A314746709 - 21627 - 3 - 2019 - Lecture - 13-14 HPLC & GC

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 48

High Performance Liquid

Chromatography
most popular, powerful and versatile
form of chromatography
Principle
Resolution Power of chromatographic column is determined by
many factors
• number of theoretical plates (N) in the column and hence plate
height (H)
• Value of N increases with column length but there are practical
limits to the length of a column owing to the problem of peak
broadening
Principle:
• Surface area of the stationary phase is directly proportional to the
number of theoretical plates

• Smaller the particle size of the stationary phase, greater will be the
surface area, hence greater the value of N

• N is inversely proportional to particle size.


Principle: Backpressure

• Unfortunately, smaller the particle size, greater is the resistance to


the flow of the mobile phase

• This resistance creates a backpressure in the column

• that is directly proportional to both the flow rate and the column
length

• inversely proportional to the square of the particle size


Principle: Backpressure
• This back-pressure may cause the structure of the matrix to
collapse, thereby further reducing eluent flow and impairing
resolution
Principle: Backpressure problem solution
• development of small particle size stationary phases (5-10 µm
diameter with a narrow range)

• which can withstand pressures up to 40 Mpa


HPLC Principle: Particle Size
• Larger particle size phases are the basis of low-pressure liquid
chromatography

• flow of the eluent through the column is either gravity-fed or


pumped by a low pressure pump, often a peristaltic pump

• LPLC is cheaper to run than HPLC but lacks the high resolution
that is the characteristic of HPLC
For gradient elution two reservoirs &
two pumps are used with liquid-
phase mixing before entry to the
sample injection loop

Components of an isocratic HPLC system


Columns:
Conventional columns Microbore or open tubular
columns
• made of stainless steel • internal diameter of 1–2mm
• can withstand pressures of up • generally 25-50 cm long
to 50MPa
• flow rates of 5-20mm3 min–1
• columns are 3–25 cm long
• approximately 4.6 mm internal
diameter
• flow rates of 1-3cm3 min–1
Columns
Microbore columns have three important advantages over
conventional columns:
• reduced eluent consumption due to the slower flow rates

• ideal for interfacing with a mass spectrometer due to the reduced


flow rate

• increased sensitivity due to the higher concentration of analytes


that can be used
Sample Loading: Injector loop

• common method of sample introduction is by use of a loop


injector

• consists of a metal loop, of fixed small volume, that can be filled


with the sample
https://www.youtube.com/watch?v=8-D-mMfDw7g
Cont…
• eluent from the pump is then channelled through the loop
• by means of a valve switching system and
• sample flushed onto the column via the loop outlet
• without interruption of the flow of eluent to the column
Guard column
• Highly impure samples viz. sera, urine, plasma or whole
blood, which have preferably been deproteinated, may
cause the column to lose its resolving power

• So a guard column is often installed between the injector


and the column
Cont…
• guard column = short (1-2 cm) column of the same internal
diameter
• packed with material similar to that present in the analytical
column.
• packing in the guard column preferentially retains
contaminating material and can be replaced at regular intervals
Mobile phases
• Choice of mobile phase depends on the type of separation to be
achieved

• Isocratic elution = single pump/single eluent or two or more


eluents premixed in fixed proportions

• Gradient elution = separate pumps to deliver two eluents in


proportions predetermined by a gradient programmer
Cont…
• All eluents must be purified

• traces of impurities can affect the column

• interfere with the detection system

• But even with pure eluent a 1-5mm microfilter is introduced


into the system prior to the pump
What is mobile phase degassing?
• Henry’s law states that ‘mass of a gas which will dissolve into a
solution is directly proportional to the partial pressure of that gas
above the solution

• When the solvents are mixed, solubility of air is less than it is


in same proportion of pure solvents

• and excess air will tend to come out of solution (bubble out – c/d
outgassing)

• Rough surfaces in HPLC system produce nucleation sites for


bubble formation
Problems due to gassing

• Air bubbles modify the flow of mobile phase through the column

• Gassing can alter column resolution

• Interfere with continuous monitoring of the eluate (Air bubbles


passing through detectors lead to spurious peaks)

• Practically it is not necessary to remove the entire dissolved air


but only a fraction can be removed
Degassing methods
• Helium purging: helium is bubbled through the solvent and removes up to
80% of dissolved air

• Vacuum degassing: solvent is exposed to a vacuum and the reduced


pressure removes more than 60% of the dissolved air

• Sonication: ultrasonic baths removes up to 30% dissolved air

• warming

• stirring vigorously with a magnetic stirrer


Pumps
• Capable of outputs of at least 50 MPa

• ideally there must be no pulses (i.e. cyclical variations in pressure); may


affect the detector response

• Flow capability of at least 10 cm3 min-1 and up to 100 cm3 min–1 for
preparative separations

