Professional Documents
Culture Documents
Maintenance Positioner
Maintenance Positioner
PREVENTIVE MAINTENANCE
2 PREVENTIVE MAINTENANCE
Daily inspection, periodic inspection, and periodic maintenance can
keep the performance of the positioner stable for a long period of time.
NOTE
The annual operating time of the 1-axis servo
positioner is assumed to be 3840 hours.
When using the positioner beyond this annual
operating time, correct the maintenance frequency
in this chapter, which is determined on assumption
that the positioner is used for 3860 hours per year,
and perform maintenance as appropriate.
-5-
2.PREVENTIVE MAINTENANCE MAINTENANCE B-82545EN/02
-6-
B-82545EN/02 MAINTENANCE 2.PREVENTIVE MAINTENANCE
-7-
2.PREVENTIVE MAINTENANCE MAINTENANCE B-82545EN/02
-8-
B-82545EN/02 MAINTENANCE 2.PREVENTIVE MAINTENANCE
You should have the following instruments and tools ready for
maintenance.
a) Measuring instruments
Instrument Condition Use
Dial gauge 1/100mm For positioning precision and
backlash measurement
Calipers 150mm
b) Tools
Phillips screwdrivers (large, medium, and small sizes)
Flat-blade screwdrivers (large, medium, and small sizes)
Box wrenches (M3 to M6)
Allen wrenches (M3 to M20)
Torque wrench See Table 2.6.
Long T wrenches (M5, M6 and M8)
Adjustable wrenches (medium and small sizes)
Pliers
Long-nose pliers
Cutting pliers
Both-ended wrench
Grease gun
C-ring pliers
Flashlight
Seal tape
Loctite No. 242, 262, 675
Sling
Gear puller
NOTE
Prepare torque wrenches that enable the torque
values indicated in Table 2.5 to be set.
NOTE
When applying LOCTITE to the important bolt
tightening points, make sure that it is applied to the
entire longitudinal portion in the engaging section of
the female threads. If it is applied to the male
threads, the bolts may be loosened because
sufficient effects cannot be obtained. Remove the
dust within the bolts and taps and wipe oil off the
engaging section. Make sure that there is no
solvent in the taps.
-9-
2.PREVENTIVE MAINTENANCE MAINTENANCE B-82545EN/02
Table 2.5
Bolt size Torque setting value
M3 1.3 Nm (13 kgfcm)
M4 4 Nm (41 kgfcm)
M5 9 Nm (92 kgfcm)
M6 15.6 Nm (159 kgfcm)
M8 37.2 Nm (380 kgfcm)
M10 73.5 Nm (750 kgfcm)
M12 128.4 Nm (1310 kgfcm)
M14 204.8 Nm (2090 kgfcm)
M16 318.5 Nm (3121 kgfcm)
- 10 -
B-82545EN/02 MAINTENANCE 3.PERIODIC MAINTENANCE
3 PERIODIC MAINTENANCE
- 11 -
3.PERIODIC MAINTENANCE MAINTENANCE B-82545EN/02
The grease for the reducers needs to be replaced every three years or
11,520 hours in the procedure shown below. For greasing points, see
Figs. 3.1 (a) and (b).
<1> Remove the seal bolts and plugs of the grease outlets shown in
Figs. 3.1(a) and (b).
<2> Attach the grease nipple supplied with the positioner.
<3> Apply the grease specified in Table 3.1(a) and (b) until the old
grease is replaced and the new grease is output from the grease
outlets. Use grease outlet 1 shown in Figs. 3.1 (a) and (b). Only
when grease outlet 1 cannot be used due to the installation state of
the jig, use grease outlet 2. At this time, confirm that the same
amount of the old grease as that of applied grease has been output
to prevent the grease bath from overflowing.
<4> After applying grease, release the residual pressure in the grease
bath as instructed by the procedure in Section 3.2.
CAUTION
When reusing a grease nipple once used, wind
sealing tape around the screw.
- 12 -
B-82545EN/02 MAINTENANCE 3.PERIODIC MAINTENANCE
Fig. 3.1 (a) Grease inlet and outlet ( 300kg load capacity)
Fig. 3.1 (b) Grease inlet and outlet ( 1000kg load capacity)
- 13 -
3.PERIODIC MAINTENANCE MAINTENANCE B-82545EN/02
CAUTION
If you grease incorrectly, the pressure in the grease
bath will increase, leading to a broken seal, which will
eventually cause grease leakage or malfunction.
When greasing, be sure to follow the cautions stated
below.
1 Before starting greasing, open the grease outlets
(remove bolts and the like from the grease outlets).
2 Using a manual greasing pump, grease gently and
slowly.
3 Avoid using a pneumatic pump driven from a factory
pneumatic line as much as possible.
Even when using a of Table 3.1 (a) and (b) less
during application of grease.
4 Be sure to use the specified grease. Otherwise,
damage to reducers or a similar abnormality may
occur.
5 After applying grease, release the residual pressure
within the grease bath as described in the
procedure in Section 3.2.
6 Wipe off any grease from the floor and positioner
completely, so no one will slip on it.
After applying grease, to release residual pressure from the grease bath,
perform a repetitive operation with a motion angle of at least 60 degrees
and OVR100% for 10 minutes or more while leaving the grease inlets
and outlets open.
Under the grease inlets and outlets, attach bags for collecting grease so
that grease does not spatter when it comes out of the inlets or outlets.
If the above operation cannot be performed due to the environment of
the positioner, prolong the operating time so that an equivalent
operation can be performed. (If only a motion angle of 30 degrees can
be set, perform an operation for 20 minutes or more.) After completion
of the operation, attach the plugs and seal bolts to the grease inlets and
outlets. When reusing the plugs and seal bolts, be sure to seal them with
seal tape.
- 14 -
B-82545EN/02 MAINTENANCE 3.PERIODIC MAINTENANCE
CAUTION
Be sure to turn on the power. If the battery is
replaced when the power is off, the current position
information is lost, so that mastering becomes
necessary.
- 15 -
3.PERIODIC MAINTENANCE MAINTENANCE B-82545EN/02
Cap
- 16 -