Professional Documents
Culture Documents
Oxyprobe PDF
Oxyprobe PDF
November 2002
After Unpacking
Compare the contents of the shipping container against the packing slip. If the instrument is damaged in
shipping, report the extent of the damage to the carrier.
1. % Oxygen Sensor Operation
In this manual, the word “probe” means the assembly consisting of the Zirconia and on the ratio of the oxygen partial pressure
of % oxygen sensor, thermocouple sensor, their protection tube on the reference side of the electrolyte to the oxygen partial
and the cold end termination head ( see Figure 4-1). pressure on the process side of the electrolyte.
Two electrodes make contact with the platinum coated Probe Vdc = 0.0125 x T x Ln(O1/O2)
zirconium element at the tip of the sensor, one outer
electrode, one inner electrode. Where:
The special alloy protection tube is one conductor for the Probe Vdc = Sensor output, in millivolts
oxygen probe signal, which eliminates the need of a small
signal wire normally exposed to the the furnace environment. T = Sensor temperature, in Degrees Kelvin
The protection tube has excellent resistance to corrosion
and oxidation at high temperature, and has good mechanical Ln = Natural logarithm
strength (since Hydrogen causes embrittlement).
O1 = Oxygen concentration on the inside surface of
The AP Probe is suitable for high temperature carburizing the sensor, namely, ambient air which is 20.9%
heat treating, carbonitriding and endothermic gas generators. oxygen
The AP Probe is NOT suitable for nitriding applications.
O2 = Oxygen concentration on the outside surface
Zirconia is a solid state electrolyte which conducts oxygen of the sensor, namely, furnace atmosphere
ions at temperatures above 1400°F. The ion conduction
results in a voltage between the two electrodes. The Two connections to the Zirconia cell conduct the voltage to
magnitude of the voltage is dependent on the temperature the four pin connector.
Thermocouple
(20 AWG)
2. Specifications
Output 1.00 to 1.20 Vdc over operating range. Weight 3 pounds, 5 ounces.
Readout % oxygen sensors should be used with controlling, Response Time Less than 1.0 second.
recording and indicating instruments having an input
impedance of 8 megohms or higher. Reference Air Uncontaminated air at maximum rate of 236
cc per minute = 0.5 SCFH (standard cubic feet per hour).
Immersion Depth 3 inches minimum.
Protection Tube Special alloy is resistant to corrosion and
Thermal and Mechanical Shock The special alloy outer oxidation up to 2012°F (1100°C).
protection tube protects against both thermal and mechanical
shock. As with any precision instrument, take reasonable Operating Temperature 1400°F (760°C) to 2012°F
care when handling and installing. (1100°C).
Accuracy ±0.05 weight percent carbon in normal operating Thermocouple Types K, R and S.
range.
Page 2 1122-IN-001-0-12
3. Probe Location on Furnace 4. Mounting the Probe and Air Supply
If the enriching gas supply has a separate inlet, locate the The probe is supplied with a compression fitting which has 1”
probe in the following manner: NPT male threads. Install a mounting fitting with 1-1/2” NPT
female threads and use a 1” x 1-1/2” reducing bushing. This
• close to atmosphere intake, where atmosphere is
permits easy removal after soot buildup on the protection tube.
well mixed and at furnace temperature
• as far away from the fumes of quench oil as possible Mount the reference air supply in an area where the ambient
• as far away from door opening as possible temperature is less than 125°F. if mounting on the furnace,
have at least 8 inches of air space between the supply and
• work load should be clean – free from oils the furnace wall.
• probe should not be located in a “hot spot”
• avoid hot spots in the furnace
• the probe can be mounted at any angle
Ports to Furnace
on each side
1” NPT Optional
Closed End Process Mtg. Air Purge
Special Alloy Fitting 1” Ref.
