Professional Documents
Culture Documents
Ee2092d Emsl Expts
Ee2092d Emsl Expts
Ee2092d Emsl Expts
ELECTRICAL MEASUREMENTS
LABORATORY MANUAL-II
APRIL 2019
1
CYCLE OF EXPERIMENTS (4th Semester)
2
Exp No: 1
Date:………….
AIM
APPARATUS REQUIRED
Sl. Name Rating Quantity
No
1 Ammeter (0-2A) PMMC 1
2 Rheostat 300Ω.1.7A 1
50Ω.5A 1
3 Fuse 2A, 2
5A 1
4 DPST 1
5 Iron ring 1
7 Autotransformer 220V/30V 1
8 Resistance box 1
9 Capacitor 2µF 1
10 CRO 1
11 Flux meter 1
THEORY
Magnetic hysteresis may be defined as the lagging of magnetization or induced flux density
(B) behind the magnetizing force (H). Alternatively it may be defined as that quantity of a
magnetic substance due to which energy is dissipated in it at the reversal of its magnetism.
From the B-H curve of a magnetic material it can be seen that B always lags behind H.
Furthermore, the area of the hysteresis loops gives the net energy dissipated in taking the
magnetic material from through one cycle of magnetization. Now if the magnetic material has
to be magnetized again, it will trace the same hysteresis loop. In fact, when H is varied a
number of times the size of loop become smaller until the material is cyclically magnetized.
3
Measurement of flux density in ring specimen
4
PROCEDURE
5
Diameter of the ring =
Diameter of the iron ring specimen=
Length of the ring=
Sl. Magnetic Flux meter reading NΔФ Ф= H= B=Ф/ B0 μr
No current initial final Range NΔФ/Ns Np I/l A = =
(I) (N) (wb) μ0 B/
H B0
RESULT
1. The B-H µr – H and µr – B curves were determined for given iron ring specimen.
2. The B-H curve is obtained on CRO
6
Exp No: 2
Date:………….
AIM:
Calibrate the magnetic flowmeter using standard air core, search coil and Hibbert‟s magnetic
standard.
APPARATUS:
Ammeter PMMC(0-10) 1
Rheostat 15 ohm,10A 1
Fuse 10A 2
Iron ring 1
Reversing switch 1
Switch 1
Flux meter 1
Dc supply 1
PROCEDURE:
7
3. By knowing the No .of turns of the search coil, the true value of the flux
density can be found out.
4. Calculate error and correction.
2. Observe the deflection on the flux meter when the coil of the Hibbert‟s
magnetic standard falls down in magnetic field.
3. Calculate the error and correction.
8
OBSERAVTION TABLE:IRON RING
9
AIR CORED SOLENOID OBSERAVTION TABLE:
10
INFERENCE:
Experimental readings shows deviation from the true values contributing to errors.
These errors may be due to instrumental errors and parallax errors. From the graphs,
we see that the calibration curves are plotted as piecewise linear curves.
RESULT:
Hence magnetic flux meter is calibrated using standard air core, search coil
and Hibbert‟s magnetic standard. The calibration curves are plotted on the graph
using measured flux on X-axis and percentage error on Y-axis and variations are
observed.
11
Exp No:3
Date:………….
b) To measure the value of given unknown low resistance using Kelvin‟s double
bridge.
APPARATUS REQUIRED:
2. Wheatstone bridge 1
4. Galvanometer 1
THEORY: The Wheatstone bridge is an instrument for making comparisons and finding an
unknown resistance by operating upon a null deflection principle. This means deflection is
independent of the calibration of the null indicating instrument or any of its characteristics.
For this reason very high degree of accuracy can be achieved using Wheatstone bridge. It has
four resistance arms, having resistances P, Q, R and S with a source e.m.f. and a null detector,
usually a galvanometer. The bridge is said to be balanced when there is no current through
the galvanometer. At balanced condition P/Q=R/S
i.e. R= PS/Q
12
The Kelvin‟s double bridge is a modification of the Wheatstone bridge and provides
increased accuracy in measurement of low value resistances. It incorporates the idea of a
second set of ratio arms, hence the name double bridge. When the ratio P/Q =p/q;
13
Observation table:
Wheatstone bridge
14
PROCEDURE:
Ig =
Where,
Ig = the out of balance current in amps.
