Professional Documents
Culture Documents
Phacoemulsification Procedur
Phacoemulsification Procedur
PHACOEMULSIFICATION
irrigation
While working within the eye, irrigation fluids maintain the
optimal depth of the anterior chamber. This decreases risk
of injury to delicate corneal tissues.
The irrigation system design is simple and based on
gravity flow. A nonvented bottle of irrigating solution is
suspended from an intravenous fluid hanger attached t o the
rear of the unit. The bottle hanger is calibrated and adjus-
table so that the hanger and bottle may be moved up or
down, depending on the patient's position, to achieve the
desired depth of the anterior chamber. During setup, the 0
point of the IV pole should be lined up with the patient's eye
level. The water level in the irrigation tubing drip chamber
is set at 65 cm above the 0 point. Balanced salt solution
(BSS, Alcon) is the most frequently used solution, and
many surgeons prefer to add the Plus (Alcon) component,.
which contains bicarbonate, dextrose, and glutathione (a
tripeptide important in cellular respiration). These addi-
tives have been reported to facilitate sustenance of corneal
e n d ~ t h e l i u m .Epinephrine
~ 1:1000 (0.3 to 0.5 ml) is often
added to the 500 ml of irrigating solution to maintain
necessary mydriasis during the procedure. Maximum dila-
tion of the pupil is required for visual exposure of eye
structures throughout the procedure and to avoid inadvert- Figure 1. Lumen diameters in millimeters of irrigationlas-
ent fragmentation o r aspiration of the iris. piration tips with corresponding identification bands.
The irrigation handpiece is a hollow-handled instrument
that connects t o the irrigation tubing. Various devices such
as bent-tipped needles for anterior and posterior capsulot- vacuum levels tested on both the I / A and ultrasound (U/S)
omy, cystitome, and capsule polishers can be attached to handpieces (see section on procedure setup).
the handpiece. Aspiration of lens material is accomplished by the peris-
T o convey fluids to the handpiece, the drip chamber on taltic pump. This pump creates a vacuum within the tubing
tubing supplied in the sterile kit is passed off the field and by the actions of a n internal roller that simply "milks" the
inserted into the fluid supply bottle on the IV pole hanger. aspiration tubing. T o aspirate material, the operating tip of
The circulator carefully secures a section of tubing into a the handpiece engages lens fragments and a vacuum build-
pinch valve (irrigation solenoid) on the front console, while up begins due to occlusion of the tip lumen. The peristaltic
a t the sterile field the luer fitting is attached t o the hand- pump continues t o run, creating the vacuum strength
piece. The irrigation solenoid valve pinches the irrigation necessary to aspirate the material into the tip and convey it
tubing closed in footswitch position 0 (zero) and stops the through the tubing and into the drainage bag located a t the
flow of fluid. On the console, the pinch valve is opened in rear of the unit. A safety mechanism is built into the PEA
footswitch positions 1,2, and 3, and fluid is allowed t o flow that limits the vacuum potential t o the preselected level so
through the tubing into the handpiece and out the surgical that vacuum strength cannot collapse the anterior chamber
tip being used. of the eye by rising t o excessive levels when the occlusion
breaks and material starts moving through the tip.
Aspiration The manual vacuum-select switch on the front console
The aspiration function of KPE serves two purposes: it controls vacuum levels during the surgical procedure. In
removes material from the eye and holds lens particles the U/S position, vacuum levels are equivalent t o 41 mm
against the ultrasonic tip t o allow efficient fragmentation of Hg, the 11A MIN mode provides a level of 65 mm Hg, and
the cataract. 1 / A MAX mode 370 mm Hg. In the I / A MAX mode, it is
The irrigation/ aspiration (I/ A) handpiece provides irri- recommended that only I / A tips of 0.3 mm or 0.2 mm be
gation in footswitch position 1 o r permits simultaneous used, as vacuum strength in a larger lumen may cause
irrigation and aspiration in footswitch position 2. The I/ A collapse of the anterior chamber.
handpiece is used to remove residual cortical material after The aspiration function includes a vent system. Lens
ultrasonic fragmentation of the lens nucleus. A variety of material is held by the tip until it is either aspirated o r
I / A tips are available with lumen diameters as follows: 0.2, released by a break in the vacuum. T o break the vacuum,
0.3, 0.5, and 0.7 mm. Identification bands on each tip the aspiration line can be momentarily opened (vented) to
indicate the size of the lumen (Figure I). the outside air by shifting the footswitch from position 2 to
During setup, the h e r fitting of the aspiration tubing is 1. This allows air t o enter the tubing at the Cam Lock
connected t o the I / A handpiece and the tubing is passed off T-fitting, which rapidly reduces the vacuum in the system.
