Large Extremity Coil / Small Extremity Coil

--- Installation & Adjustment ---

Tokyo, Japan

LZE-BM0248-2

C~~~~~I~~N’~IAl~l’~Y : This mnterini contains proprieLnry information of


Hitachi Medical Corporation and it shrill not be disclosed to the third party or
reproduced without the prior written permission or Hitachi Medical Corporation.
Copyright@ Hitachi Medical Corporation. lSRS,IS90. All rights reserved.
CONTENTS

Page
1. Adjustment of Extremity coil (large) ..........-O...._...I 1

2 . Adjustment of Extremity coil (small) . . . . ..~.....-........ 3


1 . Adjustment of Extremity coil ( .arge)

The coil adjustment is effected in the same manner as


described in the TECHNICAL GUIDE - Installation & Adjustment.
Only the different jobs from those in the TECHNICAL GUILlti
will be described herein.
(1) Check the tuned value and adjust impedance using the
phantom (small) supplied to the system.
(2) The trimmer capacitors Cm and Cr for adjusting this coil
are located as shown below.
Adjust the capacitors similarly to the other coils.

Trimmer capacitor used


for adjustment of decoupling co51

NOTE : Do not touch this trimmer


capacitor, since this 5s
\C r preadjusted before shipment.

(3) The coil adjustment values should be as follows:

Coil name Coil No. Object F.ange of tuned value hdjustment


to be checked impedance

Extremity Clnused Kickel


coil KO. -solution oc to 2. 700i100 ri
(large) phantom
(small)

- l- 2
Center of light localizer Phantom center line(uhite)

T ,A-

Adjust the height by using a pad or the like

(4) Attach the coil No. label with the szme coil No. as
used in item(3) from among those supplied nez.r the BNC
connector on t‘he tuning box.
(5) According to the instructions -for gain adjustment in
the TECHNICAL GUIDE: - Installation & Adjustment,
adjust the gain.
(6) Measure the signal-to-noise (S/N) ratio of this coil.
The slice thickness is 5mm.
.Checkthet the S/N ratio of this coil is 98+10.
-

(7) After completion 0,+ all the necessary adjus+Jnents,


check that the image with no ebnormelity is obtained.

- 2- 3 LZE .BM0248
2. Adjustment of Extremity coil (Small)

The coil adjustment is effected in the same manner as


described in theTECHNICAL GUIDE - Installation 8 Adjustment.
Only the different jobs from those in the TECHNICAL GUIDE
will be described herein.

(1) Check the tuned value and adjust impedance using the
phantom (small) supplied to the system.

(2) The trimmer capacitors Cm and Cr for adjusting this


coil are located as shown below.
Adjust the capacitors similarly to the other coils,

Trimmer capacitor used


for adjustment of decouplincr coil

NOTE: Do not touch this trimmer


capacitor, since this is
preadjusted before shipment.

(3) The coil adjustment values should be as follows:

Range of tuned value Adjustment


Coil name Coil No. Obiect to be checked imDedance
Extremity Unused Nickel
coil No. -solution oc to 2A 200+100
(small) phantom
(smalll)

- 3 -
Center of light localizer

Caution nameplate side

I
Bed side - IIc Gantry side
I
f '
Adjust the height by using a pad or the like

(-4) Attach the coil No. label with the same coil No. as
used in item(3) from among those supplied near the BNC
connector on the tuning box.

(5) According to the instructions for gain adjustment in the


TECI-INICAL GUIDE - Installation & Adjustment, adjust the
gain.
At that time, set the light localizer to the positin
as shown above and set-fzed.

(6) Measure the signal-to-noise (S/N) ratio of this coil.


The slice thickness is 1Omm.
Check that the S/N ratio of this coil is 240555.
Khen measuring the S/N ratio, set the light localizer
to the center of this coil and set-feed. Take an image
at the position of 30mm deeper that of the ISCANOGXA.!!]
image using the/SZ:SECT SLICE].

(7) After completion or' all the necessary adjustment,


check that the image with no abnormaility is obtaindedl.

I3

6 lzx-FiMU24E
- 4 -
--- Installation & Adjustment ---

Tokyo, Japan

L2E-B&40253-5

C~~FI~ENT~ALITY : This material contains proprietary information of


Hitachi Medical Corporation and it shall not be disclosed to the third party or
reproduced without the prior written permission of Hitachi Medical Corporation.
Copyright@ Hitachi Medical Corporation. 1990.1991. All rights reserved.
Contents

Page

1. General Q..............~_..........~.............. 1

2. Adjustment .-..........a.........S..,............. 2

I L2E-BM0253
1. General

This technical Guide describes how to install and adjust


the flexible body coil which can be set in accordance with
the patient's posture for body imaging.

- l-
2. Adjustment

The coil is to be adjusted in the same manner as


described in the Technical Guide - Installation and
Adjustment. So, only the different procedures from
those explained in the Technical Guide will be
described herein.

(1) Use the phantoms below to check on the TUNE and


impedance values.

NO. 1 connector l%R-FBC-31 Connectors other than No. 1,


MR-FBC-32

Nickel solution phantom No. 3 Nickel solution phantom No. 4

Trimmer capacitor \ No.1 No.2 No.3 No.4

0
for adjustment A

(2) The coil adjustment should be performed by using


connector No. L and the other connectors are used only
to check on the impedance and phase.

3 L2E-BIN0253
- 2 -
(3) Remove the cover of the coil and the polyvinyl chloride
cover (as shown in "A" in the previous page), 2nd set
the phantom on to the coil at the connector position
No. 1 (where the coil length is shortest).

Lock : Push down


,A-----'../ Unlock : Lift

_Set tlie coil center to the


approxiinate center line of
the tabletop.

(4) Adjustment of the impedance

Set the coil No. and TUNE voltage at "18". Connect


the impedance meter to the preamplifier. Adjust the
trimmer capacitor so that the phase is on O', and check
the impedance if within the specified range as described
below.
Refer to 4.7.2, "receiving coil" in the
Installation Manual.
T
-
Coil name Coil No. Object TUNE value tc Target
be checked impedance

i
Flexible Unused
Nickel solurion
phantom No. 3 oc to 72 :
I
coil for Nickel
the body NO. solution
1 phantom No. 4 oc to 72

(5) Adjustment of the RF gain value

Use the coil No. used in (4).

NOTE: Record the RF gain value when adjusting.

(6) Measure the signal-to-noise ratio value. For the body


coil (large) as well, measure the signal-to-noise ratio
value in accordance with the standard manual. Make
sure that the signal-to-noise ratio value is 0.8 to
1.6 times that of the body coil.
(Only connector No. 1 of the MR-FBC-31 takes 1.5s3.3
times)

(7) Check on the impedance and adjust the RF gain values


for the connectors from No. 2 to 4. Set the coil No.
to perform the 1 TSCANOGRAMI.
hen, check on the TUN E
value. If within the specified range in table (41,
connect the impedance meter to the preamplifier, Turn
the dial to 0" phase to change the frequency and check
on the impedance if within the range specified in table
(4) f Use the same coil No. as the connector No. used
in (4) for the coil used in (7).

NOTE: If there is any gap found between the coil and


the phantom, use a sponge pad or the like to
fill up the space for uniform lapping of the
coil as shown below.

5 LZE-BM0253
- 4 -
If any gap is found, use a spo~lge
or the like to wrnp the coil.
/

connector so that it can


cted at the right lateral
as viewed from the table.

Adjust the height by using


a mattress or the like. I

(8) After completion of all the adjustments, make a manual


entry of the RF gain value. The value to be entered
should be the mean value measured from the connector Nos.
1 to 4, and the coil No. should be the one used in (4).

(9) Set the cover of the coil, and put the accessory
coil No. label ona position as shown below. The coil
NO. should be the one used in (4) D

Atrach label of the


coil No. here.

Caution Label

6 L2E-BM02.53
- 5 -
(10) Perform the impedance adjustment with the human body.
The adjustment should be performed for one of the
connectors and the value should be the same as in (4) O
Impedance adjustment for the other connector is
unnecessary.

fll) Make sure imaging can be successfully performed.

7 L!ZE-BM0253
- 6 -
ode1

Trunk Surface Coil

Tokyo, Japan

~~TIAL~~~Y : This material contains proprietary information of


Hitachi Medical Corporation and it shall not be disclosed to the third party or
reproduced without the prior written permission of Hitachi Medical Corporation.
Copyright@ Hitachi Medical Corporation. 1989. All rights reserved.
Adjustment of the trunk surface coil

The coil adjustment is effected in the same manner as


described in the Technical Guide - Installation & Adjustment.
Only the different jobs from those in the Technical Guide
will be described herein.

1, Check the tuned value and adjust the impedance with


the small phantom supplied.

2. The tuning portion of this coil is molded with silicone


rubber for damage prevention. Do not touch any
of parts except the exposed trimmer capacitor.
Here,. this trimmer capacitor is used for phase
adjustment. Adjust the trimmer capacitor so that
the phase will be O"+lO", and check the impedance
at the phase.

3. Fold:-th-e co-i1 to adjust it.

,-

Center line of coil

1 L2E-BM0256
- l-
4. Se't the coil and phantom as shown below.

Tuning board side

NOTE: Set the coil s NOTE: Set the body under


light beam fro coil at the nearest
localizer coincides with the position to the
center line of the coil. foot side.
(X and Y-axis directions)

Fix the phantom using


the Velcro tape.
/

2 L2E-BM0256
-2 -
-.
5. Set the phantom in the center of magnetic field.
However, positioning for the height direction
(Z-axis direction) is not required.

6. The coil adjustment values should be as follow:

Coil name Coil No. Object Tuned value Range of tuned Impedance
to be adjusted value to be to be
checked checked

Nickel-
Trunk Unused solution
surface No. 18 OC to 24 200 R-tlOOR
phantom
coil
(small)

7. Attach the one of supplied coil No. labels with


the same coil No. used in item 6 to the position
shown below.

Attach here.

Caution nameplate

3 L2E-BM0256
- 3 -
8. According to the Technical Guide - Installation &
Adjustment, adjust the gain.

Take a sagittal image using the v] ~

Slice position
I 1

lO* Take a coronal image at the 30mm nearer position


from the bottom of phantom to the center. -.
The sequence shall be the same as described in the
Technical Guide - Installation & Adjustment.

11, Measure the signal-to-noise (S/N) ratio of the


trunk surface coil and head coil, respectively
using the same phantom according to the S/N ratio
Check Program, Check that the S/N ratio of the
trunk surface coil is 1.2 to 1.8 time larger than
that of the head coil.

12. After completion of all necessary adjustments,


check that the image with no abnormality is
obtained by the coil,

-4-
yste

ode1 P

ead Coil

--- Installation & Adjustment ---

Tokyo, Japan
LZE-BMOZSl-1
Contents

Page

1. General ...........................................

?
L. Procedure for modification . . . . . . . . . . . . . . . . . . . . . . . . 2
2.1 Installing the receiver coil ................. 2
2.2 Installing the connector box ................. 3
2.3 Additional receiving cable work .............. 5
2.4 Replacing the preamplifier ................... 6
2.5 Replacing the mat ............................ 6
2.6 Modifying the MRI unit ....................... 6
2.7 Modifying the filter box ..................... 7
2.8 Connecting the additional external cables .... 7
2.9 Program entry ................................ 8

3. Adjustment ........................................ 9
3.1 Entry of receiver coil data .................. 9
3.2 Impedance adjustment ......................... 10
3.3 RF gain adjustment .......... . ................ 13
3.4 Performance and operation check .............. 14

4. Coil information entry program .................... 15


4.1 Function ..................................... i5
4.2 Operating procedure .......................... 15

5. List of supplied parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

1 L2E-BM028i
1. General

This manual describes the procedures for installation and


adjustment of the HS/MR head coil which permits to produce
images of high quality than with the standard craniocervical
receiver coil.

-l- 2 LZE-BMO 281


2. Procedure for modification

2.1 Installing the receiver coil

(11 Remove the existing head rest from the table, and then
detach the sliding rail.

(2) Install to the head rest for HS/MR head coil the
sliding rail having been detached at step (1) above
(however, this is not necessary in case it is not
intended to use the craniocervical receiver coil).

i3) Install the head rest for HS/MR head coil to the HS/MR
head coil sliding rail, and install the head rest to
the tabletop.

Craniocervical
I , P18~20 screw I[RENY) receiver coil

Head rest --
(for IIS/?lR
head coil)

head coil
sliding
113~ 23 scre-x raii
(RENY)

(4) Install the receiver coil to the rail, and then check
that the rail slides smoothly. Also check that the
craniocervical coil sliding rail moves slides smoothly,

-2 - 3 LZE-BM0281
2.2 Installing the connector box

The content of this work is different in accordance with


the shape of RF cover hole.

(1) Detach the side cover and front cover from the gantry.

(2) Remove the reception connector and cabie from the


RF cover.

Grounding plate
Grcuriing pla;e

(3) Cut the rubber placed between RF cover and front cover.
(Only in the case of RF cover hole being round.)

- 3 - 4 LZE-BM0281
(4) Attach the connector box, which is sent by HMC.
(Only in case of RF cover hole being round.)

Co-axial cable

.’ / 1’/ Soideriess terraina).


/

/’
/$gilal:.
Grounding plate
'
/ \-%9
A-4
To preamolifier Co-axial cable

(5) Attach the connector attaching plate, which is sent


by HMC. The side on which the red name plate is
sticked, is to be up-side.
(Only in case of RF cover hole being oval.)

- 4 - 5 LZE-BM0281
(6) Stick the labels "A"' and "B"', which are sent by HMC
together with other labels, on the place shown in
the figure. (Only in case of RF cover hole being oval
shape.)

RF cover

NOTE E Stick the label co that


the label become the same
level with BMC connector.

(7) Attach the gantry front cover.

2.3 Additional receiving cable work

Stick the indication mark (pink color), which is sent by


mc, onto the receiving cable. Including the spare, the
works for two cables are to be done.

- 5 - 6
2.4 Replacing the preamplifier

Connect the cables as discriminated with their ccnnector


Nos.

F-/ M4xg
5gCN 5~5:r:R
/ Replace.

2.5 Replacing the mat

The auxiliary table top mat is repiaced with the mat to


be sent.

2.6 Xodifyirlg the MRI unit

(1) Detach the cover.

(2) Install the supplied reception connector (with cabies


and fixture) to the left side of the RF transmitter/
receiver unit chassis.

Left side of chassis

\ To RF amplifier

M3xlO screw
Additional reception connector

- 6 - 7 LZEBM0281
(31 Replacing the module

Replace the RF AMP module in the slot No. 1 and the


I/F module in the slot No. 14 with the supplied ones.

(4) Make wiring between the modules in the RF transmitter/


receiver unit and connect the additional reception con-
nector cable (to 581CN of RF amplifier).

3
L. 7 Modifying the filter box

Drill holes in the filter box and install the connector to


the filter box, as shown below.

Connector nameplate (27CIT inside shielded room;


26CN outside shielded room)
Connector S-PA (J8351110)

opposite side)

2.8 Connecting the additional external cables

Connect the two additional reception cables as shown below:

(1) Run #335 (5D2W; having a same length as run #328*) from
12CN of gantry (GAN) preamplifier to 27CN of filter box
(FIL)

(2) Run #3 12 (5D2W ; having a same length as run #308*) from

-7 - 6 L2E-EM0281
:2 CN of MRI unit (MRI) to 26CN of filter box (FIL)

NOTE: The additional cables should be the same in length


as the cables marked with *. If the supplied cables
have not the same length, they should be extended
or shortened so as to have the same ienqth.

2.9 Program entry

Regi.ster the programs in the FD (service program, "COIL DATA


ENTRY") , which are sent by HMC, referring to Section 2.5 in
"Technical Guide - Check & Service Program".

NOTE: In case the version-up of the system software ac-


comodates the use of the HS/MR head coil, this entry
is not required.

- 8 - 9 L2E-BMOZ31
3. Adjustment

NOTE: When the table is modified for use of the HS/MR head
coil, the existing receiver coil must be readjusted,

3-l Entry of receiver coil data

Select and run the service program COIL DATA ENTRY (48-313,
set a coil name and RF reception line and enter the
attenuation amount for adjustment of the coil gain. For the
program, refer to item 4. This entry is to be done for all
the coils used.

Setting:

- COIL NAME:

Set a name for each coil with a maximum of 16 characters,


and enter it.

- RF LINE:

Set "3" (RF LINE B) for the existing coil, and "2" (RF LINE
A AND B) for the HS/MR head coil.

- RF ATT:

Enter an attenuation amount for the HS/MR head coil alone,


This setting is not necessary for other coils. Set '4" for
the HS/MR head coil.

- 9 - 10 LZE-EM0281
Example settings:

3, 2 impedance adjustment

(1) Existing receiver coils

For the existing receiver coils, adjust the impedance


in the conventional manner by referring to item 4.7.2
in the Technical Guide - Installation & Adjustment or
the installation manual for each of the coils, taking
the following in consideration:

(a) Connect the reception cable to the connector -R


(with red label).

fb) Use the input connectors 58CN and 59CN of the pre-
amplifier.

(c) When it is intended to make a measurement and


check the tuned value, be sure to set a coil No.
as HS/MI? head coil No. by running the program
"PATIENT DATA SET" before the measurement.

- 10 - 1 1 LZE-BIvl02 8 1
(2) HS/MR head coil

(a) Detach the cover,

(b) Set the Ni solution phantom (small) as shown.


Slide the receiver coil to a position where it
abuts the head rest. Have the light mark defined
by the light localizer coincide with the slice
mark on the receiver coil (the light mark cannot
be made coincident with the cross (t) mark on the
phantom. Press the SET button to set the phantom
in the center of the magnetic field.

Ni solution phantom (165 dia.)

'Abdomen immobilizing matt


(20 mm thick)

(cl Take a same manner as with the existing coils to


set a tuned value, The value should be "18".

(d) Impedance adjustment

Using a Type T BNC cable as in the conventional


impedance adjustment, connect an impedance meter

- 11 - 12 L2E-BM0281
to the receiver coil as shown below:

Line-k adjustment: Line-B adjustment:

Line B
Line
~s:~~ BxmpedancJ z ~~~~ B

meter

For each of the lines A and B, turn the trimmer


capacitor as shown below for the phase to be
0" i 2". Then check that the impedance is 250
ohms 2 100 ohms.

Line-A adjustment

- 12 - 13 L2E-BM0281
Line-B adjustment

Disconnect the impedance meter, and connect the


lines A and B to 56CN and 58CN, respectiveiy.

(e) Effect a scanography to make sure that the tuned


value is within a range of "OC" to "48". For this
checking, be sure to set a coil No, as HSJMR head
coil No. by running the program "PATIENT DATA
SET".

(f) Attach the coil cover, and attach a coil No.


label as shown beiow.

, Cover

Ceil No. label

3.3 RF gain adjustment

Adjust the RF gain as with the existing coils by referring


to item 4-8 of the Technical Guide - Installation & Adjust-
ment. This adjustment must be done with all the existing

1 4 L2E-BM0261
- 13 -
coils as well.

3. 4 Performance and operation check

(1) Signal -to-noise ratio

(a) Set a same phantom as in the impedance adjustment.

(bl Measure the signal-to-noise ratio in the same man-


ner as with the craniocervical receiver coil,

(cl Measure the signal-to-noise ratio for the cranio-


cervical receiver coil as well.

(d) Make sure that the signal-to-noise ratio of the


HS/MR head coil is 20% to 27% higher than that of the
craniocervical receiver coii.

(2) Operation

After completion of all necessary adjustments, check


that the image with no abnormality is obtained by the
coil.

- 14 - 1 5 L2E-Bhf0281
4. Coil information entry program

4.1 Function

This is a program to change the coil information (in the


system table) a With this program, a name, RF reception line
and attenuation amount for gain adjustment of ES/MR head
coil are set for each coil.

4.2 Operating procedure

(1) Press the SELECT MODE switch, enter '4", "8"' and press
the ENTER key, and then enter "3"', "1" and press the
ENTER key.

(2) The following message and the


information list of coil
ctirrently set in the system table are displayed. Then
e n t e r a coil No. to be changed and press the ENTER key
in response to the message.

** COIL DATA ENTRY **


COIL # COIL NAME RF LIKE RF ATT
1 0000000000000000 0 0

1 6 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0

COIL NUMBER =

---This is to be entered at this step (2).

Coil information list

(3) After a coii No. is entered, the following message will


be displayed. Enter a coil name in response to the
message.

Enter a coil name to be changed with a maximum of 16


characters, and press the ENTER key- After that,
change the coil information list. By pressing the

- 15 -
1 6 LZE-BM0281
CANCEL key, the step (2) for input cf the coil number
is restored.

COIL NAME =

(4) After input of the coil name, an RF reception line cur-


rently set is displayed as default in the following
message. In response to the message and referring to
the RF reception lines list displayed in the lower
right area of the screen, enter a numeral to be changed
and press the ENTER key.

After that, change the coil information list. By pres-


sing the CANCEL key, the step (3) for input of the coil
name is restored.

Pl? LINE (1 - 3) = 0

L- Default numeral

- Display of RF reception lines list (displayed in the


lower right area of the screen):

* RF LINE *
1 s.. RF LINE A
2 . . 1 RF LINE A AND B
3 . . . RF LINE B

(5) In case the RF reception line 2 is selected in the step


(4) for input of the RF reception line, the currently
set attenuation amount for gain adjustment of the HS/MR
head coil is displayed as default in the following mes-
sage. In response to the message, enter a numeral to
be changed, and press the ENTER key.

After that, change the coil information list, By pres-


sing the CANCEL key, the step (4) for input of the RF
reception line.

- 16 - 1 7 LX-Bhf02 8 1
Also, in case other than "2" is selected in the step
(4) for input of the RF reception line, no input of the
attenuation amount for gain adjustment of the HS/MR
head coil is done, the attenuation amount is auto-
matically changed to 0 (Zero) and the co il informat ion
list is changed correspondingly.

RF ATT (0 - 7) = o

L.----- Default display (in dB)

(6) After the input of an RF reception line in the step (4)


or after the input of an attenuation amount for gain
adjustment of the HS/MR head coil in the step (51, the
following message is displayed and an input in response
to the message is awaited.

NEXT SET CR? (Y/N) =

By entering "Y" and pressing the ENTER key, the step


(2) for input of the coil No. is restored and the coii
information can be continuously changed.

Ey entering sN" and pressing the ENTER key, the opera-


tion goes to the step (7).

(71 After entering "N" and pressing the ENTER key, the fol-
lowing message is displayed and an input in response to
the message is awaited.

SET OK? (Y/N) =

Ey entering IlY" and pressing the ENTER key, all the


inputs are entered into the system table and the
process is ended. By entering "N" and pressing the
ENTER key, they are not entered into the system table
and the process is ended,

- 17 - 1 8 L2E-BM0281
5. List of supplied parts

(1) HS/MR head coil 1


(2) Head rest (for HS/MR head coil) 1
(31 Sliding rail (for HS/MR head coil) 1
(4) Preamplifier (for HS/MR head coil) 1
(5) Connector box (with cables) 1
(6) Connector box fixture 1
(7) Connector attaching plate 1
(8) Tube (for connector box) 4

(9) RF AMP module 1

(10) IF module 1

(11) Additional connector for RF transmitter/


receiver unit (with fixture and cables) 1

(12) Mat for auxiliary table top 1

(131 Type N connector (N-PA) 1

(14) External cables:


- Run #312
- Run $335
(15) Screws:
M8x20 (KEN-f) for sliding rail' 4
M8~20 (RENY) for head rest 7
&14x25 for connector box 4
M4x8 for connector box and preamplifier 8
M3,110 for RF transmitter/receiver unit 2
M3X12 for filter box 4
M3 nut for filter box 4
M3 spring washer for filter box 4
M3 plain washer for filter box 4

(16) Coil No. label 1


(17) Connector No. label (26CN, 27CN) 1
(18) Program FD 1
(19) Color mark (pink and blue) 1

- i8 - 19 LZE-BM0281
MR-QPVC-3 1

MR-QPVC-4 1

MR-QPVC-5 1

Twinsburg, OH
CONFIDENTIALITY: This material contains proprietary information
Hitachi Medical Systems America (HMSA) and it shall not be disclosed
other third parties or duced without prior written permission of
0 Copyright Hitachi cal Systems America 1996. All rights reserved.

L5E- x01-4
1.0 Outline
The QD Pelvis Coil is intended to provide high quality imaging of male and
female pelvic anatomy.

2.0 Installation and Adjustment


The following steps describe the Installation and Adjustment procedures for
the QD Pelvis Coil for the MRP-5000, MRP-7000 and AIRIS.

2.1 Installation

2.1.1 Place the QD Pelvis Coil on the patient table on top of a ‘thin’ mat.

2.1.2 Locate the two Access Ports to the tuning capacitors of the QD Pelvis Coil, and
remove the plastic plugs. These Access Ports are located along the base of the
coil, one per side; refer to Figure 1 below.

2.1.3 Position the Number 4 SNR phantom in the coil, so that the bottom of the
phantom is flush with the top and bottom of the center arc. Use the
positioning lights to center the coil and phantom; refer to Figure 1 below.

\ Access Port to
tuning capacitors
(1 per side)

Figure 1. Phantom Positioning in the QD Pelvis Coil

Page 2 L5E-BJ.‘i!IAlO1-4
2.1.4 Press set to advance the coil and phantom into the magnet for scan position.

2.1.5 plug the cable into the connector on the left inner surface
nd ensure that the two cable connectors are attached
according to the color coding.

2.1.6 plug the cable into the connector on the left


and align the D-shaped connector before
inserting.

2.1.7 Access the Coil Data Entry program, and register the coil with the following
parameters:

2.2 Adjustment

Adjustment may be difficult due to the necessity to ‘blind’ insert the adjust-
ment tool.

2.2.1 Access the Coil ata Entry program, and set up the coil as follows:
RF Line 1 (A - only)

2.2.2 Use the Signal Generation Check program to set the TUNE value as follows:
TUNE 18

2.2.3 Use the Impedance Meter to adjust the Line A impedance to the following
specifications:

200+ 30 iz
or maximum impedance with Phase < 20%

age 3 L5 AlO 1-4


2.24 Select a scanogram and ensure the TUNE is within the following value range:
TUNE 12 - 1E

2.2.5 Use the Signal Generation Check program to set the TUNE value as follows:
TUNE 18

2.26 Access the Coil Data Entry program, and set up the coil as follows:
RF LINE 3 (B - only)

2.2.7 Use the Impedance Meter to adjust the Line B impedance to the following
specifications:

200+ 30 n
or maximum impedance with Phase < 20%

2.2.8 Select a scanogram and ensure the TUNE is within the following value range:
TUNE 12 - 1E

2.2.9 Access the Coil Data Entry program, and set up the coil as follows:

Coil PELVIS
RF LINE 2 (A and B)
ATT 0
TUNE STEP 3 (MRP-7000/AIRIS only)

2.3 RF Gain

2.3.1 Perform the RF Gain (automatic) function on the QD Pelvis coil to set the
transmit power level.

2.4 SNR Test

2.4.1 Calculate the SN ratio with the following sequence:

MRP-5000 MRP-7000AIRI~
Sequence SE Sequence SE
TR 500 TR 500
TE 40 TE 40
Slice Thickness 10 Slice Thickness 5
AVE 2 AW 2
Projection 256 Projection 256
SQ No. 3 SQ No. 3
FOV 320 FOV 320

Page 4 L5E-BMAlOl-4
2.4.2 Verify that the SN Ratio is within the specified range according to the
following table; contact HMSA Technical Support if it does not:

MR-QPVC-3 1 xX+xX

MR-QPVC-41 115 k 20

MR-QPVC-51 xxkxx

2.4.3 Inspect the SNR image for asymmetrical shading or image artifacts; contact
HMSA Technical Support if any are noted.

Page 5
in. yste

Model AIRIS

Small Extzmity Coil


MR-JC-52

TECHNICAL GUI
--- Installation & Adjustment ---

Tokyo,Japan

LZE-BM5088-I

CONFIDENTIALITY : This material contains proprietary information of


Hitachi Medical Corporation ond it shall not be disclosed to the third party or
reproduced without the prior written permission of Hitachi Medical Corporntion.
Copyright @ Hitachi Medical Corporation. 1995. All rights reserved.
Contents

Page

1. Adjustment of Extremity Coil (small) . . . e . . . . . . 1

( 1 ) 1 LZE-BM5088
1, Adjustment of Extremity Coil (small)

The coil adjustment is effected in the same manner as


described in the TECHNICAL GUIDE - Installation & Adjustment.
r‘nly the different procedures from those described in the
TECHNICAL GUIDE will be mentioned herein.

(1) Check the tune value, and adjust the impedance using
supplied phantom No. 1.

(2) The trimmer capacitors Cm and Cr for adjusting this coil


are located as shown below.

Adjust the capacitors similarly to the other coils.

Cm

Trimmer capacitor used for


decoupling coil adjustment
/

Cr

NOTE : Do not touch this trimmer capacitor, since this is


preadjusted before shipment.

(3) The coil adjustment values should be as follows.


-.-
Tuned value
Voltage to Impedance Phase
Coil # Coil # Phantom # range to be
be applied adjustment (0) adjustment checked

9 Small 1 1 WI 200 jI 100 IklO” OC-36


extremity

- l- L2E-RM5088
Phantom No. 1

20 mm thick abdominal immobilization mat


75mmth

Same in 50mm thickness

(4) Set the following scan parameters, and perform three


scans without any positioning.

SCAN PLANE : TRS


SEQUENCE : SE
2D/3D : 2D
SEQ. No. :3
REPHASE : OFF
PHASE DIR. : A-P
REJECTION : OFF
RECTziNG. : OFF
ANTI - WRAP : OFF
PRESAT. : OFF
ADAPTIVE IMAGE : OFF
C. POSITION : 0
FOV :180
TR : 500
TE : 40
THICKNESS :5
MULTI SLICE : 1
ECHO :1
NSA :2
PROJECTION : 256

-2- LZE-BM5088
(5) Check to see that the SN ratio is within the specified
range as described in the following table.

Coil model Standard SN ratio

I MR-JC-52 I 220 +40 I

(6) After completion of all necessary adjustments, check that


the image with no abnormality is obtained.

-3 - 4 L2E-RM5088
MR-QCS-42

MR-QCS-52

CONFIDENTIALITY: This material contains proprietary information of


Hitachi Medical Systems America (HMSA) and it shall not be disclosed to
other third parties or reproduced without prior written permission
0 Copyright Hitachi erica 1997. AI1 rights reserved.
- ---.__.- .-~--~. ~ .-
1.0 Outline
The QD C-Spine Coil is intended to provide high quality imaging of the
cervical spine; dependent upon patient anatomy, the coil sensitivity may
additionally image upper thoracic and lower cranial structures. The QD C-
Spine Coil makes use of a split top coil design. A releaseable bridge structure
is provided for easier patient positioning. Refer to Figures 1A and 1B below.

