TABLE OF CONTENTS

SAFETY NOTES…………………………………………….2-5

FEATURES…………………………………………………..6-7
Oxygen Concentrator HANDLING……………………………………………………8
7F-5A OPERATING & INSTALLATION…………………………9-14
User’s manual MAINTENANCE……………………………………………15-20

DISASSEMBLY……………………………………………..21-23

OTHER ATTENTION ITEMS………………………………24-25

DO NOT OPERATE THIS UNIT WITHOUT FIRST


READING AND UNDERSTANDING THIS MANUAL

1
SAFETY NOTES SAFETY NOTES
WARNING: For avoiding power outage or oxygen concentrator blocked. Keep the openings free from lint, hair and the like.
operating failure, exigent oxygen user and serious patient must Ⅲ. USING
equip with reserve oxygen supply (for example: oxygen cylinder,
oxygen bag). This device is to be used as an oxygen supplement  The use of oxygen therapy requires that special care be taken to
and is not considered life supporting or life sustaining. reduce the risk of fire. Users MUST NOT SMOKE while using this
Ⅰ. BEFORE INSTALLATION device. Keep all matches, lighted cigarettes or other sources of
 The concentrator should always be kept in the upright to prevent ignition out of the room in which this product is located. NO
cabinet damage while being transported. SMOKING signs should be prominently displayed. Textiles and
 Electrical source, power voltage is instable that beyond normal other materials that normally would not burn are easily ignited and
power voltage range. Please add manostat. burn with great intensity in oxygen enriched air. Failure to observe
 Use eligible, safe power set and junction box. this warning can result in severe fire, property damage and cause
 Curbstone people do not open oxygen concentrator’s cabinet. physical injury or DEATH.
Ⅱ. PUT POSITION  For optimum performance, do not open or turn off the concentrator
 You may select a room in your house where using your oxygen frequently, reset after 3~5 minutes. Shorter periods of operation
concentrator would be most convenient. Your concentrator can be may reduce maximum product life.
easily rolled from room to room on its casters.  A spontaneous and violent ignition may occur if oil, grease or
 Be certain to place the unit so all sides are at least 10cm (4inches) greasy substances come in contact with oxygen under pressure.

away from walls, draperies, furniture, or other obstruction. Do not These substances MUST be kept away from the oxygen
concentrator, tubing and connections, and all other oxygen
place the unit in a confined area.
equipment. DO NOT use any lubricants unless recommended by
 Oxygen concentrator must avoid heat source, fire source, wetness,
manufacturer.
exorbitant or over-low extraventricular condition. Ⅳ. MAINTENANCE
 Sundries and vessel do not be placed on top of the concentrator. The oxygen concentrator was specifically designed to minimize
 NEVER block the air openings of the unit or place it on a soft routine preventive maintenance at intervals of once per year. Only
surface, such as a bed or couch, where the air opening may be professionals of the healthcare field or persons fully conversant with
2 3
SAFETY NOTES SAFETY NOTES
this process such as authorized or factory trained personnel should amount of oxygen, the supplier and/or physician should be
perform preventive maintenance or performance adjustments on the contacted immediately. No adjustments should be made to the
oxygen concentrator. flowrate unless prescribed by a physician.
Ⅴ. RADIO FREQUENCY INTERFERENCE  Close supervision is necessary when this product is used near
 Most electronic equipment is influenced by Radio Frequency
children or physically-challenged individuals.
Interference (RFI). CAUTION should be exercised with regard to
 Use this product for only intended use as described in this manual.
the use of portable communications equipment in the area around
 DO NOT use parts, accessories or adapters other than those
such equipment.
authorized by manufacturer.
Ⅵ. TO REDUCE THE RISK OF BURNS, ELECTROCUTION,
 DO NOT connect the concentrator in parallel or series with other
FIRE OR INJURY TO PERSONS.
oxygen concentrators or oxygen therapy devices.
 Avoid using while bathing. If continuous usage is required by the
 Use of some administration accessories or certain humidifiers not
physician’s prescription:
specified for use with oxygen concentrator may impair the
The concentrator must be located in another room at least
performance.
2.5m from the bath.
 In certain circumstances oxygen therapy can be hazardous.
 DO NOT come in contact with the concentrator while wet.
Manufacturer recommends that you seek medical advice before
 DO NOT place or store product where it can drop into water or
using this product.
other liquid.
 Avoid creation of any spark near medical oxygen equipment. This
 DO NOT reach for product that has fallen into water. UNPLUG
includes sparks from static electricity created by any type of
IMMEDIATELY.
friction.
 A product should NEVER be left unattended when plugged in.
 This device is to be used only in accordance with the prescription
of a physician and this User’s Manual. If at any time the patient or
attendant conclude that the patient is receiving an insufficient

