Professional Documents
Culture Documents
TEE For Impella
TEE For Impella
Special Article
Transesophageal Echocardiography for Impella
Placement and Management
Jerome Crowley, MD*, Brett Cronin, MDy,
Michael Essandoh, MDz, David D’Alessandro, MD1 x,
Ken Shelton, MD*, Adam A. Dalia, MD, MBA*,
*
Department of Anesthesiology, Pain Medicine, and Critical Care, The Massachusetts General Hospital,
Harvard Medical School, Boston, MA
y
Department of Anesthesiology, University of California San Diego, UCSD School of Medicine, San Diego, CA
z
Department of Anesthesiology, The Ohio State University Medical Center, Columbus, OH
x
Department of Cardiothoracic Surgery, The Massachusetts General Hospital, Harvard Medical School,
Boston, MA
Key Words: transesophageal echocardiography; TEE; intraoperative echocardiography; Impella; percutaneous left ventricular assist device; procedural
TEE guidance
ACUTE CARDIOGENIC SHOCK is a common problem Model Types, Features, and Indications
encountered by cardiac anesthesiologists. Current methods of
treatment include pharmacologic and mechanical circulatory The Impella heart pump is a miniaturized ventricular assist
support (MCS). MCS is used for temporary or long-term treat- device that consists of an Archimedes screw pump inserted
ment of cardiogenic shock. Temporary MCS is used as a bridge across the aortic valve and provides continuous drainage of
to myocardial recovery, a bridge to a more durable therapy blood from the left ventricle and reinfuses the blood into the
(durable MCS or heart transplantation), or a bridge to decision- ascending aorta regardless of left ventricular contractility.
making.1 Temporary MCS is becoming more common in car- There are several types of Impella devices that vary based on
diogenic shock in multiple scenarios including postcardiotomy insertion site and maximal flow rate. The Impella 2.5 (Fig 1)
shock, high-risk percutaneous cardiac intervention, acute and the Impella CP are placed most commonly percutaneously
decompensated heart failure, and ventricular tachycardia abla- via the femoral artery and are approved for a maximum dura-
tion, and for percutaneous venting of the left ventricle during tion of use of 4 days; maximum flow rates for the Impella 2.5
peripheral venoarterial extracorporeal membrane oxygen- and Impella CP are 2.5 L/min and 3.3 L/min, respectively.4
ation.2,3 The left-sided Impella (Abiomed Inc., Danvers, MA) The Impella 5.0 is placed most often by surgical cutdown to
heart pump is a commonly used temporary MCS device. This an artery and the Impella LD is placed directly into the aorta
review describes the Impella and provides guidance in the use during surgical exposure, and they are approved for a maxi-
of transesophageal echocardiography in its placement. Informa- mum duration of use of 6 days with maximum flow rates of
tion about troubleshooting deteriorating hemodynamics when 5 L/min.5 For right ventricular failure, the Impella RP can be
an Impella is being used is provided as well. used to drain blood from the right ventricle and reinfuse into
the pulmonary artery to aid right ventricular recovery but is
1
Address reprint requests to Adam A. Dalia, MD, MBA, Division of Cardiac not the focus of this article.
Anesthesiology, Department of Anesthesia, Critical Care and Pain Medicine, Despite differences in insertion technique and maximal
The Massachusetts General Hospital, Harvard Medical School, 55 Fruit St, blood flow support provided by the different Impellas, the final
Boston, MA 02114.
E-mail address: aadalia@mgh.harvard.edu (A.A. Dalia).
placement of the device is very similar. The device consists of
https://doi.org/10.1053/j.jvca.2019.01.048
1053-0770/Ó 2019 Elsevier Inc. All rights reserved.
2664 J. Crowley et al. / Journal of Cardiothoracic and Vascular Anesthesia 33 (2019) 26632668
Fig 2. Midesophageal long-axis view of Impella device. LA, left atrium; LV, left ventricle.
2666 J. Crowley et al. / Journal of Cardiothoracic and Vascular Anesthesia 33 (2019) 26632668
arrhythmias owing to device abutment of the ventricular sep- into the aorta will result in a lack of left ventricular support
tum. Advancement of the Impella too far into the left ventricle (the ventricle will not be unloaded) despite adequate flows,
may result in arrhythmias or direct myocardial injury causing thus making TEE guidance crucial during the manipulation
left ventricular puncture. Excessive withdrawal of the device of the Impella heart pump.
Fig 4. Midesophageal long-axis view of Impella device outlet with color Doppler. LA, left atrium; LV, left ventricle.
