Download as pdf or txt
Download as pdf or txt
You are on page 1of 8

Designation: G185 – 06

Standard Practice for


Evaluating and Qualifying Oil Field and Refinery Corrosion
Inhibitors Using the Rotating Cylinder Electrode1
This standard is issued under the fixed designation G185; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.

1. Scope ization Resistance Measurements


1.1 This practice covers a generally accepted procedure to G96 Guide for Online Monitoring of Corrosion in Plant
use the rotating cylinder electrode (RCE) for evaluating Equipment (Electrical and Electrochemical Methods)
corrosion inhibitors for oil field and refinery applications in G102 Practice for Calculation of Corrosion Rates and
defined flow conditions. Related Information from Electrochemical Measurements
1.2 The values stated in SI units are to be regarded as G106 Practice for Verification of Algorithm and Equipment
standard. The values given in parentheses are for information for Electrochemical Impedance Measurements
only. G111 Guide for Corrosion Tests in High Temperature or
1.3 This standard does not purport to address all of the High Pressure Environment, or Both
safety concerns, if any, associated with its use. It is the G170 Guide for Evaluating and Qualifying Oilfield and
responsibility of the user of this standard to establish appro- Refinery Corrosion Inhibitors in the Laboratory
priate safety and health practices and determine the applica- 3. Terminology
bility of regulatory limitations prior to use.
3.1 The terminology used throughout shall be in accordance
2. Referenced Documents with Terminologies G15 and D4410 and Guide G170.
2.1 ASTM Standards:2 4. Summary of Practice
D1141 Practice for the Preparation of Substitute Ocean
Water 4.1 This practice provides a method of evaluating corrosion
D4410 Terminology for Fluvial Sediment inhibitor efficiency in a RCE apparatus. The method uses a
G1 Practice for Preparing, Cleaning, and Evaluating Corro- well-defined rotating specimen set up and mass loss or elec-
sion Test Specimens trochemical measurements to determine corrosion rates in a
G3 Practice for Conventions Applicable to Electrochemical laboratory apparatus. Measurements are made at a number of
Measurements in Corrosion Testing rotating rates to evaluate the inhibitor performance under
G5 Reference Test Method for Making Potentiostatic and increasingly severe hydrodynamic conditions.
Potentiodynamic Anodic Polarization Measurements 5. Significance and Use
G15 Terminology Relating to Corrosion and Corrosion
Testing 5.1 Selection of corrosion inhibitor for oil field and refinery
G16 Guide for Applying Statistics to Analysis of Corrosion applications involves qualification of corrosion inhibitors in the
Data laboratory (see Guide G170). Field conditions should be
G31 Practice for Laboratory Immersion Corrosion Testing simulated in the laboratory in a fast and cost-effective manner
of Metals (1).3
G46 Guide for Examination and Evaluation of Pitting 5.2 Oil field corrosion inhibitors should provide protection
Corrosion over a range of flow conditions from stagnant to that found
G59 Test Method for Conducting Potentiodynamic Polar- during typical production conditions. Not all inhibitors are
equally effective over this range of conditions so that is
important for a proper evaluation of inhibitors to test the
1
inhibitors using a range of flow conditions.
This practice is under the jurisdiction of ASTM Committee G01 on Corrosion
of Metals and is the direct responsibility of Subcommittee G01.05 on Laboratory 5.3 The RCE is a compact and relatively inexpensive
Corrosion Tests. approach to obtaining varying hydrodynamic conditions in a
Current edition approved Jan. 15, 2006. Published February 2006. DOI: laboratory apparatus. It allows electrochemical methods of
10.1520/G0185-06.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
3
Standards volume information, refer to the standard’s Document Summary page on The boldface numbers in parentheses refer to the list of references at the end of
the ASTM website. this standard.

Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.

