Accurate Conductivity 7.2.8 PDF

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 17

How to Increase the Accuracy of Solution Conductivity Measurements

By Robert C. Hunt — Sensor Development, Inc. — © 1995

Measuring the electrical conductivity of aqueous solutions can provide useful information about the amount of material dissolved in solution, i.e.
chemical concentrations. The parameters of this widely used technique are straightforward, but do vary appreciably with the type of chemicals
present. An understanding of these variations will enable the conductivity instrument user to obtain greater accuracy and repeatability. This paper
will explain the basics of the measurement, discuss the variables, and then suggest methods for use and calibration that will enhance accuracy.

E
LECTRIC CURRENT will itive sodium ions will move toward its literal and figurative reciprocal,
readily flow through certain the negative voltage, the negative the mho. Europeans have recently
solids and liquids when a chloride ions will move toward the changed the name of the conductiv-
voltage difference exists between positive voltage, and current flow ity unit to the siemen. The mho
two points thereon or in. The will occur. However, the current and the siemen are identical, and
amount of current that will flow for carriers in this case have very differ- choice of usage depends on which
a given voltage is predicted by ent characteristics than the electron scientist you care to honor.
Ohm’s Law carriers in metal. Because commonly encountered
Compared to electrons, ions are dilute aqueous solutions have small
E = IR Eq.(1) huge, and much more limited in conductivities, the units are most
number per unit volume. Limita- often expressed as the millionth
Where E is the potential differ- tions on current flow will be less part, which is micromho or micro-
ence in volts, I is the current in due to the resistance of the conduct- siemen. A typical conductivity
amps, and R is the resistance to cur- ing medium than to the availability range for tap water, for instance, is
rent flow in ohms offered by the of ionic carriers. Since this current 20 to 1,000 micromhos. The latter
conductor. controlling factor is an enabler rath- figure approaches the edge of com-
In metals this current will consist er than an impedance, measure- fortable drinkability, as is the case
of free electrons moving through ments of solution current carrying in much of Southern California.
the atomic structure. These charge ability are described in terms of Sea water has a conductivity of
carriers have almost negligible mass conductivity, the opposite of resis- about 65,000 micromhos (or mi-
and move quite freely. Any limita- tivity crosiemens).
tions on current flow (electrical re- Since the primary limiting factor
sistance) are determined by the Conductivity = 1÷ Resistivity Eq.(2) in liquid conductivity is the number
characteristics of the conducting of current carriers (ions) that are
medium more so than by the car- The unit of resistance is the available, the technique immediate-
riers themselves. ohm, and the unit of conductivity is ly suggests itself as a means to deter-
Electric current will also pass
through certain liquids, via a differ-
ent mechanism. The molecular
structure of a liquid is not suitable (-) (+)
for free electron movement, and an-
other sort of charged particle must VOLTAGE SOURCE
serve this purpose if any current is
to flow at all. In solvents that will
support conduction, ionization pro-
vides the needed carriers.
When most inorganic com-
- +
pounds are dissolved in water, their - Na+
Cl- Na+ Cl- +
molecules separate into two equally - Na+ +
and oppositely charged parts called - Na+ Cl- +
- Na+ Cl- +
ions. Sodium chloride, NaCl, will - +
separate into sodium ions, Na+, and - Na+ Cl- Cl- +
- Na+ +
chloride ions, Cl-. When a voltage
is applied across a volume of water
containing dissolved NaCl, the pos- Fig.(1) Movement of ions in solution, under influence of applied voltage.
mine chemical concentration levels. When using conductivity for uid volume, and not the area of the
That capability in fact makes it one chemical analysis, the term of inter- electrodes, as is often supposed. It
of the most common analytical est is specific conductivity, which is is also common to see κ used as the
measurements encountered in water the inverse of specific resistivity, ρ, symbol for probe constant, prob-
treatment and chemical processes. and is designated as κ . It is there- ably because of the similarities be-
The oldest and still most com- fore necessary to relate specific con- tween Eq.(4) and Eq.(5).
monly used way to make a conduc- ductivity to measured conductivity If the electrodes are moved to
tivity measurement in liquid is to in a volume defined manner. This 10cm apart, the probe constant be-
immerse two conducting electrodes can be done by inverting Eq(3) and comes 10. For a constant of 0.1,
in solution, apply a known voltage stating it in terms of κ move them to 0.1cm apart. (The
across them, and measure the cur- same effect can be achieved by in-
rent which results. Since the object (1 / ρ) = κ = (1 / R) (L / A) Eq.(4) versely varying the area instead of
is to determine the number of ions the length.) It can be seen that the
per unit volume, the volume of The term (1 / R) is the measured probe constant is a multiplier figure
fluid being measured must be de- conductivity. To convert to specific that can be used to scale the rela-
fined in order to interpret the re- conductivity, the volume being tionship between measured conduc-
sults accurately. This differs tivity and the figure of actual
appreciably from the normal interest, specific conductivity.
usage of resistance measure- In instrument terms, the
ments. probe constant is a scale multi-
In the typical case, resis- plier.
tance is measured between An instrument set for a full
two points without regard scale range of 5,000 microm-
for either the shape or the hos using a probe constant of
volume specific resistivity of 1 will have a range of 50,000
the item being measured. 1cm micromhos when using a
This is because interest lies probe constant of 10, and 500
primarily in the item's total micromhos if the probe con-
effect on current flow in a stant is reduced to 0.1.
circuit, and not in determin- In practice, the separating
ing the electrical properties geometries are usually more
1
of its material of construc- 1cm complex than the defining ex-
tion. Another way of stating ample, since the closed cube is
this is that the interest is in not a practical sensor design,
the measured resistance, not Fig.(2) A probe constant of 1 is obtained when the and the electric field between
the specific resistivity. The fluid to be measured is contained entirely within a 1cm the electrodes will actively ex-
relationship between resis- cube with 1cm2 electrodes on two opposing sides. tend to all available fluid vol-
tance and specific resistivity umes between them. The rela-
is described by measured, L/A, must always be con- tionship in Eq.(2) still holds, but L
sidered. This leads to the concept and A become complex mathemati-
R = ρ (L / A) Eq.(3) of a probe constant or cell constant. cal expressions. For most if not all
A probe constant of 1 is defined as electrode arrangements, the formula
Where R is the measured resis- two electrodes, each 1cm , located for probe constant will be a scaled
2
tance in ohms, ρ is the volume spe- 1cm apart, with all current flow reciprocal of the formula for the ca-
cific resistivity in units of Ω-cm, contained within that 1cm3 of fluid pacitance between the electrodes.
and the expression in brackets is the between them as shown in Fig.(2).
defined volume of conducting ma- The probe constant is expressed by From Theory to Practice
terial. L is the distance between the
As with many measurements, the
electrodes in centimeters, and A is Probe Constant = θ = L / A Eq.(5)
the cross sectional area of the uni- idealized central premise is compro-
form conductive path separating mised by a number of side effects
Bear in mind that A is the cross- that must be taken into account. A
them in square centimeters. sectional area of the intervening liq- major factor is polarization.

