Download as docx, pdf, or txt
Download as docx, pdf, or txt
You are on page 1of 21

 D.C.

Machines current, and a graph of terminal


voltage versus load current.
8. Complete an assignment report,
including comments regarding the
shapes of the two characteristics
compared with those predicted by
theory.

Assignment 2
3. Close switch S1 , measure and
tabulate the values of field current To obtain the output characteristics for

and generated emf. a shuntwound d.c. generator.

4. Increase the field current, in discrete Apparatus :


steps, and tabulate these current 1 X shunt generator
settings together with the 1 X drive motor
corresponding emf generated. 1 X rheostat
Continue this procedure until the 1 X ammeter
generator is producing its normally 1 X voltmeter
rated emf. 1 X single-pole switch
5. Now Close S2, measure and tabulate
Method:
the corresponding values of load
1 Connect the circuit as in Fig. 6.52,
current and terminal voltage.
leaving the
6. Increase the load current, in discrete
steps, tabulating the current and
terminal voltage at each step.
Continue this procedure up to the
rated full-load output for the
machine.
7. From your tabulated data, plot a
graph of generated emf versus field
terminal voltage at each step.
Continue this procedure up to the
full-load output of the machine

5 If possible, increase the load current


beyond the full-load value, and
observe the affectson both this
current and terminal voltage.

6 Prepare an assignment report. This


should include a plotted graph of the
Switch in the open position, and the generator output characteristic, and
rheostat set to its maximum comments regarding its shapes etc.
resistance setting.
Assignment 3
2 Drive the generator and note the
value of generated voltage. Note To obatian the output characteristic for

that this voltage should build up a compound generator , using both

rapidly to its normally rated output cumulative and differential

value. If self-excitation does mot compounding.

occur, switch off the drive motor, Apparatus :


reverse the connections to the 1X compound generator
generator still fails to self-excite, Remainder of apparatus as for
consult your lecturer. assignment 2.

3 Close the switch, note and tabulate Method:


the corresponding values of load As for Assignment 2, except that the
current and terminal voltage. procedure will be carried out for both
cumulative and differential
4 Increase the load current, in steps,
compounding . In order to change
noting the values of load current and
form one form of compounding to the
other , simply reverse the connections
to the series field winding. The circuit
is shown in Fig. 6.53.

■ 7 Measuring Instruments and Measurements ■

This chapter is intended to help you understand the principal features and
limitations of some commonly used measuring instruments. On completion you
should be able to :
1] Appreciate the sources of error that can occur when taking measurement .
2] Predict the loading effects and frequency response characteristics of
instruments.
3] Predict the affect of distorted waveforms on detector used in electronic
instruments.
4] Calculate the power consumed by instruments connected into circuits.
5] Appreciate the basic principles of a.c. bridge measurements
6] Select and correctly use appropriate intsruments for a variety of
measurements
7.1 Introduction ________________________________________________________________
The operating principles of moving instruments, As part of this experience
coil and rectifier moving coil you may well have come to the
multimeters; Wheatstone bridge; important realisation that no
slidewire potentiometer; and CRO measurement should ever be
were covered in Volume 1 of this title. considered as being absolutely correct.
In addition, you will by now, have Any measurement that is made will
gained some practical experience of involve a number of possible sources
simple voltage, current and resistance of error.
measurements, using the above

■ Calibration errors ■
The calibration of an instrument is the calibration errors, and may be shown
process where by its indicated values in the form of a graph, as in fig. 7.1
are compared against a known There will be some allowable limits of
standard. Since no instrument can be error specified for the instrument. The
made prefects in all respects, there will range of these limits determines the
be discrepancies between the indicated ‘grade’ which the instrument is
values and the ‘true’ values. These allocated; the lower the limits of error
discrepancies are known as the the ‘higher’ the grade. Figure 7.1
could represent the calibration graph
for a 0 to 30 V voltmeter, with a
maximum allowable error of ± 0.6 V.
The accuracy of an instrument is
always quoted in term of the
maximum allowable error, usually as a
percentage of the full-scale deflection
(fsd). Thus for the above example, the
accuracy of the instrument would be
quoted as being ±2% of fsd. It is worth For this reason, when using pointer-
noting that if the instrument was in on-scale instruments, it is advisable to
error by ±0.6V at an indicated value use an instrument ( or the appropriate
of, say, 10V, then the actual scale on the instrument ) so that the
percentage error at this indication indicated value is as near to fsd as is
would be : possible. When this is done, the
0.6 calibration error is minimized. The
± 𝑥 100% = ± 6%
10 affect of this form of error is illustrated
This is because the allowable error of
in Example 7.1.
0.6 V is a much more significant
proportion of 10V than it is of 30V.