1
G185 – 06
estimating corrosion rates on the specimen and produces a should then be moved to monitor the center of the specimen,
uniform hydrodynamic state across the metal test surface. and the process repeated. Finally the micrometer should be
(2-21) moved to the bottom of the specimen and the process repeated.
5.4 In this practice, a general procedure is presented to The assembly should also be rotated at its maximum rotation
obtain reproducible results using RCE to simulate the effects of rate and the specimen wobble checked again using, for
different types of coupon materials, inhibitor concentrations, example, a laser indicator or vibration monitor.
oil, gas and brine compositions, temperature, pressure, and 6.7 Appropriate cylinder specimen (such as, carbon steel) is
flow. Oil field fluids may often contain sand. This practice does machined and snugly fitted into the PTFE or any other suitable
not cover erosive effects that occur when sand is present.
specimen holder (Fig. 2). The presence of gap between
6. Apparatus specimen and holder will create crevice corrosion as well as
6.1 Fig. 1 shows a schematic diagram of the RCE system. change the flow pattern. If necessary, apply a very small
The RCE apparatus consists of a rotating unit driven by a amount of epoxy to fit the specimen into the holder. Tightly
motor that is attached to a sample holder. A system with a range attach or screw an end-cap so that only the outer cylindrical
of rotational speeds from 100 to 10 000 rpm with an accuracy area of known length is exposed to the solution. The specimen
of 62 rpm is typical. It is essential to be able to rotate the holder is then attached to the rotating unit. Specimen, holder,
electrode at both low and high speeds and to be able to measure and end-cap should all have the same diameter.
the speed and maintain it at a constant. The accuracy of the 6.8 The rotating unit is attached into the experimental
rotation rate should be checked. At the side of the sample vessel, ensuring that there is no leakage through the rotating
holder where it is outside the cell, electrical connections to the shaft and the holder and that the rotating shaft is vertically
electrodes are made by a brush contact. It is important for the positioned. Even a very slight inclination could drastically
connection to be as noise free as possible. change the flow pattern.
6.2 The cylinder geometry is usually defined in terms of the 6.9 A versatile and convenient apparatus, consisting of a
length-to-diameter ratio. Both low and high ratios are used, kettle or flask (Fig. 1) of suitable size (usually 500 to
with ratios varying between 0.3 and 3.0. The rotating cylinder
5000 mL), inlet and outlet ports for deaeration, thermowell and
can also be used as a mass loss coupon when the mass loss is
temperature-regulating device, a heating device (mantle, hot
sufficiently large to be accurately measured using a conven-
plate, or bath), and a specimen support system, should be used.
tional balance (with accuracy of 0.1 mg).
6.3 The RCE geometry may have an inner cylinder and an The volume (of the solution) to surface area (of the specimen)
outer cylinder. The geometry is usually defined in terms of the ratio has some effect on the corrosion rate and hence inhibitor
radius of the inner cylinder and the radius of the outer cylinder. efficiencies. A larger volume/surface area (minimum 40 mL/
When the outer diameter is several times the diameter of the cm2) should be preferred.
inner electrode the hydrodynamics are essentially controlled by 6.10 In some cases a wide-mouth jar with a suitable closure
the diameter of the inner rotating cylinder (2). The outer can be used, but open-beaker tests should not be used because
cylinder may act as counter electrode. An RCE with only an of evaporation and contamination. Do not conduct the open-
inner cylinder may also be used. beaker test when H2S (hydrogen sulfide) is used. In more
6.4 A saturated calomel electrode (SCE) with a controlled complex tests, provisions might be needed for continuous flow
rate of leakage or a saturated calomel electrode utilizing a or replenishment of the corrosive liquid, while simultaneously
semipermeable membrane or porous plug tip or silver/silver maintaining a controlled atmosphere.
chloride or any other suitable electrode should be used as 6.11 For experiments above atmospheric pressure, a high-
reference electrode. The potential of the reference electrode temperature, high-pressure rotating cylinder electrode (HTH-
should be checked at periodic intervals to ensure the accuracy PRCE) system with an electrically isolated electrode system,
of the electrode. For experiments at higher-temperature, a an electrically isolated motor for rotating the electrode, and a
higher-pressure, reference electrode arrangement that can with- vessel that can withstand high pressure without leakage should
stand higher temperature and pressure should be used (22).
be used.
This may require special care.
6.5 Fig. 2 shows a typical rotating electrode unit. A rotating 6.12 A design of the vessel that can be used in elevated
shaft can be modified by drilling a hole in the shaft into which pressure conditions (23, 24) include a standard autoclave (Fig.
a polytetrafluoroethylene (PTFE) insulator is inserted. Inside 3) modified by lining on the inside with PTFE. The stirring rod
the PTFE insulator, a metal rod should be introduced (Fig. 2). can be modified by drilling a hole into that a PTFE insulator is
One end of the metal rod is threaded so that the cylindrical inserted. Inside the PTFE insulator, a metal rod is introduced.
electrode can be attached. The other end of the rod is attached Three O-rings are used to secure and to prevent leakage. One
directly to the rotating unit, through which the electrical end of the metal rod is threaded so that cylindrical (Fig. 3)
connection is made. electrode can be attached. The other end of the rod, projecting
6.6 After attaching the specimen to the shaft, the system slightly above the motor unit, is attached directly the rotating
should be checked for eccentricity and wobble. This can be unit, through which the electrical connection is made. The rod
accomplished by installing a dial micrometer so as to monitor is rotated by a motor connected to the rod using a belt. The
the location of the top of the rotating cylinder and rotating the counter and reference electrodes are inserted inside the auto-
shaft slowly through one complete turn. The micrometer clave.

