Safety Information
4 | Safety Information
Overview of Smart Range Systems Upload
The Smart Uploader V3.0 is a computer program that must be used when performing one
of the following cases on a Smart range Direct-Q, RiOs-DI, RiOs, Synergy, Simplicity or
ICW-3000 system.
Case 1 Case 2
PC Board Replacement Software Version Upgrade
Important: A new PC Board does not have any software or calibration data on it.
Note: The LWD Test Bench Manager Database is very large. It is advisable not to
replicate a local copy but instead download a calibration file using a network
connection only when required.
Tools Needed:
1. A Computer (with Windows XP or Windows 7).
2. An RS232 Serial Port on the Computer (or a validated RS232/USB Adaptor, refer to the
How to Order a Dell Docking Station on page 14 or Other RS232 Adaptors on page
16 of this guide.
3. The Smart Uploader Cable - FTPF07908.
The Smart Uploader V3.0 Software is required for Smart Range System uploads.
Procedure
1. To Log On Using Your SADM Account.
a) Press CTRL + ALT + DELETE and select Switch User.
Note: Make sure you have saved all work before commencing.
The Smart Uploader V3 is now installed and a shortcut will have been placed on your
desktop.
What to do next
Log off from your SADM Account and log back on using your normal account credentials.
A Software file is needed in all cases for Upload and is found on the Field Service Support
Database.
Procedure
1. Go to the Field Service Support Database.
2. In the Field Service Support Database navigate to Software > Smart Platform >
System Software and select the correct system type.
In the example above software version V3.80 for a Direct-Q or RiOs-DI has been
selected.
3. Double clik on the Software file (.s19) on the database and select Save.
4. Choose a location to save the file on your laptop.
Results
The Software File has been saved to your Laptop.
Calibration files can be downloaded from the LWD Test Bench Manager Database.
Procedure
1. Go to the LWD Test Bench Manager Database.
Note: If you do not have the icon on your Lotus Notes Wokspace, then go to: File
> Open > Lotus Notes Application. In the Open Aplication Pop-up, change the
Look in: to Strasbourg01/EU/Millipore or WWApps01/NA/Millipore and from the
list select LWD Test Bench Manager Prod.nsf.
A connection between the laptop and the Smart Range System through the Smart
Uploader Cable is needed to perform an upload.
Procedure
1. Unplug the power cable of the system (behind the system).
Important: Make sure the cable is not plugged in to a source of electric power.
2. Locate and unscrew the 4 screws at the back of the system. Remove the back cover so
you can access the inside of the system.
3. Locate the 4-pin Molex output on the PC Board. This is located on the left side of the
power entry module and to the right side of the tank level sensor cable.
Results
The Smart Uploader Cable is now connected.
Important: DO NOT power on the system yet.
Extension Lead
Docking Station
Extension Lead
Procedure
1. Install adaptor's driver as per instructions.
2. Open Control Panel.
Note: If you are not logged in using your SADM account you will get a message
to say you are logged in as a standard user. If you need to make changes to
settings you must be logged on with Administrator rights.
4. Expand Ports.
5. Check that the RS232/USB adaptor is associated with COM1 and the name of the
adaptor on the installation CD or on the adaptor's box, compares with the name in
Ports. If not, please refer to the Troubleshooting section
6. If the adaptor still cannot be found, re-install the adaptor's driver or contact your local IT
support.
Procedure for upgrading Smart software or uploading software to a new Smart system
PCB.
Important: If you are uploading software to a new Smart system PCB you will also
have to upload the calibration file.
Before performing the software upload you must have:
• Installed the Smart Uploader V3.0 tool on your computer
• Obtained the corresponding Software File from the LW FSS Database
• Obtained the Calibration File from the LWD Test Bench Manager Database (if uploading
software to a new PCB)
• Connected the Smart Uploader Cable between the system and the computer.
Important: Make sure your laptop is connected to the water system before running
the Smart Uplaoder. If not you will get the following error message.
Procedure
1. Start the Smart Uploader V3.0.
2. Plug the system power cord into the power outlet.
18 | Uploading Software
3. Click on Connect Device.
4. Go to the File Identification part of the uploader tool and click on Load File.
Uploading Software | 19
5. Select the system software file (file type.s19) and click on Open.
Note: This is the Software file copied from the Field Service Support Database.
The File name section of the Smart programmer tool becomes populated with the file
location information.
6. Click on Device Configuration.
20 | Uploading Software
7. Select your system type and check the UV present box if your system has a UV. Cick
Ok.
The Smart configuration pop-up box closes and the 'Smart Model' box in the Smart
Programmer is automatically populated.
Uploading Software | 21
9. The Smart Uploader V3.0 will start to program the PC Board.
Results
The upload is completed and a pop-up is displayed.
What to do next
Important: If you are replacing the PCB then DO NOT disconnect. Proceed to follow
the instructions for uploading the calibration file.
If you are just uploading a new version of software onto an existing PCB then click
'Disconnect Device'.
22 | Uploading Software
Uploading Calibration File
Procedure for loading a Calibration File onto a new Smart system PCB.
Procedure
1. In the Smart Uploader V3.0 click on Load File.
3. Click on Open.
The file that you choose from your computer is now listed in the Smart Uploader.
4. Click on Upload Config.
Results
The Water System restarts automatically. The Water System is now ready for use. The
upload of the calibration file is done.
1 = Direct-Q
2 = RiOs-DI
3 = RiOs
1 = UV
Value Range 1 4 = AFS
0 = no UV
5 = Synergy and
ICW-3000
6 = Simplicity
Procedure
1. Log on as an Administrator using your SADM account, as described in the Installing
Smart Uploader Software section.
2. Open the "Device Manager" as described in the Other RS232 Adaptors section.
3. In "Ports", right click on the RS232/USB port
4. Select Properties
5. Select the Port Settings tab
Note: This is also where the speed settings can be modified.
6. Click on Advanced.
7. Select COM1 in the COM Port number list, even if COM1 is already used.
8. Close the different windows with OK buttons.
Smart Range System catalogue numbers are needed for calibration file download.
The different catalogue numbers are listed below:
System Catalogue Number
Direct Q ZRQS0P030
Direct Q UV ZRQSVP030
RiOs ZR0S0P030
RiOs DI ZRDS0P030
RiOs DI UV ZRDSVP030
RiOs DI Clinical ZRDSVCL30
Simplicity SIMS00000
Simplicity UV SIMSV0000
Synergy SYNS00000
Synergy UV SYNSV0000
ICW-3000 SYNSUV0IC