Worked Example 7.1

Q The value of a 50 Ω resistor is to be verified by measuring the current


flowing
Q through it and the corresponding p.d. a cross it. Details of the two
meters, and the readings recorded are shown below. Determine the range of
values between which the actual resistor value may be said to lie.
Voltmeter : accuracy ± 2% of fsd; range 0 - 100V; indicated reading 75V
Ammeter : accuracy ± 2 % of fsd; range 0 - 3A; indicated reading 1.5A.

A
Therefore possible current = 1.5 ± 0.06
Possible voltmeter error = ± 0.02 X = 1.44 A to 1.56 A
100 = ± 2V
Therefore possible p.d = 75 ± 2 = 73 V Since R V/I ohm, then the maximum
to 77 V error of measurement will occur when
Possible Ammeter error = ± 0.02 X 3 one meter error is at its upper limit,
= ± 0.06 A and that for the other meter is at its
lower limit.
i.e when V = 77V and I = 1.44 A Therefore, the resistor value may be
or when V = 73V and I = 1.56 A said to have a value in the range
hence, R =
77
= 53.472 Ω 46.795 Ω and 53.472 Ω Ans.
1.44
73
or R = = 46.795 Ω
1.56

■ Systematic errors ■

This form of error is either avoidable, circuit is a common but easily


or may be taken into account when a avoidable systematic error. The
measurement is made. Failure to loading effect of an instrument , and
ensure that an instrument is correctly the points in the circuit to which it is
‘zeroed’ before connecting it to a connected may also be
taken into account in order to
minimize this form of error. This form
of error is illustrated in Example 7.2.

Worked Example 7.2

A 5 kΩ resistor is connected
Q across a 10 V supply. The p.d.
Q
across the resistor and the current
flowing through it are measured as
shown it fig. 7.2. The voltmeter has a
total resistance of 20 kΩ, and that of
the ammeter is 5 Ω. Assuming
negligible calibration errors, let us
determine the systematic errors in the
two meter readings, attributable to
their manner of connection
A

With neither meter connected, then the This is the value of current measured
‘true’ values for p.d and current will by the ammeter, so the error in this
be 10 V and 2 mA respectively. reading would be :
5 ×20 2.497−2
RAB = kΩ = 4 kΩ 2
×100% = +25%
5 +20

𝑅𝐴𝐵
RAc = 4.005 kΩ; and V = ×E=
𝑅𝐴𝐶
4
×10 = 9.888 V
4.005

This would be the p.d. across the


voltmeter, and assuming that it could
be accurately read down to three
The voltmeter was connected between
decimal places, then the error would
A and C, as in Fig. 7.3, then :
be :
Voltmeter reading = 10 V, i.e.
9.888−10
×100% = -1.2%
10 zero systematic error

This degree of error would be Current through ammeter = I =


acceptable. Indeed, the voltmeter 𝐸 10
amp = = 1.998 mA
𝑅𝐴𝐵𝐶 5005
reading would most likely be
So ammeter error
interpreted as 10 V. 1.998−2
= ×100% = -0.1%
The circuit current, 2

𝐸 10
I= amp = = 2.497 mA
𝑅𝐴𝐶 4005

This degree of error is most systematic error. It is left to the reader


acceptable, so this alternative to verify at if the voltmeter resistance
connection of the voltmeter is was (say) 10 MΩ, then the systematic
obviously better at reducing the total error introduced by the voltmeter
connection of Fig. 7.2 would be would have an infinite resistance
negligible. This clearly illustrates one (draw zero current). On the other hand
advantage of a digital voltmeter a perfect ammeter would have zero
compared with a moving coil resistance, and hence cause no
instrument. That is, a perfect voltmeter additional potential drop in the circuit.

