The AirFlow HVS PM 2,5 particulate sampler allow the user to perform PM
2,5 particle fraction sampling in compliance with EN-12341:2014 regulation.
The system can be interfaced with a meteorological sensors to a complete
sampling data report. It is based on an advanced microprocessor control
allowing the user to a correct sampling performance, flow and
compensation controls, ambient pressure and temperature detection,
corrected volume at standard conditions and %RSD calculation. All data
referred to the sampling periods, are stored in the memory and available for
download trough an USB port to transfer on SDC 2000 software application.
Optionally the user can carry on a PUF cartridge after filter to collect PAH
and Dioxins in vapor fractions.
Technical features
• Volumetric flow control based on ISO-5167 automatically corrected to
standard and actual temperature and pressure
• Reference temperature programmable based on user application
• Three stages brushless motor with electronic control
• Weather-proof anticorrosive anodized aluminium
• Class of insulation IP 55
• Automatic flow compensation and pressure detection on filter load PUF
• Real time %RSD calculation for best practice sampling protocols
• Wind direction and velocity sensors (optionally) available to activate or
deactivate sampling based on wind source
• Programmable alarm settings on flow control by the user
• Three user pre set program for standard custom applications
• Delayed start and sampling time up to 9999 minutes
• Easy on user calibration with any digital calibrator available
• Available to use with the high volume 5 stages cascade impactors
Tipical applications
• PM2,5 high volume samplers complying with EN-12341:2014 reference
method.
• PM10 application at 500 liters per minute with additional nozzle plate kit
www.amsanalitica.com • info@amsanalitica.com
AirFlow HVS-PM2,5 Technical Specifications
Motor pump assembly Three stages brushless motor
Maximum noise (1130 l/min - 8 meters) 65 dBA - (45 dBA with noise reduction kit)
Barometric pressure sensor 700 - 1200 mbar Res.0,1 mbar accuracy 0,2%
Optional sensors available Wind velocity and direction (ultrasonic and mechanical)
How to order
600/AFPM2501K Air Flow Sampler PM2,5-HVS (EN-12341)