2
G185 – 06

A. Reference Electrode
B. Inlet
C. Outlet
D. Luggin Capillary
E. Counter Electrode
F. Rotating Cylinder
G. Temperature Probe
H. pH Electrode
I. Rotating Cylinder Electrode or Coupon
FIG. 1 Schematic of a RCE System (18)

3
G185 – 06

A. Outside View
B. Cross-Sectional View
FIG. 2 Schematic Representation of a RCE with its Components (adapted from Ref 18)

6.13 The suggested components can be modified, simpli- 8. Test Solutions


fied, or made more sophisticated to fit the needs of a particular
8.1 All solutions (oil and aqueous) should be obtained from
investigation.
the field for which the inhibitor is being evaluated. These are
7. Materials known as live solutions. It is important that live solutions do
not already contain corrosion inhibitor. In the absence of live
7.1 Methods for preparing specimens for tests and for
solutions, synthetic solutions should be used, the composition
removing specimens after the test are described in Practice G1.
Standard laboratory glassware should be used for weighing and of which should be based on field water analysis. The compo-
measuring reagent volumes. sition of the solution should be determined and reported.
7.2 The specimen shall be made of the material (such as, Alternatively, standard brine (such as per Practice D1141)
carbon steel) for which the inhibitor is being evaluated. The should be employed. The solutions should be prepared using
specimen should have same metallographic structure as that analytical grade reagents and deionized water.
used in the service components. The specimens should be 8.2 The solutions should be deoxygenated by passing nitro-
ground to a specified surface finish (such as, 150-grit). The gen or any other inert gas for sufficient time to reduce the
grinding should produce a reproducible surface finish, with no oxygen content below 5 ppb and preferably below 1 ppb in
rust deposits, pits, or deep scratches. All sharp edges on the solution. The solution must be kept under deoxygenated
specimen should be ground. All loose dirt particles should be conditions. The oxygen concentration in solution depends on
removed. the quality of gases used to purge the solution. Any leaks
7.3 The specimens are rinsed with distilled water, degreased through vessel, tubing, and joints shall be avoided.
by immersing in acetone (or any suitable alcohol), ultrasoni- 8.3 The appropriate composition of gases is determined by
cally cleaned for 1 minute, and dried. The surface of the the composition of gases in the field for which the inhibitor is
specimens should not be touched with bare hands. The speci- evaluated. Hydrogen sulfide (H2S) and carbon dioxide (CO2)
mens are weighed to the nearest 0.1 mg (for mass loss are corrosive gases. H2S is poisonous and should not be
measurements), the dimensions are measured to the nearest 0.1 released to the atmosphere. The appropriate composition of gas
mm, and the surface area is calculated. can be obtained by mixing H2S and CO2 streams from the
7.4 Freshly prepared specimens are installed in the RCE standard laboratory gas supply. Nitrogen or other inert gases
holder. If the test is not commenced within 4 h, the prepared can be used as a diluent to obtain the required ratios of the
coupons shall be stored in a desiccator to avoid pre-rusting. corrosive gases. Alternatively, gas mixtures of the required