V = voltage applied to the circuit.
n = Q/P= S/R at balance
4. Study the Kelvin‟s double bridge. Note down the connection diagram given on the
cover of the equipment. Measure the different types of low resistances i.e. (a)
Standard 4 terminal low resistance specimen, (b) Domestic wiring cable.
15
Ans. Bridge methods gives accurate measurements. It usually makes reading at
null condition when the internal parameter have very little offered as the
measurement.
4. If you are asked to measure the resistance of a coil of wire in house wiring
which one will you choose and why?
Ans. Kelvin‟s double bridge should be used since the resistance of coil wires is
very low. Other methods will give inaccurate readings due to comparable contact
resistance.
INFERENCE
Wheatstone bridge cannot measure low resistance with accuracy. So, Kelvin‟s
double bridge is used.
RESULT
The value of the medium resistance measured by Wheatstone bridge is…………Ω
The value of low resistance measured by the Kelvin‟s double bridge is………... Ω
16
Exp No:4
Date:………….
THEORY:
Current transformer and potential transformer are used to measure high currents and high
voltages respectively at power stations and transmission lines. They can be used to extend a
wattmeter range also. Let a current transformer has a nominal ratio of and a PT has a
nominal ratio of .
= and =
PROCEDURE:
1. Use suitable current and potential transformers such that the given 5A, 125V
wattmeter is able to measure up to 10A, 250V.
2. Calibrate the wattmeter whose range has been extended using a standard wattmeter
rating 10A, 250V for resistive loads.
17
3. Repeat the second step for inductive load.
4. Plot the graph of the extended reading versus correction percentage for resistive load
and inductive load.
CIRCUIT DIAGRAM:
18
3. What are the special features incorporated in an electro dynamometer wattmeter to
make it a low power factor type of wattmeter?
Ans:
a. Pressure coil current: The pressure coil circuit is designed to have a low value of
resistance so that the current flowing through it is increased to give an increased
operating torque.
b. Compensation for pressure coil current: The power being measured in a low
power factor circuit is small and current is high. We can use compensating coils
which can be connected in series with the pressure coil circuit and is made as
nearly as possible identical and coincident with the current coil. It is so connected
that it opposes the field of the current coil.
c. Compensation for inductance of pressure coils: The error is Now
with LPF the value of Φ is large and therefore the error is correspondingly large.
Therefore, in a LPF wattmeter we must compensate for the error caused by the
inductance of pressure coil.
d. Small control torque: LPF wattmeter are designed with a small control torque
arrangement. So that they give full scale deflection for power factor as low.
4. Explain true power, reactive power, apparent power and power factor?
Ans:
a. True power: It is that power which is due to the voltage and current that are in phase.
c. Apparent power: It is the product of voltage and current. It is also called the total
power.
5. What are the instrumental transformers? How do they differ from power
transformers?
Instrument transformers are used in conjunction with measuring instruments for
measuring high currents and high voltages where the measuring circuit is emulated with
the power circuit. In construction of instrument transformers emphasis is given more to
accuracy than to efficiency as in power transformers. In this sense the effect of secondary
burden is reduced considerably in instrument transformers.
6. You are supplied with a wattmeter, the current coil of which can take a maximum
current of 5A and whose voltage coil can take a maximum voltage of 150V. you are
required to measure the total power of 10Kw in a 250V single phase supply at 0.8 pf.
What additional equipment you need? Connect them in proper circuit and find out the
multiplying factor which has to be used?
