Professional Documents
Culture Documents
Iso 7991
Iso 7991
Hkkjrh; ekud
dk¡p — eè; js[kh; rkih; çlkj xq.kkad Kkr djuk
( nwljk iqujh{k.k )
Indian Standard
GLASS — DETERMINATION OF COEFFICIENT OF
MEAN LINEAR THERMAL EXPANSION
( Second Revision )
ICS 81.040.01
© BIS 2012
BUREAU OF INDIAN STANDARDS
MANAK BHAVAN, 9 BAHADUR SHAH ZAFAR MARG
NEW DELHI 110002
NATIONAL FOREWORD
This Indian Standard (Second Revision) which is identical with ISO 7991 : 1987 ‘Glass — Determination
of coefficient of mean linear thermal expansion’ issued by the International Organization for
Standardization (ISO) was adopted by the Bureau of Indian Standards on the recommendation of the
Glass, Glassware and Laboratoryware Sectional Committee and approval of the Chemical Division
Council.
This standard was first published in 1970 and then subsequently revised in 1999 to bring it in line with
the latest version of ISO 7991 : 1987. During this revision, the Committee felt that it would be more
convenient to prepare this standard by adoption of ISO 7991 on dual number basis.
The text of ISO Standard has been approved as suitable for publication as an Indian standard without
deviations. Certain conventions are, however, not identical to those used in Indian Standards. Attention
is particularly drawn to the following:
a) Wherever the words ‘International Standard’ appear referring to this standard, they should
be read as ‘Indian Standard’.
b) Comma (,) has been used as a decimal marker while in Indian Standards, the current
practice is to use a point (.) as the decimal marker.
In this adopted standard, reference appears to certain International Standards wherein the standard
atmospheric conditions to be observed are stipulated which are not applicable to tropical/subtropical
countries. The applicable standard atmospheric conditions for Indian conditions are 27 ± 2°C and
65 ± 5 percent relative humidity and shall be observed while using this standard.
In this adopted standard, reference appear to the following International Standard for which Indian
Standard also exists. The corresponding Indian Standard, which is to be substituted in its place, is
listed below along with its degree of equivalence for the edition indicated:
The technical committee has reviewed the provision of the following International Standard referred
in this adopted standard and has decided that its is acceptable for use in conjunction with this
standard:
For the purpose of deciding whether a particular requirement of this standard is complied with, the
final value, observed or calculated, expressing the result of a test, shall be rounded off in accordance
with IS 2 : 1960 ‘Rules for rounding off numerical values (revised)’. The number of significant places
retained in the rounded off value should be the same as that of the specified value in this standard.
IS 5623 : 2012
INTERNATIONAL STANDARD ISO
ISO 7991
7991 : 1987
: 1987 (E)
Indian Standard
GLASS — DETERMINATION OF COEFFICIENT OF
MEAN LINEAR THERMAL EXPANSION
Glass - Determination ( Second
of coefficient
Revision )of mean linear
thermal expansion
1 Scope and field of application For the purposes of this International Standard, the nominal
reference temperature, to, is 20 OC; therefore the coefficient of
This International Standard specifies a method for determining mean linear thermal expansion is denoted by a(20 OC; t).
the coefficient of mean linear thermal expansion of glass in the
elastic solid-like state, well below the transformation tempera-
ture (see ISO 7884-8). .
4 Apparatus
This International Standard applies to all glasses of normal
bulk-production compositions. lt does not apply to fused silica, 4.1 Device for measuring the specimen length, to an ac-
glass ceramics or other glasses of similarly low linear thermal curacy of 0,l %.
expansion coeff icients.
4.2 Push-rod dilatometer, capable of determining changes
in length of the specimen of 2 x 10V5 l. (i.e. 2 Fm per
2 References
100 mm).
ISO 7884-8, Glass - Viscosity and viscometric fixed Points -
The contact forte of the extensometer should not exceed 1 N.