4 5
FEATURES FEATURES
Ⅰ. SUMMARY 3. Maximum Outlet Pressure: 45kPa±4.5kPa (6.52psi±0.65psi)
7F-5A oxygen concentrator is intended for individual use in the 4. Pressure Relief Mechanism Operational at:
home. It is an electronically operated device that separates oxygen 250kPa±25kPa (36.25psi±3.63psi)
from room air. It provides high concentration of oxygen directly to 5. Altitude: Up to 1828 meters (6000 ft) above sea level without
you through a nasal cannula. Clinical studies have documented that degradation of concentration levels. From 1828 meters (6000 ft)
oxygen concentrators are therapeutically equivalent to other types to 4000meters (13129 ft) below 90% efficiency.
of oxygen delivery systems. 6. Sound Level: ≤55dB (A)
This user’s manual will tell you about your concentrator and will ※ 7. Maximum Nebulization Rate (Apply in the oxygen
serve as a reference as you use your concentrator. concentrators having the function of nebulizing): ≥0.15mL/min
Ⅱ. CHRARCTERS 8. Power Supply:
 Complete plastic out shell, safe and reliable. □AC120V±10% □AC220V±10% □AC230V±10%
 Timing turns off function using convenience. □50Hz □60Hz 1.8 amps(AC220V~240V) ; 2.9 amps(AC120V)
 Exceed press safety valve helping safe ensure. 9. Input Power: 400VA 10. Weight: 28kg (55lbs)
 Install power cut warning function. 11. Dimension:
 Compressor with hot protection which ensure safe of the H71.2cm×W28.5cm×D54.5cm
compressor and concentrator. 12. Minimum Operating Time: 30minutes
 Power cord and nasal oxygen cannula etc accessories can be put  Electric classification: II class ,type B applied part.
in the store box of concentrator.  Work system: work continuously.
 ※ Having nebulizing rationale (apply in the oxygen  Keep it far from flammable and explosive gas area.
concentrators having the function of nebulizing ): May connect Ⅳ. NORMAL OPERATION CONDITIONS
with nebulizer to have nebulizing cure.  Temperature range: 5℃~40℃
Ⅲ. SPECIFICATIONS  Relative humidity:≤80%
1. Flow Range: 1~5L/min 2. Oxygen Concentration: ≥90%  Atmosphere pressure: 86kPa~106kPa (12.47psi~15.37psi)

6 7
HANDLING OPERATING & INSTALLATION
Ⅰ. UNPACKING Ⅰ. FEATURE VIEW
 Check for any obvious damage to the carton or its contents. If
damage is evident, please notify the carrier or local dealer.
Display panel Power switch
 Remove all loose packing from the carton.
Timing breaker
 Carefully remove all the components from the carton. button for
increasing Flowmeter
NOTE: Unless the oxygen concentrator is to be used immediately,
Timing knob
retain containers and packing materials for storage until use of the button for
concentrator is required. decreasing Panel

Ⅱ. INSPECTION
Flowmeter Truckle
Examine exterior of the oxygen concentrator for nicks, dents,
scratches or other damages. Inspect all components. Humidifier
bottle
Ⅲ. STORAGE
 Store the repackaged oxygen concentrator in a dry area.
 DO NOT place other objects on top of the repackaged concentrator.
Suck connect
Ⅳ.OPEN UP THE PANEL
Pry out the shaft with the special tool as follows, then lift up the Safety valve Oxygen outlet
display panel for about 45 degree, put the shaft back, fix the display
cover
panel, and the machine can be used.(Figure 1)
bottle