■ Observational and random errors ■


All other forms of measurement error multimeter such as an AVO, and
may be included under this heading. inadvertently reads the indication from
Observational (human) error can occur the wrong scale, or misinterprets the
in a number of ways. When using a scale setting applied by the rotary
pointer-on-scale instrument, the range switch. This problem with the
pointer often comes to rest between range selection can also occur with
two marked graduations on the scale. digital instruments. With this form of
The reading recorded is therefore open instrument there can be no ‘argument'
to interpretation by the individual as to the digits displayed, but the
observer, and different observers will displayed value is normally subject to
rend tend to give slightly different in a ± 1 digit random error. Temperature
interpretations. Occasionally there and other environmental factors may
may be very large observational errors. also be the cause of random errors in
This is fairly common when a student measurements.
first uses a

7.2 Accuracy and Sensitivity_________________________________________________________


These two measurement terms are features of a measuring instrument or
often confused with each other, despite system.
the fact that they describe different
■ Accuracy ■
This has already been described as the it represents the closeness to the true
error referred to the fsd. Another way value that is obtainable.
to define accuracy would be to say that
■ Sensitivity ■

The sensitivity, discrimination or change of indication(output)


resolution of an instrument describes change of measured quantity(input)

the smallest change of the measured Comparing accuracy and sensitivity, it


quantity(measured) that can be can be said that a sensitive instrument
discerned on the display. Thus is not necessarily accurate; but an
sensitivity may be defined as: accurate instrument needs to be
sensitive.`

7.3 Total Measurement Error ______________________________________________________


The in-depth study of the theory of most common techniques will be
errors is beyond the scope of the outlined, in order to give an
course of study being undertaken here. appreciation of the result of total error.
However, a simplified courage of the

■ Additional and subtraction measurements ■

Consider two meter readings which are of the two readings, D1, and the total
to be added or subtracted to obtain a errors is:
final result, let D1 and D2 be the two D1 = D1 + D2 ± (e1 + e2 )
displayed readings, and e1 and e2 the And the difference
respective maximum errors. The sun Dd = D1 - D2 ± (e1 + e2 )
Hence the total error in the sum or the applied voltage and associated total
difference of two or more readings is error is obtained thus :
equal to the sum of the individual
Since meter has an accuracy of ± 2%
error. Note that e1 and e2 are the
of fsd, then absolute possible error is
absolute error in the readings, and not
± 2V. the possible values for V1 and V2
the quoted accuracies of the
are,
instruments. For example, suppose that
V1 = 60 ± 2 V and V2 = 20 ± 2 V
two p.d.s are measured in series d.c,
so, V = V1 + V2 = (60+20) ± (2+2) V
circuit, and the sum of the two
= 80 ± 4 V
represents the applied voltage. The
Thus the total possible error is ± 4 V
voltmeter used has an accuracy of ±
or ± 5%.
2% of fsd of 100 V. If the two p.d.s
indicated are V1 = 60 V and V2 = 20 V,

■ Multiplication and division of measurements ■

Consider two measurements and their normally very much greater than the
associated errors as follows: product δ1δ2
𝑒1 Therefore,
D1 ± e1 = D1 (1 ± ) = D1(1±δ1)
𝐷1

and similarly D2(1±δ2) Dp = D1D2 [1±(δ1+ δ2)]

Where δ1 = e1/D1; δ2 = e2/D2, are the So, for a product of two


fractional errors. The product of the meansurements, the total fractional
two readings yields the result, error is the sum of individual
fractional errorss. Similarly, it can be
DP= D1D2 (1±δ1) (1±δ2)
shown that when dividing one reading
= D1D2 [1±(δ1+ δ2) + δ1δ2]
by another, the quotient,
But since δ is generally a very small
𝐷1
quantitiy, then the sum (δ1+ δ2) is Dp = [1±(δ1+ δ2)]
𝐷2
■ Power or root of a measurement ■

Since a root is simply a factional fractional error in raising to a power n


power, then the same technique times the original fractional error, such
applies to booth. For example, the that:
square-root of a quantity expressed as Reading raised to the power
a power is, it is found that the total n = Dn(1±no)