4-pin
Reference Electrical
Air Supply Connector
Connection
1122-IN-001-0-12 Page 3
70
6. Thermocouple Wiring
NEVER wire thermocouples using ordinary copper wire. Wire 150
0°F
only with thermocouple extension wire. 60
8. Furnace Startup Mercury and certain other heavy metals will attack the oxygen
Turn on reference air to the probe at furnace startup. If the probe. Consult the factory for any unusual application.
probe is either installed or replaced in a hot furnace, turn
reference air on as soon as possible. New and rebricked furnaces should be fully dried and fully
cured before installing the oxygen probe.
The tables at the end of this manual are required in order to
interpret the output voltage of the oxygen sensor in the probe. New and rebricked furnaces require several days to stabilize
These tables are not required when the 2400 Series Carbon so the new brick can absorb carbon atmosphere. During
Controller is used. Also see Figure 8-1. this time, exact carbon control will be difficult.
Page 4 1122-IN-001-0-12
It is common procedure to operate a new and rebricked Burnout Procedure (Continuous Furnace)
furnace at an elevated temperature with a carburizing Recommended air flow: 2 SCFH
atmosphere to eliminate harmful components. Purge period: 5 minutes
Burnout cycle: 3 hours
11. Maintenance
Check the air pump daily. 1. Hold the enrichment gas flow at its current value
during the purge cycle.
Check the air supply to the probe once a week. Remove the
air tube fitting from its connection at the probe head and 2. Do NOT short the probe terminals.
submerge the air tube into a container of water. A steady
flow of air bubbles indicates proper operation. Reconnect 3. If the carbon probe temperature exceeds 1900°F,
the air tube fitting to the probe head. reduce the purge air flow to 1 SCFH (or discontinue
purge air completely) until probe temperature
Check the air supply filters once a month. decreases to a level below 1700°F.
8. Check the furnace door for air leaks. 3. If the carbon probe temperature exceeds 1900°F,
discontinue purge air and close the hearth zone
9. Check the furnace completely for air leaks door until probe temperature falls to 1700°F.
(including the floor).
4. Purge air can be supplied to the probe whenever
10. Check for air blocks in the furnace. the hearth chamber door is open. The length of
time that purge air is supplied is not critical unless
11. Make certain parts are clean before loading them the probe and furnace are excessively sooted.
into the furnace.
The burn off of excessive soot should be monitored closely
12. Make certain to give a new or rebuilt furnace a with a probe thermocouple monitor. It may be necessary to
two day curing time. remove the probe from the furnace and remove excess soot
with compressed air (after first cooling the probe to ambient
temperature) or light brushing. Do not strike the unit or insert
12. Probe Burnout hard objects into the probe.
Furnaces are only capable of carrying a finite amount of
carbon. Carbon will “drop out” as the atmosphere cools, Sooted probes should not be allowed to burn off in an air
resulting in a buildup of soot inside the protection tube. The atmosphere without monitoring the temperature rise. Visually
soot is removed by purging the probe with a controlled observe the hot probe end. Sparks or white heat indicate an
amount of air, which allows the soot to burn off. Failure to excessive soot condition. Excessively sooted probes should
purge can result in a damaged probe. be returned to the factory for refurbishing.
1122-IN-001-0-12 Page 5
Probe Test Procedure Recommended Spare Parts
Air Supply Kit ........................................... Part No. A-13893
1. Record probe voltage. Air Flow Meter ........................................... Part No. 36-122
Air Hose (X = feet) .................................... Part No. X-3201
2. Place a test resistor in parallel across the voltmeter
leads (see diagram). Returning a Unit for Reconditioning
When returning a probe to the factory, package it carefully in
3. Record probe voltage. If the voltage is now less the original packaging material and mark the carton “Fragile
than 1/2 the voltage in step 1, the probe is defective. Instrument.”
Write the Entire 15-digit Model Number 5. The AP probe must be brought up to temperature
gradually, or else the Zirconia Oxide cell will be
damaged. When installing the probe into a hot
furnace, insert slowly (one inch every five minutes).