the sterile field. A Cam Lock T-fitting is inserted into the The vent stystem allows control of vacuum strength poten-
aspiration vacuum port on the console, tubing is threaded tial, as well as manipulation of lens material.
around a peristaltic pump also o n the console, and the Occasionally, lens material completely occludes the aspi-
tubing end inserted into a drainage bag. After setup, the ration line. A bulb on the aspiration manifold proximal to
irrigation and aspiration manifolds should be primed and the handpiece may be manually compressed, which will
8-THE SURGICAL TECHNOLOGIST SEPTEMBER 1990
flush the particle through and restore normal aspiration fragmentation and aspiration. F o r these reasons, a tip with
function. a large bevel a n d oval lumen may be preferred.
The KPE instrumentation must be tuned before use to
Ultrasound ensure that the electronic characteristics of the machine are
The ultrasonic handpiece provides three functions that balanced with the handpiece. Once tuned, the U/S power
occur separately o r simultaneously: irrigation (footswitch control switch setting can be changed without altering the
position I), irrigation and aspiration (footswitch position electronic tuning. When the setting is increased, the power
2). and irrigation, aspiration, and fragmentation (foot- output of the ultrasonic generator is increased and the back
switch position 3). Two significantly different handpieces and forth excursion rate of the ultrasonic tip is increased. If
are available with the PEA system. the stroke is too forceful, lens fragments will be thrown
One type of U/ S handpiece turns electrical impulses into away from the tip, scattering within the eye and causing
ultrasonic vibrations by way of magnetoscrictive mecha- possible damage. Decreasing the setting prevents this
nisms. The internal component of the handpiece is called "chattering" of small nuclear particles against the corneal
the Acoustic Vibrator and is composed of numerous, flat, endothelium.
metal bands joined at both ends. These bands elongate and
contract rapidly due to changes in the magnetic field Operative Procedure
created by electrical impulses coming from the U/S genera- Phacoemulsification is an increasingly preferred form of
tor - after receiving signals from the footswitch - within cataract extraction. A study presented a t the 1989 annual
the PEA unit. The rapid vibrations of the metal bands meeting of the American Society of Cataract and Refrac-
convey linear motion to the instrument tip, allowing effec- tive Surgery showed that among surgeons performing over
tive fragmentation of the lens nucleus. The magnetostric- 51 cataract extractions per month, 65% prefer phacoemul-
tive handpiece is assembled a t the field after sterilization sification.5 This figure continues to increase each year. In
and disassembled for cleaningafter each use. It is used with addition, 86% of ophthalmologists surveyed have taken a
a power cord that must be ETO gassed o r soaked for phaco course, and 63% plan on using phaco more often in
sterilization. the future.
A newer handpiece utilizes piezoelectric (piezo is Greek The following discussion assumes knowledge of oph-
for pressure) mechanisms t o stimulate the instrument tip. thalmic anatomy, physiology, and previous experience
This handpiece has a one-piece design, is not assembled] dis- with cataract extractions; emphasis is placed on KPE tech-
assembled for cleaning, and is autoclavable with the cor- nique, without reference to insertion of an IOL.
responding power cord. T o produce ultrasonic vibrations,
electrical impulses from the PEA unit stimulate crystals Patient Selection
that rapidly expand and contract. This mechanical energy Considerations included in patient selection for KPE are
is then conveyed t o the instrument tip imparting the same (1) extent of pupillary dilation (minimum of 6.0 mm), (2)
linear motion as the magnetostrictive model. hardness of the lens nucleus (hard nuclei are more difficult
The conversion of electrical energy into mechanical and require more time to emulsify), (3) condition of the
energy creates heat that can damage eye tissues. Therefore, cornea (clear, with no evidence of endothelial disease), and
a cooling system is provided t o ensure that the handpiece (4) depth of the anterior chamber, particularly if the ante-
and tip remain cool throughout the procedure. In older rior chamber technique is employed.
models, a reservoir for distilled water is located in the
bottom of the PEA unit. During ultrasonic activation, the Procedure Setup
cooling pump circulates this nonsterile fluid through the Preparation for KPE is similar to planned ECCE with
U / S power cord to the magnetostrictive handpiece and respect to general ophthalmic instrumentation. However,
back. Due t o the mechanics of the cooling system, several setup procedures vary according to the PEA model being
factors are crucial to the proper and safe operation of the u'sed, and instructions included in the operator's manual
U /S handpiece. Care must be taken t o assemble the hand- should be followed carefully. The following are general
piece correctly; nonsterile fluids might otherwise leak onto setup procedures.
