Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

Journal of Cardiothoracic and Vascular Anesthesia 33 (2019) 26632668

Contents lists available at ScienceDirect

Journal of Cardiothoracic and Vascular Anesthesia


journal homepage: www.jcvaonline.com

Special Article
Transesophageal Echocardiography for Impella
Placement and Management
Jerome Crowley, MD*, Brett Cronin, MDy,
Michael Essandoh, MDz, David D’Alessandro, MD1 x,
Ken Shelton, MD*, Adam A. Dalia, MD, MBA*,
*
Department of Anesthesiology, Pain Medicine, and Critical Care, The Massachusetts General Hospital,
Harvard Medical School, Boston, MA
y
Department of Anesthesiology, University of California San Diego, UCSD School of Medicine, San Diego, CA
z
Department of Anesthesiology, The Ohio State University Medical Center, Columbus, OH
x
Department of Cardiothoracic Surgery, The Massachusetts General Hospital, Harvard Medical School,
Boston, MA

Key Words: transesophageal echocardiography; TEE; intraoperative echocardiography; Impella; percutaneous left ventricular assist device; procedural
TEE guidance

ACUTE CARDIOGENIC SHOCK is a common problem Model Types, Features, and Indications
encountered by cardiac anesthesiologists. Current methods of
treatment include pharmacologic and mechanical circulatory The Impella heart pump is a miniaturized ventricular assist
support (MCS). MCS is used for temporary or long-term treat- device that consists of an Archimedes screw pump inserted
ment of cardiogenic shock. Temporary MCS is used as a bridge across the aortic valve and provides continuous drainage of
to myocardial recovery, a bridge to a more durable therapy blood from the left ventricle and reinfuses the blood into the
(durable MCS or heart transplantation), or a bridge to decision- ascending aorta regardless of left ventricular contractility.
making.1 Temporary MCS is becoming more common in car- There are several types of Impella devices that vary based on
diogenic shock in multiple scenarios including postcardiotomy insertion site and maximal flow rate. The Impella 2.5 (Fig 1)
shock, high-risk percutaneous cardiac intervention, acute and the Impella CP are placed most commonly percutaneously
decompensated heart failure, and ventricular tachycardia abla- via the femoral artery and are approved for a maximum dura-
tion, and for percutaneous venting of the left ventricle during tion of use of 4 days; maximum flow rates for the Impella 2.5
peripheral venoarterial extracorporeal membrane oxygen- and Impella CP are 2.5 L/min and 3.3 L/min, respectively.4
ation.2,3 The left-sided Impella (Abiomed Inc., Danvers, MA) The Impella 5.0 is placed most often by surgical cutdown to
heart pump is a commonly used temporary MCS device. This an artery and the Impella LD is placed directly into the aorta
review describes the Impella and provides guidance in the use during surgical exposure, and they are approved for a maxi-
of transesophageal echocardiography in its placement. Informa- mum duration of use of 6 days with maximum flow rates of
tion about troubleshooting deteriorating hemodynamics when 5 L/min.5 For right ventricular failure, the Impella RP can be
an Impella is being used is provided as well. used to drain blood from the right ventricle and reinfuse into
the pulmonary artery to aid right ventricular recovery but is
1
Address reprint requests to Adam A. Dalia, MD, MBA, Division of Cardiac not the focus of this article.
Anesthesiology, Department of Anesthesia, Critical Care and Pain Medicine, Despite differences in insertion technique and maximal
The Massachusetts General Hospital, Harvard Medical School, 55 Fruit St, blood flow support provided by the different Impellas, the final
Boston, MA 02114.
E-mail address: aadalia@mgh.harvard.edu (A.A. Dalia).
placement of the device is very similar. The device consists of

https://doi.org/10.1053/j.jvca.2019.01.048
1053-0770/Ó 2019 Elsevier Inc. All rights reserved.
2664 J. Crowley et al. / Journal of Cardiothoracic and Vascular Anesthesia 33 (2019) 26632668

discussion with the proceduralist, because other types of


mechanical support may be suited better; a full list of contrain-
dications can be found in Table 1.4