19
Total power= 10Kw
True power= 10Kw*0.8= 8Kw
Current =
Multiplying factor =
OBSERVATION TABLE:
S.no Wattmeter Corrected Standard %
reading ( wattmeter wattmeter correction
in watts reading reading
* 100
( in watts ( in
watts
Resistive 1
load 2
3
4
5
Inductive 1
load 2
3
4
5
RESULT:
INFERENCE:
20
Exp No: 5
Date:………….
Apparatus required:
1. Rheostat 500 Ω, 1.2 A 1
2. Voltmeter (0-150) V PMMC, (0-30) V PMMC 1
3. Decade Resistance Box 1
4. DC Source (0-220) V 1
5. DPST switch 1
6. Fuse 1
Circuit Diagram:
Procedure:
Calculate the value of resistance to be set in the decade resistance box so that the
given (0-30) V voltmeter can be used to measure voltage up to 150 V.
Verify result experimentally.
Calibrate the voltmeter whose range has been extended using a standard (0-150) V
voltmeter.
Plot the graph with extended reading on X-axis and (a) actual reading (b) %
correction on Y-axis.
21
II. Aim:
Conversion of (0-20) mA ammeter into voltmeter to measure voltage up to 150 V and
calibrate the same thing using a standard (0-150) V voltmeter.
Apparatus required:
1. Rheostat 500 Ω, 1.2 A 1
2. Voltmeter (0-150) V PMMC 1
3. Ammeter (0-10) mA PMMC 1
4. Decade Resistance Box 1
5. DC Source (0-220) V 1
6. DPST switch 1
7. Fuse 1
Circuit Diagram:
Procedure:
Calculate the value of resistance to be set in the decade resistance box so that the
given milliammeter can be used to measure voltage up to 150 V.
Verify result experimentally.
Calibrate the milliammeter converted to voltmeter using a standard (0-150) V
voltmeter.
Plot the graph with extended reading on X-axis and (a) actual reading (b) %
correction on Y-axis.
All the points are to be connected by straight lines.
Post -experimental report:
1. Explain absolute instruments.
2. Explain the use of shunt as a device for extending the range of ammeter.
22
3. An MC ammeter has a full scale deflection of 50 µA and a coil resistance of 100 Ω.
What will be the value of shunt resistance required for the instrument to be converted
to read full scale reading of 1 A?
4. A moving coil ammeter gives full scale deflection with 15 mA and has a resistance of
5 Ω. Calculate the resistance to be connected in
(a) Parallel to the instrument to read up to 1 A.
(b) Series to enable it to read up to 10 V.
5. Explain secondary instrument.
6. What is hot-wire ammeter?
7. What is moving coil galvanometer?
Observation Table
Extension of (0-30) V voltmeter to read up to (0-150) V:
(0-30) V Extended Actual Voltmeter % Error % Correction
Voltmeter Reading Reading (V)
Voltmeter Reading(V) (V)
Result:
Extension of range of voltmeter and conversion of ammeter into voltmeter using standard
voltmeter was performed.
23
Exp No:6
Date:………….
LLOYD FISCHER SQUARE
Aim:
To setup a circuit to determine the iron loss/kg of the given steel at various values of
minimum flux densities
1. Ammeter (0-15) A, MI 1
2. Voltmeter (0-60) V, MI 1
3. Rheostat 15 Ω/ 10 A 1
4. Wattmeter 125 V/ 5 A, UPF 1
5. Auto transformer 230 V / 15 A 1
6. Lloyd fischer square 1
Procedure:
1. What are the uses of magnetic squares and what are they? What are the different
arrangements of stamping‟s in such squares? Explain with the help of neat sketches?
2. What are the loss curves?
3. Describe briefly, how the eddy current and hysteresis losses are separated in a sample
of magnetic material?
4. What is the use of allowance of corner pieces in Lloyd fischer square unlike Epstein
square?
5. In this experiment why do you connect current coil of the wattmeter on primary
winding and pressure coil on secondary winding?
