Professional Documents
Culture Documents
Randox RANSOD PDF
Randox RANSOD PDF
Cat. No. Health and Safety data sheets are available on request.
SD 125 R1a. Mixed Substrate 5 x 20 ml
5 x 20 ml R1b. Buffer 1 x 105 ml The reagents must be used only for the purpose
R2. Xanthine Oxidase 3 x 10 ml intended by suitably qualified laboratory personnel,
CAL Standard 5 x 10 ml under appropriate laboratory conditions.
REAGENT COMPOSITION
PROCEDURE CALCULATION
Select SOD in the Run Test screen and carry out a water A2 - A1
blank as instructed. = −A/min of standard or sample
Pipette into a cuvette: 3
Standard Standards Samples Control
S1 S2 TO S6 Sample diluent rate (S1 rate) = rate of uninhibited reaction
Ransod Sample Diluent 30 µl - - - = 100 %
Standard - 30 µl - -
Diluted Sample - - 30 µl -
Diluted Control - - - 30 µl All standard rates and diluted sample rates must be
Mixed substrate (R1) 1000 µl 1000 µl 1000 µl 1000 µl converted into percentages of the sample diluent rate, and
subtracted from 100 % to give a percentage inhibition.
Mix well and add :-
(−Astd/min x 100)
Xanthine Oxidase (R2) 150 µl 150 µl 150 µl 150 µl 100 - = % inhibition
(−AS1/min)
Mix well, insert the cuvette into the Rx Monza cell holder
and press read. (−Asample/min x 100)
100 - = % inhibition
CALIBRATION (−AS1/min)
Calibrate this assay using the standard supplied with the kit.
It is recommended that the following dilutions are made of Plot percentage inhibition for each standard against Log10
the Standard CAL (or S6) to produce a standard curve:- (standard conc. in SOD units/ml)
Pipette sample and R1 into a cuvette and mix well. Add R2, (b) Converting to SOD units/ml of whole blood
mix and read initial absorbance A1 after 30 seconds and
start timer simultaneously read final absorbance A2 after 3 0.575 x 300 = 172.5 SOD units/ml
minutes.
(c) Converting to SOD units/g Haemoglobin
172.5
(SOD units/ml)/(g Haemoglobin/ml) = = 1461.9
0.118
QUALITY CONTROL
Ransod Control is recommended for daily quality control. The
control should be assayed at least once a day. Values
obtained should fall within a specified range. If these values
fall outside the range and repetition excludes error, the
following steps should be taken:
1. Check instrument settings and light source.
2. Check cleanliness of all equipment in use.
3. Check water, contaminants i.e. bacterial growth may
contribute to inaccurate results.
4. Check reaction temperature.
5. Check expiry date of kit and contents.
6. Contact Randox Laboratories Customer Technical
Support, Northern Ireland (028) 94422413.
RANDOX Laboratories Ltd., 55 Diamond Road, Crumlin, Co. Antrim, United Kingdom, BT29 4QY
ΑΒΧ
Tel: +44 (0) 28 9442 2413 Fax: +44 (0) 28 9445 2912
Email: applications@randox.com Website: www.randox.com
MANUAL/ RX MONZA - RANSOD - SD 125 PAGE 3 OF 3
NORMAL RANGES
1102 - 1601 U/g Hb
164 - 240 U/ml
LINEARITY
Samples should be diluted to give an inhibition between
30 % and 60 % of the sample diluent rate (i.e. the
uninhibited reaction).
SENSITIVITY
The minimum detectable concentration of SOD less than
standards should be reported as <S1 standard value.
PRECISION
CORRELATION
The Randox method on the Rx Monza (Y) was compared to
the Rx Daytona (X) and the following linear regression
equation was obtained:
Y = 0.9417 X + 0.1004
and a correlation coefficient of 0.965
REFERENCES
1. Woolliams JA, Wiener G, Anderson PH, McMurray
CH Research in Veterinary Science 1983, 34: 253-256.
2. Suttle NF, The Veterinary Record 1986, 119: 519-522.
3. Suttle NF, McMurray CH Research in Veterinary
Science 1983; 35: 47-52.
4. Arthur JR, Boyne R Life Sciences 1985 36: 1569-1575.
Revised 07 Oct 09 ck
RANDOX Laboratories Ltd., 55 Diamond Road, Crumlin, Co. Antrim, United Kingdom, BT29 4QY
ΑΒΧ
Tel: +44 (0) 28 9442 2413 Fax: +44 (0) 28 9445 2912
Email: applications@randox.com Website: www.randox.com