Page 6 1122-IN-001-0-12
8. If several furnaces are fed by one source of N2- 11. Close temperature control is desirable since %
Methanol, each furnace should have its own set carbon readings vary inversely with temperature
of pressure regulators and flowmeters so as to cycling. Fluctuations of temperature will cause
avoid flow rate changes as other furnaces are cyclic conditions of carbon control.
added to, or removed from production.
12. In order to provide extended, trouble-free service,
9. An accurate dewpoint analysis and a trained Platinum thermocouples should be employed in
operator are essential for setting up and checking AP probes used at 1700°F and above.
the furnace atmosphere.
13. Enriching gas should have a maximum flow rate
10. An atmosphere calibration chart from the supplier of approximately 10% of total gas flow (e.g., 90
of the N2-Methanol gas is essential. CFH natural gas to 900 CFH atmospheric gas).
A-13893-X00-X-0X
Models are available with a complete housing for surface A-13893-400-x-xx has one output with a 0.06-0.50 LPM flow
mounting or with a front panel only for installation into a meter for reference air and a solenoid activated output with
control panel. All versions are 50/60 Hz at 28 Watts. a 0.15-1.00 LPM flow meter for purge air.
A-13893-000-x-xx has a single output with a 0-2 SCFH flow A-13893-500-x-xx has a single output with a 0.06-0.50 LPM
meter for use as a reference air supply for a single probe. flow meter for use as a reference air supply for a single probe.
1122-IN-001-0-12 Page 7
Installing the Air Supply:
1. Mount the housing with a minimum of 8 inches (200mm) of air gap between the housing and the furnace wall. The air
supply should be within 50 ft. (15 meters) of the probe.
2. Remove the front cover.
3. Remove the safety cover on the terminal block.
4. Connect power leads to terminals L1 & L2. The air supply should be connected directly to the furnace control power.
The air supply should be ON whenever the furnace is powered up.
5. On models with a purge air solenoid, connect the purge control leads to the terminals marked S1 & S2. S2 may be
jumpered to L2 when the power leads and the purge control share the same circuit.
6. Connect terminal G to earth ground.
7. Reinstall the safety cover to the terminal block.
Model Number
A-13893 A-13893 A-13893 A-13893 A-13893 A-13893 A-13893 A-13893 A-13893 A-13893
-000-X-X0 -100-X-X0 -200-X-X0 -400-X-X0 -500-X-X0 -000-X-X1 -100-X-X1 -200-X-X1 -400-X-X1 -500-X-X1
Part
Item Description Qty. Qty. Qty. Qty. Qty. Qty. Qty. Qty. Qty. Qty.
Number
1 AL-431 Filter 1 1 1 1 1 1 1 1 1 1
36-151 1 to 10 SCFH Flow Meter 1 1
2
36-152 .15 to 1 LPM Flow Meter 1 1
50-1746 Air Pump 120V 50/60 Hz 1 1 1 1 1
3
50-1746-100 Air Pump 240V 50/60 Hz 1 1 1 1 1
AL-171 Solenoid 120V 50/60 Hz 1 1
4
AL-183 Solenoid 240V 50/60 Hz 1 1
36-130 .2 to 2 SCFH Flow Meter 1 2 1 1 2 1
5
36-158 .06 to .5 LPM Flow Meter 1 1 1 1
Page 8 1122-IN-001-0-12
Table 1. Millivolt Readings for Endothermic Atmospheres Generated from Methane (Natural Gas)
and Containing Approximately 20% (CO + CO2)
NOTES:
1. Carbon absorption and diffusion rates limit the actual level attainable on the steel work surface. (Carburizing time and
temperature and steel composition affect the relationship between atmospheric carbon potential and resultant surface
carbon level.) Specific conditions, such as shorter carburizing cycles, may necessitate controlling at a carbon potential
as much as 0.25 wt. %C higher than required on the work.