24
Tabular column
Circuit Diagram:
A
260T 80T
(0-15)A
160T V (0-60)V
15 Ω /10A
M C
P
15 A
230 V,50 Hz, Single phase
230 V/15 A
AC supply
DPST
Auto transformer
N 15 A
Result:
Iron loss/kg of the given iron specimen/steel stampings at various values of maximum flux
densities were obtained and tabulated. Graph between iron loss/kg Vs maximum flux
densities were plotted
25
Exp No: 7
Date:………….
1. To calibrate the given single phase energy meter using standard wattmeter by direct
loading.
Apparatus required:
Quantity Specifications
1 Ammeter (0-5) A, MI 1
2 Voltmeter (0-250) V, MI 1
3 Single phase variable resistance load 15 Ω/ 5 A 1
4 Wattmeter 250 V/ 5 A, UPF 1
5 Single phase Auto transformer 270 V / 15 A 1
6 Three phase Auto transformer 1
7 Stop watch 1
Principle
In order to check the calibration of a single phase energy meter, the reading of the energy
meter is compared with that of a standard instrument. For determining the true energy
consumption, a standard wattmeter and an accurate stopwatch is used. From the calculated
true power, the error and the percentage error in the energy meter reading is determined. In
direct loading, the current coils of the energy meter and wattmeter are connected to a single
phase supply in series with the loading device (say rheostat) whereas the pressure coils are
connected directly to the supply. The loading device is adjusted to get the required current.
Then the energy consumption is determined by observing the time for a fixed number (say N)
of revolutions. The true energy is calculated from the wattmeter reading and the time
indicated by the stopwatch.
26
Fig. Direct loading
Procedure
The connections are done as shown in the circuit diagram. Adjust the auto transformer to
minimum position. Supply is switched on and rated voltage (specified on energy meter) is
applied by varying autotransformer. Remove the loads completely (i.e. keep the loading
rheostat in maximum resistance position or if you are using single phase resistive load then
switch off all the loads in it). Now see if the disc is rotating without any load, if it is rotating
then it is having creeping error. Note down the time taken for 1 revolution. The current is
varied using loading rheostat till the rated current (specified on energy meter) step by step.
The ammeter reading, voltmeter reading, wattmeter reading and time for 5 revolutions of
energy meter disc are noted in each step. Remove all the loads step by step. Adjust the auto
transformer to minimum position and switch off supply. Plot the graph with indicated power
on X-axis and (a) true power (b) % correction on Y-axis.
DIRECT LOADING
Unity
27
*W1-Actual power (Wattmeter), W2-Indicated power (Energy meter)
Note:
kWh =
2. To calibrate the given single phase energy meter using standard wattmeter by
phantom loading without using phase shifting transformer
28
Principle
When a energy meter is designed for high current loads, it is uneconomical to arrange such
loads for testing purposes as it involves a considerable waste of time and power. To avoid
this problem "phantom” loading is done. In phantom loading, pressure coil is excited from
normal supply voltage and current coil is excited from a separate low voltage supply. The low
impedance of current coil circuit makes it possible to circulate the required current even with
low supply voltage.
Procedure
For testing energy meter at UPF condition connections are done as shown in fig 1. Keep the
autotransformer position in minimum and loading rheostat position in maximum. Supply is
given and apply rated voltage (mentioned on energy meter) across pressure coil of energy
meter and wattmeter. Current in the current coil of the circuit is adjusted by varying auto
transformer. First adjust auto transformer to low value of current (say 1A) and increase the
current to rated current (mentioned on energy meter) (say up to 5A). The voltmeter, ammeter,
wattmeter and time for 5 revolutions of energy meter are noted for various load currents.
Note: For this connection, you should not use same three phase autotransformer for both
current coil and pressure coil.
For a power factor of 0.5, connections are made as shown in figure 2. Keep the
autotransformer position in minimum. The current coil of wattmeter and energy meter is
connected in series to R-phase and pressure coil to Y-phase for lag and B-phase for lead.