Part 8: Determination o f fdila tome tric) transforma tion tem-
That forte shall act through contacts of planes with spherical
pera ture.
faces the radii of curvature of which shall be not less than the
rod diameter of the specimen. In some special assemblies (sec
IEC Publication 584-1, Thermocouples - Part 7: ßeference
figure 1) parallel planes are needed.
lt is given by the following equation: 4.3 Furnace, compatible with the dilatometer assembly, for
temperatures up to 50 OC above the expected transformation
temperature. The working Position of the furnace relative to the
1 1 - 10
a(t,; t) = 7 x - . . . (1) dilatometer assembly shall be defined with a repeatability sf
t - to
0 0,5 mm in both the axial and the radial directions.
where
Within the range of testing temperatures (i.e. up to tempera-
tures about 150 OC below the highest expected transformation
f0 is the initial or reference temperature;
temperature, t9, and at least up to 300 OC), the furnace shall be
capable of marntaining a constant temperature to + 2 OC over
t is the actual (constant or variable) specimen tem-
the whole specimen length.
perature;
l. is the length at temperature t. of the specimen of glass 4.4 Furnace control device, suitable for the desired rate of
under test (usually a rod made from the glass); increase in temperature up to (5 + 1) OC/min within the test
range (sec 6.1) and for a cooling rate of (2 + 0,2) OC/min for
I is the specimen length at temperature t. the annealing procedure according to 5.2.
1
1
IS
ISO5623
7991: 2012
: 1987 (E)
ISO 7991 : 1987
4.5 Temperature-measuring device (e.g. a thermocouple wait for about 5 min before beginning the test as described in
of type E, J or K in accordance with IEC 584-11, capable of 6.3 or 6.4.
determining the temperature of the specimen to + 2 OC in the
temperature range between t. and t. 6.3 Test at increasing temperature
NOTE - The specimen may be, for example, a rod either with a cir- As the dilatometer readings of AI,,,ea,are recorded during the
cular Cross-section having a diameter of 5 mm or with a Square cross- increase in the temperature between t. and t (values Chosen in
section 5 mm x 5 mm, and between 25 and 100 mm in length. In cer- accordance with 6.11, it should be borne in mind that a
tain cases, a Cross-section of at least 100 mm2 is more convenient (sec temperature differente will exist between the hot junction of
the annex).
the thermocouple and the test specimen; therefore a correction
shall be applied to the apparent temperature of the test
5.2 Preparation specimen.
The test specimen shall be annealed before the test by heating NOTE - The magnitude of this correction depends on the rate of
it to about 30 OC above the transformation temperature and temperature Change and the rate of heat exchange between the fur-
then cooling it to about 150 OC below the transformation nace and the test specimen. lt is essential that the correction is deter-
temperature at a rate of (2 + 0,2) OC/min, followed by further mined experimentally by comparison with measurements at constant
cooling to room temperature in draught-free air. temperatures.
The test shall be carried out with two test specimens (sec also Determine the Position of the dilatometer at the initial
7.4). temperature t. and take this reading as zero for the uncorrected
Change in length, Almeas, which will be measured. Subse-
quently heat the furnace to the selected final temperature t and
hold it constant to + 2 OC for 20 min. Then take from the
6 Procedure dilatometer reading the value of AZ,,,,.
All readings of temperatures and temperature differentes shall From the measured Change in length, Al,eas, the corrected
be taken with an accuracy to 2 OC. Though these actual values length I at temperature t is calcu lated using the following equa-
are used in the calculations in accordance with clause 7, the tion :
test range shall be expressed in terms of the nominal tem-
peratures (see 7.4). For a given coefficient a(20 OC; t) ex- I = Z. + AZ,,,, + AZQ - AIB . . . (2)
pressed in terms of the nominal temperature, no influence on
the value of the coefficient tan be detected within the limits where the correction terms AI0 and AZBare explained in 7.2
specified for the preferred actual temperatures. and 7.3 respectively.