Head lmmobilizer Pads

Alignr

Figure IA, 1B. QD C-Spine Coil (Assembled, Exploded views)

CAUTION
Do not carry the QD C-Spine Coil by the bridge.
Although the bridge latches securely to the coil base,
it contains sensitive electronics, and it is not durable
as a carry handle.

Page 2 L5E-BMA106-4
ose
This document specifies the installation test and adjustment procedure for the
D C-Spine coils intended for use with the RP-7000 andAIRIS.The test
results shall be documente for quality assurance purposes and evaluated to
ensure that each production coil meets the coil performance specifications.

3.1 Q5E-BMA106 Operator’s Instruction Manual, QD C-Spine Coil


(MRP-7000/AIRIS)

Equipment List
C-Spine Coil
P-7000 or AIRIS
Vector impedance meter, HP 4193A or equivalent (qty. 1);
Test cable (qty. 1): RG58 BNC malelBNC female, half wave length,
with.. .
Integral fine-tuning adjustable capacitor;
Hitachi RF coil loading phantom no. 2;
Non-magnetic screw drivers;
Non-metallic screwdriver tuning tool.

ustment
5.1 Coil and Phantom Setup

5.1.1 Turn on the Vector Impedance Meter, allowing it to warm-up 5 minutes before
use.

5.1.2 Set the frequency of the Impedance Meter to the system frequency (f,).
Calibrate the half wave length cable for the particular system frequency by
fine tuning the capacitor of the cable assembly. The cable should be calibrated
as:

Open load impedance : > 1000 Ohm


Phase : 0" +- 2"

5.1.3 Assemble the C-Spine coil. Ensure that the bridge is securely latched to
the coil base. se pads as necessary such that the coil and phantom are
centered vertically in the magnet bore. Place the QD C-Spine RF coil pad
inside the coil, and set the No. 2 Phantom (130mm) on top of the pad, as
shown below. Ensure that the label on the end of the phantom faces out of the
magnet. For the ne er phantoms, ensure that the phantom bottle cap faces
out of the magnet. efer to Figure 2 on the following page.

age 3 L5E- 106-4


Coil Landmark Line

Phantom Landt

#2 Phantom (130mm)
3/0” RF Coil Pad

(as needed)

AK15 Coil Podt3oning

Phantom #2

QD C-5pine RF Coil Pad

1” Thick pads from QD Flex


Body Coils

MRP-7000 Coil P&-Liming

~--‘.-- P h a n t o m #2

~-x.l”- QD C-5pine RF Coil Pad

.---x- Standard mattress pad

Figure 2. Coil and Phantom Positioning for QA Measurements


Page 4 L5E-J3MA106-4
5.1.4 Plug in the coil and position the coil in the center of the magnet on the patient
table.

5.1.5 ress set to advance the coil and phantom to magnet isocenter.

5.1.6 plug the coil cable into the connector on the left inner
surface of the front iacade, aligning the D-shaped connector before inserting.

5.1.7 remove the pre-amp assembly cover. Plug the coil cable into
the table, aligning the D-shaped connector before inserting.

5.2 Coil Tuning

5.2.1 Access the COIL program and add D C-Spine as the coil
name for coil #23. Select RF LINE 1 - A only.

5.2.2 Perform scanogram (coronal) and display the phantom image (as diagrammed
in step #1 of Figure 3).

5.2.3 Adjust Window to about 100, and Level as appropriate, to present an image
similar to phantom diagrammed in step #2 of Figure 3 below.

1 +2 -3

D C-Spine Coil Sensitivity

5.2.4 Locate the center of coil sensitivity (identified for example purposes as ‘A’ in
phantom diagrammed in step #3 of Figure 3).

5.2.5 Move the table as appropriate to align the center of coil sensitivity ‘A’ to
coincide with the magnet isocenter.

5.2.6 Perform a second scanogram (coronal) to verify correct repositioning.

5.2.7 Use the SIGNAL GENE program to set the TUNE value to 18.

5.2.8 eter an

age 5
52.9 Identify the tunable capacitor labeled “TUNE A? on the coil, to adjust the
LINE A impedance (refer to Figure 4 below).Tune the “TUNE A” capacitor so
that the LINE A impedance, at the system frequency, is specified as :

Magnitude (Z): 2 250 Ohm


Phase (a): -20” < a < 20”

,- Coil Landmark Line

Phantom Landmark Line -..,


I---------
,,~/.-.--.:” “\
w Tee. .
b
#2 Phantom (130mm)

QD C-Spine RFCoil Pad

(as needed) A : Saddle Coil


B : 5olenoid Coil
Figure 4. Coil Tuning Ports Locations

5.2.10 Remove the half wave length cable connection.

5.2.11 Select a SCANOGRAM and verify that the TUNE value is in between 12 and lE,

5.2.12 Use the SIGNAL GENERATION CHECK program to set the TUNE value to 18.

5.2.13 Access the COIL DATA ENTRY program and select the QD C-Spine coil.
Select RF LINE 3 - B only.

5.2.14 Use the Impedance Meter and the half wave length cable to monitor the coil
LINE B impedance. The frequency of the Impedance Meter should remain at
the system frequency (fo) for the measurement.

5.2.15 Identify the tunable capacitor labeled “TUNE B” on the coil, to adjust the
LINE B impedance (refer to Figure 4 above). Tune the “TUNE B” capacitor so
that the LINE B impedance, at the system frequency, is specified as :

Magnitude (Z): 2250 Ohm


Phase (a): -20” < a < 20”

5.2.16 Remove the half wave length cable connection.

5.2.17 Select a SCANOGRAM and verify that the TUNE value is in between 12 and 1E.

Page 6 L5E-BMA106-4
5.3 Registration of Coil Data

5.3.1 Access the COIL DATA ENTRY program and select the QD C-Spine coil.
Select RF LINE 2 - A and B. Set the ATT value to 2. For systems with
Operating System Software Ver. 4.0 or later, set the RF GAIN Adjust to Y.

5.3.2 Perform scanogram (coronal) to identify the center of the coil sensitivity. Move
the table as appropriate to align the center of coil sensitivity with the magnet
isocenter.

5.3.3 Perform scanogram (transverse) to check vertical centering of the phantom.


Adjust the phantom height as necessary.

5.3.4 RF Gain Adjustment

5.3.5 Perform the RF GAIN (automatic) function on the coil to find the transmit
power level for TRS and COR planes. Calculate the average of the TRS and
COR values, and use the ‘Manual’ function to enter the average as the power
for this coil number.

S eas
6.1 Calculate the SN ratio with the following sequence:

Slice Thickness 5

6.2 Verify that the SN Ratio is within the specified range according to the
following table; contact MSA Technical Support if it does not:

6.3 Inspect the SNR image for asymmetrical shading or image artifacts; contact
HMSA Technical Support if any are noted.

er e
as

Page 7 L5 106-4
MR-QMC-41

MR-QMC-51

Twinsburg, OH
CONFIDENTIALITY: This material contains proprietary information of Hitachi
Medical Systems America (HMSA) and it shall not be disclosed to other
third parties or reproduced without prior written permission of HMSA,
0 Copyright Hitachi Medical Systems America 1997, All rights reserved.
-. - - - .- -

L5E-BMA 105-2
Table of Contents

Revision History

Revision Effective
Date Comments Initial
Number Pages
L5E-BMA 105- 1 10 / 17195 All Preliminary Draft Issue JJR
-..-.___--.
Revised to include coil positioning
L5E-BMA 105-2 4/l 8197 All JJR
figures

1.0 Component List


HS/MR QD Bilateral Breast Coil. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Qty. 1
TorsoPad ...................................................... ..Qty. 1

AnklePad........................................................Qty. 1
TopFillerPad......................................................Q~ 2
BottomFillerPad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ,Qty,2
Instruction Manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Qty, 1

2.0 Product Introduction


2.1 The HS/MR QD Bilateral Breast Coil is intended to provide high quality imaging of
breast anatomy. The following steps describe the Installation and Adjustment
procedures for this coil.

Page 2 L5E-BMA 105-2


3.
3.1 The HS/MR QD Bilateral Breast Coil is intended to provide high quality imaging of
breast anatomy, The HS/MR QD Bilateral Breast Coil makes use of a dual cup,
single housing design with proprietary multiple receiver quadrature coil design for
improved S/N Ratio. A Torso pad is provided, and attaches to the caudal edge of
the breast coil with velcro material for improved patient comfort. A pair of Breast
Coil Filler Pads may be used separately, or in combination, (none, one, or two), to
reduce distortion and vibration effects on the breast. Refer to Figures 1A and 1 B
below for component description.

Assembled Breast; Coil

II HS/MR QD Bilateral Breast Coil

Page 3 L5E-BMA1052
4.0 Coil Positioning
4.1 The HS/MR QD Bilateral Breast Coil should rest directly on the Patient Table: for the
MRP- 7000, remove the Auxiliary Tabletop and pads from the Patient Table, and
for the AIRIS, remove the pads and Trough Filler Pads from the Patient Table. Refer
to Figure 2 below.

4.2 Remove any Filler Pads which may be present in the cups of the HS/MR QD
Bilateral Breast Coil.

Small Trough Filler Pad,

Breast Coil .

Figure 2. Coil Positioning on the MRP-7000 and AIRIS Patient Tables

Page 4 L5E-BMA 105-2


4.3 Place one each of the provided phantoms in each cup. If necessary, use a towel
or pads to keep the phantoms upright in the cups, and so that the phantoms do
not touch the edges of the cups

4.4 plug the cable into the connector on the left inner surface of
front facade: and align the D-shaped connector before inserting.
IS, plug the cable into the connector located at the end of the patient table,
and align the D-shaped connector before inserting.

4.5 Advance the patient table until the positioning lights are over the HS/MR QD
Bilateral Breast coil landmark lines, which indicates the center of the cups. If
necessary, center the coil Left-Right on the Patient Table, Press Set to advance
the coil to Isocenter. Refer to Figure 3 below.

Landmark Lines

. Coil Landmark Lines

5.1 Access the COIL DATA ENTRY program and register the HS/MR QD Breast as the
coil name for coil #29.

5.2 Using the COIL DATA ENTRY program,set up the coil as follows:
RF Line 1 (A - only)

5.3 Using the SIGNAL GENERATION CHECK program, set the TUNE Step to ’ 1,’ and set
the TUNE value as follows:
TUNE 24

5.4 Access the Tuning A port, and adjust for the following (refer to Figure 4):

Impedance / Phase
200 rt 50 R -15”<a< 15”

5.5 Run a SCANOGRAM (FOV = 300mm) and ensure the TUNE is within the following
value range:
TUNE 20-2A

Page 5 L5E-BMA 105-2


Figure 4. Coil Tuning Port Locations
5.6 Using the COIL DATA ENTRY program,set up the coil as follows:
RF Line 3 (B - only)

5.7 Using the SIGNAL GENERATION CHECK program, set the TUNE Step to ’ 1,’ and set
the TUNE value as follows:
TUNE 24

5,8 Access the Tuning A port, and adjust for the following (refer to Figure 4):

Impedance Phase
160+20R -15”<a< 15”

5.9 Run a SCANOGRAM (FOV = 300mm) and ensure the TUNE is within the following
value range:
TUNE 20-2A

NOTE
Ensure that the TUNE A and TUNE B values are within 6H of each other (1 tic),

5.10 Access the Coil Data Entry program, and set up the coil as follows:
Coil BREAST
RF LINE 2 (A and B)
All 0
TUNE STEP 3 (MRP-7000/AIRIS only)

Page 6 L5E-BMA 105-2


5.11 Perform the Automatic RF Gain Adjustment using transverse (TRS) and coronal
(COR) test planes. Calculate the average of these two results, and use the
Manual RF Gain Adjustment to enter the result for Coil #29.

5.12 Perform an SNR Testwith the following steps:

512.1 Move the table to the OUT limit,

512.2 Use the alignment lights and the table Lateral controls to center the left phantom,
then press Set to move the coil to Gantry Isocenter,

512.3 Run a SCANOGRAM.

512.4 Perform an SNR test scan using the following parameters:

Sequence SE
TR 500
TE 40
Slice Thickness 5
AVE 2
Projection 256
SQ No. 3
FOV 200
Phase Direction H-F
Slice Plane COR

5.13 Repeat Steps 5.12.1 - 5.12.4 for the right phantom,

5.14 Verify that the SN Ratio is within the specified range according to the following
table; contact HMSA Technical Support if it does not:

