Download as pdf or txt
Download as pdf or txt
You are on page 1of 4

OpenPath detectors - photoelectric barriers for detecting

flammable gases
Wolfgang Jessel *

The idea
A thorough examination of the rela-
tively large cuvette of the Polytron IR
Ex Transmitter (see article on page 2)
might be enough to lead the
innovative engineer to imagine that
the area of monitoring could be
significantly enlarged if the light
source and the infrared detector were Receiver and
(spatially) separated (Figure 1) - not light source
just by a few metres, but by some ten,
if not several hundred metres. If there
is to be an open cuvette, it might as
well be really open!
It is a well-known fact, after all, that
according to Lambert’s and Beer’s Receiver Light source
law the longer the path interspersed Receiver Light source
with infrared-active gas, the more the up to 120 m
infrared light will be weakened: the
more absorbing molecules there are,
the stronger is the weakening effect.
Thus a Polytron Transmitter is
transformed - theoretically speaking -
into an open system, a so-called Fig. 1: From Polytron IR Ex to OpenPath detector
OpenPath system, and should ideally
be able to replace several
transmitters along a particular path.
And since it is more likely that a line probability rises proportionately to the of being able to precisely locate the
(the infrared beam) will be increase in distance between the light gas cloud.
interspersed by a gas cloud than that source and the receiver; but at the In short, a sensible compromise to en-
this cloud will hit a particular point same time so do the demands on the sure trouble-free operation of this
(the infrared transmitter), lower costs set-up, precision and stability of this type of OpenPath system would be to
in this case actually lead to an type of optical measurement path. set up a path with a length of some
improved quality of monitoring! More Especially in outdoor applications, ten to one hundred metres.
safety for less money? weather conditions such as fog, rain
A more detailed analysis, however, and snow are disruptive factors which The unit of measurement
brings the innovative engineer down increase their effect in proportion to In order to comprehend how such
to earth. It is true that the detection the length of the measurement path. OpenPath systems work, it is
Finally, the “uncertainty relation” has important to know that these gas
to be considered: the greater the warning systems are not
distance between light source and concentration measuring instruments
* Dr. W. Jessel, Dräger Sicherheitstechnik GmbH, receiver, i.e. the higher the detection of the familiar kind. It is the weakening
Lübeck probability, the lower are the chances of the infrared radiation, which is

34 Dräger Review 81 (July 1998) – W. Jessel: OpenPath detectors – photoelectric barriers for detecting flammable gases
dependent on the concentration and
Signal: 5 LELm expansion of a gas cloud, that is
50 metres measured. Small gas clouds with high
gas concentration can produce the
IR light source IR receiver same measuring signal as large gas
clouds with low gas concentration.
For example, within the monitored
path, a propane gas cloud with 100%
10 % LEL ropane LEL and an expansion of 5 metres
causes the same signal as 10% LEL
over a distance of 50 metres. In both
Signal: 5 LELm
cases, a signal of 5 LELm is recorded
(Figure 2). It could be said that an
5 metres OpenPath system counts the propane
IR light source IR receiver
molecules within the infrared beam!
The measuring unit LELm assesses
the product of concentration and
100 % LEL propane expansion of a gas cloud. Both quanti-
ties are indications of dangerous
Fig. 2: For 5 metres or 50 metres, the signal is 5 UEGm.
operational conditions, since the
violence and effect of a potential
explosion are not just dependent on
the gas concentration, but also on its
expansion (energy content and
IR light source explosion pressure are proportional
Reflector to the volume). An OpenPath system
(not to scale) detects both:
V Gas clouds of high expansion yet
IR receiver
low concentration indicate long
lasting leakages which have already
experienced dilution. In the vicinity
System GD 4001 of the leakage, flammable concen-
trations can well be expected.
IR light source IR receiver V Gas clouds of low expansion yet
high concentration already are
flammable.
Thus the unit of measurement LELm
provides evidence of the potential
danger. On a strictly symbolic level,
System GD 4012 indeed, a scaling in “expected total
cost of damages in $million“ instead
Fig. 3: There are two different versions of the Dräger OpenPath system: of a scaling in LELm would be
GD 4001 and GD 4012 conceivable.