6.2 Determination of the reference length 7.2 Calculation of the expansion of the specimen-
holding assembly
Determine the reference length Z. of the annealed specimen
(sec 5.2) to an accuracy of 0,l % at the reference temperature In the case of a simple push-rod dilatometer, the correction
to. Subsequently insert the specimen into the dilatometer and ter *rn AlQ in equation (2) is the thermal expansion of that part of
22
IS 5623: :1987
ISO7991 2012(El
ISO 7991 : 1987
the specimen-holding assembly alongside the specimen, hav- The measurements on glass and the blank test shall be carried
ing the length I, at temperature tO. out under identical conditions. The blank test shall be repeated
at least whenever a Performance test in accordance with
In the case of a differential push-rod dilatometer, the correction clause 8 is carried out.
term AZ0 is the expansion of a reference rod with the specimen
length Z0at temperature fO. 7.4 Calculation of the coefficient of mean linear
thermal expansion
In either case, the correction term AZ0 is calculated using the
following equation In Order to calculate the coefficient of mean linear thermal ex-
pansion, ab,; t), insert the measured values of I, and Almeas,
Al0 = \* l aQ l (t - t,)
1) Enquiries about sources of certified reference materials (CRMs) may be addressed to the S ecretariat of REMCO, Inter ,national Organization for
Standardization (ISO), 1, rue de Varembe, Case postale 56, CH-121 1 Geneva 20, Switzerland
3
3
IS
ISO5623
7991: 2012
: 1987 (E)
ISO 7991 : 1987
d) type of push-rod dilatometer used; For the temperatures t. and t, use the nominal values (sec 7.4).
4
ISO7991
IS 5623 :: 1987
2012(EI
ISO 7991 : 1987
Annex
ldeally the axes of the test specimen and push-rod coincide, An example for minimizing changes in alignment in a
and the length I, should lie in the same axis. In practice, small dilatometer assembly working almost vertically is illustrated in
deviations between the axes of the test specimen and push-rod figure 1. The guiding devices made from platinum wire prevent
may occur. Such deviations are negligible only when that further lateral changes in the Position of specimen and push-
misalignment remains constant throughout the test. Similar rod once the stable Position is achieved by slight shaking. The
considerations hold true for the push-rod direction and the axial movements caused by thermal expansion, however, are
working direction of the extensometer. Changes in alignment not hindered. Exactly vertically mounted dilatometer
(e.g. caused by Vibration of the apparatus) shall be avoided by assemblies have been found to be the most sensitive with
appropriate devices as shown in the examples (figures 1 respect to changes in alignment during the test.
and 2).
5
IS 5623 : :2012
ISO7991 1987 (E)
ISO 7991 : 1987
1 Specimen-holding tube with dilatometer base and sealed end plug ground flat perpendicular to tube axis, made from fused silica
2 Push-rod made from fused silica
3 Guiding device for the push-rod, made from platinum wire, 0,5 to 1 mm in diameter
4 Guiding device for the specimen, made from platinum wire, 0,5 to 1 mm in diameter
NOTE - Between the base and the push-rod, half the tube is tut away so that the test specimen tan be easily changed.
Figure 1 - Example of a specimen-holding and push-rod assembly of a dilatometer working almost ver tically
6
IS 5623
ISO 7991 ::1987
2012(EI
ISO 7991 : 1987
An example for minimizing changes in alignment in a distance holders. The push-rod is also supported by two
dilatometer working horizontally is illustrated in figure 2. The spheres of suitable diameter guided on the same guide-rod.
support for the specimen consists of four spheres (e.g. made After the apparatus has been shaken gently the test specimen
from ruby or fused silica), a cylindrical guide-rod, and suitable and push-rod achieve a stable Position.
NOTE - Between the base and the push-rod, half the tube is tut away so that the test specimen tan be easily changed.
7
Bureau of Indian Standards
BIS is a statutory institution established under the Bureau of Indian Standards Act, 1986 to promote
harmonious development of the activities of standardization, marking and quality certification of goods
and attending to connected matters in the country.
Copyright
BIS has the copyright of all its publications. No part of these publications may be reproduced in any form
without the prior permission in writing of BIS. This does not preclude the free use, in course of imple-
menting the standard, of necessary details, such as symbols and sizes, type or grade designations.
Enquiries relating to copyright be addressed to the Director (Publications), BIS.
Amendments are issued to standards as the need arises on the basis of comments. Standards are also
reviewed periodically; a standard along with amendments is reaffirmed when such review indicates that
no changes are needed; if the review indicates that changes are needed, it is taken up for revision. Users
of Indian Standards should ascertain that they are in possession of the latest amendments or edition by
referring to the latest issue of ‘BIS Catalogue’ and ‘Standards: Monthly Additions’.
This Indian Standard has been developed from Doc No.: CHD 10 (1786).