Page 2
Ions travelling toward the
electrodes will quickly form CONDUCTIVITY INSTRUMENT REPONSES
dense clouds near the electrode
surfaces. They cannot physical- IDEAL
1000
ly cross the solid/liquid barrier,
and the voltage is usually too low
for significant electrolysis to oc- POLARIZED,

MEASURED CONDUCTIVITY
cur (forming elemental sodium 800 ADJUSTED
and chlorine gas for our exam-
ple). But the cloud of negative POLARIZED
ions will present a rising counter
emf at the positive electrode, and 600
vice-versa. The result is that cur-
rent flow rapidly and exponen-
tially drops to zero, invalidating 400
the measurement.
To overcome this, an alternat-
ing polarity voltage is applied.
This can be a sine wave, square 200
wave, trapezoid, or any number
of forms. The purpose of using
AC is to reverse the motion of
0
the ions before they build up
0 200 400 600 800 1000
near the electrode surfaces in any
quantity. ACTUAL CONDUCTIVITY
This simplified description
does not address lesser but still Fig.(3) This graph shows typical response for an instrument operating beyond its
serious problems with the reduced linear range and severely polarizing its electrodes. A normal reaction to this by
but not eliminated polarization users is to adjust for an accurate reading at some point near the center of range,
that are overcome with various in this case at 600 micromhos, and reduce the overall error somewhat. While the
electronic tricks. Frequency ma- response below 600 looks reasonable when graphed, it might be noted that the er-
ror at 200 is 10% of the reading. By standardizing at 700 instead, the error at both
nipulation, current sensing win- 200 and 1,000 would be about 15% of reading. In some cases, response actually
dows, peak detectors and other de- goes flat-line and there is no possibility of meaningful adjustment. The response
vices are used to make sure that labeled IDEAL is what might be expected if a black platinized sensor of the same
the current monitored is represen- probe constant were substituted for the plain metal sensor being polarized.
tative of the ion population and
not polarization.
The problem is not fully appre- that occurs. The technique is not unity constant ranges limited to
ciated by some instrument design- often used for on-line measure- 5,000 micromhos or less unless a
ers. It is an unfortunate fact that ments because the platinized coat- platinized sensor or a sensor with
circuits can readily be designed that ing is soft and easily removed by extremely large electrodes is used.
will work to extremely high conduc- abrasion in flowing streams of less Such limited ranges not only
tivity ranges with dummy resistive than ideal fluids. cause the user to have to go to high-
loads or idealized sensors. The lat- It is when a circuit is tied to er probe constants, with increased
ter term refers to sensors whose plain metal electrodes with limited expense and reduced reliability due
electrode surfaces have been coated surface area that its design meets a to small passageways, but they are
with black platinum. This spongy real and practical test. Well de- inordinately sensitive to electrode
coating increases the effective sur- signed instruments will give linear fouling in use. Partial occlusion of
face area of the electrode by a factor response with a 1 constant plain an electrode surface with stream
of hundreds or more, making it less metal electrode sensor up to 20,000 contaminants will produce a large
sensitive to the early stages of pola- to 100,000 micromhos. However, decline in the reading.
rization. It also catalytically speeds it is not uncommon to see some An instrument which will not
up the small amount of electrolysis popular instruments with linear take a plain metal unity probe con-

Page 3
stant sensor to at least 10,000 mi- have charges of 2 and 3 or even Ct = C25 (1 + α (t - 25)) Eq.(6)
cromhos in a linear manner will be higher, which in turn affects their
hard pressed to deliver repeatable speed in an applied electric field.
readings in on-line use. A further Chemical handbooks list mobili- Ct = conductivity at t° C
disadvantage of the limited linear ty ratings for various ions, and no C25 = conductivity at 25° C
ranging is that when temperature two are exactly the same. Further,
begins to rise, the range is reduced these mobility figures vary with t = operating temp. in ° C
even further because the uncorrect- both temperature and chemical α = temperature coefficient
ed conductivity level rises. concentration.
Non-linear response can be de- As the temperature goes up, mo-
tected by measuring standard solu- bility increases, and the conductivi- The temperature coefficient is
tions across the range of interest ty for a fixed concentration of the the amount of change in conductiv-
and graphing the results. An in- chemical rises. This is a large ity per °C , expressed as a decimal
creasing droop in response below change, on the order of 2% of the part of the conductivity at some ref-
the expected value of the solutions 25°C conductivity value for each °C erence temperature. In this case, as
as the conductivity increases is an of temperature change. Automatic in most modern instruments, the
indication that polarization is creep- compensation for this effect is a def- reference temperature is 25°C. A
ing into the reading. inite requirement for accuracy. typical value for α is 0.02, although
Another way to verify this is to it can vary from that figure by a fac-
obtain a platinized laboratory sensor Because the temperature effect is tor of 2 or more in either direction
with the same probe constant as the close to linear for some common so- and is actually a function of both
on-line sensor, and take measure- lutions, the following equation is temperature and chemical concen-
ments with it in the same solutions, used to describe it. Keep in mind, tration.
connected to the instrument in however, that the real case is non- As chemical concentration rises,
place of the on-line sensor. A linear linear, so this equation is just an ap- the ions become so numerous as to
response with the platinized sensor proximation. impede one another’s movements
absolutely confirms the sick instru- through the solution. This is analo-
ment diagnosis.

