PT
PTT
FIB
1. Pipette 50ul plasma dilution (1:10 with IBS or Owrens buffer) into cuvette
2. Prewarm plasma for 1 min.
3. Transfer cuvette to measuring position
4. Activate optic (press key “optic”)
5. Add 25ul fibrinogen reagent (the measurement starts automatically).
6. The instrument will read for a maximum of 60s.
7. The result is delayed in seconds and mg/dl.
TT
1. Pipette 50ul plasma sample into cuvette
2. Prewarm plasma for 1 min
3. Transfer cuvette to measuring position
4. Activate optic (press key “optic”)
5. Add 50ul bovine thrombin (the measurement starts automatically).
6. The instrument will read for a maximum of 300s.
7. The result is displayed in seconds and ratio.