TEE Guidance for Impella Heart Pump Placement

Implantation of the Impella heart pump is dictated by the


model (2.5, 5.0, CP) as described above. For the Impella 2.5,
5.0, and CP, the vessel of choice (commonly femoral or axil-
lary) is accessed and a sheath is placed. Next, a combination
of wires and catheters are used to cross the aortic valve to
serve as a conduit for the Impella heart pump. The descending
thoracic aorta should be interrogated in the aortic short and
long axes to confirm wire placement in the aorta as well as to
rule out aortic dissection from placement. Next, the mideso-
phageal aortic valve long-axis view can be used to visualize
the wire crossing the aortic valve and entering the left ventri-
cle. If the wire is in too deep in the left ventricular cavity, it
may cause ventricular arrhythmias; it is also important to
avoid having the wire cross the mitral valve, as inappropriate
wire placement will lead to inappropriate Impella placement.
In the case of mitral valve tethering, the guidewire will need to
be withdrawn and repositioned. Increases in aortic insuffi-
ciency (significant increase may represent aortic valve injury)
and changes in mitral valve function (the wire may be interfer-
ing with the subvalvular apparatus) should be monitored and
discussed with the proceduralists to reduce complications.
Fig 1. Impella 2.5 device labeled.
Once the wire is seen crossing the aortic valve, a midesopha-
geal short-axis view of the aortic valve can be used to rule out
perforation or significant tethering of the aortic leaflets, as the
3 major components: a screw pump that provides continuous wire will need to be withdrawn and repositioned.7 Images of the
blood flow, a flexible pigtail that extends from the distal end ascending aorta and aortic arch should be obtained to rule out fur-
of the pump, and a drive line that connects the motor to the ther aortic dissection. After the wire is confirmed to be in the
controller outside the patient. The pump has to be positioned proper location, the proceduralist then will place the Impella heart
appropriately to function properly; that is, the inlet of the pump over the wire; at this point, the best view to guide place-
pump must be in the left ventricle and the outlet must be in the ment is the midesophageal aortic valve long-axis view. If there is
ascending aorta. The initial placement of the Impella com- trouble advancing the device, views of the descending thoracic
monly is done under fluoroscopic and transesophageal aorta and the aortic arch should be obtained to aid in trouble-
echocardiography guidance; additionally, troubleshooting a shooting, because the Impella heart pump may be stuck in a
malfunctioning Impella heart pump is a scenario where TEE branch vessel or may be kinking the wire upon itself. After the
commonly is used. Although fluoroscopy can be helpful in device is visualized across the aortic valve, the wire is withdrawn
identifying the general location of the Impella in vitro, it can- and final appropriate positioning confirmed by TEE. The pigtail
not detect its position in relation to important anatomic cardiac at the end of the motor is not always visible on TEE imaging,
landmarks such as the subvalvular apparatus of the mitral but attempts to locate it in the midesophageal aortic valve
valve or the distance from the aortic valve to the Impella inlet; long-axis view can be helpful to ensure it is not abutting the
thus, TEE is a valuable tool for placement.6
It is also important that anesthesiologists have an under- Table 1
standing of the indications and contraindications for Impella Left-Sided Impella (2.5, 5.0, CP) Contraindications
heart pump placement. Indications for the Impella heart pump
Biventricular failure
include reversible myocardial ischemia, acute severe left ven-
Cardiac tamponade
tricular failure, and ventricular tachycardia ablation, or use Left ventricular rupture
during high-risk percutaneous cardiac intervention in patients Mechanical aortic valve
with severe coronary artery disease and reduced ejection frac- Mural thrombus in left ventricle
tions.4 Biventricular failure, significant aortic insufficiency, ASD or VSD
Severe aortic insufficiency
left ventricular thrombus, mechanical aortic valve, and large
Aortic dissection
intracardiac shunts may represent contraindications to Impella
heart pump placement and at the very least require additional Abbreviations: ASD, atrial septal defect; VSD, ventricular septal defect.
J. Crowley et al. / Journal of Cardiothoracic and Vascular Anesthesia 33 (2019) 26632668 2665

intraventricular septum (increased risk of arrhythmias) or entan- Troubleshooting