Measuring Concentration Conductivity vs. Concentration

900000
While the current that develops
from the applied voltage is propor-
800000
tional to the number of ions in so-
lution, the proportion varies for
700000
each kind of ion. NaCl
Ions are massive when compared
600000 KCl
to electrons, and thus have to bull
Micromhos

their way through all the other large


500000 NaOH
molecules in solution instead of flit-
ting from atom to atom at near the HCl
speed of light. 400000
Ions are hydrated, meaning that H2SO4
they have several polarized water 300000
molecules stuck to them which have HNO3
to be dragged along for the ride. 200000
The technical names for the solu-
tion phenomena which retard 100000
movement of the ions are electropho-
resis and assymetry effect. 0
0

10

20

30

40

50

In addition, electrons have a


charge of 1, while ions frequently
% by Weight
Page 4 Fig.(4)
gous to an attempt to run through a changing, or the solute of interest is linearity and exact temperature
crowd of people. The more of much higher in conductivity than compensation are less important,
them, the slower your speed, re- any others present. and distinctions between types of
gardless of the attraction coming chemicals is unimportant.
from the destination. (Stated scien- In solutions containing multiple
tifically, electrophoresis and assyme- chemicals (assuming no reactions
try effect are more pronounced at between them) the total conductivi- Neutralization Indicator
higher concentrations.) The result ty is roughly equal to the sums of
is that conductivity begins to in- the individual conductivities. This Though not often used for this
crease at a slower percentage rate rule of thumb is accurate for dilute purpose, conductivity makes an ex-
than the increase in chemical con- solutions, and increasingly less so as cellent linear indicator of the
centration. At some point, it may concentrations rise. progress of a batch neutralization.
reach a maximum and then actually Even at total concentrations of Whenever a strong acid is neutral-
decrease for further increases in 1% it is readily observable that the ized with a strong base, or vice-
concentration. Refer to the graph whole of conductivity is less than its versa, somewhat less highly conduc-
in Fig.(4) for examples. parts, and at 20% the difference is tive salts are produced as the highly
So far only the single solute case pronounced. These figures are for conductive main products react.
has been addressed. Conductivity is example only. There is no fixed The result is a linear decrease in
highly sensitive to ions in solution, starting point or relationship for conductivity toward a knee, where
but it is not selective. If more than which it can be said that the sum is it will start to rise again as the neu-
one chemical is in solution, there is any particular amount less than its tralizing chemical exceeds the con-
no way for a conductivity measure- component parts except for an ex- centration of the chemical being
ment to distinguish which of them plicitly defined mix. neutralized.
causes a change in reading, unless A drawback to this use is that
their contributions to same are vast- there is no way to tell whether the
ly different. Leak Detection & Water Quality reading is on the acid or base side of
A numerically good example of neutral. However, a backup indica-
this would be 5% acetic acid and Often it is necessary not to know tor of pH, which doesn’t work ac-
5% hydrochloric acid in the same exactly what is in solution, only curately in strong acids and bases,
solution. Conductivity would read- whether a certain level of ions from will serve to tell which side it is on
ily follow changes in the hydro- any source has been reached (single and thus enable the accuracy which
chloric levels, but would not see the point control). This is true in cool- the conductivity reading can give.
acetic at all. This is because 5% hy- ing towers and boilers, where the re-
drochloric reads about 430,000 mi- circulating water must be dumped Accuracy of Conductivity Measure-
cromhos while 5% acetic acid is a when its salt level increases due to ments
little less than 1,800 micromhos. evaporation or steam loss. Too
(Acetic is a weak acid and does not high a dissolved solids level will Major items affecting the accura-
ionize fully. cause scaling, foaming, heat transfer cy and repeatability of conductivity
The bottom lines are that in or- inefficiency, and corrosion. measurements are:
der to use conductivity to accurately Many processes use heat ex-
measure chemical concentration, changers to vary the temperature of (1) Electronic stability and linearity
several things are needed: strong, and often dangerous chemi- of the instrument.
cals. They need to know immedi- (2) Linearity of instrument/sensor
• A chart showing the relationship ately if the low conductivity heating with solutions (polarization).
between concentration and conduc- or cooling fluid is being contami- (3) Accurate temperature measure-
tivity for the chemical of interest. nated with even small amounts of ment and correction for tempera-
process chemical, so that flow can ture effects in solution.
• Temperature compensation for be shut down before damage is (4) Knowledge of the relationship
the non-linear response of the spe- done. between conductivity and concen-
cific solution. In both these cases, measure- tration for the solution of interest.
ment inaccuracies of 5% or greater
• Either a single solute or the case can often be tolerated without The suggestions that follow as-
where only the solute of interest is problem. For these applications, sume a need for precision which

