Download as pdf or txt
Download as pdf or txt
You are on page 1of 3

IEEE TRANSACTIONS ON COMPONENTS, HYBRIDS, AND MANUFACTURING TECHNOLOGY, VOL. 12, NO.

I , MARCH 1989 121

Rapid Nondestructive Testing of Ceramic


Multilayer Capacitors
OTMAR BOSER, PAULETTE KELLAWON, AND RICHARD GEYER

Abstruct-A rapid nondestructive test method for ceramic multilayer layer capacitors represent such density variations, that can
capacitors made from piezoelectric materials such as barium titanate or shift or even inhibit the appearance of the acoustic resonances
lead-containing materials is described and evaluated. The test method is
depending on the size of the delamination. The nature of the
based on the internal excitation of standing acoustic waves in the
capacitor. The standing waves are severely dampened by defects such as resonances was studied as a function of capacitor size [8]. It
delaminations and pores. An undampened resonance is a good indication was found that only longitudinal acoustic waves can be excited.
of a defect-free ceramic multilayer capacitor. The frequencies of the main acoustic resonances are deter-
In addition, a high-speed automated test setup was evaluated that has mined by the three dimensions of the capacitor. The width
the potential of the testing of 180 OOO capacitors per hour with a sorting
accuracy of about 92 percent.
resonance, i.e., the resonance produced by a standing wave
parallel to the electrodes and perpendicular to the line
connecting the contacts is usually the most pronounced. The
INTRODUCTION optimum bias voltage was found to be 50 V or 1.5 MV/m for
Z5U capacitors and 200 V or 7 MV/m for X7R capacitors. At
C ERAMIC multilayer capacitors are composite structures
made from alternating layers of metal (electrodes) and
ceramic (dielectric material). As can be imagined, joining
that voltage, the width resonance exhibits the lowest losses.
A similar test procedure was proposed recently for trans-
such different materials into a monolithic structure presents ducers [9]. The difference between transducers and capacitors
problems especially since the sintering of the ceramic takes is their poling behavior: Transducers can be permanently
place between 1200°C and 1500°C depending on the type of poled whereas capacitors are poled only as long as the dc bias
ceramic. The electrodes are generally much thinner than the is applied.
dielectric and the former can act as crack initiation sites if the To determine the resonance behavior of the capacitors, the
ceramic multilayer capacitor has not been heat-treated care- capacitance and impedance of the ceramic multilayer capaci-
fully. Therefore, it is possible to have delaminations between tors were measured with a Hewlett-Packard low-frequency
electrode and ceramic that cannot be detected by capacitance impedance analyzer, HP model 4192A, that has a frequency
measurements. Currently delaminations are detected in a range from 5 Hz to 13 MHz. The impedance analyzer was
destructive fashion, i.e., the production is sampled periodi- controlled by an HP computer, model 9836, that made it
cally and the ceramic multilayer capacitors are metallographi- possible to sweep the frequency over the frequency range of
cally inspected. If the samples tested do not show delamina- the resonance and print out the results on an HP plotter.
tions the whole batch is declared good. The impedance analyzer has a built-in dc bias of up to 35 V.
Nondestructive testing of ceramic capacitors is difficult, For measurements with a larger bias, a sample holder was
because the testing method has to be rapid, efficient, and low- constructed according to the HP manual [lo]. This sample
cost. Different test methods have been proposed: neutron holder allowed the use of bias voltages of up to 200 V. The
irradiation [11, scanning acoustic microscope [2], and acoustic setup was calibrated in the shorted and open conditions at the
emission [3]. All test methods have shown a measure of upper frequency of the swept frequency range.
success. In the following, a new nondestructive test method [4]
for Z5U and X7R capacitors will be discussed that has the TESTSAMPLES
AND MECHANICAL
RESONANCES
advantage of being rapid, efficient, and low-cost. As test samples Z5U capacitors with nominal dimensions of
0.10 x 0.05 x 0.05 in (2.5 x 1.25 x 1.25 mm), i.e., size
NONDESTRUCTIVE
TESTMETHOD 1005, were used with a capacitance value of 20 nF. The test
Most ceramic multilayer capacitors are made from piezo- batch of ceramic capacitors showed an exceptionally large
electric materials. Therefore, capacitors that are biased with a incidence of delaminations. With the sound velocity estimated
dc field become temporarily poled and can act as transducers in [8] as about 5000 m/s one obtains for the length and width
and exhibit mechanical resonances. As shown elsewhere [5]- resonance 1 and 2 MHz, respectively. However, it has to be
[7], mechanical resonances are affected by density variations remembered that the cross section of the capacitor is square
in the structures under test. Delaminations in ceramic multi- and hence different modes than the ones assumed in [8] can
appear that lead to deviations from the calculated resonance
Manuscript received February 25, 1988; revised September 20, 1988. frequency.
The authors are with Philips Laboratories, North American Philips
Corporation, Briarcliff Manor, NY 10510. The impedance of the Z5U capacitors of size 1005 was
IEEE Log Number 8824724. measured with the above described test setup in the frequency