4
G185 – 06

1. Electrical Contact Unit


2. Techometer (Rotation Speed Display)
3. Rotation Controller
4. Electrochemical Instruments
5. Working Electrode
6. Reference Electrode
7. Water Cooler Coil
8. Gas Inlet
9. Thermocouple
10. Gas Outlet
11. Counter Electrode
12. Autoclave Body
13. Solution
14. PTFE Liner
FIG. 3 Schematic Diagram of HTHPRCE System (20,21)

compositions can be purchased from suppliers of industrial solution depend on the characteristics of the inhibitor and on
gases. The concentrations of impurities, particularly oxygen, the specified test conditions.
shall be kept as low as possible with guidelines of below 5 ppb 8.6 Oil-soluble, water-dispersible inhibitor solutions are
and preferably under 1 ppb oxygen in solution. prepared by the partition method. The required amounts of oil
8.4 The solution pH before and after testing shall be and brine are placed in the partitioning vessel (usually a
measured, recorded, and reported. The solution pH should be separation funnel). The relative volumes of oil and aqueous
monitored regularly (at least once a day) during the test. phases should reflect the ratios of water and oil in the field for
8.5 Inhibitor concentrations should be measured and re- which the inhibitor is evaluated. If samples from the field are
ported in % mass/volume or parts per million (ppm). The not available, heptane, kerosine or any suitable hydrocarbon
method of injecting the inhibitor into the test solution should may be used. The corrosion inhibitor is added to the oil phase.
reflect the actual field application. Water-soluble inhibitors The vessel is vigorously shaken for one minute to mix both
may be injected neat (as-received) into the test solution phases thoroughly, and the phases are allowed to separate.
(aqueous phase). To avoid the errors associated with handling Heating to the temperature of the field helps in the separation.
small volumes of solution, an inhibitor stock solution may be The aqueous phase is removed and used as test solution.
prepared by diluting the as-received chemical in an appropriate 8.7 Oil-soluble inhibitors (usually as batch inhibitors) are
solvent. The type of solvent and the concentration of the stock dissolved in the oil phase to form an inhibited oil-phase. The