5.15 Inspect the SNR image for asymmetrical shading or image artifacts; contact
HMSA Technical Support if any are noted,

Page 7 LSE-BMA 105-2


el

lexible ody Coil

--- Installation

Tokyo, Japan
LZE-B&40508-2

~~~~~~TIALITY : This material contains proprietary information of


Xitachi MedicaI Corporation and it shall not be disclosed to the third party or
reproduced without the prior written permission of Hitachi Medical Corporation.
Copyright@ Hitachi Medical Corporation. 1992,1994. All rights reserved.
Contents

Page

1. Outline ..e . ..* . . . . . . . . . . . . ...*. 1

2. Adjustment . . e . a . * . . e . . . . . . . . - . . . o 1

3. Checking Performance . . . e . . e . v . e . - - - - . . 7

(1) 1 LZE-BM0508
1. Outline

This manual describes setting and adjustment of the flexible


QD body coil which is effective to deliver high quality image
of abdominal area.

2. Adjustment

(1) Detaching cover

Take off the upper part of the leather cover and detach
the two polyvinyl chloride covers shown below.

Detach by removing 8 M4xEf screws

Detach by removing 2 M4x15E

Fig, 1

(2) Setting phantom


Set the phantom in the coil as shown below. Pay
attention to setting way of the phantom as it differs
according to coil model.

-I-
Table 1

MRP model Coil model Coil setting

Ni solution
-- phantom No. 3
MR-QFC-3 1 Auxiliary
tabletop
Tabletop “,.
Q

- Ni solution
phantom No. 4
MR-QFC-32 Auxiliary
tabletop
MRP-5000
‘MRP-20EX-2) Tabletop
a

40mm thick mat

MR-QFC-33 Auxiliary
tabletop --- Ni solution
phantom No. 4
Tabletop
a

40mm thick mat

MRP-5000 MR-QFC-34
. Ni solution

Tabletop
-Q-1 A~~l?“,:a!~~op

Ni solution
--- phantom No. 3
MR-QFC-41 Ta bletop

Ni solution
/ phantom No. 4
M R-QFC-42 i
Ta bletop

G
vlRP-7000-2
-4Omm thick mat

-2- 3 L2E-BM0508
.center
Set the of coil to
the lateral center of the

/ I

After setting Make sure that each center of


the phantom, the top and bottom of the
fix the coil polyvinyl case are vertically
with the velcro at the same position. (Refer
tape. to the Fig. 3 below.)

Fig. 2

Coil Make sure that each center of the top


and bottom are at the same vertical line.

<Good > Tabletop <Nogood>

Fig. 3

-3- 4 L2E-BM0508
(3) Tuning

Carry out the tuning in the same procedures as those in


the Installation Manual of the main MFtP system.

Set the phase to O" and adjust the coil so that <he
impedance at that phase is within the value shown in the
following table. The location of the trimmer capacitors
is also shown in Fig. 4.

Table 2
MRP model Coil model Coil name TUNE value Impedance Phase
Saddle coil
(A side ) 500 t 200
MR-QFC-31 Solenoid coil
(B side) 35Ok 150
Saddle coil
(Aside) 200 t 100
MRP-5000 MR-QFC-32 18 O”k2”
Solenoid coil
(MRP-20EX-2) I (Bside) 35Ok 150
I

Saddle coil
(Aside) 2002 100
MR-QFC-33 Solenoid coil
( B side ) 2505 150
Saddle coil
(Aside) 250F 150
MRP-5000 MR-QFC-34 I Solenoid coil
(B side) 500 rt: 200
Saddle coil
(Aside) 350+ 150
MR-QFC-41 Solenoid coil
(B side) 2OOk 100
Saddle coil
(Aside) 3502 150
MRP-7000-2 M R-QFC-42 18 O”k 2”
Solenoid coil
( B side ) 2OOk 100
Saddle coil
(Aside) 2002 100
M R-QFC-43
Solenoid coil
( B side ) 250 IL 100
Saddle coil
(Aside) 250 + 150
MR-QFC-44 Solenoid coil
( B side ) 250t 150

-4- 5 L2E-BM0508
Trimmer capacitor for Trimmer capacitor for
adjustment of B- side adjustment of A- side
(solenoid coil side) (saddle coil side)

Fig. 4

(4) Registration of coil data

Register the coil data in the same procedures as those


described in the Installation Manual of the main system.

RF LINE = 2 ( LINE A and B )


ATT = To be decided by the following procedures.

Calculate the ATT value by measuring each SN ratio of


the solenoid coil and saddle coil and substitute the
measured SN ratio for the following equation.

Procedure :

0 Set the phantom, (similarly to the phantom (2))

@ Switch the RF line as follows by using the coil data


entry program.

Solenoid coil : RF LINE = 3


Saddle coil : RF LINE = 1

@ Using the coil not in use, m a k e the RF Gain


adjustment and measure the SN ratio.
Use the same sequence as that of (6) for S ratio
measurement with single slice.

-5- 6 1.2E-BM050R
(Refer to the Installation Manual of the main system
for RF Gain adjustment.)

@ Calculate ATT value.

ATT = 20.log (Solenoid coil SN ratio/Saddle coil SN ratio)

@ Register the ATT value calculated in @ by using the


coil data entry program. Input coil No. of the coil
not'in use.

(5) Adjustment of RF Gain

Referring to the Installation Manual of the main system,


carry out the adjustment of RF Gain. Input coil No. of
the coil in use. (However, for MRP-7000 register the
MR-QFC-41...10, MR-QFC-42...11, MR-QFC-43...12, MR-QFC-
44 . ..16)

(6) Measurement of SN ratio

Take a scanogram and check to see that the .TUNE value is


within the range of OC - 4E.
Calculate the SN ratio with the following sequence.

Sequence : SE
TR : 500
TE : 40
Slice thickness : 5 (10 for MR-QFC-31, 32, 33, 34)
AVE :2
Projection' : 256
SQ No. : 3
FOV : 260 (for MR-QFC-31, 41)
320 (for MR-QFC-32, 42, 33, 43)
400 (for MR-QFC-34)
420 (for MR-QFC-44)

Check to see that the SN ratio is within the specified


range described in the following table.

7 L2E-RM0508
Tokyo,Japan
Contents

Page

1. Outline . . . . . . . . . . . . . . . . . . . . . . . . 1

2. Coilmodelvs.systemmodel . . . . . . . . . . . . . . 1

3. Adjustment . . . . . . . . . . . . . . * . . . . . . . . 2 ..' -:

4. Supply list . . . T ._..~.. .'.. . . - . . . . . . e . . . 11 -4

. 1
1. Outline

This device is the re ional receiver coil capable to ima


the temporo mandibular ) with high sensitivity.
Referring to this manual and the, Installation Manual of the
main system, carry out the adjustment of the coil.

2. Coil model vs. system model

model of the T coil is different according to t Y


_... the main systems ke sure the coil el .4
B
referri to the folio i;
[
Table 1 TMJ coil model vs. system model

System model

I MR-TMC-21 MRP-20-2 (QD coil is not yet modified)


MRP-20-2 (QD coil is modified)
M R-TMC3 1 MRP-20EX-2
MRP-5000
MRP-7000-1
MR-TMC~l
MRP-7000-2

,.

-I- 2
3, Adjustment

(1) Setting coil No,

Use the coil No. not in use for the MR-TMC-21 and MP-
TMC-31, and '9" for the MR-TMC-41.
Use the same coil No. for each of the two coils for the
both sides. Put the designated No. seal on each of the
two coils at the position on the leather cover of the
tuning box as shown in Fig. 1. The seal should be put
on the side from which the portion of putting the velcro
_. - -. .tape and leather cover together on the coil can be seen. *' 1
1
After putting the No. seal on the coil, set the two coil
of which cable outlets are in the same direction and
make sure that the sides with the No. seals face
oppositely each other.

TMJ coil TM1 coi I

Put No. seal on the back

\ I
Set the coils so that their cable outlets are

i ‘. \ in the same direction, consequently the sides


with No. seal face ODDOSib?ly each other.

r
..

canbeseen. -

Set them facing


in the same dir&ion
P u t No.seal here.’ \ ,
\

Fig. 1 Place to be put on with No, seal

-2-
(2) Phantom

Use the PVA i solution inside


diameter 165~) for adjustment.
Pay attention to the impedance value to be adjusted and
consequent SN ratio since they vary depending on type of
the phantom.

(3) Detaching cover

Detach the leather cover of the tuning box a remove


three.sere e cover of the

) Setting coil

The impedance adjustment and checking SN ratio of the


coil must be carried out on each of the two coils
separately.

As shown in Fig. 2, mount the coil so as to closely


contact it on the edge face of the phantom and fix it by
using the velcro tape. In this case the coil can be
faced in any direction.

Next, set the phantom mounted with the coil on the


tabletop so that the coil loop faces in the X-axis
direction, In this case adjust the position of the
utting mats or the like so that the edge
ositioned about 30 from the
mark of the light localizer in the X-axis-an t h e venter
of the coil loop is set to the center of the gantr

Connect the receiver cable to the accessory junction


cable. The junction cable has two input connectors but
any of them can be used and the other can be left
unused. Connect the connector of the junction cable to
the gantry, then start the adjustment.

-3-
Side with operation caution label

Toward gantry

Set the cross mark ofthe lighi


localizer to the center of the
- - ff coil and phantom.

UII CPblr
-iye fkoupf
Fix the co$with velc~otape.
lint.)
X

Ni solution phantom No. 3 (4165)

Toward gantry

o i l with velcro tape.

Connectthe ca
to the junction cable

’ adjustthe height by putting mat.

WA phantom

-4- 6
(5) tuning

Carr out the tunin rocedures as t


described in the Installation nual of the main system.

E value to the value shown in the Table 2 and


the phase to O"flOo by adjusting the trimmer capacitor,
The trimmer capacitor is shown in Fig. 3.

Trimmer capacitor for phase adjustment

Knob for decoupling


Do not touch this
because it has been
adjusted.

Remove three screws.

Fig,, 3 Location of trimmer capacitor

justing the phase to o"+loo, make sure that the


impedance value at that time is within the range shown
in the table 2.

Table 2 Criteria for tuning

-5-
(6) Registration of coil data

This procedure should be performed for MRP-20EX and


MRP-7000. It is also necessary for MRP-20 software
version V4.OA (CPU 68000) and after, and the software
version V2.OA (CPU 68020) and after.

Carry out the registration of the coil data in the same


procedures as those of the Installation Manual of the
main system.

RF LINE = 3 (LINE B)

Use the coil No. set in the step (1).

(7) RF Gain adjustment

Carry out the RF Gain adjustment in the same procedures


as those of the Installation Manual of the main system.

Use the coil No. set in the step (1).


Keep the record of the.RF Gain after adjustment.

(8) Registration of RF Gain

After completion of the adjustments up to the step (7)


for both the two coils, calculate the arithmetic mean of
the BP Gain values of the two coils. Then, substitute
the calculated RF Gain value for the coil No. set in the
step (-1). This RF Gain value is used commonly for the
unilateral and bilateral modes.

(9) Measurement of SN ratio

Measure SN ratio for both the two coils respectively.


Take a coronal scanogram and make sure that the TUNE
value is within the range of the value shown in the
Table 3.

Then, measure SN ratio with the following sequence.

-6-
easurement conditions

1 Sequence I SE I SE I

Slice plane SAG SAG

T-R 500 500

On the scanogram set the slice plane to the position of


5th~ from the edge of the phantom. (Refer to Fig. 4)

After completion of the measurement, make sure by using


the SN ratio evaluation program that the SN ratio is
within the range shown in the Table 3.

50mm
I COR image
I

Ni solution (No. 3) PVA

Fig. Slice line for SN ratio measurement

-7-
Table 3 Criteria for TUNE value and SN ratio

Coil model Phantom TUNE value to SN ratio


be set to be checked
WA 30-48 100+20
MR-TMC-21
Ni solution No. 3 30-48 130+30
PVA oc-3c 100,+20
MR-TMC-31
Ni solution No. 3 oc-3c 130230
MR-TMC-41 Ni solution No. 3 78-48 125+20
hiZ-TM c-51
(10) Operation check in bilateral mode
i
Check the operation in the bilateral mode as follows.
Mount the.two receiver coils on the phantom as shown in
Fig. 5 and set the phantom in the gantry so that the
coil loop faces in the X-axis.

Head t
Caution plate for operation

Setthe crOn mark of the light


localizer to the center of the
Set
her mT0
into me
the gantry.
gantry.

Fix the junction box on the


LLEii%k,~:
tabletopwith velcrotape.

Connect to the gantry.

a. Ni solution phantom No. 3 (+ 165)

5 Coil setting in bilateral mode

-8-
Head band

Set the cross mark of the light


Set into the oantrv localizer to the center of the
coil.

Fix the junction


box on the
tabletop with
velcro tape. X Y
Connect to the gantry.

Fig. 5 Coil settin in bilateral mode

After taking a scanogram, ke sure that the TUNE value


is within the range shown in the Table 4.

alue to be checked

(11) Attaching covers

Attach the cover of the tuning box and leather cover in


the reverse rocedure of the step (3). When attaching
the leather cover, set it as the copper pi
coil an e seen,

-‘$jr
10
(12) Checking operation

Make .sure that normal scan can be performed with the


human body in the unilateral and bilateral modes.

-IO-
4. Supply list

(2) Junction cable a e . e e . . . 1

(3) Head band . e . * . * . e * . 1

(4) Coil number label . e . . . e 2

4.

-'dl-
MR Imaging System

Model MRP

TMJ Coil
MR-TMC-21
MR-TMC-31
MR-TMC-41
MR-TMC-51

TECHNICAL GUIDE
m-m Installation & Adjustment ---

HITACHI MEDICAL CORPORATION


Tokyo, Japan
L2E-BM0509-3

CONFIDENTIALITY : This material contains proprietary information of


Hitachi Medical Corporation and it shall not be disclosed to the third party or
reproduced without the prior written permission of Hitachi Medical Corporation.
Copyright @ Hitachi Medical Corporation. 1992,1995. All rights reserved.
Contents

Page

1. Outline . . . . . . . . . . . . . . . . . . . . . . . . 1

2. CoilModelvs.SystemModel . . . . . . . . . . . . . . 1

3. Adjustment . . . . . . . . . . . . . . . . . . . . . . . 2

4. Supply List . . . . . . . . . . . . . . . . . . . . . . 11
1. Outline

This device is the regional receiver coil capable to image


the temporo mandibular joint (TMJ) with high sensitivity.
Referring to this manual and the Installation Manual of the
main system, carry out the adjustment of the coil.

2. Coil Model vs. System Model

Since model of the TMJ coil is different according to the


model of the main system, make sure the coil model by
referring to the following table.

Table 1 TMJ Coil Model vs. System Model

TMJ coil model System model

I M R-TMC-21 I MRP-20-2 (QD coil is not yet modified)


MRP-20-2 (QD coil is modified)
MR-TMC-31 MRP-20EX-2
M RP-5000
M RP-7000-0
MR-TMC-41
MRP-7000-1
I M RP-7000-2

I MR-TMC-51 I AIRIS
3. Adjustment

(1) Setting coil No.

Use the coil No. not in use for the MR-TMC-21 and MR-
TMC-31, and "9" for the MR-TMC-41, '1101' for the MR-TMC-
51.
Use the same coil No. for each of the two coils for the
both sides. Put the designated No. seal on each of the
two coils at the position on the leather cover of the
tuning box as shown in Fig. 1. The seal should be put
on the side from which the portion of putting the velcro
tape and leather cover together on the coil can be seen.

After putting the No. seal on the coil, set the two coil
of which cable outlets are in the same direction and
make sure that the sides with the No. seals face
oppositely each other.

TMJ coil TMJ coil

Put No. seal on the back

’ Set the coils so that their cable outlets are


in the same direction, consequently the sides
with No. seal face oooositely each other.

The portion
the velcro ta
leather cove
can be seen.

Set them facina


in the same direction.
\ Put No. ieal here.’

Fig. 1 Place to be Put on with No. Seal


(2) Phantom

Use the PVA phantom or Ni solution phantom No. 3 (inside


diameter 165mm) for adjustment.
Pay attention to the impedance value to be adjusted and
consequent SN ratio since they vary depending on type of
the phantom.

(3) Detaching cover

Detach the leather cover of the tuning box and remove


three screws to detach the cover of the box.

(4) Setting coil

The impedance adjustment and checking SN ratio of the


coil must be carried out on each of the two coils
separately.

As shown in Fig. 2, mount the coil so as to closely


contact it on the edge face of the phantom and fix it by
using the velcro tape. In this case the coil can be
faced in any direction.

Next, set the phantom mounted with the coil on the


tabletop so that the coil loop faces in the X-axis
direction. In this case adjust the position of the
phantom by putting mats or the like so that the edge
face of the phantom is positioned about 30mm from the
mark of the light localizer in the X-axis and the center
of the coil loop is set to the center of the gantry in
the Y and Z-axes.

Connect the receiver cable to the accessory junction


cable. The junction cable has two input connectors but
any of them can be used and the other can be left
unused. Connect the connector of the junction cable to
the aantrv, then start the adjustment.
Side with operation caution label

Set the cross mark of the


localizer to the center of
coil and phantom.

* UM cable
(Active
decoupling

to the junction cabl

* *NC /cable
Adjust the height by putting mat.
(Signal line)

* Only for MR-TMC-51 Ni solution phantom No. 3 (~$165)

Toward the gantry


et the cross mark of
he light localizer to
h e center of the
oil and phantom.

e coil with velcro tape.

Connect the cable


to the junction cable

Adjust the height by putting mat.

PVA phantom
Tuning

Carry out the tuning in the same procedures as those


described in the Installation Manual of the main system.

Set the TUNE value to the value shown in the Table 2 and
the phase to O"+lOo by adjusting the trimmer capacitor.
The trimmer capacitor is shown in Fig. 3.

Trimmer capacitor for phase adjustment

Knob for decoupling


Do not touch this
because it has been
adjusted.

Remove three screws.

Fig. 3 Location of Trimmer Capacitor

Adjusting the phase to O"rtlOo, make sure that the


impedance value at that time is within the range shown
in the table 2.

Table 2 Criteria for Tuning

TUNE value Phase to be Impedance to


Coil model Phantom
to be set adjusted bechecked

PVA 54 0”+10” 1500*500


M R-TMC-21
Ni solution No. 3 54 0”*10” 2000+500
PVA 30 0”*10” 350&l 00
M R-TMC-31
Ni solution No. 3 30 0”+10” 500t100
1MR-TMC-41 1 Ni solution No. 3 1 I 0”*10” 550&l 50
MR-TMC-51 Ni solution No. 3 1.9 [V] 1 0”+10” 550+150
(6) Registration of coil data

This procedure should be performed for MRP-20EX and


MRP-7000 and AIRIS. It is also necessary for MRP-20
software version V4.OA (CPU 68000) and after, and the
software version V2.OA (CPU 68020) and after.

Carry out the registration of the coil data in the same


procedures as those of the Installation Manual of the
main system.

RF LINE = 3 (LINE B)

Use the coil No. set in the step (1).

(7) RF Gain adjustment

Carry out the RF Gain adjustment in the same procedures


as those of the Installation Manual of the main system.

Use the coil No. set in the step (1).


Keep the record of the RF Gain after adjustment.

(8) Registration of RF Gain

After completion of the adjustments up to the step (7)


for both the two coils, calculate the arithmetic mean of
the RF Gain values of the two coils. Then, substitute
the calculated RF Gain value for the coil No. set in the
step (1). This RF Gain value is used commonly for the
unilateral and bilateral modes.

(9) Measurement of SN ratio

Measure SN ratio for both the two coils respectively.


Take a coronal scanogram and make sure that the TUNE
value is within the range of the value shown in the
Table 3.

Then, measure SN ratio with the following sequence.


Measurement conditions

M R-TMC-21
Scan parameter MR-TMC-41,51
M R-TMC-31

I Sequence SE SE

1Slice plane
I TR

17-E
I Thickness I IO I 5
I AVE I 2 I 2 I

On the scanogram set the slice plane to the position of


50mm from the edge of the phantom. (Refer to Fig. 4)

After completion of the measurement, make sure by using


the SN ratio evaluation program that the SN ratio is
within the range shown in the Table 3.

50mm 50mm
COR image image

- _._/_” -s-- ---w--


\
\

Slice l i n e

Ni solution (No. 3) WA

Fig. 4 Slice Line for SN Ratio Measurement


Table 3 Criteria for TUNE Value and SN Ratio

TUNE value SN ratio to be


Coil model Phantom
to be set checked

PVA 30 - 48 100+20
MR-TMC-21
Ni solution No. 3 30 - 48 13ort30
PVA oc - 3c 100+20
MR-TMC-31
Ni solution No. 3 oc - 3c 13ort30
MR-TMC-41, Ni solution No. 3 18-48 125+20
MR-TMC-51

(10) Operation check in bilateral mode

Check the operation in the bilateral mode as follows.


Mount the two receiver coils on the phantom as shown in
Fig. 5 and set the phantom in the gantry so that the
coil loop faces in the X-axis.

Caution plate for operation

’ Light iocalizer

Set the cross mark of the light


localizer to the center of the
Set into the gantry. coil.

Fix the junction box on the Z


tabletop with velcro tape.
x Y
Junction cable

\ Connect
Connect to the gantry.

a. Ni solution phantom No. 3 (~$165)

Fig. 5 Coil Setting in Bilateral Mode


Light localizer ,

\ C^CIL^ r”^rr -^“I, ^*&I-^ ,:-kc


3cc LllrC,“S, ,lldlK”, LllC l l c j , l L

Set into1the gantry. localizer to the center of the


u A

Fix the junction


box on the Z
tabletop with

\ Connect to the gantry.


velcro tape.

b. WA phantom
x&y
Fig. 5 Coil Setting in Bilateral Mode

After taking a scanogram, make sure that the TUNE value


is within the range shown in the Table 4.

Table 4 TUNE Value to be Checked

I Coil model
I
TUNE value to
be checked I
1 MR-TMC-21 1 OC - IE (
1 MR-TMC-31 1 OC - 48 1
1MR-TMC-41,51 1OC - 48 1

(11) Attaching covers

Attach the cover of the tuning box and leather cover in


the reverse procedure of the step (3). When attaching
the leather cover, set it as the copper pipes of the
coil and the PCB cannot be seen.

(12) Checking operation

Make sure that normal scan can be performed with the


human body in the unilateral and bilateral modes.
4. Supply List

(1) TMJ coil . . . . . . . . . . . . . . 2

(2) Junction cable . . . . . . . . . . . 1

(3) Head band . . . . . . . . . . . . . 1

(4) Coil number label . . . . . . . . . 2

E
MR Imaging System

Model AIRIS, MRP

QD Wrist Coil

TECHNICAL GUIDE
--- Installation & Adjustment ---

HITAC EDICAL CO
Tokyo, Japan
L2E-BM5171-5

CONFIDENTIALITY : This material contains proprietary information of


Hitachi Medical Corporation and it shall not be disclosed to the third party or
reproduced without the prior written permission of Hitachi Medical Corporation.
Copyright @ Hitachi Medical Corporation. 1997,1998,1999. All rights reserved.
Contents

Page

1. General . , . . . . . . . . . . . . 1

2. Composition . . . . . . . . . . . . 1

3. Adjustment . . . . . . . l . . . . 2

4. Confirmation . . . . . . . . . . . . 8

5. Recording sheet . . . . . . . . . . . . . . . . 9
1, General

This manual describes the Installation and Adjustment for


the QD wrist coil with which a high quality image can be
produced as compared with the standard joint coil.
Refer to
the Installation and Adjustment for the main unit together
with this Installation Manual.

2. Composition

Coil . . . . . . .
Sponge . . . . . .
Mata . . . . . .
Mata . . . . . .
Band . . . . . . .
phantom No.6 ...

Mat

phantom No .6
3. Adjustment

(1) Removing the cover

Remove the lower and upper cover.


(2) Phantom setting

Set the phantom onto the coil and connect the upper and
lower coil as illustrated in the diagram. Take care
not to be tilted in the X, Y, and Z axis.

Phantom
(Accessory pha

(3) Tuning

Adjust the phase and impedance as in the tuning of


HS/MR head coil in the Installation and Adjustment for
the main unit. The adjustment for each trimmer
capacitor is indicated in the following. The phase and
impedance to be adjusted are indicated.

-4
Tuning voltage at Impedance to be Phase to be
Coil name
adjustment adjusted adjusted
I I Saddle coil I
(line A)
MR-QWC-41
18 (1 [VI) 200 ?I 1 oon 0”3-30”
MR-QWC-51 Solenoid coil
(line B)

MR-QWC-31A 78 (1 [VI) 15oJc1oon O”i30”


Line B (solenoid coil)

Line A (saddle coil) adjusting place

When adjusting QD wrist coil, put the coil so that the


saddle PCB comes up side and adjust it in the main
unit.
After the adjustment has finished, set the coil so that
the saddle PCB comes below side as shown in the figure
above and make sure to check that each adjusted value
is within the standard range.

(4) Coil data registration

Register the coil data of the coil in accordance with


the Installation and Adjustment for the main unit.

RF LINE = 2 (LINE A and B)


ATT = It is determined from the following
procedures
TUNE STEP = 3

Fordetermination of attenuation value (ATT value) of


QD wrist coil, measure S/N ratio of solenoid and saddle
side of QD wrist coil respectively, and obtain the ATT
value from the S/N ratio.

Procedures:

0 Set the phantom (in the same way as in tuning


adjustment).
@ Change RF LINE by COIL DATA ENTRY program.
Solenoid side : RF LINE = 3
Saddle side : RF LINE = 1

@ Scan an image for S/N ratio measurement and measure


the S/N ratio. Refer to item (6) of this Manual for
sequence of S/N ratio measurement.
Scan is to be performed in single slice.
Be sure to scan only after the RF GAIN adjustment is
done for each scan. (Refer to the Installation
Manual for RF GAIN adjustment.)

@ Calculate ATT value as follows:


ATT value = 2Oelog (Solenoid side S/N / Saddle side
S/W

@ Round off the ATT value obtain in @ and set this


value by COIL DATA ENTRY program.
If ATT value is 7 or more, input 7.

(5) RF GAIN Adjustment

Adjust the RF GAIN in accordance with the


Installation and Adjustment for the main unit. The
Coil No. should be the same as used in (4).

(6) S/N ratio measurement

Check the tuning voltage if within the range OC N 4E


by taking a SCANOGRAM. The measured S/N ratio
sequence is indicated in the following:
Sequence : SE
TR : 500
TE : 40
Slice thickness : 5 ("10" in case of MR-QWC-31A)
AVE. (NSA) :2
Projection : 256
SQ No. :3
FOV : 120
Check the S/N ratio if it is within the range in table
below.

Type SIN Standard Coil No. 1


I MR-QWC-41 I 2OOIfi-30 I 21 I

MR-QWC-51 2OOrt30 12

MR-QWC-31A 2OOk30 12
Reference range of S/N ratio
I I
I I

240

230

220

--------C----------------~-----------------~-------------

210 1
I
I
I
I
I
I I I
I I I
I I I
I f I
I I I
I I I
--------t-----------------~-----------------~-------------

200 t I I
f I I
I I I
I I I
I I I
I I
I I i
I I I
--------+----------------~-----------------~-------------

190

180

170

- --------C----------------~-----------------~-------------

160 I
I I
I I
I I I
/-
r 1 I
I
/ I

----&? 20 25 30 Temperature
(“Cl

Reference range of S/N ratio while using the QD-Wrist Coil


(SE 500/40, one slice, one echo)
4. Confirmation

Confirm if a normal image can be obtained in the human body.


5. Recording sheet:

Adjustment & check data recording sheet

(1) MR -QWC- -

- Impedance : ohms (200 +/- 100 ohms)


(150 +/- 100 ohms)

- Phase : deg. (O" +/- 30°)

- Tuned value: (OC to 4E)

(2) S/N ratio check

- S/N ratio : (200 +/- 30)

- Imaging room temperature : OC


MR Imaging System

Model AIRIS Series / MRP Series

Latchable Joint Coil

MR-JCL

TECHNICAL GUIDE
--- Installation & Adjustment ---

HITACHI MEDICAL CORPORATION


Tokyo, Japan
LZE-BM5136-4

CONFIDENTIALITY This material contains proprietary information of Hitachi


Medical Corporat,ion and it shall not be disclosed to the third party or reproduced
without the prior written permission of Hitachi Medical Corporation.
Copyright 0 Hitachi Medical Corporation. 1996, 1997, 1999,ZOOO. All rights reserved.
CONTENTS
Page

1. Adjustment of the MR-JCL Coil . . . . . . . . . . . . . . . . . . . .._.............................................................................. 1

2. Registration of Coil Data ................................................................................................................ 2

2.1 AIRIS Series .......................................................................................................................... 2


2.2 AIRIS-II Series ...................................................................................................................... 3
2.3 MRP-7000 Series.. ................................................................................................................. 4
2.4 MRP5000AD.. ....................................................................................................................... 4
2.5 AIRIS-mate.. .......................................................................................................................... 4

3. Check of S/N Ratio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..5


3.1 AIRIS and MRP-7000 Series.. .............................................................................................. 5
3.2 AIRIS-II ................................................................................................................................. 6
3.3 MRP5000AD.. ....................................................................................................................... 7
3.4 AIRIS-mate.. .......................................................................................................................... 8

4. Operation Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..~.. 9

5. Recording Sheet . . . . . . ..._.................................................................................................................... 9


I. Adjustment of the MR-JCL Coil
The coil adjustment is effected in the same manner as described in the Technical Guide -
Installation & Adjustment. Only the different jobs from those in the Technical Guide will be
described here.
(1) Check the TUNE value and adjust the impedance with the phantom supplemented to the
system (nickel solution phantom No.2)
(2) The trimmer capacitors Cm and Cr for adjusting this coil are located as shown below.
Adjust the capacitors similarly to the other coils.

Cr

Cm
(3) Set the phantom and coil to the gantry.

Example Setup of Phantom

(4) The coil adjustment values should be as follows.

Tuned voltage Range of tuned value Adjustment


System name Phase
to be set to be checked impedance
AIRIS
OC - 36
MRP-7000 (1 Fl)
3002 100 Q
AIRIS-II 12 - 54
(G-l)
ot30°
MRP-5000AD OC-36
(1 ‘i,
2002 100Q
AIRIS-mate 12 - 54

2. Registration of Coil Data


2.1 AIRIS Series
Model: AIRIS- Coil: MR-JCL-52
Start the coil data entry of service mode and enter the coil name. Enter “3” for LINE in this i
case.

Coil name Coil No.


LATCH JOINT 24

Model: AIRIS- Coil: MR-JCL-62


Start the coil data entry of service mode and enter the coil name. Enter “3” for LINE in this
case.

I Coil name 1 Coil No. 1


1 LATCH JOINT 1
2.2 AIRIS-II Series
Model: AIRIS II-2 Coil: MR-JCL-52
After starting the system table editor and selecting the coil data entry, enter each parameter
referring to the table below:
Coil Name LATCH JOINT I
NO. 1 Any No. not reserved
Physical Coil No. 62 I
Attenuation Value I 00 I
Tuning Measure Points 22
RF Gain Start Value OeO
RF Gain Adjustment Step 16
Adjustment Average 1
Sensitivity Correction Coefficient 64
First Modulate Decoupling 64
Second Modulate Decoupling
- - 00
Third Modulate Decoupling 2f
Forth Modulate Decoupling 00
Selector I 1 I
Number of Channels I 1 I
Shading Correction I No I
Bed Movement I No I
Recen tion Line I 1 I
QD Attenuation Value I 0 I
Method A I

Model: AIRIS II-1 Coil: MR-JCL-62


After starting the system table editor and selecting the coil data entry, enter each parameter
referring to the table below:
Coil Name LATCH JOINT I
NO. Any No. not reserved
Physical Coil No. 75
Attenuation Value 00
Tuning Measure Points 22
RF Gain Start Value OeO
RF Gain Adjustment Step 16
Adjustment Average 1
Sensitivity Correction Coefficient 64
First Modulate Decoupling 64
Second Modulate Decoupling 00
Third Modulate Decoupling 2f
Forth Modulate Decounline 00
Selector 1
Number of Channels 1
Shading Correction No
Bed Movement I No I
Reception Line 1
BD Attenuation Value 0
Method A
2.3 MRP-7000 Series
Model: MRP-7000 Series Coil: MR-JCL-42
Start the coil data entry of service mode and enter the coil name. Enter “3” for LINE in this
case.

Coil name Coil No.


LATCH JOINT 17

2 . 4 MRPdOOOAD
Model: MRP-5000AD Coil: MR-JCL-32A
Start the coil data entry of service mode and enter the coil name. Enter “3” for LINE in this
case.

Coil name Coil No.


LATCH JOINT 15

2.5 AIRIS-mate
Model: AIRIS-mate Coil: MR-JCL-72
After starting the system table editor and selecting the coil data entry, enter each parameter
referring to the table below:
Coil Name LATCH JOINT
NO. Any No. not reserved
Physical Coil No. 61
Attenuation Value 00
Tuning Measure Points 22
RF Gain Start Value OeO
RF Gain Adjustment Step 16
Adjustment Average 1
Sensitivity Correction Coefficient 64
First Modulate Decounline: 64
3. Check of S/N Ratio
3.1 AIRIS and MRP-7000 Series
(1) Adjust the RF gain in the similar manner to that described in the Technical Guide -
Installation &Adjustment of the main unit.
(2) Set the following scan parameters and make a scan.

SLICE PLANE : TRS SEQUENCE : SE


2D/3D : 2D SEQ No. : 3
REPHASE : OFF PHASE DIR : A-P
REJECTION : OFF RECTANG : OFF
ANTI-WRAP : OFF PRESAT : OFF
ADAPTIVE IMAGE : OFF C. POSITION : 0
FOV : 260 TR : 500
TE : 40 THICKNESS : 5
MULTI SLICE : 1 ECHO : 1
NSA : 2 PROJECTION : 256

(3) Check if the S/N ratio is within the range shown below:

270

250

230

210

.g
g 1 9 0
i -3

170

150

130 . . . . . ...-.r--------------.---------,------------------------*,____-___________

110 -~---‘----:‘---‘------------------~------------------------.---------.----.-

..
:

\\ I I I
)I
20 25 30
Phantom temperature (“C)
3.2 AIRIS-II
(1) Set the following scan parameters and make a scan.
Basic group Technique group
Slice Plane : TRS S. Presat : Off Matrix group
Sequence : 2D SE Rephase : off Gating group
FOV : 260mm Phase Dir : A-P Dynamic group
TR : 500 AM1 : 25 Filter group
TEl : 40 Half Echo : off Rect.FOV : Off
FA : 90 Bandwidth : 7.2 Anti. Aliasing : off
Multi Slice : 1 MTC : off Reconmatrix : Auto
Multi Echo : 1 Gating : off
Angle# : 1 Particular group Cont. Agent : Off
Thickness : 5mm Multi Echo : 1 Breath-hold : off
Interval : 50mm Dynamic Scan : off
Dual Slice : Off Trancation : off
Freq# : 256 T2 Correction : off
Phase# : 128 Shading : off
Half Scan : Off Adaptive : off
NSA : 1

(2) Check if thc : S/N ratio is within the range shown below:
1YCI - ;8 ; I

1110/--

170

1.60

.s t 1
i!! 1.50 ---------:---------------~---------------~-----------
I
I I
? I I
t t
0 I I I
I
1 ,
I 1 :
6 8 8
---------~--------------- +-------- ----+-----------

0 It 20 25 30

Phantom temperature (“C)


3.3 MRP-5000AD
(1) Adjust the RF gain in the similar manner to that described in the Technical Guide -
Installation &Adjustment of the main unit.
(2) Set the following scan parameters and perform a scan.

SLICE PLANE : TRS SEQUENCE : SE


2D/3D : 2D SEQ No. :3
REPHASE : OFF PHASE DIR : A-P
REJECTION : OFF RECTANG : OFF
ANTI-WRAP : OFF PRESAT : OFF
ADAPTIVE IMAGE : OFF C. POSITION : 0
FOV : 260 TR : 500
TE : 40 THICKNESS : 10
MULTI SLICE : 1 ECHO : 1
NSA : 2 PROJECTION : 256

(3) Check if the S/N ratio is within the range shown below:

290

280

270

260
.o
E
2 250
a

240 . . . ------.------------------------.---------.--------~~---~;-~~~.-.---------.
230
m ::::::::~~~:::::::::::::::::
220
j Lower limit I

210
\\

\\ I I I
20 25 30
Phantom temperature (“C)
3.4 AIRIS-mate
(1) Set the following scan parameters and make a scan.
Basic group Technique group
Slice Plane : TRS S. Presat :. Off Matrix group
SecIuence : 2D SE Rephase : Off Gating group
FOV : 260mm Phase Dir : A-P Dynamic group
TR : 500 AM1 : 25 Filter group
TEl : 40 HalfEcho : Off Rect.FOV : Off
FA : 90 Bandwidth : 7.2 Anti. Aliasing : Off
Multi Slice : 1 MTC : Off Reconmatrix : Auto
Multi Echo : 1 Gating : Off
Angle# : 1 Particular group Cont. Agent : Off
Thickness : 1Omm Multi Echo : 1 Breath-hold : Off
Interval : 50mm Dynamic Scan : Off
Dual Slice : Off Trancation : Off
Frecl# : 256 T2 Correction : Off
Phased : 128 Shading : Off
Half Scan : Off Adaptive : Off
NSA : 1

(2) Check if the S/N ratio is within the range shown below:

22C

210

200

190 1” . . . . r~ ,._.,.__._..._____.._ _ ___.___ TI ..-..__..l._._” --..--_---_ i ------.--._.--...

i
0

‘3 1 8 0
z
3

170

160

150 ..-. “I _....I . ..- “...“.““” . .. I “...“_’,“..----..-----..--.-....~......~.-------..

&\\ !
140 ..,.“.....: WI
.._. _ -...Ai.....-.. _ __.. ;- .-.- _ _.____.._______ ---A.-.....-.
__ .-___.
\ Ldwer llmrt I
\I

-“.J~L--J-.--.*.---L
20 25 30

Phantom temperature (“C)


4. Operation Check
Check if normal images can be acquired for the human body.

5. Recording Sheet
( 1 ) MR-JCL-

l Impedance (MRP-7000, AIRIS, AIRIS II: 3001fr 100 Sz)


(MRP5000AD, AIRIS-mate: ZOO& 100 Q)
0 Phase o (O-t-30”)
0 Tuned value (MRP5000AD, MRP-7000, AIRIS: OC-4E)
(AIRIS II, AIRIS-mate: 12-54)
(2) S/N ratio check
l S/N (MRP-7000, AIRIS: 190+ 40)
(MRP5000AD: 250*25)
(AIRIS II: 145 i- 30)
(AIRIS-mate: 180+ 35)
l Scan room “C
temperature

E
ect:

OOOSPA~ES ............-..~~.......~..~.~..~..~...........~.~.......................~.........~..~.....................~............*.......-.......
.. 2

P7000(2)SPARES(SERIALN 7133 AND VP) ....I.......................*.....................*..................... 10

AIRISSPARES ......~..............~..~...........*.............~................................~...~.......~..~.................~.........................
. 13

. . . . . . ..~..~.~I.........................~....~...........~.............................................~.............~.................. 16

S ...............~...............~.......~.....*..............................*..............*.............*............................
.... 17

N T .0....................*.......~............................................................,..................................... 9

TEST EQV NT,TOOLS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..*..................................................... 19

RDWARE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...~....................................................................................
.................... 20

OO~SVPPLIES ......................~.....~................~.................................................................................
.... 21

~AGNETICFIELD COMPENSATIONUNIT ............................................................................................ 21

COIL eOMPONENTS ........................................................................................ 22


M R P 5 0 0 0 COIL COMPONENTS.. ......................................................................................................... 22
M R P 7 0 0 0 COIL COMPONENTS ............................................................................................................ 24
AIRIS COIL COMPONENTS ................................................................................................................. 26

~RP~AIRISRE~E~VERCOILEXPLODEDVIEWS ................................................................................ 29

Copyright@ 1996 Hitachi Medical Systems America, Inc.. All rights reserved.
I
MRP5000 SPARES

___._ - -_..-- . _ -- _.. . . . . . . .-.-- __...._.. _I-..-..- ....I...-..... -._.


Console 14” Monitor 1117165A
Console 3M Universal Pixel Clock Interface Kit PCKIOOO
Console 5” Blank OD Cartridge SW1046
Console AD SEL PCB 1111342A
Console Air Filter (Control Console) R62 12207
Console Backplane 1220738A
Console Control Console Pwr Sw Cover 56480512
Console COM FD PCB 111580lA
Console DC PS Es-600-S K6211011
Console DC Ps Hk- 1 O-24 1327196D
Console DC PS Hk-1 l-5 1327 196F
Console DC PS Hk-12-12 1327 196E
Console DC Ps Hk-8-5 1327 196A
Console DISP 5B PCB 1111073A
Console ECG Keycap , 1222377D
Console Fan Motor / 1420628B
Console Fan Motor (AC fan for card rack) 1420628A
Console Fan Motor (Md & Od) L5420308
Console Floppy Disk Drive 13 16077C
Console Interphone Amplifier PCB 1314431B
Console IPALU PCB 113934XA
Console IPDM2 PCB 114884XA
Console IPDM3 PCB 1111582A
Console IPDM4 PCB 1118708A
Console IPSM PCB (Interchange with XA is OK) 113933XB
Console IPSM PCB (Interchange with XB is OK) 113933XA
Console IPSQ PCB 113935XA
Console Keyboard (S/N 1 - 83) 128939XB
- Console
- - Keyboard (S/N 84 & UP) 1236241A
Console Led Hdsp-7803 J3761708
Console Led Hdsp-7808 J3761709
Console Led Hlmp-2500 J3761702
Console Magnetic Disk 140mb R6210112
Console Magnetic Disk 250mb R6210117
Console Md Mounting Bracket 1447045A
Console MPU 2 PCB 1112693A
Console MR PNL3 PCB 111715319
Console MR PNL4 PCB 1117154A
Console MT-ODT PCB 115132XB
Console OD Cartridge Cleaning Kit OC-CK 1
Console OD Drive Cleaning Kit OCKlOOO
Console Opt Disk Formatter PCB R62 10307
Console Optical Disk Drive R62 10306
Console PNL CNT PCB - ALL S/N’S 11167165:
Console PNL, CNT PCB - S/N 73 & BELOW 1116716Q
Console Protective Screen for CRT 137064x4
Console PSC PCB 1111341A
Console RESP Keycap 1222377E
Console SCSI FD PCB 1115800A
SCSI2 PCB 1 1118814A
Track Ball 2” - Above 4068 - Above 7023 1225942A
1223476A

Copyright@ 1996 Hitachi Medical Systems America, Inc.. All rights reserved.
2
Copyright@ 1996 Hitachi Medical Systems America, Inc.. All rights reserved.
3
System Shim Plates MR-SP- 1
System Transformer 1327814E
Table 1111 Arm Rest Pad 1325065A
Table Bearing WBCS-1022A i MR-BP-I
Table 1 Collar B 1 MR-CB-1 I
Table EM Bearing Pin MR-EMB- 1
Table EM Lever A
Table Head Fixing Pad
1116387A
Table -.. Head Support ..- 119594XA
Table Head Support Cover 1115220A
Table Knee Support Pad 1227400A
Table Magic Tape (Velcro...Loop) 1327204B
Table Mat For Aux Table Top AM 1003
Table Mat For MR-TP- I (Thick) EM1005
Table Mat For MR-TP- 1 (Thin) / EM1004
Table Microswitch 156510158 I
- 1 Patient Table Motor SS40E4-E-C-75 MR-LDM- 1
1Patient Table Rotary Encoder K65 1 A003
1 Res Mattress-Body 1433335A
1Restraint Stran / 1326575B
Restraint Strap 2 1 1432571A
Restraint Strap A 1 1432572A
Restraint Strap B / 1432573A I
Table 1Roller For Table 1 1431861A
Table 1 Shim j 1432574A
Table Shim
Table
-._--- Solid State Relay D2W2 1 OEd
-____----

Table T-Cont PCB (4013 and up, SIB#I 2)


Table T-Cont PCB (400 I-401 2, SIB#l2) 1112690A
1
Table .-.II 1Table Speed Up Mod MR-SP-MOD
Table Table Top Ext (1 Board only - no pads) MR-TP- IP
Table Table Top Extension Set (2 Boards & 4 Pads) MR-TP- 1
Table Table Top Mattress j 1325555A
Table 1Table Top Mattress 1 / 1324485B
Table
l-._-._ll--ll_ Table Top Mattress 1 1324485C
_-Table
I ___- - - - . -
Table Top Mattress 2 1324486B
Table Table Top Mattress 2 1324486A
1 Table Table Top Mattress-Bd 1433335B

Copyright@ 1996 Hitachi Medical Systems America, Inc.. All rights reserved.
5
MRP7000 SPARES (serial number 7001-7132 ref SIB #175, 179)

Console14” Monitor 1117165A


Console3M Universal Pixel Clock Interface Kit PCKlOOO
Console5” Blank OD Cartridge SW1046
Console682MPU PCB 1117701A
1 ConsoleADBM PCB 1121253A
Console
1 CP PCB [ 1121252A
I
1 Console
1DC PS Ews 1500 K62 11026
ConsoleDC PS EWS 600 K62 11024
ConsoleDC PS Hk-l( I-5 1327 196Q
ConsoleDC Ps Hk- 11-24 12243OOC
ConsoleEL Displav Panel 1 1339774A
Console
1 EL-IF PCB ) 1226935A
IC onsole
1ETL Backplane Fuse 1 1456687A I
I
I
ConsoleFan Motor (Md & Od) ’ L.5420308
Console
HBCON PCB 1120060A
Console
HDSPS PCB (Non NTSC capable) 1120160A
ConsoleHDSPSA PCB (NTSC capable) 1123835A
LConsole HIPDM-A PCB / 1118421A
Console
1 HIPDM-B PC 23 1 111842lB
1 Console1 HIPMBM PCB (Backplane) ( 1123993A
ConsoleInterphone Amp PCB 1 1339347A
ConsoleKeyboard (S/N l-89) 1 1226374A
ConsoleKeyboard (S/N 90 A;ND up) 1236242A
ConsoleMagnetic Disk 1. lgb Q224000 1 S
ConsoleMagnetic Disk 250mb R6210117
ConsoleMd Mounting Bracket 1447045A
ConsoleMR CAM PCB 1226687A
1
1 1338910A
tConsole 1MR EL SW PCB
I I
ConsoleMR PNLS PCB lllYY08A
ConsoleMR PNL6 PCB 1119911A
ConsoleOD Cartridge Cleaning Kit OC-CK 1
ConsoleOD Drive Cleaning Kit OCKIOOO
Console
Opt Disk Formatter PCB R62 10307
1 Console Optical Disk Drive R62 10306
Console
1 PCON-C PCB 1121142C
1 Console f PIP ALU PCB 1118420A
ConsolePIP SQ PCB 1118419A
ConsoleProtective Screen for CRT 137064x4
Console PSC2 PCB 1120501A
Console SCCNT PCB / 1119094A
Console Track Ball 2” - Above 4068 - Above 7023 1225942A
1 Console 1Track Ball 3” - Up To 4068 - Up To 7023 1223476A
Console Window Controller [ 1323720A
Filter Box
Filter Box 1 MR-FB-2
FLT BOX BUI LKHEAD CONN N TYPE FEMALE
Filter Box J8351110
IGantrv I +/- .I deg Thermistor J3610301
390uH Choke Coil - Preamp Mod CL390
Gantry I Aux Heat I/F PCB MR-AHM-5
Gantry 1Capacitor - EC
Gantry Capacitor - ECMK 4YC 36 1
Gantry Capacitor - ECMK-4YC 39 1 J90 / 51751374
, _--._^
Gantry C*anacitor -
Gantry
i--I--~. - Capacitor -.-- Fe<
CopyrightO 1996 Hitachi Medical Systems America, Inc.. All rights reserved
6
Gantry Center light projector j 1338414A
Gantry 1Fan Motor j 1433524A
Gstlwv
-_..__ Floor Heat Insulator 1338736A
.-Gs Halogen Lamp J8510201
I
frantrv Heat Insulation C a n 1440667B

1 Heater Element 35V. 75W 1213037F


Gantry MIC PCB Cz2069-F 1324572B
Gantry MRI Warning Label 1328321A
Gantry Pre Amp Module 122897 1A
Gantry ( Preamp Coaxial Cable Assembly / MR-PCL-7
Gantry 1Preamp Protection Diode 1 J312WOlO
Gantry 1 Preamp2 PC Board / 1223311B
Genrrv
--.._. !3~Receiver Connecting Harness / 1341994A 4
Gantry 1 S-CD PCB ( 1119860A
Gantry 1 S-LD PCB j 1119859A
Gantry 1 S-RD PCB / 1119861A
Thermostat Cs-7Sa / K2351301

MRI 1 OK Trimnot (GC CONT PCB)


MRI ( 12 Amp Fuse - Gc Amp 1 J8211954
MRI 1 1M TrimDot (I TURN GC CONT PCB) / MR-TRIM-1M 1
MRI 1ADC Module / 121309714
/ I 1AF Amn Module 11213096H
i
MRI Backup Relay K235 1202
MRI CAP (HV) 68OOuF 450VDC (l-72,1-38) 51471201
MRI CAP (HV) 82OOuF 450VDC (73 up, 39 up) J1471401
1 DC PS Fs-600A-48 1 K6211008

/ Diode P5 1 C / J3421021
1Fan Motor 1 L5420303
MRI 1 Fuse Holder
1- - Gc Amp - ~---.- 1 58221123 4
MRI- 1G CONT2 PCB - Non Fs Echo j 1113493A 1
- ~~
-MRI 1 G CONT2 PCB - With Fs Echo / 1113493c 1
MRI 1Gc Amolifier (Gradient) j 1117323A
MRI 1 GC Amplifier Fan 1 12171233
1 Heater PCB 1 1111521C
IF Module / 1213098B
Indication PCB j 1214123B
.PD
MRI PD Cont PCB 1214124A
MRI RF Amp Module 12130966
MRI SYNTH 1 Module 1213099E
MRI SYNTH2 Mod(Syn2V2) 1213100C
MRI SYNTH3 Module 1213099F
MRI SYNTH4 PCB 12 13724A
/MRI Trans 1 Module / 1 2 1 3 0 9 4 F
1__l_---.-_-l-
MRI ..-- Trans2 Module _l___.-~_.^ 12 13094B
MRI Trans3 Module 12 130946
MRI Trans4 Module 1227212C
Copyright@ 1996 Hitachi Medical Systems America, Inc.. All rights reserved
7
’ RF Power Amp Fan Motor For MR-PA-3 K7300728
RF Power Amp Fan Motor For MR-PA-4 K7300737
RF Power Amp Fan Motor For MR-PA-4 K7300738
RF Power Amp MR-PA-3 CNTL PCB K7300727
RF Power Amp Power Amp Module for MR-PA-3 K7300723
RF Power Amp Pre Amp Module for MR-PA-3 K7300724
RF Power Amp RF Power Amplifier (new type two piece) MR-PA-4
RF Power Amp RF Power Amplifier (old type one piece) MR-PA-3
SIB 1Serv Info Bulletin #1.58 SIB 158
SIB Serv Info Bulletin #98 SIB98
SIB Serv Info Bulletin #99 SIB99
SIB Service Info Bulletin #125 SIB 125
SIB Service Info Bulletin #182(7000) SIB 182-7
System Accessory Rack 1215482A
System Cable 1323940B
System Cable 1323940A
System Caution Label 1339108A
System Dry Chemical For FOV Phantom MR-CH.M-5
System Dry Chemical For Phantom #l MR-CHM- 1
System Dry Chemical For Phantom #2 j MR-CHM-2
System Dry Chemical For Phantom #3 MR-CHM-3
System Dry Chemical For Phantom #4 MR-CHM4
MR-ETL-M7
System Fan Motor For UPS(CPC) FM-UPS- 1
System FOV Fixture Assy 1113287A
System Fov Phantom 1325OllA
System Phantom #3 1220161A
System Phantom #l 1220162B
System- Phantom #2 1220162A
System ( Phantom #4 / 1117719A
Svstem 1 RFRoom 1 MR-RFR-I

Table Cam Follower / CF1767


Table 1 Cam Follower / CF1760 I
Table Guide Rail MR-GDR- 1
Table Head Fixing Pad 1326816A
Table Head Support 119594xA
Table Head Support 1116387A
Table Head Support Cover 1115220A

Table Patient Table Fuse GDL 2.5A 58211955


Table Restraint Strap 2 1432571C
Table Roller For Table 1451679A
Table 1 Shim ( 1432574A
Table 1 Shim / 1434032B
CopyrightO 1996 Hitachi Medical Systems America, Inc.. All rights reserved.
8
Table 1Solid State Relay D2W210Ed 1K1241602
m 1. Ps.m 1,
I aDIe I I op rvrarrrc :ss-Rear j 1228885A
Table TCNT3 PC1B / 1120910A
Table
__-- Table Guide Retrofit Kit MR-TBL- 1
Table Table Top Mat-Front 134 1694A
Table 1Table Top Mat-Rear 134 1693A J
Table Table Top Mattress-Bd 1433335B
Table Table Top Mattress-Ft 1346029A
Table Table Top Spacer 1333868A
Table Tape Switch Assembly MR-TSW- 1

Copyright@2 1996 Hitachi Medical Systems America, Inc.. All rights reserved
9
MRP7000(2) SPARES (serial number 7133 and up)

Console ) 14” Monitor 1 1117165A


Cnnwle 1 3M Universal Pixel Clock Interface Kit / PCKlOOO
1 >-‘-Blank OD Cartridge 1 SW1046 I
1 682MPU PCB j 1117701A I
~683MPUPCB-S/NA013&up-S/N7145&up / 1128670A I
1 ADBM PCB 1 1121253A 1
1 CP PCB 1121252A
’ DC PS EWS 600 K62 11024
Console DC PS Hk- 1 O-5 1327 196Q
Console DC Ps Hk- 1 l-24 12243OOC
Console EL Display Panel 1339774A
Console EL-IF PCB 1226935A
Console ETL Backplane Fuse 1456687A
Console Fan Motor (Md & Od) L5420308
Console HBCON PCB 1120060A
Console HBCON PCB - S/N A013 & up - S/N 7145 & up / 1120060B
Console HDSP.5 PCB (Non NTSC capable) j 1120160A
HDSPSA PCB (NTSC capable) 1 1123835A
HIP PCB 1 1124316A
Console HIPDM-A PCB 1118421A
Console HIPDM-B PCB 1118421B
Console HIPMBM PCB (Backplane) 1123993A
Console ) Interphone Amp PCB 133934719
Cnnsnle 1Kevhoard (S/N 90 AND UP) 1236242A
Console Magnetic Disk 1. lgb / Q2240001 S
Console Magnetic Disk
Magnetic Disk 250mb
2~ R6210117
Console Md Mounting Bracket 1 1447045A
1447045A
Console I MD Signal ” Cable Controller-OD(SIB 206) 1226677D
1 Console
Console 1MD MD Signal
Signal Cable
Cable OD-MD(SIB
OD-MD(SIB 206)
2b6) ’ 1226677C
‘1 1226677C I
Console MR CAM PCB 1226687A
Console 1MR MR EL SW PCB ( 1338910A
1338910A I
Console MR PNLS PCB 1119908A
_ Console 1 MR PNL6 PCB 1 1119911A
1119911A .I
Console OD Cartridge Cleaning Kit 1 OC-CKl
Console OD Drive Cleaning Kit OCKlOOO
Console Opt Disk Formatter PCB R62 10307
Console 1 Ontical Disk Drive R62 10306
Console 1 PCON-C PCB 1 1121142C
Console 1 PCON-DPCB-S/NA013&up-S/N7145&up / I 126826B
Console Protective Screen for CRT j 137064XA
Console PSC2 PCB 1 1120501A
r Console __ SCCNT PCB 1 1119094A
Console l-SCCNT2 PCB - S/N A013 & uo - S/N 7145 & uo
Console 1Track Ball 2” - Above 4068 - Above 7023
Console 1 V3.8A &grade Kit
Console Window Controller 132372OA
Filter Box FLT BOX BULKHEAD CONN N TYPE FEMALE 583.51110
Gantry ~. . ..--.-- 390uH Choke Coil - Preamp Mod CL390
-.!?%!Y. I-Aux Heat I/F PCB / MR-AHM-5 1
Gantry 1 Capacitor - ECMG YC 470590 1 J1751365 I
Gantry ~.-_- 1I Canacitor
1~ - ECMK 4YC 36 1 J90 i 51751373 I
Gantry Capacitor - ECMK-4YC 391 J90 j 51751374
Gantry -.- Capacitor - Ecmk-5Yc 431 J90 j 51751381
Copyright@ 1996 I- Hitachi Medical Systems America, Inc.. All rights reserved.
10
Gantry 1Capacitor - Fe922H 15 1 J 1 51751204
Gantry Fan Motor 1 1433524A
Gantry Floor Heat Insulator 1338736A
Gantry Halogen Lamp J8510201
Gantry Heat Insulation Cap 1440667B
Gantry Heat Insulation Cover 1338730A
Gantry Heater Element 35V, 75W 1213037F I
Gantry Jack Boss Screws MR2040
Gantry MIC PCB Cz2069-F 1324572B
Gantry MRI Warning Label 1328321A
_ -... - -... -
’ Gantry Preamp Protection Diode J312WOlO
Gantry 1 Preamp2 PC Board 12233llB
Gantry 1Thermostat Cs-7Sa K2351301
MRI 1 10K Trimpot (GC CONT PCB) MR-TRIM- 1 OK
MRI 12 Amp Fuse - Gc Amp 58211954
MRI 1 M Trimpot (I TURN GC CONT PCB) MR-TRIM- 1 M
MRI Backup Relay K235 1202
MRI 1CAP (HV) 82OOuF 450VDC (73 up, 39 up) J1471401
MRI DC PS Fs-600A-48 K62 11008
MRI DC PS GT 3W (+/- 15V-3A) K6211502
MRI DC PS GT-2.5 (.5V-3.7A) K62 11507
MRI DC PS GT3 (5V - 3.7A) K62 11543
MRI f
DC PS Hk-10-15 1327196J
MRI DC PS Pws-350 K6211410
MRI DC PS Pws-750 K6211411
MRI DC PS Ves-8-24-Pc K6211408
\ MRI Diode P5 1 C J3421021
I
[ MRI - - J Fan Motor L5420303
MRI / Fuse Holder - Gc Amp -~ J822 1123 .-.
1117323A

Indication PCB
PDCNT2 PCB 1126408A
MRI Power Supply EWS-600-48 (Airis up to s/n 3 I) K6211021
MRI Power Supply HKTl OO-5 FF K6211314
MRI Receiv Module 1235567A
j .~-----l...
MRI Rectifier “-~
RM30D2-M H 1350003
MRI Solid State Relay SF30DPS-HIT--- K1240001
MRI SYNTH4 PCB 12 13724A
MRI SYNTHE Module 1235569A
MRI SYNTHE Module 1235569C I
MRI Trans Module 1235568A
RF Power Amp Fan Motor For MR-PA-3 K7300728
RF Power Amp Fan Motor For MR-PA-4 K7300738
RF Power Amp Fan Motor For MR-PA-4 K7300737
RF Power Amp MR-PA-3 CNTL PCB K7300727

Copyright@ 1996 Hitachi Medical Systems America, Inc.. All rights reserved.
I1
Copyright@ 1996 Hitachi Medical Systems America, Inc.. All rights reserved.
12
_____.. . . -
I Console 14” Monitor 1 1117165A
_l..l-__
I Cnnsole it-3M Universal Pixel Clock Interface Kit I PCKlOOO
Console 15” Blank OD Cartridpe 1 SW1046 1
cnnwle rb82MPU PCB I1117701A
Console /683MPUPCB-S/NA013&up-S/N7145& : UD , 1 1128670A
Console [ ADBM PCB 1 1121253A
Console CP PCB 1 1121252A
Console DC PS EWS 600 1 K6211024
Console DC PS Hk- 10-5 1 13271960
Console DC Ps Hk-1 l-24 1 1224300C
Console EL Display Panel 1339774A
1
Console EL-IF PCB 1226935A
Console ETL Backplane Fuse 1456687A
Console Fan Motor (Md & Od) L5420308
Console HBCON PCB 1 1120060A
Console HBCON PCB - S/N A013 & up - S/N 7145 & up 1 1120060B 1
1
Console 1 HDSPS PCB (Non NTSC capable) j 1120160A
Console HDSPSA PCB (NTSC capable) 1123835A
Console HIP PCB 112431644
Console HIPDM-A PCB 1118421A
t
Console j HIPDM-B PCB 1 1118421B 1
Console / HIPMBM PCB (Backplane) 1 112399314 1
Console 1 Interphone Amp PCB 1 1339347A I
11Console
Keyboard (S/N 90 AND UP) 1 1236242A 1
Sb 1 Q2240001S
CCUlWJlP
-“..-./.- 1 Md MnllntinP
___ _ _-- ..__.. Bracket
_ _ _-
. ..-. 1447045A
Console / MD Signal Cable Controller-OD(SIB 206) 1226677D
. ^A, ,,..,.,-
_---._l--l.--_l_ MD Signal Cable OD-MD(SIB 2061
Console
Console
-_..-_ MR CAM PCB
Console MR EL SW PCB
Console MR PNL5 PCB
Console MR PNL6 PCB
Console OD Cartridge Cleaning Kit
Console OD Drive Cleaning Kit
-Console Opt Disk Formatter PCB
Console Optical Disk Drive 1 R6210306 I
Console PCON-C PCB 1121142C
Console PCON-DPCB-S/NA013&up-S/N7145&up 1126826B
Console Protective Screen for CRT 137064XA
Console PSC2 PCB 1120501A
Console SCCNT PCB 1119094A
Console SCCNn PCB - S/N A01 3 & up - S/N 7 145 & up 1125323A
Console Track Ball 2” - Above 4068 - Above 7023 1225942A 1
I
Console V3.8A Upgrade Kit 15 1365412
Console Window Controller 1323720A
B
Filter - ox 1 FLT BOX BULKHEAD CONN N TYPE FEMALE ( J83.5 111 0
Gantry 1 19011H
_ _ _.___
Choke
_..-..-
Coil
--..
- Preamn
~~~ Mod I CL390
Gantry 1Gantry Top Cover ) 1128825A
-Gantry - - 1Jack Boss Screw