0148-641 1/89/03OO-0121$01.OO O 1989 IEEE


I22 I F F t TRANSACTIOKS O S COMPOUESTS. HYBRIDS. .Ah[) \t.ANCFACTUKING I ECHNOLOGY. VOL 1 2 . NO I . MARCH I989

200 GOOD

-
2
160

-
I
0
W
0
120 ~

z
4 60
0 I .

w - 005"
a
2 40
(a)
0 ~- BAD

zoo I

160 -
E
-
I
0
120
W
0 TYPE Z5U
z
a 60 SIZE . 0.10"x 0.05'
x 0.05"
0 CAPACITANCE : IOOnF
?U
a
2 40 (b)
Fig. 2 . (a) Micrograph of capacitor of size I005 that is free of delamina-
tions. (b) Micrograph of capacitor of size 1005 that contains a delamination.
0 -
0

B R E A K D O W N TEST DELAMINATION TEST

Fig. I . (a) Impedance versus frequencq for a delamination-free capacitor (20


nF) of size 1005 biased with SO V . (b) Impedance versus frequency for a
R1
capacitor with delaminations (20 nF) of sire 1005 biased with SO V .

range between 0.5 and 3.5 MHz with a 50-V bias. The results
are shown in Fig. 1. In Fig. l(a) the impedance versus
frequency curve of a good capacitor is shown whereas in Fig.
I(b) the same is shown for a delaminated capacitor. The
difference in the impedance at antiresonance between the good
and the delaminated capacitor is quite pronounced. This
difference in impedance is a good measure to determine the
quality of a ceramic capacitor.
Due to the square cross section. the width and the thickness
resonances occur at the same frequency (nominally 2 MHz). Fig. 3. Circuit diagram of the automated test oscillator.
The coincidence of the two standing waves perpendicular to
each other leads to a new vibrational mode that has a be no oscillations. The circuit is shown in Fig. 3. The op-amp,
resonance frequency of 1.8 MHz, slightly lower than the one and R3 and R4 form a noninverting amplifier with a gain of
predicted for single standing waves.
To confirm the suggestion that the change in impedance at Gain = ( R3+ R 4 ) / R 4 . (1)
antiresonance is caused by delaminations, the capacitors were
The resistor R5 lowers the gain of the amplifier at low
cut and polished. Fig. 2 shows the micrographs of the
frequencies to prevent spurious oscillations, and C , bypasses
capacitors and it confirms that the capacitor tested in Fig. l(a)
R5 at the antiresonance frequency of the capacitor under test
is good and the one tested in Fig. l(b) is delaminated.
(C,). The capacitor C2is a blocking capacitor and prevents the
OsCIu.-\mK TESTSETUP polling voltage from reaching the op-amp. At the antireso-
nance frquency the resistor R I and the impedance Z,, divide
To accelerate the testing of ceramic multilayer capacitors an
the output signal of the amplifier by the following factor:
oscillator test circuit was proposed that is combined with a
breakdown and poling test setup. The test circuit was designed
as a marginally stable oscillator circuit. where the antireso-
nance frequency of the capacitor determines the frequency of
oscillation and the magnitude of the antiresonance impedance where Z,, is the impedance of the capacitor under test at the
determines whether oscillations occur: If the impedance is antiresonance frequency. This divided signal is applied to the
high the capacitor is good and oscillations occur, however, if input of the amplifier. The condition for oscillation is that the
delaminations are present the impedance is low and there will gain of the amplifier and the voltage divider is greater than
BOSER er a1 TESTING OF CERAMIC MULTILAYER CAPACITORS 123

capacitors of the 50 with delaminations indeed had delarnina-


i tions and only one of the 50 that were supposed to be
delamination-free had a delamination. That means that the
VOLTAGE ON
above testing resulted in a correct identification of the
TEST CAPACITOR delaminated and delamination-free capacitors with a 92-
percent accuracy.
I
CONCLUSION