gling in the subvalvular apparatus of the mitral valve8 (Video 1).
The inlet to the motor, which is visualized as a teardrop shape It is important to remember that cardiogenic shock is a
in the midesophageal aortic valve long-axis view, should be in dynamic process that is highly dependent on the current state
the left ventricle 3.5 cm from the aortic valve (Fig 2). Color and of biventricular function, volume status, and adjunctive phar-
spectral Doppler imaging can be helpful in identifying the loca- macologic therapy, and TEE can be used to verify some of
tion of the inlet, as can 3D echocardiography (Video 2, Fig 3). these acute changes. A common evaluation involves patients
Once the inlet is seen in the appropriate position, then the outlet who are failing MCS with evidence of worsening multiorgan
of the motor in the ascending aorta also can be visualized in the perfusion. Using a midesophageal 4-chamber view, the anes-
midesophageal long-axis view with the aid of color Doppler thesiologist can observe for new causes of shock, such as acute
(Video 3, Fig 4). The transgastric long- or short-axis view is right ventricular failure. Interventricular septal position is also
also helpful to confirm appropriate positioning of the Impella an important indicator of optimal Impella heart pump place-
heart pump in the ventricle as well as to verify further no entan- ment and function and should be interrogated. In the setting of
glement with the mitral valve apparatus (Video 4, Fig 5). Proper a well-functioning Impella, the interventricular septum should
positioning includes the Impella heart pump curved away from be close to midline on TEE evaluation. Deviation of the inter-
the mitral valve facing the left ventricular apex. ventricular septum to the right implies worsening left ventricu-
After the Impella heart pump is identified as being in the cor- lar decompression that can be treated with increasing the
rect position, the function of the mitral valve and aortic valve Impella pump speed and/or inotropes. Deviation of the inter-
again should be interrogated (Video 5). Color Doppler interro- ventricular septum to the left implies inadequate filling of the
gation of the mitral valve to evaluate for the presence of or left ventricle, which may be owing to right ventricular dys-
worsening of mitral regurgitation related to device malposition function (treated with inotropes or right ventricular support) or
is useful; attention should be paid to any leaflet tethering that suctioning of the left ventricle by high Impella pump speed
may be owing to the Impella heart pump.9 The aortic valve is and can be treated by reducing the Impella pump speed.
difficult to image owing to device artifact; however, severe aor- New valvular lesions (aortic or mitral) detected on TEE may
tic insufficiency can be suggestive of valve injury, so visualiz- represent iatrogenic injury from the Impella device and may
ing the device crossing the aortic valve in short axis and provide an explanation for the patient’s worsening status.
observing leaflet motion are helpful (Video 6). The aorta should New-onset severe mitral regurgitation could be owing to sub-
be imaged at the end of device placement to ensure further no valvular apparatus injury from device migration. Ventricular
iatrogenic dissection occurred. The pericardium also should be arrhythmias commonly are owing to irritation of the ventricu-
assessed for any evidence of a new pericardial effusion, which lar septum by migration of the Impella, and position should be
may represent ventricular perforation. The correct Impella heart verified as described above. Improved left ventricular contrac-
pump final position is described in Table 2. tility also can cause increased ventricular ectopy or result in

Fig 2. Midesophageal long-axis view of Impella device. LA, left atrium; LV, left ventricle.
2666 J. Crowley et al. / Journal of Cardiothoracic and Vascular Anesthesia 33 (2019) 26632668

Fig 3. 3D midesophageal long-axis view of Impella device.

arrhythmias owing to device abutment of the ventricular sep- into the aorta will result in a lack of left ventricular support
tum. Advancement of the Impella too far into the left ventricle (the ventricle will not be unloaded) despite adequate flows,
may result in arrhythmias or direct myocardial injury causing thus making TEE guidance crucial during the manipulation
left ventricular puncture. Excessive withdrawal of the device of the Impella heart pump.

Fig 4. Midesophageal long-axis view of Impella device outlet with color Doppler. LA, left atrium; LV, left ventricle.
J. Crowley et al. / Journal of Cardiothoracic and Vascular Anesthesia 33 (2019) 26632668 2667

Fig 5. Transgastric midpapillary view with x-plane of Impella device free from papillary or subvalvular mitral apparatus. LV, left ventricle.