Copyright@ 1996 Hitachi edical Systems America, Inc.. All rights reserved
13
Gantry MRI Warning Label 1328321A
Gantry PDCNT2 PCB 1126408B
Gantry Preamp Protection Diode J312WOlO
Gantry Preamp2 PC Board 122331 IB
Gantry I RPANL 1 PCB - Serial No A001 Thru A019 only 1242849B
Gantry RPANLI PCB - Serial No A001 Thru A019 only 1242848A
Gantry RPANL2 PCB - S/N A013 & Up 1245610A
Gantry TUNBD-A PCB 124287 1 A
Gantr, 1 TUNBD-B PCB / 124287lB
Gantry 1Upper RF Cover 1 1358815A I
MRI 1 IOK Trimuot (GC CONT PCB) 1 MR-TRIM-IOK 1
MRI 12 Amp Fuse - Gc Amp J8211954
MRI 1M Trimpot (I TURN GC CONT PCB) MR-TRIM- 1 M
MRI Active Decoupling Cntrl Circuit PCB (s/n 37 & up) 1130424A
1 MRI 1Active Decoupling Cntrl Circuit PCB (up to s/n 37) 1 1129088A 1
MRI Backup Relay K235 1202
MRI CAP (HV) 8200uF 450VDC (73 up, 39 up) Jl471401
MRI DC PS Fs-600A-48 K6211008
MRI DC PS GT 3W (+/- 15V-3A) K62 11502
MRI 1DC PS GT-2.5 (5V-3.7AI 1 K6211507
MRI 1DC PS GT3 (5V - 3.7A) ’ 1 K6211543 1
MRI 1 DC PS Hk-IO-15 f 13271965 1
MRI DC PS Pws-350 K6211410
MRI DC PS Pws-750 K6211411
MRI DC PS Ves-8-24-Pc K6211408 I
MRI 1Diode P5lC [ J3421021
. __._
MRT 1Fan Motor 1 L5420303
MRI Fuse Holder - Gc Amp J8221123
MRI G CONT2 PCB 1126603F
MRI Gc Amplifier (Gradient) 1117323A
MRI I GC Amolifier Fan / 1217123s 1
MRI Heater PCB 1128573A
MRI Indication PCB 1214123B
MRT Power Sunnlv EWS-600-48 (Airis s/n 32 & UD) I K62lA016
MRI Power Supply EWS-600-48 (Airis up to s/n 3 1) 1 K6211021
MRI Power Supply HKTI 00-5 FF [ K6211314
MRI Receiv Module 1235567A
MRI Rectifier RM30D2-M H 1350003
MRI Solid State Relay SF30DPS-HI -4 K 124000 1
( MRI SYNTH4 PCB 1213724A
MRI SYNTHE Module 1235569C
MRI Trans Module 1235568A
RF Power Amp Final-PA Module 1244269B
RF Power Amp PA-Fan Set 1244269D
RF Power Amp Preamp Module 1244269C
RF Power Amp PS-Fan 1 1244269E
RF Power Amp RF Power Amplifier 1 MR-PA-5
SIB Service Info Bulletin #I90 SIB 190
SIB Service Info Bulletin #I94 SIB 194
SIB Service Info Bulletin #I98 SIB198
SIB Service Info Bulletin #206 SIB206
SIB Service Info Bulletin #210 SIB210
SIB Service Info Bulletin #211 SIB21 1
SIB Service Info Bulletin #212 SIB212
SIB Service Info Bulletin #214 SIB214
/ SIB / Service Info Bulletin #218 1 SIB218 I
Copyright@ 1996 Hitachi Medical Systems America, Inc.. All tights reserved.
14
SIB Service Info Bulletin #22 1 1 SIB221
SIB Service Info Bulletin #222 1 SIB222
SIB Service Info Bulletin 217 1 SIB217
- System _ Accessory } 1Rack 2 - - 1
System Dry Chemical For FOV Phantom j MR-CHM-5
System Dry Chemical For Phantom #I 1 MR-CHM-1
System Dry Chemical For Phantom #2 1 MR-CHM-2
J
System Dry Chemical For Phantom #3 MR-CHM-3
System Dry Chemical For Phantom #4 MR-CHM-4
System Fan Motor For UPS(CPC) FM-UPS- 1 I
’ System FOV Fixture Assy 1 1113287A 1
System Fov Phantom 132501 IA
System Phantom #3 1220161A
*stem Phantom #l 1220162B
System 1 Phantom #2 1 1220162A
J
System Phantom #4 1117719A
System RF Room MR-RFR- 1
System
“- .-- SCSI Cable Set For 68030 Processor MR-SCS- 1
System Shield Cloth
.__ -______ -__--.---.- MR-SC- 1 .--_f
System Shim Plates MR-SP-2
Table Arm Rest Pad 1325065A
Table 1 Cam Follower CFI 767
Table
~11--- Fixing-.-.-----
Band __-__- S3700801
..-Table B
B e l
^~~ t [ 136679OD -i
Table Fixing Belt 1 1366790A
. ^ , ,-fir.- I
Table Fixing Belt J 15bbIYUL

Table Fixing Belt 1 1366790B I


Table
c~II_ Head Fixing Pad 1 1326816A J
Table
_.-- 1 Head Support 3 2 2 9 7 4 1 3 B I
Table Head Sup port Base 1 1345548B 1
Head Support Cover ( 1115220B i
Knee Sup port Pad 1 1227400B
Knee Support Pad 1 1227400A --.--I
Table ..-- -.-- - a sic
..-._.-- . - Tape (Velcro...Loop) / 1327204B 1
Table Matress Pad 1 1246546D
Table Matress Pad 1246546C
Table 1 Matress Pad Ii 1246546A
Table ~_- -Matress - Pad ~I~ 1 1246546B 1
- Table- - -_---” Matress Pad - 150mm x IOOmm x 20mm 1 1366418A 1
Table Matress Pad - 3OOmm x 250mm x 63mm 1350563D
Table Mattress Pad 1246546E
1
Table Mattress Pad 1350563C
Table Mattress Pad - 1455mm x 600mm x 43mm j 1360693A
Table Mattress Pad - 600mm x 280mm x 43mm 1 1350559B
Table 1Mattress Pad - 900m x 360mm x 18mm 1 1350562B
__-.x_-- i
Table Microswitch 56510158
Table Motor/gear Assembly - lateral motion K5 14A007
Table Photo sensor EE-SX67 1 A
Tnhle Photo sensor FF-‘X67
--Y-V,_ 1
__^_^^. - I
Table I Preamp (Solenoid) 1 12358Y4ci 1
Table j Prei Imp (Saddle) / 1235894H
Table 1Preamp Side Coaxial Cable 147 1994B
Table ___- Table Side Coil Connector Assy
l-.-l - - 1243561C
Table Tnne Switch 1466410A
Table / TDRIVI PCB / 1129498A 1

Copyright@ 1996 Hitachi Medical Systems America, fnc.. All rights reserved.
15
MRH1500 SPARES

Console EDSP 400 PCB Rev A 1243153A


Console EIDC 400 271-1541
Console EPSC 400 CCTB 271-1853
Console 1EPSC 4 10 CCTB 271-1858
Console / Fan 764-3016
Console 1Fan (For O/D) L542006
Console 1 Fan (Other) L542048
Console / Full Keyboard 762-6803
Console [ H/D (Dk515) 764-6093
Console 1 Monitor (14”) 762-6150
Console Optical Disk Drive R6210306
Console SCPU 200 CCTB 762-6387
Console 1TFT (Display Panel) 271-1934
Console 1 Track Ball 762-6805
Gantry / Marker Unit 1362052A
Gantry Operator Call 764-03 10
Gantry SW Cover For EMRD 823-1358
GC Unit Control Panel 764-2332
GC Unit / Control Unit 1 764-2031
GC Unit Fan Assy 764-2 100
GC Unit GC Power Amp (Hitachi Spec) 764-2330
Oxford Bursting Disk 3 1 l-5626
Oxford Fan Assy 764- 1109
RF Unit ADC Unit 764-3014
RF Unit Booster Unit 764-3286
RF Unit CNT Unit 764-30 11
RF Unit Dif, Amp Unit 764-3288
RF Unit [ GEN Unit 764-3012
RF Unit 1Neutralizing Capacitor (Hitachi Spec) 764-3062
RF Unit PAC Unit 764-3010
RF Unit PCNT Unit 764-3270 i
RF Unit 1 PCNT-P Unit 764-3271
RF Unit Power Supply 764-2030
RF Unit Power Supply For RF Unit MR-PRF- 1

RF Unit Probe Switching Circuit PCB 1242161A


Copyright@ 1996 Hitachi Medical Systems America, Inc.. All rights reserved.
16
RF Unit j PSC Unit 1 764-3015
RF Unit 1 REC Unit 764-30 13
RF Unit 1 RF Amplifier Chassis MR-RFA- I
RF Unit 1 Vacuum Tube (Hitachi Spec) 764-3003
System 1 Head Phantom 762-8009

J&Control
- Panel
- Assy (Left)

CopyrightO 1996 I-iitachi Medical Systems America, Inc.. All rights reserved
17
Gantry Control Panel Assy (Right) 1129808A
Gantry Marker Unit 1362052A
Gantry Operator Call 764-03 10
SW Cover For EMRD 823-1358
GC Unit Control Panel

RF Unit f Dif, Amp Unit 764-3288


RF Unit 1 GEN Unit 764-3012
RF Unit 1Neutralizing Capacitor (Hitachi Spec) 764-3062
RF Unit 1PAC Unit 764-3010
RFUnit [ PCNTUnit 764-3270
RF Unit 1 PCNT-P Unit 764-327 1
RF Unit 1Power Supply 764-2030
RF Unit / Power Supply For RF Unit MR-PRF- 1
RF Unit / Pre Amp 764-3344
I
RF Unit 1Probe Switching Circuit PCB 1242161A
RF Unit / PSC Unit 764-3015
RF Unit REC Unit 764-3013
RF Unit RF Amplifier Chassis MR-RFA- 1
RF Unit Vacuum Tube (Hitachi Spec) 764-3003
System 1 Filter Box MR-FL- 15A2
Svstem / Head Phantom 762-8009
System 1 Oxygen Monitor MR-OXM- 1
System 1 Oxygen Sensor Element 65-0609 MR-OSE- 1
System / RF Room MR-RFR- 1
System / TV Camera 761-0120
Table j Bias Plug MR-BC- 1

Copyright@ 1996 Hitachi Medical Systems America, Inc.. All rights reserved.
18
5-7-7(2)-A 1ECG Monitor - DS 504 DS504
5-7-7(2)-A Tel Unit Lx3 lOO(Ds502) MR-3 loo-LX
15-S Telemetry Unit MR-3220-LX
5-7-7(2)-A Pulse Cat Sensor (Ds504) MR-GS- 1
5-7-7(2)-A Pulse Cat Sensor (Ds502) MR-GS-1A
5-7-7(2)-A Ecg Monitor (Ds 502) R5300205
5-7-7(2)-A Tel Unit Lx20 (Ds504) R5300206 i
5-7-7(2)-A Ecg Pads (60 Ea Per Box) TE-1119A
5 Resprobe (20 Ea Per Box) TR-04
15-s / Electrode Cable CM8 1 H 147 1530A
15-s 14M Ext Cable CJ320H 1471527A
-~-’ Optical Fiber Gating Sensor
5-7-7(2)-A 1356604B
5-7-7(2)-A ECG Electrodes MR-ELE- 1 I

Half Wave Length Cable (MRP7000 - Airis) j 1343818A 7-7(2)-A


Chart Ret Pen / CP2000 5-7-7(2)-A
Chart Ret Paper (Astro) j CP3000 5-7-7(2)-A
Terminal LUIg l/4” - j E6-14 5-7-7(2)-A
Terminal Lug 5/l 6”
-. .-- j E6-516 5-7-7(2)-A
Temp Control Relay For Startup Heater 1K235 1203 5-7-7(2)-A
t
Coax Connector Removal Tool 1 MR-EXT-I 7-7(2)
Test Leads MR-TST- 1 5-7-7(2)-A
C h a r t Ret P a p e r SSIOOF 5-7-7(2)-A
Tie Wrap 4” TWlOOO-4 5-7-7(2)-A
Tie Wrap 6” TW 1000-6 5-7-7(2)-A
T Handle Shim Tool ST1000 5-7-7(2)
Aqueous Fiber Pen MR-AFP- 1 5-7-7(2)-A-15-S
BNC Adapter for Impedance Meter HP-BNC- 1 5-7-7(2)-A
T Handle Tool - 17mm TH17 A
T Handle Tool - 23mm TH23 A
Power Cord MR-PWC- 1 5-7-7(2)-A-15-S
Shim Fixture Probe 1335258C 7-7(2)-A
Shim Fixture Remote Control Assy 1222990B 7-7(2)-A
Phillips Screwdriver #2 - Titanium MR-SDT-2 5-7-7(2)-A- 15-S
T Handle Tool for Shim Steel MR-STH- 1 S

Scope cover for Tek 2235A j MR-TKC-1 1 5-7-7(2)-A-15-S

Copyright@ 1996 Hitachi Medical Systems America, Inc.. All rights reserved
19
T Handle Shim Tool ’ ST2000 A
Coaxial Connector Removal Tool - Airis only MR-CXT-1 A
BNC - SMB Coax Test Cable MR-CTC- 1 5-7-7(2)-A
Transmit Pulse Pickup Coil Fixture MR-PCF- 1 5-7-7(2)-A
, Scope Probe MR-PRB- 1 , 5-7-7(2)-A- 15-S

Hardware

, --Plastic Screw - FOV Phantom Holder


PM4 Washer - Cover 1 1426595A
( 1429662A
I
Washer 1 1430636C
Bushing 1439492C
Hex Head Bolt Ml430655
Hex Skt Bolt M4 X 8 M3320408

1Grin Anchor 12M I M3401037 I


Floating Nut / M4991002
Plastic Screw M3 X 8 1 M5490308
Plastic Screw M3 X 12 M54903 12
Plastic Screw M4 X 8 M5490408
Plastic Screw M4 X 12 M5490412
Plastic Screw M4 X 20 M5490420
Plastic Screw M6 X 15 M5490615
Plastic Screw M6 X 20 / M5490620
PIstir Screw M6 X 2S / M5490625 I
Plastic Screw M8 X 20 j M5490820
Sems Screw M2 X 10 j M6440210

Sems Screw M6 X 12 1 M6440612


Screws M4 X1 0 / M6513410
Spring Washer 12M M8711012
Plain Washer 12M M8722012
Washer (For Fuse Holder J8221123) W8221123
Nvlon Screws M5490406
Nylon Screws M54904 16
Nylon Screws M5490316
Nylon Screws M5500308
Table Bracket - Gantry Foot Bolt Ml421230
Table Top Covers (Flattened) Screw M564AOOl
Table Side Covers Screw M554A002
Table Side Covers Flathead Screw M5540406

Copyright@ 1996 Hitachi Medical Systems America, Inc.. All rights reserved
3” Wide Co er Ta e j CT3 1
Fingerstock 24in Copper With Adhesive FSTlOOO
Fingerstock 24in Stainless With Adhesive FST2000
Fingerstock 16in Copper FST3000

SC3 if

517-7(2)-A 1AC Probe


..2----- ( 1333089A
5-7-7(2) / Compensating Unit ] MR-FC-2DS
’ 5-7-7(2) 1Cables For Fld Comp Unit MR-FCC- 1
5-7-7(2)-A 1AC MFCU Adjusting Jig MR-JIG- 1
5-7-7(2)-A
..-_--_l_l 1 Compensating
---1 Unit (Ceiling Moun_l MR-FC-2ES
i __l__l_--
5-7-7(2)-A
/ MFCU Extension Cable
---.-_-~---- i K4751102

Copyright@ 1996 Hitachi Medical Systems America, Inc.. All rights reserved
21
MRP / AIRIS Receiver Coil Components

This section contains a numerical listing of all the available MRP and AIRIS
receiver coil components as well as exploded views of each receiver coil.