5
G185 – 06
coupons are exposed to this solution for a certain amount of displayed by the tachometer, within 30 s. Alternatively, the
time (usually 30 min). The coupons are then removed and rotation speed can be set prior to pumping the solution into the
introduced into the experimental vessel. vessel.
9.12 The electrodes (working, counter, and reference) are
9. Experimental Procedure for Atmospheric Pressure connected to the electrochemical instruments. After the poten-
Experiments tial has reasonably stabilized (usually within 1 h), perform the
9.1 Detailed procedures to determine corrosion rates using electrochemical measurements (LPR or EIS) at regular inter-
electrochemical instruments are described in Practice G3, Test vals. Procedures to determine corrosion rate from electro-
Method G5, Test Method G59, Guide G96, Practice G102, and chemical measurements are described in Guide G96.
Practice G106, and a procedure to determine corrosion rates 9.13 After the predetermined time or after the corrosion rate
from mass loss is described in Practice G31. is stabilized, the continuous inhibitor may be injected and the
9.2 Solutions are usually prepared in a separate container corrosion rate can be monitored.
called the preparation vessel, pre-saturated with the required 9.14 For mass loss experiments, terminate the experiment
gas mixture, and preheated to the required temperature. Pre- (typically after 24 h), and determine the corrosion rate from the
treatment described in 8.5-8.7 is usually carried out in the amount of metal loss (after proper cleaning as described in
preparation vessel. Transfer solutions from the preparation Practice G1) as described in Practice G31. The samples are
vessel to the experimental vessel (described in Section 6) under examined and evaluated for pitting corrosion as in Guide G46.
positive nitrogen or other inert gas pressure to minimize air If the experiments are repeated two or more times, then
contamination during the transfer operation. calculate the average, standard deviation and coefficient of
9.3 Depending on the size of experimental vessel, heating variation using the methods provided in Guide G16.
unit (mantle, bath, or wrapper around the vessel), difference 9.15 Determine the Inhibitor efficiency at each rotation
between room and experimental temperatures, a range of speed and at each inhibitor concentration using the following
temperature may prevail within the vessel. Take care to avoid equation:
or minimize the temperature differentials. Heat the test vessels @C.R]No.inhibitor – [C.R] Inhibitor 3 100
Inhibitor Efficiency, % 5 (1)
slowly (usually at a rate of 0.1°C/s) to avoid overheating. The @C.R]No Inhibitor
exact protocol depends on the controller, the size and output of
the heater, and parameters such as vessel size, amount of where:
[C.R]No.inhibitor = the corrosion rate in absence of inhibi-
liquid, thermal conductivity of liquid, and agitation. Maintain
tor, and
the test temperature within 2°C of the specified temperature.
[C.R]inhibitor = the corrosion rate in the presence of
9.4 Insert pre-weighed coupons (pretreated as necessary, inhibitor.
such as with batch inhibitors), thermometer, and pH probes (as
appropriate). Position the liquid inlet and outlet such that none 10. Experimental Procedure for High-Temperature,
of them are protruding into the solution. High-Pressure Experiments
9.5 Initially all other ports of the experimental vessel,
except inlet and outlet ports are closed. The inlet tube should 10.1 A general procedure to carry out corrosion experiments
have a Y-joint. Attach one end to the experimental vessel. at elevated pressure and temperature is described in Guide
Attach the other two ends to the preparation vessel and to an G111. Before the experiments, check the autoclave for safety
inert gas, for example, argon or nitrogen cylinder. and integrity at a pressure that is about 1.5 times the pressure
at which the experiment is planned. For example, if the
9.6 First, pass the inert gas to expel oxygen from the
experiment is planned at 250 psi (1724 kPa), test the system at
experimental vessel.
about 400 psi (2759 kPa). This testing is required to ensure the
9.7 After 15 min, stop the gas flow, and close the passage safety of the personnel and the equipment, and also to detect
between the experimental vessel and the gas cylinder. any leak. After pressurizing the autoclave to the preset value,
9.8 Close the passage between the experimental and prepa- close the inlet and outlet valves. After 30 min, test the
ration vessels. autoclave for any leaks. If any leak is found, tighten the
9.9 Open the passage between the experimental and prepa- autoclave head screws and recheck for any further leak. If there
ration vessels, and pump the gas-saturated brine, which may or is still any pressure loss, then the system is faulty. Stop the
may not contain inhibitor prepared as per 8.6 or 8.7, into the experiment and have the autoclave inspected and repaired. If
experimental vessel. Maintain the experimental vessel with the the pressure holds constant for more than 30 min, release the
heater or the water bath at the required temperature. pressure by opening the outlet. Once the pressure gauge shows
9.10 The additional gas-inlet on top of the vessel should that the pressure has been released, close the outlet valve.
allow for keeping the gas mixture blanket on top of the 10.2 As in the atmospheric pressure experiments, follow the
solution, which is required when the experiment is planned for steps in 9.2-9.11 to charge the autoclave.
longer duration, for example, more than 24 h. Keep the gas 10.3 For high-temperature, high-pressure experiments using
flow rate to a minimum. Take care that the gas does not entrain a premixed gas composition, pressurize the autoclave using the
with the solution. specified gas composition, and depressurize to approximately
9.11 The speed controller is used to preset the rotation speed 0.2 bars above atmospheric pressure. Repeat this cycle of
and to start the motor. The rotation speed usually stabilizes, as pressurizing/depressurizing at least twice to ensure that the gas