Page 5
justifies both the cost and time they Check for stability at both full has to deal with. For example, if
will require. They are not proposed scale and at zero deflection (open the range of interest is 0 to 5,000
as standard procedures for using sensor leads). There may have to be micromhos at 75°C, then extend
conductivity instruments, since a resistor in place of the tempera- the tests to 10,000 micromhos.
many applications do not demand ture compensation element if there Conductivity will increase about
absolute accuracy, and single point is no manual TC setting available. 2% per°C, so an increase of 50°C
control is not affected by non- Make sure that the resistor has a will double the nominal full scale
linearities away from the setpoint. low temperature coefficient value, value. Despite what the corrected
No pass or fail values are given for less than 50 ppm, or attach leads so reading appears as on the display,
the error sources described, since that it can be located outside the the instrument will be seeing dou-
that will be a function of the accu- box and not see the temperature ble that, or 10,000 micromhos, in
racy required by the individual user. swings. Carbon resistors with its input stages.
broad tolerances can have tempera-
Electronic Stability & Linearity ture coefficients as large as several Lead Wire Resistance
hundred ppm, well in excess of
While it would be heartening to what needs to be monitored in the Wires are not perfect conduc-
recommend a simple reference to instrument. tors. Even heavy gauge copper
the instrument manufacturer’s spec- Next, use precision resistors to wires offer some small amount of
ifications, this would fall short of check the linearity across the range resistance to the passage of electric
verifying the actual on-line charac- of interest at four or more points. current, as illustrated in the table
teristics. Practically all instrument To determine the required resis- below.
specs refer to electronic response to tance to achieve a given reading, di-
resistor simulations for both the vide the micromho value desired Wire Resistance Values
conductivity and temperature sen- into one million times the probe
sors. Because no polarization oc- constant. Thus, 1,000 micromhos Wire Gauge Ohms/1000 Ft.
curs with resistors, optimum spec requires a resistance of 1,000 ohms,
figures can be obtained from circuit 2,000 micromhos corresponds to 8 0.654
designs that are ill suited to working 500 ohms, 4,000 micromhos to 10 1.018
with the solid/solution interface 250 ohms, and so on. An instru- 12 1.619
problems. ment that will not achieve the re- 14 2.575
However, it is advisable to look quired accuracy and repeatability 16 4.094
at the published linearity tolerance, with resistors has no chance of im- 18 6.510
which should be no more than a provement with a sensor in solu- 20 10.35
few tenths of a percent across the tion. 24 16.46
range. Another important figure is 26 41.62
stability with changes in ambient 28 66.60
temperature. The instrument Linearity in Solution Measurements
should exhibit zero and span chang- From data published by the General
es of no more than 200ppm (of full After verifying that the basic Electric Company for solid round cop-
scale) from 0° to 60°C. If there are electronics are satisfactory, check per conductors.
no published ambient figures, it is for linearity with the sensor that
advisable to put a resistive load on will be used on-line. This can be Since conductivity measurement
the instrument to simulate full scale done in one of two ways — using is in essence also a resistance meas-
of the range of interest, and put it solutions of known conductivity urement, the series resistance of the
through a temperature change while value, or direct comparison with a lead wires between sensor and in-
carefully monitoring the output. A platinized sensor with an accurately strument will cause errors. At 100
simple way to get the high end tem- known probe constant similar to micromhos, with a 1 probe con-
perature condition is to place a the on-line sensor’s. stant, the solution resistance is
cardboard box over the instrument In either case, cover a conductiv- 10,000 ohms and lead resistance is
and a 100 watt standard light bulb ity range which reflects the temper- not a serious factor. But at 100,000
(lit, of course). Monitor the tem- ature effects on the solution con- micromhos the solution resistance
perature to make sure it does not ductivity, since this increases the drops to 10 ohms and now lead re-
exceed 60°C and cause damage. dynamic range that the instrument sistance can cause gross errors.
Page 6
This problem can be solved in If the lead length is doubled to temperature sensitive and automatic
several ways. Some instruments of- 200 feet and all other factors re- temperature correction is a source
fer automatic lead resistance com- main the same, the formula answer of considerable error if not done
pensation circuitry, typically using mandates 18 gauge wire in order to properly.
4 leads from sensor to instrument. keep lead resistance error below the Despite these drawbacks, howev-
In cases where the raw measure- stipulated 1%. In practice, since er, they can give good results if cer-
ment is being fed to a computer for careful calibration can split the er- tain precautions are taken. Make
scaling and temperature compensa- ror to both sides of ideal response, a sure that the instrument has been
tion, the lead resistance can be com- 2% figure for error could have been calibrated at full scale with a preci-
pensated for with a calculation. A used. sion resistor before doing these
very few microprocessor instru- tests.
ments have this last feature built Conductivity Standard Solutions (1) Allow the solutions to tempera-
into their programming. Inserting ture equilibrate in a location that is
a figure for lead length allows auto- Standard solutions present sever- as close to 25°C (77°F) as possible,
matic correction for same. al problems. First is simply obtain- in order to minimize automatic
In the majority of cases however, ing them. While available, they are temperature compensation errors.
the only available solution is to tail- less common than pH buffers and a It may be preferable to simulate
or either wire gauge size and/or lead ready source can be difficult to lo- 25°C with a resistor in place of the
length in a manner that keeps lead cate. automatic compensation element or
resistance below some predeter- They can be prepared with con- select manual compensation at
mined maximum allowable level. siderable accuracy using a precise 25°C, and monitor the solution
The calculations necessary to do “recipe”, reagent grade chemicals, temperature with a lab thermome-
this are straightforward but labori- distilled water, a scale accurate to ter. Reference to the solution
ous. For convenience, the equation within a thousandth of a gram or chemical’s temperature response
below will specify directly the re- better, and a good volumetric flask will then allow slight corrections to
quired wire gauge based on inputs of one or two liters. If no scale is be calculated and applied to the
of allowable error as percent of full available, precise solutions can be measured readings.
scale conductivity, probe constant, purchased from a lab chemical sup- Automatic compensation can be
maximum conductivity, maximum ply house and diluted using good used if enough time is taken to in-
temperature, and lead wire length. volumetric glassware. This discus- sure temperature equilibration, and
sion concludes with a listing of in- the proper α value is selected. If α
for the standard solution is not
Wire Gauge = 9.943 Log10 [ 5x108θ %E / d (κ25 (1 + α (t - 25)))(100 - %E )] + 9.9 known, it can be determined by
Eq.(9) heating the solution slightly and ad-
%E = Allowable error in reading, expressed as % of full scale conductivity justing the instrument’s α setting,
assuming that it has one. If it does
d = Distance in feet between instrument and sensor (formula takes into account 2 leads)
not, then use manual simulation or
κ25 = Full scale conductivity value at 25°C and θ = Probe Constant consider using a precision lab sensor
comparison instead (next section).
t = Maximum operating temperature, °C and α = Temperature coefficient, decimal
(2) Metal sensors go through a pas-
sivation stage in solution, as various
oxides form on their electrode sur-
Cable Resistance Example formation that will allow accurate faces. Before making measure-
solutions to be prepared from a ments, clean the electrode surfaces
Assume the application calls for wide choice of chemicals. well with strong detergent or de-
a full scale of 2000 micromhos, Because they are simple salt solu- greasing agent, rinse extremely well
max. temperature of 75°C, a 100 tions, conductivity standards do not with running water, then soak the
foot cable run, and an allowable er- have the stability of pH buffers and sensor for at least 24 hours in a so-
ror of 1% of full scale. Not know- are easily contaminated, particularly lution whose conductivity is near
ing α, assume it to be 0.02. Plug- at low values. the middle of the range of interest.
ging these values into Eq.(9) tells us Last, and far from least, they, The apparent probe constant of
that 21 gauge lead wire is needed. like all conductive solutions, are a plain metal electrode sensor will
Page 7
increase slightly during the first 24 tainer to see what effect positioning from time to time before taking
hours in solution as the oxides are has on the reading, and locate it in readings. (This also turns out to be
formed. Detergents and degreasers a region where several inches of a common and little recognized
have high conductivities, and small movement in any direction has no problem on-line.)
amounts may be carried over in cre- observable effect on the reading. In
vices of the sensor body even with some cases, this may require a large (7) Never mix used conductivity
good rinsing. If this or their chemi- container and a large amount of standard solutions with the remain-
cal constituents are problematic, fluid, as well as a holding device to ing unused portions. Discard the
then clean the surfaces by wiping suspend the sensor in a fixed loca- solutions after use. With great care
with a high purity solvent such as their integrity can be maintained
MEK or trichlorethane. Spray for a time, but contamination
cans of circuit board cleaning sol- happens easily and there is no
vent work well for this purpose. warning flag to tell the user that
Whatever method is used, it is the standard is no longer accu-
most important to insure there is rate.
no grease on the electrode surfaces
from handling or production. Carefully plot the measured
values against the known solution
(3) Start with the lowest conduc- values to determine the amount
tivity solutions first to reduce con- of non-linearity. This should
tamination from carryover traces take the form of measured values
on the sensor. A small amount of that are increasingly lower than
strong solution can seriously con- the known at some point. Errors
taminate a weaker one, but a weak alternating to both high and low
solution carryover to a stronger sides indicate problems with the
one can do no more than dilute technique, and the test should be
the stronger with a nearly negligi- repeated with greater care.
ble amount of liquid. For best re- Since calibration at mid-scale
sults, pour some of the stronger will split the top end error to
solution over the sensor to rinse it Fig.(5) Precision laboratory sensors both sides of ideal (see Fig.(3)) a
before taking a measurement. with platinized electrodes. These were droop that is double the allowable
manufactured by the former Beckman error can be tolerated.
(4) Use a lab mixer to keep the so- Instruments and similar units are now
lution in constant motion. If not available from Rosemount Analytical,
located in Irvine, California.
available, then stir the solution well Comparisons With Precision Plati-
with the sensor until there is no ob- nized Sensors
servable difference between the mix- tion. It may, in fact, be impractical
ing and stationary modes. A gradu- to test certain sensor designs on the A much easier way to check on-
al change in reading with a non- bench if their electrode geometry is line linearity is by using a laborato-
stirred sample may indicate leaching too open (such as two pins project- ry type platinized sensor whose
of a previous solution from sensor ing from the end of an insulator, probe constant is similar to that of
crevices, lack of temperature equili- with no shroud around them). the on-line sensor. It is best to use
bration between sensor and sample, These sensors must be calibrated identical probe constants, but ac-
or air bubbles forming on electrode on-line in their actual working con- ceptable results can still be obtained
surfaces. figuration, and standard solutions with a platinized probe constant
will be difficult or impossible to ap- that is no more than ten times
(5) With some designs part of the ply. smaller or larger. Larger is prefer-
current flow will be outside the sen- able since it avoids the polarization
sor body and the conductivity read- (6) Don’t allow bubbles to form on threshold.
ing will fall as the sensor is moved the electrode surfaces or be trapped The great advantages here are
closer to the sides or bottom of a inside the sensor. Solution mixing that no temperature compensation
non-conducting fluid container. helps here, but it may be necessary is needed (and indeed should not
Move the sensor around in the con- to give the sensor a vigorous shake even be attempted) and there is no