MRP5000 Coil Components

BNC COAX CONN (COIL SIDE) J8351114


Bnc Coax Conn (Gantry Side) J83.5900 1
Body Coil (Lg rigid solenoid) 1117303A
Body Coil (Sm rigid solenoid) 1117303B
Body Coil Connector (Lower) 58341911
Body Coil Connector (Upper) 58341912
I
Body Under Coil (Solenoid only) 1 1116765A
BREAST COIL 1 MR-MC-31s
Cable Boot (black strain relief) 145944 1 A
Capacitor - Tuning (Black) lo-90pF 51930202
Catch Block 1438666A
Catch Block 1438667A
Catch Block 1438665A
Center Shaft 2 1452714A
Cervical Coil (non reinforced old type) 1220 15oc
Cervical Coil Cover 1236315C
Cervical Coil Latch 1433558A
Coaxial Cable (Bulk RG58) MR-CBL- 1
Coil Cover - Extra Large - Quad Flex 1234453D
Coil Cover - Large - Quad Flex 1234453C
Coil Cover - Large Solenoid 1336196B
Coil Cover - Medium - Quad Flex 1234453B
Coil Cover - Small - Quad Flex 1234453A
i Coil Cover - Small Solenoid 1336196A
Connect Plate 1345885A
Connector Holder HS/MR Flex 1345886A
Cover - solenoid tuning PCB(Quad Flex Coil) 1345879A
Decoupling Diode - Saddle Side (Model ERA-91) J3130901
Decoupling Diode - Solenoid Side MR-DECUP- 1
Ejector 3 1455600A
Ejector 3 1455600B
Ejector L 1452144A
Ejector R 1452144B
Flex Coil Connector W/Guide Pin (FEMALE) 58341948
Flex Coil Connector W/Guide Pin (Male) 58341946
Frame L - Quad Knee Coil 1122143B
Frame U - Quad Knee Coil 1122143A
Guide Pin 1452654A
Guide Pin (Qd Body Coil) 1449013A
Joint Coil Connector (Female) 58341943
Joint Coil Connector (Male) J8341941
Knob 1452145A
Knob 3 1455601A
L Plate 1345884A
Large Flex Coil 1 1118330B
Large Joint Coil ) 1220294C
Latch (Knee Coil) 1 L9370115

Copyright@ 1996 Hitachi Medical Systems America, Inc.. All rights reserved.
22
Latch Grommet ] L9376104
1 L9376105
1 MR-QFC-33s
~I- 1 MR-QFC-32s
Pelvis Coil 1MR-QPC-3 1 S
QD Coil Cover Plate - top ] 1345877A 1
Quad Body Coil 1 1121389B i
Quad Coil Cover Plate 1 1345878A
1 105258XA
Quad Knee Coil MR-QKC-32s
Quad Knee Coil Stand -. QKIOOO
Quad Knee Frame (L) ( 1122961B 1
Quad Knee Frame (U) 1 1122961A I
Receiver Coil Cable (Coil - Gantry Jumper) 1 1325237D
Replacement Coil Coax Cable Assy 1 CLC5000
Ribbon Cable (Bulk) 1 RClOOO
Ring ---ri%2147B /
Ring 1452147A
Seal
._~--l.l___ - - L4560217
Seal L4560 11.5
Shaft 1 1452148A i
Slide Block 3 1341683B
Small Flex Coil 1118330A
Small Joint Coil 133105oc
Small Quad Flex Coil 1 MR-QFC-31s 1
Spine Surface Coil 1 MR-SFC-1 1
1 1452711A
Stopper
.-- - 3 _I -1 1 1452079A
Stopper A --~-.- - 1 1452655A I
Stopper Shaft 3 1 1452078A I
TMJ Coil 1MR-TMJ-3 1 S
mzn,c\1
Tunm Ca acltor (Green ~-JV~JL I JI/J”LI”I
TlO?n?nl

Tuning Diode (All Coils - Gray/Yellow)


.-~ ( 53150202
..TTuning Diode (Solenoid Flex -Black) ] J3150302
i- Under Body Coil Cove
1 XLg Quad Flex Coil

Copyright@ 1996 Hitachi Medical Systems America, Inc.. All rights reserved
23
MRP7000 Coil Components

Body Coil Cable Assy 1121830B


Body Coil Connector (Lower) J8341911
Body Coil Connector (Upper) J8341912
BREAST COIL MR-MC-4 1 S
Cable Boot (black strain relief) 145944 1 A
Capacitor - Tuning (Black) IO-90pF J1930202
Catch Block 1438665A
Catch Block 1438667A
Catch Block 1 1438666A
1 Center Shaft 2 1 1452714A
Cervical Coil (old style non-reinforced) 1121333A
Cervical Coil (reinforced) 1121333B
Cervical Coil Cover 1236315C
Cervical Coil Latch 1433558A
Coaxial Cable (Bulk RG58) MR-CBL- 1
Coil Cover - Extra Large - O u a d Flex 1234453D I
Coil Cover - Large - Quad Flex 1234453C
Coil Cover - Large Solenoid 1336196B
1234453B
Coil Cover - Small - Quad Flex
Coil
__.. Cover
- - - Small
.~ Solenoid )6A
’ Coil D Connector J833273 1
Coil Side Coax Pin (Female) 58332796
Connect Plate 1345885A +
r~C&ector Holder HS/MR Flex 1 1345886A 1
Cover - solenoid tunin

Ejector 3 14556OOA
E.jector L 1452144A
Ejector R 1452144B
Flex Coil Cable (Large) 11218306
Flex Coil Cable (Small) 1121830F

I Flex Coil Connector W/Guide Pin (FEMALE


Flex Coil Connector W/Guide Pin I (Male)
Frame L - Quad Knee Coil
J8341948
1 J8341946
1122143B
Frame U - Quad Knee Coil 1122143A
Gantry D Connector J8332732
Gantry Side Coax Connector (Male) 58332798
Guide Pin 1452654A
Guide Pin (Qd Body Coil) 1449013A
Head Coil Cable Assy 112183OA
Joint Coil Connector (Female) 5834191 1 3
Joint Coil Connector (Male) J834194 41
Knob 1452145A
Knob 3 1455601A
Large Flex Coil 11183306
Latch (Knee Coil) 1 L9370115
I
L Latch Grommet 1 L9376104
/,-Latch Plunger jt - ---“05
LY3/bl
1 Lg Extremity Coil 1 1121334A
Copyright0 1996 Hitachi Medical Systems America, Inc.. A right reset -ved
24
Lg Extremity Coil - Latchable MR-JCL-42s
Lg Quad Flex Coil MR-QFC-43s
Med Qu_ad Flex Coil - MR-QFC-42s
Pelvis Coil MR-QPC-4 1 -S -
QD Coil Cover Plate - top 1345877A
Quad Body Coil 1121387A
Quad Coil Cover Plate
Quad Head Coil
Quad Knee Coil
Q u a d Knee Coil Cover
Quad Knee Coil Stand
Replacement Coil Coax Cable Assembly CLC7000
Ribbon Cable (Bulk)

Seal L4560217
Seal L456011.5

I Shaft
Slide Block 3
Sm Extremity Coil 112133514 d
Small Flex Coil 111833OF
Small Quad Flex Coil MR-QFC-4 1 S
._mzLL- ----1 145271 IA
i
St-Per 3
--_-_____ ~--- 1452079A
Stopper A 1452655A
Stopper Shaft 3 1452078A
TMJ Coil MR-TMJ-4 1 S
Tuning Capacitor (Green 5-30pF) J1930201
Tuning Diode (All
- Coils
- - - -GrayNellow)
- J3150202
Tuning Diode (Solenoid Flex -Black) j 5 3 1 5 0 3 0 2
Under Body Coil Cover 1 1228827A
i
Under Body Coil Substrate j 1120654B
I XLg Quad Flex Coil 1 MR-QFC-44s -t

Copyright@ 1996 Hitachi Medical Systems America, Inc.. All rights reserved
25
AIRIS Coil Components

Breast Coil 1 MR-MC-5 IS 1


11Cable
1 4 5 9 4 4 1 A 1
Boot (black strain relief)
Capacitor - Tuning (Black) IO-90pF J1930202 j
Catch Block 1438665A ’
Center Shaft 2 1452714A
Cervical Coil Latch 1433558A
Coaxial Cable (Bulk RG58) MR-CBL- 1
Coil Cable 1 1367126A
1
__-- Cover
cl&l -- - Extra Large - Ouad Flex 1 12344531)
1 Coil Cover - Large - Quad Flex f 1234453C
Coil Cover - Medium - Quad Flex 1234453B
Coil Cover - Small - Quad Flex 1234453A
l-Connect Plate 1 1345885A j
Connector Assy for Joint Coil (Cervical) 1129772C ’
Connector Assy for Joint Coil (Large Extremity) 1129772D
i
C o n n e c t o r Afor
s s Joint
y Coil ~(Sm Extremitv) 11297725
Connector Assy for Latchable Coil (L8: E x t r e m i t y ) 1 1129772L i
Connector Assy for Quad Flex Coil (Lg) 1129772F I
1 Connector Assy for Quad Flex Coil (Med) 1129772B
/Connector Assv for Ouad Flex Cc oil (Sm) 1129772H
Connector Assy for Quad Flex Coil (XL& 11297726 1
Emrector Assy for Quad Head Coil 1 1129772A 1
Connector Assy for Quad Knee Coil 1129772ET
!
Connector Assy for TMJ Coil 1129772K
Connector Holder HS/ MR Flex / 1345886A
_ .--_^^-
Ejector 3 1 14556UUB
Ejector 3 [ 1455600A

Fixing Band For Head Coil / 1362514A


1 Fixing Pin 1 145074714 I
Fixing Shaft / 1341701A
/
Flex Coil Connector W/Guide Pir I (FEMALE) j J8341948 1
/Flex Coil Connector W/Guide Pir I (Male) j 58341946 I
Frame L - Quad Knee Coil / 1122143B
Frame U - Quad Knee Coil / 1122143A
Guide Pin 1 1452654A
Joint Coil Connector (Female) 58341943
Joint Coil Connector (Male) J8341941
Knnh 1452145A
Knob 3 - 1455601A
Large Extremity Coil MR-JC-5 1 S
Large Quad Flex Body Coil MR-QFCJ3S
Latch (Knee Coil) I L9370115 I
Latch Grommet j L9376104
Latch Plunger 1L7.2
n9-7fz/“I”J
rnc
Latchable Extremity Coil 1MR-JCL-5 1 S
Medium Quad Flex Body Coil 1 MR-QFC-52s
Neck Coil / MR-JC-53s j
PPlVic: rni1 / MR-OPC-51s /
CopyrightO 1996 Hitachi Medical Systems America, inc.. All ights reser v e d .
26
QD Coil Cover Plate - top 1345877A ’
MR-QHC-5 1 S
Quad Knee Coil MR-QKC-5 1 S
Receiver Coil Decoupling Diode H113A003 j
1
Ribbon Cable (Bulk) / RClOOO /
Ring j 1452147B

Seal
Shaft 1 1452148A
Slide Block 3 1 1341683B
Small Extremity Coil MR-JC-52s
Small Quad Flex Coil MR-QFC-5 1 S
Stopper 2 145271 IA
Stopper 3 j 1452079A
Stopper A / 1452655A
i
Stopper Shaft 3 j 1452078A
TMJ Coil jMR-TMJ-5 1 S
Tuning Cgacitor (Green 5-30pF) 51930201
-_---__l_- -.-._------.,
Tu% Diode (All Coils - Gray/Yellow) J3150202 j

CopyrightO 1996 Hitachi Medical Systems America, Inc.. All rights reserved
27
Copyright@ 1996 Hitachi Medical Systems America, Inc.. AII rights reserved.
28
Copyrighta 1996 Hitachi Medical Systems America, Inc.. All rights reserved
29
MU-WS-4
1 D a t e I Sep.30.1993 I
Applied
at S/N
1R e f e r e n c e S . I . / I
105044x.4 QD HEAD COIL

D r a w n bv C h e c k e d b y A p p r- o v- e d. br
- -
, ,; ,I. ’ >k- .xj’ ’ 47, /c,
MR-WS-4
Date Sep.30.1993
Applied
at S/N
rnr;nnnvn nn urnn f-n,, Reference S.I. I
1 “J”f-tAr? w.” 11L.rxv L”IL.

ITEM PART NO. D E S C R I P T I O N Q '7-Y

I 106001XA QD Head Coil 1


2 1337191A Guide I
3 1445445A Guide Rail 2
4 1225871D QD H-I P.C.B. I
5 1225872A QD H-2 P.C.B. 1
6 1225873D QD H-3 P.C.B. I
7 1225874A QD H-4 P.C.B. I
8 1338169A Receiver Cable 1
9 1339108A Caution Label 2 1
10 1327118A Coil Number Label 1
11 1337723A Connector Box Label 1
12 1225874D QD H-4 P.C.B. 1

Drawn by j Checked by / Approved by


,,p /I'
522-l
MR-WS-4

~1
1220150H JOINT COIL (MI

Drawn by Checked by Approved by


..___~~
i ?\ ,\' J$(,/. ;.,,:'7/
523/l
MR-WS-4
D a t e Sep.30,1993
Applied
a t S / N
R e f e r e n c e S.I.
122015OH J O I N T C O I L CM)

ITEM PART IN 0 . D E S C R I P T I O N Q ‘TY

I 1117177c Middle Coil Unit 1

1-l 12350948 Middle Coil 4 1

1-2 1235094A Middle Coil 4 I

l-3 1333009D Copper Board 3 2

l-4 1333009E Copper Board 3 2

l-5 1333009F Copper Board 3 2

1-6 13414728 Tunning Box 5 1

l-7 1450568A Cover 5 I

l-8 1440367A CZ21GM-0 P B Assy 4

l-9 1220086H Tune Board 1

l-10 1441297A Cover 210 3

l-11 1440894A Cover 210-A 1

l-12 58341943 Connector HI F7-40DA I!


-1.2DS(31)

l-13 58341941 Connector /H I F7-40PA 2


j- 1 .27DS(31 I

1-14 1433558A PC Hook 2


j
l-15 13282698 Caution Label E / I

l-16 1431690A Gilt Plate A


l-17 1327118A Coil Number Label 1
I
1-18 1458291A Middle Coil Cover-2 1 ;!
1
l-19 13355388 Coaxial Cable 2
/
l-20 1459441A Cable Boots 1 1

l-21 38390742 TY-Rap : T18R I /

l-22 G1960305 Tube F2 /W h i t e I

2 1327266L Sponge I

3 1327266!.1 Sponye / 5;
!
4 1327266U Sponge .! :

5 12363151) Middle Coil Cover-3 / :I


6 1236315C Middle Coil Cover-3 I ’ i
I I

I !

I
I 1
I
! i i
D r a w n b y /C h e c k e d by 1A p p r o v e d by

s23- 1
MR-QBC-3 1

11213898 QD BODY COIL


~‘-“;I”’
- ,5

/’ 3. /

\
2

2-l
2-5
2-4 n 2-11
/ -$ ,2-5
n
2-l I
MR-QBC-3 1
Date 1 Sep.30.1993
Applied 1
a t S / N

,121389B QD BODY COIL I Reference S. I. 1

Inner Body Coil Unit


Inner Body Coil

Decoupler 2
Tune Board 3
DIVISION 3 P.C.B.
Tune Board 4
Copper Board
Copper Board
Copper Board
Copper Board
Copper Board
wer Board

pper Board

ceiver Cable
hesive Backed Mount

der Body Co1 I Unit

er Body Coil 4
er Body Coi 1 Cover 2

h Block (‘2)
S I O N 4 P.C.B.
S I O N 5 P.C.B.

r Body Coil

ion Label 2

D r a w n b y 1C h e c k e d b y i A p p r o v e d by

_ J/l,,
Cl-l
MR-QBC-31

121389B QD BODY COIL

Center Shaft 2
Coil No. Label

Cable Boots
MR-JC-31
D a t e / Sep.30.1993

I!----4
1220294E JOINT COIL (L)

1 Drawn by Checked by Approved by


MR-JC-31
Date Sep.30.1993
Applied
at S/N
Reference S.I.
1220294E JOINT COIL (L)
ITEM PART NO. D E S C R I P T I O N Q 'TY

1 1115772E Large Coil Unit I


1-l 1214008A Large Coil 1
1-2 13414728 TunnincJ Box 5 1
1-3 1317793c Box Spacer I
1-4 1450568A Cover 5 I
1-5 1330794A Copper Board 2 2

l-6 1431690A Gilt Plate A


1-7 12200866 Tune Board 7 1
l-8 13391088 Caution Label 2 I
l-9 14505568 Cable Cleat 1
l-10 13355388 Receiver Cable 1
l-11 58390742 TY-Rap I
1-12 G1960305 Tube F2 1
1-13 1 3 2 7 1 18A Coil No.LabeI 1
1-14 1459441A Cable Boots I
2 1220293A Large Coil Cover 1
3 1327266H Sponqe I

01-i
MR-JC-32
D a t e Sep.30,1993
-.
Applied
a t S/N
-.
Reference S.I.
13310501) J O I N T C O I L U N I T (S)

El/l
MR- JC-32
1D a t e I SeP.30.1993 I
Appl i ed
at S/N
Reference S.I.
13310501) JOINT COIL UNIT (S)

Small Coil Unit


Small Coil

Copper Board C1020R-1/4H


Gilt Plate A
Coil No. Label
Tune Board Unit 3
Caution Label 2

Tuning Box 5
Cable Cleat
Receiver Cab1 e
Cable Boot
Connector Box Label

Small Coil Cover

Drawn by C h e c k e d b y / A p p r o v e d by

I,‘. A”!
El-l
MR-FBC-31
D a t e / Sep.30.1993
APPI Ied
a t S / N
R e f e r e n c e S . -I---,--
I .
1118330E FLEXIBLE COIL UNIT (S)

t-27

l-8
MR-FBC-31
1D a t e I Sep.30.1993 i

I Appl i ed
at S/N I
Reference S.I. 1
I

1118330E FLEXIBLE COIL UNIT (S) I

ITEM PART NO. D E S C R I P T I O N Q ‘TY

1 1118155E Flexible Coil 1


1-1 1333096A P.C.B. Cover 3 1
l-2 1333095A P.C.B. Cover 2 I
1-3 1333094B P.C.B. Cover 1 4
1-4 1333237B Flat Cable 4
1-5 1333237E Flat Cable 4
Flat Cable 12
1
1
3
1
1
3
5
Rubber Packing 3 I
P.C.B. Cover 1
Rubber Packing 2 1
Plate Nut 2 1
Rubber Packing 1 4
Plate Nut 1 4
Latch Plunger H322-4-l-2
Latch Grommet H322-4-I-2
F BODY 3 P.C.B.
F BODY IA P.C.B.
F BODY IA P.C.B.
F BODY 1A P.C.B. 1
F BODY IA P.C.B. 1
F BODY 2A P.C.B.
Receiver Cable

Cable Boots
Cab1 e Guard
F l e x i b l e C o i i C o v e r (1)
F l e x i b l e C o i l C o v e r (21
Coil No. Label
Caution Label 2

1 D r a w n by j C h e c k e d b y / A p p r o v e d by
MR-FBC-32

~“::133”9

11 183308 F L E X I B L E C O I L U N I T (L)

/
I-31

Checked b y Approved by 1

GI/l
MR- FBC- 3’2
1D a t e I Sep.30.1993 I
Applied
at S/N
Reference S. I.
11 1 R??llR
- - _---II
FI FYIRI
.I-..----
F C-I-III
--.I
IlNlT
- . . . .
I I
._I
1

ITEM PART NO. D E S C R I P T I O N Q ‘TY

1 1118155B Flexible Coi I 1


1-I 13330966 P.C.B. Cover 3 1
1-2 13330958 P.C.B. Cover 2 1
1-3 13330948 P.C.B. Cover 1 4
1-4 1333237C Flat Cable 4
1-5 1333237A Flat Cable 4
l-6 1333237D Flat Cable 12
l-7 1335607A Sponge I
I
3
I
I
3
5
Rubber Packing 3 1
P.C.B. Cover 1
Rubber Packing 2
Plate Nut 2
Rubber Packing 1
Plate Nut 1
L a t c h Pluqer H323-1-l-2
Latch Grommet H322-4-1-2
F BODY 3 P.C.B.
F BODY 1 P.C.B.
F BODY 1 P.C.B.
F BODY I P.C.B.
F BODY 1 P.C.B.
F BODY 2 P.C.B.
Receiver Cable

Cable Boots
Cable Guard
F l e x i b l e C o i l C o v e r (II
F l e x i b l e C o i l C o v e r (21
Coil No. Label
Caution Label 2

Drawn by : Checked by ! Approved by


,,JJA+i i 212 I -;“;:/c-
MR-QFC-3 I
D a t e Sep.30.1993
Applied
a t S / N -
Reference 5.1.
112413lA FLEXIBLE QD COIL (51

/”

HI/l
MR-QFC-31
Date / Sep.30.1993
Applied
at S/N
Reference S.I.
__-_-- *.. .__.. _-__ _- _-_- .-.
ITEM PART NO. D E S C R I P T I O N Q'TY

I 1124132A Flexible QD Coil Assembly 1


l-1 1345908C Sponge (FQDB) 1
l-2 1345908H Sponge (FQDB) 1
1-3 1345908G Sponqe (FQDB) 3
1-4 1345908K Sponge (FQDB) 2
1-5 1345908L Sponge (FQDB) 2
2 1'234453A FQDB Cover I
Caution Label 2 I
Coil No. Label 1
2

Drawn by Checked by Approved by

?Y /I/ 1 ,A! d 4. /CT


Hi-i
MR-QFC-3 1

~I
1124132.4 FLEX1 BLE QD CO1 L ASSEMBLY

// 24

1D r a w n b y I Checked by Approved by 1

H2/1
MR-QFC-31
D a t e Sep.30.1993 ’
Applied
a t S / N
Reference S.I.
124:i321 FLEX1 BLE QD CO1 L ASSEMBLY -
7
ITE M P A R T N O . D E S C R I P T I O N
7; 2 ‘7‘Y

1 1232477A FQD Sheet 1

2 1345882A Base 1 1
3 134588lA Base I

4 1345883A Base 2 1

5 1345881B Base I

6 1345877A Cover 1 1
7 1345879A Cover 3 I
8 1345878A Cover 2 I

9 1345880A Cover 4 1
1 0 1345886A Connector Holder I
1 1 1345884A L Plate I
1:2 1345885A Connect Plate I
1: 3 1121830F Receiver Cable 1
li 1 1124129A =LXQDA P.C.B. I
1:5 1124127A ‘QDB 1 P.C.B. 1
If j 1124130A ‘LXQDB P.C.B. 1
1; 7 1124128A ‘QDB 2 P.C.B. I
II 3 1347155A :QDB 3 P . C . B . 1
1: 1352751A :lat Cable 2
2c 1352751E :lat Cable 2
21 1352751E :lat Cable 2
2; 13527515 :lat C a b l e 2
23 1352751N :lat C a b l e 2
24 13527515 ‘lat Cable 2
25 D2131203 :opper B o a r d 1020R-l/.&H i3
.3Ts-50x-10

26 58390805 ‘etai ner c-102 4


27 58390742 Y-Rap 18R i
i>
28 L9376105 atch Plunger 322-4-l-2
29 L9376104 atch G r o m m e t 322-J-1-2 1:)
30 1459441.4 able Boot 5

--

1 D r a w n by c h e c k e d by 1 A p p r o v e d by 1

H2- I
MR-QFC-32

D a t e Sep.30,1993

Applied
a t S / N
Reference S.1.
1124131B FLEXIBLE QD COIL (M)
5
MR-QFC-32
Date Sep.30.1993
Applied
at S/N

1124131B FLEXIBLE QD COIL (MI I Reference S.I. ! 1

ITEM PART NO. D E S C R I P T I O N Q 'TY

1 11241328 Flexible QD Coil Assembly 1


l-l 13459088 Sponge (FQDB) 1
1-2 1345908F Sponge (FQDB) 1
1-3 1345908M Sponge (FQDB) 3
l-4 1345908K Sponge (FQDB) 2
l-5 1345908L Sponge (FQDB) 2
2 12344530 FQDB Cover
3 1339108A Caution Label 2
4 1327118A Coil No. Label
5 13504508

Drawn by Checked by Approved by


MR-QFC-32
1D a t e I Sep.30.1993 1
Applied
a t S / N I I
1 Reference S.I. / I
11241328 FLEXIBLE QD COIL ASSEMBLY

24

\
2

/ D r a w n b y Checked by Approved by 1

J’/l
MR-QFC-32
D a t e Sep.30.1993
Applied
a t S / N
Reference S.I.
19” t 13R
,illULU CI
.--. FYIRI
I lyll F On
_- f-fill
--_- ASSEMRI.Y
..---.----

ITEM P A R T N O . D E S C R I P T I O N Q ‘TY

1 12324778 FQD Sheet 1


2 1345882A Base 1 I
3 1345881A Base 1
4 1345883A Base 2 I
5 13458815 Base 1
6 I345877A Cover 1 1
7 1345879A Cover 3 1
8 1345878A Cover 2 1
9 1345880A Cover 4 I
10 1345886A Connector Holder 1
11 1345884A L Plate 1
12 1345885A Connect Plate 1
13 1121830F Receiver Cable I
14 1124215H FLXQDA 1 P.C.B. I
15 112.l127H FQDB 1 P.C.B. 1
16 1124130A FLXQDB P.C.B. 1
17 11241286 FQDB 2 P.C.B. 1
18 1347155A ’ FQDB 3 P.C.B. 1
19 13527518 Flat Cable 3
20 1352751F Flat Cable 2
21 1352751F 1 Flat Cable 2
22 1352751K i Flat Cable 2
23 1352751P Flat Cable 2
24 1352751T Flat Cable 2
25 D2131202 ’ Copper Board Cl020R-1/4H 8
0.3Tx”50x-10
26 58390805 1 Retainer / TC-102 4
27 58390742 TY-Rap / Tl8R 4

28 L9376105 Latch Plunger i H322-J-l-2

29 ’ L9376104 Latch Grommet 1 H322-J-1-3 I ”_’


30 1459441.4 i ! Cable Boot 5
/
I
,
/
/ /
/ I
,
I I 1
I

I I ’ I

D r a w n b y C h e c k e d b y / Approve& by I

/lJ. A<!

J2- 1
MR-QFC-33
D a t e Sep.30,1993
Appl i ed
a t S / N I
Reference ST-
1124131C FLEXIBLE QD COIL (L)

5
MR-QFC-33
Date Sep.30.1993
APPI i ed
at S/N
Reference S.I.
1124131c FLEXIBLE QD COIL (L)
ITEM PART NO. D E S C R I P T I O N Q'TY

1 1124132C Flexible QD Coil Assembly I


I-1 1345908A Sponge (FQDB) I
1-2 1345908E Sponge (FQDB) I
l-3 13459081) Sponge (FQDBf 3
l-4 1345908K Sponge (FQDB) 2
1 1124132C Flexible QD Coil Assembly 1
2 1234453C FQDB Cover I
3 1339108A Caution Label 2 1
4 1327118A Coil No. Label 1
5 135045oc FQDB Mat 2

1 Drawn by i Checked by i Approved by 1


MR-QFC-
D a t e Sep.30.1993
Applied
a t S/N
j - -
Reference S.I.
1124132C FLEXIBLE QD COIL ASSEMBLY

28

29
h

/4 24
MR-QFC-33
D a t e Sep.30.1993
Applied
a t S / N
Reference S.I.
124132C FLEXIBLE QD COIL ASSEMBLY

ITEM P A R T N O . D E S C R I P T I O N Q ‘TY

1 1232477C FQD Sheet I

2 1345882A Base 1 1

3 1345881A Base 1

4 1345883A Base 2 I
5 13458818 Base 1

6 13458776 Cover 1 1
7 1345879.4 Cover 3 1

8 1345878A Cover 2 I
9 1345880A Cover 4 1

10 1345886A Connector Holder I


11 1345884A L Plate I
12 1345885A Connect Plate 1
13 1121830F Receiver Cable I
14 1124129c FLXQDA P.C.B. 1
15 1124127C FQDB I P.C.B. I
16 1124130B FLXQDB P.C.B. ’ I
I
17 1124128C FQDB 2 P.C.B. 1
18 1347155A FQDB 3 P.C.B. 1
19 1352751c Flat Cable 2

20 1352751G Flat Cable 2


21 1352751G Flat Cable 2
22 1352751L Flat Cable 2
23 1352751Q Flat Cable 2
24 1352751u Flat Cable I 2

25 D2131202 Copper Board 8


I ;: o,;;-“;;;“y,

26 58390805 Retainer j TC-102 4

27 38390742 TY-Rap Tl8R 4

28 L9376105 Latch Plunger H323-J-1-2 2

29 L937610-1 Latch Grommet h’322-4..,-2 I2


30 / 1459441.-\ Cable Boot 51
I /
I I
I
i

i !

I I 1

D r a w n by ! C h e c k e d b y Approved by

i Ja A. /

K2- I
MR-TMC-31

15113398 TMJ COIL UNIT

6
5

) Drawn bL' Checked by 1 Approved by 1


MR-TMC-3 1
D a t e Sep.30.1993
Appl i ed
a t S / N
Reference 5.1.
TMJ COIL UNIT L
15113398

ITEM P A R T N O . D E S C R I P T I O N Q ‘TY

1 1342427C TMJ Coil 1

2 1342427D TMJ Coil I

3 1342424A TMJ Cable 1

4 1342426A TMJ Head Band 1

5 1327118A Coil No. Label 1

6 1337723A Connector Box Label 1

D r a w n b y I Checked by Approved by

m. k.
MR-TMC-3 1
D a t e Sep.30.1993
Applied

at S/N
-Reference 5. i.
1342427C TMJ COIL

Dratin by Checked by Approved by


MR-Th/lC-31
D a t e Sep.30,1993
Applied-
at S/N
Reference S.I.
t 943437r
I”~‘..1&.,”
TM.1
. . ..-
rnt I
----

ITEM PART NO. D E S C R I P T I O N Q'TY

1 1230908B TMJ Coil ASSY I

2 1342422A TMJ Coil Cover 2 I


3 1342423A TMJ Coil Cover I I
4 1451341A TMJ Coil Spacer 1
5 1456608A TMJ Reinforcing Plate 3
6 13260585 Magic Tape 2
7 1233773A TMJ Leather Cover 1

Drawn by Checked by Approved by

Li:V ; 2L.J. m /c
L2-1
MR-TMC-3 1
D a t e-
Applied
at S/N
Reference S.I. /
1342427D TMJ COIL

Dracn by Checked by j A p p r o v e d by
MR-TMC-31
IDate I sep.30.1993 I

I Applied
at S/N
1I Reference S.I. 1
I

1342427D TMJ COIL 1

ITEM PART NO. D E S C R I P T I O N Q'TY

1 12309088 TMJ Coil Assv I

2 1342422A TMJ Coil Cover 2 1


3 1342423A TMJ Coil Cover 1 1
4 1451341A TMJ Coil Spacer 1
5 1456608A TMJ Reinforcing Plate 3

6 13260585 Magic Tape 2


7 1233773A TMJ Leather Cover 1

Drawn by Checked by 1 Approved by


\
. 1. A//
MR-QKC-32
D a t e
Sep.30.1993
APP~ i ed

i231940A a t S/IV
QD KNEE COIL UNIT 2
Reference S.I.
4
MR-QKC-32

Cl

1231940A QD KNEE COIL UNIT 2 ..---. -_.-- -. _. 1

ITEM PART NO. D E S C R I P T I O N Q'TY

1 12302558 QD Knee Outside Frame Unit I


2 1231941A QD Knee Inner Frame Unit 2 I
3 13391088 Caution Label 2 I
4 1233045E Receiver Cable I
5 1459441A Cable Boot 2
6 1345548A QD Knee Stand 2 1
7 1231943A QD Knee Cover 2 1
8 12319438 QD Knee Cover 2 1
9 12201628 Knee Phantom Assembly I

Drawn by j Checked by Approved by


MR-QKC-32
IDate I Sep.30.1993 I

I Applied
at S/N I
1 Reference S.I. 1
I
I
12302555 QD KNEE OUTSIDE FRAME UNIT

1 Drawn by 1 Checked by ; Approved by 1


MR-QKC-32
Date Sep.30.1993
APPI i ed
at S/N
Reference S.I.
230255B QD KNEE OUTSIDE FRAME UNIT
ITEM PART NO. D E S C R I P T I O N Q'TY

1 1122961A QD Knee Outside Frame 2 I


2 11229618 QD Knee Outside Frame 2 I
3 1452654A Guide Pin 4
4 1452655A Stopper A 4
5 1452144A Ejector 1
6 14521448 Ejector 1
7 1452145A Knob 2
8 1452148A Center Shaft 2
9 1452147A 2
10 L9370115 Duracon Catch TLllS I

Drawn by 1 Checked by / Approved by

is' A' / 4w-k


MR-QKC-32
D a t e
j See.30.1993
APF~ i ed
I
QD K N E E I N N E R F R A M E UNIT 2 a t S / N
Reference S.I. _-.--

12
15
MR-QKC-32
Date Sep.30.1993
Applied
at S/N
Reference S.I.
1231941A QD KNEE INNER FRAME UN1 T 2
ITEM PART NO. D E S C R I P T I O N Q 'TY

1 1122962A QD Knee Inner Frame 2 1


2 1122962B QD Knee Inner Frame 2 I
3 D2131201 Copper Board C1020R-1/4H 6
0.3Tx15x8
4 1231942A QD Knee Copper Board 1 I
5 1231942A QD Knee Copper Board 2 1
6 1231942A QD Knee Copper Board 3 1
7 1231942A QD Knee Copper Board 4 1
8 1345550A QD N-l P.C.B. 2 1
9 1343733A QD N-2 P.C.B. I
10 1343734A QD N-3 P.C.B. 1
11 1345551A QD N-4 P.C.B. 2 I
12 1345552A QD N-5 P.C.B. 2 1
13 1345553A QD N-6 P.C.B. 2 1
14 13455538 QD N-6 P.C.B. 2 1
15 1345554A QD N-8 P.C.B. 2 1
16 1345554B QD N-8 P.C.B. 2 I

Drawn by 1 Checked by Approved by

A *A,' //r, k,
M3- 1
MR-WS-7
D a t e SeP.30.1993 I
Applied

ITEM / PART N O . / I D E S C R I P T I O N IQ’TYI

106001XA QD Head Coi 1


13416838 Slide Block 3

1453788A Cable Cleat


1225871C QD H-l P.C.B.