Figure 2: view of the humidifier


Figure 1

8 9
OPERATING & INSTALLATION OPERATING & INSTALLATION
Ⅱ. PREPARE WORK Ⅲ. OXYGEN ABSORBING OPERATION
1. Unscrew the bottle from the humidifier clockwise. Fill the bottle 1. FLOWRATE (Figure 4 )
with pure water (or distilled water) to the level indicated by the  Turn the flowrate knob to the setting Flowmeter
magnitude sign
manufacturer. prescribed by your physician or
Flowrate knob
DO NOT OVERFILL humidifier bottle. (Figure 3) therapist.
2. Screw securely the bottle back NOTE: To properly read the Flowmeter

counterclockwise. flowmeter, locate the prescribed


MAX Black ball
3. Connect the power supply. flowrate line on the flowmeter.
WARNING: MIN Next, turn the flow knob until
Figure 4
1) If the concentrator has a damaged the ball rises to the line. Now,
cord or plug, if it is not working Figure 3 center the ball on the L/min line prescribed.
properly, if it has been dropped or CAUTION: If the flowrate on the flowmeter ever falls below
damaged, or dropped into liquid, call Qualified Service Personnel 0.5L/min, check tubing or accessories for blocked or kinked
for examination and repair. tubing or a defective humidifier bottle.
2) Keep the cord away from HEATED or HOT surfaces.  Meanwhile, humidifier bottle will have air bubble around the

3) Do not move or relocate concentrator by pulling on the cord. sieve core. Then, oxygen comes from the oxygen outlet.
4) Do not use extension cords with this unit. 2. Connect nasal oxygen cannula to oxygen outlet, the other end
4. Press down the power switch while the display shows elapsed time match the patient.
and “— — ”. After a few seconds, the concentrator will work. Note: The time of absorbing oxygen and the flowrate range
NOTE: Concentrator may be used during the initial start warm-up according to the doctor’s advices.
time (approximately 30 min) while waiting for the O2 purity to ※ Ⅳ .NEBULIZING OPERATION (Apply in the oxygen

reach maximum. concentrators having the function of nebulizing.)


 Fill proper medicinal liquid in the nebulizing cup (Please follow
the doctor’s advices or don’t exceed the maximum reticle of the
10 11
OPERATING & INSTALLATION OPERATING & INSTALLATION
nebulizing cup).
 Upward pull out the nebulizing cover on the nebulizing interface.
(Figure 5) Figure 5
Mouthpiece T-piece Ripple
 Connect the air hose to neblizing cup and tube
nebulizing interface, then turn on the power Cap
Air hose
of the oxygen concentrator, now can begin
Baffle
nebulizing cure immediately.
 When drug nebulizaion have finished, Nebulizing Nebulizing
interface cover Cup
turn the nebulizing cover right to
nebulizing interface for tightening. If you won’t breathe oxygen,
Figure 6
please turn off the oxygen concentration.
 Pull out the air hose, pull the mouthpiece, downward pull the
Ⅴ. SETTING TIMING UP
nebulizing cup’s cap, empty remained medicinal liquid in the
User can press the timing button in 00-99 minutes.
nebulizing cup, then wash the air hose, mouthpiece, nebulizing
 When it start working “--”shows on display, it says timing
cup’s cap, nebulizing baffle, nebulizing cup, ripple tube, T-piece,
etc , with pure water or immerge them in warm water for about function not open. It will work continually until power supply
15 minutes. In order to wash them healthy, you may add some cuts off.
vinegar in water. (NOTE: DO NOT cook the accessories above  Press the “+”button once, operation time increases one minute,
for wash or wash them with boiled water, in case they’ll distort holding the button more than 1.5 seconds it will increase
when heating).
continually.
 After finishing washing you must dry all the accessories before
 Press the “-”button once, operation time decrease one minute,
storing. (Nebulizing installment is shown in figure 6).
holding the button more than 1.5 seconds it will decrease
NOTE: the time for nebulizing should follow the doctor’s
continually.
advices.
 It will auto turn off when finishing timing meanwhile “00” show
on display. Reset timing function.

12 13
OPERATING & INSTALLATION MAINTENANCE
Ⅵ. TURNNING OFF NOTE: Power should be disconnected before beginning preventive
Take off nasal oxygen cannula from oxygen outlet first, close power maintenance on the concentrator. The concentrator is specifically
switch, and then cut off power source. designed to minimize routine preventive maintenance at intervals of
Ⅶ. SYMBOLS once per year. In places with high dust or soot levels, maintenance
Symbol Description Symbol Description may need to be performed more often. The following must be
performed at a minimum of one year in service to assure years of
~ Alternating current Consult the manual additional reliability.
Ⅰ. CLEANING CABINET
WARNING: CUT OFF the power supply first. To avoid electrical
Class Ⅱ Equipment Type B applied part shock, DO NOT removes cabinet.
Clean the cabinet with a mild household cleaner and non-abrasive
OFF(power. disconnection ON(power. connection cloth or sponge at least one time each month.
from the mains) to the mains) Ⅱ. CLEANING OR REPLACING FILTERS
Please clean and replace the filters in time, it’s very important to

breaker protect compressor and delay machine’s life.


