May 10, 1990
- Specification Sheets -
Figaro cas sensor TGS 800
FIGARO ENGINEERING IHC,1. Model : Figar
2. Structure and
Structure =
0 Gas Sensor TGS 800
Materials
Refer to Fig.1 in page 3
Part
Material
Standard
[Sensing element
Sintered Sn0z semiconductor
Refer to Article 4
Characteristics”
“Electric
Flame arrester
SUS 316, 100 mesh stainless
steel gauze
Double layer
Heater coil
Chrome alloy wire
# = 0.06
Resistance value of heater:
Ru = 38Q + 39
(measured at room temperature)
Lead wire Gold alloy wire
gp = 0.08
Housing Glass Fiber reinforced Approved by 94HB of UL standard
polyamide resin Durable temperature: Max 240°C
Pin Mi The pins can withstand a with-
Jdrawal force in exceed of 5kg.
in the pins direction
3. Circuit Condition
Item Standard Remarks
Heater power Pn = 65080 Typical value
consumption
Sensor power Ps S 15aW Ps = Vo® - Rs/(Rs#Ri)?
consumption
Heater voltage (Va) [Vu = 5.0V + 0.2V
A.C. or D.C.
Circuit voltage (Vc)|Ve S 24v Ps S 15a¥
A.C. or D.C.
Load Resistor (Ri) [variable Same as above4. Electric Characteristics
Ttem
Sensor Resistance
Test Condition
Basic measuring circuit is shown
in Fig. 4 in page 4
Test gas : H2 (purity 99%)
Test chamber capacity
: 10 or more per sensor
Standard atmospheric condition:
clean air
temperature: 20°C
relative humidit:
Before testing, 48 hours
preheating under standard
atmospheric condition is
necessary
Ro(Hz 1OppR
Standard
in clean air)
Rs = 10KQ ~ 130KQ
in He 10ppm)
Rs (Air)
= 0.2 ~ 0.6
Response time
‘After exposure to Hz 10ppm,
sensor resistance must reach
90% of the stable level in
He 10pm
Within 20 seconds
Tine for initial
stabilization
After switch on, sensor resistance
must reach 80% of the stable level
in clean air
Within 1 hour
Insulation Apply DC 100V between live and 5MQ or more
resistance dead metal part
Dielectric ‘Apply 500V for one minute between | No dielectric breakdown
strength live and dead metal part is caused
(Current leakage
under 1a)Fig. 1. Structure
Flame arrester
Lead wire
Sensing element
Heater coil
Housing
Pin
(a) Cut View
Flame arrester
(b) Bottom View The numbers in the figure
indicate the six pins.
Fig. 2. Dimensions
95¢—+
12 005
<. iL.1(4) 3(6)
gS
2(5)O>
st2joL7 __,
oO
. A(1) 6(3)
Fig. 3. Symbol of TGS 800 in
Electric Circuit
* 2, 5: Meater pin
1, 3: Connected electrode pair
4, 6 : Connected electrode pair
Fig. 4 Basic Measuring Circuit
# Standard Test Condition
Circuit voltage : Ve = 10.0V + 0.1 (A.C. or D.C.)
+
Heater voltage : Vw = 5.0V + 0.05V (A.C. or D.C.)
Load Resistor : R= 4.0KQ + 1%Temperature and Humidity Dependency
Temp. (C) —>(eee) ‘2u0p srg
ot oT it
= 7 soo ==
[ spexotiod F
ystra}oeI
ey) AITATITSU
35
008 SOL