the sterile field. In addition, the appropriate level of coolant 1. Drape the tray holder and form the tray pouch.
must be maintained, and bacterial growth in the cooling 2. Place the sterilized phaco instrument tray on tray
system must be controlled. Refer to the operator's manual holder.
for the specific procedures. I n contrast, the newer piezoe- 3. Assemble the ultrasonic handpiece.
lectric handpieces are air cooled, avoiding the hazard of On the magnetostrictive model, make sure the O-rings
nonsterile fluids. are properly and securely placed (Figure 2, see p lo), insert
The U/ S instrument tip is a removable, single-use, hol- the Acoustic Vibrator into the ultrasonic handpiece, attach
low, titanium needle. T o protect adjacent eye tissues, the the ultrasonic nose cone, attach the ultrasonic tip to the
needle is covered before use with a soft silicone sleeve to acoustic vibrator using the tip wrench, attach the tip cap/ s-
expose only the last millimeter of the sharp tip. Linear leeve, and place the tip protector on assembled handpiece.
motion produces vibrations longitudinally along the axis of Hand off the ultrasonic handpiece power cord connector to
the needle, forward and back, a t a frequency of 40 kHz the circulator.
(40,000 times/ second). The U/ S needle is available in vary- On the piezoelectric model, connect the ultrasonic tip to
ing bevels: 15,30,45, and 60 degrees, with either a round or the handpiece, attach the tip cap; sleeve, place the tip pro-
oval lumen. In general, bevels with a greater degree of tector on assembled handpiece, and pass off the power cord
angulation allow the surgeon t o view the entire cutting edge connector to the circulator.
of the tip for safer manipulation, yet proper occlusion of the On both models, thesilicone cap! sleeve is placed over the
lumen is more difficult. However, the oval-shaped lumen titanium tip so that the two irrigation ports on the sleeve are
restores occlusion mechanics to the tip providing effective facing laterally and medially to the flat bevel surface of the
tip. lrrigatingfluids pass between the tip and sleeve, exitirfg the tuning meter is deflected to the right as far as possible.
through these two ports. As the surgical tip is used in a At this point, a characteristic sound is heard that the well-
bevel-up position, this allows irrigating fluids to be directed trained ear will immediately recognize.
laterally and medially, not toward the cornea and posterior 9. After operational checks are completed, place irriga-
capsule. tion manifold on irrigation handpiece and attach irrigating
4. Assemble the I/ A handpiece. cystitome tip.
Insert the interbody in handpiece housing, tighten
securely, attach the I / A tip using the tip wrench, and attach Surgical Technique
the tip caplsleeve. The irrigation ports of the sleeve should Incision. After anesthesia, a conjunctival flap is created.
be oriented 90 degrees away from the aspiration port to Scissors expose a 4.0 mm zone of sclera that is then cleaned
avoid conflict in fluid dynamics. and vessels cauterized. An incision along the limbus allows
5. Attach the irrigation and aspiration manifolds t o the entry into the anterior chamber. The incision is measured
I / A handpiece and pass off drip chamber and drainage with callipers to equal a chord length of 3.0 mm, parallel
tubing ends to the circulator. with the iris plane.
6. Manifolds are properly secured into appropriate fit- Anterior Capsulotomy. The irrigating cystitome is de-
tings on the PEA console and into fluid receptacles by the signed t o supply irrigating solution t o the anterior chamber;
circulator (Figure 3). This can be accomplished on recent this maintains the distance between the cornea and lens
PEA models by insertion of a disposable cassette that while the anterior lens capsule is being opened. Some sur-
houses all manifold connections. geons also use sodium hyaluronate (Healon). In footswitch
7. Prime the manifolds on the I / A handpiece using the position 1, the cystitome tip is inserted into the anterior
test chamber. Check vacuum levels. For all operational chamber and used to incise the capsule using an 0- or
checks, hold the handpiece horizontally at the PEA tray H-shaped "can-opener" incision. Numerous small cuts are
level. made in the anterior capsule, close together, radial to the
Priming Manifolds/Vacuum Check. For all operational optic axis (Figure4,A). Alternatively, some surgeons prefer
checks, hold the handpiece horizontally at the PEA tray a "single tear" method with the cystitome, while others use
level. T o prime the manifolds, the circulator sets the the Nd:YAG laser for anterior capsulotomy. After opening
vacuum t o I/ A MAX; depress the footswitch to position 2 the capsule, the tip of the cystitome is used t o snag a free
to allow irrigating fluids to fill tubing and test chamber, edge of the anterior capsule to determine if it is fully separ-
removing all air. T o check vacuum levels, activate the I/ A ated. The anterior capsule is either coaxed through the
function by depressing the footswitch to position 2 and incision and cut, or later aspirated.