• Constant displacement pumps maintain a constant flow rate through


the column irrespective of changing conditions within the column.
Cont…
• Reciprocating pump is the most commonly used form of constant
displacement pump
• Such pumps produce small pulses of flow and pulse dampeners
are usually incorporated into the system to minimise this pulsing
effect
• All constant displacement pumps have inbuilt safety cut-out
mechanisms
• so that if the pressure within the column changes from pre-set
limits the pump is inactivated automatically
Detectors
• Since the quantity of material applied to an HPLC column is
normally very small

• it is imperative that the sensitivity of the detector system is


sufficiently high and stable to respond to the low concentrations of
each analyte in the eluate
Detectors: commonly used detectors

• Variable wavelength detectors


• Scanning wavelength detectors
• Fluorescence detectors
• Electrochemical detectors
• Mass spectrometer detectors
• NMR spectrometer detectors
• Refractive index detectors
Variable wavelength detectors
• based upon ultraviolet–visible spectrophotometry
Scanning wavelength detectors
• record the complete absorption spectrum of each
analyte, thus aiding identification
Fluorescence detectors
• greater sensitivity (10–12 g cm–3) than UV detectors
• Limited by the fact that few analytes fluoresce
Electrochemical detectors
• These are selective for electroactive analytes and
are potentially highly sensitive
Mass spectrometer detectors
• These enable the analyte to be detected and its
structure determined simultaneously.
NMR spectrometer detectors
• These give structural information about the analyte
Refractive index detectors
• These rely on a change in the refractive index of the
eluate as analytes emerge from the column.
GAS CHROMATOGRAPHY
Principle

• exploits differences in the partition coefficients of the


volatilised analytes between a stationary liquid phase and a
mobile gas phase

• confined to analytes that are volatile but thermally stable

• most volatile elute first


Cont…
• temperature of the column is raised to 50-300 oC to facilitate
analyte volatilisation

• stationary phase consists of a high-boiling-point liquid material


such as silicone grease or wax
Principle
• widely used for the qualitative and quantitative analysis of a large
number of low-polarity compounds

• because it has high sensitivity, reproducibility and speed of


resolution
Components
of a GC
system
Columns: two types
• Packed conventional columns
• Capillary (open tubular) columns
Packed conventional columns
• consist of a coiled glass or stainless steel column 1-3m long
• 2-4mm internal diameter
• packed with stationary phase coated on an inert silica
support.
Packed conventional columns
Commonly used stationary phases include the
• polyethylene glycols (Carbowax)
• Or methylphenyl- and methylvinylsilicone gums
• Commonly used support is Celite (diatomaceous silica)
• Celite is often treated so that –OH groups gets modified (to prevent
support-sample interaction)
• Modification is done by silanisation of the support with
hexamethyldisilazane
Capillary (open tubular) columns:
• made of high-quality fused quartz
• are 10-100m long
• 0.1-1.0mm internal diameter.
• Two types: wall-coated open tubular (WCOT) and
support-coated open tubular (SCOT), also known as
porous layer open tubular (PLOT) columns
• WCOT = SP is thinly coated directly onto the walls of the
column

• SCOT = matrix is bonded to the walls of the column and SP is


coated onto the support
Application of sample
• majority of non- and low-polar compounds are directly amenable
to GC

• compounds possessing polar groups viz. -OH, -NH2 and -COOH


are generally retained on the column for excessive periods of time

• Poor resolution and peak tailing usually accompany this excessive


retention.
Application of sample: Derivetization
• Problem can be overcome by derivatisation of the polar groups

• This increases the volatility and effective distribution coefficients


of the compounds

• Methylation, silanisation and perfluoracylation are common


derivatisation methods for fatty acids, carbohydrates and amino
acids
Application of sample
• Test sample is dissolved in a suitable solvent such as acetone,
heptane or methanol

• For packed and SCOT columns, sample is injected onto the column
using a microsyringe through a septum in the injection port
Application of sample
• Normally 0.1 to 10mm3 of solution is injected

• As there is only a small amount of stationary phase present in WCOT


columns, only very small amounts of sample may be applied to the
column

• Splitter system has to be used at the sample injection port so that only
a small fraction of the injected sample reaches the column

• remainder of the sample is vented to waste


Application of sample
• maintain the injection region of the column at a slightly higher
temperature (+20 to 50 oC) than the column itself as this helps to
ensure rapid and complete volatilisation of the sample.
Mobile phase
• mobile phase consists of an inert gas such as nitrogen for packed
columns or helium or argon for capillary columns
• gas from a cylinder is pre-purified by passing through a variety of
molecular sieves to remove oxygen, hydrocarbons and water
vapour
• It is then passed through the chromatography column at a flow
rate of 40-80 cm3 min–1
• A gas-flow controller is used to ensure a constant flow
irrespective of the back-pressure and temperature of the column.
Detectors
• Flame ionisation detector (FID) – detect all organic compounds
• Electron capture detector (ECD) – For analyte which captures
electron e.g. halogen containing compounds. Used for
polychlorinated compounds such as pesticides, DDT, dieldrin,
aldrin
• Nitrogen–phosphorus detector (NPD)
• Flame photometric detector
• Rapid scanning Fourier transform infrared detector
• Mass spectrometer detector

You might also like