle distortion and hum generatedwithin the ingtru-
voltmeter dependsprimarily on the exactnessof ment. The overall accuracy should faU well
tlte voltage divider resistors, which are wlthln within 5% of FULL SCALE + 0.1% distortion.
lft, and the accuracy of calibration. The accura-
cy of meter indications of less than full scale is The accuracy of the distortion measurement is
affected by linearity of the crystal diodebridge, independentof the accuracy of the calibration of
the amplifiers and ttremeter. The overall accur- the voltmeter, because the same meter sensi-
acy should fall well within 5%of FULL SCALE. tivtty is used in the SET LEVEL posttion.

R E P L A C E M E N TP A R T S P R I C EL I S T
To order parte, use the Parts Order Form fur- not availeble, refer to ReplacementParts in the
nished with this kit. u a Parts Order Form is Kit Builders Guide.

PART PRICE DESCRIPTTON PART PRICE DESCRIPTION


No. Eaclr No. Each

CONDENSERS
RESISTORS
2L-29 .15 4.7 ppf ceramic tubular
t-45 .10 220 A 20-52 .35 7.5 p,rf mica
1 -1 1 .10 1500r| 2t-7 .10 33 ppf ceramic disc
1 -1 3 .10 2700 s} 2t-ll .10 150 ppf ceramic disc
1 -1 6 .10 4?00 f} 21-31 .10 .02 pfd ceramic disc
L-20 .10 10 KO 29-28 .20 .1 p.fd paper tubular
-- t-22 .10 22 K{t 23-58 .40 .5 pfd paper tubular
t-25 .10 4? KA 23-t7 1.40 2 ufd metalite
t-26 .10 100 l(f,, 25-30 1.00 20-20 tlfd 350 v. electrolytic
t-27 .10 150 Kf, 25-38 1.50 300 pfd 150 v. electrolytic
t-29 .10 220 K{l 26-s 2.20 Dual 420 ppf tuning
1-35 .10 1 megohm 31-6 t.20 5-20 ttpf trimmer
t-40 .10 10 megohm
1-21.-r . .10 680 Q 1 watt
1-46-1 .10 27 l(f,, I watt
t-15-2 .15 1000 O 2 watt
2-50 .20 10 KSIpreclslon CONTROLS-SWITCHES
2-39 ,20 21,62 KSIpreci8ion
2-40 .20 68.38 Ktrl preclsion L0-255 .80 500 Ktr audio control
2-42 ,20 216,2 KSIprecision 11-1? .70 100 Q wire-wound control
2-45 .20 683.8 IK, precision 11-102 1,23 50 Q vrire-wound control
2-54 ,20 200 IK, precision tt-22 1.00 500 O wire-wound control
2-55 .25 2 megohmprecision 63-506 2,25 5-position switch RANGE
2-26-l 1.20 20 megohmprecislon 63-50? 2.40 g-position switch
3-1-5 .15 2500 Q 5 watt SENSITIYITY
3-8-? .15 4000 A ? watt 63-505 1.30 On-off switch