General information Sirona Dental Systems GmbH
Installation Requirements C8+ / C8+ Turn

General information
About this document
This document describes the installation requirements for
the C8+ / C8+ Turn / C8+ Turn treatment centers.
The following sets of Installation Instructions describe the
subsequent installation procedure:
• REF 59 42 912 (C8+)
• REF 59 78 197 (C8+ Turn)
You will also need the following drilling template in order to
fasten the treatment center securely to the floor:
• REF 59 66 911 (C8+, C8+ Turn)

kÉï=~ë=çÑW= NOKOMNO

Changes since the last version (09/2009):

Chapter or section, page


Does not apply to the following products:
C8+ Solo, C8+ Stand-Alone, C8+ Side delivery,
C8+ Rear delivery

60 28 000 D3417
2 D3417.021.04.10.02
Sirona Dental Systems GmbH Table of contents
Installation Requirements C8+ / C8+ Turn

Table of contents

båÖäáëÜ
1 Preparations....................................................................................................................... 5
1.1 Safety........................................................................................................... 6
1.2 Media quality................................................................................................ 7
1.3 Supply lines in the termination panel ........................................................... 8
1.4 Underfloor installation of SIVISION connections ......................................... 11
1.5 Mounting plates C8+ (not for C8+ Turn)....................................................... 12
1.6 Demonstration chair plate C8+ Turn ............................................................ 13

2 Dimensions, technical data ............................................................................................ 15


2.1 C8+ 1:20 ...................................................................................................... 16
2.2 C8+ Cart 1:20............................................................................................... 21
2.3 C8+: Technical data ..................................................................................... 23
2.4 C8+ Turn 1:20 .............................................................................................. 24
2.5 C8+ Turn Cart 1:20 ...................................................................................... 31
2.6 C8+ Turn: Technical data ............................................................................. 33

3 Electromagnetic compatibility ....................................................................................... 35


3.1 Accessories ................................................................................................. 36
3.2 Electromagnetic emission............................................................................ 37
3.3 Immunity to interference .............................................................................. 38
3.4 Working clearances ..................................................................................... 40

60 28 000 D3417
D3417.021.04.10.02 3
1 Preparations
`UH / C8+ Turn

1.1 Safety .................................................................................................................................... 6

1.2 Media quality ......................................................................................................................... 7

1.3 Supply lines in the termination panel ..................................................................................... 8

1.4 Underfloor installation of SIVISION connections ................................................................... 11

1.5 Mounting plates C8+ (not for C8+ Turn) ................................................................................ 12

1.6 Demonstration chair plate C8+ Turn ..................................................................................... 13

60 28 000 D3417
D3417.021.04.10.02 5
1 Preparations Sirona Dental Systems GmbH
Installation Requirements C8+ / C8+ Turn

1.1 Safety
ATTENTION ATTENTION
It is essential that you comply with the warning and safety The floor must be flat and level (DIN 18 202). A mounting
information contained in the Installationsvoraussetzungen. plate must be used for uneven floors (see Section 1.5,
All such information is highlighted by the signal words "Mounting plates C8+ (not for C8+ Turn)" on page 12).
NOTE, CAUTION or WARNING.

ATTENTION
The unit may only be operated by properly trained person-
nel. On-site installation must be performed in accordance
with our specifications.
In our capacity as a manufacturer of medical electrical
equipment, we only consider ourselves to be responsible for
the safety characteristics of the unit if we perform mainte-
nance and repair work on it ourselves or if this is performed
by centers that we have expressly authorized to do so; and if ATTENTION
defective components are replaced with original spare The floor must have a minimum loading capacity of 0.5N/
parts. cm².
When this work is carried out, we recommend that you ask
those responsible for the work to provide a certificate indi-
ATTENTION
cating the nature and scope of the work, including details of
changes to rated data or the working area where applicable. Wireless phone interference with medical electrical equip-
This certificate should also be signed and dated, and should ment:
show information about the relevant company. To ensure safe operation of medical electrical equipment,
the use of mobile wireless phones in practice or hospital en-
vironments is prohibited.
ATTENTION
By law, any changes to this unit which may compromise the
ATTENTION
safety of operators, patients or third parties are not permis-
sible. Electromagnetic compatibility: The unit should not be op-
For reasons of product safety, this product may only be op- erated in the immediate vicinity of other devices. If this
erated with original Sirona accessories or accessories from proves to be unavoidable, the unit should be monitored to
third parties approved by Sirona. The user assumes the risk ensure that it is operating properly.
of using non-approved accessories.
If any devices not approved by Sirona are connected, they
must comply with the applicable standards, e.g.:

– IEC 60950 for information technology equipment


(e.g. PCs) and
– IEC 60601-1 for medical equipment.

The monitor used at the treatment center must meet the re-
quirements of IEC 60950. The loudspeaker port of the mon-
itor may only be connected to a device that complies with
IEC 60950 (e.g. a PC) or IEC 60601-1. Under no circum-
stances may it be connected to a stereo system, for exam-
ple.
If a system is being assembled during installation, the re-
quirements of IEC 60601-1 (3rd edition) must be adhered
to. The person assembling the system assumes responsi-
bility for ensuring that it conforms to Directive 93/42/EEC.

