KLA-3-TS Operation Manual PDF
KLA-3-TS Operation Manual PDF
KLA-3-TS Operation Manual PDF
Table of Contents
1. Introduction ............................................................................................................... - 1 -
1.1 Koehler's Commitment to Our Customers .........................................................................................- 1 -
1.2 Recommended Resources and Publications.....................................................................................- 1 -
1.3 Instrument Specifications ...................................................................................................................- 1 -
1.5 Delivery Condition and Procedure of Instruments .............................................................................- 2 -
1.6 Assistance..........................................................................................................................................- 2 -
1.7 Packing and Shipment .......................................................................................................................- 2 -
4. Descriptions .............................................................................................................. - 4 -
4.1 Software Features..............................................................................................................................- 4 -
4.2 Measuring Principle............................................................................................................................- 4 -
4.2.1 ASTM D2500 ..................................................................................................................................- 4 -
4.2.2 ASTM D97, IP15, ISO 3016............................................................................................................- 5 -
4.2.3 ASTM D5853, IP441 .......................................................................................................................- 6 -
4.3 Instrument Controls and Connections................................................................................................- 6 -
5. Installation ................................................................................................................. - 6 -
5.1 Installing the Analyzer ........................................................................................................................- 6 -
5.2 Main Software Interface .....................................................................................................................- 7 -
6. Operation................................................................................................................. - 10 -
6.1 Analysis Parameters ........................................................................................................................- 10 -
6.2 Cloud & Pour Point Analysis Step by Step ......................................................................................- 12 -
6.2.1 Cloud Point Analysis .....................................................................................................................- 12 -
6.2.2 Pour Point Analysis.......................................................................................................................- 13 -
6.2.3 Pour Point Analysis of Crude Oils.................................................................................................- 13 -
6.3 Results Browser ...............................................................................................................................- 13 -
8. Troubleshooting...................................................................................................... - 15 -
8.1 Replacement of Bath Heater/ Thermoswitch/PT 100 ......................................................................- 15 -
8.2 Replacement of the sample measuring probe (PT 100) ..................................................................- 16 -
8.3 Adjustments .....................................................................................................................................- 16 -
8.4 Connection Box................................................................................................................................- 17 -
KLA-3-TS-Manual
KLA-3-TS Automatic Cloud and Pour Point Analyzer
Operation and Instruction Manual
Notes ............................................................................................................................. - 19 -
KLA-3-TS-Manual
KLA-3-TS Automatic Cloud and Pour Point Analyzer
Operation and Instruction Manual
KLA-3-TS-Manual -1-
KLA-3-TS Automatic Cloud and Pour Point Analyzer
Operation and Instruction Manual
KLA-3-TS-Manual -2-
KLA-3-TS Automatic Cloud and Pour Point Analyzer
Operation and Instruction Manual
o Total weight.
KLA-3-TS-Manual -3-
KLA-3-TS Automatic Cloud and Pour Point Analyzer
Operation and Instruction Manual
Assembly. Safety is very important during Integrated Panel PC Operation and Instruction
assembly and operation of this instrument. Manual
4. Descriptions
4.1 Software Features
User friendly interface
Windows® based system
Analytical progress and results recorded and
stored
Results Journal is accessible during analysis
Analytical progress can be displayed in both
graphical and numeric form.
Diagnostic panel with access to all
inputs/outputs
Analysis results automatically saved in Excel®
compatible format.
Guided probe calibration in 3 easy steps
3.1 Packing List Scope. This test method covers only petroleum
products which are transparent in layers 40 mm
The instructions for preparing the equipment in thickness, and with a cloud point below 49°C.
assume that the user is aware of the contents of
this document, which lists the warranty Summary of method. The sample is cooled at
conditions and important precautions. a specified rate and examined automatically.
The temperature at which a cloud is first
KLA-3-TS Auto Cloud & Pour Point Analyzer observed at the bottom of the test jar is recorded
115V/230V 50/60 Hz as the cloud point.