2. This chart pertains only to endothermic base atmospheres generated from predominantly methane (i.e., 20% CO).
3. Controlling carbon potential at a level exceeding the carbon content of saturated austentite is discouraged. An
optimum combination of productivity and control might be realized by a boost-diffuse carburizing technique.
1122-IN-001-0-12 Page 9
Table for 20% (CO + CO2)
% 1500°F 1510°F 1520°F 1530°F 1540°F 1550°F 1560°F 1570°F 1580°F 1590°F
Carbon 816°C 821°C 827°C 8832°C 838°C 843°C 849°C 854°C 860°C 886°C
0.10 1016
0.15 1032 1032 1033 1033
0.20 1042 1043 1044 1044 1045 1046 1046
0.25 1051 1051 1052 1053 1053 1054 1055 1056 1056 1057
0.30 1059 1060 1060 1061 1062 1063 1064 1065 1065 1066
0.35 1066 1067 1068 1069 1070 1071 1071 1072 1073 1074
0.40 1073 1074 1075 1076 1076 1077 1078 1079 1080 1081
0.45 1079 1080 1081 1082 1083 1084 1085 1086 1087 1088
0.50 1084 1085 1086 1087 1088 1089 1090 1092 1093 1094
0.55 1089 1090 1091 1093 1094 1095 1096 1097 1098 1099
0.60 1094 1095 1096 1097 1099 1100 1101 1102 1103 1104
0.65 1099 1100 1101 1102 1103 1104 1105 1107 1108 1109
0.70 1103 1104 1105 1106 1108 1109 1110 1111 1112 1114
0.75 1107 1108 1109 1110 1112 1113 1114 1115 1117 1118
0.80 1111 1112 1113 1114 1116 1117 1118 1119 1121 1122
0.85 1115 1116 1117 1118 1119 1121 1122 1123 1125 1126
0.90 1118 1119 1121 1122 1123 1124 1126 1127 1128 1130
0.95 1122 1123 1124 1125 1127 1128 1129 1131 1132 1133
1.00 1125 1126 1127 1129 1130 1131 1133 1134 1135 1137
1.05 1132 1133 1135 1136 1137 1139 1140
1.10 1139 1141 1142 1143
1.15
1.20
1.25
1.30
1.35
1.40
Page 10 1122-IN-001-0-12
Table for 20% (CO + CO2)
% 1700°F 1710°F 1720°F 1730°F 1740°F 1750°F 1760°F 1770°F 1780°F 1790°F
Carbon 927°C 932°C 938°C 943°C 949°C 954°C 960°C 966°C 971°C 977°C
0.10 1021 1022 1022 1023 1023 1024 1025 1025 1026 1027
0.15 1041 1041 1042 1043 1044 1045 1045 1046 1047 1048
0.20 1055 1056 1057 1058 1059 1060 1061 1062 1062 1063
0.25 1067 1068 1069 1070 1071 1072 1073 1074 1075 1076
0.30 1077 1078 1079 1080 1081 1082 1083 1084 1085 1086
0.35 1085 1087 1088 1089 1090 1091 1092 1093 1094 1095
0.40 1093 1094 1095 1096 1098 1099 1100 1101 1102 1103
0.45 1100 1101 1102 1103 1105 1106 1107 1108 1109 1111
0.50 1106 1107 1109 1110 1111 1112 1114 1115 1116 1117
0.55 1112 1113 1115 1116 1117 1118 1119 1121 1122 1123
0.60 1117 1119 1120 1121 1123 1124 1125 1126 1128 1129
0.65 1123 1124 1125 1126 1128 1129 1130 1132 1133 1134
0.70 1127 1129 1130 1131 1133 1134 1135 1137 1138 1139
0.75 1132 1133 1135 1136 1137 1139 1140 1141 1143 1144
0.80 1136 1138 1139 1140 1142 1143 1144 1146 1147 1148
0.85 1140 1142 1143 1144 1146 1147 1149 1150 1151 1153
0.90 1144 1146 1147 1148 1150 1151 1153 1154 1155 1157