manually occlude both manifolds simultaneously for five Posterior Chamber Phacoemulsification. The lens can be
seconds; release. The test chamber, representing the ante- effectively fragmented with ultrasound i.n the posterior
rior chamber of the eye, should remain inflated to indicate chamber. This is preferred if the anterior chamber is shal-
that vacuum levels and fluid mechanics are operating nor- low as it causes less corneal endothelial cell loss; however, it
mally and anterior chamber collapse will be avoided. Refer involves risk of rupturing the posterior capsule.
to the troubleshooting section of the operator's manual if Anterior Chamber Phacoemulsification. Kelman's tech-
the test chamber collapses. nique originally involved an anterior chamber method.
8. Repeat the aforementioned operational check with After capsulotomy, the lens is prolapsed into the anterior
the manifolds on U / S hand piece. In addition, check ultra- chamber using the tip o r flat edge of the cystitome. Once the
sonics using the test chamber. irrigation and aspiration manifolds are secured on the
Ultrasonics Check. If automatic tuning is not available, a ultrasonic handpiece, the titanium tip is inserted bevel
U /S power setting of 8 is selected, the footswitch depressed down, rotated upward, and used t o fragment and aspirate
to position 3, and the tuning control switch is rotated until the lens nucleus in footswitch position 3. As a safety mea-
10-THE SURGICAL TECHNOLOGIST SEPTEMBER 1990
Figure 4. A, anterior capsulotomy; B, anterior chamber phacoemulsification; C, aspiration of cortical material off
posterior capsule; and D, aspiration of cellular debris; posterior capsule polishing.
sure, the U /S power setting on the console is not selected engaged, peeled off the posterior capsule, and aspirated
until successful entry into theanterior chamber is complete. (Figure 4,C).
Three different methods of fragmenting the lens nucleus At this point, the posterior capsule is cleaned using the
can be used. One method involves fragmenting a pie- I/ A tip, o r the irrigating manifold can be transferred to the
shaped sector of nucleus, rotating the lens to another sec- irrigation handpiece and a capsule polishing/ scraping tip
tor, and repeating this process until fragmentation is com- used to remove particle debris on the posterior capsule
plete. The "carousel" or "cartwheel" technique is effective (Figure 4,D).
with a soft nucleus and involves rotating the lens around the In some cases, a posterior capsulotomy is performed on
tip of the stationary titanium needle using a spatula o r opaque capsules, and a peripheral iridectomy or iridotomy
nucleus rotating instrument that has been inserted through is performed to decrease the risk of pupillary block. Finally,
a side port. The "croissant" technique involves boring to complete the KPE, one or two closing sutures are placed.
through the center of the nucleus in a sagittal direction
using short energy bursts (Figure 4,B). Complications
KPE is the procedure of choice for soft nuclei. However, Studies show that KPE compares favorably with other
a pretreatment procedure allows surgeons to utilize KPE methods.+" However, as with any surgical procedure, com-
on even the hardest, most sclerotic lenses in grades 111 and plications may result. Several factors contribute to the
IV. Nd:YAG laser photodisruption of the lens nucleus complexity of KPE.
before KPE has been reported to decrease the phaco time Machine
for lenses in all grades.6a7 Laser pulses create shock and Proper depth of the anterior chamber is vital to the success
pressure waves that mechanically disrupt tissues adjacent of KPE. T h e anterior chamber may become too shallow or
to areas disintegrated by the focused laser beam. This dis- even collapse if the height of the irrigation bottle is too low
persion property effectively softens any hard nucleus which or if the vacuum level is set too high for the 1,'A tip lumen
might otherwise be considered too dense for effective KPE. size being used. Conversely, the anterior chamber may
Irrigation/Aspiration. After the nuclear fragments have become too deep and cause iris prolapse if the irrigation
been aspirated, the manifolds are transferred t o the li A bottle is too high. Faulty power settings can also disrupt
handpiece. and the power setting placed on 1/A MAX to KPE. All of these factors are easily eliminated by proper
allow high vacuum power t o aspirate residual cortical maintenance, testing, and troubleshooting of the PEA
material from beneath the iris. The vacuum setting is then unit's functions and associated instrumentation prior to the
changed to I / A M IN and the remaining edge of cortex is operative procedure.