Professional Documents
Culture Documents
5991-8780EN Poroshell HILIC-OH5 Vitamins Application
5991-8780EN Poroshell HILIC-OH5 Vitamins Application
5991-8780EN Poroshell HILIC-OH5 Vitamins Application
Food Testing
Author Abstract
Anne Mack
Eight water-soluble vitamins are baseline separated in 4 minutes on an
Agilent Technologies, Inc.
Agilent InfinityLab Poroshell 120 HILIC-OH5 column. The column is a 2.1 × 100 mm
format with 2.7 µm superficially porous particles. An ammonium acetate and
acetonitrile gradient was used to accomplish the separation on an Agilent 1290
Infinity LC.
Introduction
Superficially porous particle LC
columns are a popular tool in liquid
chromatography. These columns
generate high efficiency at lower Thiamine, B1 Riboflavin, B2 Nicotinic acid, B3 Pyridoxine, B6
pressure compared to their totally porous
particle column counterparts1. This is
primarily due to a shorter mass transfer
distance and substantially narrower
particle size distribution of the particles
in the column. The higher efficiency can
be used to speed up analyses or improve
results by increasing resolution and Folic acid, B9 PABA, B10 Cyanocobalamin, B12
sensitivity.
To date, superficially porous
particles have primarily focused on
reversed‑phase separations. With the
maturation of superficially porous Ascorbic acid, C
particle technology, applications for Figure 1. Water-soluble vitamins.
further chemistries and chromatographic
techniques, such as hydrophilic
interaction liquid chromatography Table 1. Agilent 1290 Infinity LC System configuration.
to baseline resolve eight water‑soluble Agilent 1290 Infinity • Heat exchanger, low dispersion, 1.6 μL, double (G1316-60005)
• Heater & Column: Agilent InfinityLab Quick Connect assembly, 105 mm, 0.075 mm
vitamins. Figure 1 shows these Thermostatted Column
(5067-5961)
Compartment G1316C • Column & Flow cell: Capillary, stainless steel, 0.075 × 220 mm, SV/SLV (5067‑4784)
compounds.
Agilent 1290 Infinity Diode
Agilent Ultra-Low Dispersion Max-Light Cartridge flow cell, 10 mm (G4212-60038)
Experimental Array Detector (G4212A)
• G4220A: B.06.53 [0013]
Agilent OpenLAB CDS
• G4226A: A.06.50 [003]
An Agilent 1290 Infinity LC configured for ChemStationEdition
• G1316C: A.06.53 [002]
RevisionC.01.05 [35] • G4212A: B.06.53 [0013]
low dispersion was used for this work.
Table 1 shows the details. Table 2 shows Agilent LC column Agilent InfinityLab Poroshell 120 HILIC-OH5, 2.1 × 100 mm, 2.7 μm (685775-601)
2
The eight vitamins analyzed in this work Table 3. Sample preparation
were purchased as individual powders
Water-soluble vitamin (B,C) Concentration (mg/mL) Sample solvent
from Sigma-Aldrich. Ammonium
Thiamine (B1) 0.4 CH3CN/H2O (9:1)
acetate and acetic acid were also
Riboflavin (B2) 0.1 CH3CN/DMSO (9:1)
from Sigma‑Aldrich. Acetonitrile was
Niacin/nicotinic acid (B3) 0.4 CH3CN/H2O (9:1)
purchased from Honeywell (Burdick and Pyridoxine/pyridoxol(B6) 0.4 CH3CN/H2O (9:1)
Jackson). Water was 0.2 µm filtered, Folate/folic acid (B9) 0.4 CH3CN/DMSO (9:1)
18 molecular weight, from a Milli-Q 4-Aminobenzoic acid/PABA (B10) 0.2 CH3CN/H2O (95:5)
system (Millipore). Cyanocobalamin (B12) 0.4 H2O
Figure 3. Separation of water-soluble vitamins on alternative Agilent InfinityLab Poroshell 120 HILIC
phases.
3
Conclusion
The Agilent InfinityLab Poroshell 120
HILIC-OH5 column is well suited for the
separation of water-soluble vitamins.
This column offers good resolution
and peak shape for all compounds.
Other HILIC chemistries were also
evaluated for this separation. However,
the resolution on other phases was not
adequate, and would require further
method development to assess optimal
conditions for a suitable separation.
References
1. Gratzfield-Hugsen, A.; Naegele,
E. Maximizing efficiency using
Agilent InfinityLab Poroshell 120
Columns, Agilent Technologies
Application Note, publication number
5990‑5602EN, 2016.
www.agilent.com/chem