Page 8
need for precision in making the quirement, but it does offer a solid corrected as an assumed part of the
test solutions. Simply add salt until base to determine the absolute val- solute TC with reasonable accuracy.
a 10% or so increase is noted, then ues of test solutions. Price of an in- The accurate correction for this
connect each sensor alternately to strument of this caliber is about portion that is needed below 10 mi-
the instrument and take readings $7,000. Sources are Rosemount, cromhos, where it becomes signifi-
from each with no more than 2 and Radiometer of Copenhagen. cant, utilizes a presumed value for
minutes between them to avoid The latter have offices in the United the probe constant in its calcula-
temperature drift. States. Yellow Springs Instruments tions, and a small error in in this
Since the intent is only to com- also offers lab instruments to go presumption will translate to much
pare responses, it is useful but not with their sensors. larger errors in the temperature cor-
necessary to know the constant of A lab instrument/sensor combi- rected reading.
the platinized sensor to a high de- nation of high accuracy offers the Knowing the probe constant ac-
gree of accuracy. ability to make precise adjustments curately is only useful if the instru-
While it is probable that a poor of on-line instruments from grab ment contains some provision for
instrument will produce non-linear samples. inputting that information into its
readings at some point even with a Some manufacturers are offering correction scheme. Some instru-
platinized sensor, such a device will plain metal electrode sensors with ments do not, and simply assume
give distinctly low comparison read- probe constants that are tested to a that an accurate probe constant is
ings with the plain metal electrodes, high degree of accuracy and marked being used.
possibly even getting into flat-line on the sensor. This offers potential Accurately standardizing the
response. problems because of polarization ef- reading to correct for probe con-
In other words, it will fail the fects. stant deviations will not alleviate
comparison test miserably, so the If the instrument and sensor are the temperature compensation error
fact that the referee readings are also being operated into a non-linear re- described.
non-linear will be unimportant. gion, then the highly accurate probe
A rough rule of thumb for the constant which was measured at a Temperature Compensation
platinized sensors would be to use lower conductivity value offers
0.1 constants to no more than nothing except a false sense of se- After assuring non-polarized op-
1,000 micromhos, 1 constants to a curity. This approach will be more eration, the greatest remaining
maximum of 10,000 micromhos, successful for solutions under 100 source of error in a conductivity
and so forth. Plot the comparative micromhos, where polarization ef- measurement is improper correc-
readings as outlined previously in fects are less prevalent because of tion for temperature effects. There
order to check for non-linearity the low ionic density. At higher are three distinct sources of this
(polarization droop). ranges, an “accurate” determination problem—incorrect readings from
One of the better sources for lab of the plain metal electrode cell the temperature element, a poor
type sensors with highly accurate constant is of dubious value. temperature correction scheme, and
constants is Rosemount Analytical, There is an area of application a lack of knowledge about the tem-
located in Irvine, California. They where knowing the probe constant perature characteristics of the solu-
acquired Beckman Instruments, to a high degree of accuracy is par- tion.
who had previously acquired Indus- ticularly important, and that is in
trial Instruments of Cedar Grove, high purity water solutions of less Temperature Measurement Accuracy
New Jersey. They offer a variety of than 10 micromhos. This is be-
platinized sensor types with probe cause temperature correction is Most auto TC elements are built
constants certifiable to 1/2% and 1/ large, non-linear and done in two into the conductivity measuring
4% accuracy. Prices range from parts — one for the solutes, and an- sensor. In addition to the basic tol-
about $300 to over $700, depend- other for the self-ionized portion of erance of the element, which can be
ing on style and accuracy. Yellow the water. surprisingly large in some cases,
Springs Instruments of Yellow The latter is always a part of any there is the problem of the effect of
Springs, Ohio is another source. conductivity reading, but since it is sensor mass on response to stream
For really top-drawer calibration, small (less than 1 micromho at temperature.
a research grade bridge or instru- 100°C) it can be ignored in the The element may be buried in
ment can be purchased to go with higher ranges. Even in the region the main body of the sensor, or
the accurate sensor. This isn’t a re- of 10 to 100 micromhos it can be contained in a thermowell project-