7.4 Wattmeter Corrections ___________________________________________________________


When a wattmeter is used to measure drawn by the voltage or pressure coil.
the power in a circuit. Systematic The wattmeter reading is therefore:
errors are introduced. The amount of
P = VL (IL + IP) watt
error this introduced depends upon the
= VLIL + VLIP
way in which the wattmeter is
= load power + VLIP watt.
connected into the circuit. Consider
The wattmeter reading is therefore too
Fig 7.4, which shows a
high, by the amount of power
dissipated in the pressure coil. This
coil of the wattmeter will have a
relatively high resistance, so IP will be
correspondingly small. Thus, if IL ≥ 10
× IP, then the error introduced
negligible. If, on the other hand, the
Wattmeter connected between a
load current is of a comparable value
supply and its load. The wattmeter
to IP, then the error can be significant.
reading depends on the p.d. across its
In this situation the wattmeter reading
voltage coil. However, the current
should be corrected by subtracting the
through the current coil is the load
pressure coil power from the meter
current, IL, plus the small current IP
reading. This is possible provided that
the resistance of the pressure coil is plus the potential drop due to the
known, such that: resistance of the current coil. The total
𝑉𝐿 wattmeter reading is therefore:
PP = watt, or PP = IP2RP watt
𝑅𝑝

P = (VL + IRc)IL watt


The wattmeter may be connected so
= VLIL + IL2Rc
that the pressure coil is connected to
= load power + IL2Rc watt.
the supply side of the circuit as in Fig
7.5. in this situation the current This means that the meter reading will
again be high, and may be corrected
provided that the current coil
resistance is known. This form of
connection will minimize the error
when the load current (and hence the
current coil p.d) is relative small.

Through the current coil will be the


true load current. However, the p.d.
across the pressure coil is the load p.d

7.5 Electromagnetic Voltmeter _____________________________________________________


It has been shown that the main amplifier is normally based on a field
source of error for analogue effect transistor (FET), which can
voltmeter, such as the AVOmeter, have an input resistance in the order
occurs when measuring the p.d. of hundreds of megohms. This
across a high value resistor. This error amplifier may also be connected as a
is due to the loading effect of the DC (directly coupled) amplifier,
meter’s input resistance. Additionally, which enables it to have a bandwith
when used on the a.c. ranges the extending from 0Hz up to many
upper frequency limit will be in the megahertz. Since this amplifier can
order of 20 kHz. An electronic have considerable voltage gain, there
voltmeter reduces these sources of is a limitation on the size of input
error by employing a transistor voltage that may be applied to it. For
this reason a switchable attenuator much simplified diagram of such an
network is interposed between the instrument in shown in Figs. 7.6 (a)
meter terminals and the amplifier and (b).
input terminals. One disadvantage of
this attenuator network is to reduce
the overall input resistance of the
instrument to the region of 10-20
MΩ. However, this is still far greater
than that of AVO, so the loading
effect of the electronic voltmeter is
minimal. Another advantage of
employing an amplifier is that an
electronic voltmeter may be capable
of measuring voltages as low as
microvolts. The amplified voltage is
normally displayed on a conventional
moving oil movement. For this
reason, when the instrument is used
for a.c. measurements, a rectifier is
switched into the input circuit. A
Fig 7.6

7.5 Digital Voltmeter (DVM) ________________________________________________________


As with the electronic voltmeter, a frequency dependent components such
DVM has a high input resistance as capacitors, the bandwith of such an
because it employs a transistor arrangement is much less than that
amplifier and input attenuator. The achievable from an electronic voltmeter
amplifier is usually in the form of an indeed, some of the commonly used
operational amplifier connected as an (and ‘cheaper’) DVMs may have a
integrator. This, coupled with an bandwith of only 2-4 kHz. It is
oscillator, a counter and associated therefore most important to consult the
logic control circuitry forms an instrument data sheer to check on this
analogue-to-digital converter (ADC) aspect. Since an AVO has a better
circuit. Due to the presence of frequency response than the above
figure, then this instrument could prove d.c. measurements any DVM will
to be more reliable for an a.c. normally be preferable to an analogue
measurements at frequencies greater instrument.
than 2 kHz. For power frequency and

