Flight 60® iO2

Quick Reference Guide

Alarms & Power Management

Patient Monitoring

Controlled Parameters

Alarm Reset
ON/OFF This quick reference guide does not replace the Flight
60 Operator’s Manual. The Flight 60 Operator’s Manual
containing instructions, warnings, cautions and other
important information for the safe and proper use of this
Nebulizer Port product must be carefully read prior to
Patient Circuit Connectors using the Flight60 ventilator.
Circuit TEST
A Circuit Test is to be performed each time a patient circuit or components
of a patient circuit are replaced:

1. Connect an adult (500ml) test lung with a 90 degree


elbow or occlude the patient side of the circuit.

2. Press the Technical button once, and then


tap the Circuit TEST button.

3. Press Enter

2
Starting Ventilation
Turn on the ventilator by pressing
the On/Off button.

Start ventilation by pressing and holding


the On/Off button for 2 seconds.

Stopping Ventilation/
Shutting Down
Press the On/Off button once, then
again for 2 seconds.

3
Parameters Set Up

Trigger (sensitivity) Pressure Support Volume/Pressure


setting setting Setting

Ptrig PEEP PSV above peep Rate Ti PCV above peep

-2.0 5 10 14 1.4 10
Screen %O2 100% O2

ALARMS 21% OFF SIMV PCV Volume or


Pressure

To Alarms Screen 2 Minutes Ventilation Modes


100% O2 ACMV, SIMV, SPONT/CPAP, VtG/MVG,
B-LEV
4
Extended Screen

Pressure or Flow Non-Invasive Square or Descend


Trigger Selection ON/OFF Waveform Selection

SIGH Trigger Mode NPPV Trends Trigger Mode

OFF Ptrig OFF ON OFF


Rise Profile PSV Flow Term. PSV Ti Waveform VG mode

3 15 1.4 Square VtG

Flow Termination/ Cycling- VtG or MVG Trigger Delay ON/OFF -


from 10% to 70% of the In MVG mode doubles the delay
peak flow Limit the inspiratory time before mandatory ventilation
time in Pressure Support is activated

5
Trigger Mode NPPV

Ptrig OFF
Alarms Screen
Rise Profile PSV Flow Term PSV Ti Waveform ALARMS

3 25 1.0 Square Screen

Buzzer Level Apnea Interval


Selection Setting

Buzzer High Rate LOW Rate LOW Vte LOW Vti Apnea Interval

LOW OFF OFF OFF OFF 20


LOW P HIGH P LOW MV HIGH MV FiO2 Low FiO2 High

3 99 0.1 20.0 OFF OFF

FiO2 Alarms ON/OFF

6
Quick-Start - quick access to the preset ventilation
configurations directly from the main screen

Load preset #1 Load preset #2 Load preset #3

Last SPONT
Resume ventilation Patient
Adult Pediatric (CPAP)
of the current patient
(default) To start ventilation press
ON/OFF key for 2 seconds

* Quick-Start is disabled by factory default and requires activation through the passcode protected
service screen.

7
For Local Sales Contact:

Flight Medical Innovations Ltd.


www.flight-medical.com
Local support: +1866 6587793

LIT-0017-REV.A01