Page 9
ing into the stream. In either case, measured temperature droop will be impressive graph showing that α
the mass of the sensor will slow re- more linear. falls from 0 to 100°C, and further,
sponse to changes in stream temper- It also tends to be repeatable. that it does this in a distinctly non-
ature, since it is built of materials This allows the differential to be ef- linear manner. Their α values dif-
with low thermal conductivity fectively corrected by calibration, if fer by more than 0.014 at 100°C
(plastics, epoxy, stainless steels). If and only if the temperature adjust- which translates to a huge correc-
the stream being measured has rap- ment is a gain, or two-point adjust- tion error.
id swings in temperature, there is a ment, rather than a simple zero ad- Neither references a concentra-
good chance that the TC element is just. tion for which the values are sup-
reading incorrectly a majority of the Take readings from both the TC posed to apply, and neither explici-
time. element and the referee device at in- ty states the reference temperature
A plot of sensed temperature ver- creasing temperatures. If the diffe- for compensation. While most in-
sus actual would reveal that the ele- rential becomes pronounced, it may struments use 25°C as a reference,
ment sees each swing, but at a trun- be necessary to insulate the back of scientific work and published data
cated level out of time phase with the sensor, use a separate thermo-
frequently use both 18° and 20°C.
the fluid changes. The larger and well for the temperature element, or
more rapid the swings, the more se- select a sensor design with more As with the probe constant,
rious this problem. It is therefore a rapid temperature response. knowing precise values for α across
good idea to make tests on-line wide ranges of temperature and
with a rapid response temperature Correction Schemes and the Data concentration means nothing if
sensing device to see what the rela- That Enables Them there is no way to input that infor-
tionship is between sensor TC ele- mation to the instrument. While a
ment response and the actual Many, perhaps most, instru- number of instruments offer the
stream temperature. ments offer only a single value line- ability to dial in a value for α to be
The John Fluke Company, best ar correction that assumes α from used in a linear correction scheme,
known for its excellent line of digi- Eq.(6) to be exactly 0.02, or 2% per few can handle α values that are
tal voltmeters, also sells temperature °C, for a reference temperature of non-linear with temperature. Vari-
sensing modules which can be 25°C. Fortunately, this is nearly ations of α with concentration re-
plugged into any DVM to give a true for many common salt solu- quire sophisticated digital calculat-
readout in °C or °F. A variety of tions at ambient temperature levels. ing ability to perform iterative
sensors are offered which give both Unfortunately, it is exactly true for circular reference solutions for the
rapid and accurate readings. Cost none, and quickly loses validity for temperature correction.
of the Fluke Model 80TK thermo- wide swings in concentration and/
couple module is in the neighbor- or temperature. An error of only a Limited Range = Higher Accuracy
hood of $100. half percent per °C in the α assump-
The heat sink influence of the tion will give a corrected reading in That being the case, high accura-
sensor can also be considerable, es- cy can only be obtained in instances
error by 25% at 75°C. Tempera-
pecially for units that are inserted of limited ranges of temperature
ture correction coefficients vary and concentration changes, or by
into a pipe or the wall of a vessel. from a half percent per °C to almost
Since the back part of the sensor is characterizing the temperature re-
4%, and very pure water approaches sponse of a specific solution in great
at a different temperature than the 10%.
front, heat is constantly being detail and doing the correction in a
Reference literature on α values computer.
pumped in or out of the tempera-
ture element through its leads and is both scarce and, in many cases, It should be noted that a small
through contact with other parts of suspect. In reviewing literature for but definite part of the temperature
the sensor. this paper, two currently published effect is due to thermal expansion
The differential this creates be- data sources for NaCl temperature and contraction of sensor parts,
tween actual stream temperature coefficients were found that offer causing a small but very real change
and measured will increase as solu- graphs of the change in the α value in the probe constant. Any charac-
tion temperature increases. Its pat- across the range of 0 to 100°C. terization of a particular fluid
tern is similar to the graph shown in One shows α increasing from should also include the sensor and
Fig.(3) which illustrates polarized 0°C to 100°C in a perfectly linear instrument so that all appropriate
electrode response, except that the manner. The other gives an equally factors enter into the picture.

Page 10
A good method of obtaining ac-
curate temperature compensation is HCl Conductivity Temperature Coefficients,
to measure and plot the tempera- per NBS # NSRDS-NBS33, 1970
ture responses of the actual solution
to be measured over the expected 0.0155
ranges of temperatures and concen-
trations. Once this is done, it is rel-
atively simple to obtain linear as- 0.0150

Temperature Coefficient
sumptions for α, or to derive a 0.001M
polynomial function for it which
can be used in a computer correc- 0.0145 0.01M
tion scheme.
0.1M
On-site temperature characteriz-
ing is advisable even for single 0.0140 1M
chemical solutions because of the
data sourcing problems. For multi- 3M
chemical solutions this is absolutely 0.0135
necessary, since there will be about
zero chance of finding relevant pub-
lished data. 0.0130
If the sensor plumbing allows, 0