Table 2
dynamic process affected by the level of anesthesia, volume
Correct Position of the Impella and Corresponding TEE Views shifts, and pharmacologic intervention (Table 3). Finally, TEE
is useful during the MCS weaning process and may reveal
Device inlet 3.5 cm below the aortic valve (midesophageal aortic valve evidence of myocardial recovery.
long-axis view)
Device outlet above the aortic valve (midesophageal aortic valve long-axis
view) Conclusion
Device angled away from the mitral valve toward the left ventricular apex
(transgastric long-axis left ventricular view or midesophageal aortic valve
long-axis view)
Temporary MCS devices such as the Impella are becoming
more common owing to benefits in decreased myocardial oxy-
Abbreviations: TEE, transesophageal echocardiography. gen consumption compared with pharmacologic support.10 It is
critical that cardiac anesthesiologists are familiar with the func-
Table 3 tion of these devices as well as the important role that TEE
Improper Positions of the Impella Device plays in placing and managing these devices. As these devices
mature, more procedures may be done with Impella heart pump
Device too far into left ventricle (inlet >3.5 cm from aortic valve or outlet
across the aortic valve) support, particularly temporary support for high-risk patients.
Outlet too far into ascending aorta (outlet area well above the aortic valve) Understanding how to guide placement and troubleshoot these
Device pigtail in the subvalvular apparatus of the mitral valve or papillary devices using echocardiography certainly will be a skill set
muscle required of cardiac anesthesiologists in the future.
Device angled away from the left ventricular apex

Supplementary materials

Although less common, suction events may occur if the ven- Supplementary material associated with this article can be
tricle is decompressed excessively owing to improved native found in the online version at doi:10.1053/j.jvca.2019.01.048.
ventricular function, a sign that the patient no longer may need
ventricular support with the Impella. Although not diagnosed
References
by TEE but rather laboratory findings, extended use of the
Impella heart pump also can lead to anemia or hemolysis, as is
1 Naidu SS. Novel percutaneous cardiac assist devices: The science of and
seen with durable left ventricular assist devices. indications for hemodynamic support. Circulation 2011;23:533–43.
Changes in patient status or device function perioperatively 2 Fiedler AG, Dalia A, Axtell AL, et al. Impella placement guided by echo-
should prompt investigation via TEE, as this is commonly a cardiography can be used as a strategy to unload the left ventricle during
2668 J. Crowley et al. / Journal of Cardiothoracic and Vascular Anesthesia 33 (2019) 26632668

peripheral venoarterial extracorporeal membrane oxygenation. J Cardio- 7 Hong E, Naseem T. Coiled Impella drive line in the left ventricle: A rare
thorac Vasc Anesth 2018. complication of a left ventricular assist. J Cardiothorac Vasc Anesth 2018;
3 Higgins J, Lamarche Y, Kaan A, et al. Microaxial devices for ventricular In press.
failure: A multicentre, population-based experience. Can J Cardiol 2011; 8 Pantin EJ, Chyu D, Mungekar SS, et al. Coiled Impella drive line in the left
27:725–30. ventricle: A rare complication of a left ventricular assist device. J Cardio-
4 A.B.I.O.M.E.D. Available at: http://www.abiomed.com/impella. Accessed thorac Vasc Anesth 2015;29:1308–10.
December 3, 2018. 9 Elhussein TA, Hutchison SJ. Acute mitral regurgitation: Unforeseen new
5 Griffith BP, Anderson MB, Samuels LE, et al. The RECOVER I: A multi- complication of the Impella LP 5.0 ventricular assist device and review of
center prospective study of Impella 5.0/LD for postcardiotomy circulatory literature. Heart Lung Circ 2014;23:e100–4.
support. J Thorac Cardiovasc Surg 2013;145:548–54. 10 Flaherty, et al. Hemodynamic support with a microaxial percutaneous left
6 Patel KM, Sherwani SS, Baudo AM, et al. Echo rounds: The use of transe- ventricular assist device (Impella) protects against acute kidney injury in
sophageal echocardiography for confirmation of appropriate Impella 5.0 patients undergoing high-risk percutaneous coronary intervention. Circ
device placement. Anesth Analg 2012;114:82–5. Res 2017;120:692–700.

You might also like