Cloud & Pour Point head
Cord cable without plug Significance and use. The cloud point of a
Test jar with silver bottom petroleum product is an index of the lowest
KLA-3-TS Cloud & Pour Point Analyzer with temperature of its utility for certain applications.
KLA-3-TS-Manual -4-
KLA-3-TS Automatic Cloud and Pour Point Analyzer
Operation and Instruction Manual
If the sample reaches 27°C before the detection, the If the first test detects no pouring then the
bath temperature will be controlled at 0°C. analyzer sets an alarm because the selected
If the sample reaches 9°C before the detection, the first tilt test temperature was too low.
bath temperature will be controlled at -17°C.
If the sample reaches - 7°C before the detection, This test procedure will continue every 3°C until
the bath temperature will be controlled at -34°C. the oil in the jar does not flow for 5 seconds
If the sample reaches -24°C before the detection, when tilted in the horizontal position.
the bath temperature will be controlled at -51°C.
If the sample reaches -42°C before the detection, If the oil has not ceased to flow when its
the bath temperature will be controlled at -68°C. temperature has reached 27°C, the bath is
controlled according the following schedule:
4.2.2 ASTM D97, IP15, ISO 3016
Oil is at +27°C, the bath temperature will be
Scope. This test method covers and is intended controlled at 0°C.
for use on any petroleum oil. Oil is at +9°C, the bath temperature will be
controlled at -18°C.
Summary of method. After preliminary heating, Oil is at -6°C, the bath temperature will be
the sample is cooled at a specific rate and controlled at -33°C.
examined at intervals of 3°C and a multiple of Oil is at -24°C, the bath temperature will be
3°C for flow characteristics. The lowest controlled at -51°C.
temperature at which movement of the oil is Oil is at -42°C, the bath temperature will be
observed is recorded as the pour point. controlled at -69°C.
Significance and use. The pour point of a For black oil, cylinder stock and non distillate fuel
petroleum product is an index of the lowest oil, the result is obtained by the procedure
temperature of its utility for certain applications. described above. If required, determine the lower
(minimum) Pour Point by heating the sample while
Test Procedure. stirring to 105°C, pouring it into the jar, and
determining the Pour Point as described.
1. Pour accurately the oil into the test jar to the
level mark. When necessary, heat the oil in
a water bath until it is just sufficiently fluid to
KLA-3-TS-Manual -5-
KLA-3-TS Automatic Cloud and Pour Point Analyzer
Operation and Instruction Manual
4.2.3 ASTM D5853, IP441 Oil is at +9°C, the bath temperature will be
controlled at -18°C.
Scope. This test method is intended for use on Oil is at -6°C, the bath temperature will be
any petroleum oil. controlled at -33°C.
Summary of method. After preliminary heating, For crude oils, the result is obtained by the
the sample is cooled at a specified rate and procedure described above.
examined at intervals of 3°C and a multiple of
3°C for flow characteristics down to -36°C. The
lowest temperature at which movement of the oil 4.3 Instrument Controls and
is observed is recorded as the pour point. Connections
Test Procedure.
KLA-3-TS-Manual -6-
KLA-3-TS Automatic Cloud and Pour Point Analyzer
Operation and Instruction Manual
KLA-3-TS-Manual -7-
KLA-3-TS Automatic Cloud and Pour Point Analyzer
TOUCH THIS FIELD TO CHANGE THEand
Operation OPERATOR’S NAME
Instruction Manual
The operator’s name will be recorded in the results file.
TOUCH THIS
BUTTON TO
START THE
ANALYSIS
TOUCH THIS
TOUCH THIS BUTTON TO
FIELD TO SET ACCESS THE
THE ANALYSIS RESULTS
PARAMETERS JOURNAL
TOUCH THIS
THIS FIELD BUTTON TO
SHOWS THE ACCESS THE
ANALYSIS DIAGNOSTIC
STATUS PANEL
The data grid will not be filled during the initial pre-cooling or
pre-warming of bath or sample, so it will be shown as a blank
field until the “active part” of analysis starts.
