Electrosurgical Unit

(100Watt)

User’s Manual
Catalogue
Ⅰ Electrosurgical Unit Front Panel…………………………………………………………………………… - 1 -

Ⅱ Monopolar Cut Setting Panel……………………………………………………………………………… - 2 -

Ⅲ Monopolar Coagulation Setting Panel……………………………………………………………………… - 3 -

Ⅳ Electrosurgical Unit Operating Instructions………………………………………………………………… - 4 -

Ⅴ Safety Instructions…………………………………………………………………………………………… - 4 -

Ⅵ Assembly Diagram………………………………………………………………………………………… - 5 -

Ⅶ Use of Neutral Plate (Skin Plate)…………………………………………………………………………… - 6 -

Ⅷ Common Troubles and Troubleshooting of the Electrosurgical Unit……………………………………… - 9 -


Electrosurgical Unit User Manual

Ⅰ. ELECTROSURGICAL UNIT FRONT PANEL

1 2

3 4

5
7 6

Figure 1-1

Description:
1: Cutting indicator
2: Coagulating indicator 5: Power switch
3: Alarm indicator 6: Neutral pad socket
4: Power indicator 7: Monopolar surgical pencil socket

-1-
Electrosurgical Unit User Manual

Ⅱ. Monopolar Cut Setting Panel

2.1

2.2
2.3
2.4

2.6 2.5

Figure 2-1

(All the parameter settings for the cutting mode is within the above panel)

2.1 Cutting indicator which means the is working in the cutting mode when it is on
2.2 button of Pure Cut
2.3 Cutting power indicator which shows the setting power of cutting mode
2.4 Button of Blend Cut
2.5 Cutting power adjuster which allows you turn up or down the cutting power between 0W and 100W for Pure
cut mode. (0W to 70W for Blend cut mode)
2.6 Cutting mode switching button which allows you to switch among Pure Cut and Blend Cut

-2-
Electrosurgical Unit User Manual

Ⅲ. Monopolar Coagulation Setting Panel

3.1

3.2
3.3
3.4

3.6 3.5

Figure 3-1
(All the parameter settings for the coagulation mode is within the above panel)

3.1 Coagulating indicator which means there is working in the coagulation mode when it is on
3.2 Button of Standard coagulation
3.3 Coagulating power indicator which shows the setting power of coagulation mode
3.4 Button of Spray coagulation
3.5 Coagulation power adjuster which allows you to turn up or down the coagulating power between 0W and
70W for Standard coagulation mode. (0 to 50W for Spray coagulation mode )
3.6 Coagulation mode switching button which allows you to switch between Standard Coagulation and Spray
Coagulation
(Spray Coagulation is more effective than Standard Coagulation)

-3-
Electrosurgical Unit User Manual

Ⅳ. Electrosurgical Unit Operating Instructions

1 Connect the accessories

Please connect the neutral electrode cable, neutral pad, applied surgical pencil, bipolar forceps
and footswitch according to the surgery need.

2 Power up

Turn on the main generator and it will be ready after the “Beep”.

3 Mode select and power settings

Please select the operating mode and the according power set.

4 Perform the surgery

You can activate the cutting mode by stepping on the yellow pad of the footswitch or pressing
the yellow handswitch on the surgery pencil;
You can activate the coagulation mode by stepping on the blue pad of the footswitch or pressing
the blue handswitch on the surgery pencil;
Only can you activate the bipolar mode by stepping on the blue pad of the footswitch.

Ⅴ. Safety Instructions

1. Patient Isolation:The rubber mattress normally is not good enough for the isolation between the
patient and the operating table. It is strongly recommended to apply a piece of dried cloth
between the rubber mattress and the patient. Meanwhile, since the dried cloth can easily get wet
during the operation, a waterproof layer shall be applied over the dried cloth.Last but not the
least, there shall be dried cloth between the limbs to prevent direct contact.
2. Heart Pacemaker: Since the electrosurgical unit can interrupt even damage the heart pacemaker,
Please consult the cardiology department before electrosurgery if the patient is equipped with a
heart pacemaker.
3.Explosion hazard: The anesthetic or disinfectant that applied to the patient and the electrosurgical
unit must be inflammable and inexplosive, and they must be vaporized or wiped out completely
before turning on the main generator. Besides, please note that the abdominal cavity and the
internal organs can also produce flammable or explosive gas.
4 Active electrodes: It is mandatory to put the electrodes in a safe place away from the patient
when they are not needed so potential accident like mistriggering can be avoided.