7.6 Complex Waveforms _____________________________________________________________


The concept of complex waveforms is the bandwith of the internal amplifier
dealt with in Volume 1, Chapter 9. may introduce error.
The implications regarding the
The best means of measuring a
bandwith required for the transmission
complex waveform is probably by
of complex waveforms was also
the use of an oscilloscope. This gives
described. The measurement of a
the added advantage that the actual
complex waveform voltage using an
waveshape can also be observed.
analogue moving coil meter will
However, for the displayed
normally result in a large error. This is
waveform to be a faithful
because such an instrument is
reproduction of the measured
calibrated to indicate the rms value of
waveform all of the harmonics
a sinusoid, by employing a correction
present in the original must be
factor equal to the form factor for a
amplified (or attenuated) by the same
sinewave. Depending upon the type of
amount. If this is not the case then
circuitry employed, electronic and
waveshape’ will be changed. Thus
digital instruments may introduce
the amplifiers employed must have a
either zero error or large error due to
very wide bandwith, with a flat
the complexity of the waveform. An
frequency response curve. The other
instrument that is quoted as being a
main problem concerns any time
‘true rms’ meter should introduce no
delay (phase shift) that the circuitry
error due to waveshape. However, in
imposes on different harmonics. A
the case of this group of instruments,
slight change of phase of a harmonic
can result in a dramatic change in component, it is important that this is
waveshape. For these reasons the treated in the same way as all the
amplifiers used in oscilloscopes are harmonics. In order to obtain an
usually directly coupled. This avoids appreciation of the changes produced
the use of coupling capacitors and in an complex wave due to variations
resistors (which can introduce time in the amplitude and phase of
delays and phase shift) and also harmonics, it is suggested that
allows the amplification of signals experimentation with a waveform
down to 0 Hz. Since complex synthesiser and oscilloscope be
waveforms contains a.d.c undertaken.

7.7 Measurement of Phase and Frequency____________________________________


This two characteristics of performed with an oscilloscope, as
alternating quantities may be will now be describe.
measured by meters designed The measurement of the phase
specifically for this, i.e. a relationship between two waveforms
phasemeter and a frequency (of the same frequency) may be
meter/counter. These instruments are carried out with a CRO using either
not always readily available, and one of two methods.
these measurements are more often
■ Dual trace-method ■

The two waveform are displayed on setting. Since the periodic time c time
the screen, and by means of the ‘Y’- T of a waveform corresponds to 360o,
shift controls the two traces are then the phase angle ϕ May be
aligned along a common horizontal determined from :
axis, as shown in Fig. 7.7. The time ϕ 𝑡
=
interval T and t are then measured, 360 𝑇

using the graticule and timebase


ϕ 𝑡
=
360 𝑇

Fig 7.7

■ Lissajous Figures ■

In this method the CRO timebase is


turned off, and one of the two
waveforms is connected to the ‘X’
input. The second waveform is
connected to ‘Y’-amplifier input as
normal. Depending upon phase angle
between the two waveforms the
resulting display will be either a
Fig 7.8
straight line, a circle, or an ellipse.
Examples are shown in Fig. 7.8, where
(a) Φ = 0o (in phase)
the phase relationships are :
(b) Φ = 90o
(c) Φ = 180o (in antiphase)
(d) 0o < ϕ < 90o
Where the display is an ellipse, the
phase angle may be calculated by
measuring the dimensions A and B, as
in Fig. 7.9, and applying the following:
𝐴
Sin ϕ =
𝐵

Fig 7.9

7.8 A.C. Bridges__________________________________________________________________________


An a.c. bridge utilities the same basic balance condition is more difficult
principles as those of the Wheatstone than for the simple Wheatstone bridge
bridge used for the d.c. measurement because the potential at B and D must
of resistance. The bridge supply, and be equal in both amplitude and phase.
the detector in the central limb must be There are a number of different forms
sensitive to a.c. The balance condition of a.c. bridge for the measurement of
occurs when the current through the inductors and capacitors, including the
detector is zero. The balance condition Owen, Hay’s, Maxwell, and Schering
may also be expressed in terms of the bridges. Each of these may be used to
ratio of the impedances of the four measure either inductance or
outer limbs, as below. The basic capacitance. More commonly,
arrangement is shown in Fig. 7.10 ‘Universal’ bridges that may be used
𝑍1 𝑍3 to measure inductance, capacitance,
=
𝑍2 𝑍4
resistance and Q-factor are employed.
One of the limbs will contain the It is suggested that the student carries
impedance to be measured (e.g. an out practical experience thus gained
inductor), and the other arms will will be far more instructive than a
contain variable reactance and
resistance elements. Obtaining the
description of the procedure given
here.