OUTPUT OF
OP-AMP

1
The test results show that a ceramic multilayer capacitor can
be nondestructively tested in a time span of about 20 ms. This
translates into 180 OOO capacitors an hour. In addition. the
nondestructive test setup can be integrated with the standard
voltage breakdown test reducing capital investment to a
minimum. The accuracy of the test is good and hence it can be
’1 SWITCH 2
OPEN
SWITCH 2
CLOSED
1
1
concluded that the above method is a rapid, efficient, and low-
cost nondestructive testing method for ceramic multilayer
I capacitors.
I I
k-- - 20ms ~ 4
SWITCH 1 OPEN SWITCH 1 CLOSED
ACKNOWLEDGMENT
Fig. 4 . Schematic of the timing necessary during testing. The authors want to thank Dr. S. Long, Mepco/Centralab,
Los Angeles, CA for the samples and the support of this
investigation.
unity or
REFERENCES
(3) G . F. Kiernan, “Comparison of screening techniques for ceramic
capacitors,” in Proc. Symp. (held at Marshall Space Flight Center),
NASA Conf. Publ. 2186, Marshall Space Flight Center, Alabama,
It should be remembered that resistors R2 and R6 shunt the 1981.
impedance Z,, of the capacitor under test, so that the effective C. L. Vorres, D. E. Yuhas, and L. W. Kessler “Non-destructive
evaluation of ceramic chip capacitors by means of the scanning laser
value of Zar is somewhat lower than the impedance at acoustic microscope, ” in Proc. 2nd Capacitor and Resistor Technol-
antiresonance. ogy Symp., pp. F31-1 to 3-6, Components Technology Inst., Suite
1122-K, 303 Williams Ave., Huntville, AB 35801, 1982.
HIGH-SPEED
NONDESTRUCTIVE
TESTING
AND RESULTS S . R. Kahn and R. W. Checkaneck, “Acoustic emission testing of
multilayer ceramic capacitors,” IEEE Trans. Comp. Hybrids,
In a nondestructive test the capacitor to be tested (CJ is Manuf. Technol., vol. CHMT-6, pp. 517-526, 1983.
inserted in the test setup with both switches closed. The test 0. Boser, US Patent 4 644 259, Feb. 1987.
V. M. Baranov and E. M. Kudryavtsev, “Application of the ultrasonic
starts by opening switch 1 to apply the poling voltage to the resonance method in the inspection of small products.” Sov. J .
capacitor. A period of 5 ms was found to be enough to let all Nondestruct. Testing, vol. 15, pp. 750-755, 1980.
transients die out in Z5U capacitors. Then switch 2 was L. R. Testardi, S . I. Norton, and T. Hsieh, “Determination of
inhomogeneities of elastic modulus and density for one-dimensional
opened to start the oscillations. It was left open for 12 ms. structures using acoustic dimensional resonances.” J . Appl. fhys.,
Then switch 2 and switch 1 are closed again. This sequence of vol. 56, pp. 2681-2685, 1984.
events is illustrated in Fig. 4. It shows that a nondestructive -, “Acoustic dimensional resonance tomography: Some examples
in one-dimensional systems,” J . Appl. Phys.. vol. 59, pp. 55-58,
test can be performed in less that 20 ms. For the actual test the 1985.
critical impedance Z,, to start oscillations was set at 127 Q 0. Boser, “Electromagnetic resonances in ceramic capacitors and
(effective value 112 0). evaluation of the piezoelectric materials’ properties,” Adv. Cer. Mat.,
vol. 2 , pp. 167-172, 1987.
A batch of 954 capacitors of size 1005 was tested. Of these, W. W. Wolney and K. Breboel, “An NDT-method of detection of
244 tested free of delaminations and 710 had delaminations. internal defects in piezoceramic transducers,” IEEE Trans. Ultrason.
From each of these two groups 50 capacitors were chosen at Ferroelec. Freq. Contr., vol. UFFC-33, p. 817, 1986.
Operations and Service Manual for Model #4192A Impedance
random for destructive testing. Metallographic preparation Analyser, Yokogawa-Hewlett-Packard Ltd., Tokyo, Japan, Mar.
techniques at different hights in the capacitor showed that 46 1983.

You might also like