12258728 QD H-2 P.C.B.

1225873C QD H-3 P.C.B.


1225874B QD H-4 P.C.B.
1225874C QD H-4 P.C.B.

1121830A Receiver Cable

1339108A Caution Label 2


1337723A Connector Box Label
1327118A Coil No. Label

Drawn by 1C h e c k e d by j Approved by

iJA/

S28- 1
RIR-WS-7
Date Sep.30, I993 I
Applied
at S/N I
1 Reference S.I. 1 I
1121332A QD HEAD COIL

Drawn by / Checked by / Approved by 1


MR-WS-7
Date / Sep.30.1993
Applied I-
at_-~
S/N -..- --

1121410A HEAD SUPPORT UNIT 2

Drawn by / Checked by i Approved b\


--.~-
-~.-

531/l
MR-WS-7
D a t e Sep.30.1993
Appl i ed
at S/N
Reference S.I.
,,7tn,nn
1 lill‘
”l ““Ui
S . l
-...
UPA~ SIIPPnRT IINIT
“... 1
7
-

1 TECI PART NO. D E S C R I P T I O N Q ‘TY

1 1341706A Head Support 4 I


1-I 1121975A Head Support 5 I
l-2 1314425D Magic Tape 2
1-3 13144256 Magi c Tape 2
2 1121408A Head Support Base 1
3 13417036 Guide Rail 4 I
4 13417038 Guide Rail 4 I
5 1341702A Base Block I
6 1341701A Fixing Shaft 2
7 13417046 Fixing Shaft Holder 2
8 1450747A Fixing Pin 2
9 1450746A FixinrJ Seat 2
10 L4560217 Seal 1 Class A P9 I
1I 143147XA Set Mark 4
12 L45601 15 Seal 1 Class A P 1 5 2
13 1343033A Head Support Spacer I
14 1452079A Stopper 3 1
15 1452078A Stopper Shaft 3 I

D r a w n by / C h e c k e d bv A p p r o v e d by
--p--.L-

s31-1
MK- w3- i

D a t e Sep.30.1993
Applied
at S/N
Reference S. I.
1121389A QD BODY COIL

- I-18

.- 6

c-1,

2-2

D r a w n by C h e c k e d b y 1 Approved b?
__---~-- - ----- .iL--------
,t.Y, A”/ ) p 1 -Gp /c

5291 I
MR-WS-7

1121389A QD BODY COIL I I I

ITEM PART NO. D E S C R I P T I O N Q ‘TY

1 105665XA Inner Body Coil Unit 1


l-1 105412XC Inner Body Coil 1
1-2 58341912 Connector 4
l-3 1341673A Decoupl er 1
l-4 1341669A Tune Board 1 1
l-5 1341671A DIVISION 1 P.C.B. 1
l-6 1341670A Tune Board 2 1
l-7 1122116L Copper Board 1
1-8 1122116M Copper Board 1
I-9 11221165 Copper Board 2
l-10 1122116F Copper Board 2
l-11 11221166 Copper Board 2
l-12 1122116A Copper Board I
1-13 11221168 Copper Board 1
l-14 1122116C Copper Board I
1-15 1122116D Copper Board 1
l-16 1122116H Copper Board 8
l-17 11218308 Receiver Cable 1
1-18 58390808 Adhes i ve Backed Mount TMS-20 3
1-19 58390742 TY-Rap T18R 3
l-20 1449013A Guide Pin 4
l-21 1459441A Cable Boots 2
2 1121388A Under Body Coil Unit 1
2-l 1438665A Catch Block a
2-2 11206548 Under Body Coil I
2-3 1228827A Under Body Coil Cover I 1
2-4 1340146A Guide Block 2
I
2-5 1438667A Catch Block (2) I 4
2-6 1341672A DIVISION 2 P.C.B. 1
2-7 13416728 DIVISION 2 P.C.B. 1
2-8 1122116K Copper Board 2
2-9 1122116N Copper Board L
x-10 1122116E Copper Board
1-l 1 5834191 1 Connector 1 1’
1
‘-12 1122116H Copper Board / 8
3 1120655A Outer Body Coil 1
4 1438666A Catch Block (1 1 4
5 1339108A Caution Label 2 1 1
6 1455600A Ejector 3 2

Drawn by ( C h e c k e d by / A p p r o v e d by

s29- 1
MR-WS-7

a t S/K
Reference S. I .-- i
-I I7I3RclA
_ _ _ _ ..-
wn --._. rni
Rl~l-lY_-.._I

Center Shaft 2
Auxiliary Top Plate
Magic Tape
Coil No. Label

Drawn b y j C h e c k e d by j Approved by
J/I
Ml?-WS-7
Date Sep.30.1993
Applied -
at S/N
Reference S.I.
11213338 JOINT COIL (LI)

Drawn by Checked by I Approved by

J,n' j 4qy k',


MR-WS-7
fDate I Sep.30.1993 I
Applied /- -

11213338 J O I N T C O I L (Ml I f
I-

ITEM P A R T N O . D E S C R I P T I O N 9 ‘TY

1 1235094A Middle Coil 210 I


2 1235094B Middle Coil 210 1
3 1333OOSD Copper Board 3 2
4 1333009E Copper Board 3 2
5 1333009F Copper Board 3 2
6 1341472B Tunninq B o x 5 1
7 1450568A Cover 5 I
8 1440367A CZ’LIGM-0 P B Assv 4
9 1228837A Tune Board 7 I
10 14505568 Cable Cleat I
11 1441297A Cover 210 3
12 1440894A Cover 210-A 1
13 58341943 Connector HIF7-40DA 2
- 1 .27DS(31 1
14 58341941 Connector HlF7-40PA 2
-1.27DS(31)
15 1433558A PC Hook 2
16 13391088 Caution Label 2 I

17 1431690A Gilt Plate A

18 1327118A Coil No. Label


19 1327266L Sponge 2
20 132726681 Sponge 2
21 1327266N Sponse 2

22 1236315D Middle Coil Cover-2 I

23 1236315C Middle Coil Cover-2 I


24 112183OC Receiver Cable 1

25 58390742 TY-Rap TlBR 1 ’


26 G1960305 Tube F2 White I

27 1459441A Cable Boots I

28 1458291A Middle Coil Cover 2 I

/ /

/ i

L-“r awn by / Checked by 1A p p r o v e d by 1

s30- 1
MR- JC- -11
[Date I Sep.30.1993 I

1121334A JOINT COIL (L)

( D r a w n b y C h e c k e d by : Approved by 1

Dl/l
MK-JC-42
D a t e S e p . 31 0 . 1 9 9 3
Applied
at S/N
1121335A JOINT COIL (S) Reference S.17

D r a w n by Checked by i Approved by

El/l
MR-JC-42
D a t e Sep.30.1993
Applied
at S/N
Reference S.I.
1121335A JOINT COIL c.5)
ITEM PART NO. D E S C R I P T I O N Q 'TY

1 1228836A Small Coil 2 I


2 12288368 Small Coil 2 1
3 13414728 Tunning Box 5 1
4 1450568A Cover 5 1
5 1341676A Copper Board 4 2
6 13416768 Copper Board 4 2
7 1341676C Copper Board 4 2
Gilt Plate
Tune Board 7 1
Small Coil Cover 2 1
Small Coil Cover 2 1
Caution Label 2 1
Receiver Cable I
CZ21GW0 PB Assy
HIF7--LODA

Connector HIF7-1OPA 21
-l.'27DS(31)
I 2
Cable Cleat 1

Coil No. Label


Cable Boots

1 Drawn by / Checked bY j Approved bY (

El-l
CIR-FBC-4 1
D a t e
- SeP.S0,1993
APPI i e d
at, S/N
-.----_____--- __ 1.--- 1
R e f e r e n c e S.1 /
1118330F FLEXIBLE COIL UNIT (S)

‘-5
l-8
MR-FBC-4 1
D a t e Sep.30,1993
Applied
a t S / N
R e f e r e n c e 5. I .
1 I1 8330F FLEXIBLE COIL UNIT (5)

ITEM P A R T N O . D E S C R I P T I O N Q ‘TY

1 1121336A Flexible Coil I

I-I 1333096A P.C.B. Cover 3 I

1-2 1333095A P.C.B. Cover 2 1

I-3 1333094B P.C.B. Cover 1 4

1-4 13332378 Flat Cable 4

I-5 13332378 Flat Cable 4

l-6 1333237D Flat Cable 12

Sponge I

1-8 1335607H Sponge I

I-9 1335607C Sponge 3

I-IO 13356075 Sponge I


l-11 1335607K Sponge 1

l-12 1335607F Sponge 3

l-13 1335607L Sponge 5

I-14 1333103A Rubber Packing 3 I

I-15 1333097.4 P.C.B. Cover I

l-16 1333102A Rubber Packing 2 I

l-17 13330996 Plate Nut 2 1

Rubber Packinq I 4
Plate Nut I 4

l-20 L9376105 L a t c h Plunger H322-4-l-2 21


l-21 L9376104 Latch Grommet H332--1-l-2 I2

F BODY 3 P.C.B. 1

1-23 1228842A F BODY 18 P . C . B .

l-24 12288428 F BODY 1B P . C . B .


I-25 1228842C F BODY 18 P . C . B .

l-26 1228842D F BODY 1B P . C . B .

1-27 1224350F F BODY 2A P.C.B.

l-28 112183OG Receiver Cable

I-29 58390742
Cable Boots

Cable Guard
F l e x i b l e C o i l C o v e r (1)

F l e x i b l e C o i l C o v e r (2)

Coil No. Label


Caution Label 2

1D r a w n b y C h e c k e d b y /A p p r o v e d by 1

Fl-I
MR-FBC-12
D a t e Sep.30.1993
Applied -
a t S/N
11 I833OG FLEX1BL.E C O I L U N I T (L) Reference S.I.

I-31 /F IX

D r a w n b y Checked by Approved by
MR-FBC-42
D a t e Sep.30.1993
Applied
a t S / N
R e f e r e n c e S.I.
1 I1 83306 FLEXIBLE COIL UNIT (L)

ITEM P A R T N O . D E S C R I P T I O N Q ‘TY

1 1121336B Flexible Coil 1


l-l 13330966 P.C.B. Cover 3 I
l-2 13330958 P.C.B. Cover 2 1

l-3 13330948 P.C.B. Cover I 4

1-4 1333237c Flat Cable 4

l-5 1333237A Flat Cable 4


l-6 1333237D Flat Cable 12

l-7 1335607A Sponge I


l-8 13356078 Sponge 1

l-9 1335607C Sponge 3


l-10 1335607D Sponge 1
l-11 1335607E Sponge 1
l-12 1335607F Sponge 3
l-13 1335607L Sponge 5
l-14 1333103A Rubber Packinq 3 I
I-15 1333097A P.C.B. Cover I
l-16 1333102A Rubber Packiny 2 1 1
I-17 1333099A Plate Nut 2 I’
l-18 1333101A Rubber Packiny 1 ;I
l-19 1333098A Plate Nut 1
il
l-20 L9376105 L a t c h Pluger H322-i-l-2 2 /
1-21 L93761O-t Latch Grommet H322--1-I-2 1’ /
1 --22 1335775A F BODY 3 P.C.B. I I

l-23 1228842E F BODY IB P . C . B . ;i


l-24 / 1228842F F BODY IB P . C . B . 1 I
l-25 / 12288426 F BODY IB P . C . B .

1-26 1228842H F BODY IB P . C . B . 1 j.

l-27 12243506 F BODY -‘A P . C . B . I


1-28 1121830F Receiver Cable I 1’
j ;
I-29 J83907-12 TY-Rap / T18R 1
l-30 1459431A Cable Boots I :
I-31 1459838A Cab1 e Guard 20 ;
2 11183298 F l e x i b l e C o i l C o v e r (11 I !
I
3 i 1336196B / F l e x i b l e C o i l C o v e r (21
I ~
‘I ’ 1327118A Coil No. Label !
/
5 1339108A Caution Label 2 1 I
I
I

Drawn b y 1C h e c k e d b y Approved by

-4Tf!,

Gl-1
MR-QFC-4 1
D a t e - 1 Sep.30.1993
--.-

1121131D FLEXIBLE QD COIL (S)


pg
s . i:j-----
;, -..-- 1
5

Hi/l
MR-QFC-41
Date Sep.30.1993
Applied
at S/N

1124131D FLEXIBLE QD COIL (S) L Reference S.I. /

ITEM PART NO. D E S C R I P T I O N Q 'TY

1 1124132D Flexible QD Coil Assembly 1


l-1 1345908C Sponge (FQDB) I
1-2 1345908H Sponqe (FQDB) I
1-3 1345908G Sponge (FQDB) 3
1-4 1345908K Sponge (FQDB) 2
1-5 134590811 Sponge (FQDB) 2
2 1234453A FQDB Cover 1
3 1339108A Caution Label 2 1
4 1327118A Coil No. Label 1
5 13504506 FQDB Mat 2

Drawn by 1 Checked by 1 ,\pproved by

LP.,L! I @? Lt.
HI-1
MR-QFC-11
Date
,";p;:;d
~ Sep'30'1993 '

Reference S.I.
1124132D FLEXIBLE QD COIL ASSEMBLY

Drawn by Checked by iApproved bv


- I
u: 4, ,- j ,+zq k-.
H2/1
MR-QFC-4 I
D a t e / Sep.30.1993 I
Appl i e d
a t S / N
R e f e r e n c e S . I .
1241 1 FLEX B L E QD COIL i
-- --
I TEI P A R T N O D E S C R I P T I O N

1232477A FQD Sheet

1345882A Base 1
1345881.4 Base

1345883A Base 2
13458818 Base

1345877A ,C o v e r 1

1345879A ,C o v e r 3

1345878A tC o v e r 2

1345880A IC o v e r 4

1345886A ( Connector Holder

1345884A I_ P l a t e
1345885A ( Zonnect P l a t e

1121830K I ieceiver Cable


1124129D I ZLXQDA P.C.B.

15 1124127D f‘ Q D B 1 P.C.B.

If 112413oc E‘ L X Q D B P.C.B.

li 1124128D F :QDB 2 P.C.B.

1E 1347155A F :QDB 3 P.C.B.

1: 1352751A F :lat C a b l e

2c 1352751E F:lat C a b l e

21 1352751E F:lat C a b l e

22 13527513 F :lat C a b l e

23 1352751N F‘lat Cable

24 1352751s F ‘lat Cable

25 D2131202 C ‘opper Board 1 CL020R-l/IH


j G..3Tx-50x-10

58390805 R etainer T7-IO2

58390742 T Y-Rap Tl8R

L9376105 L atch P l u n g e r hp”-4-1-2

L9376104 L atch Grommet H322-1-1-z


1459441A Cable Boot

D r a w n bv C h e c k e d by 1 A p p r o v e d by

H2- I
MR-QFC-42
D a t e / SEtP. 3 0 * 1 9 9 3
APFI i e d
a t S / N 1

1124131E FLEXIBLE QD COIL (M) Reference s . I y-----


MR-QFC-42
D a t e Sep.30.1993
Applied
a t S/N

v-
R e f e r e n c e S . I . /
113Al’?lF
. . . . . . Y.Y
FI FXlRl
. --,.----
F QD f-fill
_-
I RI I
--.-

ITEM P A R T N O . D E S C R I P T I O N Q’TY

I 1124132E Flexible QD Coil Assembly 1

l-1 13459088 Sponge (FQDB) 1

1-2 1345908F Sponge (FQDB) I

1-3 1345908M Sponge (FQDB) 3

1-4 1345908K S p o n g e (FQDB) 3

1-5 1345908L Sponge (FQDB) 2

2 12344538 FQDB Cover 1

3 1339108A Caution Label 2 I


4 1327118A Coil No. Label I

5 13504508 FQDB Mat 2

D r a w n b y Checked by / Approved by -

J PA I 34&! m, /c ,

Jl-1
iMR-QFC-42

1124132E FLEX1 BLE QD CO1 L ASSEMBLY

,38

D r a w n b y C h e c k e d bv ,Approved b y
MR-QFC-42

1124132E F L E X I B L E Q D COIL A S S E M B L Y
, / I I
I I
ITEM j PART NO. / D E S C R I P T I O N IQ’TYI

1 12324778 FQD Sheet I

2 i 345882A Base 1

3 1345aalA Base I :
4 i 345883A Base 2 1
5 13458818 Base 1
6 1345877A Cover I 1
7 1345879A Cover 3 1
a 1345878A Cover 2 1
9 1345aaoA Cover 4 I
10 1345886A Connector Holder 1
II 1345884A L Plate 1
12 i 345885A Connect Plate I
13 112183OL Receiver Cable 1

14 11241296 FLXQDA P.C.B. I ’


15 1124127E FQDB 1 P.C.B. I

16 112413OC FLXQDB P.C.B.


I ’
17 1124128D FQDB 2 P.C.B. 1
18 1347155A FQDB 3 P.C.B. 1

19 1352751B Flat Cable

2201 11 33 55 22 77 55 11 FF F
F ll aa tt C aa bb ll ee
C 1I 21
f,

22 1352751K Flat Cable 2,

23 1352751P Flat Cable /I *


24 13527511 Flat Cable
I i ?/
25 D2131202 Copper Board Cl020R-l/-&H
j 0.3Tx”50x-10 / a;
26 38390805 Retainer iTC-102

27 58390742 TY-Rap ’ Tl8R i :,


28 L9376105 Latch Plunger i H322-4- l-2 -7

L937610-1 / Latch Grommet ! H322-4-l-2 I ’


2g I
30 145944lA Cable Boot I 5

Drawn b y )C h e c k e d b y / A p p r o v e d by
^ * I

J2- I
MR-QFC-43
'Date 1 Sep.30.1993
-.--+.
Applied
at S/N I
Reference S.I. /
1124131F FLEXIBLE QD COIL (L)

1 D rawn by ' Checked by : Approved by 1


MR-QFC-43
[Date I Sep.30,1993 I
Appl i ed
at S/N
Reference 5.1.
1124131F FLEXIBLE QD COIL (Ll
-. .-
ITEM PART N0 . D E S C R I P T I O N ‘-I-
Y
I 1124132F Flexible QD Coil Assembly 1

I-1 1345908A Sponge (FQDB) 1


;
I-2 1345908E Sponge (FQDB) I

1-Z 1 345908D Sponge (FQDB) 3


1-4 345908K Sponge (FQDB) 2
I-E I 345908L Sponge (FQDB) 2
i 234453C FQDB Cover 1
cc 339108A Caution Label 2 1
4 13271186 Coil No. Label 1
5 135045oc FQDB Mat 2

!
/
I
/

1 Drawn by ! Checked by ! Approved by

Kl-1
MR-QFC-13

~~1

1124132F FLEX1 BLE QD CO1 L ASSEMBLY

Drawn by Checked by I Approv”d b\

K?/i
MR-QFC-43
D a t e / Sep.30.1993
Applied
1
a t S / N
R e f e r e n c e S . I . /
1124132F FLEXIBLE QD COIL iI -
- .- -. -
I TEI !A P A R T N O D E S C R I P T I O N Q ' TY
-

1 1232477C FQD Sheet 1

2 1345882A Base 1 1

3 1345881A Base 1

4 1345883A Base 2 1

5 13458818 Base 1
6 1345877A Cover 1 1

7 1345879A Cover 3 1

a 1345878A Cover 2 1
3 1345880A Cover 4 I

II 3 1345886A Connector Holder 1

1 1 1345884A L Plate 1

1: 2 1345885A Connect Plate


, 1
1:3 1121830M Ie c e i v e r C a b l e
R 1
11 t 1124129F ICLXQDA P.C.B. 1

If 5 1124127F I‘ Q D B 1 P . C . B . 1

16 ; 1123130D ITLXQDB P.C.B. 1

li , 1124128E I‘ Q D B 2 P.C.B. 1

12 I 1347155A 1‘ Q D B 3 P.C.B. 1

1C I 1352751C “lat Cable 2

2cI 13527516 1: ‘lat Cable 2


21 13527516 1 F :lat Cable 2
22 1352751L F:lat Cable ! i

23 13527519 F:lat Cable ? 1

2.l 135275111 F‘ l a t Cable ! /

25 D2131202 C :opper B o a r d %

26 J8390805 i F;letainer 1
27 J8390742 T Y-Rap

28 L9376105 atch Plunger H322-1- l - 2

29 L9376104 ILL a t c h Grommet H3.22--1- l - 3 12

30 1459441A ,C a b l e Boot 5

- -- -
I
D r a w n b y Checked by /Approved by

I 4?7/c,

K2- I
MK- TCIC- 4 1
D a t e SZP.30,1993
I
Appl led
at S/N I ~.____
-..-----A
It Reference
.
S. I. ; i
1511339c TllJ C O I L U N I T

1, 2
.-_ _---- _.._ --_- - ..__
iI’1RTNnr-- D E S C R I P T I O N Q 'TY

1 1342427E TMJ Coil I


2 1342427F TMJ Coil 1
3 13424240 TMJ Cable I
4 13424266 TMJ Head Band I
5 1327118A Coil No. Label I
6 1337723A Connector Box Label I

i ,

Drawn by Checked by Approved by

J? ,t' ,A=?, k
MR-TMC-41
Date
~---- 1 Sep.30.1993
APPI 1 ed
a t 5/N
__- I
----.~-
Reference S. I. /
1342427E TMJ COIL

Drawn b y Checked by 1 A p p r o v e d by
- -v-
,P k’ i m;k.
L2/1
MR-TMC-41

Z]

.--,.- . . . - --- _..- --.-


I T E Ed! PART NO. D E S C R I P T I O N a 'TY
/
1 12309088 TMJ Coil Assv 1
2 1342422.4 TMJ Coil Cover 2 1
3 1342423A TMJ Coil Cover 1 I 1
4 1451341.1. TMJ Coil Spacer I
I
5 1456608A TMJ Reinforcing Plate
6 13260585 Magic Tape I :
7 1233773A TMJ Leather Cover
7 1233773A TMJ Leather Cover 1 1
,
I

Drawn by j Checked by Approved by

liW i ezy’iB. / rnk.


LZ-1
MU-TMC-41

~-""1
1342427F TMJ COIL

1 Drawn by Checked by Approved by 1


MR-TMC-41
--.--
D a t e Sep.30.1993
try;;;d 1

Reference S.I. 1
!342427F- - -TMJ - COIL - - - - -
-
I TEL1 PART NO. D E S C R I P T I O N Q'TY

I 12309038 TMJ Coil Assy 1

2 1342422A TMJ Coil Cover 2 1


3 1342423A TMJ Coil Cover I I l
4 1451341A TklJ Coil Spacer 1
5 1456608A TMJ Reinforcing Plate 3
6 13260585 Magic Tape 2
7 1233773A TMJ Leather Cover 1

Drawn by 1 Checked by i Approved by

APA' i >d / d+? k.


L:3- I
MR-QKC-41
Date
Sep.30,1993
Applied
1230257A at S/N
QD KNEE COIL UNIT
Reference 5.1.
4

8
/
MR-QKC-41

)I

230257A QD KNEE COIL UNIT 1 1

ITEM PART NO. D E S C R I P T I O N Q'TY

1 1230255A QD Knee Outside Frame Unit 1


2 1230256A QD Knee Inner Frame Unit I
3 1339108B Caution Label 2 I
4 1121830H Receiver Cable 1
5 1459441A Cable Boot 2
6 13399868 QD Knee Stand 1
7 12304876 QD Knee Cover 1
8 1230487B QD Knee Cover I
9 12201628 Knee Phantom Assembly I

Drawn by Checked by Approved by


MR-QKC-4 1
1D a t e I Sep.30.1993 I
Applied
a t S / N
Reference S.I. ,
1230255A QD KNEE OUTSIDE FRAME UNI’I

1D r a w n bY c h e c k e d bY A p p r o v e d by 1
MR-QKC-41
Date Sep.30.1993
Applied
at S/N
Reference S.I.
2302 1 QD KI OUTSIDE Fli!Ab!E Z UNIT
-
ITEI PART NO D E S C R I P T I O N Q *' TY
-
11221434 QD Knee Outside Frame I
1122143B QD Knee Outside Frame 1
1452654A Guide Pin 4
1452655A Stopper A 4
1452144A Ejector 1
14521448 Ejector 1
1452145A Knob 2
1452148A Center Shaft 2
, 14521476 iing 2
II L9370115 )uracon Catch TL119 1

Drawn by j Checked by 1 Approved by

ii? /I.'
M2- 1
MR-QKC-41
Date / Sep.30.1993
Applied
at S/N
Reference S.I. /
1230256A QD KNEE INNER FRAME UNIT

(C Drawn
h e by
c k e d by n\pproved by 1
MR-QKC-41
Date j Sep.30.1993
Appl i ed
at S/N
Reference S.I.
1230256A QD KNEE INNER FRAME UNIT
ITEM PART NO. D E S C R I P T I O N Q 'TY

1 11220446 QD Knee Inner Frame I


2 11220448 QD Knee Inner Frame I
3 D2131201 Copper Board C1020R-1/4H 6
0.3Tx30x8
4 1231235A QD Knee Copper Board I I
5 1231235A QD Knee Copper Board 2 I
6 1231235A QD Knee Copper Board 3 1
7 1231235A QD Knee Copper Board 4 1
a 1343732A QD N-l P.C.B. 11
9 1343733A QD N-2 P.C.B. 1
10 1343734A QD N-3 P.C.B. 1
11 1343735A QD N-4 P.C.B. 1
12 1343736A QD N-5 P.C.B. 1
13 1343737A QD N-6 P.C.B. I
14 13437378 QD N-6 P.C.B.
15 1343738A QD N-8 P.C.B. 1
'I
16 13437388 QD N-8 P.C.B. I

Drawn by Checked by j Approved bY

c13-1
MR-QHC-51

108438XA A D tiEAD C O I L
MR-QHC-51

>I

108438XA AD IlEAD COIL I -.-------- - 1 I

ITEM P ART NO. DESCR;PTION Q 'TY

1 106923XC QD Head Coil 1

2 1459538A Boss 1 2

3 1459539A B o s s 2 4

4 1350098A Lever 1

5 13500988 Lever 1

6 1350099A Operating Plate 2

I 1350100A Bar 1

a 1350126A Plate upper 1

9 13501260 Plate upper 1

10 1350127A Plate Lower 1

11 13501278 Plate Lower 1

12 1248676A Receiver Cable AD 1

13 1350381E Base 1

14 1350381F B a s e 1

15 1339108A Caution Label 1

16 1438665A Catch Block 4

17 1244002A AD QDH-1 P.C.B. 1

18 1244003A AD QDH-2 P.C.B. 1

19 1244004/I AD QDH-3 P.C.B. 1

20 12364228 QDH-4 P.C.B. 1

21 12364238 QDH-5 P.C.B. 1

22 147062‘lA WF2009-0 PB Assy

23 131442511 Magic Tape 2

24 58341912 Connector MIF63-80 DA 1


-1.27SH

25 58341911 Connector MlF63-80 PA 1


-1.27SH

26 1362513A Head Band 1 2

27 1362514A Head Band 2 2

28 13144251 Magic Tape 4

29 1470712A Stop Cover 2

30 1470713A Sponge 2

Drawn by Checked by Approved by

52&J

Sl-1
MR-QFC-52
D a t e May. 25.1996
App I ied
a t S / N
R e f e r e n c e S . I.
1121131L FLEXIBLE 40 C O I L (MI

<‘,‘.- . -:-.,
..;,:. -I. .I:,,‘‘... : ;
6 . .- \

s2/1
MR-QFC-52
D a t e May.25.1996
App I ied
a t S / N
R e f e r e n c e S.I. 1
1124131L FLEXIBLE QD C O I L (MI 1

ITEM P A R T N O . D E S C R I P T I O N Q ‘TY

1 1129495A Flexible QD Coil Assembly 1

l-1 13459080 S p o n g e (FQDB) 1

l - 2 1345908F S p o n g e (FQDB) 1

l - 3 1345908M S p o n g e (FQDB) 3

l-4 1345908K S p o n g e (FQDB) 2

1-5 1345908L S p o n g e (FQDB) 2

2 12344530 FQDB Cover I

3 1339108A Caution Label 2 I

4 1 3 2 7 1 18A Coil No. Label 1

5 13504508 FQDB Mat 2

1 D r a w n b y j C h e c k e d b y -1 A p p r o v e d by 1
MR-QFC-5:
1 D a t e / Mav. 2 5 . 1 9 9 6

I App I i c d
a t S / N
R e f e r e n c e S. I . 1
I

I129495A FLEXIBLE QD COIL ASSEMBLY (M)

\
5

D r a w n by C 11 c c k e d h Y Approved by

s3/1
MR-QFC-52
D a t e May.25.1996
Applied
a t S / N
R e f e r e n c e S.I.
1129495A FLEXIBLE QD COIL ASSEMBLY (M)

ITEM P A R T N O . D E S C R I P T I O N Q’TY
1
1 1232477B FQD Sheet 1

2 1345882A B a s e 1 1

3 1345881A Base I

4 1345883A Base 2 1

5 13458818 Base 1

6 13458770 Cover 1 1

7 13458790 Cover 3 1

8 13458788 Cover 2 1

9 13458800 Cover 4 1

10 1367780/I Connector Holder 2 1

11 1345884A L Plate 1

12 1367781A connect Plate 2 1

13 12486760 Receiver Cable AD 1

14 1 1 2 9 4 96A FQDAAD P . C . B . 1

15 3244238A FQDlAD P . C . B . I

16 1244237A FQDBAD P.C.B. 1

17 1124128G FQDB 2 P.C.B. 1

18 1347155A FQDB 3 P.C.B. 1

19 1352751B Flat Cable 2

20 1352751F Flat Cable 2

21 1352751F Flat Cable 2

22 1352751K Flat Cable 2

23 135275113 Flat Cable 2

24 135275lT Flat Cable 2

25 D2131202 Copper Board CIOZOR-1/4H 8


0.3Tx50xlO

26 38390805 Retainer TC-102 4

27 58390742 TY-Rap T18R 2

28 L9376105 Latch Plunger H322-4-1-2 2

29 L9376104 Latch Grommet H322-4-1-2 12

30 E6600121 Coaxial Cable ULi553-1.5D-2V 1

31 Cl920122 Tube UL8hWC Black 8

Drawn br C h e c k e d by Approved by

GA\

s3-1
MR-QFC-53
D a t e May.25.1996
App I ied
a t S / N
Reference s . I
FLEXIBLE QD COIL (L)
MR-QFC-53
D a t e May.25.1996
Applied
a l S / N

I 1 ‘ 3, 1 3 I I), I’L.c*lt)LL bl” L”,L CL, ,

ITEM P A R T NO. D ESCRIPTION Q ‘TY

1 1129495B F l e x i b l e Q D Coil Assembly 1

1-l 1 3 4 5 9 0 8 1 2 S p o n g e (FQDB) 1

1-7. 1345908E S p o n g e (FQDB) 1

1-3 1 3 4 5 9 0 8 D S p o n g e (FQDB) 3

l-4 1345908K S p o n g e (FQDB) 2

1-5 1345908L S p o n g e (FQDB) 2

2 1234453c FQDB Cover 1

3 1339108/l C a u t i o n L a b e l 2 1
4 1327118A Coil No. Label I

5 135045oc F Q D B M a t 2

D r a w n b y Checked by Approved by

4A.l -9 Z@

s4-I
MR-QFC-52
D a t e May.25.1996
App I icd
a t S / N
R e f e r e n c e S. I .
11291950 FLEXIBLE QD COIL ASSEMBLY (L)
MR-QFC-53
IDale 1 May.25.1996 1
Applied
a t S / N
R e f e r e n c e S.1.
11294958 F L E X I B L E QD COIL A S S E M B L Y (L)

ITEM P A R T N O . D E S C R I P T I O N Q’TY

I 1232477C FQD Sheet 1’


2 1345882A B a s e 1 1
3 1345881A Base 1
4 1345883A Base 2 1
5 13458810 Base 1
6 13458770 Cover 1 1
7 13458798 Cover 3 1
8 13458788 Cover 2 1
9 13458808 Cover 4 1
10 1367780A Connector Holder 2 1
11 1345884A L Plate 1
12 1367781A Connect Plate 2 1
13 1248676F Receiver Cable AD 1
14 11294960 FQDAAD P.C.B. 1
15 12442380 FQDlAD P . C . B . 1
16 12442370 FQDBAD P.C.B. 1
17 112412811 FQDB 2 P . C . B . 1
18 1347155A FQDB 3 P.C.B. 1
19 1352751C Flat Cable 2
20 1352751G Flat Cable 2
21 1352751G Flat Cable 2
22 13527511. Flat Cable 2
23 1352751Q Flat Cable 2
24 135275111 Flat Cable 2
25 D2131202 Copper Board C1020R-1/4H 8
0.3Tx50xlO
26 58390805 Retainer TC-102 4
27 58390742 TY-Rap TlSR 2
28 L9376105 Latch Plunger H322-4-1-2 2
29 L9376104 Latch Grommet 1-1322-4-1-2 12

30 E6600121 Coaxial Cable UL1553-1.5D-ZV I


31 Cl920122 Tube ULIAWG B l a c k 6
32 E6600122 Coaxial Cable UL1553-3D-2V 1
33 G1920122 Tube UL8AWG B l a c k 2
34 G1960310 Tube FZ. W h i t e 1OP 1

D r a w n b y Checked by Approved by

GzGJ 32%

s5-1
MR-QFC-5.