60 28 000 D3417
6 D3417.021.04.10.02
Sirona Dental Systems GmbH 1 Preparations
Installation Requirements C8+ / C8+ Turn

,
1.2 Media quality

båÖäáëÜ
Suction machine

N
oil-free
Air

>80 µm

Cold water
(drinking water quality)
Dilution hardness
1 bar = 100 kPa = 14.5 psi

Water quality – Avoid long stub lines to the treatment center.


Lime deposits and corrosion residues in tap water can lead – Carry out the installation so that, where possible,
to the following malfunctions: other main consumers (such as the sink) are fed
from the same line downstream of the treatment
• Premature clogging of the filters in the unit
center connection.
• Rapid clogging of the fine water paths and jets in the
– Avoid laying the supply line parallel to hot water
treatment instruments
pipes.
For these reasons, the following points must be observed:
• Observe EN 1717 for the protection of public drinking
• Permitted water pressure: 2.5 bar (36.25 psi) to 6 bar water.
(87 psi) Provided a disinfection system is also installed, the
• Permitted minimum flow volume: 3 l/min treatment center fulfills the requirements of EN 1717
• For water hardness (total hardness) of 2.2 mmol/l (= and of the DVGW (German Technical and Scientific As-
12° dH ), install water softeners. sociation for Gas and Water). It can be connected di-
Set the blend hardness to 1.4 mmol/l (= 8° dH). rectly to the public drinking water system.
• Install a conventional fine filter; fineness: > 80 µm (0.08
mm). Air quality
• Installation must be performed in compliance with the The air for driving the turbines, for cooling the drives and for
recommendations of the national installation require- the cooling spray must be free from oil, dry and hygienically
ments (e.g. EN 1717/DIN 1988). faultless.
• The water quality must comply with the national re-
quirements for drinking water. Suction pipe
• The connection must be made to cold water. Install steam trap K.
• When laying the water pipe to the treatment center, With a vacuum of pu >0.18 bar back pressure, the treat-
comply with the following instructions to reduce the ment center must be retrofitted with the "Vacuum limiter"
quantity of micro-organisms in the feed pipe: retrofit kit
(REF 59 68 826).

60 28 000 D3417
D3417.021.04.10.02 7
1 Preparations Sirona Dental Systems GmbH
Installation Requirements C8+ / C8+ Turn

1.3 Supply lines in the termination panel

9
4

m
m

m
m

.5
8“
3/
36
50

1
5
2 1
6
6
6 7
m
m
8“
.5
3/
36
1

Center base plate


opening to floor
3

Foot end

7, 6
3/8”
min.
max. max. 500mm
max.
60mm 5 mm 5 mm 20”
2 3/8” 3/16” 3/16”
10 x 1mm

10 x 1mm
3/8” x 1/32“

3/8” x 1/32“

2 3 1 4 5

60 28 000 D3417
8 D3417.021.04.10.02
Sirona Dental Systems GmbH 1 Preparations
Installation Requirements C8+ / C8+ Turn

Supply lines in the termination panel

båÖäáëÜ
• Observe the national regulations for electrical installa- Table 1: Supply lines
tions (e.g. VDE 0100, VDE 0100, Part 710).
Item Description
• Comply with the national regulations for water supply
1 Water inlet pipe 10x1 mm,
installations (e.g. EN 1717, DIN 1988) and sewage in-
corner valve outlet 3/8"
stallations (e.g. EN 12056-1).
2 Compressed air inlet pipe 10x1mm, corner
• For the suction pipe, observe the instructions in the
valve outlet 3/8"
Suction Machine Installation Instructions.
3 Suction pipe DN40 HT-PP DIN 19560 (polypro-
• For fastening the pipe ends in the installation field, we
pylene, inner diameter 36.5mm!)
recommend using an installation template. They can
be ordered from Sirona under REF 33 15 830. 4 Water drain DN40 HT-PP DIN 19560 (polypro-
pylene, inner diameter 36.5mm!)
If necessary, you can also prepare the template your-
self based on the above sketch (not true to scale!). 5 Installation pipe, DN40 HT-PP DIN 19560
(polypropylene, 40mm!)
6 Suction machine control cable ( )and call
cables ( , ) 3x1.5mm2
7 Power cable 3x1.5mm2
Fuse: 16A slow-blow
Recommended: Type B automatic circuit
breaker
8 Not applicable
9 Installation pipe (or corresponding flat duct) for
additional requirement e.g. practice network
connecting cable
60 28 000 D3417
D3417.021.04.10.02 9
1 Preparations Sirona Dental Systems GmbH
Installation Requirements C8+ / C8+ Turn