Dräger Review 81 (July 1998) – W. Jessel: OpenPath detectors – photoelectric barriers for detecting flammable gases 35
In short, an OpenPath system is not a
concentration measuring instrument
but rather a hazard alarm instrument.

The product
Dräger offers two different versions Fig. 4: GD4001
of the OpenPath system (Figure 3).
The GD 4001 system (Figure 4) is a
“single-ended system”, i.e. the light
source and receiver are combined
and are situated opposite a reflector.
Depending on the size of the
reflector, monitoring paths of up to
20, 35 or 60 metres are possible. The
essential advantage of this type of Electrical data: 24 V DC voltage (via factory and is permanently programm-
“single-ended” system is the fact that safety barriers), analogue output ed into it. In order to check the
electrical installation has to be carried 4...20 mA (for connection to any OpenPath system, polypropylene foil
out only on one side, which is control system). can be used, which - like telescope
convenient in cases where the In short, the OpenPath system is an and adjustment aid - is included in the
monitoring path stretches over intrinsically safe detection instrument installation set. This type of foil
virtually inaccessible areas like for for flammable gases and for operation produces a signal of approx. 1 LELm.
example inshore waters. The standard with any 4...20 mA control devices. For the sceptic, a set of special
measuring range is 0 ... 5 LELm cuvettes filled with 100 % V/V
propane, i.e. 5 LELm propane Operation propane or methane is also provided.
produces a signal of 20 mA. Installation of the GD 4001 and 4012 The propane cuvette has a layer
In all other cases it is advisable to use OpenPath systems is very simple. For thickness of 8 cm and the methane
the GD 4012-system, a “double- rough adjustment a small telescope is cuvette a thickness of 20 cm. A quick
ended” system in which receiver and used, with additional fine tuning calculation reveals which signal the fil-
light source are situated opposite one through pitch control being carried led cuvettes generate in the beam
another. The monitoring path can be out using earphones. For this, an path1):
as much as 120 metres long. The (intrinsically safe, battery-driven)
standard measuring range is 0 ... 10 acoustic adjustment aid is used, Propane: LEL = 2.0 % V/V,
LELm propane. which is temporarily plugged in to the thus 100% V/V = 50 LEL,
Since the GD 4001 and the GD 4012 infrared receiver. 50 LEL x 0.08 m = 4 LELm.
systems were specially developed for Automatic zero point calibration is Methane: LEL = 5.0 % V/V,
the offshore industry, both are then initiated by installing a module thus 100% V/V = 20 LEL,
designed for continuous operation in instead of the adjustment aid: the 20 LEL x 0.20 m = 4 LELm.
the roughest surroundings: impact OpenPath receiver is told, as it were,
resistant special plastic, protection that the actual infrared intensity is in
class IP 66, intrinsically safe as per future to correspond to the zero point. 1
This is based on the standard US lower explosive
EEx ib e IIB T4 (CENELEC), A sensitivity calibration using gas is limits for propane and methane used in offshore
applications. According to the IEC standard 79-20,
operational temperature range from not performed. This is already carried the lower explosive limits are 1.7 % V/V and 4.4 % V/V
–30 to +60 °C. out before the instrument leaves the respectively, producing signals of 4.7 and 4.5 LELm.

