TECHNICAL DATA :
Generator Shock Wave
Type generator Shock Wave : EMSE 140f
Description : Effective for Theraputic and minimization of side effects were the design objectives
The EMSE 140f is ideal for a spectrum of applications – from low-energy analgesic therapy to low-anesthesia
stone treatment to high-energy pseudarthrosis application
Principle : Electromagnetic
Aperture : 140 mm
Aperture Area : π (D/2)2 , With D = 140 mm
Aperature Angle : 50°
Energy Flux Density ED : 0.05 to 0.88 mJ/mm²
Effective Energy E12mm : 4.1 to 61 mJ
Focal Pressure : 8 to 51 MPa
Average Focus size (FHWM) : 6.7 x 79 mm
Penetration Depth : 150 mm
Number of Energy Levels : Max 9
Triggering (shocks/min) : 60 to 180 Manual
Patient Coupling : 5 Levels
Mobile/Fixed : Mobile
Therapy Head Positioning : Mechanical setting and locking in 36 positions and parking position
Localization Arm : Isocentric Ultrasound Arm positioning technology
: Vertical movement: 88°
: Horizontal movement: 60° / 290°
: Transversal movement: 140 mm
: Scanner rotation: > 200°
Theraphy Interface system technology : Degassed Water System with Temperature Control
Typical shock needed for theraphy : 1,500 to 3,000 shock/patient (estimation)
Outline positioning ultrasound : Yes, default using outline positioning to get clearer image and reduce shock impacted to ultrasound probe