Supply lines in the termination panel

3/8“
m
,5 m
Ø 36 <6
0
<2 mm
3/
8“
Ø
36
,5
m
m
3/8“

m
<5m “
6
<3/1

0mm
>50 0“
>2

m
1m
mm 10x 1/32”
>5
00 ” x
3/8
0’’
>2
Installation template
m
6,6m
Ø3

40
DN

mm
200 ’’ mm
8 250 0’’
1

Supply above the floor, "above-floor installation" Supply through the floor, "underfloor installation"

The supply pipe ends, corner valves and cables must be 1. The top edges of the corner valves for air and water
routed as shown above. must not protrude more than 60mm above the upper
surface of the finished floor.
The retrofit kit for above-floor installation (REF 33 17 265)
is required for connection. 2. The suction and drain pipes must be flush with the up-
per surface of the floor (a deviation of +5mm is permis-
ATTENTION sible).
For cleaning, flush the air and water pipes thoroughly (metal Internal diameter for both pipes: 36.5mm.
chips!). 3. The electrical cables must protrude at least 500mm.

60 28 000 D3417
10 D3417.021.04.10.02
Sirona Dental Systems GmbH 1 Preparations
Installation Requirements C8+ / C8+ Turn

1.4 Underfloor installation of SIVISION connections

båÖäáëÜ
SIVISION

2.

L15, L37, L38, audio


Protective earth conductor 1.

Important information for the installer 1. Bend the wire at the front end of cables L37 and L38 to
form a hook.
Depending on the prevailing local conditions, the existing
cable set can be installed in the cable duct of an underfloor 2. Pull cables L15, L37, and L38 as well as the audio ca-
installation by an installer prior to the installation of the ble and the protective earth conductor for the treatment
treatment center. In this case, please observe the following: center through the cable duct to the location of the SIV-
ISION PC.
Proceed with extreme care when running the cables. Par-
ticularly cables L15 and L38 are very sensitive, and must Save the accessory parts for final installation!
never be kinked or twisted. The cables must not overlap or
cross one another.
RS232 (L37) and the XGA cable (L38) are not yet cut to
length and terminated on the PC side. It would be impossi-
ble to pull the cables through when installing them under
floor level if a sub D connector were already connected.
These cables should always be pulled.
Free length A of cables at the treatment center end:
Length A = 600mm
If S video cable L15 is equipped with both a female and a
male connector, make sure that the female connector
(socket) points to the connection box of the treatment cen-
ter.

60 28 000 D3417
D3417.021.04.10.02 11
1 Preparations Sirona Dental Systems GmbH
Installation Requirements C8+ / C8+ Turn

1.5 Mounting plates C8+ (not for C8+ Turn)

Mounting plate
REF 58 26 511

For especially uneven floor conditions, a steel mounting


plate is available to compensate for the irregularities of
the floor (REF 58 26 511).

Demonstration chair plate


REF 47 08 074

For floors which do not permit permanent connection of the unit (e.g.
demo operation at a fair), installation on a steel demonstration chair
plate is possible (REF 47 08 074).

60 28 000 D3417
12 D3417.021.04.10.02
Sirona Dental Systems GmbH 1 Preparations
Installation Requirements C8+ / C8+ Turn

1.6 Demonstration chair plate C8+ Turn

båÖäáëÜ
Demonstration chair plate
C8+ Turn
REF 59 46 269

For floors which do not permit permanent connection


of the unit (e.g. demo operation at a fair), installation
on a steel demonstration chair plate is possible
(REF 59 46 269).

60 28 000 D3417
D3417.021.04.10.02 13
H
`U / C8 Turn +
2 Dimensions, technical data
2.1 C8+ 1:20................................................................................................................................. 16

2.2 C8+ Cart 1:20 ......................................................................................................................... 21

2.3 C8+: Technical data ............................................................................................................... 23

2.4 C8+ Turn 1:20 ........................................................................................................................ 24

2.5 C8+ Turn Cart 1:20................................................................................................................. 31

2.6 C8+ Turn: Technical data ....................................................................................................... 33

60 28 000 D3417
D3417.021.04.10.02 15
2 Dimensions, technical data Sirona Dental Systems GmbH
2.1 C8+ 1:20 Installation Requirements C8+ / C8+ Turn

2.1 C8+ 1:20


Recommended distances from cabinet or wall

1700 1300
67” 51 3/16”

2500
!
98 7/16”

Hazard warning:
The lamp installed here and the dentist
element have a swivel range which ex-
ceeds the specified distances!

250
9 7/8”

250
9 7/8”
!
Center of floor opening

60 28 000 D3417
16 D3417.021.04.10.02
Sirona Dental Systems GmbH 2 Dimensions, technical data
Installation Requirements C8+ / C8+ Turn 2.1 C8+ 1:20

C8+ 1:20

båÖäáëÜ
With long support arm for the dentist element

LEDview SIROLUX

10°

90
3 1/2” Height of the lamp with
SIVISION on the swivel
arm or lamp support
tube
65° 1950
76 3/4”

45°

Height of the lamp


without SIVISION or
SIVISION on the tray
1875
73 3/4”

max. 1230 48 3/8”


45 min. 760 29 7/8”
1 3/4”

* max. 755 29 3/4”


min. 395 15 1/2"

1212
47 3/4”