CAUTION: DO NOT operate the concentrator without the filters
installed.
 Disassembly
(1) The first mesh in filter cleans
once half a month. (Figure 7)
Ⅱ-stage filter door

Ⅰ-stage filter mesh


B
Figure 7
14 15
MAINTENANCE MAINTENANCE
Ⅲ.TROUBLESHOOTING GUIDE
Symptom Probable cause Solution Remark
1) Power cord NOT 1) Insert the power
Figure 8 plugged In. cord
Press the switch, 2) Use normal
the concentrator is 2) No power outlet. power.
alarm of 3) Reset switch open. 3) Press it.
continuous buzzer
(2)Ⅱ-stage filter mesh in filtrater cleans once month. (Figure 8) but no display and 4)Down-lead plug-in of 4) Check and Repairs by
working. the transformer loose fasten it.
Open filter access door; qualified
5) The transformer is personnel.
5) Replace it.
Unscrew the filter corn; broken.

Ⅱ-stage filter mesh cleaning time depends on practically using



1) -stage filter blocked.
1) Clean
replace it.
or

and condition. If the filter core becomes black, please replace or Ⅱ


2) -stage filter blocked.
2) Clean
replace it.
or

3) Do not use
clean it at once, no matter how long it used. extension cords.
Concentrator is 3) Insufficient power at Move to another
 Cleaning working, the outlet.
electrical outlet or
turning sound
1) Clean the filters with a vacuum cleaner or wash in warm circuit.
of assemble
4) The air intake of
valve is weak 4) Smooth them.
soapy water and rinse thoroughly. compressor kinked.
and the
5) Repair or
2) DRY the filters thoroughly before reinstallation. flowmeter can 5) Pressure relief valve of
replace the
turn down but compressor leaked gas.
valve.
Damage: Do not operate the concentrator without the filters can’t turn up at Repairs by
a certain value. 6) Tube on the sieve 6) Tighten it. qualified
bed leaked gas.
installed, or while filters are wet. These actions could personnel.
7) Insufficient air
7) Replace the
permanently damage the concentrator. exchanging of assemble
valve.
valve.
8) Failure of electronic 8) Repair or
control circuit. replace it.
Concentrator is 1) Repair or
1) Failure of pilot valve.
working with a replace it.
sound of 2) Failure of assemble
2) Replace it.
phut-phut, and valve. Repairs by
the ball in the qualified
flowmeter personnel.
3) Failure of electronic 3) Repair or
located on “0”
control circuit board. replace it.
position or float
unstable.
16 17
MAINTENANCE MAINTENANCE
Symptom Probable cause Solution Remark
Symptom Probable cause Solution Remark
1) Reinstall and 1) Make sure the
Concentrator is 1) Humidifier leaked. 1) It’s not ventilate
tighten it. machine is at least
working, but completely around
2) Safety valve of 2) Safety valve 10 cm away from
oxygen from machine, so operating
humidifier opens. restoration. the walls, other
outlet is little or temperature is too high.
3) The nasal cannula jam or heater.
no. 3) Replace it.
kinked or blocked. 2) Temperature of the
Concentrator is 2) Add cold water
water added in
working with 1) The nasal cannula in bottle, but don’t
1) Smooth it. humidifier bottle is
normal sound, folded or blocked. add overhot water.
overhigh.
but flowmeter
can turn down 3) Water added
but can’t turn 2) The filter corn in 2) Clean or replace 3) The water added in should between
down at a certain humidifier bottle
the corn. humidifier bottle is too the maximum and
blocked.
value. The nasal much. minimum of the
Concentrator is 1) The air intake of cannula has water curve.
1) Tighten it.
working with flowmeter is loose. more mirage or 4) Stop breathing
normal sound, but 2) Flowmeter has fault. 2) Replace it. blobs. oxygen
flowmeter’s 4) The machine stops
immediately.
display is “0” and 3) The pressure regulator 3) Replace it. suddenly during
ruptured or blocked. Restart the
can’t turn up. breathing oxygen.
machine to eject
Concentrator is 1) The connect line of 1) Tighten the
the hydrosphere.
working with display is loose. connecting line.
normal sound, 2) The connect line of 2) Replace the 5) The machine stops
5) Smooth the
but the display display’ is broken. connect line. suddenly when folding
nasal cannula.
has no display or the nasal cannula.
3) Display circuit board
appears 3) Replace it. 6)a. Take out the
has fault. 6) Fan inside the
disorderly code. eyewinker that
machine can’t run or Repairs by
Press the 1)Operate the lock the fan.
1) Turning on/off the running rate turns slow qualified
concentrator after

switch,
machine too frequently make the operating personnel.
displaying is turning off 3 5
makes it breakdown. temperature overhigh. 6) b. Replace it.
normal, but minutes.
compressing 2) Replace the
pump/fan can’t 2) The main controlling main controlling
work. circuit board has fault
circuit board.