0.95 1148 1149 1151 1152 1154 1155 1156 1158 1159 1161
1.00 1152 1153 1155 1156 1157 1159 1160 1162 1163 1164
1.05 1155 1157 1158 1160 1161 1162 1164 1165 1167 1168
1.10 1159 1160 1162 1163 1164 1166 1167 1169 1170 1172
1.15 1162 1163 1165 1166 1168 1169 1171 1172 1174 1175
1.20 1165 1167 1168 1170 1171 1173 1174 1176 1177 1178
1.25 1168 1170 1171 1173 1174 1176 1177 1179 1180 1182
1.30 1174 1176 1177 1179 1180 1182 1183 1185
1.35 1183 1185 1186 1188
1.40
1122-IN-001-0-12 Page 11
Table for 20% (CO + CO2)
% 1900°F 1910°F 1920°F 1930°F 1940°F 1950°F 1960°F 1970°F 1980°F 1990°F
Carbon 1038°C 1043°C 1049°C 1054°C 1060°C 1066°C 1071°C 1077°C 1082°C 1088°C
0.10 1035 1036 1036 1037 1038 1039 1040 1041 1041 1042
0.15 1058 1059 1059 1060 1061 1062 1063 1064 1065 1066
0.20 1074 1075 1076 1077 1078 1079 1080 1081 1083 1084
0.25 1087 1089 1090 1091 1092 1093 1094 1095 1096 1097
0.30 1099 1100 1101 1102 1103 1104 1105 1107 1108 1109
0.35 1108 1109 1111 1112 1113 1114 1115 1117 1118 1119
0.40 1117 1118 1119 1120 1122 1123 1124 1125 1126 1128
0.45 1124 1125 1127 1128 1129 1131 1132 1133 1134 1136
0.50 1131 1132 1134 1135 1136 1138 1139 1140 1141 1143
0.55 1137 1139 1140 1141 1143 1144 1145 1147 1148 1149
0.60 1143 1145 1146 1147 1149 1150 1151 1153 1154 1156
0.65 1149 1150 1152 1153 1154 1156 1157 1158 1160 1161
0.70 1154 1155 1157 1158 1160 1161 1162 1164 1165 1167
0.75 1159 1160 1162 1163 1165 1166 1167 1169 1170 1172
0.80 1164 1165 1167 1168 1169 1171 1172 1174 1175 1177
0.85 1168 1170 1171 1172 1174 1175 1177 1178 1180 1181
0.90 1172 1174 1175 1177 1178 1180 1181 1183 1184 1186
0.95 1177 1178 1179 1181 1182 1184 1185 1187 1188 1190
1.00 1181 1182 1183 1185 1186 1188 1189 1191 1192 1194
1.05 1184 1186 1187 1189 1190 1192 1193 1195 1196 1198
1.10 1188 1190 1191 1193 1194 1196 1197 1199 1200 1202
1.15 1192 1193 1195 1196 1198 1199 1201 1202 1204 1205
1.20 1195 1197 1198 1200 1201 1203 1204 1206 1207 1209
1.25 1198 1200 1202 1203 1205 1206 1208 1209 1211 1212
1.30 1202 1203 1205 1206 1208 1209 1211 1213 1214 1216
1.35 1205 1206 1208 1210 1211 1213 1214 1216 1217 1219
1.40 1208 1210 1211 1213 1214 1216 1218 1219 1221 1222
Page 12 1122-IN-001-0-12
Table 2. Millivolt Readings for Endothermic Atmospheres Generated from Propane
and Containing Approximately 20% (CO + CO2)
NOTES:
1. Carbon absorption and diffusion rates limit the actual level attainable on the steel work surface. (Carburizing time and
temperature and steel composition affect the relationship between atmospheric carbon potential and resultant surface
carbon level.) Specific conditions, such as shorter carburizing cycles, may necessitate controlling at a carbon potential
as much as 0.25 wt. %C higher than required on the work.