10

20

30

40

50

60

70
there is a simple way to do this.
Valve isolate the sensor and moni-
tor the temperature of the trapped °C Fig.(6)
sample accurately. This can be
done with either a separate tempera-
ture measurement, or by monitor- terest are needed to calculate α, but sure that it is not undergoing some
ing the temperature element in the it’s a good idea to take several as a reactive change during the test
sensor. self-check. Even with care it is easy which will invalidate the conductiv-
If the trapped sample is already to get slightly erroneous readings, ity data. Tests should always be
hot, allow it to cool slowly and especially if the temperature chang- done as quickly as possible to mini-
record the temperature/ es rapidly and the monitoring ele- mize changes due to chemical reac-
conductivity reading pairs at inter- ment has any time lag in its re- tion, evaporation and the like.
vals. If the line is at ambient tem- sponse. Temperature elements
perature, obtain a heating element built into sensors often show a lag The following formula calculates
that can be placed in contact with because of the considerable mass an assumed linear α:
the piping (wrap-around strips they are buried in.
work nicely) and heat the sample Bench checks with low conduc- α = (Ct1 - Ct2 ) / [Ct2 (t1 - 25) - Ct1 (t2 - 25)]
from the outside, taking care not to tivity solutions are difficult because Eq.(7)
build up excessive pressure in the of contamination by carbon dioxide
trapped fluid or melt plastic piping. from the atmosphere. Absolutely Where t1 and t2 are two different
The same sort of manipulation can pure water has a 25°C conductivity solution temperatures and C is the
be done to the cold side with value of 0.054 micromhos, but will measured conductivity value at the
strapped-on ice packs. rise to a little more than 2 microm- temperature indicated by the sub-
In the event that the sensor is hos within minutes of exposure to script. For using the formula, it
not accessible enough for this tech- atmosphere. Bench testing is not does not matter which of the two
nique, the tests will have to be done advisable for solutions under 20 mi- temperatures is higher.
on the bench. In either case, make cromhos unless they are known to
sure that automatic temperature Because α is likely to be non-
already be carbon dioxide saturated
compensation is disabled, and that linear, greater accuracy will be ob-
by exposure to atmosphere in the
25°C is being simulated. tained if one of the temperature val-
circulating system.
In theory only two points brack- Attention should also be given to ues is close to 25°C and the other is
eting the temperature range of in- the chemistry of the solution to in- near the expected center of the op-
erating temperature range.
Page 11
If computer correction is possi- Edition, page D-165), HCl conduc-
ble, then α should be calculated at tivity values at that concentration Despite the fact that approxima-
numerous points across the entire are: tions are still being used, tempera-
temperature range. The results can 0.5M, 25°C = 180,350 µmhos ture compensation has been re-
be entered into a number of com- duced to a negligible error
mercially available personal com- 0.5M, 55°C = 254,000 µmhos contributor. It will remain negligi-
puter programs to obtain a curve fit ble only if the temperature and con-
equation, most likely a polynomial The referenced listing gives what centration excursions remain limit-
of the 2nd to 5th order. is called equivalent conductivity val- ed.
The graph in Fig.(6) is for hy- ues, which allow more ready com- It is a straightforward exercise to
drochloric acid, using data pub- parison of conductance efficiency continue the error calculations
lished by the National Bureau of between various chemicals. To con- across wider ranges of concentration
Standards, applied to Eq.(7). A vert to specific conductivity, multi- and temperature to see how broad a
concentration of 1 molar HCl is ap- ply by 1000 times the molarity. range can be covered while still
proximately 3.65% by weight. Solving Eq.(6) for the corrected maintaining the required accuracy.
This level of detail about the be- conductivity value, C25 , gives: If multiple chemical process so-
havior of the temperature coeffi- lutions are reasonably consistent in
cient allows accurate temperature makeup, then an α analysis can be
corrections to be made, and thus ac- C25= Ct / (1 + α (t - 25)) Eq.(8) done in the same manner as shown
curate conductivity readings to be for HCl.
obtained, even in the absence of If this correction scheme is ap-
non-linear correction capability. plied to the example with the “stan- Summary of Conductivity Measure-
dard” α value of 0.02, the instru- ment Error Reduction Techniques
An example will help to illustrate ment reading would be:
the degree of improvement offered ✔ Do a one-time test to make sure
by this attention to detail. C25= 254,000 / (1 + 0.02 (55 - 25)) the instrument selected is capa-
ble of meeting the accuracies be-
ing sought. Verify the electronic
HCl Auto TC Example C25= 158,750 µmhos stability and linearity of the in-
strument with resistive loads
This result is in error by -12%. over the ambient temperature
Chemical: Hydrochloric acid, HCl range of 0-60°C ( or whatever it
is likely to see in service). If the
Concentration: 10,000-20,000 ppm Now make the same calculation, instrument will not perform
(1% - 2% by Weight) first referring to α values in Fig.(6). properly with resistors, attempts
The traces for 0.1M and 1M are to achieve on-line accuracy will
Temperature Range: 50-60°C parallel to one another in the region be a total waste of time and ef-
of 50 to 60°C. Since 0.5M is fort. A bad instrument can be
Instrument Auto TC: Accepts roughly half-way between them, a adjusted for accuracy at one
dialed in values for α for use in a line drawn there gives an α range of point only, and will not main-
linear correction scheme. 0.01385 to 0.01355. The middle tain that for long, nor give ac-
value of this range is 0.0137. Using ceptable accuracy even over lim-
For calculation, assume a con- this α value in the correction gives: ited range excursions. Systems
centration of 18,100 ppm, which is, operating in polarization can of-
conveniently enough, a 0.5 molar C25 = 254,000 / (1 + 0.0137 (55 - 25)) fer only single point control, not
solution. Temperature is 55°C. analytical accuracy.
C25 = 180,014 µmhos
Per the data in NBS publication ✔ Determine the linear measure-
#NSRDS-NBS 33, Electrolytic Con- ment range of the instrument/
ductance and the Conductances of the This result is in error by only sensor combination by measure-
Halogen Acids in Water, page 24, two-tenths of a percent to the low ments of known value solutions
(also published in CRC Handbook side (-0.2%), a sixty-fold improve- over the range of interest, or by
of Chemistry and Physics, 67th ment. comparison with platinized sen-
Page 12
sor responses with the instru- ductivity reading. If it jumps racy of the lab reading. If the lab
ment over the same range. suddenly, odds are good that the test is done at some temperature
rap dislodged bubbles from the other than 25°C or the on-line
✔ If the purpose of the measure- electrode surfaces. temperature, then correction ef-
ment is to determine chemical fects can enter because of the
concentrations, make certain ✔ Air entrapment is often an unrec- non-linear nature of the temper-
that accurate data is used to ognized problem when using lab ature coefficient, α. The best
make the correlation to conduc- type platinized sensors for com- way to do grab sample checks is
tivity. Although some of the parison or calibration. Air per- with a portable instrument, a
published data in this area was meates the spongy coating and is constantly flowing sample
obtained with great care and pre- usually not detectable by visual straight from the process line
cision, there is no guarantee of inspection. It is best to presoak next to the on-line sensor, and
that, and it almost never states the platinized sensor for a at least real-time access to the on-line in-
the possible error spread. In the 24 hours in the fluid that will strument’s reading. For samples
case of multiple chemical solu- serve as the test solution. Run below 20 micromhos this tech-
tions, on-site measurement often the conductivity up by adding nique is mandatory if large errors
will be the only source of data. salt and stirring, but do not re- are to be avoided.
move the platinized sensor even
✔ Characterize the temperature re- momentarily during the test run. ✔ Limit the ranges of temperature
sponse of the measured fluid Offsets as large as 5% can occur, and conductivity as much as pos-
over the entire ranges of interest especially in low constant sensors sible. Smaller dynamic ranges
of concentration and tempera- where the distance between elec- limited to the areas of true inter-
ture. Calculate α values using trodes is short and surface occlu- est allow much more precise cali-
Eq.(7) and plot them for refer- sion therefore directly affects the bration. Large errors often result
ence when calibrating the instru- effective volume between elec- from needless attempts to align
ment for service over a specified trodes. Heating the sample to for maximum accuracy at all
chemical and temperature range. start will drive off much of the points on the scale when in prac-
gas, but you must then allow it tice only a small percentage of
✔ Make sure the sensor does not to cool to avoid temperature gra- the scale will be used.
trap or collect air bubbles on its dients during comparison tests.
surfaces when on-line. This is a To verify results, take them sev- ✔ If the process coats surfaces with
common and little appreciated eral times over a period of several any sort of material, sensor
problem because you cannot see hours. Continue until variations cleaning will have to be done at
through most piping. To test are less than ±0.5%. regular intervals. Coatings typi-
for surface bubbles, rap the back cally cause the reading to fall off
of an on-line sensor sharply with ✔ When calibrating by collecting with time.
a screwdriver handle and look grab samples, be aware of the ef-
for a step change up in the con- fects of temperature on the accu- ❄