Switch on the supply and apply rated voltage (mentioned on energy meter) across pressure
coil of energy meter and wattmeter. If the wattmeter reads negative the pressure coil
connections are interchanged. Current in the current coil of the circuit is adjusted by varying
auto transformer. First adjust auto transformer to low value of current (say 1A) and increase
the current to rated current (mentioned on energy meter) (say up to 5A). The voltmeter,
ammeter, wattmeter and time for 5 revolutions of energy meter are noted for various loads
current. Plot indicated power vs actual power. Also plot the indicated power vs % correction.
29
PHANTOM LOADING
Unity
0.5
lag
0.5
lead
PRE-EXPERIMENT QUIZ
a. Aluminium is preferred to copper in making the rotating disc in energy meter. Why?
b. What is the purpose of providing copper shading bands on the central limb of the core
on which the pressure coil is wound in induction type energy meters?
c. What do you mean by inductive load adjustment? What are the methods to do this?
Briefly explain.
d. What is meant by „creeping‟ in energy meters? How is it avoided?
e. What do you mean by „phantom‟ loading and why is it resorted to?
f. Draw a neat phasor diagram to suit this loading situation in your experiment?
Answer the following Questions:
1. Is it possible to reverse the direction of rotation of the disc in energy meter? If yes, how
?
2. If the induction type energy meter runs fast, what is to be done to slow it down?
3. After conducting this experiment, will you be able to check whether the energy meter in
your house is reading correctly, you can use only a stop watch and no additional meters
can be connected (at least approximately).
4. Is it possible to calibrate the energy meter at 0.5 lead and lag? If yes, how?
5. Even in phantom loading, a resistance is connected in series with the current coil. Is it
necessary for the conduct of the experiment?
Result:
30
Exp No: 8
Date:………….
Apparatus required:
Quantity Specifications
1. Ammeter (0-5) A, MI 1
2. Voltmeter (0-500) V, MI 1
3. Three phase variable resistance load 15 Ω/ 10 A 1
4. Dynamometer type Wattmeter UPF 2
5. Stop watch Digital 1
6. Three phase energy meter Induction 1
Procedure:
1. Choose a three phase energy meter, note down its rating and choose apparatus
accordingly.
2. Connect the circuit as shown in the figure.
3. Initially keep the three phase autotransformer at the minimum potential position and
turn off all the loads in three phase variable resistance load.
4. Switch on the supply and gradually increase the supply by varying three phase
autotransformer till voltmeter reads rated voltage (specified on the three phase energy
meter).
31
5. Now see the disc, if it is rotating without turning on any load then it is having
creeping error. Note down the time taken for 1 revolution.
6. Gradually increase the load in steps up to the rated current (indicated on the three
phase energy meter) and note down the readings of voltmeter, ammeter, wattmeters,
and time taken for 5 revolutions in each step.
7. Switch off all the loads in three phase variable resistance load, bring back the three
phase autotransformer to its minimum potential position, and switch off the supply.
8. Convert the energy meter reading into indicated power in watts and plot indicated
power vs true power. Also plot the indicated reading in watts Vs % error.
Observation Table
P Voltmeter Ammeter Wattmeter Wattmeter True power Time for 5 Indicated % Error
F WT =W1 + W2 revolutions
Reading Reading Reading Reading Power *100
(V) (W) t (Sec)
(A) W1 (W) W2 (W) WI (W)
U
N
I
T
Y
Note:
kWh =
1. How is compensation for static friction provided in induction type energy meters?
32
3. Explain the necessity and principle of creep test at higher voltage?
4. Explain clearly the adjustments to be made on a multi element meter used on three phase
energy circuit?
1. Is it necessary that the meter disc should rotate in normal direction for energy to be
recorded correctly?
2. In many houses, three phase supply is provided even though there are no three phase loads
in use. Why?
3. When three phase supply is used in a domestic circuit, is it better to use three single phase
energy meter instead of a three phase energy meter?
4. Is it possible to calibrate the energy meter at 0.866 lag and 0.866 lead. If yes, how?
5. Study the various types of error adjustments provided in the given three phase energy
meter.