Standard upholstery
* max.: backrest and chair at highest setting
min.: backrest and chair at lowest setting

60 28 000 D3417
D3417.021.04.10.02 17
2 Dimensions, technical data Sirona Dental Systems GmbH
2.1 C8+ 1:20 Installation Requirements C8+ / C8+ Turn

C8+ 1:20

250
9 7/8”

770 750
30 3/8” 29 1/2”

90°
180
7” 60°

90°
90°
2070 +50***
81 1/2” +2“ ***

430 +15
17“ +5/8”
1210
47 1/2”

860
33 7/8”

78
3” 45
1 3/4”
max. 380 +15
15“ +5/8” 459
18”

310
12 1/4”

1540
60 5/8”

*** Depending on the


chair height

60 28 000 D3417
18 D3417.021.04.10.02
Sirona Dental Systems GmbH 2 Dimensions, technical data
Installation Requirements C8+ / C8+ Turn 2.1 C8+ 1:20

C8+ 1:20

båÖäáëÜ
With short support arm for the dentist element

LEDview SIROLUX

10°

Height of the lamp with


90 SIVISION on the swivel
3 1/2” arm or lamp support
tube
65° 1950
76 3/4”

45°

Height of the lamp


without SIVISION or
SIVISION on the tray
1875
73 3/4”

max. 1440 56 3/4”


45 min. 970 38 1/8”
1 3/4”

* max. 755 29 3/4”


min. 395 15 1/2"

1212
47 3/4”

Standard upholstery
* max.: backrest and chair at highest setting
min.: backrest and chair at lowest setting

60 28 000 D3417
D3417.021.04.10.02 19
2 Dimensions, technical data Sirona Dental Systems GmbH
2.1 C8+ 1:20 Installation Requirements C8+ / C8+ Turn

C8+ 1:20

250
9 7/8”

770 750
30 3/8” 29 1/2”

90°
180
7” 60°

90°
90°
2070 +50***
81 1/2” +2“ ***

430 +15
17“ +5/8”

1210
47 1/2”
860
33 7/8”

45
150° 1 3/4”
max. 380 +15
15“ +5/8” 459
18”

140°

310
12 1/4”

1170
46”

*** Depending on the chair height

60 28 000 D3417
20 D3417.021.04.10.02
Sirona Dental Systems GmbH 2 Dimensions, technical data
Installation Requirements C8+ / C8+ Turn 2.2 C8+ Cart 1:20

2.2 C8+ Cart 1:20

båÖäáëÜ
LEDview SIROLUX

10°

90
3 1/2”

Height of the lamp with


65° SIVISION on the swivel
arm or lamp support
tube
45° 1950
Height setting in 5 steps, 26 mm each 76 3/4”

Height of the lamp


without SIVISION or
SIVISION on the tray
930
36 5/8” 1875
73 3/4”
45
800 1 3/4”
31 1/2”
* max. 755 29 3/4”
min. 395 15 1/2"

1212
47 3/4”

Standard upholstery
* max.: backrest and chair at highest setting
min.: backrest and chair at lowest setting

60 28 000 D3417
D3417.021.04.10.02 21
2 Dimensions, technical data Sirona Dental Systems GmbH
2.2 C8+ Cart 1:20 Installation Requirements C8+ / C8+ Turn

C8+ Cart 1:20

250
9 7/8”

770 750
30 3/8” 29 1/2”

90°
180
7” 60°

90°
90°

2070 +50***
81 1/2” +2“ ***

1210
430 +15 47 1/2”
17“ +5/8”

860
33 7/8”

45
1 3/4”
max. 380 +15
15“ +5/8” 459
18”

310
12 1/4”

Supply cable
≤ 2300
480 ≤ 90 1/2”
18 7/8”

420
16 1/2”

430
17”

*** Depending on the chair height

60 28 000 D3417
22 D3417.021.04.10.02
Sirona Dental Systems GmbH 2 Dimensions, technical data
Installation Requirements C8+ / C8+ Turn 2.3 C8+: Technical data

2.3 C8+: Technical data


Technical data

båÖäáëÜ
Power supply line 100 V / 115 V / 127 V / 230 VAC 50 / 60 Hz
Nominal current 11.5 A .....................at 100 V AC
10 A .....................at 115 V AC
9A .....................at 127 V AC
5A .....................at 230 V AC
In addition, max. 6 A for external devices
Main unit fuse T 10 A H, 250 V AC , REF 10 77 460
Transport and storage conditions Temperature: -40°C – +70°C (-40°F – +158°F)
Relative humidity: ......................10% – 95%
Air pressure: ......... 500hPa – 1060hPa
Operating conditions Ambient temperature: ......... 10°C – 40°C (50°F – 104°F)
Relative humidity: ............ 30% – 85% (no condensation)
Air pressure: ..................................... 700hPa – 1060hPa
Maximum load capacity of the patient chair 135kg, including accessories
Installation site ≤ 3000 m above sea level
Pollution degree 2 acc. to IEC 60664-1
Protection class Class I device
Device class in accordance with Directive 93/ Class IIa equipment
42/EEC
Degree of protection against electric shock
Type B applied parts,
except SIROCAM 3/SIROCAM C:
Type BF applied parts