18 19
MAINTENANCE DISASSEMBLY(CUT OFF the power supply first)
Symptom Probable cause Solution Remark
1) Check nebulizer is
Ⅰ.CUT OFF THE POWER SUPPLY FIRST
distortional or not if it Put the power switch on ‘0’,pull out the plug from the socket.
was disinfected. 1) Replace it.
Ⅱ.DISASSEMBLE THE DISPLAY PANEL
Unscrew 4 screws in the back of the display
2) Ensure that it
※ 2) Check air hose
whether it is bended or
isn’t bended and
airflow is flowed
panel and take off the panel.(Figure 9 )
The nebulizer not. Apply in 4 screws
smoothly when it
doesn’t run. the Figure 9
is used.
oxygen Ⅲ.DISASSEMBLE THE OUTSIDE CABINET
3) Check whether 3) Add appropriate
concentrat
medication was used up amount
ors having 1. Poke the filter door slowly from inner to outer near the
or not. medication to cup.
the
4)Fix it according
4)Check the nebulizer is function orifice,unscrew the filter core.(Figure 10)
the method of
fixed properly or not. of
nebulizer’s fixing. 2.Unscrew 6 screws in the back cover and take off the back cover.
nebulizing
Add appropriate .
amount medication
※ The medication was
to cup. Connect air
hose to main body
Water drops over-filled or the and turn on. Close 6 screws
have formed on nebulizer was washed. and open
the air hose. repeatedly by finger
in order to get out
the water drops. Figure 10

3.Unscrew 9 fixing screws in the left and right cabinet(5 in the left
If you have all other problems, TURN OFF the concentrator first, side,2 in the right side,2 in the back side)(Figure 11)
use your reserve oxygen supply (if provided), and contact dealer
or manufacturer immediately.

Figure 11

5 screws 2 screws 2 screws


in the left in the back in the right

20 21
DISASSEMBLY(CUT OFF the power supply first) DISASSEMBLY(CUT OFF the power supply first)
.DISASSEMBLE THE SIDE BOARD OF THE INSIDE
CABINET Sieve bed
1 .Unscrew 10 fixing screws in the inside cabinet, take off the inside Pressure
regulator
cabinet.(Figure 12)
Connection-base Transformer

Figure 14

10 screws

Figure 12

2.Unscrew the screws in both sides (9 screws each side) and take
out it. (Figure 13, Figure 14)

Electronic control
circuit board
Air valve
Fan
Inlet sound
buffer box
Exhaust sound
muffler
Compressor
Figure 13

22 23
OTHER ATTENTION ITEMS OTHER ATTENTION ITEMS
Ⅲ. ELECTRICAL RATIONALE
Ⅰ. LIMITATION OF TRANSPORT AND STOCK
Transformer Low power supply
 Temperature range: -40℃~70℃(-104℉~158℉)
 Relative humidity: 10%~100%
 Atmosphere pressure range: 80kPa~106kPa (12.47psi~15.37psi) Electronic control
~ circuit
 Shipping weight: 31 kg
The concentrators must be stored in aeration without erode gas,

Ⅰ Ⅱ
avoided shaking and inversion in transport. Assembled Assembled
Fan Compressor
valve valve
or

Ⅱ. GAS PASS OPERATION SKETCH MAP Ⅳ. ACCESSORIES LIST

Air Filter Compressor Heat exchanger 1. Nasal oxygen cannula: 1 set 2. Filter core: 1 set
3. Fuse tube: each 2 pieces 4. User manual: 1 piece
※ Nebulizer (apply in the
Air hose Nebulizing ※5. Nebulizer
oxygen concentrators
outlet
having nebulizing function) (Apply in the concentrator having nebulizing function): 1 set
All specifications and product configurations are subject to change
Oxygen azote Molecular Assembled
separation sieve valve without notification.

Oxygen Azote Commute valve Muffler

Unilateralism
Product Pressure Bacilli
valve
tank regulator filter

Nasal oxygen Humidifier Oxygen


cannula outlet Flowmeter

24 25