A smiley icon may appear for 1 minute each 10: if the analysis is idle, an
automated procedure will cycle the cooling to prevent oil settling in gas
circuit, this will ensure a greater lifetime to the cooling system.
KLA-3-TS-Manual -8-
KLA-3-TS Automatic Cloud and Pour Point Analyzer
Operation and Instruction Manual
The Arm Position field features four button The Back button will close the diagnostic panel.
to control the arm of the analytical head
and it’s position: by selecting a position, the For troubleshooting purpose, while the
near field will change it’s scale so that the diagnostic panel is open, the detection, sample,
value label (showing 1000 in the above and bath fields will also show the probe’s VOLT
picture) will move up and down according signal.
the head’s actual position.
Calibration Panel
The Bath field allows the user to turn on/off
the cooling or heating, note that they
cannot be activated together, so by turning
on the Cooler while the heater is active,
then the heater will be automatically
disengaged and vice-versa; since the
instrument is equipped with a safety cut-off:
despite the Heater status, the heating will
be disabled if the bath reaches a
temperature of 72°C or above.
KLA-3-TS-Manual -9-
KLA-3-TS Automatic Cloud and Pour Point Analyzer
Operation and Instruction Manual
6.1 Analysis Parameters The three buttons on the top left let the user
choose which kind of analysis will be performed.
By touching the main screen’s OPERATOR field In Figure 8, the selected buttons Cloud Point
or the FILE NAME field a standard keyboard will and Pour Point are selected, the Bath
appear on the screen, you can insert the desired Management is defined as prescribed by
text and then push the OK button to confirm. Method and the After completing reheat
sample to (°C) is enabled: by touching the OK
button the analyzer will be ready to perform both
Cloud Point and Pour Point analysis in
complete conformity with official methods.
Cloud Point analysis will be performed, once
Cloud Point has been obtained, the analyzer
automatically switches to Pour Point mode in
order to continue the analysis to obtain the Pour
Point; once both analyses are completed, the
KLA-3-TS-Manual - 10 -
KLA-3-TS Automatic Cloud and Pour Point Analyzer
Operation and Instruction Manual
heater is activated to reheat the sample to the step below the obtained cloud point, by following
specified temperature. normal scheduled sequence.
KLA-3-TS-Manual - 11 -
KLA-3-TS Automatic Cloud and Pour Point Analyzer
Operation and Instruction Manual
The fiber optic is an optical device, the tip should 1. Once the analysis has been setup as
be clean but avoid the removal of sample’s described in previous section, press the
residual by the means of scratching devices Start Analysis button, the panel shown in
such as knives or sand paper, use a mild solvent Figure 11 will pop up and the head will be
instead. moved to the loading position; the user can
now easily remove the test jar, fill it as
Once an analysis has been completed, if the required then refix the test jar in its position.
After Completing Reheat Sample is enabled,
the heater will be activated to reheat the sample
to the selected temperature.
KLA-3-TS-Manual - 12 -
KLA-3-TS Automatic Cloud and Pour Point Analyzer
Operation and Instruction Manual
When the Cloud Point is detected, the analysis Once the analysis results have been obtained,
data is stored, but if the selected method even if the heater is active, the START button
foresees both Cloud Point and Pour Point may be released safely, this will not causes the
analysis, the analysis status show the Cloud loss of analysis results, but if the reheat
Point results and the analyser will perform also procedure is in progress, it will be interrupted.
the Pour Point analysis, an audible alert will be
played once.
Once the analysis is complete, the analyzer will
6.2.2 Pour Point Analysis automatically lift out from the bath the analytical
If the analysis has been set-up as Pour Point head to the vertical position: you can now safely
only and a Preheat sample has been selected, release the “START” button, the analyzer is
the sample will be warmed to the selected ready for the next test.