D a t e May.25.1996

A p p I ied
a t S / N
R e f e r e n c e S . I .

1124131N FLEXIBLE QD COIL (EXL)


MR-QFC-54
D a t e May.25.1996
Appi ied
a t S / N
1 R e f erence S.I. I I
I1241. 31h CID C O I L (EXL)
- I -
ITEM P A R T N O D ESCRIPTION

I 1129495c Flexible QD Coil Assembly


I-1 1345908N Sponge (FQDB)
l-2 1345908P S p o n g e (FQDB)
l - 3 13459088 Sponge (FQDB)
I - 1345908K S p o n g e (FQDB)
I - 13459081. S p o n g e (FQDB)
12344531) FQDB Cover
1339108A Caution Label 2
1327118A Coil No. Label
135045oc FQDB M a t

D r a w n b y Checked by Approved by
MR-QFC-5l
D a t e May.25.1996
Applied
a t S / N
R e f e r e n c e S.I.
1129495c FLEXIBLE QD COIL A S S E M B L Y (EXL)

/’

D r a w n b y Checked by Approved by
MR-QFC-54
D a t e Mar.25.1996
A p p l i e d
a t S / N

,.^n,,.r^ -*....,....- ^- ^^.. .^^-..-... ttXL1


.-.... R e f e r e n c e 5. I
1 1 LY4Y3L I’LC~iOLiz CaU LUIL AbbtMtlLY ,

ITEM P A R T N O . D E S C R I P T I O N Cl ‘ T Y

I 1231234A FQD S h e e t 2 1

2 I345882A Base 1 1

3 1345881A B a s e 1

4 1345883/I B a s e 2 1

5 1 3 4 5 8 8 1 0 B a s e 1

6 1 3 4 5 8 7 7 0 C o v e r 1 1

I 1 3 4 5 8 7 9 8 C o v e r 3 1

8 1 3 4 5 8 7 8 0 C o v e r 2 1

9 1 3 4 5 8 8 0 8 C o v e r 4 1

10 1367780A Connector Holder 2 1


11 1345884A L Plate I

12 1367781A C o n n e c t P l a t e 2 1

13 12486766 Receiver Cable AD 1

14 1129496C F Q D A A D P . C . B . 1

15 1 2 4 4 2 3 8 C FQDlAD P . C . B . 1

16 1 2 4 4 2 3 7 C FQDBAD P . C . B . 1

17 1 1 2 4 1 2 8 5 F Q D B 2 P . C . B . 1

18 1347155/I F Q D B 3 P . C . B . I

19 1 3 5 2 7 5 1 D F l a t C a b l e 2

20 1 3 5 2 7 5 1 H F l a t C a b l e 2

21 1352751H F l a t C a b l e 2

22 1 3 5 2 7 5 1 M F l a t C a b l e 2

23 1 3 5 2 7 5 1 R F l a t C a b l e 2

24 1 3 5 2 7 5 1 V F l a t C a b l e 2

25 02131202 Copper Board C1020R-1141~ 8


0.3Tx50xlO

26 5 8 3 9 0 8 0 5 R e t a i n e r TC-102 4

27 5 8 3 9 0 7 4 2 T Y - R a p T18R 2

28 i-9376105 Latch Plunger t1322-4-1-2 2

29 L9376104 L a t c h G r o m m e t 11322-4-l-2 12

30 E6600121 C o a x i a l C a b l e UL1553-l.SD-2V 1

31 C l 9 2 0 1 2 2 T u b e ULIAWC B l a c k 8

32 E6600122 C o a x i a l C a b l e UL1553-3D-2V 1

33 C l 9 2 0 1 2 2 T u b e ULBAWG B l a c k 2

34 C1960310 T u b e F2 W h i t e 1 OP 1

35 tll 13A003 D i o d e UM6001B 2

D r a w n b y C h e c k e d by Approved by

4q.J .

s7-I
MR-‘.?FC-S I
D a t e May.25.1996
App I ied
a t S / N
Re’fe rcncc S. I
1121131P F L E X I B L E QD C O I L (S)
MR-QFC-51
D a t e May.25.1996
Appi ied
a t S / N
Reierence S . 1
124131P FLEXIBLE QD C O I L (S)

ITE P A R T N( D E S C R I P T I O N

11294950 Flexible QD C o i l Assembly

1345908C Sponge (FQDB)


1345908H Sponge (FQDB)

1345908G Sponge (FQDB

1345908K Sponge (FQDB


1345908L Sponge (FQDB
1234453A FCIDB Cover

1339108A Caution Label 2

1350450A FQDB Mat

D r a w n b y Checked by Approved by

4./v! =7?f3%

S8-1
MR-QFC-51
Date May.25.1996
App I icd
a t S / N
R e f e r e n c e s . I .
1129495D FLEXIBLE QD COIL ASSEMBLY (S)
MR-QFC-51
D a t e May.25.1996
App I ied

.\ ,. . -. _ ., > + ” .
1129495D FLEXIBLE QD COIL ASSEMBLY (S)

ITEM P A R T N O . D ESCRIPTION Q ‘TY

1 1232477A FQD S h e e t 1

2 134588Z.A Base 1 1

3 1345881/I Base 1

4 1345883A B a s e 2 1

5 1 3 4 5 8 8 1 0 Base 1

G 1345877B C o v e r 1 1

1 1 3 4 5 8 7 9 8 C o v e r 3 I

8 1 3 4 5 8 7 8 0 C o v e r 2 1

9 1 3 4 5 8 8 0 0 C o v e r 4 1

10 1367780A Connector Holder 2 I

11 1345884/I L Plate 1

12 1367781A C o n n e c t Plate 2 1

13 12486761~ R e c e i v e r C a b l e A D 1

14 1 1294960 FQDAAD P.C.B. 1

15 1 2 4 4 2 3 8 0 FPDlAD P . C . B . 1

16 1244237A FPDBAD P . C . B . 1

17 1124128G FQDB 2 P . C . B . 1

18 1347155A F Q D B 3 P . C . B . 1

19 1352751A F l a t C a b l e 2

20 1352751E F l a t C a b l e 2

21 1352751E F l a t C a b l e 2

22 1 3 5 2 7 5 1 J F l a t C a b l e 2

23 1352751N F l a t C a b l e 2

24 1 3 5 2 7 5 1 s F l a t C a b l e 2

25 D2131202 Copper Board ClOZOR-l/411 8


0.3Tx50xlO

2G J8390805 Retainer i-C-102 4

27 J8390742 T Y - R a p T18R 2

28 L9376105 L a t c h P l u n g e r H322-4-l-2 2

29 L9376104 L a t c h G r o m m e t H322-4-l-2 12

30 E6600121 C o a x i a l C a b l e UL1553-1.5D-2V 1

31 C l 9 2 0 1 2 2 T u b e ULBAWG Black 8

D r a w n b y C h e c k e d b y Approved by

?GR/’ J

s9-I
MR-JC-52
D a t e May.25.1996
A p p l ied
a t S / N
R e f e r e n c e S. I.
1129856A J O I N T COIL (S)

19

I I 16

I7

SlO/l
MR-JC-52
D a t e May.25.1996
Applied
at S/N
Reference S.I.
1129856A JOINT COIL (S)

ITEM P ART NO. D E S C R I P T I O N Q 'TY

1 1228836/I Small Coil 2 1

2 12288360 Small Coil 2 I

3 1341472C Tunning Box 5 1

4 1450568A Cover 5 1

5 1341676/I Copper Board 4 2

6 1341676B Copper Board 4 2

1 1341676C Copper Board 4 2

a 1431690A Gilt Plate

9 124494oc AD TUN Unit 1

10 1229398A Small Coil Cover 2 1

11 12293988 Small Coil Cover 2 1

12 13391088 Caulion Label 2 1

13 1248676J Receiver Cable AD 1

14 1440367A CZZlGM-0 PB Assy 4

15 58341943 Connector HIF7-40DA-1.27DS 2


(31)

16 58341941 Connector HlF7-40PA-1.27DS 2


(31)

17 1433558A PC Hook 2

18 1450556C Cable Cleat 1

19 1450728A Cover 6 3

20 1450729A Cover I 1

21 1436F71C Cover 3 1

22 58390742 TY-Rap T18R 1

23 1327118A Coil No. Label 1


MR-JC-5:
Date May.25.1996
APPI ied
at S/N

1129541A Reference S.I.


JOINT COIL (M)
. .._
MR-JC-52

I I
1129541A J O I N T COIL (MI

ITEM P A R T N O . D ESCRIPTION Q ‘TY

1 1235094A Middle Coil 4 I

2 1235094B Middle Coil 4 1

3 13330091) Copper Board 3 2

4 1333009E Copper Board 3 2

5 1333009F Copper Board 3 2

6 13414720 Tune Box 5 1

7 1450568A Cover 5 1

8 1440367A CZZICH-0 PB Assy 4

9 1458291A Middle Coil Cover 2 1

10 1441297/I Cover 210 3

11 1440894A Cover 210-A 1

12 38341943 Connector HIF7-40DA-1.27DS 2


(31)

13 58341941 Connector HlF7-40DA-1.27DS 2


(31)

14 1433558A PC Hook 2

15 13391080 Caution Label 2 1

16 143169OA G i l t P l a t e A 1

17 1327118A C o i t No.Label 1

18 1450556C Cable Cleat 1

19 1327266L Sponge 2

20 1327266M Sponge 1

21 1327266N Sponge I

22 1236315D Middle Coil Cover-3 I

23 1236315C Middle Coil Cover-3 1

24 38390742 TY-Rap T18R 1

25 1248676C Receiver Cable AD 1

26 1244940/I A D T U N Unit. I

D r a w n b y Checked by Approved by

Sll-1
MR-JC-51
Date May.25.1996
Appl icd
at S/N
Reference S. I
11213348 JOINT COIL (L)
MR-JC-51
D a t e May.25.1996
Applied
at S/N
RrrPrPncr
__._. - ..__ s-...
I
1121334B JOINT COIL (L)

ITEM P A R T NO. D E S C R I P T I O N Q'TY

1 1214008A Large Coil 1

2 1341472C Tune Box 5 1

3 1317793c Box Spacer 1

4 1450568A Cover 5 1

5 1330794A Copper Board 2 2

6 143169OA Gilt Plate A 1

7 13391080 Caution Label 2 1

8 1450556C Cable Cleat 1

9 1220293A Large Coil Cover 1

10 132726611 Sp0llge 1

11 58390742 TY-Rap T18R 1

12 Cl960305 Tube F2 White 1

13 1248676D Receiver Cable 2 1


14 12449400 AD TUN Unit 1

Drriwn b y Checked by Approved by

c.ti

SlZ-I
MR-TMC-51
D a t e May.25.1996
App 1 ied
a t S / N
Refcrcnce S.1.
151 1339F TMJ COIL UNIT
MR-TMC-5 I
1 D a t e 1 May.25.1996 1

I App I ied
a l S / N
R e f e r e n c e g. I .
I I

1511339F TMJ COIL UNIT I

ITEM P A R T N O . D ESCRIPTION Cl ‘TY

1 13424276 TMJ Co i I 1

2 13424271-1 I-MJ Coil 1


3 13642558 TMJ Cable 2 1

4 1342426A TMJ Head Band 1

5 1327118A Coil No. Label 1

D r a w n b y Checked by Approved by

c.hi

s13-1
MR-TMC-51
1D a t e / Mav.25. 1996
App I ied
a t S / N
t R e f erence S.I. I
13424276 TMJ COIL

\
4

:
MR-l-MC-51
D a t e May.25.1996
A p p l i e d
a t S / N
R e f e r e n c e S.I.
13424276 TMJ COIL
-
ITE P A R T N O D E S C R I P T I O N
--I- (1 .
-

1364253A T M J C o i l Assy

1342422B TMJ Coil Cover 2

1342423B TMJ Coil Cover 1


145134iA TMJ Coil Spacer

1456608/I TMJ Reinforcing Plate

13260585 Magic Tape


1233173A TMJ L e a t h e r Cover

D r a w n by Checked by Approved by

5.d
MR-TMC-51
D a t e May.25.1996
App I ied
a t S / N
Reference S. I.
134242711 TMJ COIL

5
MR-TMC-51
D a t e May.25.1996
Appl ied
a t S / N
R e f e r e n c e S . I
1342427H T M J COIL

ITE P A R T N O D E S C R I P T I O N

1364253A TMJ Coi I Assy

13424220 TMJ Coil Cover 2

13424238 TMJ Coi I Cover 1

1451343A TMJ Coi I Spacer


1456608A TMJ Reinforcing Plate

13260583 Magic Tape

1233773A TMJ Leather Cover

D r a w n b y Checked by Approved by

G/d

SIS-1
MR-QKC-51
D a t e
MaY.25.I996
APP 1 ied
a t S / N
Refe rence s . I .

/
6

8
MR-QKC-5 1
D a t e May.25.1996
A p p l i e d
a t S / N
R e f e r e n c e S.I. (
12319408 QD Kb I E E COIL UNIT I

-I
ry
ITE * r
P A R T N O D ESCRIPTION z

1230255A QD Knee Outside Frame Unit

1230256C QD Knee Inner Frame Unit

13391088 Caution Label 2

1248676E Receiver Cable


13399860 QD K n e e Stand

1230487A QD Knee Cover

12304870 CID Knee Cover

12476820 MRI P h a n t o m No.1

D r a w n by Checked by Approved by

-5.i~

S16-1
MR-QKC-5 1
D a t e M a y . 25,1996
App I ied
a t S / N
R e f e r e n c e s. I
1230255A QD KNEE OUTSIDE FRAME UNIT
MA-QKC-51

1 Z3lJZSSA UU KNEE OUTS.IDE FRAME UNIT , I

ITEM P A R T N O . D E S C R I P T I O N Q 'TY

1 1122143A QD Knee Outside Frame 1

2 1122143B QD Knee Outside Frame 1

3 1452654A Guide Pin 4

4 1452655A Stopper A 4

5 1452144A Ejector 1

6 1452144B Ejeclor 1

I 1452145A Knob 2

8 1452148A Center Shaft 2

9 1452147/I Ring 2

10 L9370115 Duracon Catch TLIl9 1

Drawn by Checked by Approved by

S.ti

s17-1
MR-QKC-5
Date’ May. 25. 1996
A p p l i e d
I a t S / N
1 R e f erence S.I. I
1230256C QD K N E E I N N E R F R A M E UNIT

\ 5
MR-QKC-51
D a t e May.25.1996
A p p l i e d
a t S / N
Reference
_ _ _ . . S . 1 . 1
1230256C QD KNEE INNER FRAME UNIT f

I l-EM P A R T N O . D ESCRlPTlON Q ‘TY

1 1122044A QD Knee Inner Frame 1

2 11220448 QD Knee Inner Frame 1

3 D2131201 Copper Board CIOZOR-1/4H 6


0.3Tx30x8
4 1475220A QD K n e e C o p p e r Board 2 2

5 1475219A CID K n e e C o p p e r Board 2 4

G 1475221A QD Knee Copper Board 3 4

1 1363031A ADNQD-1 P.C.B. 1

a 13437338 QD N - 2 P . C . B . I

9 13437340 QD N-3 P.C.B. 1

10 1363032A ADNCID-4 P . C . B . 1

11 1343736C QD N-5 P.C.B. 1

12 1363033A ADNQD-6 P.C.B. I

13 13630338 ADNQD-6 P.C.B. 1

14 1363034A ADNQD-8 P.C.B. I

15 13630348 ADNQD-8 P . C . B . 1

I6 1475220B QD Knee Copper Board 2 2

D r a w n b y Checked by Approved by

SiS-1
MR Imaging System

M o d e l AIRIS

Open Head Coil Up-grade

TECHNICAL GUIDE

HITACHI MEDICAL CORPORATION


Tokyo, Japan
L2E-BM5208- 1
r 7
CONFIDENTLAZITY : This material contains proprietary information of Hitachi
Medical Corporation and it shall not be disclosed to the third party or reproduced
without the prior written permission of Hitachi Medical Corporation.
Copyright Hitachi Medical Corporation. 1998. All rights reserved.
General

This manual relates to up-grading Closed Type Head coil (Model MR-QHC-51) to Open type
Head coil (Model MR-QHC-52).
Work for this up-grading referring together to the Instruction Manual of main unit.

Outline of Up-grading Procedures

Lower coil Upper coil

I Removing FRP covers


I

Replacing PCBs

Attaching coaxial cables

Positioning connectors

I + I
Adjusting impedance and phase Adjusting isolation

1NG
I OK

Completed

L2E-BM5208
Then remove three PCBs mounted on the lower coil. Use the soldering iron to remove
the copper plates soldered to each PCB. The parietal side connector shown in the
figure above is included in the accessories supplied. If it is hard to remove it, connector
pins may be cut off.
Then attach PCBs supplied (three in total).
At this point, however, the screws fixing PCBs should be temporarily tightened.

1.3 Connecting cables


Attach the coaxial cables to PCB.
Refer to the figure below for soldering locations.
P C B a t t h e of coil
Rbd-Shirld

2. Preparing the upper coil


As shown in the figure, remove the FRP cover of upper coil and then the side plate (hatched).
Then loose the screws fixing PCBs.

L2E-BM5208
-3-
@ When the connector at the parietal side does not fit
Remove the copper plate soldered to the right and left PCBs for the moment.
Then file off the PCB mounting hole and FRP cover windows (6 locations).

Foot side

Parietal side

To

CBS are shifted to Lhe

4. Attach all the FRP covers and check if the connectors can be successfully connected.

5. Checking connection of the lock


Check if there is no positional shift of the lock in the up/down direction.
If any shift in the fore/back direction is found, follow the steps below for adjustment.
(JiJ Remove the lock currently attached.
@ With the upper or lower hole as a reference, file off the hole of the side where the lock
is shifted.

To be filed off

At this time, take care of the hole to be widened laterally, and not to be widened
vertically.

0 Use brass screws @I3 flat-head screw) and hexagonal nuts included in the accessories
supplied to fix the lock at the side which holes have been filed off.
@ Restore the FRP covers (U) and (IJ to their positions.

-5- L2E-BM5208
@ Fold aluminum foil (or spacer for adjusting table height: aluminum plate) to a size
of about 1Ocm square, and slowly bring it from the left or right side of coil.

Aluminum foil or the like

At this time, measure amplitude of signal displayed on the oscilloscope and find a
position where the amplitude becomes smallest.

I I

Amplitude of oscilloscope

Then, remove the copper tape if it is stuck at the side opposite to the side where the
amplitude became smallest.
If the copper tape has not been stuck at the opposite side, stick the copper tape at
the side of aluminum foil.
When sticking the copper tape, do so to the location shown below so as not to touch
the copper plate of coil.

Location where the copper


tape is to be stuck.

After sticking the copper tape, try to bring aluminum foil from the left or right side
of coil. When amplitude of signal displayed on the oscilloscope becomes larger
even if bringing it from either direction, adjustment has been completed.

-7- L2E-BM5208
MR Imaging System

Joint Motion Devices


Installation and Adjustment Manual

Knee/Cervical Device
Shoulder Device
Wrist Device
Ankle Device
Lumbar Device
Hip Device

Model MRP-5000
Model MRP-7000
Model AIRIS

HITACHI MEDICAL SYSTEMS AMERICA


Twinsburg, OH

0 Copyright Hitachi Medical Systems America 1997, All rights reserved,

L5E-BMA102- 1
1. Conte
No, Description Page

1 .o Table of Contents, Revision History. . , , . . . . . , . ,* 2


2.0 Component List and Quantity Checklist. . a. . 3
3.0 Product Introduction a. . I a, , I , , . . . , . . . . . . I 5
4.0 Unpacking and Inventory Procedures , a. a. . 6
5.0 Component Assembly (MRP-5000, MRP-7000) 11
6.0 Component Assembly (AIRIS) , I aI I . . . . . . . . 12

Revision History
Revision Effective
Date Comments Initial
Number Pages
LSE-BMA 102- 1 l/15/97 All Preliminary Draft Issue JJR

Page 2 LSBMA 102- 1


Component h-f and Quanfirly CheckhY Page 3

.l Knee/Cervical Device
2.1.1 Knee/Cervical Device , . . . , . , , , . . , . , , , . . . , . . , , . , , . Qty. 1 ___
2.1.2 Pads

2”x42”.,..,.....,....,.......,...,....,...,..,..Qty.2 _ _
2”x24”.............,...,...........,..........,.Qty.l -
2.1.4 Feet Kit
l/4” x l/2” Machine screw and rubber bumper. . . , . . . . Qty. 8 -

2.2 Ankle Device


* . . * Qty. 1 ___

2”x34”.,.,,....,,.........,.......,......~,.
2”~24”.....,.........,,.,..,..,.....,......
2”x20”,,.................,,..,....,........
2.2.4 RFCoil support , . , , , , . . . . , . . , , , , I , , . . . . . . . , . .
2,2.5 Locking Pins . , , , , , . , I I . , . . . . , , , , , , , , . . , . , , . a.
2.2.6 Feet Kit
T/4” x l/2” Machine screw and rubber bumper. . . 0 . . I Qty. 8 -

2.3 Wrist Device


2.3.1 Wrist Device +I , Qty. 1 -
2.3.2 P a d s
Finger Pad . . a. Qty. 1 -
2.3.3 S t r a p s
2” x 20” , . . . . . * . * , Qty. 2 ~
2” x 14” . . * . I I Qty. 1 -
2.3.4 RF Coil support ,,* Qty. 1 -
2.35 Locking Pins . . . Qty. 2 _ _

LSBMA 102- 1
Componen f Lisf and Quan fify Checklisf Page 4

2.4.1 Lumbar Device . , . , . . . , . ,,. ,. ,. Qty. 1 -


2.4.2 Pads (Egg-crate Sponge)
6”x 8” . . . . . . . . . , . . , . . a Qty. 1 -
%“x 8” #. . . , . , , . , *. . . , * * * . . . Qty. 1 _ _
2,4.3 S t r a p s
Qty. 3 -
Qty, 1 -
Qty. 2 __

2.5 Hip Device


25.1 Hip Device I . a . . * . . , , Qty. 1 __
25.2 P a d s
Heel Pad . , . . . . I . . Qty. 2 __
2.5.3 Straps
2” x 34” . . , . . * . , Qty.4 -
2” x 24” . . . . . s . . . . Qty.4 -
2” x 20” . I I * * . . . . , Qty. 2 __
25.4 Locking Pins. . * * . . Qty. 6 -

2.6 Shoulder Device


2.6.1 Base Plate . , , . . . . , . . , , . . . . . a . . . I I I * Qty. 1 -
Abduction / Adduction Unit . . * . . I I . * . Qty. 1 ___
Internal / External Unit . , . , . . . . . * . . Qty. 1 ___
2.6.2 Pads
. . . Qty. 1 ___

. * I Qty. 1 -
. . . . . a Qty. 1 -
,. * . . * * , * Qty. 2 ~
. * . . . * * * . Qty.4 __

2.7 Optional Accessories


2.7.1 Storage Cabinet.. , , . , , . , , . , , , , . , . . q, . . . . . . , . , . . , .Qty. 1 __

L5E-BMA 102- 1
Producf In froducfion Page 5

The Joint Motion Devices are a family of six mechanical supports intended to
assist in the MR imaging of seven anatomical areasThe six Joint Motion Devices,
the seven anatomical areas of interest,and the RF Coils recommended for use
with the Joint Motion Devices are listed in the chart below,

Joint Motion Device Anatomical Joint RF Coil


HS/MR (Quad) Extremity
Knee
Knee / Cervical Joint/Cervical
Cervical Spine Joint/Cervical
Ankle Ankle ~__ ..-..- Small Extremity -.--__
-___
_~--..-..--. Wrist Wrist Small Extremity
Lumbar Lumbar- Spine Large Extremity
HS/MR (Quad) Flexible --
(Small, Medium, Large)
Hip Hip (Bilateral) Rigid Quad Body
Solenoid Flex Body
(Small, Large)
Shoulder Shoulder Large Extremity

The Joint Motion Devices are intended to provide repeatable imaging of the
above joints through dynamic and kinematic motion. The Joint Motion Devices
enable incremental, stepped motion of the joint under study, and constrain joint
motion to a single user-designated plane of rotation, These user-designated
planes of rotation emulate the natural motion of the desired articular anatomy.

L5E-BMA 102-l
Un-packing and In veniory Procedures Page 6

cki tor
The following steps describe the un-packing and inventory procedures for the
Joint Motion Devices for the MRP-5000, MRP-7000 and AIRIS. Select a suitable work
area for the following steps.

Unpacking

4.1.1 Remove each device from the crating and packaging material,
4.1.2 Take care when using a knife so as not to nick or cut the patient pads
and straps.

Inventory

4.2.1 The Joint Motion Devices are shipped from HMSA generally fully
assembled. Using the figures below, and the Component List and
Quantity Checklist in Section 2, inventory and inspect each device for
completeness and/or damage. Contact HMSA Service for
replacement or repair, as necessary.
Alignment Marker
User Control Panel and Angle Indicator
Knee/Cervical not shown
Device
Hinge
Knee lmagkg \ /

Base Rail P a d ’ /
Thigh Strap - 42”
RF Coil Strap -’42”
Pivot Plate and Strap- 24”

L5E-BMA102-1
Un-packing and ln ventory Procedures Page 7

Alignment Marker
and Angle Indicator
Device

Cervical Spine
lmaghg

Base Rail Pad

Pivot Plate

Fi Base Plate Pivot Supports Foot Grip, Heel Pad,


Ankle Device & Straps - 24”, 20”

Leg Holder & straps - 34” L RF coil Support Foot Plate -!l
(noi’ sbo wn)

L5E-BMA102- 1
L/n-packing and Inventory Procedures Page 8

Finger Pad
Wrist Device /-land Rest and Strap - 14”
\

- Hand Ring

---Base Plate
Locking Pit -l
RF Coil Support
(nofshown)

Arm Rest and Straps- .20”

L/__ Rotation Rockers

Figure 4-4 Storage Locking Pin , Body Rotation Plate


Lumbar Device

Body Support Plate

- Base Plate

7 Body Straps- 66”

Carry Handle I- - Carry Handle

Angle Indicator

Egg-crate
?E?-%Z,;‘” I Leg/Thigh Strap -
66 ”
RF Coil Strap
(notshown)

L5E-BMA102- 1
Unpacking and Inventory Procedures Page 9

Figure 4-5 - Storage Locking Pins


Hip Device

Foot Grip, Heel I


-& Straps - 24”) 21

Base Plate

Leg Strap - 34”

- - 10” Abduction 1-ocking

Figure 4-6A
Shoulder Device

Base Plate
’ & Patient Pad

Abduction/Adduction Rotation
‘Locking Pin Sockets
(Total Qty 7)

Rotator Wheel

Internal/External Unit
Fixing Locking Pins
Accessory Arm

Abduction/Adduction Unit
Fixing Locking Pins

L5E-BMA 102- 1
Unpacking and ln venicory Procedures Page 10

Abduction/Adduction Unit
Shoulder Device
Alignment Marker
&se Hate wifh
Angle indicator
Aba’uc fion/
Ad&.&ion Un?

Arm Length-
Base Plate -/ Locking Pin
Arm Rail
(adjustable)

L- Arm Elevation
Adjusting Screw

Abduction/Adduction Unit
Fixing Locking Pins

L5E-BMA102-1
m(MRP--5000,
P a g MRP-7000) e 11

Figure 4-6C Internal/External Unit ‘-7 Arm Rest7


She lulder Device Arm Strap
\ r (Upper Arm) - 24” I/- $!~~?~~- 24”
Base Plate-, 1 HarEd Grip

5.0 Component Assembly (MRP-5000, MRP-7000)


Several of the Joint Motion Devices require the following specific assembly
procedures in order for proper use with the MRP-5000 and MRP-7000.