Results
33
Exp No: 9
Date:………….
Apparatus required:
1. Iron ring specimen
2. Ammeter
3. Rheostat
4. DPST Switch
5. Fluxmeter
6. DC source
7. Fuse
Circuit Diagram:
Procedure:
Make connections as shown in the figure.
Demagnetise the iron specimen.
The flux meter is reset and a current is passed through a coil . The fluxmeter reading
is noted and current is changed to i2 in the same direction.
34
After fluxmeter is reset after a series of demagnetising method of opening and
reversing DPST switch,the current is taken through following series without any
fluxmeter resetting in between fluxmeter reading is noted down
0-1A,1-2A,2-1A,1-0A,0 to -1A, ,-1 to -2A, ,-2 to -1A, ,-1 to 0A, 0-1A,1-2A.
The change in the fluxmeter reading is noted down ,plotted and hysteresis curve is
then obtained by suitable adjustments.
Pre-experimental report:
1. List different types of a flux meter, which type of flux meter are you using in the
laboratory?
2. How will you extend the range of a flux meter?
3. What is linkage factor?
4. What is a magnetic potentiometer?
5. What is permeameter, what is its importance?
6. Discuss how a flux meter is differ from a moving coil ammeter or galvanometer?
7. What is the importance of the area under hysteresis loop?
Tabular column
Sl no. Current Flux meter reading H Φ B
Result:
35
The hysteresis loop of given iron specimen was determined by six point method and
hysteresis loop was plotted .The total energy loss due to hysteresis is ………J
36
Exp No:10
Date:………….
PHOTOMETRY
Aim:
i. To determine the candle power of a given test lamp to various operating voltage up to
230V, using photometric bench.
ii. To determine the polar curves of the candle power distribution in a horizontal plane
passing through the lamp source operating at a voltage of 230V and hence the
determination of MHCP of lamp.
iii. To determine the MSCP of the given test lamp of 230V using Rousseau‟s
construction and hence calibrate the given integrating sphere.
Apparatus Required:
wattmeter (0-250)V 1
DPST switch 1
Fuse 5A 1
Procedure:
1. Two lamps (standard and test) are placed at some distance apart with photometric
head in between them.
2. “S” is standard lamp whose intensity is known and T is the test lamp whose intensity
is to be measured.
3. Photometric head is moved between the lamps till the illumination on the both the
sides is equal.
4. Distances are noted from the photometric bench.
5. For different values of applied voltage null power is measured.
37
To determine MHCP and polar curve:
1. This may be determined for a given source by turning it about a vertical axis and
measuring intensity in the direction of the line joining the source under test and
standard source.
2. The readings are taken after 10 to 15 rotations
3. MHCP may be obtained by plotting intensity against roatation on rectangular axis
4. Mean ordinate of the curve given the value of MHCP gives the value to MSCP.
a) Factory floors
b) Rooms
c) Street lighting
Q3.How do you select light fittings and lamp shaders in domestic lighting? What are the
criteria used?
Q4.On what all factors does illumination from a given lamp depends?
Q9.Define point source, line source, extended source, and solid angle point?
Circuit Diagram:
38
[1] To determine candle power of test lamp
Result
39
Exp No: 11 A
Date:………….
To measure the value of unknown capacitance using Schering bridge and dissipation
factor
APPARATUS REQUIRED
3 Function Generator 1
4 Galvanometer 1
5 Multimeter 1
THEORY
Schering bridge are most popular used these days in industries for measurement of
capacitance, dissipation factor and loss angles.
) =( )( )
j/ =( )( )
40
j/ )( j/ )= )( j/ )
(- / )-( )= / )-
41
PROCEDURE
INFERENCE
Error in the resistor and capacitor values may be because of the resistance of standard
capacitor used.
RESULT
The value of unknown capacitance has been measured using Schering Bridge
42
Exp No: 11 B
Date:………….
