RED LION Calibration-1

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 12

Signal & Display Parameters

Action Flow Display Description Options/Range Remarks


Power ON 000.00 Current Value

Press - Pro
Press - 1 - Inp Signal Input Parameter

rANGE / 0.02u 0.002u


Press - Load Cell mV o/p range
0.02u Select ‘0.002u’
Press - dECPt/0 Decimal Point 0.0000
00000
Aggregate - 00000
0/0.0/0.00/0.000
Cement – 0000.0
Water – 0000.0
Additive – 00.000

Press - round/1 Least Count 1 to 100 Select ‘1’


1, 2, 5, 10….. 1
0.0 to 25
Press - FiLtr/1 Filter Value Select 0.5

Press - band Filter Band 0.0 to 25

Select 5
Press - Pts Scaling Points 2 to 16 Generally select 2
In the event of non linear
response selection may
Press - StYLE Calibration Style be suitably increased

Press - Apply/kEY Select – Apply


Apply-For auto calibration
Key – for Manual mV keying

Exit From Signal Input


Press - End Parameter

2
Secondary Parameters
Action Flow Display Description Options/Range Remarks
Press - Pro
Signal Input
Press - 1 - Inp Not Used
Parameters
User Input/Function
Press - 2 - Fnc Not Used
Key Parameters
Display/ Program
Press - 3 - Loc Not Used
Lock-out Parameters
Secondary
Press - 4 - Sec Not Used
Function Parameters
Max. Capture
Press - Hi - t Not Used
Delay Time
Min. Capture
Press - Lo - t Not Used
Delay Time
Press - Dsp - t
Display Update 1,2,5, 10, 20
Rate

Select 1

3
Auto - Zero Tracking
Press - At - t Not Used
Delay Time
Press - At - b
Auto - Zero 1 to 4095
Tracking Band

Select 2
Units Label On/ Off
Press - b - Lit
Back Light Off
Press - OFFSt
Display Offset -19999 to 19999
Value

Enter 0.00
Exit From Secondary
Press - End Parameters

4
Analog Output Parameters
Action Flow Display Description Options/Range Remarks

Press - Pro
Signal Input Not Used
Press - 1 - Inp
Parameters
Not Used
Press - no
Factory Setting Not Used
Press - 9 - FCS
Parameters
Analog Output
Press - 8 - Out
Parameters
Press - type Analog Type 0-20/0-10/4-20

Select 0-10
AS In Inp/HI/LO/tot
Press - Analog Assignment Select Inp
Press - An - LO
Analog Low -19999 to 99999
Scale Value
Aggregate = -30
Cement = -10
Water = -5
Additive = -0.5

5
Press - An - Hi
Analog High -19999 to 99999
Scale Value
Enter Value equal
To Maximum
Weigher
Capacity
udt
Analog Update 0.0 to 10.0 Sec.
Press -
Time Select 0.0
Press - End Exit From Analog
Output Parameters

6
Calibration
Pre Calibration Check

Load Cell & Cables:


• Ensure that all Load Cells are of Same Make, Same Capacity and of Same mv/v o/p
• Ensure that all Load Cells at same level
• Carryout cold checks for each load cell (Resistance Measurement)
• Ensure all Load Cells to be firmly bolted.
• Ensure the o/p of all Load Cells are more or less same on no load and then on load
• Trace Load Cells Cables till panel and check for joints, loose contacts, dry solder etc.

Mechanical Observations:

• Check for the hopper to be free suspending without touching any other part.

• Check for hopper to be sufficiently free from Jacks.

• The discharge chute/pipe needs to be soft and clean

• Ensure sufficient gap between hopper and mixer/inclined belt etc.


7
Action Flow Display Description Options/Range Remarks

Power ON 000.00 Current Value

Press - Pro
Press - 1 - Inp Signal Input Parameter

Not Use
Press - rANGE / 0.02u

Not Use
Press - dECPt/0

Not Use
Press - round/1

Not Use
Press - FiLtr/1

8
Not Use 0.0 to 25
Press - band 10
Pts
Not Use 2 to 16
Press - 2
Press - Style/Key

Press - APLY Select Apply

Zero Calibration:
Input Value For
Inp 1
•Displays the current mV reading
Press -
Scaling Point 1 From load cells.
•Ensure hopper to be empty, free from
obstructions and stable reading.

Press - 2.1234
Display Value For
Press - dSP 1
Scaling Point 1

• Enter ‘0’ value

9
Span Calibration:

Press - Inp 2 •Keep Standard Weights on


Weigher upto 80% to 90% of
12.354 Weigher Capacity
• Ensure the Weigher is free from
obstructions & display reading is stable
• Note that display reading (mV) should
be greater than zero calibration value

Press - dSP 2 After ensuring stable mV reading on


Display enter the value equal to the
1200 Standard Weights on the Weigher.

Press - 1200 • Confirm the display reading to be


Equal to the standard weights value
• Remove all weights and ensure ‘0’
Value on display.

10
PC Calibration:
• Run i-batch software
• Ensure PLC is in ‘Connected’ Status
• Go to ‘Calibration Screen’ in Technical Folder, as Shown below

Enter Parameters for each Scale as below,


• Minimum value: Value equal to ‘An – LO’ (Analog Low Scale Value) Field in respective Weight Indicator.
• Maximum value: Value equal to ‘An - Hi’ (Analog High Scale Value) Field in respective Weight Indicator.
• Range: Value equal to Maximum Possible Weighment in a batch. Care should be taken not to exceed 90% of
respective ‘Maximum value’
Note:
• Image: Displays the digital equivalent of analog in put for reference only.
• Actual Value: After completion of above procedure, confirm the actual values displayed for each scales match with their
respective weight indicator values.
11
Calibration Report
After Completion of weight indicator and PC calibration prepare the following calibration chart and record it.

Plant Details: Scale Type: Date:

Loading Value Unloading Value

Sr. Standard Indicator Actual Indicator PC Indicator Actual Value Indicator PC


No Weight Reading Value in Error % Error % Reading in PC Error % Error %
in Kgs. PC

1 0 0 0.00 0 0.0 0 0 0.0 0.0


2 20 20 20.00 0 0.0 20 20

3 40 40 40.00 0 0.0 40 40

.. .. .. .. .. .. .. .. .. ..

.. .. .. .. .. .. .. .. .. ..

.. .. .. .. .. .. .. .. .. ..

Up to Range of respective scale