6
G185 – 06
space has the required composition. Finally, pressurize the 11.3.7 Chemical composition or trade name of metal.
autoclave to the test pressure. 11.3.8 Form and metallurgical conditions of specimen.
10.4 For high-temperature, high-pressure experiments (25, 11.3.9 Exact size, shape, and area of specimen.
26) using individual gases, pressurize the autoclave with H2S 11.3.10 Inhibitor type and concentration.
to the required partial pressure and leave for 10 min. If there is 11.3.11 Treatment used (including e.g., batch inhibitor) to
a decrease of pressure, pressurize the autoclave again. Repeat prepare specimens.
this process until there is no further pressure loss. Then, 11.3.12 Type of corrosion measurements, such as mass loss
pressurize the autoclave with CO2 by opening the CO2 gas or electrochemical (that is, linear polarization resistance (LPR),
cylinder at a pressure equal to the CO2 and H2S partial electrochemical impedance spectroscopy (EIS)). Practice G1
pressures, and leave it for 10 min. If there is a decrease of and Practice G102 provide methods to determine corrosion
pressure, pressurize the autoclave again with CO2. Repeat this rates from mass loss and electrochemical measurements re-
process until no further pressure loss is reported. Finally, spectively.
pressurize the autoclave with the inert gas by opening the inert 11.3.13 For electrochemical measurements, size, shape, and
gas cylinder at the total gas pressure at which the experiments distance (from working electrode) of counter and reference
are to be carried out. electrodes, Tafel constants, solution resistance, number and
10.5 Carry out experiments using the same procedure as that frequency of measurements, and EIS model. Test Method G5
for atmospheric pressure experiments (see 9.12-9.14). and Practice G106 provides information for making polariza-
11. Report tion and impedance measurements.
11.3.14 Test method used to clean specimens after experi-
11.1 Record all information and data as completely as
ment and the extent of any error expected by this treatment.
possible. Practice G31 provides a checklist for reporting
11.3.15 Initial and final masses and actual mass losses for
corrosion data.
mass loss measurement
11.2 Average corrosion rates and the standard deviation of
each concentration of inhibitor at each rotation rate should be 11.3.16 Inhibitor efficiency at each concentration and at
reported. each rotation speed.
11.3 The following checklist is a recommended guide for 11.3.17 Evaluation of attack if other than general, such as
reporting important information: pit depth and distribution, standard deviation and coefficient of
11.3.1 Solution chemistry and concentration (any changes variation, crevice corrosion, and results of microscopical
during test). examination.
11.3.2 Volume of test solution.
11.3.3 Volume of the experimental vessel. 12. Keywords
11.3.4 Temperature (maximum, minimum, average). 12.1 corrosion inhibitor; electrochemical; high-pressure;
11.3.5 Pressure (maximum, minimum, average). high-temperature; laboratory evaluation; mass loss; oil-field
11.3.6 Duration of each test. inhibitors; RCE; refinery inhibitors; rotating cylinder electrode