AIM
APPARATUS REQUIRED
2 Rheostat 3
3 Function Generator 1
4 Galvanometer 1
5 Multimeter 1
THEORY
[ ]
And
43
The expression for Q factor of the coil is,
L1 – Unknown inductance
R1 – Effective resistance of inductor,
C4 – Variable standard capacitor
R2, R3, R4 – Known non-inductive resistance
Phasor Diagram
PROCEDURE
44
INFERENCE
Error in the inductor and resistor values may be because of the resistance of variable
capacitor used.
RESULT
The value of unknown inductance has been measured using Maxwell‟s capacitance
bridge.
45
Exp No: 12
Date:………….
Aim: To calibrate the given PMMC type ammeter and dynamometer type wattmeter using
vernier potentiometer.
Vernier Potentiometer:
This is a precision instrument for the measurement of voltages in the range of 0 to 1.901 volts
with a least count of 1 microvolt or 0.1 microvolt according to two different models.
Standardization:
1. Connect a freshly charged 2 volts lead accumulator to terminals marked “2 volts”
3. Connect a cadmium standard cell of E.M.F. 1.0186 Volts to pair of terminals marked “STD
CELL” An auxiliary scale is provided to take into consideration temperature variation of
standard cell voltage. Set this scale equal to standard cell E.M.F. at the working temperature.
Please note that standard cells have a negative temperature coefficient of 40 microvolts per
degree centigrade.
5. Standardization is independent of setting of 3 voltage dials and range multiplier used. Set
the range switch at X1, or X0.1, or X0.01. Standardization is same at all the 3 ranges.
6. Get the null point of galvanometer by pressing two galvanometer keys and by adjusting the
3 rheostats namely coarse, medium and fine. First press the coarse galvo key to get the null
point. When the null point approaches, press the fine key only to get the exact null point.
Now standardization is complete. And rheostats should not be disturbed now. But during
measurements, standardization should be checked off and rheostats should be slightly
adjusted if found necessary.
46
Measurements of Voltages:
1. The unknown voltage is connected to either of the test circuits marked 1, 2, 3.
2. The selector switch should be set according to the test circuit used.
4. Check standardization as explained before and do not disturb the rheostat after
standardization.
5. Get the null point by pressing the two galvanometer keys one after another. To get the null
point, adjust three voltage dials only. The choice of voltage range depends on the magnitude
of the unknown voltage.
6. The sum of three voltage dials multiplied by the range multiplier chosen gives the value of
the unknown voltage in volts.
Schematic Diagram:
Apparatus required:
47
Rheostat 50 Ω, 5 A
Ammeter (PMMC) (0-5) A
Standard Resistor 0.1 Ω, 10 A
Procedure:
Ammeter is of PMMC type. Standardize the potentiometer by setting the vernier dials to
marked position corresponding to the emf of the standard cell and adjust the potentiometer
current, until galvanometer shows the null deflection. Keep the current constant through the
entire range of the instrument.
Calculate,
i. % Error
ii. % Correction
Draw the curves with ammeter reading along x-axis and % error along y-axis and actual
reading vs true value plot. All points are to be joined with straight line.
Circuit Diagram:
48
Tabular column
Apparatus required:
Rheostat 3000 W, 5 A
Ammeter (PMMC) (0-5) A
Voltmeter (PMMC) (0-250)V
Wattmeter (dynamometer type) 250 V, 5A
Standard Resistor 0.1 Ω, 10 A
Voltage ratio box Ouptut of 1.5 V
Procedure:
Wattmeter, W of dynamometer type and all other instruments are of PMMC type.
Standardize the potentiometer by setting the vernier dials to marked position corresponding to
the emf of the standard cell and adjust the potentiometer current, until galvanometer shows
the null deflection. Keep the current constant through the entire range of the instrument.
Calculate,
i. % Error
49
ii. % Correction
Draw the curves with ammeter reading along x-axis and % error along y-axis and actual
reading vs true value plot. All points are to be joined with straight line.
50
Circuit Diagram:
Result:
51