A. INSTRUMENT SPECIFICATIONS
RECORD TIME: 25, 50, 100, 250, 500, 1000, 2000 milliseconds (user-selectable).
25 25
50 50
100 100
200 200
500 500
1000 1.000
2000 2000
SIGNAL ENHANCEMENT: Successive shots may be summed in digital memory, to aid detection of weak signals.
The maximum stack count (nlirober of summed shots) that can be displayed is 255.
STORE/RECALL: Up to 12 sets of Channel Settings (each set consists of gain and trace size settings for all 12
channels) may be
stored in and retrieved from nonvolatile memory
DISPLAY: Built-in daylight-visible CRT, 5 in. E12 cm] diagonal; wiggle-trace or vat-iable-area display
capability. 5 1 pixels horizontal by 256 pixels vertical.
HARD COPY: Built-in printer/plotter, produces fadeproof reproducible record 4 in. wide by
14.25 in. long 110 cm x 36 crii] Dn eleett-osensitive paper.
AMPLIFIER GAINT 0, 6, 12,, 18, 24, 30, 36, 42, 48, 541 60, 66 'db (user--selectable).
QUANTIZATION LEVEL: 95 millivolts. This is the smallest signal which can be added to memory after amplification.
FILTER: Lowcut, 40-hz corner frequency, butterworth filter (Optional higherfrequency filters
available). Damping factor is 0.5; rolloff is 12 db per octave.