temperature. To remove the test jar and perform cleaning, you
The sample is cooled as prescribed and the may use the Diagnostic Maintenance panel
Pour Point test is repeated every 3°C (or and move the analytical head to the loading
according the selected “Tilt Out Each”) from the position, if you plan to perform another test, you
start temperature of the analysis. The test may proceed to setup the test and then hit the
procedure will continue every prescribed step Start Analysis button so that the analyzer will
until the oil in the jar doesn't flow for 5 seconds lift the head to the loading position: you can now
when tilted in the horizontal position. remove the test jar and perform cleaning, then
When the Pour Point is detected, the analysis fill the sample and start the analysis as usual,
data is stored. you can also hit the Start Analysis button just to
remove the test jar and perform cleaning then
6.2.3 Pour Point Analysis of Crude Oils use the provided Cancel button.
As prescribed by the method, if the expected
6.3 Results Browser
pour point is below 36°C, the sample will be
heated to 45°C otherwise it will be heated 12°C The result browser may be accessed even while
over the expected pour point but never above the analysis is in progress.
60°C, once the sample has reached the proper
temperature, the analysis begin. The Results browser allows the user to easily
The sample is cooled as prescribed and the browse and display the performed analysis, in
Pour Point test is repeated every 3°C from the the top area you will see the selected month. By
start temperature of the analysis. The test means of the arrow buttons you may select the
procedure will continue every 3°C until the oil in desired month. If an analysis was performed on
the jar doesn't flow for 5 seconds when tilted in a certain day in a month, then the correspondent
the horizontal position. day will turn from grey to black as shown in
Figure 16 below for December 1st, 3rd and 4th.
If the sample’s temperature reaches -36°C
without detection, the analysis will be
interrupted.
When the Pour Point is detected, the analysis
data is stored.
KLA-3-TS-Manual - 13 -
KLA-3-TS Automatic Cloud and Pour Point Analyzer
Operation and Instruction Manual
KLA-3-TS-Manual - 14 -
KLA-3-TS Automatic Cloud and Pour Point Analyzer
Operation and Instruction Manual
Cloud 8. Troubleshooting
4.PT100 11. PT100 Bath
Sample Pour Use our set of specific keys and screw drivers
5.PT100 12. Solenoid Valve Gas for the following procedures:
Detection Pour
6. Test Jar 13. Air 8.1 Replacement of Bath Heater/
Thermoswitch/PT 100
7. Cooling Gas
1. Disconnect the power cable from the bottom
System Description of the analyzer.
1. Sample quantity corresponds with ASTM 2. Disconnect the 12 pin connector and the
D97 and D2500 (± 1 mm height) motor connector from the head. Pull up the
head.
2. Measurement element of the sample
temperature for Cloud Point. 3. Remove the screws from the main cover of
the analyzer.
3. Fiber optic for Cloud Point detection 4. Remove the main cover.
4. Measurement element of the sample 5. Remove the black support from the top of
temperature for Pour Point the cooling bath.
6. Carefully cut the insulation on the back of
5. Measurement element for Pour Point the bath.
detection
WARNING: Cut it slowly and do not touch the
6. Test jar with bottom mirror bath with the knife to avoid to cutting the heater
or wires
7. The analyzer is equipped with a solenoid
valve, on cooling gas, activated by the 7. Remove the insulation.
application software
No. on
Art. Description Article Number
Scheme
Sample PT 100
2 KLA-1S-008-12
(Cloud)
Figure 19: Insulation
3 Double fiber optic
KLA-1S-008-06
(Cloud) 8. Disconnect and remove the defective item
Sample PT 100 (Heater, Thermoswitch or PT100)
4 KLA-2S-008-12
(Pour)
2 x Detection PT 9. Clean the cooling bath surface with solvent.
5 KLA-2S-008-13
100 (Pour)
6 Test Jar KLA-1S-008-04 10. Dry the cooling bath surface with acetone.
10 Heater KLA-3S-005-03 11. Heater: Pull out the heater and replace with
new one properly.