J. Crowley et al. / Journal of Cardiothoracic and Vascular Anesthesia 33 (2019) 26632668 2667
Fig 5. Transgastric midpapillary view with x-plane of Impella device free from papillary or subvalvular mitral apparatus. LV, left ventricle.
Table 2
dynamic process affected by the level of anesthesia, volume
Correct Position of the Impella and Corresponding TEE Views shifts, and pharmacologic intervention (Table 3). Finally, TEE
is useful during the MCS weaning process and may reveal
Device inlet 3.5 cm below the aortic valve (midesophageal aortic valve evidence of myocardial recovery.
long-axis view)
Device outlet above the aortic valve (midesophageal aortic valve long-axis
view) Conclusion
Device angled away from the mitral valve toward the left ventricular apex
(transgastric long-axis left ventricular view or midesophageal aortic valve
long-axis view)
Temporary MCS devices such as the Impella are becoming
more common owing to benefits in decreased myocardial oxy-
Abbreviations: TEE, transesophageal echocardiography. gen consumption compared with pharmacologic support.10 It is
critical that cardiac anesthesiologists are familiar with the func-
Table 3 tion of these devices as well as the important role that TEE
Improper Positions of the Impella Device plays in placing and managing these devices. As these devices
mature, more procedures may be done with Impella heart pump
Device too far into left ventricle (inlet >3.5 cm from aortic valve or outlet
across the aortic valve) support, particularly temporary support for high-risk patients.
Outlet too far into ascending aorta (outlet area well above the aortic valve) Understanding how to guide placement and troubleshoot these
Device pigtail in the subvalvular apparatus of the mitral valve or papillary devices using echocardiography certainly will be a skill set
muscle required of cardiac anesthesiologists in the future.
Device angled away from the left ventricular apex
Supplementary materials
Although less common, suction events may occur if the ven- Supplementary material associated with this article can be
tricle is decompressed excessively owing to improved native found in the online version at doi:10.1053/j.jvca.2019.01.048.
ventricular function, a sign that the patient no longer may need
ventricular support with the Impella. Although not diagnosed
References
by TEE but rather laboratory findings, extended use of the
Impella heart pump also can lead to anemia or hemolysis, as is
1 Naidu SS. Novel percutaneous cardiac assist devices: The science of and
seen with durable left ventricular assist devices. indications for hemodynamic support. Circulation 2011;23:533–43.
Changes in patient status or device function perioperatively 2 Fiedler AG, Dalia A, Axtell AL, et al. Impella placement guided by echo-
should prompt investigation via TEE, as this is commonly a cardiography can be used as a strategy to unload the left ventricle during
2668 J. Crowley et al. / Journal of Cardiothoracic and Vascular Anesthesia 33 (2019) 26632668
peripheral venoarterial extracorporeal membrane oxygenation. J Cardio- 7 Hong E, Naseem T. Coiled Impella drive line in the left ventricle: A rare
thorac Vasc Anesth 2018. complication of a left ventricular assist. J Cardiothorac Vasc Anesth 2018;
3 Higgins J, Lamarche Y, Kaan A, et al. Microaxial devices for ventricular In press.
failure: A multicentre, population-based experience. Can J Cardiol 2011; 8 Pantin EJ, Chyu D, Mungekar SS, et al. Coiled Impella drive line in the left
27:725–30. ventricle: A rare complication of a left ventricular assist device. J Cardio-
4 A.B.I.O.M.E.D. Available at: http://www.abiomed.com/impella. Accessed thorac Vasc Anesth 2015;29:1308–10.
December 3, 2018. 9 Elhussein TA, Hutchison SJ. Acute mitral regurgitation: Unforeseen new
5 Griffith BP, Anderson MB, Samuels LE, et al. The RECOVER I: A multi- complication of the Impella LP 5.0 ventricular assist device and review of
center prospective study of Impella 5.0/LD for postcardiotomy circulatory literature. Heart Lung Circ 2014;23:e100–4.
support. J Thorac Cardiovasc Surg 2013;145:548–54. 10 Flaherty, et al. Hemodynamic support with a microaxial percutaneous left
6 Patel KM, Sherwani SS, Baudo AM, et al. Echo rounds: The use of transe- ventricular assist device (Impella) protects against acute kidney injury in
sophageal echocardiography for confirmation of appropriate Impella 5.0 patients undergoing high-risk percutaneous coronary intervention. Circ
device placement. Anesth Analg 2012;114:82–5. Res 2017;120:692–700.