Degree of protection against ingress of water Ordinary equipment (not protected)


The foot switch has an IPX1 degree of protection against liquids
(drip-proof).
Operating mode Continuous operation with intermittent loading corresponding to the
dental mode of working.
Permanently connected unit.
Year of manufacture ..............................................................
........................ (on rating plate of chair)

Dimensions of the packaging


Dentist element, assistant element, 120cm x 63cm x 110cm
C8+ Cart dentist element 97.5cm x 50.5cm x 50.5cm
Water unit 120cm x 63cm x 96cm
Chair 157cm x 64cm x 82 cm
Upholstery 120cm x 52cm x 40cm
SIROLUX / LEDview 10.5cm x 56cm x 28cm
Weight (1kg = 2.2lbs) incl. / without packaging
Dentist element, assistant element, 37.5kg / 10kg + 3kg
C8+ Cart dentist element 17kg / 13kg
Water unit 44kg / 16.5kg
Chair 120kg / 90kg
Upholstery 13kg / 9kg
SIROLUX / LEDview 14kg / 10kg
On-site pressure readings (1 bar = 100 kPa = 14.5 psi)
Air min./max. 550 kPa / 750 kPa (5.5 / 7.5bar)
Water min./max. 250 kPa / 600 kPa (2.5 / 6bar)

60 28 000 D3417
D3417.021.04.10.02 23
2 Dimensions, technical data Sirona Dental Systems GmbH
2.4 C8+ Turn 1:20 Installation Requirements C8+ / C8+ Turn

2.4 C8+ Turn 1:20


Recommended distances from cabinet or wall

1600 1600
63” 63”

2500
98 7/16”

Hazard warning:
The lamp installed here and the dentist
element have a swivel range which ex-
ceeds the specified distances!

250
9 7/8”

250
9 7/8”

Center of floor opening

60 28 000 D3417
24 D3417.021.04.10.02
Sirona Dental Systems GmbH 2 Dimensions, technical data
Installation Requirements C8+ / C8+ Turn 2.4 C8+ Turn 1:20

C8+ Turn 1:20

båÖäáëÜ
Minimum distances for conversion from right-handed
to left-handed treatment center and vice versa
2700x2650mm

1050 1050
41 3/8” 41 3/8”
1600 1600
63” 63”
1234
48 1/2”
Cabinet on the
right or left
max. height:
900mm

2750
108 1/4”

250
9 7/8”

250 Center of floor opening


9 7/8”

60 28 000 D3417
D3417.021.04.10.02 25
2 Dimensions, technical data Sirona Dental Systems GmbH
2.4 C8+ Turn 1:20 Installation Requirements C8+ / C8+ Turn

C8+ Turn 1:20


Minimum distances for conversion from right-handed
to left-handed treatment center and vice versa in
narrow spaces 3650x2100mm

1050 1050
41 3/8” 41 3/8”
Installation area
max. 900mm high

2134
84”
1250
49 1/4”

3650
143 3/4”

250
9 7/8”

250
9 7/8” Center of floor opening

60 28 000 D3417
26 D3417.021.04.10.02
Sirona Dental Systems GmbH 2 Dimensions, technical data C8+ / C8+ Turn 1:20
Installation Requirements C8+ / C8+ Turn 2.4 C8+ Turn 1:20

C8+ Turn 1:20

båÖäáëÜ
Identical working conditions for right-handed and
left-handed operators:
Position for right-handed operators

1050 1050
41 3/8” 41 3/8”
1600 1600
63” 63”
1234
48 1/2”
17°

2750
108 1/4”

250
9 7/8”

250
Center of floor opening 9 7/8”

60 28 000 D3417
D3417.021.04.10.02 27
2 Dimensions, technical data C8+ / C8+ Turn 1:20 Sirona Dental Systems GmbH
2.4 C8+ Turn 1:20 Installation Requirements C8+ / C8+ Turn

C8+ Turn 1:20


Identical working conditions for right-handed and
left-handed operators:
Position for left-handed operators

1050 1050
41 3/8” 41 3/8”
1600 1600
63” 63”
1234
48 1/2”
17°

2750
108 1/4”

250
9 7/8”

250
Center of floor opening 9 7/8”

60 28 000 D3417
28 D3417.021.04.10.02
Sirona Dental Systems GmbH 2 Dimensions, technical data C8+ / C8+ Turn 1:20
Installation Requirements C8+ / C8+ Turn 2.4 C8+ Turn 1:20

C8+ Turn 1:20

båÖäáëÜ
LEDview SIROLUX

10°

210 Height of the lamp with


8 1/4” SIVISION on the swivel
arm or lamp support
tube
65° 1950
76 3/4”

45°
Height of the lamp
without SIVISION or
SIVISION on the tray
1875
73 3/4”

max. 1230 48 3/8”


45 min. 760 29 7/8”
1 3/4”
* max. 770 30 3/8”
min. 410 16 1/8"