2. This chart pertains only to endothermic base atmospheres generated from predominantly propane (i.e., 23% CO).
3. Controlling carbon potential at a level exceeding the carbon content of saturated austentite is discouraged. An
optimum combination of productivity and control might be realized by a boost-diffuse carburizing technique.
1122-IN-001-0-12 Page 13
Table for 23% (CO + CO2)
% 1500°F 1510°F 1520°F 1530°F 1540°F 1550°F 1560°F 1570°F 1580°F 1590°F
Carbon 816°C 821°C 827°C 8832°C 838°C 843°C 849°C 854°C 860°C 886°C
0.10 1009
0.15 1025 1026 1026 1027
0.20 1036 1037 1037 1038 1039 1039 1040
0.25 1044 1045 1046 1046 1047 1048 1048 1049 1050 1050
0.30 1053 1053 1054 1055 1055 1056 1057 1058 1059 1060
0.35 1060 1061 1061 1062 1063 1064 1065 1066 1067 1067
0.40 1066 1067 1068 1069 1070 1071 1072 1073 1074 1074
0.45 1072 1073 1074 1075 1076 1077 1078 1079 1080 1081
0.50 1078 1079 1080 1081 1082 1083 1084 1085 1086 1087
0.55 1083 1084 1085 1086 1087 1088 1089 1090 1091 1092
0.60 1088 1089 1090 1091 1092 1093 1094 1095 1096 1097
0.65 1092 1093 1094 1095 1096 1097 1099 1100 1101 1102
0.70 1096 1097 1098 1100 1101 1102 1103 1104 1105 1106
0.75 1100 1101 1103 1104 1105 1106 1107 1108 1110 1112
0.80 1104 1105 1106 1108 1109 1110 1111 1112 1114 1115
0.85 1108 1109 1110 1111 1113 1114 1115 1116 1117 1119
0.90 1111 1113 1114 1115 1116 1117 1119 1120 1121 1122
0.95 1115 1116 1117 1119 1120 1121 1122 1124 1125 1126
1.00 1118 1119 1121 1122 1123 1124 1126 1127 1128 1129
1.05 1120 1122 1124 1125 1126 1128 1129 1130 1133 1133
1.10 1130 1132 1134 1135 1136
1.15 1138
1.20
1.25
1.30
1.35
1.40
Page 14 1122-IN-001-0-12
Table for 23% (CO + CO2)
% 1700°F 1710°F 1720°F 1730°F 1740°F 1750°F 1760°F 1770°F 1780°F 1790°F
Carbon 927°C 932°C 938°C 943°C 949°C 954°C 960°C 966°C 971°C 977°C
0.10 1014 1014 1015 1016 1016 1017 1017 1018 1019 1019
0.15 1033 1034 1035 1036 1036 1037 1038 1039 1039 1040
0.20 1048 1049 1050 1050 1051 1052 1053 1054 1055 1056
0.25 1060 1061 1061 1062 1063 1064 1065 1066 1067 1068
0.30 1070 1070 1071 1072 1073 1074 1075 1076 1078 1079
0.35 1078 1079 1080 1081 1082 1083 1084 1085 1086 1088
0.40 1086 1087 1088 1089 1090 1091 1092 1093 1094 1096
0.45 1093 1094 1095 1096 1097 1098 1099 1100 1102 1103
0.50 1099 1100 1101 1102 1103 1105 1106 1107 1108 1109
0.55 1105 1106 1107 1108 1109 1111 1112 1113 1114 1115
0.60 1110 1111 1112 1114 1115 1116 1117 1119 1120 1121
0.65 1115 1116 1118 1119 1120 1121 1123 1124 1125 1126
0.70 1120 1121 1122 1124 1125 1126 1127 1129 1130 1131
0.75 1124 1126 1127 1128 1130 1131 1132 1133 1135 1136
0.80 1129 1130 1131 1133 1134 1135 1137 1138 1139 1141
0.85 1133 1134 1135 1137 1138 1139 1141 1142 1143 1145