11 Bath PT 100 KLA-3S-005-06
WARNING: Avoid air bubbles below the
Solenoid Valve
12 KLA-3S-006-01 heater. In order to protect the heater, fix two
Cooling Fluid
KLA-3-TS-Manual - 15 -
KLA-3-TS Automatic Cloud and Pour Point Analyzer
Operation and Instruction Manual
KLA-3-TS-Manual - 16 -
KLA-3-TS Automatic Cloud and Pour Point Analyzer
Operation and Instruction Manual
10. Service
Under normal operating conditions and with
routine maintenance, the KLA-3-TS Analyzer
does not require service. Any service problem
can be quickly resolved by contacting Koehler’s
technical service department either by letter,
phone, fax, or email. In order to assure the
fastest possible service, please provide us with
Figure 23: Connection Box the following information.
Model Number:
Serial Number:
Date of Shipment:
9. Maintenance and Precautions
11. Storage
Regular Maintenance
This laboratory test instrument is equipped
with electrical components. Storage facilities
Every year, check the elevation system of the
should be consistent with an indoor laboratory
analytical head.
environment. This testing equipment should not
be subjected to extremes of temperature and/or
Regularly clean the fixation system of the test
moisture. This equipment was shipped from the
jar.
factory in a corrugated cardboard container. If
long term storage is anticipated, re-packing the
Regularly clean the cooling tank. instrument in a water-resistant container is
recommended to ensure equipment safety and
Do NOT analyze other products than those longevity.
described in the method.
12. Warranty
Do NOT use solvent like acetone to clean any
element of the analyzer. We, at Koehler, would like to thank you for
your equipment purchase, which is protected by
Do NOT leave the same sample in the test jar the following warranty. If within one (1) year from
for several days. the date of receipt, but no longer than fifteen
(15) months from the date of shipment, Koehler
Avoid shock from the PT100. equipment fails to perform properly because of
defects in materials or workmanship, Koehler
Instrument Company, Inc. will repair or, at its
sole discretion, replace the equipment without
charge F.O.B. its plant, provided the equipment
has been properly installed, operated, and
maintained. Koehler Instrument Company must
be advised in writing of the malfunction and
KLA-3-TS-Manual - 17 -
KLA-3-TS Automatic Cloud and Pour Point Analyzer
Operation and Instruction Manual
authorize the return of the product to the factory. 13. Returned Goods Policy
The sole responsibility of Koehler Instrument
Company and the purchaser’s exclusive remedy To return products for credit or replacement,
for any claim arising out of the purchase of any please contact Koehler Customer Service with
product is the repair or replacement of the your purchase order number, our packing
product. In no event shall the cost of the list/invoice number, the item(s) to be returned
purchaser’s remedy exceed the purchase price, and the reason for the return. You will be issued
nor shall Koehler Instrument Company be liable a Returned Authorization (RA) number, which
for any special, indirect, incidental, must be prominently displayed on the shipping
consequential, or exemplary damages. container when you return the material to our
KOEHLER INSTRUMENT COMPANY, INC. plant. Shipping containers without an RA
DISCLAIMS ALL OTHER WARRANTIES, number prominently displayed with will be
EXPRESSED OR IMPLIED, INCLUDING ANY returned to the sender. Goods must be returned
IMPLIED WARRANTIES OF FITNESS FOR A freight prepaid. Returns will be subject to a
PARTICULAR PURPOSE. Please save the restocking charge, the application of which will
shipping carton in the event the equipment depend upon the circumstances necessitating
needs to be returned to the factory for warranty the return. Some returns cannot be authorized,
repair. If the carton is discarded, it will be the including certain products purchased from
purchaser’s responsibility to provide an outside vendors for the convenience of the
appropriate shipping carton. customer, products manufactured on special
order, products shipped from the factory past
ninety (90) days, and products which have been
used or modified in such a way that they cannot
be returned to stock for future sale.
KLA-3-TS-Manual - 18 -
KLA-3-TS Automatic Cloud and Pour Point Analyzer
Operation and Instruction Manual
Notes
KLA-3-TS-Manual - 19 -
KLA-3-TS Automatic Cloud and Pour Point Analyzer
Operation and Instruction Manual
Notes
KLA-3-TS-Manual - 20 -
KLA-3-TS Automatic Cloud and Pour Point Analyzer
Operation and Instruction Manual
Notes
KLA-3-TS-Manual - 21 -