1266
49 7/8”

Standard upholstery
* max.: backrest and chair at highest setting
min.: backrest and chair at lowest setting

60 28 000 D3417
D3417.021.04.10.02 29
2 Dimensions, technical data C8+ / C8+ Turn 1:20 Sirona Dental Systems GmbH
2.4 C8+ Turn 1:20 Installation Requirements C8+ / C8+ Turn

C8+ Turn 1:20

250
9 7/8”

770 750
30 3/8” 29 1/2”

90°
180
7” 60°

90°
90°

2070 +50***
81 1/2” +2“ ***

430 +15
17“ +5/8”

1270
50”
920
36 1/4”

45
max. 380 +15 1 3/4”
15“ +5/8”
400
15 3/4”

310
12 1/4”

1480
58 1/4”

800 +50***
31 1/2” +2”

*** Depending on the chair height

60 28 000 D3417
30 D3417.021.04.10.02
Sirona Dental Systems GmbH 2 Dimensions, technical data C8+ / C8+ Turn 1:20
Installation Requirements C8+ / C8+ Turn 2.5 C8+ Turn Cart 1:20

2.5 C8+ Turn Cart 1:20

båÖäáëÜ
LEDview SIROLUX

10°

210
8 1/4”

Height of the lamp with


65° SIVISION on the swivel
arm or lamp support
tube
45° 1950
Height setting in 5 steps, 26 mm each 76 3/4”

Height of the lamp


930 without SIVISION or
36 5/8” SIVISION on the tray
1875
73 3/4”
45
800 1 3/4”
31 1/2”

* max. 770 30 3/8”


min. 410 16 1/8"

1266
49 7/8”

Standard upholstery
* max.: backrest and chair at highest setting
min.: backrest and chair at lowest setting

60 28 000 D3417
D3417.021.04.10.02 31
2 Dimensions, technical data C8+ / C8+ Turn 1:20 Sirona Dental Systems GmbH
2.5 C8+ Turn Cart 1:20 Installation Requirements C8+ / C8+ Turn

C8+ Turn Cart 1:20

250
9 7/8”

770 750
30 3/8” 29 1/2”

90°
180
7” 60°

90°
90°

2070 +50***
81 1/2” +2“ ***

430 +15
17“ +5/8”

1270
50”
920
36 1/4”

45
max. 380 +15 1 3/4”
15“ +5/8”
400
15 3/4”

310
12 1/4”
480
18 7/8”

Supply cable
420
16 1/2” ≤ 2300
≤ 90 1/2”

430
17”

*** Depending on the chair height

60 28 000 D3417
32 D3417.021.04.10.02
Sirona Dental Systems GmbH 2 Dimensions, technical data C8+ / C8+ Turn 1:20
Installation Requirements C8+ / C8+ Turn 2.6 C8+ Turn: Technical data

2.6 C8+ Turn: Technical data


Technical data

båÖäáëÜ
Power supply line 100 V / 115 V / 127 V / 230 VAC 50 / 60 Hz
Nominal current 11.5 A ...........................................................at 100 V AC
10 A ..............................................................at 115 V AC
9 A ................................................................at 127 V AC
5 A ................................................................at 230 V AC
In addition, max. 6 A for external devices
Main unit fuse T 10 A H, 250 V AC , REF 10 77 460
Transport and storage conditions Temperature: -40°C – +70°C (-40°F – +158°F)
Relative humidity: ......................10% – 95%
Air pressure: ......... 500hPa – 1060hPa
Operating conditions Ambient temperature: ......... 10°C – 40°C (50°F – 104°F)
Relative humidity: ............ 30% – 85% (no condensation)
Air pressure: ..................................... 700hPa – 1060hPa
Maximum load capacity of the patient chair 135kg, including accessories
Installation site ≤ 3000 m above sea level
Pollution degree 2 acc. to IEC 60664-1
Protection class Class I device
Device class in accordance with Directive Class IIa equipment
93/42/EEC
Degree of protection against electric shock
Type B applied parts,
except SIROCAM 3/SIROCAM C:
Type BF applied parts

Degree of protection against ingress of water Ordinary equipment (not protected)


The foot switch has an IPX1 degree of protection against liquids
(drip-proof).
Operating mode Continuous operation with intermittent loading corresponding to the
dental mode of working.
Permanently connected unit.
Year of manufacture ..............................................................
........................ (on rating plate of chair)

Dimensions of the packaging


Dentist element, assistant element, 120cm x 63cm x 110cm
C8+ Turn Cart dentist element 97.5cm x 50.5cm x 50.5cm
Water unit 120cm x 63cm x 96cm
Chair 157cm x 64cm x 82 cm
Connection box 72cm x 53.2cm x 44cm
Upholstery 120cm x 52cm x 40cm
SIROLUX / LEDview 10.5cm x 56cm x 28cm
Weight (1kg = 2.2lbs) incl. / without packaging
Dentist element, assistant element, 37kg / 10kg + 3kgkg
C8+ Turn Cart dentist element 17kg / 13kg
Water unit 58 kg / 36kg
Chair 177kg / 140kg
Connection box 31kg / 25kg
Upholstery 13kg / 9kg
SIROLUX / LEDview 14kg / 10kg
On-site pressure readings (1 bar = 100 kPa = 14.5 psi)
Air min./max. 550 kPa / 750 kPa (5.5 / 7.5bar)
Water min./max. 250 kPa / 600 kPa (2.5 / 6bar)