36 Dräger Review 81 (July 1998) – W. Jessel: OpenPath detectors – photoelectric barriers for detecting flammable gases
Naturally, any optical measuring followed up by action: the leakage because the so-called primary explosi-
system can only function if the beam must be located and eliminated. on protection2) is based on the local
of light hits the receiver without Depending on the size of the fenced concentration measurement. Also the
interference. off area, a more or less extensive fact that the OpenPath systems
Therefore it must be ensured that the search begins, which calls for quick detect far fewer substances than the
monitoring path of the OpenPath and effective counter-measures. It is point detector - which is capable of
system is free of objects that could clear that in view of the size of the detecting even heavy flammable
interfere with the infrared beam. fenced off areas a sensible com- vapours reliably – will continue to
Short interruptions of up to 10 promise is needed. make infrared transmitters
seconds can be tolerated; longer An interesting advantage of an indispensable.
interruptions, however, are reported OpenPath system lies in the fact that It is rather the case that both methods
by a special status signal (2 mA). This this type of system can even detect – the OpenPath systems on the one
so-called “beam block” alarm leakages resulting from hairline hand and the point detectors on the
indicates that the OpenPath system cracks on containers. The gas other – have proved that they have a
is, at least temporarily, not opera- concentrations which escape very right to exist and that they do not
tional, and that monitoring is not slowly have time to become rarefied crowd the other out. The opposite is
guaranteed. In order to maximize the and to form a large cloud of low indeed the case – a good safety
availability of the OpenPath system, concentration. If this cloud then hits a concept is possible precisely
the monitoring path should be point detector (for example, a catalytic because both methods of measure-
arranged so that hardly any “beam combustion sensor), the instrument ment are used! Hence the GD 4001
block” can occur. The central device will not detect the leakage due to the and GD 4012 OpenPath systems are
used should be able to distinguish a fact that the concentration is far too the ideal complement to our Polytron
system fault (current < 1 mA) from a low. The OpenPath detector, on the transmitters and round off a gas
“beam block“ alarm (current = 2 mA), other hand, is perfectly well able to detection system.
since initially each have different detect the cloud. As a rule, the
priorities in the safety concept. Only a reading will be the same regardless
long-lasting “beam block” should be of the degree of rarefaction of a gas
considered as tantamount to a system cloud within the monitoring path,
fault. since the number of gas molecules
In outdoor applications, fog must be remains constant to all intents and
regarded as the main disruptive purposes along this path.
factor. However, tests have shown The OpenPath system is used for the
that the OpenPath system continues detection of flammable gases such as
to work flawlessly even when fog methane, ethane, propane or butane.
conceals the light source from the In principle, even the vapours of
human eye. This is due to the high flammable liquids (for example,
dynamic reserves of the system that pentane, hexane, methanol, ethanol,
accept a weakening in intensity of up or ethyl acetate) can be detected;
to 95 percent without generating a though leakages of heavy vapors (e.g.
“beam block“ alarm. xylene, nonane, methyl pyrrolidine)
Vibrations, too, can disrupt the are not detectable, since an average
system: if the light source is turned of more than 100 % LEL along each
just 1 mm on a 100 m path, this leads metre is necessary to trigger the
to an inaccuracy of half a metre! alarm (the recommended threshold
Although this calculation is idealized value being 1 LELm). However, many
(fortunately the infrared beam is not flammable liquids are not able to
linear but conical) it nevertheless generate such concentrations on
illustrates how precise the adjustment account of their evaporation
and how stable the fixation must be. properties. Due to convection, heavy
In short, an OpenPath system offers vapours (which have a relatively high
high availability as long as the optical flash point of > 30 °C) are carried off
path is not disrupted. more quickly than they can be
replaced by evaporation, and this is
Practical experience particularly true of outdoor
To think of an OpenPath detector as applications.
an (analogue) photoelectric barrier for
flammable gases is not all that far off Conclusion
the mark. A photoelectric barrier, after When planning to use and when
all, also says nothing about the loca- using OpenPath systems, the focus
tion of the detected object but simply should never be on trying to save
indicates its presence. So-called money by replacing several point
fence-monitoring, i.e. limiting a detectors. Although this is a frequent
danger zone with one or more claim, OpenPath systems will not 2
Primary explosion protection with gas warning
OpenPath systems, does justice to make point detectors redundant in the instruments is ensured by means of concentration
these facts, since it reports the cross- future, especially not infrared trans- measurement and the deployment of counter
measures (e.g. ventilation, eliminating ignition
ing of flammable gas from one system mitters, because the application of sources) in the event of a predefined concentration
to another. Naturally, this must be these two products differs greatly and (e.g. 20% LEL) being exceeded.

Dräger Review 81 (July 1998) – W. Jessel: OpenPath detectors – photoelectric barriers for detecting flammable gases 37

You might also like