© 1995 — Sensor Development, Inc.

All rights reserved. This publication is fully protected by copyright and nothing that appears in it may
be reprinted for commercial purposes, either wholly or in part, without special permission. Duplication
in any degree for personal use is fully permitted.

Page 13
APPENDIX 1 - CONDUCTIVITY OF STANDARD SOLUTIONS
The data listing which follows was taken from the CRC Handbook of Chemistry and Physics, 31st Edition, page 1994. The listing for
sodium chloride is particularly useful, as it is a saturated solution (add salts until excess crystals remain at all times) and thus offers a very accurate
standard which can be prepared with no equipment save reagent grade NaCl, deionized water and a beaker. Other chemicals are listed, and will
be added to in succeeding appendices to offer alternatives whenever a particular ionic presence cannot be tolerated by the on-line system, and thus
even the small carryover from testing an on-line sensor must be avoided. An example of this would be high purity solutions in the semiconductor in-
dustry, where both sodium and chlorides are objectionable.

Conductivity values are given in micromhos. Solutions are as follows —


H2SO4 — maximum conductivity (18°C); dissolve 378 g. of 97% acid in pure water and dilute to 1 liter. Density at 18°C = 1.223.

MgSO4 — maximum conductivity (18°C); dissolve in 1 liter of distilled water 552 g. of MgSO4⋅7H2O. Density at 18°C = 1.190.
NaCl — solution saturated at all temperatures given. An excess of NaCl in distilled water, about 450 g. per liter. D18 = 1.2014.

KCl — normal solution, 74.59 g. per liter of solution at 18°C. Dissolve 74.555 g. (weighed in air) KCl, dilute to 1 liter. D18 = 1.04492.
1N KCl = 71.3828 g. KCl per kg. solution — 0.1N KCl = 7.43344 g. KCl per kg. solution — 0.01N KCl = 0.746558 g. KCl per kg.

TEMP., °C H2SO4 MgSO4 NaCl KCl, 1N KCl, 0.1N KCl, 0.01N

0 518,400 28,770 134,500 65,410 7,150 776


5 579,200 34,020 155,500 74,140 8,220 896
10 640,800 39,630 177,900 83,190 9,330 1,020
15 702,800 45,550 201,400 92,520 10,480 1,147
16 715,100 46,760 206,200 94,410 10,720 1,173
17 727,500 47,990 211,100 96,310 10,950 1,199
18 739,800 49,220 216,000 98,220 11,190 1,225
19 752,200 50,460 220,900 100,140 11,430 1,251
20 764,500 51,710 225,900 102,070 11,670 1,278
21 776,800 52,970 230,900 104,000 11,910 1,305
22 789,000 54,240 236,000 105,940 12,150 1,332
23 801,300 55,510 241,100 107,890 12,390 1,359
24 813,500 56,790 246,200 109,840 12,640 1,386
25 825,700 58,080 251,300 111,800 12,880 1,413
26 837,800 59,370 256,500 113,770 13,130 ..........
27 849,900 60,670 261,600 115,740 13,370 ..........
28 862,000 61,970 266,900 ............. 13,620 ..........
29 874,000 63,280 272,100 ............. 13,870 ..........
30 886,000 64,590 277,400 ............. 14,120 ..........

Temperature Responses of Standard Solutions


0.0230
Temperature Coeffiecient

H2SO4
0.0220
0.0210 MgSO4
0.0200
0.0190 Saturated NaCl
0.0180
KCl, 1n
0.0170
0.0160 KCl, 0.1n
0.0150
KCl, 0.01n
0.0140
0 5 10 15 20 25 30

°C
Page 14
APPENDIX 2 - PREPARING ACCURATE LOW RANGE CONDUCTIVITY STANDARDS
The data listing which follows was taken from Handbook of Electrochemical Constants, Parsons; Academic Press/Butterworths, 1959. It
gives a formula and a listing of plug-in values for same which will allow accurate correlation between conductivity and low concentrations of a wide
variety of solutes. No guarantees of accuracy are available, but it appears from a number of actual trials that attention to detail will give results to
within 1% of the absolute. The formula is valid only for concentrations between 0.001 and 0.1 normal. Density correction is a minor factor for
most calculations, but a good source of density values can be found in current copies of the CRC Handbook of Chemistry and Physics, Section D,
“Concentrative Properties of Aqueous Solutions: Conversion Tables”. This section also contains a considerable listing of conductivity values, general-
ly but not entirely of high accuracy.

Conductivity at 25° C = 1,000 Λ N ( 1 - a √Λ + b Λ )


Λ = Equivalent Conductivity
N = Concentration as Normality
“a” and “b” are constants from the tables which follow
Normality = Molarity / Valence = (ppm) (Solution Density at 25° C) ÷ 1,000 (Equivalent Weight)

Chemical Eq.Wt. Λ, 25C a b Valid ppm Range

Page 15
Chemical Eq.Wt. Λ, 25C a b Valid ppm Range

Page 16
Chemical Eq.Wt. Λ, 25C a b Valid ppm Range

Page 17

You might also like