60 28 000 D3417
D3417.021.04.10.02 33
3 Electromagnetic compatibility
`UH / C8+ Turn

3.1 Accessories ........................................................................................................................... 36

3.2 Electromagnetic emission ...................................................................................................... 37

3.3 Immunity to interference ........................................................................................................ 38

3.4 Working clearances ............................................................................................................... 40

i NOTE
The C8+ / C8+ Turn fulfills all requirements for electromagnetic compatibility (EMC) compliant with IEC 60601-1-2.
The C8+ / C8+ Turn is referred to as "UNIT" in the sections that follow.

Observance of the following information is necessary to ensure safe operation regarding EMC aspects.

60 28 000 D3417
D3417.021.04.10.02 35
3 Electromagnetic compatibility Sirona Dental Systems GmbH
Installation Requirements C8+ / C8+ Turn

3.1 Accessories
Making the PC connection

Designations of interface cables Supplier


for PCs

XGA cable, 10 m (L38) Sirona

S video cable, 10 m (L15) Sirona

RS232 cable, 10 m (L37) Sirona

Audio cable, 10m Sirona

2nd protective ground wire, 2.5 mm2, Sirona


10m

• The UNIT may only be operated with accessories and


spare parts approved by Sirona. Unapproved accesso-
ries and spare parts may lead to an increased emission
or to a reduced immunity to interference.
• The UNIT should not be operated in the immediate vi-
cinity of other devices. If this proves to be unavoidable,
the UNIT should be monitored to check and make sure
that it is used properly.

The EMC measurements were performed with the following


PCs:

PC as peripheral device for Siemens Fujitsu, Pentium IV,


checking the interfaces with: 3.0 GHz

Extension of the PC

Graphics card NVIDIA GeForce 7300 LE

Frame grabber card PicPort Color frame grabber


card (Leutron)
REF 46 93 961

60 28 000 D3417
36 D3417.021.04.10.02
Sirona Dental Systems GmbH 3 Electromagnetic compatibility
Installation Requirements C8+ / C8+ Turn

3.2 Electromagnetic emission


The UNIT is intended for operation in the electromagnetic The customer or user of the UNIT should make sure that it

båÖäáëÜ
environment specified below. is used in such an environment.

Emission measurement Conformity Electromagnetic environment guidelines

RF emissions according to CISPR Group 1 The UNIT uses RF energy only for its internal function. There-
11 fore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.

RF emissions according to CISPR Class B The UNIT is intended for use in all facilities, including residen-
11 tial areas and in any facilities connected directly to a public
power supply providing electricity to buildings used for resi-
Harmonics Class A dential purposes.
according to IEC 61000-3-2

Voltage fluctuations/Flicker accord- Complies


ing to IEC 61000-3-3

60 28 000 D3417
D3417.021.04.10.02 37
3 Electromagnetic compatibility Sirona Dental Systems GmbH
Installation Requirements C8+ / C8+ Turn

3.3 Immunity to interference


The UNIT is intended for operation in the electromagnetic The customer or user of the UNIT should make sure that it
environment specified below. is used in such an environment.

Interference immu- IEC 60601-1-2 test level Compliance level Electromagnetic environment
nity tests guidelines

Electrostatic dis- ± 6kV contact discharge ± 6kV contact discharge Floors should be made of wood or
charge (ESD) accord- concrete or finished with ceramic til-
ing to IEC 61000-4-2 ± 8 kV air discharge ± 8kV air discharge ing. If floors are covered with syn-
thetic material, the relative humidity
should be at least 30%.

Electrical fast tran- ± 1kV for input and output ± 1kV for input and output The quality of the line power supply
sient/burst lines lines should be that of a typical commercial
according to IEC or hospital environment.
61000-4-4 ± 2kV for power cables ± 2kV for power cables

Surge voltages ± 1kV differential mode ± 1kV differential mode The quality of the line power supply
according to IEC should be that of a typical commercial
61000-4-5 ± 2kV common mode ± 2kV common mode or hospital environment.

Voltage dips, short < 5 % UT for ½ period (> 95 < 5 % UT for ½ period (> 95 The quality of the line power supply
interruptions and vari- % dip of UT) % dip of UT) should be that of a typical commercial
ations of the power or hospital environment.
supply 40% UT for 5 periods (60% 40% UT for 5 periods (60%
according to IEC dip of UT) dip of UT) If the user of the UNIT requires it to
61000-4-11 continue functioning following inter-
70% UT for 25 periods (30% 70% UT for 25 periods ruptions of the power supply, it is rec-
dip of UT) (30% dip of UT) ommended to have the UNIT
< 5 % UT for 5 sec. (> 95 < 5 % UT for 5 sec. (> 95 powered by an uninterrupted power
% dip of UT) % dip of UT) supply or a battery.