0.90 1137 1138 1139 1141 1142 1143 1145 1146 1148 1149
0.95 1141 1142 1143 1145 1146 1147 1149 1150 1151 1153
1.00 1144 1146 1147 1148 1150 1151 1152 1154 1155 1157
1.05 1148 1149 1150 1152 1153 1155 1156 1157 1159 1160
1.10 1151 1153 1154 1155 1157 1158 1160 1161 1162 1164
1.15 1154 1156 1157 1159 1160 1162 1163 1164 1166 1167
1.20 1158 1159 1161 1162 1163 1165 1166 1168 1169 1171
1.25 1161 1162 1164 1165 1167 1168 1169 1171 1172 1174
1.30 1165 1167 1168 1170 1171 1173 1174 1176 1177
1.35 1174 1176 1177 1179 1180
1.40 1181 1183
1122-IN-001-0-12 Page 15
Table for 23% (CO + CO2)
% 1900°F 1910°F 1920°F 1930°F 1940°F 1950°F 1960°F 1970°F 1980°F 1990°F
Carbon 1038°C 1043°C 1049°C 1054°C 1060°C 1066°C 1071°C 1077°C 1082°C 1088°C
0.10 1027 1028 1028 1029 1030 1031 1031 1032 1033 1034
0.15 1049 1050 1051 1052 1053 1054 1055 1056 1057 1058
0.20 1066 1067 1068 1069 1070 1071 1072 1073 1074 1075
0.25 1079 1080 1081 1082 1084 1085 1086 1087 1088 1089
0.30 1090 1091 1093 1094 1095 1096 1097 1098 1099 1100
0.35 1100 1101 1102 1103 1105 1106 1107 1108 1109 1110
0.40 1108 1110 1111 1112 1113 1114 1116 1117 1118 1119
0.45 1116 1117 1118 1120 1121 1122 1123 1125 1126 1127
0.50 1123 1124 1125 1127 1128 1129 1130 1132 1133 1134
0.55 1129 1130 1132 1133 1134 1136 1137 1138 1140 1141
0.60 1135 1136 1138 1139 1140 1142 1143 1144 1146 1147
0.65 1141 1142 1143 1145 1146 1147 1149 1150 1151 1153
0.70 1146 1147 1149 1150 1151 1153 1154 1155 1157 1158
0.75 1151 1152 1153 1155 1156 1158 1159 1160 1162 1163
0.80 1155 1157 1158 1160 1161 1162 1164 1165 1167 1168
0.85 1160 1161 1163 1164 1166 1167 1168 1170 1171 1173
0.90 1164 1166 1167 1168 1170 1171 1173 1174 1176 1177
0.95 1168 1170 1171 1173 1174 1175 1177 1178 1180 1181
1.00 1172 1174 1175 1177 1178 1179 1181 1182 1184 1185
1.05 1176 1177 1179 1180 1182 1183 1185 1186 1188 1189
1.10 1180 1181 1183 1184 1186 1187 1189 1190 1192 1193
1.15 1183 1185 1186 1188 1189 1191 1192 1194 1195 1197
1.20 1187 1188 1190 1191 1193 1194 1196 1197 1199 1200
1.25 1190 1192 1193 1195 1196 1198 1199 1201 1202 1204
1.30 1193 1195 1196 1198 1199 1201 1203 1204 1206 1207
1.35 1197 1198 1200 1201 1203 1204 1206 1207 1209 1210
1.40 1200 1201 1203 1204 1206 1207 1209 1211 1212 1214
WARRANTY
This product is warranted according to the General Warranty Statement that is part of the Terms and Conditions of
Sale of the Industrial Instruments Division, Barber-Colman Company.
The warranty for this product is effective for twelve months after shipment from Barber-Colman Company.
Eurotherm/Barber-Colman
741-F Miller Drive, Leesburg, VA 20175
Phone: 703-443-0000
Fax: 703-669-1300
http://www.barber-colman.com http://www.eurotherm.com