-4-
Electrosurgical Unit User Manual

Ⅵ. Assembly Diagram
1 Connection of Monopole Mode
1) Connect the foot switch as Figure 6-1:

Figure 6-1

2) Connect the neutral plate (skin plate) as Figure 6-2:

Figure 6-2

-5-
Electrosurgical Unit User Manual

3) Connect the Electrosurgical pencil as Figure 6-3:

Figure 6-3

Ⅶ. Use of Neutral Plate (Skin Plate)


1 Category and Name of Neutral Plate
1) Overview

Neutral plate is the electrode of monopole suspended output of the machine. It should be used properly

on the human body to form a circle circuit with the electrosurgical pencil to cut and coagulate for

biological tissue.

The common neutral plates show as Figure 7-1/7-2, whose middle part is composed of metal foil.

2) Category

According to times of use: Disposable Neutral Plate and Reusable Neutral Plate.

According to structural use: Monopole Neutral Plate and Bipolar Neutral Plate.

Monopole Neutral Plate is as Figure 7-1.

Bipolar Neutral Plate is as Figure 7-2.

-6-
Electrosurgical Unit User Manual

Figure 7-1 Monopole Neutral Plate

Figure 7-2 Bipolar Neutral Plate

-7-
Electrosurgical Unit User Manual

3) Name

A> The neutral plate can be called neutral electrode, muscle plate or skin plate.

B> The monopole plate is short for monopole neutral plate, generally named single-piece skin plate,

or named disfissured skin plate.

C> The bipolar plate is short for bipolar neutral plate; general named double-piece skin plate, or

named fissured skin plate.


2 Use of Monopole Neutral Plate (single-piece, disfissured)
Stick the monopole neutral plate smoothly to the body which is clean, hair-off, muscles rich, no bones

prominent and near the surgery area, then use the plate plywood to clamp the plate handle(as Figure 7-1)

and start the machine, the machine will work normally.


3 Use of Bipolar Neutral Plate (double-piece, fissured)
Use bipolar neutral plate the same as using the monopole neutral plate. Stick the monopole neutral plate

smoothly to the body which is clean, hair-off, muscles rich, and no bones prominent and near the surgery

area, then use the plate plywood to clamp the plate handle and start the machine, the machine will work

normally.

However, if the plate has not been snapped to the patient or the patient’s skin is too dry, it will increase

impedance between the human body and the plate, then the machine will alarm.

If the plate has not been attached to the patient, the machine will also alarm, and the correct methods

above described should be followed to discharge the alarm.


4 Recommend Bipolar Neutral Plate

When the impedance between the plate and the body reaches a certain degree because of some

reasons such as looseness or dryness, the human tissue attached to the plate will be burned to cause the

serious medical accident. However, since the bipolar plate can detect impedance between the human

body and the plate, and when using this plate the machine can also detect related information if the

impedance is too high, then it will stop working and start alarming to protect the patients.

Since the monopole plate doesn’t have the function above, the use of bipolar plate is recommended.

Statement: According to the criterion (6.8.2), if impedance between the bipolar neutral plate and the

patient is over 140Ω, the machine will work abnormally with sound and light alarm.

Warning: If not using the compatible and monitored neutral electrode, the contact quality monitoring

equipment won’t produce audible alarm when safe contact between the electrode and patient is lost.

-8-
Electrosurgical Unit User Manual

Ⅷ. Common Troubles and Troubleshooting of the Electrosurgical Unit


Table 8-1
No Error Code Code Meanings Troubleshooting

1. Check and make sure good contact between


the generator and the neutral pad, so it is with
1 E01 Neutral Pad connection problems
the neutral pad and the patient.
2. Replace the entire set of neutral pad.
Loose the handswitch and the footswitch for a
2 E02 Short circuit or overload protection few seconds and the alarm will turn off after no
output for a few seconds.

Loose the handswitch and the footswitch for a


Main generator has been working continuously
3 E03 few seconds and the alarm will turn off after no
for over 30s or misoperating that long.
output for a few seconds.

Shut off the generator and reboot, if the failure


4 E04 Over-voltage protection of the switch power
remains, please contact the manufacture.
Shut off the generator and reboot, if the failure
5 E05 Lose control of power output
remains, please contact the manufacture.

6 E06 Surgical pencil errors Check or Replace it.

7 E07 Footswitch errors Check or Replace it.

Shut off the generator and reboot, if the failure


8 E08 No output of the switch power
remains, please contact the manufacture.

-9-