Magnetic field of 3 A/m 3 A/m Power frequency magnetic fields


power frequencies should be at levels characteristic of a
(50/60 Hz) typical location in a typical commer-
according to IEC cial or hospital environment.
61000-4-8

Remarks: UT is the AC supply voltage prior to application of the test level.

60 28 000 D3417
38 D3417.021.04.10.02
Sirona Dental Systems GmbH 3 Electromagnetic compatibility
Installation Requirements C8+ / C8+ Turn

Interference immu- IEC 60601-1-2 test level Compliance level Electromagnetic environment
nity tests guidelines

båÖäáëÜ
Portable and mobile radio equipment
must not be used within the recom-
mended working clearance from the
UNIT and its cables, which is calcu-
lated based on the equation suitable
for the relevant transmission fre-
quency.
Recommended working clearance:

Conducted RF inter- 3Veff 3Veff


ference 150 kHz to 80 MHza d = [ 1, 2 ] P
IEC 61000-4-6

Radiated HF interfer- 3V/m 3Veff


ence 80MHz to 800MHza d = [ 1, 2 ] P
IEC 61000-4-3 at 80MHz to 800MHz
3V/m 3Veff
800MHz to 2.5GHza d = [ 2, 3 ] P
at 800MHz to 2.5GHz

where P is the nominal transmitter


output in watts (W) specified by the
transmitter manufacturer and d is the
recommended working clearance in
meters (m).
Field strengths from fixed RF trans-
mitters, as determined by an electro-
magnetic site surveyb, should be less
than the compliance levelc in each
frequency range.
Interference is possible in the vicinity
of equipment bearing the following
graphic symbol.

a. The higher frequency range applies at 80MHz and 800MHz.

b. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM/FM radio and TV broadcasts, cannot be predicted theoretically with accuracy. An investiga-
tion of the location is recommended to determine the electromagnetic environment resulting from stationary RF trans-
mitters. If the measured field strength in the location in which the UNIT is used exceeds the applicable RF compliance
level above, the UNIT should be observed to verify normal operation. If unusual performance characteristics are
observed, it may be necessary to take additional measures such as reorientation or repositioning of the UNIT.

c. Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.

60 28 000 D3417
D3417.021.04.10.02 39
3 Electromagnetic compatibility Sirona Dental Systems GmbH
Installation Requirements C8+ / C8+ Turn

3.4 Working clearances


Recommended working clearances between
portable and mobile RF communication devices
and the UNIT
The UNIT is intended for operation in an electromagnetic
environment, where radiated RF interference is checked.
The customer or the user of the UNIT can help prevent
electromagnetic interference by duly observing the mini-
mum distances between portable and/or mobile HF com-
munication devices (transmitters) and the UNIT. These
values may vary according to the output power of the rele-
vant communication device as specified below.

Nominal transmitter output Working clearance according to transmission frequency [m]


[W]
150kHz to 80MHz 80MHz to 800MHz 800MHz to 2.5GHz

d = [ 1, 2 ] P d = [ 1, 2 ] P d = [ 2, 3 ] P
0.01 0.12 0.12 0.23

0.1 0.38 0.38 0.73

1 1.2 1.2 2.3

10 3.8 3.8 7.3

100 12 12 23

For transmitters whose maximum nominal output is not


specified in the above table, the recommended working
clearance d in meters (m) can be determined using the
equation in the corresponding column, where P is the max-
imum nominal output of the transmitter in watts (W) speci-
fied by the transmitter manufacturer.
Remark 1
The higher frequency range applies at 80 MHz and 800
MHz.
Remark 2
These guidelines may not be applicable in all cases. The
propagation of electromagnetic waves is influenced by
their absorption and reflection by buildings, objects and
persons.

60 28 000 D3417
40 D3417.021.04.10.02
tÉ=êÉëÉêîÉ=íÜÉ=êáÖÜí=íç=ã~âÉ=~åó=~äíÉê~íáçåë=ïÜáÅÜ=ã~ó=ÄÉ=êÉèìáêÉÇ=ÇìÉ=íç=íÉÅÜåáÅ~ä=áãéêçîÉãÉåíëK

«=páêçå~=aÉåí~ä=póëíÉãë=dãÄe=OMMNJOMNO i~åÖì~ÖÉW=ÉåÖäáëÅÜ= mêáåíÉÇ=áå=dÉêã~åó


aPQNTKMONKMQKNMKMO=NOKOMNO `Ü~åÖÉ=kçKW=NNR=PVO fãéêáã¨=Éå=^ääÉã~ÖåÉ

páêçå~=aÉåí~ä=póëíÉãë=dãÄe

c~Äêáâëíê~ëëÉ=PN lêÇÉê=åçK SM=OU=MMM=aPQNT


SQSOR=_ÉåëÜÉáã
dÉêã~åó
ïïïKëáêçå~KÅçã

You might also like