3 Background Correction Optional Yes, Built-in Useful feature for GFA/ FIAS upgrade as well as few elements
studies in low UV region
4 Coded Lamp Yes Yes Conforms GLP
5 Lamp Turret Holder with Yes, Four as Yes, Four
No. of Lamp standard
6 Adjustment Manual Automated More flexibility with using different lamps and also automation,
each of operation.
7 Auto Alignment / energy No Yes
optimization
8 EDL Power Supply Optional Built-in Useful feature for Hydride forming elements and longer lamp life
as well as good intesity of emmission light.
9 Assymetric Modulation Yes Yes -----------------------------
10 Sample to Reference Signal 2:1 3:1 Higher time is spent in AA200 for sample beam signal
Ratio measurement.
11 Wavelength Range (nm) 185 - 900 189-900
12 Monochromator Design Czerny - Turner Echelle Echelle Grating Monochromator has highest light throughput,
uniform light intensity in all orders, highest resolution and hence,
best detection limits. for Eg : As, Ba elements echelle gives
better performance in Detection Limits than any other models.
Excellent performance in lower UV region
13 No. of Lines/mm 1200 79 In Echelle Grating, it is always used at it's blaze angle i.e.63
degree which gives highest light thru put and (0.12 nm)
dispersion. Higher ruling density have no significance in Echelle
Grating.
14 Blazed At (nm) 240 63 degree
15 Focal Length (mm/degree) 250 300 Higher focal length gives better dispersion (resolution)
16 Resolution Not mentioned 0.12 nm PerkinElmer offers highest resolution which leads to lesser
Interference in sample analysis.
17 Slit Width (nm) 0.2, 0.5 & 1.0 Variable More flexibility for HGA900 using AA200.
18 Background Correction with 2.3 2.5 Higher absorbance correction and hence difficult matrix can be
Gr.Furnace (Abs) addressed. This is also possible due to full STPF
implementation.
19 Area of Grating (mm x mm) Not mentioned 36 X 185 Higher the area, higher the light thru put.
20 Detector PMT SSD PMT has 3 times poor quantum efficiency than SSD. (PMT is
outdated Technology). Hence, SSD offers better Detection
Limits.This is proven showing our HGA 900 presentation.