Fig 7.10

■ 8 Control Principles ■

This chapter builds upon and extends the basic principles of control systems that
were dealt with in Volume 1. On completion you should be able to :
1] Appreciate the different forms of system element, and analogous elements.
2] Identify a range of first-order systems, and describe their response to step
disturbance inputs.
3] Distinguish between first-order and second-order systems.
4] Describe the response of second-order systems to both step inputs and
forcing functions, particularly with reference to system damping.
5] Explain the effect of mechanical resonance, and methods of reducing its
affects.
6] Understand the concepts of proportional, differential and integral control
strategies.
8.1 System Elements and Analogies_____________________________________________
All types of system consist of elements kinetic energy (or its equivalent); the
or components that are interconnected others being associated with potential
in such a way that the ‘behaviour’ of energy (or its equivalent). In order to
each element has an affect on the put these concepts into perspective
‘behaviour’ of the whole system. consider Table 8.1, which shows the
Some system elements are capable of basic mechanical and electrical
storing energy. Of the energy storage elements classified into the three types
elements, some are associated with mentioned above.
Table 8.1 Energy storage and dissipative elements

System ‘KE’ storage ‘PE’ storage Dissipative


Mechanical 1 1 𝑑𝑠
mass ( 𝑚𝑣 2 ) spring ( 𝑘𝑠 2 ) damper (𝑅 )
2 2 𝑑𝑡

Electrical Inductor capacitor 𝑑𝑞


resistor (𝑅 )
𝑑𝑡
1 1
(2 𝐿𝐼 2 ) (2 𝐶𝑉 2 )

Considering Table 8.1, the following Both the spring and the capacitor
analogies can be made : store energy by virtue of position or
Both the mass and the inductor store potential—that is, potential energy.
energy in a form associated with For example, the further the spring is
movement. In the case of the mass, displaced the more energy it will
this will be kinetic energy by virtue of store. Similarly, the more charge
its velocity. For the inductor, the placed on the plates of a capacitor the
energy is stored in the magnetic field larger its p.d., hence more energy
produced by the current flowing. stored.
Remember, v = ds/dt metre per Both the damper and the resistor have
second; and I = dq/dt coulomb/second a ‘slowing down’ effect in their
(amp), i.e. the current is the rate at respective systems. As a result they
which charge is displaced. cause heat energy to be dissipated.
The analogies can now logically be current, I ≡ velocity, v
extended to the system variables and resistance, R ≡ damper coefficient, R
constants such that : inductance, L ≡ mass, m
voltage, V ≡ force, F capacitance, C ≡ reciprocal of spring
charge, Q ≡ displacement, s stiffness, 1/k

8.2 First-order Systems________________________________________________________________


A first-order system is one which Chapter 4. In this situation the
contains only one of the two possible systems (circuits) were subject to a
types of energy storage element, in step input, by connecting to or
combination with the associated disconnecting from a d.c. supply. In
energy dissipative element. For the each case it was found that the
elements so far considered, the system response followed an
following combinations form first- exponential law. Let us review the
order systems: salient points regarding the C-R and
Mass-damper; spring-damper; L-R series circuits, and see how these
inductor-resistor; and capacitor- results can be translated in terms of
resistor. The last two combinations the response of analogous
we have already studied in detail mechanical systems.
when considering d.c. transients in
■ C-R series circuits and spring-damper system ■
■ C-R series circuits and spring-damper system ■

Figure 8.1 shows a circuit that is charge, and current vary exponentially
subjected to a sudden step input with time until the capacitor reaches
disturbance when the switch is closed. its fully-charged state, such that :
In this case the capacitor will charge 𝑣𝐶 = E ( 1 - e−t/τ ) volt; i =
up until the voltage between its plates 𝐼𝑜 e−𝑡/𝜏 amp
is equal to the applied voltage. The q = Q (1 - e−t/τ ) coulomb; and 𝜏 =
relevant equations for this system are CR second
as follows.
E = 𝑣𝑅 + 𝑣𝐶 volt
𝑅𝑑𝑞 𝑞
since 𝑣𝑅 = 𝑅𝑖 = ; and 𝑣𝐶 =
𝑑𝑡 𝐶
𝑅𝑑𝑞 𝑞
then, E = + volt …………….[1]
𝑑𝑡 𝐶

the above equation is known as the


differential or system equation for the
C-R circuit we have seen that its
solution shows that the circuit p.d.s.,
Fig 8.1

You might also like