Tank consumption with Qbox system

Last update : 2011-05-06

Estimated tank consumption to perform calibrations and tests with Qbox System.

Number of Calibration / Test with


Estimated gas amount (ml)
a Standard Tank1

Polytropic factor calibration 90 ≈6220

Leakage Check 380 ≈1473

Test2 500 ≈1120

Standard Daily routine3 8000 70gg

Intensive Daily routine4 15560 36gg

1
Standard tank : 560 Litres
2
Standard condition: vent valve opened 2 times to allow thermal equilibrium
3
Standard Daily Routine: Polytropic Factor Calibration / Leakage Check / 15 Tests
4
Intensive Daily Routine: Polytropic Factor Calibration (x2) / Leakage Check / 30 Tests

\\Serverdati\Cosmed\R&D\Project management\Projects\Body Box\Techical Specs\Tank_consumption_with_Qbox.doc Last Update: May 6, 2011