No additional specific assembly procedures are required for use with the AIRIS.
The Joint Motion Devices are shipped from HMSA configured for immediate use
with the AIRIS,

5.1 Knee/Cervical Device Modification

5.1,l Flip the Knee/Cervical Device upside down, so that the device is
resting on the patient pad and pivot plate,Take care to prevent dirt
or damage to the patient pad.
51.2 Each of the eight (8) feet of the Knee-Cervical Device is made up of
a rubber bumper, a l/2” spacer, and a 1 l/2” plastic l/4” machine
screw (total height = 3/4” when installed). A separate Feet Kit is made
up of a rubber bumper and l/2” plastic l/4” machine screw (total
height = l/4” when installed).
5.1.3 Remove each bumper-spacer-machine screw assembly, and replace
it with the shorter bumper-machine screw assembly.
51.4 Store the removed bumper-spacer-machine screw assemblies as
spares in a safe location.

L5E-BMA 102-l
Component Assembly (AIRS) Page 12

nkl

5.2.1 Flip the Ankle Device onto its side, Take care to prevent dirt or
damage to the device.
5.2.2 Each of the five (5) feet of the Knee-Cervical Device is made up of a
rubber bumper, a 5/8” spacer, and a 1 l/2” plastic l/4” machine
screw (total height = 7/8” when installed). A separate Feet Kit is made
up of a rubber bumper and l/2” plastic l/4” machine screw (total
height = l/4” when installed).
5.2.3 Remove each bumper-spacer-machine screw assembly, and replace
it with the shorter bumper-machine screw assembly.
5.2.4 Store the removed bumper-spacer-machine screw assemblies as
spares in a safe location.

6.0 Component Assembly (AIRIS)


No additional specific assembly procedures are required for use with the AIRIS
The Joint Motion Devices are shipped from HMSA configured for immediate use
with the AIRIS.

L5E-BMA102-1
MR Imaging System

QD Shoulder Coil
Installation and Adjustment Manual


Model MRP-7000, AIRIS
MR-QSC-41
MR-QSC-51

HITACHI MEDICAL SYSTEMS AMERICA


Twinsburg, OH
CONFIDENTIALITY: This material contains proprietary information of
Hitachi Medical Systems America (HMSA) and it shall not be disclosed to
other third parties or reproduced without prior written permission of HMSA.
 Copyright Hitachi Medical Systems America 2000. All rights reserved.

L2E-BMA108-1
1.0 Introduction
This document describes the installation and adjustment procedure for the QD Shoulder
coils intended for use with the MRP-7000/AIRIS. The test results must be documented
for quality assurance purposes and evaluated to assure that each production coil meets the
coil performance specifications.

Page 2 of 16 L2E-BMA108-1
2.0 Reference Documents
Q1E-BM5003 MRP-7000 Instruction Manual
L2E-BM0434 MRP-7000 Installation and Adjustment Manual
Q1E-BM5057 AIRIS Instruction Manual
L2E-BM5034 AIRIS Installation and Adjustment Manual
MR-HH-METER Handheld Impedance Meter Technical Guide

3.0 Materials
This procedure requires the following tools and equipment.
• QD Shoulder Coil, including right and left support pads
• Impedance Meter, hand-held
• RG58 test cable, BNC male/BNC female, half wavelength
• Hitachi RF coil loading phantom no. 3
• Non-magnetic screwdrivers
• Non-magnetic screwdriver tuning tool
• QD Flex Coil pads (wings) or other appropriate padding

Page 3 of 16 L2E-BMA108-1
4.0 Installation and Adjustment
4.1 Coil and Phantom Setup
4.1.1 Turn on the impedance meter and allow it to warm up five minutes before use.
4.1.2 Place the coil on the right shoulder pad, place it on the patient table and plug it
in to the pre-amp.
4.1.3 Use a screwdriver to remove the plugs over the tuning capacitors in the coil.

Page 4 of 16 L2E-BMA108-1
4.1.4 Place the No. 3 phantom into the coil. Make sure the base is flush with the
shoulder coil pad. Use another pad to force the phantom against the rear of
the coil.

Phantom should be flush with the coil pad


and tight against the inside surface

Flex Coil “wings”

4.1.5 Set the orientation switch on the coil for the right shoulder.

Set Switch for


Right Shoulder

Page 5 of 16 L2E-BMA108-1
4.1.6 Use the laser locator to position the coil at isocenter. Note the alignment cross
hairs on the top of the coil. Rotate the coil/phantom to align the crosshair with
the laser.

Positioning Crosshair

4.1.7 Verify that the laser bisects the phantom in the coronal plane. Adjust the coil
vertical position with pads or bedsheets as necessary.

Laser Locator bisects


phantom as shown

Page 6 of 16 L2E-BMA108-1
4.1.8 Press [SET] to move the coil to isocenter. After the coil is at isocenter, press
the [IN] button move the coil in until the display reads –65.
4.1.9 Register a New Patient. Enter a name and ID and then proceed.
4.1.10 Access the Coil Data Entry program. Add “QD/Shoulder” as the coil name for
physical coil #11. Access the tool as follows:
• Press: SELECT MODE
• Type: 48 ENTER
• Type: 31 ENTER
The following information is displayed:

** COIL DATA ENTRY **

# COIL NAME LINE ATT TUNE # COIL NAME LINE ATT TUNE
1 HS/MR HEAD COIL 2 4 3 17 3 0 3
2 HS/MR BODY COIL 3 1 3 18 3 0 3
3 JOINT 3 0 3 19 3 0 3
4 LARGE EXTREMITY 3 0 3 20 3 0 3
5 SMALL EXTREMITY 3 0 3 21 3 0 3
6 LARGE FLE. SPINE 3 0 3 22 3 0 3
7 SMALL FLE. SPINE 3 0 3 23 3 0 3
8 3 0 3 24 3 0 3
9 3 0 3 25 3 0 3
10 3 0 3 26 3 0 3
11 3 0 3 27 3 0 3
12 3 0 3 28 3 0 3
13 3 0 3 29 3 0 3
14 3 0 3 30 3 0 3
15 3 0 3 31 3 0 3
16 3 0 3 32 3 0 3

COIL NUMBER = 1
COIL NAME = HS/MR HEAD COIL *RF LINE*
RF LINE (1-3) = 2 1..RF LINE A
RF ATT (0-7) = 4 2..RF LINE A AND B
TUNE STEP (0-8) = 3 3..RF LINE B

NEXT SET? (Y/N) =

Page 7 of 16 L2E-BMA108-1
4.1.11 Enter the Coil Data elements as shown below. To modify a value, do
the following:
• Type the coil number and press ENTER.
• Type the coil name and press ENTER.
• Type the RF parameters as shown below and press ENTER.

Coil Parameter Data Value


Physical Coil No. 11
Coil Name QD Shoulder
RF Line 2
RF Att 0
Tune Step 3
RF Gain Adj. Yes

4.1.12 After the parameters are entered, exit the Coil Data Entry program:
• When prompted: NEXT SET (Y/N), type N and press ENTER.
• When prompted: SET (Y/N), type Y and press ENTER.
The Coil Data Entry program terminates.

Page 8 of 16 L2E-BMA108-1
4.2 Coil Tuning
4.2.1 Set the frequency of the impedance meter to the system frequency (f0).
4.2.2 Calibrate the half wavelength cable for the particular system frequency by
fine-tuning the capacitor of the cable assembly. The cable should be calibrated
as follows:

Open load impedance: 1.4 - 1.7 kΩ


Phase: 0°°±1°°

! NOTICE

Do not plug the impedance meter into the pre-amp before setting the Tune
voltage. Failure to comply may cause meter damage.

4.2.3 Use the Signal Generation Check program to set the Tune value to 18h.
• Press: SELECT MODE
• Type: 47 ENTER to display the check program menu
• Type: 5 ENTER to execute the PCS program
• Type: 5 ENTER to execute the Signal Generation Check program
4.2.4 Enter the following values for the Signal Generation Check program:

SGC Parameter Data Value


FREQ (Resonant Frequency)
TUNE 18
GAIN 14
COIL NO. 11

4.2.5 Start the SGC program:


• When prompted: START OK (Y/N), type Y and press ENTER.
• Wait 5-10 seconds and then press the SPACE BAR to stop the program.

Page 9 of 16 L2E-BMA108-1
4.2.6 Connect the impedance meter and half wavelength cable into the channel
connected to the solenoid coil element.

For the AIRIS:

Line B - Solenoid

For the MRP-7000:

Line B - Solenoid

Page 10 of 16 L2E-BMA108-1
4.2.7 Using the Solenoid tuning capacitor on the coil, adjust the impedance. Tune
the capacitor to the impedance range shown below. Record the value on the
Installation Report (refer to page 16).

Line Magnitude (Z): Phase:


(Solenoid/Saddle) 250 ±100 ohms 0°°±10°°

Saddle Adj.

Solenoid Adj.

4.2.8 Move the half wavelength cable connection to the other coil element as shown
below:

For the AIRIS:

Line A - Saddle

Page 11 of 16 L2E-BMA108-1
For the MRP-7000:

Line A – Saddle

4.2.9 Using the Saddle tuning capacitor on the coil, adjust the impedance. Tune the
capacitor to the same impedance range shown in step 4.2.7. Record the value
on the Installation Report.
4.2.10 Remove the half wavelength cable connection.
4.2.11 Record the individual coil impedance values in the section for the right
shoulder in the Coil Installation Report (refer to page 16). Refer to the
example below:

Type: QD Shoulder
Model (circle one): AIRIS MR-QSC-51 MRP-7000 MR-QSC-41
S/N: 305
Line B Line A Combined
Calibration Data
Solenoid Saddle
Impedance 245 270 N/A
Right: Tune
Receive Gain
RF Gain
SNR Enter the impedance data

Page 12 of 16 L2E-BMA108-1
4.3 Coil Element Verification
Both elements of the coil must be checked for proper operation. To do this, each
element must be tested for SNR, Tune, and gain values. Results must be recorded in
the Coil Installation Report.
4.3.1 Open the Coil Data Entry Tool as you did in step 4.1.10. Select
“QD/Shoulder”.
4.3.2 Isolate the Solenoid element by changing the RF Line value. Apply the
change and close the program.

RF Line 3 (selects B-Solenoid)

4.3.3 Perform a Scanogram in the coronal plane.


4.3.4 Display the image.
Note the angled orientation and the brighter areas at the edges of the phantom.

Select the Sagittal plane


as shown.

4.3.5 Choose the SAG plane as shown above. Execute the SNR protocol as
described below:

Parameter Value Parameter Value


Slice Plane SAG C. Position 0
Sequence SE FOV 260
2D/3D 2D TR 500
Sequence No. 3 TE 40
Rephase Off Slice Thickness 5mm
Phase Direction A/P Multislice 1
Rejection Off Echo 1
Rectangle Off NSA 2
Anti-wrap Off Projection 256
Presatuation Off

Page 13 of 16 L2E-BMA108-1
4.3.6 When the scan is complete, press DISPLAY. Use the S/N Ratio Check
program under the Analysis menu.
4.3.7 Record the SNR value in the B-Solenoid column for the right side in the
report (refer to Section 6.0).
4.3.8 Inspect the SNR image for asymmetrical shading or image artifacts. Contact
HMSA Technical Support if any problems are noted.
4.3.9 Examine the TUNE & GAIN display under IMAGE ANALYSIS and record
the Tune, Receive gain and RF gain values as well.
4.3.10 Using the Coil Data Entry Tool, isolate Line A- Saddle by changing the RF
Line value:

RF Line 1 (selects A-Saddle)

4.3.11 Execute the SNR protocol once again. When complete, measure and record
the SNR value in the A-Saddle column (for right side) of the report.

! NOTICE

The SNR values for the individual coil elements will differ significantly.
Typically, the ratio is 2 to 1 or greater.

4.3.12 Examine the TUNE & GAIN display under IMAGE ANALYSIS and record
the Tune, Receive gain and RF gain values.

! NOTICE

The Tune values of the individual coil elements must match closely (within
one step). If not, readjust the tuning capacitors and re-test until the values
are correctly matched.

4.3.13 Enter the Coil Data Entry Tool once again and return the RF Line to its
original value:

RF Line 2 (selects both lines)

4.3.14 Verify the Attenuation Value is set to zero. Exit the Coil Data Entry Tool.

Page 14 of 16 L2E-BMA108-1
4.3.15 Execute the SNR protocol once again. When complete, measure and record
the SNR value in the Combined column of the report. Examine the TUNE &
GAIN display under IMAGE ANALYSIS and record the Tune, Receive gain
and RF gain values.
4.3.16 Verify that the combined (Solenoid/Saddle) SNR is within the specified range
listed below:

Coil Model SNR*


MR-QSC-51 140±30
MR-QSC-41 140±30

* You will note a difference in average SNR values from right to left.
4.3.17 Flip the orientation of the coil/phantom to the left shoulder. The new position
should mirror that of the right shoulder.
4.3.18 Change the switch position on the coil to the Left Shoulder.
4.3.19 Execute the protocol and record the Solenoid, Saddle, and Combined values
for the left side on the Installation Report.

5.0 End of Installation


5.1.1 Remove the impedance meter, half wavelength cable, and all tools from the
magnet room.
5.1.2 Replace the pre-amp cover.
5.1.3 Verify that all blocks of the Coil Installation Report are completed, including
coil serial number.
5.1.4 Copy the report for your records and return a copy with your Field
Service Report.

Page 15 of 16 L2E-BMA108-1
6.0 QD Shoulder Coil Installation Report

Type: QD Shoulder
Model (circle one): AIRIS MR-QSC-51 MRP-7000 MR-QSC-41
Serial Number:
Line B Line A
Calibration Data Combined
Solenoid Saddle
Impedance N/A
Right: Tune
Receive Gain
RF Gain
SNR
Left: Tune
Receive Gain
RF Gain
SNR

Print Name:
Signature:
Date:

Page 16 of 16 L2E-BMA108-1
MR Imaging System

Model AIRIS

SHOULDER COIL

MR-SC-51
.

TECHNICAL GUIDE
--- Installation & Adjustment ---

HITACHI MEDICAL CORPORATION


Tokpo , Japan
LZE-BM5177-2

CONFIDENTIALITY : This material contains proprietary information of


Hitachi Medical Corporation and it shall not be disclosed to the third party or
reproduced witbout the prior written permission of Hitachi Medical Corporation.
Copyright @ Hitachi Medical Corporation. 1997,1998. AU rights reserved.
CONTENTS

Page

1. Adjustment of the MR-SC-51 (AIRIS) . . . . . , . . . 1

2. Check of S/N ratio MR-SC . . . . . . . . , . . . . . 2

3. Recording sheet . . . . . . . . . . . . . . . . . . 4

( 1 1 1 L2E-BM5177
1. Adjustment of the MR-SC-51 (AIRIS)

The coil adjustment is effected in the same manner as


described in the Technical Guide - Installation & Adjustment.
Only the different jobs from those in the Technical Guide
will be described herein.

(1) Check the tuned value and adjust the impedance with the MR-
SC-51 (AIRIS) set on the Phantom No. 3.

(2) The trimmer capacitors Cr and Cm for adjusting this Coil are
located as shown below. Adjust the capacitors similarly to
the other coils.

Cr

Cm

(3) The coil adjustment values should be as follows

Coil Tuned voltage Range of tuned value Adjustment


Coil Name No Object Phase
to be set to be checked impedance
Phantom
MR-SC- 51 13 1V OC-4E 400+100 0°+300
No.3

-l- L2E-BM51’77
2. Check of S/N ratio MR-SC

Similarly to the check of the head coil, determine the S/N

ratio.

(I) Place the adjusting phantom (No. 3) and cover the phantom
with the receiver coil.

Phantom No. 3

J /i/

Example phantom setting

(2) Set the following scan parameters and make three scans
without any positioning.

SCAN PLANE : TRS


SEQUENCE : SE
2D/3D : 2D
SEQ. No. : 3
REPRASE : OFF
PHASE DIR. : A-P
REJECTION : OFF
RECTANG. : OFF
ANTI WRAP : OFF
PRESAT. : OFF
C.POSITON : 0
FOV : 260
TR : 500
TE : 40

-2- LZE-BM5177
THICKNESS : 5
MULTI SLICE : 1
ECHO : 1
NSA : 2
PROJECTION : 256

(3) Measure the S/N ratio and confirm that S/N is within the
range described bellow.

S/N 180 f 40

-3- L2E-BM6177
3. Recording sheet:

Adjustment & check data recording sheet

(I) MR-SC-51 (AIRIS)

- Impedance : ohms (200 +/- 100 ohms)

- Phase : deg. (O" +/- 30°)

- Tuned value : (flC to 4E)

(2) S/N ratio check

- S/N ratio : (180 +/- 40)

- Imaging room temperature: OC

-4- S L2E-BM6177
MR Imaging System

Flexible Body Coil


MR - FBC - 31,32

TECHNIC’AL GUIDE
S-BInstallation & Adjustment ---

HMSA
HITACHI MEDICAL CORPORATlON
Tokyo,Japan

L2J3-BM0253-5

CONF’IDEWl’IALITY : ~hie material containe proprietary information of


Hitachi Medical Corporation and it &all not be disched to the third party or
reproduced without the prior nritkn perrnieah of Hitachi Medical Corporation.
CopgrigWOHitachiMediealCorporatioa1990,1991.AllrigMe~~rved.
Contents
Page

1. General ............................ ........... ... 1


2. Adjustment . . . . . . ..*.............................. 2
1. General

This technical Guide describes how to install and adjust


the flexible body coil which can be set in accordance with
the patient's posture for body imaging.

-l-
2. Adjustment

The coil is to be adjusted in the same manner as


described in the Technical Guide - Installation and
Adjustment. So, only the different procedures from
those explained in the Technical Guide will be
described herein.

(1) Use the phantoms below to check on the TUNE and


impedance values.

No. 1 connector MR-FBC-31 Connectors other than No. 1,


I MR-FIX-32
phantom No. 3 Nickel solution phantom No. 4

Trimmer capacitor \ No. 1 No. 2 No.3 No. 4


for adjustment A
0
(2) The coil adjustment should be performed by using
connector No. 1 and the other connectors are used only
to check on the impedance and phase.

j3 LSE-BiIOi63
-2-
(3) Remove the cover of the coil and the polyvinyl chloride
cover (as shown in "A" in the previous page), and set
the phantom on to the coil at the connector position
No. I (where the coil length is shortest).

Lock : Push down


Lift

to the
.ine of

(4) Adjustment of the impedance

Set the coil No. and TUNE voltage at "18". Connect


the impedance meter to the preamplifier. Adjust the
trimmer capacitor so that the phase is on O", and check
.the impedance if within the specified range as described
below.
Refer to 4.7.2, "receiving coil" in the
Installation Manual.
..

4 L?lE-BMo253
-3-
Coil name Coil No. /Object TUNE value to Target
be checked impedance

Nickel solution
Flexible Unused phantom No. 3 OC to 72 1000 1 200 R
coil for Nickel
the body No. solution 200 2 100 l-i
phantom No. 4 OC to 72

(5) Adjustment of the RF gain value

Use the coil No. used in (4).

NOTE: Record the RF gain value when adjusting.

- (6) Measure the signal-to-noise ratio value. For the body


coil (large) as well, measure the signal-to-noise ratio
value in accordance with the standard manual. Make
sure that the signal-to-noise ratio value is 0.8 to
1.6 times that of the body coil.
(Only connector No. 1 of the MR-FBC-31 takes 1.5Q3.3
times)

(7) Check on the impedance and adjust the RF gain values


for the connectors from No. 2 to 4. Set the coil No.
$0 perform the 1 SCANOGRAMI. Then, check on the TUNE
value. If within the specified range in table (4),
connect the impedance meter to the preamplifier. Turn
the dial to O" phase to change the frequency and check
on the impedance if within the range specified in table
(4). Use the same coil No. as the connector No. used
in (4) for the coil used in (7).

NOTE: If there is any gap found between the coil and


the phantom, use a sponge pad or the like to
fill up the space for uniform. lapping of the
coil as shown below.

6 L&-BMW3
-4-
If any gap is found, use a sponge
,or the like to wrap the coil.

Adjust the height by using


a matcress or the like.

(8) After completion of all the adjustments, make a manual


entry of the RF gain value. The value to be entered
should be the mean value measured from the connector Nos.
1 to 4, and the coil No. should be the one used 'in (4).

(9) Set the cover of the coil, and put the accessory.':
coil No. label ona position as shown below. The coil
No. should be the one used in (4).

Attach label of the


coil No. here.

Caution label

- 5 -
(10) Perform the impedance adjustment with the human body.
!Che adjustment should be performed for one of the
connectors and the value should be the same as in (4).
Impedance adjustment for the other connector is
unnecessary.

111) Make sure imaging can be successfully Performed.

7 UlE-BMOSSS
-6-
MR Imaging System

Model MRP - 5000

HS /MR Body Coil

TECHNICAL GUIDE
--- Installation & Adjustment ---

HMSA
HITACHI MEDICAL CORPORATlON
Tokyo, Japan

L2E-BM0445-1

CONFIDENTIALITY : TIGS material contains proprietary information of


Hit&i Medical Corporation and it shall not be disclosed to the third party or
reproduced without the prior written pmniefsion of Hitachi Medical Corporation.
Copyright@ Hitachi Medical Corporatiou 1991. AlI rigbta reserved.
CONTENTS

Page
1. General . . . . . . . . . . . . . . . . . . . . . . . ..a.................... 1

2. Installing procedure for HS/MR Body coil . . . . . . . . . ..I. 2

3. Operation check . . . . . . . . . . ..*......................... 6

1 L2EBMO44.6
1. General

This manual describes the procedures for installation and


adjustment of the NS/MR Body coil which perinits produce
images of high quality than with the standard Body coil.

-l- 2 LZE-BM0445
2. Installing procedure for HS/MR Body coil

2.1 Replacing the under-body coil

Remove the under-body coil of the existing rigid body


coil from the auxiliary tabletop, and install the under-body
coil of the supplied HS/MR body coil.

2.2 Detach,ing cover and setting phantom

Detach the outer cover (made of FRP) of the HS/MR body


coil. Place a Ni-solution phantom (LARGE) on the
auxiliaty tabletop, project a light beam from the light
localizer to the center of magnetic field and move the
phantom to the magnetic field center. Then set the
HS/MR body coil, and slide the coil, while visually
checking, so that the coil center coincides with the
light beam projected from the light localizer.

Cover

Q .

(6 pieces)
l$i/MR Body coil

Ni-solution phantom I Mattress (15 mm thick)


/ (LARGE) Auxiliary \
Patient tableto

-2- 3 L2E-BMW45
2.3 Adjusting the impedance

Use the check mode PSC C/K "SIGNAL GENERATIOBl CHECK" to set
TUNE voltage to "18~".
According to the instructions given in the Technical Guide -
Installation & Adjustment - for the standard version of MR
Imaging system Model MHP-5000, adjust the impedance in the
same manner for the conventional or existing body coil.
However, the connectors of A- and B-lines should be connected
as shown below:

Line-A adjustment Line-B adjustment


'Impedance
meter Preamplifier Preamplifier

C 56CN Line A 56CN


A A
Line A c .57CN c 57CN

Line B C 58CN B C58CN


B
c 59CN Line B c 59cN

Impedance
meter

Adjust the following trimmer capacitors for the A- and


B-lines for the phase to be within 0 +/- 2', and make
sure that the impedance is 300 +/- 100 ohms.

-3- 1 L2E-BMO445
n R s,
II I‘
II -ll-l~

Saddle Bide
(Line - B)

Solenoid side
(Line - A)

Disconnect the T-type connector and Impedance meter probe,


and connect the A- and B-lines to 56CN and 58CN, respectively.

-4- 5 LZJGBMO445
2.4
For determination of attenuation value(ATT value) of HS/MR
Body coil, measure SN ratio of solenoid and saddle side of
MS/MR Body coil respectively, and obtain the ATT value
from the SN ratio.

Procedures:
@ Set the phantom (in the same way as in tuning adjustment).

02 Change RF LINE by COIL DATA ENTRY program.


Solenoid side : RF LINE =3
Saddle side : RF LINE =l

@ Scan an image for SN ratio measurement and measure SN ratio.


Refer to Item 2.5 of this Manual for sequence of SN
measurement.
Scan is to be performed in single slice.
Be sure'to scan only after the RF CAIN adjustment is done
for each scan. (Refer to the Installation Manual for RF
CAIN adjustment.)

04 Calculate ATT value as follows:


ATT value = 20.log (Solenoid side SN/Saddle side SN)

@ Round off the ATT value obtained in @ and set


this value by COIL DATA ENTRY program. At this
time reset the RF LINE to "2".

-5- 6 L2EBMO446
2.5 Measuring the signal-to-noise ratio

01 Effect a scanogram measurement to check that the TUNE


value is within a range of "OC to 48". At this time,
use the program “PATIENT DATA SET” to set the coil No.
to a one entered at step 2.4 above.

02 Adjust RF GAIN referring to the Installation Manual.

03 Measure SN ratio. Measuring sequence is shown below.

S/N measuring sequence


SE
TR: 500 RFCT.: OFF
TE: 40
FOV: 320
Slice thickness: 10 mm
SQ#: 3
AVE.: 2
Proj.: 256

Make sure that SN ratio is within 135+25.

04 Re-install the FRP cover in place, and attach the


supplied Coil No. nameplate carrying the No. having
been entered at step 2.4 in an arbitrary position
near the signal cable lead-out hole.

3. Operation check

After completion of the installation of the HS/MR body


coil, make scans of a human body to check that the
system runs normally.

-6- 7 L.2EBMO445
I List of supplied parts:

1. HS/MR body coil 1


2. Coil No. nameplate 1

- 7 - 8 LZE-BMW45
MR Imaging System

Model MRP

HS / MR Knee Coil
MR - QKC - 31
MR - QKC - 32
MR - QKC - 41

TECHNICAL GUIDE
--- Installation & Adjustment ---

Tokyo , Japan
L2E-BM0449-2

CONFIDENTIALITY : This material contains proprietary information of


Hitachi Medical Corporation and it shall not be disclosed to the third party or
reproduced without the prior written permission of Hitachi Medical Corporation.
c Hitachi Medical Corporation. 1991. All rights reserved.
Copyright ⃝
Contents

Page

1. General ....................... 1

2. Adjustment ..................... 1

3. Confirmation .................... 4

(1) 1 L2E-BM0449
1. General

This manual describes the Installation and Adjustment for


the HS/MR knee coil with which a high quality image can be
produced as compared with the standard joint coil. Refer to
the Installation and Adjustment for the main unit together
with this Installation Manual.

2. Adjustment

(1) Removing the cover

Remove the leather cover for the upper coil by


releasing the lock for the upper and lower coil
connection.

Disassemble the upper and


lower coil by the ejector.
aaaaaaaa
aaaaaaaa
aaaaaaaa
aaaaaaaa
aaaaaaaa
aaaaaaaa
aaaaaaaa
aaaaaaaa
aaaaaaaa
aaaaaaaa
aaaaaaaa
aaaaaaaa
aaaaaaaa
aaaaaaaa
Release the lock.

-1- 2 L2E-BM0449
(2) Phantom setting

Set the phantom onto the coil and connect the upper and
lower coil as illustrated in the diagram. Take care
not to be tilted in the X, Y, and Z axis.

Phantom No. 1
(Accessory phantom
for the coil) aaaaaaaaaa
aaaaa

Z
aaaaaaaa
aaaa
1
aaaaaaaaaa
aaaaaaaa
aaaa

Y X

Caution Name Plate

(3) Tuning

Adjust the phase and impedance as in the tuning of


HS/MR head coil in the Installation and Adjustment for
the main unit. The adjustment for each trimmer
capacitor is indicated in the following. The phase and
impedance to be adjusted are indicated.

Tuning voltage at Impedance to be Phase to be


Coil name
adjustment adjusted adjusted

Saddle coil
18 500±200Ω 0°±2°
(line A)

Solenoid soil
18 500±200Ω 0°±2°
(line B)

-2- 3 L2E-BM0449
aaaaaaaaaaaaaa
aaaaaaaaaaaaaa
aaaaaaaaaaaaaa
aaaaaaaaaaaaaa
aaaaaaaaaaaaaa
aaaaaaaaaaaaaa
aaaaaaaaaaaaaa
aaaaaaaaaaaaaa
aaaaaaaaaaaaaa
aaaaaaaaaaaaaa
aaaaaaaaaaaaaa
aaaaaaaaaaaaaa
aaaaaaaaaaaaaa
aaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaa
aaaaaaaaaaaaaa
aaaaaaaaaaaaaa
line B (solenoid coil)
adjusting place
line A (saddle coil)
adjusting place

(4) Coil data registration

Register the coil data of the coil in accordance with


the Installation and Adjustment for the main unit.

RF LINE = 2 (LINE A and B)


ATT = It is determined from the following
procedures
TUNE STEP = 3 (※ for registration of MRP-7000 only)

For determination of attenuation value (ATT value) of


HS/MR knee coil, measure SN ratio of solenoid and
saddle side of HS/MR knee coil respectively, and obtain
the ATT value from the SN ratio.

Procedures:
① Set the phantom (in the same way as in tuning
adjustment).

② Change RF LINE by COIL DATA ENTRY program.


Solenoid side : RF LINE = 3
Saddle side : RF LINE = 1

-3- 4 L2E-BM0449
③ Scan an image for SN ratio measurement and measure
the SN ratio. Refer to item (6) of this Manual for
sequence of SN ratio measurement.
Scan is to be performed in single slice.
Be sure to scan only after the RF GAIN adjustment is
done for each scan. (Refer to the Installation
Manual for RF GAIN adjustment.)

④ Calculate ATT value as follows:


ATT value = 20.log (Solenoid side SN/Saddle side SN)

⑤ Round off the ATT value obtain in ④ and set this


value by COIL DATA ENTRY program.

Use the coil No. which is not yet used.


(※ use the coil No. “8“ for MRP-7000 only)

(5) RF GAIN Adjustment

Adjust the RF GAIN in accordance with the Installation


and Adjustment for the main unit. The Coil No. should
be the same as used in (4).

(6) S/N ratio measurement

Check the tuning voltage if within the range OC 4E by


taking a SCANOGRAM. The measured S/N ratio sequence is
indicated in the following:

SE
TR : 500
TE : 40
Slice thickness :5
AVE. (NSA) :2
Projection : 256
SQ No. :3
FOV : 180

-4- 5 L2E-BM0449
Check the S/N ratio if it is within the range in table
below.

Type S/N Standard

MR - QKC - 31 125±25

MR - QKC - 32 135±25

MR - QKC - 41 215±40

(7) Coil No. plate lining

Return back the leather cover that was removed in (1).


The coil No. plate which was sent should be lined onto
the location as illustrated in the diagram. The No. to
be lined should be the No. as selected in (4).

Handling the caution plate

Line the coil No. as used in (4).

3. Confirmation

Confirm if a normal image can be obtained in the human body.