REFERENCES

(1) Papavinasam, S., “Corrosion Inhibitors,” Uhlig’s Corrosion Hand- (9) Wranglen, G., Berendson, J., Karlberg, G., “Apparatus for Electro-
book, Revie, R. Winston, ed., 2nd ed., John Wiley & Sons, Inc., 2000, chemical Studies of Corrosion Processes in Flowing Systems,”
p. 1089, and p. 1169. Physico-Chemical Hydrodynamics, Spalding, B., ed., Adv. Publica-
(2) Gabe, D. R., “The Rotating Cylinder Electrode (Review),” Journal of tions, London, England, 1977, p. 461.
Applied Electrochemistry, Vol 4, 1974, p. 91. (10) Chilton, T. H., and Colburn, A. P., “Mass-Transfer (Absorption)
(3) NACE-5A195, “State-of-the-Art Report on Controlled-Flow Labora- Coefficients,” Ind. Eng. Chem., Vol 26, 1934, p. 1183.
tory Corrosion Test,” NACE International Publication, Item No.24187, (11) Holser, R. A., Prentice, G., Pond, Jr., R. B., and Guanti, R., “Use of
Houston, TX, December 1995. Rotating Cylinder Electrodes to Simulate Turbulent Flow Conditions
(4) Silverman, D. C., “The Rotating Cylinder Electrode for Examining in Corrosion Systems,” Corrosion, NACE, Vol 46, No. 9, 1990, p.
Velocity-Sensitive Corrosion—A Review,” Corrosion, NACE, Vol 764.
60, No. 11, 2004, p. 1003. (12) Chen, T. Y., Moccari, A. A., and Macdonald, D. D., “Development of
(5) Silverman, D. C., “Rotating Cylinder Electrode for Velocity Sensitiv- Controlled Hydrodynamic Techniques for Corrosion Testing,” Cor-
ity Testing,” Corrosion, NACE, Vol 40, 1984, p. 220. rosion, NACE, Vol 48, No. 3, 1992, p. 239.
(6) Silverman, D. C., “Rotating Cylinder Electrode - Geometry Relation- (13) Harriott, P., Hamilton, R. M., “Solid-Liquid Mass Transfer in
ships for Prediction of Velocity-Sensitive Corrosion,” Corrosion Turbulent Pipe Flow,” Chem. Eng. Sci., Vol 20, 1965, p. 107.
NACE, Vol 44, No. 1, 1988, p. 42. (14) Nesic, S., Solvi, G. T., and Skjerve, S., “Comparison of Rotating
(7) Eisenberyg, M., Tobias, C. W., and Wilke, C. R., “Ionic Mass Transfer Cylinder and Loop Methods for Testing CO2 Corrosion Inhibitors,”
and Concentration Polarization at Rotating Electrodes,” Journal of British Corrosion Journal, Vol 32, No. 4, 1997, p. 269.
Electrochemical Society, Vol 101, 1954, p. 306. (15) Berger, F. P., and Hau, K. F. F. L., “Mass Transfer in Turbulent Pipe
(8) Silverman, D. C., “On Estimating Conditions for Simulating Velocity- Flow Measured by the Electrochemical Method,” Journal of Heat
Sensitive Corrosion in the Rotating Cylinder Electrode,” Corrosion, Mass Transfer, Vol 20, 1977, p. 1185.
NACE, Vol 55, No. 12, 1999, p. 1115. (16) Denpo, K., and Ogama, H., “Fluid Flow Effects on CO2 Corrosion

7
G185 – 06
Resistance of Oil Well Materials,” Corrosion, Vol 49, No. 6, 1993, Vol 59, No. 10, 2003, p. 897.
p. 442. (22) Ives, D. J. G., Janz, G. J., “Reference Electrodes,” Academic Press,
(17) Ellison, B. T., and Schmeal, W. R., Journal of Electrochemical New York, NY, 1961.
Society, Vol 125, No. 4, 1978, p. 524. (23) Papavinasam, S., and Revie, R. W., “High-Temperature, High-
(18) Roberge, P., “Corrosion Testing Made Easy Series,” Erosion- Pressure Rotating Electrode System,” International Pipeline Confer-
Corrosion, NACE International, 2004. ence, ASME, Vol 1, No. 341, 1998.
(19) Kappesser, R., Cornet, I., and Greif, R., “Mass Transfer to a Rough (24) Papavinasam, S., and Revie, R. W., “Synergistic Effect of Pressure
Rotating Cylinder,” Journal of Electrochemical Society, Vol 118, and Flow on Corrosion Rates: Studies Using High-Temperature,
No. 12, 1971, p. 1957. High-Pressure Rotating Electrode System,” NACE Corrosion Con-
(20) Papavinasam, S., Revie, R. W., Attard, M., Demoz, A., Sun, H., ference, Paper No. 30, 1999.
Donini, J. C., and Michaelian, K. H., “Laboratory Methodologies for (25) Crolet, J. L., and Bonis, M. R., “How to Pressurize Autoclaves for
Corrosion Inhibitor Selection,” Material Performance, Vol 39, No. Corrosion Testing under Carbon Dioxide and Hydrogen Sulfide
8, 2000, p. 58. Pressure,” Corrosion, Vol 56, 2000, p. 167.
(21) Papavinasam, S., Revie, R. W., Attard, M., Demoz, A., and Michae- (26) Hausler, R. H., “Methodology for Charging Autoclaves at High
lian, K., “Comparison of Laboratory Methodologies to Evaluate Pressures and Temperatures with Acid Gases,” Corrosion, Vol 54,
Corrosion Inhibitors for Oil and Gas Pipelines,” Corrosion, NACE, 1998, p. 641.

ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org).

You might also like